6.11. TESTING A PACEMAKER

Size: px
Start display at page:

Download "6.11. TESTING A PACEMAKER"

Transcription

1 Purpose of experiment To examine a pacemaker s signals. Tasks of experiment TESTING A PACEMAKER Using an oscilloscope to measure pacemaker s parameters: voltage amplitude, pulse repetition rate, pulse duration and duration of front and back slopes, pulse energy, pacemaker s lifetime. Theoretical topics Electrical phenomena of human body. Heart electrical system. Structure of a pacemaker. Pacemakers electrodes. Electroshock. Electrical safety. Equipment and materials Oscilloscope, pacemaker, connecting wires. Theoretical part The heart is a single muscle, which produces an electrical signal when the muscle depolarizes. The sinus node on the wall of the right atrium of the heart of generates an action potential at a repetition rate depending on the work done by the body. Those potentials travel through a conduction system of the heart forcing myocardium contraction and thus pumping blood. In each case, the pulse parameters should depend on the patient's age, body temperature, hemodynamic and other factors. The measurement of this electrical signal (the electrocardiograph) is one of the oldest medical techniques. The ECG/EKG was first recorded by Einthoven in the Netherlands and Waller in England in 1895 (please, read more in the description of lab. work No 6.10.). When the heart cannot by itself generate natural electrical signals or they are not strong enough, electrostimulation is used. The main idea of electrical stimulation is that natural biological electrical pulses are replaced by external artificial electrical pulses. The idea of electrically stimulating the heart dates back to the beginning of the last century with the work of Burns and Aldini (a nephew of Galvani). But it was 1927 by the time Hyman built the first functioning external pacemaker, a small electric clockwork-driven generator. In 1948, Shockley, Bardeen, and Brattain invented the transistor and made it possible to drastically reduce the size of electric switching units, which also advanced the development of pacemaker design. The first implantable pacemaker was implanted in 1958 by a Swedish doctor named Elmquist. This device was made of 20 discrete components and weighed about 180 g. Pacemakers today weigh FBML TESTING A PACEMAKER 1

2 about 60 g and have the functionality of a small computer. Pacemaker therapy is based on the delivery of current pulses, which lead to the artificial depolarization of some cardiac cells. They go across the conduction system of the heart, as well as across the gap junctions (intercellular ionic bonding channels, which serve to transmit the excitement directly from cell to cell), triggering a complete contraction of the heart. Based on this triggering effect of artificial pulses, this is referred to as the all-or-nothing law. Nowadays, the electrocardiostimulation system consists of the following main components: of Pulse generator Batery Connecting wires Electrode Fig Electrocardiostimulation system. Heart an external programming device and the actual cardiac pacemaker (Figure ). The pacemaker consists of a battery, electronic elements (pulse generator and wires) and electrodes, and must match the parameters of the action potential produced by a real sinus node in the heart: amplitude, frequency, duration of the pulse and its slopes. The electrode is advanced almost exclusively via a venous access (e.g., the vena cava sup.) underneath the collarbone into the right half of the heart and anchored there. A permanent pacemaker is enclosed in a case and inserted into the body. Afterward, the pacemaker is connected and implanted subcutaneously in a pocket of skin. In addition, it must conform to the medical and biological requirements: all parts must be sealed, and the materials should be non-reactive for body tissues. A tightly sealed housing made of stainless steel or titanium prevents contact between the internal electrical circuit and the heart tissue. The electronics are usually designed as a hybrid, multichip module. The once-used thick film technique is being increasingly replaced by multilayer ceramics due to the increasing complexity of the circuits. Recently, brand new PCB (printed circuit board) technologies have been attracting interest, which, thanks to their flexibility, allow electric connection technology to be considerably simplified, thus improving the device quality. When the pacemaker is programmed, the system should be optimally adapted to the needs of the patient, and the lifetime of the battery should be maximized. With the long running times of today, it cannot be assumed that the conditions on the day of implantation will remain constant over the entire lifetime of the pacemaker. It is therefore necessary to be able to adapt the pacemaker to changing conditions. External devices that are usually based on a PC platform and drive manufacturer-specific shortrange telemetric equipment are used for programming. This data transmission, usually based on inductive nearfield coupling in a range below 100 khz, allows the implant to be queried/reprogrammed. To do this, the programming head is first placed on the skin above the pacemaker housing. Afterward, the current pacemaker parameters, as well as diagnostic values (markers, event histograms, ECG, etc.), are transmitted out and evaluated. Finally, the updated parameters are reloaded into the implant. The pacemaker electrode or probe establishes the connection between the pacemaker and the heart. It consists of a connector, i. e., the plug for connection to the pacemaker, the electrode conductor, and the electrode tip (Fig ). Fig Schematic structure of a pacemaker electrode. The connector configuration has been standardized according to the IS-1 standard, so that all modern electrodes can be connected FBML TESTING A PACEMAKER 2

3 to any pacemaker. One of the main technical challenges is the high bending stress of about 40 million load changes per year. For this reason, electrode conductors nowadays are made of four individual coiled wires, which provide the highest possible reliability with high flexibility. With a coiled wire, the bending radius is reduced, which reduces the alternating load stress on the feed line material (usually special stainless steel). For this reason, electrode breakage is no longer a major problem. Special demands are placed on the insulation material. It must be biocompatible and also be able to withstand the mechanical and chemical loads in the body. Right now, insulation made of silicone and polyurethane are being used. Polyurethane has a much lower friction coefficient, which is why it is especially preferred when implanting several electrodes via the same vessel. Furthermore, a distinction is made between uni- and bipolar pacemaker electrodes). Unipolar means that the electrode tip acts as the cathode and the pacemaker housing (or another counterelectrode with a large surface area) acts as the anode. Due to the considerably greater surface area of the pacemaker compared to the electrode tip, the housing is also called an indifferent pole. Bipolar probes also work with a cathodic tip. The anode, however, is also placed in the distal electrode area (about 2.5 cm away from the tip). Bipolar probes are somewhat thicker and more rigid than unipolar electrodes since they have two feed line coils. It is hardly possible to repair electrode breakage. A widespread compromise is the implantation of bipolar probes in the atrium of the heart (a smaller potential requires a higher sensitivity and the suppression of interference signals) and unipolar probes in the ventricle area. In addition, there are two types of the electrodes connections: endocardial (Figure a) and epicardial (Figure b). a b Fig Types of the electrodes connections: endocardial (a) and epicardial (b). Usually a pacemaker produces a constant sequence of short semirectangular pulses (Figure ). The electrical pulses produced by a pacemaker travel to the heart muscles through insulated wires ending in needle electrodes which touch the heart tissue. The effective amplitude of the pulses should generate an action potential in tissue touched by an electrode. The effectiveness of T that process depends strongly also on the front U a,v slope of the pulse, being higher for an extremely steep slope (slope of a short duration). In general, the duration of the slope is of the order of microseconds, while i requirements for the back slope of the pulse are very minor. As for the potential Time, t Fig Sequence of pacemaker s output pulses. FBML TESTING A PACEMAKER 3

