PD233: Design of Biomedical Devices and Systems

Similar documents
HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

Photomultiplier Tube

SYLLABUS. 1. Identification of Subject:

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS

COMPUTED TOMOGRAPHY 1

X-Rays and endoscopes

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage

Beam-Restricting Devices

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

X-rays in medical diagnostics

1-1. GENERAL 1-2. DISCOVERY OF X-RAYS

80 Physics Essentials Workbook Stage 2 Physics

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah

X-ray Tube and Generator Basic principles and construction

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

V SALAI SELVAM, AP & HOD, ECE, Sriram Engg. College, Perumalpattu 1 MEDICAL ELECTRONICS UNIT IV

Medical Imaging: A Look inside. Medical Imaging. Medical Imaging. Visible Human Project

X-RAYS - NO UNAUTHORISED ENTRY

X-ray Tube and Generator Basic principles and construction

JEFFERSON COLLEGE COURSE SYLLABUS BET220 DIAGNOSTIC IMAGING. 3 Credit Hours. Prepared by: Scott Sebaugh Date: 2/20/2012

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE

CT Basics: Data Acquisition Module 3

Ludlum Medical Physics

Visibility of Detail

Seminar 8. Radiology S8 1

LECTURE 1 The Radiographic Image

Radionuclide Imaging MII 3073 RADIONUCLIDE IMAGING SYSTEM

QC Testing for Computed Tomography (CT) Scanner

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

China Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW, 150kV Radiography System - HF50-R

HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE

- KiloVoltage. Technique 101: Getting Back to Basics

Principle of X-Ray Systems

Exposure System Selection

Overview. Professor Roentgen was a Physicist!!! The Physics of Radiation Oncology X-ray Imaging

An Activity in Computed Tomography

I. Introduction.

STUDENT REVIEW QUESTION SET K CR/DR CONTENT AREA

Explain what is meant by a photon and state one of its main properties [2]

Unit thickness. Unit area. σ = NΔX = ΔI / I 0

Photons interaction with matter

Calibration of KAP meters

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question.

Acquisition, Processing and Display

An Activity in Computed Tomography

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II

CR Basics and FAQ. Overview. Historical Perspective

Today s Outline - January 25, C. Segre (IIT) PHYS Spring 2018 January 25, / 26

1. Carlton, Richard R., and Arlene M. Adler. Principles of Radiographic Imaging: An Art and a Science, 5th edition (2013).

CT Basics: Equipment and Instrumentation Module 2

High Energy Digital Radiography & 3D-CT for Industrial Systems

SCANNING ELECTRON MICROSCOPY AND X-RAY MICROANALYSIS

used to diagnose and treat medical conditions. State the precautions necessary when X ray machines and CT scanners are used.

NDE SOLUTIONS RADIOGRAPHY COURSE OUTLINE

BASICS OF FLUOROSCOPY

LECTURE 20 ELECTROMAGNETIC WAVES. Instructor: Kazumi Tolich

Radiographic Testing (RT) [10]

Mammography: Physics of Imaging

Diffraction-enhanced X-ray Imaging (DEXI) Medical Solutions. More information using less radiation

PET/CT Instrumentation Basics

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Test Equipment for Radiology and CT Quality Control Contents

X-ray Imaging (XRI) Lectures on Medical Biophysics. Dept. Biophysics, Medical Faculty, Masaryk University in Brno. X-ray Imaging (XRI)

PET Detectors. William W. Moses Lawrence Berkeley National Laboratory March 26, 2002

Phase Imaging Using Focused Polycapillary Optics

RADIOGRAPHIC EXPOSURE

How are X-ray slides formed?

Optimization of Energy Modulation Filter for Dual Energy CBCT Using Geant4 Monte-Carlo Simulation

Radiation Imaging. By: Engr. Joseph Ronald Cañedo

Digital Imaging CT & MR

DALLA LUCE VISIBILE AI RAGGI X: NUOVI RIVELATORI DI IMMAGINI PER RAGGI X A DISCRIMINAZIONE IN ENERGIA ED APPLICAZIONI

Computed Tomography. The Fundamentals of... THE FUNDAMENTALS OF... Jason H. Launders, MSc. Current Technology

LlIGHT REVIEW PART 2 DOWNLOAD, PRINT and submit for 100 points

DIGITAL RADIOGRAPHY ARTIFACTS

CT Basics: Image Quality Module 6

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development

Maximum Performance, Minimum Space

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

CyberKnife Iris Beam QA using Fluence Divergence

Introduction. MIA1 5/14/03 4:37 PM Page 1

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of

Spectroscopy in the UV and Visible: Instrumentation. Spectroscopy in the UV and Visible: Instrumentation

Medical Images Analysis and Processing

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions

Lunar Technology Advantages

PhysicsAndMathsTutor.com 1

Partial Replication of Storms/Scanlan Glow Discharge Radiation

Related topics Beam hardening, cupping effect, Beam hardening correction, metal artefacts, photon starvation

MaxRay Handheld X-ray Systems Operator Training Exam

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT)

RADIOGRAPHY TERMS TO KNOW SELF STUDY DENTALELLE TUTORING

Dental Radiography. One of the problems of dental radiography is having different dimensions than normal.

