Introduction of Computed Radiography in Two Mammography Services: Image Quality and Dose Analysis

Similar documents
Imaging Technique Optimization of Tungsten Anode FFDM System

Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

Quality Control for Stereotactic Breast Biopsy. Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc

Published text: Institute of Cancer Research Repository Please direct all s to:

CR Basics and FAQ. Overview. Historical Perspective

Quality Control of Full Field Digital Mammography Units

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II

Mammography: Physics of Imaging

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

While digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them.

THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY

8/2/2017. Radiologist Responsibilities. Radiologist Responsibilities. Medical Physicist Mammography Equipment Evaluation and Annual Survey

Digital Imaging Considerations Computed Radiography

Exposure System Selection

Nuclear Associates

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

Determination of the tube voltage from clinic mammographic system using two types of detectors

Optimization of Digital Mammography Resolution Using Magnification Technique in Computed Radiography 1

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

History of digital imaging

MAMMOGRAPHY - HIGH LEVEL TROUBLESHOOTING

Overview of Safety Code 35

Test Equipment for Radiology and CT Quality Control Contents

KODAK DIRECTVIEW CR Mammography Feature User s Guide

Introduction. Digital Mammography QA: Comparing the Manufacturers Recommendations. What is QC and why is it important? Review & compare QC tests

Quality control for digital mammography: Part II recommendations from the ACRIN DMIST trial

Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities

Beam-Restricting Devices

Aspire HD. Program Manual. 2nd Edition - October 2012

REQUIREMENTS FOR LICENCE HOLDERS WITH RESPECT TO QUALITY CONTROL TESTS FOR DIAGNOSTIC X-RAY IMAGING SYSTEMS

LECTURE 1 The Radiographic Image

Y11-DR Digital Radiography (DR) Image Quality

MILADY. Product Data. Page 1 of 8

7/20/2014. Outline. Outline. Disclosures. Learning Objectives. SBB: Practical Aspects of ACR Accreditation, QC and ACR On Site Surveys

A Practical Overview of the Clinical and Operational Impact of Computed Radiography(CR) Implementations. Shirley Weddle, RT(R)(M), CIIP, BBA

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance

Acquisition, Processing and Display

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah

X-RAYS - NO UNAUTHORISED ENTRY

Ansur TNT Users Manual. Plug-In

RADIOGRAPHIC EXPOSURE

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development

Patient-Assisted Compression Impact on Image Quality and Workflow

FFDM -FCRm QC Requirements- What You REALLY Need to Know

Open. the Digitized world. Fuji Computed Radiography

Learning Objectives: What s my motivation? (unknown screen actor) Workshop Overview

Comparison of computed radiography and filmõscreen combination using a contrast-detail phantom

Nuclear Associates

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Digital radiography (DR) post processing techniques for pediatric radiology

Teaching Digital Radiography and Fluoroscopic Radiation Protection

Outline. Digital Radiography. Understanding Digital Modalities: Image Quality and Dose. Image Quality. Dose Control

4/19/2016. Quality Control Activities for the RadiologicTechnologist. Objectives. 3D Tomosynthesis QC differences

Evaluation of a quality control phantom for digital chest radiography

Comparison between film-screen and computed radiography systems in Brazilian mammography

DIGITAL RADIOGRAPHY ARTIFACTS

Breast Imaging Basics: Module 10 Digital Mammography

Overview. Professor Roentgen was a Physicist!!! The Physics of Radiation Oncology X-ray Imaging

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage

Practical Medical Physics Session: TG-151 Dose Monitoring. August 5, 2013 Katie Hulme, M.S.