4 (amplitude of pulse) it should be considerably bigger than the action potential of a cell (100 mv), for most pacemakers it is in the range of several volts ( V) to overcome possible losses in the electrode-tissue contact. A pacemaker stimulates at a defined rate approximately from 25 beats/min to 155 beats/min. Frequency-adaptive pacemakers allow the heart rate to be increased in stressful situations, even in patients with sinus node dysfunction. They are therefore indicated for physically fit patients who have an insufficient heart rate increase under stress (chronotropic insufficiency). The frequency adaptation is achieved via sensors, which determine the stress-related parameters and relate these to the stimulation frequency (so-called interference coupling). Nowadays, lithium-iodide batteries (opencircuit voltage about 2.8 V, capacitance about 1Ah, inner resistance a few 100 and 50 kω) are usually used as a power source. With this type of battery, the anode is made of lithium and the cathode of iodine. In addition to its extraordinarily low selfdischarge (< 1%/year), the lithium-iodide battery offers high stability of the inner resistance beyond the discharge time. Only at the end of the service life does the open-circuit voltage rapidly drop. Lithium-iodide batteries thus offer the greatest possible reliability with their long lifetime, small dimensions, and light weight. Today, they have running times ranging from 5 to 10 years with this technology. The trend toward primary cells, or rechargeable batteries, which are at the beginning of their technical development and are occasionally discussed even today, offer no major advantages. The supposed prolonged service life must be weighed against the drastic increase in aging-related failures after about 10 years, which makes an implant change recommendable after this time, anyway. Research has recently been carried out with nuclear power sources, but in this case there is a problem of introducing radioactive material into the body. The lifetime of each source depends on its capacity - accumulated electrical charge q, which is usually expressed in ampere-hours (Ah). Knowing this value, it is possible to evaluate the maximum lifetime of a pacemaker. The pulse shape should be nearly rectangular with an extremely steep front slope. The energy required to generate a single pulse Ei can calculated by: E i 2 U m, (6.11.1) R k here Um is the amplitude of the pulse measured in Volts, Rk is the resistance of the tissue, τ duration of a pulse measured according to Figure can change from 0.1 to 2.3 ms. It is possible to evaluate the life time of a pacemaker taking into account only the battery capacity: QT t, (6.11.2) U m R k here t lifetime of the pacemaker (battery), Q initial charged energy of the battery, T period of pulses. Note that 10% of the amplitude is neglected, to prevent uncertainty when fixing the start and end of the pulse slope due to the rounded start or end of the pulse slope. The repetition rate of modern pacemaker pulses is adjustable, responding to the workload of the body and reacting in the same way as a normal heart. Fig Shape of a single pulse (image on the oscilloscope screen). FBML TESTING A PACEMAKER 4 1 0,9 0,1 0 D C E

5 Methodology We can gain information about processes in the human body by measuring such dimensions as: temperature, blood flow rate, lung volume, air flow velocity in the airways, chemical composition of inhaled and exhaled gas, and blood sugar. These and other data are recorded by the equipment based on their physical properties and finally are converted to electrical signals. The on-line device for thorough examination of electrical signals is an oscilloscope. The main technical data for the oscilloscope DSO TDS210 or TDS220 used in the laboratory: deflection values of vertical (y axis) channel: 5 V/grid to 2 mv/grid, X - axis scan speed: from 5.00 s/grid to 5.00 ns/grid. Procedures 1. Plug the power cable into the mains socket and switch the oscilloscope "POWER" on. Wait for confirmation that the oscilloscope s selfmonitoring is complete. 2. Connect the pacemaker s output wires to the oscilloscope. 3. Choose the desired x and y axis values to get a stable image of about three pulses well fitted to the oscilloscope screen (fig ). 4. Copy the pulse image into your laboratory digest for further examination. 5. Make a readout of the parameters: a) voltage amplitude Um : B A k 1 k 3 Fig Pulse signal oscillogram. U m Ak 1, (6.11.3) here A is the vertical length (in grid) of an amplitude on the screen and k1 is the value of the voltage indicator. b) peacemaker pulse period T : T Bk 3, (6.11.4) here B is the horizontal distance (in grid) between two pulses on the screen and k3 is the value of the time scale indicator. 6. Calculate pulse frequency ν : 1/T. (6.11.5) 7. Repeat the measurements of each parameter several times to improve accuracy. Calculate the mean values of the parameters. 8. Select an x axis value so that only one expanded pulse is visible on the entire screen (as shown in Figure ). FBML TESTING A PACEMAKER 5

6 9. Copy the pulse image into your laboratory digest for further examination. 10. Measure C, D, E characteristics and calculate pulse duration τ : Ck 3, (6.11.6) and also leading slope τ f1 and ending slope τ f2 : f 1 Dk 3 and f 2 Ek 3. (6.11.7) 11. Calculate the energy Ei required to generate a single pulse using equation (6.11.1). The resistance Rk used in this experimental setup is 700 Ω. 12. Evaluate the expected maximal lifetime of this pacemaker using equation (6.11.2). Initial battery energy in this experimental setup is 0,5 Ah. Table 1. Um, V T, s, Hz, ms Ei, J t, years References 1. Irving P. Herman, Physics of the human body, Berlin; Heidelberg: Springer (2007). 2. Ruediger Kramme...[et al.], Springer handbook of medical technology, Berlin etc.: Springer, 1500 p., 2011, ISBN: B H Brown, R H Smallwood, D C Barber, P V Lawford and D R Hose, Medical Physics and Biomedical Engineering, New York N.Y.; London: Taylor and Francis (1999). 4. Arthur C. Guyton, John E. Hall, Textbook of medical physiology, Philadelphia Pa.: Saunders/Elsevier (2011). FBML TESTING A PACEMAKER 6

