Biomedical Instrumentation (BME420 ) Chapter 6: Biopotential Amplifiers John G. Webster 4 th Edition

Size: px
Start display at page:

Download "Biomedical Instrumentation (BME420 ) Chapter 6: Biopotential Amplifiers John G. Webster 4 th Edition"

Transcription

1 Biomedical Instrumentation (BME420 ) Chapter 6: Biopotential Amplifiers John G. Webster 4 th Edition Dr. Qasem Qananwah BME 420 Department of Biomedical Systems and Informatics Engineering 1 Biopotential Amplifiers Basic function to increase the amplitude of a weak electric signal of biological origin typically process voltages but in some cases also process currents Typical bio-amp requirements high input impedance -greater than 10 MΩ safety: protect the organism being studied careful design to prevent macro and microshocks isolation and protection circuitry to limit the current through the electrode to safe level output impedance of the amplifier should be low to drive any external load with minimal distortion gain greater than 1000 biopotentials are typically less than a millivolt most biopotential amplifiers are differential signals are recorded using a bipolar electrodes which are symmetrically located high common mode rejection ratio biopotentials ride on a large offset signals rapid calibration of the amplifier in laboratory conditions BME 420 Department of Biomedical Systems and Informatics Engineering 2 1

2 Figure 6.16: Voltage and frequency ranges of some common biopotential signals; dc potentials include intracellular voltages as well as voltages measured from several points on the body. EOG is the electro-oculogram, EEG is the electroencephalogram, ECG is the electrocardiogram, EMG is the electromyogram, and AAP is the axon action potential. (From J. M. R. Delgado, "Electrodes for Extracellular Recording and Stimulation," in Physical Techniques in Biological Research, edited by W. L. Nastuk, New York: Academic Press, 1964) BME 420 Department of Biomedical Systems and Informatics Engineering 3 Electrocardiogram The electric dipole, consists of two equal and opposite charges, separated by some (usually small) distance The potential differences arising in the heart (cardiac dipoles) can be represented by electrical vectors Amplitude and direction All basic vector operations can be applied to the cardiac vectors Each depolarizing myocardial cell is in fact a dipole and thus can be represented by a vector = elementary vector The sum of all elementary vectors will create an instantaneous vector the potential differences generated by the heart change from moment to moment during the cardiac cycle Once a single cell is stimulated the depolarization will propagate in every direction: a propagating wave of depolarization will be created Each of these moments can be described by an instantaneous vector (with a different size and orientation) All these vectors can be brought to a single common point: electrical center of the heart BME 420 Department of Biomedical Systems and Informatics Engineering 4 2

3 Figure 6.1: Rough sketch of the dipole field of the heart when the R wave is maximal. The dipole consists of the points of equal positive and negative charge separated from one another and denoted by the dipole moment vector M. BME 420 Department of Biomedical Systems and Informatics Engineering 5 By recording the magnitude and direction of the electrical forces that are generated by the heart by means of a continuous series of vectors that form curving lines around a central point one can record the vectorcardiography The projection of this curve as function of time on an axis corresponding to a lead is actually the ECG in that particular lead M a Cardiac vector Lead vector Figure 6.2: Relationships between the two lead vectors a 1 and a 2 and the cardiac vector M. The component of M in the direction of a 1 is given by the dot product of these two vectors and denoted on the figure by v a1. Lead vector a 2 is perpendicular to the cardiac vector, so no voltage component is seen in this lead. BME 420 Department of Biomedical Systems and Informatics Engineering 6 3

4 BME 420 Department of Biomedical Systems and Informatics Engineering 7 Electrocardiogram Electrodes for recording the potential changes of the heart are placed on the body surface in a standard way Potential changes recorded by specifically connected electrodes is called a lead(a pair of electrodes, or combination of several electrodes through a resistive network that gives an equivalent pair, is referred to as a lead) Each lead will be assigned with an axis and each of the axes will have an orientation The projection of the cardiac vectors as function of time on the axis corresponding to a lead is actually the ECG trace in that particular lead BME 420 Department of Biomedical Systems and Informatics Engineering 8 4