Digital Images & Image Quality

DENTAL RADIOGRAPHY KAMARUL AMIN BIN ABU BAKAR

COMPUTATIONAL IMAGING. Berthold K.P. Horn

Transcription:

PD233: Design of Biomedical Devices and Systems (Lecture-8 Medical Imaging Systems) (Imaging Systems Basics, X-ray and CT) Dr. Manish Arora CPDM, IISc Course Website: http://cpdm.iisc.ac.in/utsaah/courses/

Medical Imaging Systems X- ray Computed Tomography (CT) Magnetic Resonance Imaging (MRI) Ultrasound (US) Photoacoustics (PA) Optical Coherence Tomography (OCT) Image credit http://www.sprawls.org/resources/ Provide a window into the body to see anatomy and signs of pathology No window is perfect

Image credit http://www.sprawls.org/resources/

Imaging Test Result - + Accuracy of Diagnostic System Clinical questions: Is the bone fractured? Is a kidney stone present? Is their a blockage in the artery? Disease Present + - a True Positive b False Positive c False Negative d True Negative Sensitivity Probability of positive test given patient is sick Specificity Probability of negative test given patient is well What is Total accuracy?

Imaging Test Result - + Accuracy of Diagnostic System Positive Predictive Value: If the test is positive what is the probability what is the probability that the disease is present. Disease Present + - a b Negative Predictive Value: If the test is negative what is the probability what is the probability that the disease is absent. c d Prevalence: Number of diseases present in a given population at a given time

X-Ray Imaging "First medical X-ray by Wilhelm Röntgen of his wife Anna Bertha Ludwig's hand " by Wilhelm Röntgen. Reading material: Chapter 1, Kirk Shung

Electromagnetic (EM) wave Spectrum

X-Ray as Particle h = Plancks Constant = 4.13 x10-18 kev-sec What is ev? Calculate energy of single 1nm X-ray Photon

Attenuation of X-Ray beam I I-dI dx Beam of intensity I and cross-sectional area A di = -βidx β = Linear attenuation coefficient At what distance will the Intensity become half? What will happen is material changes state/density?

Attenuation of X-Ray beam Half Layer Value = 0.693/β Material HVL (mm) 30 kev 60 kev 120 kev Tissue 20.0 35.0 45.0 Aluminum 2.3 9.3 16.6 Lead 0.02 0.13 0.15 Mass-attenuation coefficient = β/ =density β=nσ Material has n atoms per unit volume each with cross section σ

Intensity of X-ray beam Intensity energy of the photons number of photons X-Ray Dose should also account for time of exposure Roentgen (R): total number of ions produced in 1cc of air at (760mm Hg and 0 o C) Radiation Absorbed Dose(rad): X-Ray energy absorbed per kg of material 1rad = 0.01 Joules absorbed per kg 1 gray (Gy) = 100 rad

X-ray Generation X Rays can be generated by bombarding metal targets with high energy electron White Radiation: Energy lost by striking electron interact with the positivity charged metal targets inelastically Also know as Bremsstrahlung or stopping radiation Characteristic Radiation: When inner shell electrons are removed by interaction striking electrons This phenomenon similar to photoelectric effect

X-ray Generators X Rays can be generated by bombarding metal targets with high energy electron X ray Tube Characteristics Target material Tube voltage Tube current Filament current Striking electrons heat up the metal target Line Focus Principle Large focal spot on the surface but small effective spot F= f sin( ) Rotating Anode 3000 to 10000 rpm

Beam Restrictors Needed to regulate size and shape of the x-ray beam Beam Restrictors: Aperture diaphragms Cone and cylinders Collimators Cllimators provide moveable opening Light used a guide to see the region to be exposed by x-ray Note Finite focal spot leads to penumbra along edges

Grids Used to remove effect scattered emissions Early image of x-ray with grid Snap on grid, attaches to the x-ray film cassette

X-ray Detectors X-Ray (Photographic) films X-Ray produces free electrons, which reduces silver halide in the exposed region Silver halide is black, hence region less exposed appear bright Digital Radiography (DR) Uses reversible chemistry Exposed film is scanned by variety of means -camera, drum scanner, laser scanning Alternatively, x ray detectors can be electronics/digitals Self study! X Ray film characteristics response curve, speed, fog, speed

