Electromagnetic compatibility Guidance and manufacturer s declaration DIN EN :2007 (IEC :2007)

Similar documents
Guidance and Declaration - Electromagnetic Compatibility (EMC) for the Delfi PTS ii Portable Tourniquet System

BIODEX MULTI- JOINT SYSTEM

HeRO duet

Powered Traction Unit OPERATION MANUAL

General Safety/EMC and Electrical Information for i-limb ultra and i-limb digits

English

Biological Safety. Electromagnetic Compatibility (EMC) Observe the following precautions related to biological safety.

Technical Specifications Micromedical VisualEyes 505 by Interacoustics

D C 01/2019 3

VN415/VO425 Specifications

(c) Medisave UK. Notice. Edition 1. i-pad Operator s Manual

Nursing Beds with Dewert drive system

MedRx Avant Polar HIT AH-I-MPHITS-5 Effective 11/07/11

User Instruction Computer Assisted Local Analgesia. 337 Marion, Le Gardeur QC, Canada, J5Z 4W8

OtoRead - Technical Specifications Page 0. Technical Specifications. OtoRead D A 2017/06

Technical Data. Electrocardiograph ECG-1250K TD.ECG1250_L. This technical data may be revised or replaced by Nihon Kohden at any time without notice.

TH008F Multi-function Infrared Forehead Thermometer

Rolyan Splint Pan OPERATION MANUAL. Item # Small Item # Large

USER MANUAL MHS-2500I. Please take time to read these instructions before starting to use the scale. Version /17

INSTALLATION MANUAL AVANT HIT+ Hearing Instrument Test Chamber.

INTRODUCTION. 4 SAFETY INSTRUCTIONS. 5 ABOUT THIS DEVICE. 6 FIRST OPERATION. 7 THE DISPLAY. 8 ATTACHING THE WRIST SLEEVE. 9 CORRECT MEASUREMENT.

PHYSIOFLOW Q-LINK TM

#

DENTAL X-RAY OPERATOR'S INSTRUCTIONS. (for USA) Wall Mount Type...WK

Ceiling Model Ceiling Model for LEDview Plus and Heliodent Plus. Installation Requirements. New as of: English

Operating Manual Infrared thermometer

USER MANUAL M-200. Please take time to read these instructions before starting to use the scale. Version /06

Not for print. Microphone Test Device with SONNET MTD Adapter. User Manual. Cochlear Implants NOT FOR PRINT

M-400 M-410 M-420 M-430

The following symbol indicates that the device is MR-unsafe:

IEC Second Edition

`bob`=p=^åèìáëáíáçå=råáí

The Symbol means that the Vet-Tech (200, 250 or 300HF) system intentionally applies some radioelectric energy for diagnostic or treatment purposes.

`bob`=p=^åèìáëáíáçå=råáí=bma

: 0089 GTIN

Rotating Anode X-Ray Tube Housing Assembly. General Data

ENG en. Operating instructions. Iris Magneton MF Wellness therapy

The following languages can be found on our website: French. German. Spanish. Italian. Swedish. Portuguese. Russian. Dutch.

Laboratory Accreditation Programmes

WRIST BLOOD PRESSURE MONITOR

#0086.

SAVI SCOUT Surgical Guidance System. Console Operation Manual

H2 Check Operating Manual

MINISCAV Vacuum Pump Manual

L NKTEMP. Non-contact. Infrared Thermometer. User Manual LMP001

OPERATOR'S INSTRUCTIONS DX-073 DENTAL X-RAY

OPERATOR'S INSTRUCTIONS DENTAL X-RAY MODEL 096

EMC Seminar Series All about EMC Testing and Measurement Seminar 1

USER S MANUAL Volk Pictor Plus Fluorescein Angiography (FA) Module For Use with: Volk Pictor Plus

One-day Conference 18 March Power Supply, EMC and Signalling, in Railway Systems

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

By order of ZHONGSHAN LIANGYI LIGHTING CO., LTD. at Zhongshan, China

Power Quality Issues from an EMC Point of View

GTIN:

INSTRUCTIONS FOR YOUR NEW PLASMAFLOW. Vascular Therapy System (Compressible Limb Sleeve Device)

EMC Test report for LED Panel Light Models , , , , ,

This annex is valid from: to Replaces annex dated: Locations where activities are performed under accreditation