New spectral benefi ts, proven low dose

Assessment of Beam Alignment, Collimation and Half Value Layer of Some Selected X-Ray Machines in Plateau State, Nigeria

Features and Weaknesses of Phantoms for CR/DR System Testing

Half value layer and AEC receptor dose compliance survey in Estonia

Image Quality. HTC Grid High Transmission Cellular Grid provides higher contrast images

Calibration of KAP meters

ACPSEM Position Paper RECOMMENDATIONS FOR A DIGITAL MAMMOGRAPHY QUALITY ASSURANCE PROGRAM V4.0

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing

NJDEP Medical Physicist s Radiographic QC Survey Registration Number:

The importance of radiation quality for optimisation in radiology

China Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW, 150kV Radiography System - HF50-R

Quality assurance: a comparison study of radiographic exposure for neonatal chest radiographs at 4 academic hospitals

Ask EuroSafe Imaging Tips & Tricks. Paediatric Imaging Working Group. Dose Management in Digital Radiography

Radiographic Techniques, Contrast, and Noise in X-Ray Imaging

Image Quality. HTC Grid High Transmission Cellular Grid provides higher contrast images

Collimation Assessment Using GAFCHROMIC XR-M2

Estimation of signal transfer property for wireless digital detector in different measurement schemes

Ludlum Medical Physics

GE Healthcare. Senographe 2000D Full-field digital mammography system

X-ray Tube and Generator Basic principles and construction

Performance and care. all in one

Digital Breast Tomosynthesis

Digital Quality Control Phantom for X-ray Mammography Image

Disclosures. Outline 7/31/2017. Current Implementation Status of IEC Standard : Exposure Index (EI) for Digital Radiography

PEDIATRIC AND STAFF DOSE EVALUATION IN FLUOROSCOPY UPPER GASTROINTESTINAL SERIES

Introduction. Sam R. Kottamasu Lawrence R. Kuhns

2012 :15th SESSION of ESMP

SYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology

Effect of Backscattered Radiation on X-Ray Image Contrast

diagnostic examination

Visibility of Detail

Transcription:

Introduction of Computed Radiography in Two Mammography Services: Image Quality and Dose Analysis Rosangela Requi Jakubiak* a, Humberto Remigio Gamba a, Maria Manuela Ramos a, Gislene Gabrielle Faversani a e João Emílio Peixoto b a Universidade Tecnológica Federal do Paraná UTFPR, Av. Sete de Setembro 3165 CEP 80230-901, Curitiba, Brazil b Instituto Nacional do Câncer- INCA, Praça Cruz Vermelha, 23 CEP 20230-130, Rio de Janeiro, Brazil. Abstract. This study has evaluated the impact of the introduction of CR technology in the routine of two mammography services operating with Lorad Affinity x-ray units and CR Fuji Profect One system. Mean glandular dose was determined to set Automatic Exposure Control parameters according to optimal image quality. Reject analysis of patient images was made to establish the overall impact in routine work. As regards to dose evaluation, the results show that in 44% of mammography exams the MGD were above action levels. It is due to inadequate AEC adjustment and insufficient training of the staff in operating CR systems. In addition, common errors with screen/film systems like wrong breast positing and selection of exposure parameters, as well as x-ray unit failures like insufficient anti scatter grid movement still occurs. This is an indication that ongoing efforts should be concentrated in the constancy of AEC adjustment for CR image plates and staff training. Until now, brazilian health authorities did not implement any program to evaluate digital systems performance, but efforts are being done in order to provide guidelines to breast screening aiming the control of breast dose compatible with optimal image quality. KEYWORDS: Digital Mammography; Quality Control; Image Quality 1. Introdution Mammography is considered the main method for detection of breast cancer in early stages and due to its importance for the health system a quality assurance program in every mammography ervice is needed [1,2,3]. However, the mammography effectiveness depends on the ability of the radiologist to detect malignity signals [3]. In this sense, digital mammography has numerous potential advantages hen compared with screen/film (S/F). It has a large dynamic range (linear response in the range of 1000:1 compared with 40:1 in screen/film mammography) [4]. Also, image processing of digital data allows the manipulation of contrast in such a way that higher contrast image resolution can be obtained, especially for denser breasts images [4-5-6]. In digital mammography the image acquisition process, displaying and storage follow a different line of that of conventional mammography. Although this dissociation made it possible to optimize each component individually, it also introduced new challenges [7]. In this work it was used the technology of Computed Radiology (CR) where the imaging plates (IP), available in standard mammographic cassette sizes, are located in cassettes for exposure in standard screen-film Bucky trays. In response to the absorption of X rays, electronic charges are stored proportionally in traps in the material of the phosphor, where they remain stable for some time [8]. After irradiation, the screen is stimulated by a scanning laser beam, to release the deposited energy in the form of visible light. The released photostimulated light is captured by a light detector, converted to digital signals, and registered with the location on the screen from which it has been released. The digital data are then postprocessed for appropriate presentation, and are sent to a hard-copy printer or a soft-copy display monitor for medical evaluation [9-10-12]. The automatic exposition control (AEC) system ensure the optimal exposure of the image receptor, but The response of computed radiography (CR) imaging plates (IPs) to X-ray beams of different energy is substantially different from that of conventional film screen combinations. Automatic exposure control (AEC) systems programmed for film maintain a constant optical density (OD) regardless of the beam quality incident upon the cassette [11]. requi@utfpr.edu.br 1

