Measurement of table feed speed in modern CT

Similar documents
CT radiation profile width measurement using CR imaging plate raw data

CR Basics and FAQ. Overview. Historical Perspective

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

QC Testing for Computed Tomography (CT) Scanner

A comparison of two methods for the determination of freein-air geometric efficiency in MDCT

X-RAYS - NO UNAUTHORISED ENTRY

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D.

The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System

I. Introduction.

Key words: fluoroscopy, dose-area-product, kerma-area-product, calibration of KAP meters, patient exposure

Learning Objectives: What s my motivation? (unknown screen actor) Workshop Overview

Half value layer and AEC receptor dose compliance survey in Estonia

Digital Imaging Considerations Computed Radiography

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of

A new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

Quality assurance: a comparison study of radiographic exposure for neonatal chest radiographs at 4 academic hospitals

7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline

Development of new dosimeter for measuring dose distribution in CT

Nuclear Associates

Features and Weaknesses of Phantoms for CR/DR System Testing

Test Equipment for Radiology and CT Quality Control Contents

Practical Medical Physics Session: TG-151 Dose Monitoring. August 5, 2013 Katie Hulme, M.S.

SYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology

Overview of Safety Code 35

Computed Tomography. The Fundamentals of... THE FUNDAMENTALS OF... Jason H. Launders, MSc. Current Technology

While digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them.

RaySafe X2. Effortless measurements of X-ray

T h e P h a n t o m L a b o r a t o r y

DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE TG-148 RECOMMENDED TESTS 1. V.B.1.C. - Y-JAW DIVERGENCE/BEAM CENTERING

Maximizing clinical outcomes

8/3/2017. Use of EPIDs for Non-Routine Linac QA. Disclosure. Learning Objectives. Parts of this project received support from Varian Medical System.

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC?

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Do you have any other questions? Please call us at (Toll Free) or , or

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

Artefacts found in computed radiography

Outline ASRT Changes Impact on current curriculum Potential new courses WECM Changes Last update Resources and needs

Comparison of computed radiography and filmõscreen combination using a contrast-detail phantom

Maximum Performance, Minimum Space

Performance evaluation of the RITG148 + set of TomoTherapy quality assurance tools using RTQA 2 radiochromic film

The Current State of EPID-Based Linear Accelerator Quality Assurance. Disclosures. Purpose of this First Talk 8/3/2017

A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup

Nuclear Associates

Reducing Radiation Exposure from Survey CT Scans

Visualization of sources of scattered radiation from x-ray equipment used for interventional radiology

SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT. Exam Content Specifications 8/22/2012 RADT 3463 COMPUTERIZED IMAGING

INNOVATION BY DESIGN. Toshiba A History of Leadership REMOTE CONTROL R/F SYSTEM

computed tomography, computed tomography dose index, dosimetry

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London

Digital radiography (DR) post processing techniques for pediatric radiology

DIAGNOSTIC ACCREDITATION PROGRAM. Radiology and CT Quality Control Procedures Workbook

GafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications

2217 US Highway 70 East Garner, NC Main: Fax:

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter

The Versatile and Powerful ACLxy. ACLxy

Diagnostic X-Ray Shielding

HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE

Dose Reduction in Helical CT: Dynamically Adjustable z-axis X-Ray Beam Collimation

Truly flexible to meet your clinical needs

Exposure System Selection

COMPUTED TOMOGRAPHY 1

Evaluation of no-grid radiography using the digital scattered x-ray removal processing

diagnostic examination

Teaching Digital Radiography and Fluoroscopic Radiation Protection

ISO INTERNATIONAL STANDARD

Overview. Professor Roentgen was a Physicist!!! The Physics of Radiation Oncology X-ray Imaging

Estimation of signal transfer property for wireless digital detector in different measurement schemes

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance

Introduction. Sam R. Kottamasu Lawrence R. Kuhns

High Energy Digital Radiography & 3D-CT for Industrial Systems

Wide-Detector CT for TAVR Planning:

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

Digital radiography. bucky table and wall stand as a dual detector or wireless system. Amadeo DR Systems

Essentials of Digital Imaging

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Unfors EDD-30 Radiation Protection in Fluoroscopy

ISPFILMQATM STATE-OF-THE-ART RADIOTHERAPY VERIFICATION SOFTWARE. Supports all major radiotherapy technologies! FilmQA TM

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II

The Flash IIP Console is the heart of every FCR system. It s designed to maximize productivity in the busiest environments.

DIGITAL RADIOGRAPHY. Digital radiography is a film-less technology used to record radiographic images.

