DICOM Correction Proposal Form

Similar documents
29 CP Define CT Reconstruction Diameter more precisely and correct Enhanced CT illustration Page 1

DICOM Correction Item

26 CP Correct order of reference to pixel spacing values in SR Image Library

DICOM Correction Proposal Form

23 CP Clarify Enhanced US Volume Image and Frame Type Values 3 and 4

23 CP Clarify Enhanced US Volume Image and Frame Type Values 3 and 4

DICOM Correction Proposal

DICOM Correction Item

IHE Radiology Technical Framework Supplement. Stereotactic Mammography Image (SMI) Trial Implementation

20 CP Transverse positioning of pre-clinical research small animal subjects

Digital Imaging and Communications in Medicine (DICOM) Supplement 188: Multi-energy CT Images

4.0 How to Turn On the Selenia Dimensions

DICOM Correction Proposal

DigiMam Conformance Statement for DICOM V3.0

Digital Imaging and Communications in Medicine (DICOM) Supplement 188: Multi-energy CT Images

Tomosynthesis and Motion

Digital Imaging and Communications in Medicine (DICOM)

DICOM3.0 Conformance Statement

REGIUS CONSOLE CS-3. DICOM 3.0 Conformance Statement CODE NO Manufacturer: 1 Sakura-machi, Hino-shi Tokyo , Japan

Digital Imaging and Communications in Medicine (DICOM)

A proposal of a digital cephalogram standard in DICOM

Digital Imaging and Communications in Medicine (DICOM) Supplement 39: Add Stored Print Media Storage - Retire Normalized Print Media Storage

Digital Imaging and Communications in Medicine (DICOM) Supplement 56: Ultrasound Waveform

DICOM Conformance Statement

Mammography Solution. AMULET Innovality. The new leader in the AMULET series. Tomosynthesis, 3D mammography and biopsy are all available.

GE Healthcare. Senographe 2000D Full-field digital mammography system

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

DICOM Enhancements Current and Future

13 Compressed RGB components (rather than YBR) really are used by some WSI vendors in order to avoid the loss in conversion of 14 color spaces.

Graphing and Describing Reflections

Collimation Assessment Using GAFCHROMIC XR-M2

DICOM Conformance Statement

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

DICOM & IHE : An Update On -> the DICOM Standard -> the IHE Radiology Framework

2 nd generation TOMOSYNTHESIS

DICOM Conformance Statement

33 CP Clarify handling of ICC profiles in WADO-RS for encapsulated images such as JPEG Page 1

21 CP Clarify Photometric Interpretation after decompression of compressed Transfer Syntaxes Page 1

ADVANCED MEDICAL SYSTEMS PTE LTD Singapore Malaysia India Australia

IHE Radiology Technical Framework Supplement Nuclear Medicine Image Profile NMI with Cardiac Option

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions

TECHNICAL DATA. GIOTTO IMAGE SDL/W is pre-arranged for Full Field Digital Biopsy examination with the patient in prone position.

MEDICAL X-RAY 2D AND 3D IMAGE VIEWER:ROLE FOR THE MEDICAL IMAGE IN DICOM STANDARD

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London

1. Figure A' is similar to Figure A. Which transformations compose the similarity transformation that maps Figure A onto Figure A'?

Mammography: Physics of Imaging

The following clarifiction is made to the tender dossier CONTRACT NOTICE

A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup

Improved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques

Tmypacs. DICOM Conformance Statement 1.0. Document Version: Tmypacs. Product Name(s): 2.0. Release:

New spectral benefi ts, proven low dose

TORNIER BLUEPRINT. 3D Planning + PSI SCAN PROTOCOL

1. Queries are issued to the image archive for information about computed tomographic (CT)

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

Picasso-Trio 3-D Dental Imaging

35 CP JPEG-LS Planar Configuration constraints conflict with WSI, US, VL, Enhanced Color MR and Page 1 36 compressed RGB images

Thermionic x-ray. Alternative technologies. Electron Field Emission. CNT Based Field Emission X-Ray Source

QC Testing for Computed Tomography (CT) Scanner

(12) Patent Application Publication (10) Pub. No.: US 2017/ A1

INNOVATION BY DESIGN. Toshiba A History of Leadership REMOTE CONTROL R/F SYSTEM

Distributed source x-ray tube technology for tomosynthesis imaging

DICOM Implementations for Digital Radiography

7/20/2014. Outline. Outline. Disclosures. Learning Objectives. SBB: Practical Aspects of ACR Accreditation, QC and ACR On Site Surveys

Maximizing clinical outcomes

DICOM Implementations for Digital Radiography

Patient-Assisted Compression Impact on Image Quality and Workflow

Veraviewepocs 2D High Speed Panoramic X-Ray Crystal Clear Images with Reduced Radiation

Best choice in all DR. Jumong M. SG HealthCare Jumong M Digital radiography specifications

Protocol for the Quality Control of the Physical and Technical Aspects of Digital Breast Tomosynthesis Systems

The first task is to make a pattern on the top that looks like the following diagram.