Mean Glandular Dose (MGD) calculations normally are executed according to the method of Dance, that, in 1990, based in Mount Carlo calculation, with factors that allow the conversion of kerma of incident air in the breast surface in absorbed dose [13]. Analysis of rejected films, which consists in counting of the rejected films in accordance with of the cause of rejection of each one, yields information about the efficacy of a department and is the basis for quality control and education of technologists [13]. Because digital technology still being at its early development stages, setting a credible rejection analysis program became too a task of the study [14]. The rejection causes of one image in system of computed radiography observed were related, especially: To the patient, due to not following orientations before and during the exam, such as the use of products that contain metallic substances, i.e. deodorant, ointment and powders; to the technologist, for the lack of technical knowledge, inexperience and lack of attention; to responsible person for the images manipulation and impression at console, which presents the deficiency same of the technologist. The objective of this work is to evaluate the impact of introduction the CR technology in routine and patient dose. 2. Materials and methods 2.1 Mammography Equipment Tests Two mammographic equipments from Lorad Affinity were tested just after acceptance. In each equipment we used the system S/F to assess the: 1) collimation, accuracy and reproducibility of tube tension; 2) reproducibility and linearity of kerma rate in the air; 3) accuracy and reproducibility of the exposition time; 4) reproducibility of the AEC; 5) control of density performance; 6) half-value layer (HVL); 7) focal spot, 8) compression system, 9) image quality, 10) entrance surface air kerma and 11) MGD. The performance of the AEC system was assessed in terms of reproducibility and accuracy under different object thickness and beam quality. According to the literature the AEC system should adjust target (tube anode, i.e., molybdenum or rhodium), filter (molybdenum or rhodium) and tube voltage such that image quality is sufficient and dose is within an acceptable range [8-11]. However, it is important to mention that the Lorad Affinity equipments does not provide rodhium target. The evaluation of AEC and image quality were done with the acrylic plates (poly(methylmethacrylate) - PMMA) and the breast simulator object (Mammographic Accreditation Phantom - ACR), respectively. The latter contains insertions that simulates the fibers, specks and tumor masses in the breast s tissue. The exposition was always made in the automatic mode. The AEC and image quality tests were performed with the equipments calibrated form the S/F and CR in years 2007 and 2008, respectively. Each CR maker has its own exposure coefficient value, which is a computer-generated parameter relating to the response of the IP phosphor [11], and so the maintenance team must come to re-adjust the AEC and check the image quality and dose. The parameter employed by Fuji is the sensitivity number (S) given by equation 1: S (4 Sk ) = 4x10 (1) 2 requi@utfpr.edu.br