EXPERIMENT 5 Bioelectric Measurements

EXPERIMENT 5 Bioelectric Measurements Objectives EXPERIMENT 5 Bioelectric Measurements 1) Generate periodic signals with a Signal Generator and display on an Oscilloscope. 2) Investigate a Differential Amplifier to see small signals in a noisy

More information

Lab E5: Filters and Complex Impedance

Lab E5: Filters and Complex Impedance E5.1 Lab E5: Filters and Complex Impedance Note: It is strongly recommended that you complete lab E4: Capacitors and the RC Circuit before performing this experiment. Introduction Ohm s law, a well known

More information

Laboratory Activities Handbook

Laboratory Activities Handbook Laboratory Activities Handbook Answer Key 0 P a g e Contents Introduction... 2 Optical Heart Rate Monitor Overview... 2 Bare Board Preparation... 3 Light Indicator... 5 Low Pass Filter... 7 Amplifier...

More information

instead we hook it up to a potential difference of 60 V? instead we hook it up to a potential difference of 240 V?

instead we hook it up to a potential difference of 60 V? instead we hook it up to a potential difference of 240 V? Introduction In this lab we will examine the concepts of electric current and potential in a circuit. We first look at devices (like batteries) that are used to generate electrical energy that we can use

More information

Electrocardiogram (ECG)

Electrocardiogram (ECG) Vectors and ECG s Vectors and ECG s 2 Electrocardiogram (ECG) Depolarization wave passes through the heart and the electrical currents pass into surrounding tissues. Small part of the extracellular current

More information

CHAPTER 6. Motor Driver

CHAPTER 6. Motor Driver CHAPTER 6 Motor Driver In this lab, we will construct the circuitry that your robot uses to drive its motors. However, before testing the motor circuit we will begin by making sure that you are able to

More information

Ques on (2): [18 Marks] a) Draw the atrial synchronous Pacemaker block diagram and explain its operation. Benha University June 2013

Ques on (2): [18 Marks] a) Draw the atrial synchronous Pacemaker block diagram and explain its operation. Benha University June 2013 Benha University June 2013 Benha Faculty of Engineering Electrical Department Hospital Instrumentations (E472) 4 Th year (control) Dr.Waleed Abdel Aziz Salem Time: 3 Hrs Answer the following questions.

More information

*Notebook is excluded

*Notebook is excluded Biomedical Measurement Training System This equipment is designed for students to learn how to design specific measuring circuits and detect the basic physiological signals with practical operation. Moreover,

More information

Biomedical Instrumentation (BME420 ) Chapter 6: Biopotential Amplifiers John G. Webster 4 th Edition

Biomedical Instrumentation (BME420 ) Chapter 6: Biopotential Amplifiers John G. Webster 4 th Edition Biomedical Instrumentation (BME420 ) Chapter 6: Biopotential Amplifiers John G. Webster 4 th Edition Dr. Qasem Qananwah BME 420 Department of Biomedical Systems and Informatics Engineering 1 Biopotential

More information

Fundamentals of Pacing Therapy

Fundamentals of Pacing Therapy 1 4 7 2 5 8 3 6 9 Fundamentals of Pacing Therapy This presentation is provided with the understanding that the slide content must not be altered in any manner as the content is subject to FDA regulations.

More information

A Design Of Simple And Low Cost Heart Rate Monitor

A Design Of Simple And Low Cost Heart Rate Monitor A Design Of Simple And Low Cost Heart Rate Monitor 1 Arundhati Chattopadhyay, 2 Piyush Kumar, 3 Shashank Kumar Singh 1,2 UG Student, 3 Assistant Professor NSHM Knowledge Campus, Durgapur, India Abstract

More information

EXPERIMENT 7 The Amplifier

EXPERIMENT 7 The Amplifier Objectives EXPERIMENT 7 The Amplifier 1) Understand the operation of the differential amplifier. 2) Determine the gain of each side of the differential amplifier. 3) Determine the gain of the differential

More information

Laboratory 3 (drawn from lab text by Alciatore)

Laboratory 3 (drawn from lab text by Alciatore) Laboratory 3 (drawn from lab text by Alciatore) The Oscilloscope Required Components: 1 10 resistor 2 100 resistors 2 lk resistors 1 2k resistor 2 4.7M resistors 1 0.F capacitor 1 0.1 F capacitor 1 1.0uF

More information

* Notebook is excluded. Features KL-720 contains nine modules, including Electrocardiogram Measurement, E lectromyogram Measurement,

* Notebook is excluded. Features KL-720 contains nine modules, including Electrocardiogram Measurement, E lectromyogram Measurement, KL-720 Biomedical Measurement System Supplied by: 011 683 4365 This equipment is intended for students to learn how to design specific measuring circuits and detect the basic physiological signals with

More information

Lab 0: Orientation. 1 Introduction: Oscilloscope. Refer to Appendix E for photos of the apparatus

Lab 0: Orientation. 1 Introduction: Oscilloscope. Refer to Appendix E for photos of the apparatus Lab 0: Orientation Major Divison 1 Introduction: Oscilloscope Refer to Appendix E for photos of the apparatus Oscilloscopes are used extensively in the laboratory courses Physics 2211 and Physics 2212.