5 Electrocardiogram 1. Two different points on the body (bipolar leads) 2. One point on the body and a virtual reference point with zero electrical potential, located in the center of the heart (unipolar leads) The standard ECG has 12 leads: 1. 3 Standard Limb Leads 2. 3 Augmented Limb Leads 3. 6 Precordial Leads The axis of a particular lead represents the viewpoint from which it looks at the heart. Limb leads Precordial leads Bipolar Lead I, II, III (Standard limb leads) - Unipolar avr, avl, avf (augmented limb leads) V 1 -V 6 BME 420 Department of Biomedical Systems and Informatics Engineering 9 Figure 6.3: Cardiologists use a standard notation such that the direction of the lead vector for lead I is 0, that of lead II is 60, and that of lead III is 120. An example of a cardiac vector at 30 with its scalar components seen for each lead is shown. BME 420 Department of Biomedical Systems and Informatics Engineering 10 5

6 Generation of ECG signal in Einthoven limb leads BME 420 Department of Biomedical Systems and Informatics Engineering 11 Generation of ECG signal in Einthoven limb leads BME 420 Department of Biomedical Systems and Informatics Engineering 12 6

7 BME 420 Department of Biomedical Systems and Informatics Engineering 13 In clinical electrocardiography Electrocardiograph Leads more than one lead must be recorded to describe the heart's electric activity fully several leads are taken in the frontal plane and the transverse plane 1. frontal plane: parallel to the back when lying 2. transverse plane: parallel to the ground when standing standing Frontal plane lead placement (called Eindhoven s triangle) Additional leads Additional leads unipolar measurements potential measured at electrodes with reference to a reference; average of the 2 electrodes Wilson central terminal three limb electrodes connected through equal-valued resistors to a common node augmented leads some nodes disconnected increase the amplitude of measurement using Figure 6.4: Connection of electrodes to the body to obtain Wilson s central terminal BME 420 Department of Biomedical Systems and Informatics Engineering 14 7

8 Example 6.1: Show that the voltage in lead avr is %50 greater than that in lead VR at the same instant. Solution: Considering the connections for avr and VR, we can draw the equivalent circuits. The voltages between each limb and ground v a, v b, v c Figure E6.1 (a) avr (b) VR (c) Simplified circuit of (a) BME 420 Department of Biomedical Systems and Informatics Engineering 15 R should be very high (~5 MΩ) or buffers (voltage followers) are used between each electrode so that the loading of any particular lead will be minimal. Figure 6.5: (a), (b), (c) Connections of electrodes for the three augmented limb leads, (d) Vector diagram showing standard and augmented lead-vector directions in the frontal plane. BME 420 Department of Biomedical Systems and Informatics Engineering 16 8

9 Figure 6.6 (a) Positions of precordial leads on the chest wall, (b) Directions of precordial lead vectors in the transverse plane. BME 420 Department of Biomedical Systems and Informatics Engineering 17 6 limb leads define electrical activity in frontal plane 6 precordial leads define electrical activity in transverse plane The 3 augmented leads compare one limb electrode to the average of the other two. (avr, avl, avf). Leads are made of a combination of electrodes that form imaginary lines in the body along which the electrical signals are measured. BME 420 Department of Biomedical Systems and Informatics Engineering 18 9

10 Figure 6.7: Block diagram of an electrocardiograph BME 420 Department of Biomedical Systems and Informatics Engineering 19 Problems Frequently Encountered 1. Frequency Distortion a. High frequency distortion if band pass filter has 0 25 Hz response b. Low frequency distortion if band pass filter has Hz response In low frequency distortion, the baseline is no longer horizontal, especially immediately following any event in the tracing. Monophasic waves appear to be biphasic. BME 420 Department of Biomedical Systems and Informatics Engineering 20 10