X-ray Detectors Scintillation Detectors X-Ray photon can produce visible photon in scintillation material (NaI, Th) Visible photons and amplified by photomultiplier tube (PMT) -85% efficient Ionization Chamber Detectors X Ray ionizes inert gases in confined chamber place between charged electrodes. Amount of ions produced result in a current which digitized

Limitation of Conventional X-ray imaging 1) 2d Projection of 3D object i.e. multiple planes are mapped on to one plane depth information is lost 2) Limited use to distinguish soft tissue 3) Conventional X ray is not quantitative -Image intensity/size depend on source-object-detectors distance

Biological Effects of X-Ray Factors effecting biological effects: Threshold: Quantitative level above which there is an tissue damage happens Exposure Time: Exposure Area: Biological Variation: Response varies from varies from species to species, tissue to tissue Lethal dose vs short term effects

Biological Effects of X-Ray LD 50/30: Dose of substance or radiation which will kill 50% of the individual over a 30 day period. Lethal dose for humans is ~450 rad Short term effects like nuseua, vomiting can happen at dose of 100rad + carcinogenic effects + genetic effects Even diagnostic X ray is harmful!!

Conventional Tomography

1 st Gen. Computed Tomography -Few minutes for each scan -Pencil beams -Motion artifacts -Translate and Rotate Scanner Images source: www.kau.edu.sa/files/0008512/files/19500_2nd_presentation_final.pdf

2 st Gen. Computed Tomography Multiple detector for single beam Initial versions with 3 detectors later upto 50+ detectors Still uses translate and rotate scanner Images source: www.kau.edu.sa/files/0008512/files/19500_2nd_presentation_final.pdf

3 st Gen. Computed Tomography 300-500 detectors Designed for pure rotational scanning X ray tube collimated for fan-beam Scanning time reduced to 2 sec per slice Got rid of translate and rotate scanning even used in most recent configurations Images source: www.kau.edu.sa/files/0008512/files/19500_2nd_presentation_final.pdf

4 th Gen. Computed Tomography Circular array of fixed detectors Only source rotates 600-4800 detectors Less efficient as only ¼ of detectors used at any point in time. Images source: www.kau.edu.sa/files/0008512/files/19500_2nd_presentation_final.pdf

Images source: www.kau.edu.sa/files/0008512/files/19500_2nd_presentation_final.pdf

5 th Gen. Computed Tomography Cine CT/ millisecond CR/ultrafast CT Stationary-Stationary configuration no mechanical scanning X ray source single tube with array of tungsten targets Reduced scanning time to 50ms, cardiac scanning made possible Images source: www.kau.edu.sa/files/0008512/files/19500_2nd_presentation_final.pdf

6 th Gen. Computed Tomography Spiral/Helical CT Table translation with source rotation Slip ring technology X ray source continuously Volume data interpolation algorithms developed Whole abdomen in 30sec (1BH) Images source: www.kau.edu.sa/files/0008512/files/19500_2nd_presentation_final.pdf

6 th Gen. Computed Tomography Spiral/Helical CT Table translation with source rotation Slip ring technology X ray source continuously Volume data interpolation algorithms developed Whole abdomen in 30sec (1BH) Images source: www.kau.edu.sa/files/0008512/files/19500_2nd_presentation_final.pdf

7 th Gen. Computed Tomography MDCT/ Cone beam CT Multi-row Detector CT Collimator opened even more Key advance in detector technology 2D arrays rather then one 1D array Images source: www.kau.edu.sa/files/0008512/files/19500_2nd_presentation_final.pdf

7 th Gen. Computed Tomography MDCT/ Cone beam CT Multi-row Detector CT Collimator opened even more Key advance in detector technology 2D arrays rather then one 1D array Images source: www.kau.edu.sa/files/0008512/files/19500_2nd_presentation_final.pdf

War on slices!! Images source: www.kau.edu.sa/files/0008512/files/19500_2nd_presentation_final.pdf

Images source: www.kau.edu.sa/files/0008512/files/1 9500_2nd_presentation_final.pdf 5 th,6 th,7 th Gen. Computed Tomography

Gen. Source Source Collimation Detector 1st Single X-ray Tube Pencil Beam Single 2nd Single X-ray Tube Fan Beam (not enough to cover FOV) Multiple 3rd Single X-ray Tube Fan Beam (enough to cover FOV) Many 4th Single X-ray Tube Fan Beam covers FOV Stationary Ring of Detectors 5th Many tungsten Fan Beam Stationary anodes in single Ring of large tube Detectors 6th 3G/4G 3G/4G 3G/4G 7th Single X-ray Tube Cone Beam Multiple array of detectors