More information

Content. User Instruction - Pointoselect digital. Intended use...3 Safe usage of the device...3 Description...4

Content. User Instruction - Pointoselect digital. Intended use...3 Safe usage of the device...3 Description...4 2 Content Intended use...3 Safe usage of the device...3 Description...4 Operating the device Setting up the device...4 Starting the device...4 Mode selection...5 Measurement...5 Automatic...5 Manual...6

More information

EXPERIMENT 8 Bio-Electric Measurements

EXPERIMENT 8 Bio-Electric Measurements EXPERIMENT 8 Bio-Electric Measurements Objectives 1) Determine the amplitude of some electrical signals in the body. 2) Observe and measure the characteristics and amplitudes of muscle potentials due to

More information

Transcutaneous Energy Transmission Based Wireless Energy Transfer to Implantable Biomedical Devices

Transcutaneous Energy Transmission Based Wireless Energy Transfer to Implantable Biomedical Devices Transcutaneous Energy Transmission Based Wireless Energy Transfer to Implantable Biomedical Devices Anand Garg, Lakshmi Sridevi B.Tech, Dept. of Electronics and Instrumentation Engineering, SRM University

More information

The Tuned Circuit. Aim of the experiment. Circuit. Equipment and components. Display of a decaying oscillation. Dependence of L, C and R.

The Tuned Circuit. Aim of the experiment. Circuit. Equipment and components. Display of a decaying oscillation. Dependence of L, C and R. The Tuned Circuit Aim of the experiment Display of a decaying oscillation. Dependence of L, C and R. Circuit Equipment and components 1 Rastered socket panel 1 Resistor R 1 = 10 Ω, 1 Resistor R 2 = 1 kω

More information

Digital Clamp Meter Model: &

Digital Clamp Meter Model: & Digital Clamp Meter Model: 72-7224 & 72-7226 1 SAFETY INFORMATION Please read these instructions carefully before use and retain for future reference. This meter is designed to meet IEC61010-1, 61010-2-032,

More information

B. Equipment. Advanced Lab

B. Equipment. Advanced Lab Advanced Lab Measuring Periodic Signals Using a Digital Oscilloscope A. Introduction and Background We will use a digital oscilloscope to characterize several different periodic voltage signals. We will

More information

total j = BA, [1] = j [2] total

total j = BA, [1] = j [2] total Name: S.N.: Experiment 2 INDUCTANCE AND LR CIRCUITS SECTION: PARTNER: DATE: Objectives Estimate the inductance of the solenoid used for this experiment from the formula for a very long, thin, tightly wound

More information

BME 3113, Dept. of BME Lecture on Introduction to Biosignal Processing

BME 3113, Dept. of BME Lecture on Introduction to Biosignal Processing What is a signal? A signal is a varying quantity whose value can be measured and which conveys information. A signal can be simply defined as a function that conveys information. Signals are represented

More information

Bio-Potential Amplifiers

Bio-Potential Amplifiers Bio-Potential Amplifiers Biomedical Models for Diagnosis Body Signal Sensor Signal Processing Output Diagnosis Body signals and sensors were covered in EE470 The signal processing part is in EE471 Bio-Potential

More information

Biopotential Electrodes

Biopotential Electrodes Biomedical Instrumentation Prof. Dr. Nizamettin AYDIN naydin@yildiz.edu.tr naydin@ieee.org http://www.yildiz.edu.tr/~naydin Biopotential Electrodes 1 2 Electrode electrolyte interface The current crosses

More information

MPS Electrical Test Laboratory, Larsen & Toubro Limited, Mysore Campus, KIADB Industrial Area, Hebbal Hootagalli, Mysore, Karnataka

MPS Electrical Test Laboratory, Larsen & Toubro Limited, Mysore Campus, KIADB Industrial Area, Hebbal Hootagalli, Mysore, Karnataka Last Amended on - Page 1 of 6 SOURCE 1. DC VOLTAGE $ 1mV to 10 mv 0.13 % to 0.014% Using Multifunction Calibrator 10 mv to 1 V 0.014% to 0.0016% By Direct Method 1 v to 10 V 0.0016% to 0.0017% 10 V to

More information

Manufacturing Process - I Dr. D. K. Dwivedi Department of Mechanical and Industrial Engineering Indian Institute of Technology, Roorkee

Manufacturing Process - I Dr. D. K. Dwivedi Department of Mechanical and Industrial Engineering Indian Institute of Technology, Roorkee Manufacturing Process - I Dr. D. K. Dwivedi Department of Mechanical and Industrial Engineering Indian Institute of Technology, Roorkee Module - 3 Lecture - 5 Arc Welding Power Source Part 2 Welcome students.

More information

Exercise 12. Semiconductors EXERCISE OBJECTIVE DISCUSSION OUTLINE DISCUSSION. Introduction to semiconductors. The diode

Exercise 12. Semiconductors EXERCISE OBJECTIVE DISCUSSION OUTLINE DISCUSSION. Introduction to semiconductors. The diode Exercise 12 Semiconductors EXERCISE OBJECTIVE When you have completed this exercise, you will be familiar with the operation of a diode. You will learn how to use a diode to rectify ac voltage to produce

More information

EE ELECTRICAL ENGINEERING AND INSTRUMENTATION

EE ELECTRICAL ENGINEERING AND INSTRUMENTATION EE6352 - ELECTRICAL ENGINEERING AND INSTRUMENTATION UNIT V ANALOG AND DIGITAL INSTRUMENTS Digital Voltmeter (DVM) It is a device used for measuring the magnitude of DC voltages. AC voltages can be measured

More information

Switching Time and Conduction Voltage Drop

Switching Time and Conduction Voltage Drop Exercise 4-1 EXERCISE OBJECTIVES At the completion of this exercise, you will be able to describe the behavior of a bipolar transistor when it turns on or turns off, and during conduction. DISCUSSION To

More information

Test No. 2. Advanced Scope Measurements. History. University of Applied Sciences Hamburg. Last chance!! EEL2 No 2

Test No. 2. Advanced Scope Measurements. History. University of Applied Sciences Hamburg. Last chance!! EEL2 No 2 University of Applied Sciences Hamburg Group No : DEPARTMENT OF INFORMATION ENGINEERING Laboratory for Instrumentation and Measurement L1: in charge of the report Test No. 2 Date: Assistant A2: Professor:

More information

Conventional Paper-II-2011 Part-1A

Conventional Paper-II-2011 Part-1A Conventional Paper-II-2011 Part-1A 1(a) (b) (c) (d) (e) (f) (g) (h) The purpose of providing dummy coils in the armature of a DC machine is to: (A) Increase voltage induced (B) Decrease the armature resistance