where Sk is the central value of logarithmic expression of the X ray dose range reads as digital image data [12]. In general the CR maker specifies for S a range of values. Therefore, the mammography equipment accesses the look-up-table (LUT), into the X-ray unit, and reads the values of target, filter, tube voltage and dose, according to the compressed breast thickness (we selected thickness varying between 20 e 70 mm), to determine an exposure according to the S Fuji values (20-80). During the tests doctors, maintenance team from Lorad and the physicist of the clinical service worked together to establish the standard exposition for the dose and image quality, approximately according to the Fuji S values specified. The definition of the standard exposition for the dose and image quality was only possible because we had in the service a great amount of images of the phantom, which were acquired during the period of transition between the systems. Unfortunately, an ionizing chamber was not available and we did not registered the dose values for these images, only the mas exposure was registered. The value of dose was measured, with an ionizing chamber, annually or when a significant alteration of the equipment was carried through. 2.2 Patient Dose The first registers of dose of entrance surface air kerma and MGD they had been carried through with mammography LORAD Affinity 4 months after its installation, with Ionizing Chamber 6M n/s 8470 Radcal Corporation) using S/F system. The adopted methodology and limits for of entrance surface air Kerma it s 10 m Gy and for MGD: acceptable < 2,5 mgy, achievable < 2 mgy, [8] The pattern value of MGD was calculated by applying the Dance method, described in equation 2. MGD=K.g.c.s (2) Where: K is the entrance surface air kerma (without backscatter) calculated at the upper surface of the PMMA. The factor g, corresponds to a glandularity of 50%, and is derived from the values calculated by Dance et al 2000 [8]. After 5 months of the complete transition to CR technology was made and the AEC was recalibrated for CR Fuji Profect One and IP s use. Then, it was done a new test to evaluate the kerma values of skin penetration (utilizing Ionizing Chamber CI-Keithley 96035 B) and MGD. The values of dose were high and therefore a new AEC calibration of the two rooms it was requested the dose evaluations had been repeated using RAD-CHECK MAMMO 06-526 and with Fluke Ionizing Chamber 600 529 model. 2.3 Rejection Analisys In order to optimize the service after the implementation of system CR and to identify problems that had been not yet solved after the period of adaptation to the new technology, it was carried through a comparative study of two analyses of rejection, the one soon after installation of the system (2007) and another pasts 7 months (2008). In these four months of collection of data, each technician was responsible for writing down the film rejection cause, in the case of printed images, or in tables, in the case of rejected examinations in the soft copy system. The rejection causes were classified in accord to the following error sources: Patient: Motion artifact, Deodorant, ointment and powders artifact. Technologist: Positioning errors, Fold artifact, and Failure to use of grid. requi@utfpr.edu.br 3

Responsible for the manipulation and impression of the images at console: artifact in IP, window (dark), window (bright), duplicate film, no centralization breast at film, absence of cut around of the breast, extraction of region of interest, during processing, reversed breast in film, improper size of images, improper size of film, incoherent assembly with the protocol defined, identification in region of interest, reverse identification, absence of identification, patient wrong name. Equipment: Problems in the mammography units. Radiologist: Exam out of the quality pattern. Some rejection causes are related to problems at mammography unit and to less quality pattern than the established by radiologist, however those present a smaller rejection index. 3.Results 3.1 Mammography Equipments Tests It s recommended that, before the system is put into clinical use, it must undergo acceptance testing to ensure that the performance meets these standards. But the prices of the quality control equipment in Brazil are so expensive, and the acceptance routine normally is not realized and the tests, never are realized in al most of clinics. But for this work, the tests were accomplished four months after the installation. Table 01: Resume of results obtained in required tests on mammography equipment, room 1 and 2, 4 moths after install. According with MS- Required Tests 453/98 Room 1 Room 2 Tube Voltage- Accuracy- Large Focal Spot Accord Accord Tube Voltage- Accuracy- Small Focal Spot Accord Accord Tube Voltage- Reproducibility- Large Focal Spot Accord Accord Tube Voltage- Reproducibility- Small Focal Spot Accord Accord Reproducibility of Exposure Time- Small Focal Spot Accord Accord Reproducibility of Exposure Time- Large Focal Spot Accord Accord Half Value Layer Determination Accord Accord Reproducibility of Air Kerma Rate Accord Accord Tube Output - Large Focal Spot Accord Accord Linearity of air Kerma rate Accord Accord Compression System Accord Accord Collimation System Evaluation- Image Field Congruency Accord Accord Collimation System Evaluation - X Ray Field Extension Accord Accord Collimation System Evaluation -Compression Plate Alignment Accord Accord Automatic Exposure Control System Performance Accord Accord Focal Spot Size Large Focal Spot 0,52-0,37 0,50-0,32 Focal Spot Size Small Focal Spot 0,15-0,16 0,14-0,14 The work of the maintenance team it was concentrated in establishing the values of product current versus time (mas) that the values of dose would be adequate S and for the quality of the image, showed in the table 2, in accordance with the previous registers, together with the doctors, and the physicist of the clinical service. The results had shown that it is possible to make de AEC calibration in both equipments, leaving a minimum difference between the two rooms. This results are very relevant, because the examinations patterns are maintained and the technicians can apply the same standards of assessment in both the rooms Table 02: Comparative results of AEC calibration of Room 1 and Room 2 4 requi@utfpr.edu.br