RADspeed Pro DIGITAL RADIOGRAPHIC SYSTEM. IBD-RADProAeroDR

of sufficient quality and quantity

160-slice CT SCANNER / New Standard for the Future

Acquisition, Processing and Display

Mammography: Physics of Imaging

X-RAY MEDICAL EQUIPMENT

Quantitation of clinical feedback on image quality differences between two CT scanner models

History of digital imaging

Open. the Digitized world. Fuji Computed Radiography

Amorphous Selenium Direct Radiography for Industrial Imaging

Y11-DR Digital Radiography (DR) Image Quality

RAD. Experiences Using the RADspeed Pro EDGE Package. 1. Hospital Description. 2. Background of Adoption. Hirohito Tanaka

Digital Image Management: the Basics

China Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW, 150kV Radiography System - HF50-R

Transcription:

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 3, 2014 Measurement of table feed speed in modern CT Atsushi Fukuda, 1,2a Pei-Jan P. Lin, 3 Kosuke Matsubara, 2 Tosiaki Miyati 2 Department of Radiology, 1 Shiga Medical Center for Children, Moriyama, Shiga, Japan; Division of Health Sciences, 2 Graduate School of Medical Sciences, Kanazawa University, Kanazawa, Ishikawa, Japan; Department of Radiology, 3 Virginia Commonwealth University Medical Center, Richmond, VA, USA ntoki@blue.plala.or.jp Received 16 September, 2013; accepted 31 December, 2013 The purpose of this study was to develop and evaluate a noninvasive method to assess table feed speed (mm/s) in modern commercial computed tomography (CT) systems. The table feed (mm/rotation) was measured at selected nominal table feed speeds, given as low (26.67 mm/s), intermediate (48.00 mm/s), and high (64.00 mm/s), by utilizing a computed radiography (CR) cassette installed with a photo stimulable phosphor plate. The cassette was placed on the examination table to travel through the isocenter longitudinally, with a total scan length of over 430 mm. The distance travelled was employed to determine the total table feed length. To calculate the table feed speed, gantry rotation time was measured concurrently at a preselected nominal rotation time of 750 ms. Upon completion of data acquisition, the table feed and gantry rotation time were analyzed and used to calculate the actual table feed speed (mm/s). Under the low table feed speed setting, the table feed speed was found to be 26.67 mm/s. Similarly, under the intermediate and high table feed speed settings, the table feed speed was found to be 48.10 and 64.07 mm/s, respectively. Measurements of the table feed speed can be accomplished with a CR system and solid-state detector, and the table feed speed results were in excellent agreement with the nominal preset values. PACS number: 87.57.Q- Key words: CT scanner, table feed speed, gantry rotation time, solid-state detector, computed radiography I. Introduction With the advent of slip ring technology, helical scanning computed tomography (CT) was introduced in 1988. (1,2) Helical scan technology, utilizing continuous gantry rotation and table feed, provided for longer anatomic coverage in a shorter time. Accurate control of the table feed speed is one of the important factors in modern CT system because scan pitch is a key factor in coronary CT angiography. (3,4) High-pitch (P = 3.2) CT angiography has been recently developed, and this technique requires a table feed speed of up to 430 mm/s. (4) Because scan pitch is defined as the ratio of the table feed per gantry rotation to the total nominal scan width, (5) a comprehensive quality assurance (QA) program should include assessing not only the beam width, but also the table feed speed and gantry rotation time. While the table feed speed is an important scan parameter of any CT scanner, the accuracy and precision of the table feed speed are not necessarily known to medical physicists, whose responsibilities may include the QA of CT scanners. The American Association of Physicists in Medicine (AAPM) published two reports, Report 39 (6) and Report 83, (7) on the QA of CT scanners which evaluate specifications and test procedures. Both reports included evaluation a Corresponding author: Atsushi Fukuda, Shiga Medical Center for Children, 5-7-30, Moriyama, Shiga prefecture, Japan 524-0022; phone: +81-77-582-6200; fax: +81-77-582-6304; email: ntoki@blue.plala.or.jp 275 275