More information

PHY152 Experiment 4: Oscillations in the RC-Circuits (Measurements with an oscilloscope)

PHY152 Experiment 4: Oscillations in the RC-Circuits (Measurements with an oscilloscope) PHY152 Experiment 4: Oscillations in the RC-Circuits (Measurements with an oscilloscope) If you have not used an oscilloscope before, the web site http://www.upscale.utoronto.ca/generalinterest/harrison/oscilloscope/oscilloscope.html

More information

10 Electromagnetic Interactions

10 Electromagnetic Interactions Lab 10 Electromagnetic Interactions What You Need To Know: The Physics Electricity and magnetism are intrinsically linked and not separate phenomena. A changing magnetic field can create an electric field

More information

3.003 Lab 3 Part A. Measurement of Speed of Light

3.003 Lab 3 Part A. Measurement of Speed of Light 3.003 Lab 3 Part A. Measurement of Speed of Light Objective: To measure the speed of light in free space Experimental Apparatus: Feb. 18, 2010 Due Mar. 2, 2010 Components: 1 Laser, 4 mirrors, 1 beam splitter

More information

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE. ISO-STIM 01D STIMULUS ISOLATION UNIT ±100 V / ±10 ma, bipolar output

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE. ISO-STIM 01D STIMULUS ISOLATION UNIT ±100 V / ±10 ma, bipolar output OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE ISO-STIM 01D STIMULUS ISOLATION UNIT ±100 V / ±10 ma, bipolar output VERSION 4.0 npi 2014 npi electronic GmbH, Bauhofring 16, D-71732 Tamm, Germany

More information

Used to overcome ventricular fibrillation may be due to coronary occlusion, shock, or abnormalities in blood chemistry

Used to overcome ventricular fibrillation may be due to coronary occlusion, shock, or abnormalities in blood chemistry Used to overcome ventricular fibrillation may be due to coronary occlusion, shock, or abnormalities in blood chemistry Main problem: heart muscle fibers are continuously stimulated by adjacent muscles

More information

Testers: / Ohmmeters Insulation Measuring Instruments. n Compact and rugged for service calls and laboratory use. Scope of delivery:

Testers: / Ohmmeters Insulation Measuring Instruments. n Compact and rugged for service calls and laboratory use. Scope of delivery: Table of Contents Testers: / Ohmmeters Insulation Measuring Instruments 51 METRAOHM 413 Article: M630A Digital Low-Resistance Measuring Instrument for Measuring Protective Conductors, Equipotential Bonding

More information

Biomedical Signal Processing and Applications

Biomedical Signal Processing and Applications Proceedings of the 2010 International Conference on Industrial Engineering and Operations Management Dhaka, Bangladesh, January 9 10, 2010 Biomedical Signal Processing and Applications Muhammad Ibn Ibrahimy

More information

Name Kyla Jackson, Todd Germeroth, Jake Spooler Date May 5, 2010 Lab 3E Group 3 Experiment Title Project Deliverable 3

Name Kyla Jackson, Todd Germeroth, Jake Spooler Date May 5, 2010 Lab 3E Group 3 Experiment Title Project Deliverable 3 Name Kyla Jackson, Todd Germeroth, Jake Spooler Date May 5, 2010 Lab 3E Group 3 Experiment Title Project Deliverable 3 Objective The objective of this project was to design and construct an ECG measurement

More information

Potential Risks of MRI in Device Patients

Potential Risks of MRI in Device Patients Outline Potential Risks of MRI in Device Patients Redha Boubertakh r.boubertakh@qmul.ac.uk MRI and cardiac implantable electronic devices (CIED) Components of an MRI scanner MRI implant and device safety

More information

OPERATING CHARACTERISTICS OF THE GEIGER COUNTER

OPERATING CHARACTERISTICS OF THE GEIGER COUNTER OPERATING CHARACTERISTICS OF THE GEIGER COUNTER OBJECTIVE The objective of this laboratory is to determine the operating voltage for a Geiger tube and to calculate the effect of the dead time and recovery

More information

Electronic Speed Controls and RC Motors

Electronic Speed Controls and RC Motors Electronic Speed Controls and RC Motors ESC Power Control Modern electronic speed controls regulate the electric power applied to an electric motor by rapidly switching the power on and off using power

More information

University of Jordan School of Engineering Electrical Engineering Department. EE 204 Electrical Engineering Lab

University of Jordan School of Engineering Electrical Engineering Department. EE 204 Electrical Engineering Lab University of Jordan School of Engineering Electrical Engineering Department EE 204 Electrical Engineering Lab EXPERIMENT 1 MEASUREMENT DEVICES Prepared by: Prof. Mohammed Hawa EXPERIMENT 1 MEASUREMENT

More information

Laboratory Exercise 6 THE OSCILLOSCOPE

Laboratory Exercise 6 THE OSCILLOSCOPE Introduction Laboratory Exercise 6 THE OSCILLOSCOPE The aim of this exercise is to introduce you to the oscilloscope (often just called a scope), the most versatile and ubiquitous laboratory measuring

More information

UCE-DSO212 DIGITAL OSCILLOSCOPE USER MANUAL. UCORE ELECTRONICS

UCE-DSO212 DIGITAL OSCILLOSCOPE USER MANUAL. UCORE ELECTRONICS UCE-DSO212 DIGITAL OSCILLOSCOPE USER MANUAL UCORE ELECTRONICS www.ucore-electronics.com 2017 Contents 1. Introduction... 2 2. Turn on or turn off... 3 3. Oscilloscope Mode... 4 3.1. Display Description...

More information

AS SUPER 4 digital. Elektrischer Nadelstimulator / Electrical needle stimulator. Art.-Nr

AS SUPER 4 digital. Elektrischer Nadelstimulator / Electrical needle stimulator. Art.-Nr AS SUPER 4 digital Elektrischer Nadelstimulator Electrical needle stimulator Art.-Nr. 200510 Gebrauchsanweisung Instruction Manual - Art.-Nr. 101477 20 Contents Foreword...21 Purpose for use...21 Safety

More information

SENSOR AND MEASUREMENT EXPERIMENTS

SENSOR AND MEASUREMENT EXPERIMENTS SENSOR AND MEASUREMENT EXPERIMENTS Page: 1 Contents 1. Capacitive sensors 2. Temperature measurements 3. Signal processing and data analysis using LabVIEW 4. Load measurements 5. Noise and noise reduction

More information

P a g e 1 ST985. TDR Cable Analyzer Instruction Manual. Analog Arts Inc.