Mo Room 1 Room 2 Rh Room 1 Room 2 mas mas % var mas mas %var. kvp kvp 24 184 182 1,09 28 62,7 60,5 3,51 25 138 136 1,45 29 51,5 51,2 0,58 26 106 109-2,83 30 85,2 84,8 0,47 27 80,1 81,4-1,62 31 74,1 73,8 0,40 28 62,9 63,6-1,11 32 61,7 60,6 1,78 29 51,7 52,4-1,35 33 54 53,9 0,19 30 45,5 46,2-1,54 34 47,7 45,2 5,24 Adjustments had been carried through large and small spot, always searching to keep answers of AEC that are similar for the selected thicknesses, shown in the tables 3 and 4. Table 03 Results to large spot calibration Large Spot - Auto kv Room 1 Room 2 Tracking kvp mas Tracking kvp mas % Var 2 cm 25 28,7 2 cm 25 29,2 2 cm -1,74 3 cm 28 35,7 3 cm 28 36,4 3 cm -1,96 4 cm 25 135 4 cm 25 132 4 cm 2,22 5 cm 29 113 5 cm 29 110 5 cm 2,65 6 cm 30 195 6 cm 30 192 6 cm 1,54 7 cm 30 289 7 cm 30 284 7 cm 1,73 Table 04 - Results to small spot calibration Small Spot - Auto kv Room 1 Room 2 Tracking kvp mas Tracking kvp mas % Var 2 cm 25 30,5 2 cm 25 30 2 cm 1,64 3 cm 28 43,2 3 cm 28 39,5 3 cm 8,56 4 cm 30 54,8 4 cm 30 54,9 4 cm -0,18 5 cm 30 102 5 cm 30 98 5 cm 3,92 6 cm 30 130 6 cm 30 132 6 cm -1,54 7 cm 30-7 cm 30-7 cm N/A The Table 4 shows the results of image quality evaluation. Table 04: Image Quality of Phantom ACR. Visible Structures Room 1 a Room 2 a Room 1 b Room 2 b Room 1 c Room 2 c Fibers 5 4 5 5 5 5 Masses 4 4 5 4 4,5 5 Speck groups 4 4 3 3 3,5 3,5 Total 13 12 13 12 13 13,5 (a) Results of SF, (b) Results of IP 2007, (c) Results of IP 2008 after last calibration requi@utfpr.edu.br 5