High precision measurement system for current and voltage IHC-A/B-RM01/03

Exergen TAT 5000S RS232 TTL Supplemental Instructions for Use

CardioMessenger II-S Transmitter for BIOTRONIK Home Monitoring. Technical Manual

SPAC265-8W. AC-DC power supply module. Features. Description. Applications

SJM MRI Activator. Handheld Device. User's Manual. Model EX4000

EMC standards. Presented by: Karim Loukil & Kaïs Siala

By order of American Power Conversion Holdings Inc. at New Taipei City, Taiwan

EN61326 EMC COMPLIANCE REPORT on the LP Series Ultrasonic Transmitter Remote Amplifier and Transducer for Hawk Measurement Systems Pty Ltd

An Introduction to EMC Testing (what can be done with scopes) Vincent Lascoste EMC Product Manager - RSF

OPERATOR S INSTRUCTIONS (for USA & Canada)

Transcutaneous Electrical Nerve Stimulator

OPSJ1900 J1900 Open Pluggable Specification Products

Multi-function (MI, ME, M2, M1) 2 NO (2 SPST-NO) 12 / / 400 3,000 1, / 0.3 / (5 / 5) AgNi

BodyclockStimplusManual.qx8_Layout 1 17/08/ :53 Page 1. Body Clock Stimplus. Instructions for use

2620 Modular Measurement and Control System

Central Blood Pressure Meter Model cbp301. Operating Manual

Operator s Manual External Remote Controller (ERC)

DIGITAL BLUETOOTH THEROMETER PBT

EMC Test Report. Report Number: M030826

TEST SUMMARY. Prüfbericht - Nr.: Test Report No.: Seite 2 von 25. Page 2 of 25

EICKEMEYER INFUSION WARMER

R&TTE Testing For EU Market

OPERATOR S MANUAL AN APPLIED DIGITAL COMPANY

EMC test report AU01+E04

User Manual. Before Using Your WiTouch Pro Device

TEST SUMMARY. Prüfbericht - Nr.: Test Report No.: Seite 2 von 27. Page 2 of 27

Overview of EMC Regulations and Testing. Prof. Tzong-Lin Wu Department of Electrical Engineering National Taiwan University

This annex is valid from: to Replaces annex dated: Location(s) where activities are performed under accreditation

User instructions for the ABI-system 100

AS/NZS CISPR 14.2:2015

MID energy meter B21 Single-phase energy meter, 65 A

POCKET AIR. Portable Nebulizer. Instruction Manual MBPN002 / MB / MB05006

AUTOMATIC BLOOD PRESSURE MONITOR (Arm Cuff Type) INSTRUCTION MANUAL

Harmonic Current emission EN :2014 Class A Pass. Voltage Fluctuation and Flicker EN :2013 Clause 5 Pass

Relief. At Last. User Manual For an effective treatment, read the entire User Manual before using Kyrobak.

C series Product brief

Discontinuous Disturbance (Click) EN :2006+A1:2009+A2:2011 Clause N/A** Radiated Emission, 30MHz to 1000MHz

Welch Allyn Home Scale (T- RPM-SCALE100)

National Voluntary Laboratory Accreditation Program

Series Rev 1.2 Dec. 2015

Accr. 002-Test EN-ISO (2005)

Draft. User s Manual. Transmitter Model EX1150

ANCOM s Laboratory for Electromagnetic Compatibility and Radio Equipment Testing

Transcription:

Compressor set Equipment Under Test (EUT) Type 028 Type 047 Type 052 Type 085 Electromagnetic compatibility Guidance and manufacturer s declaration DIN EN 60601-1-2:2007 (IEC 60601-1-2:2007) 2017 PARI GmbH Spezialisten für effektive Inhalation, 041D0466-C 02/17

3 1 ELECTROMAGNETIC COMPATIBILITY Medical electrical devices are subject to special precautionary measures in particular regarding EMC with the installation and operation. Portable and mobile HF-communication devices e.g. mobile phone can affect medical electrical devices. Use of accessories and lines other than those indicated can lead to increased transmission or reduced noise immunity of the equipment. The equipment has to be operated exclusively with original accessories. The device should not be used beside or on top of other equipment. If such use is unavoidable, the device should be observed to verify normal operation in the configuration in which it will be used. The EUT is intended for use in the electromagnetic environment specified below. The customer or the user of the EUT must ensure that it is used in such environment. There is no essential performance. 2 ELECTROMAGNETIC EMISSIONS Emission test Compliance Electromagnetic environment Guidance RF emissions CISPR 11 Group 1 EUT uses RF energy only for its internal function. Harmonic emissions IEC 61000-3-2 Voltage fluctuations / flicker emissions IEC 61000-3-3 Class B Not applicable Passed Table 1 of DIN EN 60601-1-2:2007 (IEC 60601-1-2:2007) Therefore, RF-emission is very low and not likely to cause any interference with nearby electronic equipment. The EUT is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.

