SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only

Similar documents
OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE. by Doug Bray Department of Biological Sciences University of Lethbridge

Scanning Electron Microscope FEI INSPECT F50. Step by step operation manual

FE-SEM SU-8020 Operating manual (Preliminary version)

Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014

Using the Hitachi 3400-N VP-SEM

SEM Training Notebook

Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400.

SEM Training Notebook

Title: Amray 1830 SEM#2 Semiconductor & Microsystems Fabrication Laboratory Revision: D Rev Date: 03/18/2016

Operating Checklist for using the Scanning Electron. Microscope, JEOL JSM 6400.

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah

Model SU3500 Scanning Electron Microscope

JEOL 6500 User Manual

Check that the pneumatic hose is disconnected!!!! (unless your using the BSE detector, of course)

1. Preliminary sample preparation

Dickinson College Department of Geology

JEOL 6700 User Manual 05/18/2009

SCIENTIFIC INSTRUMENT NEWS. Introduction. Design of the FlexSEM 1000

Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below.

Secondary Electron Detector

Introduction to Scanning Electron Microscopy

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM

RAITH e-line OPERATING INSTRUCTIONS

Instructions for Tecnai a brief start up manual

HEX02 EMBOSSING SYSTEM

Operation Guide. Hitachi S-3400N. Variable Pressure Scanning Electron Microscope. with. Deben Peltier Coolstage

SEM OPERATION IN LOW VACUUM MODE

2. Raise HT to 200kVby following the procedure explained in 1.6.

Figure 1 The Raith 150 TWO

ZEISS EVO SOP. May 2017 ELECTRON OPTICS

JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES

LEO 912 TEM Short Manual. Prepared/copyrighted by RH Berg Danforth Plant Science Center

The light microscope

1.2. Make sure the viewing screen is covered (exposure to liquid N 2 may cause it to crack).

Basic Users Manual for Tecnai-F20 TEM

Cressington 108 Auto/SE Sputter Coater Standard Operating Procedures (S.O.P)

Schottky Emission VP FE-SEM

1.1. Log on to the TUMI system (you cannot proceed further until this is done).

Scanning Electron Microscope in Our Facility

Strata DB235 FESEM FIB

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope

OMM300. Inverted Metallurgical Microscope

Scanning Electron Microscopy SEM. Warren Straszheim, PhD MARL, 23 Town Engineering

MSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope

LASER ENHANCED REVOLVER GRIP OWNER S MANUAL RED LASER GREEN LASER

General information. If you see the instrument turned off, notify MIC personnel. MIC personnel will help you insert your samples into the instrument.

Standard Operating Procedure for Cutting Stainless Steel on the Rofin Laser Machining Station

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right

05/20/14 1. Philips CM200T. Standby Condition

(Refer Slide Time: 00:10)

1.3. Before loading the holder into the TEM, make sure the X tilt is set to zero and the goniometer locked in place (this will make loading easier).

TABLE OF CONTENTS. Safety notes i. Care and Maintenance. ii. 1. Components Illustration Installation of Components.. 4

MSE 595T Transmission Electron Microscopy. Laboratory III TEM Imaging - I

Transmission Electron Microscopy 9. The Instrument. Outline

LSM 800 Confocal Microscope Standard Operation Protocol

Scanning Electron Microscope. Instructions for Use

RENISHAW INVIA RAMAN SPECTROMETER

ML7520 ML7530 DIOPTER ADJUSTMENT RING BINOCULAR BODY, INCLINED 30. (a) Field Iris Control Lever. (c) Filter Slots EYEPIECES, KHW10X

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com

Standard Operating Manual

FEI Tecnai G 2 F20 Operating Procedures

1. Specimen Holder Removal, Loading, and Insertion

Marine Invertebrate Zoology Microscope Introduction

The Prior School Microscope Graham Matthews

Scanning electron microscope

Scanning Electron Microscopy. EMSE-515 F. Ernst

Leica SP8 TCS Users Manual

SWIFT SERIES M2252DGL MICROSCOPE

SECONDARY ELECTRON DETECTION

SCANNING ELECTRON MICROSCOPY AND X-RAY MICROANALYSIS

JEOL JEM 2010 TRAINING TRANSMISSION ELECTRON MICROSCOPE USER MANUAL

Quick and simple installation and no maintenance needed. 3 Times More affordable Than a normal SEM. Obtaining results in less than 4 minutes