The radiologist is provide with images that have the best possible diagnostic information and the images should at least contain the defined acceptable level of information, necessary to detect the smaller lesions. The minimum score of four fibers, three speck groups, and three masses are defined by the scoring system of the ACR Phantom. 3.2 Patient Dose In the three stages of system transition, the dose values of entrance surface air kerma and MGD were measured. However, in Brazil, the availability of equipment for usual dose measured is not frequent. The monitoring of the mas can be a dose indicator, since that initial conditions of the system are know. The table 5 shows the values of the mas which were obtained in the same conditions described in table 2. Based on the mas changes due the S/F transition which called the attention since the dose values were too higher than before the transition. The difference of exposition between Room 1 and Room 2 would be also evaluated, because it implies too different image quality patterns between the two rooms. High dose give good image quality. However, they are harmful to the patient. Lower doses can lower image quality, harming the diagnoses. Neither situation is adequate, and both need corrections. Table 05: Indicadores de dose nas Room 1 and Room 2 Indicators Room 1 a Room 2 a Room 1 b Room 2 b Room 1 c Room 2 c kvp 26 26 26 25 25 25 mas 86,0 90,0 120, 0 169, 0 92,0 91,0 Entrance surface 12,68 9,39 10,03 11,02 13,83 17,38 air Kerma (mgy) MGD(mGy) 2,25 1,71 2,70 3,61 2,48 2,81 3.3 Rejection analysis The main causes of rejection had been. These causes are presented in figure 1, to following Wrong positioning of patient and movement of the patient, that they are inherent factors in such a way to the analogical system as to the digital one Film duplicate, that it is an error of the operator and not of the printer; mage without adequate clipping of the anatomical structures, harming the aesthetic one of the images; Figure 01: Comparison of major causes of rejection between 2007 and 2008. Comparison of Rejection 2007/2008 Quantity of rejeted films 50 45 40 35 30 25 20 15 10 5 0 12 44 7 0 15 36 14 12 13 6 2 0 2 0 2007 2008 Positioning errors Fold artifact Duplicate film Extraction of region of interest, Incoherent assembly with the Reverse identification Patient wrong name Major Causes of Rejections 6 requi@utfpr.edu.br

The errors due to the responsible for the manipulation and impression of the images at console, showed in the figure 2 and in the table 6, they show that this is the critical phase of the process. In case that the images are printed with errors, the cost of the examination will be bigger for the clinic, therefore films of printers Dry View are more expensive, on average, the triple of the film for system S/F. And the same manner, if the examinations are not printed matters and the patient does not need to repeat the examination, the lost time in the manipulation of the images, it means loss of time for the attendance of new patients. Figure 02: Rejections due errors in manipulation and impression at console. Rejection related at responsible for the manipulation and impression of images at console Quantity of rejected films 42 35 28 21 14 7 0 8 1 1 41 36 15 14 12 13 5 4 9 6 2 4 2 3 1 2 2 2 3 1 6 2007 2008 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 Cause of rejection Table 06: Quantity of reject films of according of source of errors: Origin of Errors 2007 (%) 2008(%) Patient 8,7 6,8 Technologist 20,3 32,9 Responsible for the images at console 71,0 54,0 Equipment 0 0,6 Radiologist 0 5,7 Total(%) 100 100 4. Discussion and Conclusion The initial tests of conformity of the two mammography units had revealed adequate. However, with the modality transition S/F for CR, the AEC response to the calibration one revealed inadequate, also with values of dose 44% above of the recommended one. Then the problem increased due to the difficulties of the maintenance staff to define the LUT s which were in accordance with the CR fabricant recommendations. For an adequate calibration of the AEC, the maintenance staff needs to know the equipment s software very well. Data from the doses and image quality are important sources of knowledge which corroborate with the maintenance staff. The data show that it is possible to establish an answer pattern of the AEC, compatible with dose, image quality and adequate S value. Calibration difficulties of the AEC also can collaborate with the exam rejection. However, the results show that, in spite of the positioning criteria holds on being the same, positioning mistakes, requi@utfpr.edu.br 7