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

icad, Inc. TotalLook TM DICOM Conformance Statement

CXI 1 micron Precision Instrument Stand

4/19/2016. Quality Control Activities for the RadiologicTechnologist. Objectives. 3D Tomosynthesis QC differences

ART Plus CdTe-Sensor Technology

Protocol for the Quality Control of the Physical and Technical Aspects of Digital Breast Tomosynthesis Systems

China Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW, 150kV Radiography System - HF50-R

DICOM Conformance Statement

Digital Breast Tomosynthesis

COMPUTED TOMOGRAPHY 1

P a n o r a m i c a n d C e p h a l o m e t r i c X - r a y s

ECOVIEW 9 / ECOVIEW 9 PLUS Digital Radiographic System

DICOM Conformance Statement

P a n o r a m i c a n d C e p h a l o m e t r i c X - r a y s

Truly flexible to meet your clinical needs

Outline. Digital Radiography. Understanding Digital Modalities: Image Quality and Dose. Image Quality. Dose Control

(12) United States Patent

Translating Protocols Between Scanner Manufacturer and Model

CT Scanner Dose Survey

RADspeed Pro DIGITAL RADIOGRAPHIC SYSTEM. IBD-RADProAeroDR

A Fast Monolithic System for Proton Imaging. Fritz DeJongh ProtonVDA Inc October 2017

25 CP Generalize Concepts in Abstract Multi-dimensional Image Model Component Semantics Page 1

Digital radiography. bucky table and wall stand as a dual detector or wireless system. Amadeo DR Systems

Image Quality. HTC Grid High Transmission Cellular Grid provides higher contrast images

DICOM Educational Conference Brisbane, Australia

Digital Panoramic X- Ray unit. Product Data

China Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW Digital RF System - HF51-5

1.1.Clinical Artefacts

Transcription:

DICOM Correction Proposal Form STATUS Final Text Date of Last Update 2014/09/04 Person Assigned Submitter Name Janet Keyes <Janet.Keyes@hologic.com> Makoto Suzuki (Toshiba) Submission date 2009.10.06 Correction Number CP-1032 Log Summary: Fix Position Angle sign inconsistency in Mammography Image Name of Standard PS 3.3 2013 + sup165_ft

Rationale for Correction: The text in the attribute description and the subsequent diagram for the Primary Positioner Angle seem to contradict each other with respect to which direction of rotation is positive. The left diagram seems to be drawn from the view of the patient facing into the page. (Or maybe I m misremembering my drafting conventions for placing views). Based on feedback that people were having trouble seeing the issue here is a more detailed explanation: The vertical component of the gantry has been omitted in the left view so it is unclear which way the view is facing. By the US/UK "third-angle" convention for orthgraphic views, the left view is looking from the patient/operator side of the gantry and the vertical is behind it. In that case, the labelling of both views are consistent with a right-handed coordinate system and both rotation directions follow the right hand rule. But then the attribute text for Primary Angle seems to conflict by saying that motion from the patient's right to vertical is positive. By the European/Asian "first-angle" convention for orthographic views, the left view is looking from the back of the gantry and the patient is behind the gantry facing us. In that case, the labelling of both views are consistent with a left-handed coordinate system and both rotation directions follow the left hand rule. Since DICOM convention is right-handed coordinate systems, we should either fix/clarify the attribute text, or clarify that Mammo uses a left-handed coordinate system. In either case it would help to add the vertical component and/or patient to the diagram in at least the left view. DICOM WG 15 notes: The referenced Figure C.8-8 was not intended to represent Positioner Primary Angle, but rather to clarify Distance Source to Detector and Distance Source to Patient. Those are the only attributes that reference C.8.11.7.1.1. Digital mammography modality manufacturers interpreted the Positioner Primary Angle description differently, so there is a large installed base of clinical digital mammography image data where the sign of the value for the same C-arm position differs depending on the manufacturer. WG 15 addressed this sign inconsistency in Positioner Primary Angle values in Supplement 165 in the Breast Projection X-Ray Image IOD (for digital breast tomosynthesis projection images), and proposes to apply the same clarification to the Digital Mammography X-Ray Image IOD (for conventional 2D mammography images) and the Breast Tomosynthesis Acquisition module of the Breast Tomosynthesis Image IOD (for digital breast tomosynthesis projection image characteristics included for reference with reconstructed slices). Correction Wording: Amend PS 3.3, C.8.11.7 Mammography Image module as follows C.8.11.7 Mammography Image Module Table C.8-74 contains IOD Attributes that describe a Digital Mammography X-Ray Image including its acquisition and positioning. Table C.8.-74. Mammography Image Module Attributes Positioner Primary Angle (0018,1510) 3 Signed Pposition in degrees of the X-Ray beam vector in the coronal anatomical plane as if the patient were standing facing the equipment where movement of the X-Ray source from right to vertical is