4 3 ELECTROMAGNETIC IMMUNITY Immunity test Electrostatic discharge (ESD) IEC 61000-4-2 Electrical fast transient / burst IEC 61000-4-4 Surge IEC 61000-4-5 Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11 Power frequency (50 Hz/60 Hz) magnetic field IEC 61000-4-8 Conducted RF IEC 61000-4-6 IEC 60601 test level ±6 kv contact ±8 kv air ±2 kv for power supply lines ±1 kv for input / output lines ±1 kv common mode ±2 kv differential mode <5% U T 1) (>95% dip of U T ) for 1/2 cycle 40% U T (60% dip of U T ) for 5 cycles 70% U T (30% dip of U T ) for 25 cycles <5% U T (>95% dip of U T ) for 5 s Compliance level ±6 kv contact ±8 kv air ±2 kv for power supply lines ±1 kv for input / output lines ±1 kv common mode ±2 kv differential mode <5% U T 1) (>95% dip of U T ) for 1/2 cycle 40% U T (60% dip of U T ) for 5 cycles 70% U T (30% dip of U T ) for 25 cycles <5% U T (>95% dip of U T ) for 5 s Electromagnetic environment Guidance Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%. Mains power quality should be similar to that of a typical commercial or hospital environment. Mains power quality should be similar to that of a typical commercial or hospital environment. Mains power quality should be similar to that of a typical commercial or hospital environment. When the user of the Medical Electrical Equipment continued function also calls in the event of disruption of supply, it is recommended the EUT from an uninterruptible power supply or a battery. 3 A/m 3 A/m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment. 3 V eff 150 khz to 80 MHz 3 V eff 150 khz to 80 MHz Portable and mobile RF communication equipment should be used no closer to any part of the EUT, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.

5 Immunity test Radiated RF IEC 61000-4-3 IEC 60601 test level 3 V/m 80 MHz to 2,5 GHz Compliance level Electromagnetic environment Guidance 3 V/m Recommended separation distance: d = 3,5/3 * SQRT (P) d = 3,5/3 * SQRT (P) 80 MHz to 800 MHz d = 7/3 * SQRT (P) 800 MHz to 2,5 GHz where "P" is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey 2), should be less than the compliance level in each frequency range 3). Interference may occur in the vicinity of equipment marked with the following symbol: Tables 2 and 4 of DIN EN 60601-1-2:2007 (IEC 60601-1-2:2007) 1) U T is the AC mains voltage prior to application of the test level. 2) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the EUT is used exceeds the applicable RF compliance level above, the EUT should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the EUT. 3) Over the frequency range 150 khz to 80 MHz, field strengths should be less than 3 V/m. Note 1: At 80 Hz and 800 MHz, the higher frequency range applies. Note 2: These guidances may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

4 RECOMMENDED SEPARATION DISTANCES For portable and mobile RF communication equipment and the EUT. The EUT is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The customer or user of the EUT can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communication equipment (transmitters) and the EUT as recommended below, according to the maximum output power of the communications equipment. Rated maximum output power of transmitter [W] Separation distance according to frequency of transmitter in [m] 150 khz to 80 MHz d = (3.5/3) SQRT(P) 6 80 MHz to 800 MHz d = (3.5/3) SQRT(P) 800 MHz to 2.5 GHz d = (7/3) SQRT(P) 0.01 0.2 m 0.2 m 0.3 m 0.1 0.4 m 0.4 m 0.8 m 1 1.2 m 1.2 m 2.3 m 10 3.8 m 3.8 m 7.3 m 100 12 m 12 m 23 m Table 6 of DIN EN 60601-1-2:2007 (IEC 60601-1-2:2007) For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres [m] can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts [W] according to the transmitter manufacturer. Note 1: At 80 MHz and 800 MHz, the higher frequency range applies. Note 2: These guidances may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

PARI GmbH Spezialisten für effektive Inhalation Moosstraße 3 82319 Starnberg GERMANY Tel.: +49 (0) 8151 279-0 Fax: +49 (0) 8151 279-101 info@pari.de www.pari.com 2017 PARI GmbH Spezialisten für effektive Inhalation, 041D0466-C 02/17