276 Fukuda et al.: Measurement of table feed speed in CT 276 of performance of electromechanical components such as table to gantry alignment, table vertical and longitudinal digital indicators, radiation dose profiles, and the X-ray-generator kvp accuracy. While the table feed speed was not explicitly included in these reports, the authors opine that verification of the table feed speed is essential for the proper assessment of a modern CT system. To the best of our knowledge, no published methods of measuring the table feed speed are available. To determine the table feed speed, it is necessary to measure the gantry rotation time. Fukuda et al. (8) reported a simple noninvasive technique to measure the rotation time using two different types of solid-state detectors. On the basis of their investigation, we considered that a computed radiography (CR) system and solid-state detector would allow accurate measurement of the table feed speed. This study aims to provide a simple noninvasive approach to assess table feed speed in a modern commercial CT system. II. Materials and Methods A. CT scanner, CR system, and solid-state detector system A 4-channel multidetector CT scanner was employed for this study. The scan parameters employed were as follows: 1) tube potential, 80 kvp; 2) tube current, 10 ma; 3) total collimation width, 32.0 mm (8.0 mm 4-channel); 4) nominal table feed, 20.0, 36.0, or 48.0 mm/rotation (with nominal table feed speed, 26.67, 48.00, or 64.00 mm/s); 5) nominal gantry rotation time, 750 ms; and 6) nominal scan pitch, 0.625, 1.125, or 1.500. A CR cassette (354 mm 430 mm) installed with a photostimulable phosphor plate (Imaging Plate Cassette Type CC and IP ST-VI; FUJIFILM Holdings Corporation, Tokyo, Japan) was employed for this study. Thus, the maximum imaging length available was 430 mm for the measurement of tabletop travel. This CR system has three exposure data recognizer (EDR) modes: auto, semi, and fixed. The fixed mode was employed to avoid any raw data manipulation by CR system software. Furthermore, the CR images were processed in fixed EDR mode, with latitude = 4 and sensitivity = 5 using the AVE4.0 test menu, as described by Liu et al. (9) The Xi solid-state detector (Unfors Raysafe, Billdal, Sweden), which is designed for application in conventional radiography and fluoroscopy, (10) was employed for the measurement of gantry rotation time. As such, the Xi detector is backed with a lead support to limit the detection of backscatter. The signal obtained is sent to a laptop computer wirelessly via Bluetooth (Bluetooth SIG, www.bluetooth.com), and data analysis is handled by the Xi View software. B. Measurement of table feed per rotation and gantry rotation time The CR cassette in conjunction with an antiscatter grid (8:1, 60 l/cm, and focus distance 100 cm; Mitaya Manufacturing Corporation, Tokyo, Japan) was placed on the examination table top on a 1.0 mm lead plate to pass longitudinally through the isocenter, as shown in Fig. 1. Although the CR cassette is installed with a 0.13 mm thick lead support, an additional 1.0 mm thick lead sheet was added to minimize the possibility of radiation reaching the photostimulable phosphor plate from the back side of the cassette. A total scan length of over 430 mm was performed to acquire a helical CT radiation profile. The photostimulable phosphor plate was processed as described above, and the data were exported to a personal computer and analyzed with ImageJ software. (11) While the helical CT radiation profile was being obtained, the Xi detector was positioned face down on the gantry cover, as shown in Fig. 1. Because the Xi detector is backed with a lead support to shield it from backscatter, the detector can measure only the primary radiation as the X-ray tube passes by the detector at the bottom of the gantry where the detector is located. After the data were acquired, the peak times were determined with the Xi View software, and the gantry rotation time was determined as the time between two successive signal peaks. (8)

277 Fukuda et al.: Measurement of table feed speed in CT 277 Fig. 1. Experimental arrangement for measurements of table feed per rotation and gantry rotation time. The CR cassette was placed on the examination table on a 1 mm thick lead plate to travel through the isocenter longitudinally for the measurement of table feed per rotation under the helical scan mode. Concurrently, the solid state detector was positioned on the inner bottom of the gantry cover for gantry rotation time measurements. III. Results Figure 2 depicts the radiation dose profile from the CR cassette as it was scanned using a nominal scan pitch of 1.125 and nominal table feed of 36.0 mm/rotation. This figure includes ten complete radiation profiles. A feed length of 362.6 mm was measured over ten gantry rotations using the plot profile function provided in the ImageJ software. Thus, the actual table feed was determined to be 36.3 mm/rotation. Similarly, the actual table feed was determined to be 20.1 and 48.3 mm/rotation under the nominal table feed of 20.0 and 48.0 mm/rotation, respectively. The dose rate profile obtained from the Xi detector, after processing with the Xi View software, is shown in Fig. 3. There are two peaks located at rotation times of 44 and 799 ms, respectively. Therefore, the gantry rotation time is determined to be 755 ms. Table 1 shows the comparison between the nominal preset values and measured values for the parameters included in this study. To obtain the table feed speed, the table feed was divided by the gantry rotation time. As a result, the measured table feed speeds under the low, intermediate, and high table feed settings were determined as 26.67, 48.10, and 64.07 mm/s, respectively. It was a simple exercise to measure the distance between the first and subsequent scan peaks derived from the CR cassette, as shown in Fig. 2. After all distances were measured, the nominal distances were subtracted from the measured distances to calculate the scanning location error. Figure 4 shows the relation between the nominal scan length and scan length error for the three table feed speed settings. For all table feed speed settings, the errors increased linearly as a function of nominal scan length.