Brightfield Microscopy and Image Acquisition on Spotcam1. by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer

INTRODUCTION We believe that every laboratory working in the field of nanotechnology needs an SEM, therefore we would like to introduce to you our IEM

Dremel 3D Printer Digital Fabrication Lab College of Design, Iowa State University

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Introduction: Why electrons?

S200 Course LECTURE 1 TEM

FEI Quanta 200 ESEM Basic instructions

Zeiss Axiovert 135 Fluorescence Microscope Quick Guide / Operations Manual (v. 1.0 February 09)

Electron Microscopy Sciences

Scanning Electron Microscopy Laboratory Portfolio

OM FL400. Reflected Light Fluorescence Microscope. Instruction Manual. Please read instructions carefully before using microscope.

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the Microscope Control interface.

ILFORD SPORTSVIEW PROJECTOR INSTRUCTION BOOK

Swift M2252DGL Series Microscope Use and Care Manual

Protective Equipment Nitrile gloves for handling sample holder and safety glasses for filling liquid nitrogen dewar.

STRUCTURE OF THE MICROSCOPE

Amray 3600 FESEM. Standard Operating Procedure. v2.2 modified by Bryan Cord. General Notes...3. Sample Loading...5. System Loading...

Olympus Digital Microscope Camera (DP70) checklist

Introduction of New Products

Nikon Ti-E Microscope Manual. Rightmire Hall Ohio State University. Director: Tony Brown Rightmire

Scanning electron microscope

MSE 460 TEM Lab 4: Bright/Dark Field Imaging Operation

Leica DMi8A Quick Guide

User instructions Compound laboratory microscope

Standard Operating Procedure Hitachi UHR CFE SU8230 SEM

Components of the Microscope

JEOL 5DII. Operation introduction. By Serge Charlebois

Electron Microscopy Sciences

Transcription:

SOP for Hitachi S-2150 Scanning Electron Microscope For review purposes only Version 1.0 Prepared by D. Turnbull February 21, 2007. Please submit any omissions to the Author Note: This SEM is a recent donation to Niagara College from Luxcell. It is still in need of some TLC and the state of the electrical wiring is not fully known. Extra caution should be taken at rear of the instrument for electrical hazards until instrument is fully commissioned. This SOP is intended as a starting point for qualified individuals who are familiar with this type of equipment and the associated hazards. This SOP has been prepared primarily for faculty who have been trained and are familiar with the instrument. The operator is referred to the, Model S-2150 Scanning Electron Microscope Manual for more complete instructions.

Startup / Shutdown / Power failure Report any issues immediately to one of the department Technologists. See the attached document from Hitachi. Note: as set-up in V13A we do not have a main switch the isolation transformer is unplugged at the wall. Note: Please review and be familiar with the procedure, measure against power failure so that you can quickly perform this in the event of a power failure.

Start-up notes: 1. Open water valve ~ ¼ open. This will provide adequate cooling water (1-1.5 L/min). This is interlocked and an alarm will sound if flow is too low. 2. It will take about 20 minutes for the high vacuum light to come on. At this point, the ready lamp will start flickering do not press ACC voltage on until this lamp stays on steady. Failing to do so will greatly shorten the filament lifetime. Shut down notes: Turn the accelerating voltage and magnification down to the lowest setting, turn off the ACC voltage, complete the log book, and turn off as described above. Note: there is a twenty-minute cool down time before the water is turned off. Note: there is no main switch at this step unplug isolation transformer from the wall. Sample installation: Note: On the front of the sample compartment there are default settings for the x, y, z and tilt setting. These must be set before opening the compartment. Failing to do so could cause serious damage to the instrument, particularly the tilt setting. The Z- axis default is 15 mm. If this is set greater than 15 mm, i.e. 20mm, it is because a thicker sample is currently installed. It is suggest the sample compartment be opened without resetting to 15mm when a greater than 15 mm Z setting is set).

i.e. in the above table, cannot rotate past 40 tilt if Z=7.5mm or less (note: care should be taken as we do not have the height adjust tool and do not absolutely know the correct height).