P a g e 1 ST985. TDR Cable Analyzer Instruction Manual. Analog Arts Inc. P a g e 1 ST985 TDR Cable Analyzer Instruction Manual Analog Arts Inc. www.analogarts.com P a g e 2 Contents Software Installation... 4 Specifications... 4 Handling Precautions... 4 Operation Instruction...

More information

I. Introduction to Animal Sensitivity and Response

I. Introduction to Animal Sensitivity and Response I. Introduction to Animal Sensitivity and Response The term stray voltage has been used to describe a special case of voltage developed on the grounded neutral system of a farm. If this voltage reaches

More information

Biomedical Engineering Electrophysiology

Biomedical Engineering Electrophysiology Biomedical Engineering Electrophysiology Dr. rer. nat. Andreas Neubauer Sources of biological potentials and how to record them 1. How are signals transmitted along nerves? Transmit velocity Direction

More information

Elizabethtown College Department of Physics and Engineering PHY104. Lab # 9- Oscilloscope and RC Circuit

Elizabethtown College Department of Physics and Engineering PHY104. Lab # 9- Oscilloscope and RC Circuit Elizabethtown College Department of Physics and Engineering PHY104 Lab # 9- Oscilloscope and RC Circuit Introduction This lab introduces you to very important tools, the oscilloscope and the waveform generator.

More information

electrical noise and interference, environmental changes, instrument resolution, or uncertainties in the measurement process itself.

electrical noise and interference, environmental changes, instrument resolution, or uncertainties in the measurement process itself. MUST 382 / EELE 491 Spring 2014 Basic Lab Equipment and Measurements Electrical laboratory work depends upon various devices to supply power to a circuit, to generate controlled input signals, and for

More information

Ph 3455 The Franck-Hertz Experiment

Ph 3455 The Franck-Hertz Experiment Ph 3455 The Franck-Hertz Experiment Required background reading Tipler, Llewellyn, section 4-5 Prelab Questions 1. In this experiment, we will be using neon rather than mercury as described in the textbook.

More information

Introduction to Medical Electronics Industry Test Analysis and Solution

Introduction to Medical Electronics Industry Test Analysis and Solution Background and development status of the medical electronics industry Background Introduction to Medical Electronics Industry Test Analysis and Solution As the global population ages, increasing health

More information

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, Unit 8 The Centre Holywell Business Park Northfield Road Southam Warwickshire CV47 0FP Contact: Mr N Morgan Tel: +44 (0)1926 812066 Fax: +44

More information

Probes and Electrodes Dr. Lynn Fuller Webpage:

Probes and Electrodes Dr. Lynn Fuller Webpage: ROCHESTER INSTITUTE OF TECHNOLOGY MICROELECTRONIC ENGINEERING Probes and Electrodes Dr. Lynn Fuller Webpage: http://people.rit.edu/lffeee 82 Lomb Memorial Drive Rochester, NY 14623-5604 Tel (585) 475-2035

More information

ECE 53A: Fundamentals of Electrical Engineering I

ECE 53A: Fundamentals of Electrical Engineering I ECE 53A: Fundamentals of Electrical Engineering I Laboratory Assignment #1: Instrument Operation, Basic Resistor Measurements and Kirchhoff s Laws Fall 2007 General Guidelines: - Record data and observations

More information

SPARK OF LIFE. How does your body react to electricity?

SPARK OF LIFE. How does your body react to electricity? SPARK OF LIFE How does your body react to electricity? WHO WAS FRANKENSTEIN? What do you know about Victor Frankenstein and his creature? Victor Frankenstein and the monster he created were invented 200

More information

Lab 1: Basic Lab Equipment and Measurements

Lab 1: Basic Lab Equipment and Measurements Abstract: Lab 1: Basic Lab Equipment and Measurements This lab exercise introduces the basic measurement instruments that will be used throughout the course. These instruments include multimeters, oscilloscopes,

More information

Understanding & Using The HA2500's Sub Drives

Understanding & Using The HA2500's Sub Drives Understanding & Using The HA2500's Sub Drives When horizontal drive to the horizontal output stage is missing, expensive horizontal output stage components cannot be determined good or bad. If horizontal

More information

I p = V s = N s I s V p N p

I p = V s = N s I s V p N p UNIT G485 Module 1 5.1.3 Electromagnetism 11 For an IDEAL transformer : electrical power input = electrical power output to the primary coil from the secondary coil Primary current x primary voltage =

More information

A Device for the Recording of Isolated Muscle Cell Contractions Using Silicone Tensometer

A Device for the Recording of Isolated Muscle Cell Contractions Using Silicone Tensometer Gen. Physiol. Biophys. (1986), 5, 567 572 567 Short communication A Device for the Recording of Isolated Muscle Cell Contractions Using Silicone Tensometer M. MARKO, Ľ. LACINOVÁ and J. POLEDNA Centre of

More information

Surge Generator for MIL-STD 1275

Surge Generator for MIL-STD 1275 Surge Generator for MIL-STD 1275 This generator PG1275 is specially designed for the test of the susceptibility to surges and spikes of military 28 Vdc electric circuits. The equipment allows performing

More information

Study of DBD electrostatic precipitator under different high voltage waveforms

Study of DBD electrostatic precipitator under different high voltage waveforms Study of DBD electrostatic precipitator under different high voltage waveforms R. Gouri Department of Electrical Engineering, University of Béchar, 8, Béchar, Algeria r.gouri@gmail.com N. Zouzou, E. Moreau,

More information

LAB I. INTRODUCTION TO LAB EQUIPMENT

LAB I. INTRODUCTION TO LAB EQUIPMENT 1. OBJECTIVE LAB I. INTRODUCTION TO LAB EQUIPMENT In this lab you will learn how to properly operate the oscilloscope Agilent MSO6032A, the Keithley Source Measure Unit (SMU) 2430, the function generator