278 Fukuda et al.: Measurement of table feed speed in CT 278 Fig. 2. Illustration of the radiation dose profile and table feed distance under the intermediate table feed setting. The CR cassette located at the isocenter was scanned with a nominal gantry rotation time of 750 ms and nominal table feed per rotation of 36.0 mm/rotation, with X-ray settings of 80 kvp, 10 ma, 32 mm total collimation width. The table feed length over ten gantry rotations is 362.6 mm and the measured table feed per rotation is 36.3 mm/rotation. Fig. 3. Illustration of the detector output signals/peaks using an Xi detector for measurement of the gantry rotation time. The Xi detector was located on the inner gantry cover, and the measurements were conducted with a nominal gantry rotation time of 750 ms. The two peaks are registered at 44 and 799 ms, respectively. The measured gantry rotation time is 755 ms. Ta b l e 1. Nominal preset values and measured values for the parameters considered in this study. Table Feed Speed Table Feed Gantry Rotation Time Table Feed (mm/s) (mm/rotation) (ms) Speed Setting Preset Measured Preset Measured Preset Measured Low 26.67 26.67 20.0 20.1 750 755 Intermediate 48.00 48.10 36.0 36.3 750 755 High 64.00 64.07 48.0 48.3 750 755

279 Fukuda et al.: Measurement of table feed speed in CT 279 Fig. 4. Relation between the nominal scan length and scan length error. For all table feed speed settings, the errors increased linearly as a function of nominal scan length. A scan length of approximately 400 cm would result in an error of over 2 mm. Note that the scan length error is the nominal scan length subtracted from the measured scan length. IV. DISCUSSION The table feed speed is an important parameter that CT users have to manipulate when they are setting up scanning protocols. The ratio of scan length to scanning duration is used to determine the slowest table feed required to image the selected anatomical range within the desired time. To make the scanning protocols appropriate, it is important to understand not only the variable setting, but also the accuracy of the table feed speed. In this study, the CR system and solid-state detector were employed to measure the table feed and gantry rotation time. The table feed speed was easily obtained by dividing the table feed by the rotation time. We believe that our technique is useful to physicists because 1) the table feed speed and gantry rotation time can be measured without a time-consuming process, and 2) the technique does not require complicated QA tools or software as the CR system and solid-state detector are devices familiar to many departments of radiology. The Xi solid-state detector was employed to determine the gantry rotation time. Moreover, many detectors with a similar temporal resolution are available in the market. Most of these detectors would appear to be applicable for assessing gantry rotation time, although no definite claim can be made without verification. The table feed speeds for the three table speed settings were measured successfully with an error within 0.1 mm/s. Although the measured values were in excellent agreement with nominal preset values, the results show that measured table feed speeds were faster than nominal settings. On the other hand, the measured gantry rotation time (755 ms) was longer than the nominal setting (750 ms). As a result, scan length error was introduced, as shown in Fig. 4. In this study, the scan length error increase was approximately linear with nominal scan length (0.6% 0.9%). A scanning distance of approximately 400 cm would result in an error of over 2 mm, as shown in Fig. 4. To reduce the error, strict control of table feed speed and gantry rotation time is necessary. Unfortunately, to the best our knowledge, no published studies exist that evaluate the impact of table feed error. The American College of Radiology published the 2012 CT quality control manual, which includes a test of table travel accuracy. (12) The objective of the test is to verify that the patient table translates in accordance with nominal settings. A phantom with two sets of external fiducial markers of known separation is set on the couch and the distance between the fiducial markers as determined by table travel is compared with the known distance. The ACR test also does not account for potential variable table feed speeds. Therefore, we believe that to