related to the technologist, are the second main cause of exam rejection. It happens even if the personnel pass through continued education programs. Fails related to the technologist contribute to the increase in dose in the patient, for the increase in costs due to the reduction of available time to help other patients and even due to the fact that Fuji IP has guaranteed development until an average number of 40,000 expositions. The exams repetition reduces the useful life time of the IP, resulting also in the increase of the costs for the clinic. Since that, unfortunately, it is still necessary to elaborate new continuous education programs and the theme must be treated again. Taking in consideration those data, it is fundamental which the member states shall ensure that all staff involved in a radiological procedure has adequate theoretical and practical training for the purpose of radiological practices, as well as relevant competence in radiation protection. It also states that continuing education and training after qualification is provided, with a provision of appropriate training for practitioners conducting special practices. Unfortunately, the equipment for the conformity tests developing are still very expensive and are not always available to the tests realization right after the maintenance is done. This allows that in many times the patient is exposed to higher radiation doses that it is recommended. The technology in digital mammography is still very recent and the maintenance staffs are also having difficulties to do the AEC s calibration correctly. Multidisciplinary personnel with maintenance technicians, doctors and physicists must work together to establish, inside the clinic, the pattern of image quality which defines an adequate dose pattern to the patient without harming the diagnostic. Still it will be necessary to carry through new adjustments in AEC, however, this first experience samples that it is necessary to work together with the maintenance team, in order to establish the parameters of quality and dose of the health service. Acknowledgements We are thankful to the DAPI for being in the vanguard of the radioprotection in Brazil. REFERENCES [1] TARDIVON, A. A., et al., Imaging and management of no palpable lesions of the breast. European Journal of Radiology 42 (2002), 2-9. [2] POULOS, A., MCLEAN, D, The application of breast compression in mammography: a new perspective, Radiography 10 (2004), 131-137. [3] TAKEO, H. SHIMURA, K. IMAMURAK. T. Detection System of Clustered Micro calcifications on CR Mammogram. Ieice Trans. Inf. & Syst., E88 D, No.11 (2005) [4] FEIG, S. A., YAFFE, M.J. Digital Mammography. RadioGraphics, 18, (1998), 893-901. [5] YAFFE, M.J.BLOOMQUIST, A. K., Quality control for digital mammography: Part II recommendations from the ACRIN DMIST trial. Med. Phys. 33, (2006),737-752. [6] PISANO, E, YAFFE, M.J. Digital mammography, Radiology (2005), 353-361. [7] SIEGEL, E., et al, Digital Mammography Image Quality: Image Display, J Am Coll Radiol;3, (2006),615-627. [8] THE EUROPEAN PROTOCOL FOR THE QUALITY CONTROL OF THE PHYSICAL AND TECHNICAL ASPECTS OF MAMMOGRAPHY SCREENING, ADDENDUM ON DIGITAL MAMMOGRAPHY - to chapter 3 of the: European Guidelines for Quality Assurance in Mammography Screening, EUREF (2003). [9] MAHESH,M. Digital Mammography: An Overview. AAPM/RSNA Physics Tutorial. RadioGraphics 24, (2004), 1747-1760. [10] SAMEI, E.,et al. Performance evaluation of computed radiography systems. Med. Phys. 28 (3), ( 2001),361-371. [11] DOYLE, P., et al. Optimizing automatic exposure control in computed radiography and the impact in patient dose. Radiation Protection Dosimetry, 2005, Vol. 114, Nos 1-3, pp. 236 239 [12] FCR (FUJI COMPUTED RADIOGRAPHY) General Description of Image Processing. Fuji Photo Film Co. Ltd 8 requi@utfpr.edu.br

[13] DANCE, D. R. Monte Carlo calculation of conversion factors for the Estimation of mean glandular breast dose. Phys. Med. Biol., 1990, Vol. 35, No 9, 1211-1219. [14] PEER, S., et al, Comparative reject analysis in conventional film-screen and digital storage phosphor radiography Eur. Radiol. 9, (1999 ),1693±1696. [15] HONEA, R.,BLADO, M. E., MA, T. Is reject analysis necessary after converting to Computed Radiography?, Journal of Digital Imaging, 15, Suppl 1, (2002), 41-52. requi@utfpr.edu.br 9