More information

Pointoselect Digital DT

Pointoselect Digital DT Pointoselect Digital DT Digital Point Finder for Auricular and Body Acupuncture With Integrated PuTENS Stimulation Item no. 200506 Instruction Manual Item no. 100858 Content 1. Intended Use... 3 2. Safety

More information

Test No. 1. Introduction to Scope Measurements. Report History. University of Applied Sciences Hamburg. Last chance!! EEL2 No 1

Test No. 1. Introduction to Scope Measurements. Report History. University of Applied Sciences Hamburg. Last chance!! EEL2 No 1 University of Applied Sciences Hamburg Group No : DEPARTMENT OF INFORMATION ENGINEERING Laboratory for Instrumentation and Measurement L: in charge of the report Test No. Date: Assistant A2: Professor:

More information

Existing and Emerging Opportunities in Printed Electronics For Printers

Existing and Emerging Opportunities in Printed Electronics For Printers Existing and Emerging Opportunities in Printed Electronics For Printers Don Banfield Conductive Compounds, Inc. Hudson, New Hampshire, USA www.conductivecompounds.com Presentation Outline Summary of some

More information

Pre-Laboratory Assignment

Pre-Laboratory Assignment Measurement of Electrical Resistance and Ohm's Law PreLaboratory Assignment Read carefully the entire description of the laboratory and answer the following questions based upon the material contained

More information

OPTICAL EMISSION CHARACTERISTICS OF HELIUM BREAKDOWN AT PARTIAL VACUUM FOR POINT TO PLANE GEOMETRY

OPTICAL EMISSION CHARACTERISTICS OF HELIUM BREAKDOWN AT PARTIAL VACUUM FOR POINT TO PLANE GEOMETRY OPTICAL EMISSION CHARACTERISTICS OF HELIUM BREAKDOWN AT PARTIAL VACUUM FOR POINT TO PLANE GEOMETRY K. Koppisetty ξ, H. Kirkici 1, D. L. Schweickart 2 1 Auburn University, Auburn, Alabama 36849, USA, 2

More information

Extended analysis versus frequency of partial discharges phenomena, in support of quality assessment of insulating systems

Extended analysis versus frequency of partial discharges phenomena, in support of quality assessment of insulating systems Extended analysis versus frequency of partial discharges phenomena, in support of quality assessment of insulating systems Romeo C. Ciobanu, Cristina Schreiner, Ramona Burlacu, Cristina Bratescu Technical

More information

EECE Circuits and Signals: Biomedical Applications. Lab ECG I The Instrumentation Amplifier

EECE Circuits and Signals: Biomedical Applications. Lab ECG I The Instrumentation Amplifier EECE 150 - Circuits and Signals: Biomedical Applications Lab ECG I The Instrumentation Amplifier Introduction: As discussed in class, instrumentation amplifiers are often used to reject common-mode signals

More information

USER MANUAL 600A AC Clamp Meter + NCV Model MA610

USER MANUAL 600A AC Clamp Meter + NCV Model MA610 USER MANUAL 600A AC Clamp Meter + NCV Model MA610 Additional User Manual Translations available at www.extech.com Introduction Thank you for selecting the Extech MA610 Clamp Meter. This meter measures

More information

1. The coulomb is a unit of. A. charge B. voltage C. energy D. capacitance E. current. 2. The following is not true about voltage:

1. The coulomb is a unit of. A. charge B. voltage C. energy D. capacitance E. current. 2. The following is not true about voltage: BioE 1310 - Review 1 - DC 1/16/2017 Instructions: On the Answer Sheet, enter your 2-digit ID number (with a leading 0 if needed) in the boxes of the ID section. Fill in the corresponding numbered circles.

More information

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, Unit 8 The Centre Holywell Business Park Northfield Road Southam Warwickshire CV47 0FP Contact: Mr N Morgan Tel: +44 (0)1926 812066 Fax: +44

More information

Current Probes. User Manual

Current Probes. User Manual Current Probes User Manual ETS-Lindgren Inc. reserves the right to make changes to any product described herein in order to improve function, design, or for any other reason. Nothing contained herein shall

More information

Generation of Sub-nanosecond Pulses

Generation of Sub-nanosecond Pulses Chapter - 6 Generation of Sub-nanosecond Pulses 6.1 Introduction principle of peaking circuit In certain applications like high power microwaves (HPM), pulsed laser drivers, etc., very fast rise times

More information

Multifunction Digital

Multifunction Digital MS2009A Multifunction Digital Clamp Meter User Manual 200/600 OFF 2/20 NCV SEL MAX V RAN HOLD OFF MS2009A AC CLAMP METER AUTO MAX C F kmω μmva CONTENTS Safety requirements...1 Safety signs...1 Notes...1

More information

Lab 1: Pulse Propagation and Dispersion

Lab 1: Pulse Propagation and Dispersion ab 1: Pulse Propagation and Dispersion NAME NAME NAME Introduction: In this experiment you will observe reflection and transmission of incident pulses as they propagate down a coaxial transmission line

More information

LAB II. INTRODUCTION TO LAB EQUIPMENT

LAB II. INTRODUCTION TO LAB EQUIPMENT 1. OBJECTIVE LAB II. INTRODUCTION TO LAB EQUIPMENT In this lab you will learn how to properly operate the oscilloscope Keysight DSOX1102A, the Keithley Source Measure Unit (SMU) 2430, the function generator

More information

Current sensor by IZM

Current sensor by IZM Current sensor by IZM TYPICAL APPLICATIONS Current measurement in commutation cell Monitoring of switching behavior of Si, SiC, GaN, or similar semiconductors Measuring of current pulses Analysis of power

More information

Fluke 192/196/199. MS 190 and MA 190. Users Manual Supplement

Fluke 192/196/199. MS 190 and MA 190. Users Manual Supplement Fluke 192/196/199 MS 190 and MA 190 Users Manual Supplement 4822 872 00979 April 2000, Rev.2, 9/00 2000 Fluke Corporation. All rights reserved. Printed in the Netherlands. All product names are trademarks