280 Fukuda et al.: Measurement of table feed speed in CT 280 maintain the accuracy of table travel, it is important to measure not only the accuracy of the indicator, but also the table feed speed accuracy. The ACR recommends a limit of 2 mm (1.7%) over the 120 mm length of the phantom for its table travel accuracy test. In this study, the measured error over 120 mm scan length was within 1.0 mm (0.8%), which was well within the ACR limits. It may be necessary that the measurement of table feed speed be conducted with appropriate tabletop loading. AAPM Report 83 recommends that scanner table tests be conducted with the tabletop loaded with at least 150 lb (75 kg) of distributed weight to simulate a patient. (7) In this study, the measurements of the table feed speed were performed without tabletop loading, but the same methodology could be performed with table loading, if desired. Note that this study has two limitations. First, we had no access to a state-of-the art CT system having a high table feed speed of up to 430 mm/s. Under these high-speed conditions, it would be difficult to measure the table feed speed accurately using the CR system (354 mm 430 mm) employed. A long-view CR cassette and stitching software may be a promising technique to overcome this limitation. Second, we determined the radiation beam profile at 80 kvp using a CR cassette in accordance with the recommendation of Liu et al. (9) However, the CR photostimulable phosphor plate is not an adequate replacement for radiographic films to measure the CT beam profile at 120 kvp because the radiation dose is significantly high to delineate a clear radiation profile in fixed EDR mode with latitude = 4 and sensitivity = 5. Many pixels become saturated with the use of such high peak voltages even if the tube current is at its lowest setting (10 ma). Theoretically, there could be differences in the actual beam width of the radiation profile produced at 120 and 80 kvp; however, we believe that the scan length per rotation is invariant. V. Conclusions Despite some limitations, we proposed a simple noninvasive method for the measurement of table feed speed in modern commercial CT systems. We showed that measurements of CT scanner table feed speed can be accomplished with a CR system and solid-state detector. In addition, it is noteworthy to point out that the measurement results of table feed speed are in excellent agreement with the nominal preset values. However, the measured gantry rotation time in this study is higher than the nominal setting by 5 ms. Slower gantry rotation caused a scan length error that increased linearly as a function of the nominal scanning length. Finally, the data suggest that clinical medical physicists should consider adding periodic assessment of the accuracy and precision of the table feed speed to their QA program and consider this in their QC testing, where and when applicable. References 1. Zeman RK, Fox SH, Silverman PM, et al. Helical (spiral) CT of the abdomen. AJR. 1993;160(4):719 25. 2. Kalender WA and Polacin A. Physical performance characteristics of spiral CT scanning. Med Phys. 1991;18(5):910 15. 3. Sun Z, Choo GH, Ng KH. Coronary CT angiography: current status and continuing challenges. Br J Radiol. 2012;85(1013):495 510. 4. Flohr TG, Leng S, Yu L, et al. Dual-source spiral CT with pitch up to 3.2 and 75 ms temporal resolution: image reconstruction and assessment of image quality. Med Phys. 2009;36(12):5641 53. 5. International Electrotechnical Commission (IEC). Medical electrical equipment. Part 2-44: Particular requirements for the safety of x-ray equipment for computed tomography. IEC publication No. 60601-2-44, edition 2.1. Geneva, Switzerland: IEC; 2002. 6. Lin PJ, Beck TJ, Borras C, et al. Specification and acceptance testing of computed tomography scanners. AAPM Report No. 39. Report of Task Group 2 Diagnostic X-Ray Imaging Committee. NY: American Institute of Physics; 1993. Retrieved Aug 1, 2013 from http://www.aapm.org/pubs/reports/rpt_39.pdf

281 Fukuda et al.: Measurement of table feed speed in CT 281 7. Mutic S, Palta JR, Butker EK, et al. Quality assurance for computed-tomography simulators and the computedtomography-simulation process: report of the AAPM Radiation Therapy Committee Task Group No. 66. Med Phys. 2003;30(10):2762 92. 8. Fukuda A, Lin PJ, Matsubara K. Miyati T. Measurement of gantry rotation time in modern CT. J Appl Clin Med Phys. 2014;15(1):4517. 9. Liu HL, Liu RR, Reeve DM, Shepard SJ, Willis CE. Measurement of CT radiation profile width using CR imaging plates. Med Phys. 2005;32(9):2881 87. 10. Fukuda A, Miyati T, Matsubara K. Where should we measure the entrance air kerma rate during acceptance testing of the automatic dose control of a fluoroscopic system? Radiol Phys Technol. 2013;6(2):313 16. 11. ImageJ [website Home Page]. Note: ImageJ is available as free software. Retrieved Aug 1, 2013 from http:// rsbweb.nih.gov/ij/ 12. Cody DD, Pfeiffer D, McNitt-Gray MF, Ruckdeschel TG, Strauss KJ. CT quality control manual. Reston, VA: American College of Radiology; 2012.