positive, and vertical is zero. Positioner Primary Angle Direction (0018,9559) 3 The interpretation of the sign of the Positioner Primary Angle (0018,1510). Enumerated Values: CW = Clockwise. Positioner Primary Angle movement is positive when patient s right (with patient standing, facing equipment). CC = Counter-clockwise. Positioner Primary Angle is positive when patient s left (with patient standing, facing the equipment). See C.8.X.2.1. <in sup165_ft, X not yet assigned> Positioner Secondary Angle (0018,1511) 3 Position in degrees of the X-Ray beam vector in the sagittal anatomical plane as if the patient were standing where movement of the X-Ray source from anterior to posterior is positive, and vertical is zero. C.8.11.7.1 C.8.11.7.1.1 Mammography Image Attribute Descriptions Mammography X-Ray Beam and X-Ray Beam Vector Definition Figure C.8.11.7-1 shows the X-Ray beam for a digital mammography system. The X-Ray beam vector is defined from the Focal Spot to the center of the chest wall line of the Image Detection device. Z Z X-Ray Beam X-Ray Beam Vector Image plane Positive rotation / Y O X O Positive rotation / X Figure C.8.11.7-1 X-Ray Beam for a Digital Mammography System Amend PS 3.3, C.8.21.3.4 Breast Tomosynthesis Acquisition module as follows Y

C.8.21.3.4 Breast Tomosynthesis Acquisition Module This section describes the Breast Tomosynthesis Acquisition Module. Table C.8.21.3.4-1. Breast Tomosynthesis Acquisition Module Attributes Attribute Name Tag Type Attribute Description X-Ray 3D Acquisition Sequence (0018, 9507) 1 Each Item represents an acquisition context related to one or more reconstructions. The values of the Acquisition Index (0020,9518) attribute may be used as index in this sequence. One or more Items shall be included in this sequence. >Include Table C.8.21.3.1.3-1 X-Ray 3D General Positioner Movement Macro Attributes >Per Projection Acquisition Sequence (0018,9538) 1 Sequence containing detailed acquisition context of each individual projection used in this acquisition context. One or more Items shall be included in this sequence. >>Include Table C.8.21.3.1.2-1 X-Ray 3D General Per Projection Acquisition Macro Attributes >>Positioner Primary Angle (0018,1510) 1 Signed Pposition in degrees of the X-Ray beam vector in the coronal anatomical plane as if the patient were standing facing the equipment where movement of the X-Ray source from right to vertical is positive, and vertical is zero. >>Positioner Primary Angle Direction (0018,9559) 3 The interpretation of the sign of the Positioner Primary Angle (0018,1510). Enumerated Values: CW = Clockwise. Positioner Primary Angle movement is positive when patient s right (with patient standing, facing equipment). CC = Counter-clockwise. Positioner Primary Angle is positive when patient s left (with patient standing, facing the equipment). See C.8.X.2.1. <in sup165_ft, X not yet assigned> >>Positioner Secondary Angle (0018,1511) 1C Position in degrees of the X-Ray beam vector in the sagittal anatomical plane as if the patient were standing where movement of the X-Ray source from anterior to posterior is positive, and vertical is zero.

Required if secondary positioner was used during acquisition. >>Exposure Time in ms (0018,9328) 1 Duration of X-Ray exposure in milliseconds. See C.8.7.2.1.1. Referenced figure and its context from sup165_ft: C.8.X.2 Breast X-Ray Positioner Macro Table C.8.X.2-1 specifies the Attributes of the Breast X-Ray Positioner Functional Group macro. C.8.X.2.1 Breast X-Ray Positioner Attribute Descriptions Figure C.8.X.2-1 illustrates the Defined Terms for Positioner Primary Angle Direction (0018,9559), indicating which C-arm movement direction defines a positive value for Positioner Primary Angle (0018,1510). Figure C.8.X.2-1 Illustration of Positioner Primary Angle Direction