More information

Interface Electronic Circuits

Interface Electronic Circuits Lecture (5) Interface Electronic Circuits Part: 1 Prof. Kasim M. Al-Aubidy Philadelphia University-Jordan AMSS-MSc Prof. Kasim Al-Aubidy 1 Interface Circuits: An interface circuit is a signal conditioning

More information

THE AMPLIFIER. A-B = C subtractor. INPUTS Figure 1

THE AMPLIFIER. A-B = C subtractor. INPUTS Figure 1 OBJECTIVES: THE AMPLIFIER 1) Explain the operation of the differential amplifier. 2) Determine the gain of each side of the differential amplifier. 3) Determine the gain of the differential amplifier as

More information

EXPERIMENT 1 PRELIMINARY MATERIAL

EXPERIMENT 1 PRELIMINARY MATERIAL EXPERIMENT 1 PRELIMINARY MATERIAL BREADBOARD A solderless breadboard, like the basic model in Figure 1, consists of a series of square holes, and those columns of holes are connected to each other via

More information

IMPROVEMENTS IN ELECTROCARDIOGRAPHY SMOOTHENING AND AMPLIFICATION

IMPROVEMENTS IN ELECTROCARDIOGRAPHY SMOOTHENING AND AMPLIFICATION IMPROVEMENTS IN ELECTROCARDIOGRAPHY SMOOTHENING AND AMPLIFICATION Manan Joshi, Sarosh Patel, Dr. Lawrence Hmurcik Electrical Engineering Department University of Bridgeport Bridgeport, CT 06604 Abstract

More information

QUASAR PP & PPR RECTIFIERS

QUASAR PP & PPR RECTIFIERS 1 9 is a line of switch-mode rectifiers for surface treatment processes, electro-winning and water treatment, that adopts pulse width modulation (PWM) technique for the controlling of current amplitude.

More information

Air Coupled Ultrasonic Inspection of Steel Rubber Interface

Air Coupled Ultrasonic Inspection of Steel Rubber Interface Air Coupled Ultrasonic Inspection of Steel Rubber Interface More Info at Open Access Database www.ndt.net/?id=15204 Bikash Ghose 1, a, Krishnan Balasubramaniam 2, b 1 High Energy Materials Research Laboratory,

More information

EE 210: CIRCUITS AND DEVICES

EE 210: CIRCUITS AND DEVICES EE 210: CIRCUITS AND DEVICES LAB #3: VOLTAGE AND CURRENT MEASUREMENTS This lab features a tutorial on the instrumentation that you will be using throughout the semester. More specifically, you will see

More information

Making sense of electrical signals

Making sense of electrical signals Making sense of electrical signals Our thanks to Fluke for allowing us to reprint the following. vertical (Y) access represents the voltage measurement and the horizontal (X) axis represents time. Most

More information

The Heart Rate Exercise sensor can be connected to the all einstein Tablets, einstein LabMate, and einstein LabMate+.

The Heart Rate Exercise sensor can be connected to the all einstein Tablets, einstein LabMate, and einstein LabMate+. Understanding how the heart works is basic to biology studies and is one of the first experiments any science student should learn to perform. The Heart Rate Exercise sensor bundle includes a Polar belt,

More information

EDL Group #3 Final Report - Surface Electromyograph System

EDL Group #3 Final Report - Surface Electromyograph System EDL Group #3 Final Report - Surface Electromyograph System Group Members: Aakash Patil (07D07021), Jay Parikh (07D07019) INTRODUCTION The EMG signal measures electrical currents generated in muscles during

More information

Kentech Instruments Ltd. CPS3/S pulser Serial No. J03*****

Kentech Instruments Ltd. CPS3/S pulser Serial No. J03***** Notes on the use of Kentech Instruments Ltd. CPS3/S pulser Serial No. J03***** 17th April 2003 Kentech Instruments Ltd.,Unit 9,Hall Farm Workshops, South Moreton,Oxon OX11 9AG, UK UK Tel: 01235 510748

More information

EE 230 Experiment 10 ECG Measurements Spring 2010

EE 230 Experiment 10 ECG Measurements Spring 2010 EE 230 Experiment 10 ECG Measurements Spring 2010 Note: If for any reason the students are uncomfortable with doing this experiment, please talk to the instructor for the course and an alternative experiment

More information

Implementation of wireless ECG measurement system in ubiquitous health-care environment

Implementation of wireless ECG measurement system in ubiquitous health-care environment Implementation of wireless ECG measurement system in ubiquitous health-care environment M. C. KIM 1, J. Y. YOO 1, S. Y. YE 2, D. K. JUNG 3, J. H. RO 4, G. R. JEON 4 1 Department of Interdisciplinary Program

More information

Telemetrie-Messtechnik Schnorrenberg

Telemetrie-Messtechnik Schnorrenberg Telemetrie-Messtechnik Schnorrenberg MTP-IND-PWR User Manual Inductive power supply set Power supply for power head 25 and 50mm mounting tape to fix coil on shaft Ferrite tape 30mmx3m CUL 1.00 mm (Enamelled

More information

Experiment 2 Determining the Capacitive Reactance of a Capacitor in an AC Circuit

Experiment 2 Determining the Capacitive Reactance of a Capacitor in an AC Circuit Experiment 2 Determining the apacitive eactance of a apacitor in an A ircuit - Objects of the experiments: a- Investigating the voltage and the current at a capacitor in an A circuit b- Observing the phase

More information

Automotive Scope/GMM/OBD II Code Reader

Automotive Scope/GMM/OBD II Code Reader Finest 1006 Automotive Scope/GMM/OBD II Code Reader Automotive On-board diagnostic systems of thesedays have gone a long way toward helping you locate a problem. But those systems aren't perfect and they

More information

PULSED BREAKDOWN CHARACTERISTICS OF HELIUM IN PARTIAL VACUUM IN KHZ RANGE

PULSED BREAKDOWN CHARACTERISTICS OF HELIUM IN PARTIAL VACUUM IN KHZ RANGE PULSED BREAKDOWN CHARACTERISTICS OF HELIUM IN PARTIAL VACUUM IN KHZ RANGE K. Koppisetty ξ, H. Kirkici Auburn University, Auburn, Auburn, AL, USA D. L. Schweickart Air Force Research Laboratory, Wright

More information