In images below is the specimen stage showing the X, Y, Z, tilt, and rotation (located in center of X control) stage controls and shows the stage opened. Note that specimen stage should slide easily along the rails to open. It should not be forced open (i.e. opened before it reaches atmospheric pressure). Sample installation: 1. Make sure ACC voltage is off (turn down and then off) 2. Press air switch wait for chamber to reach atmospheric pressure. Then slide open. 3. Loosen Allen screw and install specimen stub with sample mounted or mount onto specimen stub that is present. A good electrical connection is important. This is still currently under development. However, for microchips (even on a ceramic base) and metal coated glass slides, aluminum tape to mount the sample and as a grounding strap, attaching the sample to the specimen stub seems to work. Bonding with super glue then a bonding strap also works. Metallic coatings and conductive pastes and epoxies are being investigated at the time of writing. Sample should fit onto the specimen stub. Sample height is important to set so that damage to the objective lens does not occur. Unfortunately, the SEM was received without the height gauge. Typically a flat sample should not be of any concern. Use 15 mm Z setting. If sample is thicker the Z setting should be increased approximately the sample diameter (i.e. Higher numbers 15 to 20 moves the specimen away from the objective lens). Refer to the manual for more information on sample installation and movable range and tilt limitations based on the sample size (copy included above).

Image Observation: The following is just an overview with some suggestions for a trained user. The user is referred to the manual for more details. Appended to this SOP are manual pages 2-25, 2-26,2-27 that are a set of tables with some guidelines for obtaining clear image quality. 1. High Vacuum light is on and ACC voltage ready lamp is on and not flickering, press ACC voltage ON/OFF switch. Press the up arrow to increase. a. For microchip sample tested an image can be seen at as low as 1.5 kv. However, 5-10 kv seem to be a good starting point. Emission current should be seen on the readout. 2. CRT screen will start in a raster mode. Press to switch to TV mode. 3. If instrument was left in working order. Pressing the auto then ABC buttons a few times should bring the image into view. Switching to manual and adjusting the brightness and contrast seems to work better than auto. 4. As above, auto focus coarse and fine and auto Stigmator followed by fine manual adjustment should bring a fairly clear image into view. 5. At this point one can increase the magnification adjusting the above at each step. 6. At some point the accelerating voltage may have to be increased, the aperture adjusted, and C. Lens adjusted to obtain a clear image that does not wander on the CRT too much. a. Note: The operator should be trained in these adjustments and refer to the manual for further details. Reducing the condenser lens current excessively with the post HV = on (as in normal operation) can cause too much secondary electron emission which may damage the scintillator. (p.2-27) 7. Tables 2-2 to 2-5 are appended to this document and provide some guidelines for settings.

Example #1 effect of working distance on image AMD chip ACC=10kV Aperture = 3 Mag = 1000X C lens = 9:30 Tilt = 45 WD = 25 mm, 17 mm in pictures 1 and 2 Picture 1 Picture 2

Example #2 effect of tilt on image AMD chip ACC=10kV Aperture = 3 Mag = 1000X C lens = 10:00 Z=17mm Tilt = 0, 45, 60 - pictures 3, 4, 5, respectively Picture 3 Picture 4

Picture 5

Example #3 effect of accelerating voltage on image AMD chip ACC=10kV and 20 kv pictures 6 and 7, respectively Aperture = 3 Mag = 1000X C lens = 10:00 Z=17mm Tilt = 60 Picture 6 Picture 7

Example #3 effect of magnification on image AMD chip ACC=20 kv picture 8 and 25 kv pictures 9-11 Aperture = 4 Mag = 2K, 8K, 20K, and 30K, respectively C lens = 10:00 to 10:30 Z= 17 mm Tilt = 60 Picture 8 Picture 9

Picture 10 Picture 11