11 Problems Frequently Encountered 2. Saturation or Cutoff Distortion The peaks of QRS can be cutoff because the output of the amplifier cannot exceed the saturation voltage or even the lower portion of the S wave will be cutoff. 3. Ground Loop BME 420 Department of Biomedical Systems and Informatics Engineering 21 Problems Frequently Encountered 4. Open Lead Wire Always make sure that the electrodes are in good contact. Otherwise, usually high potentials can often be induced in the open wire as a result of electric field. Coming from a power line, it causes a wideconstant-amplitude deflection on the ECG trace. BME 420 Department of Biomedical Systems and Informatics Engineering 22 11

12 Problems Frequently Encountered 5. Artifacts From Large Electric Transients Occurred due to cardiac defibrillation Figure 6.8 Effect of a voltage transient on an ECG recorded on an electrocardiograph in which the transient causes the amplifier to saturate, and a finite period of time is required for the charge to bleed off enough to bring the ECG back into the amplifier's active region of operation. This is followed by a first-order recovery of the system. BME 420 Department of Biomedical Systems and Informatics Engineering Interference from Electric Devices Figure 6.9 (a) 60 Hz power-line interference, (b) Electromyographic interference on the ECG. BME 420 Department of Biomedical Systems and Informatics Engineering 24 12

13 6. Interference from Electric Devices For a 9 m cable, i d 6 na, the electrodes impedance can differ as much as 20kΩ. Then, This can be minimized by shielding the leads. Figure 6.10 A mechanism of electric-field pickup of an electrocardiograph resulting from the power line. Coupling capacitance between the hot side of the power line and lead wires causes current to flow through skin electrode impedances on its way to ground. BME 420 Department of Biomedical Systems and Informatics Engineering Interference from Electric Devices Look at this situation. This is noticeable on the ECG and very objectionable on EEG. Figure 6.11 Current flows from the power line through the body and ground impedance, thus creating a common-mode voltage everywhere on the body. Zin is not only resistive but, as a result of RF bypass capacitors at the amplifier input, has a reactive component as well. BME 420 Department of Biomedical Systems and Informatics Engineering 26 13

14 6. Interference from Electric Devices Figure 6.12: Magnetic-field pickup by the electrocardiograph (a) Lead wires for lead I make a closed loop (shaded area) when patient and electrocardiograph are considered in the circuit. The change in magnetic field passing through this area induces a current in the loop, (b) This effect can be minimized by twisting the lead wires together and keeping them close to the body in order to subtend a much smaller area. BME 420 Department of Biomedical Systems and Informatics Engineering 27 Transient Protection Figure 6.13: A voltageprotection scheme at the input of an electrocardiograph to protect the machine from highvoltage transients. Circuit elements connected across limb leads on left-hand side are voltage-limiting devices. Figure 6.14: Voltage-limiting devices (a) Current voltage characteristics of a voltagelimiting device. (b) Parallel silicon-diode voltage-limiting circuit. (c) Back-to-back silicon zener-diode voltage-limiting circuit. (d) Gas-discharge tube (neon light) voltage-limiting circuit element. 28 BME 420 Department of Biomedical Systems and Informatics Engineering 28 14

15 BME 420 Department of Biomedical Systems and Informatics Engineering 29 BME 420 Department of Biomedical Systems and Informatics Engineering 30 15

16 BME 420 Department of Biomedical Systems and Informatics Engineering 31 Figure 6.15 Driven-right-leg circuit for minimizing common-mode interference. The circuit derives common-mode voltage from a pair of averaging resistors connected to v3 and v4 in Figure 3.5. The right leg is not grounded but is connected to output of the auxiliary op amp. BME 420 Department of Biomedical Systems and Informatics Engineering 32 16

17 Figure E6.3 Equivalent circuit of driven-right-leg system of Figure BME 420 Department of Biomedical Systems and Informatics Engineering 33 Figure 6.18: This ECG amplifier has a gain of 25 in the dc-coupled stages. The high-pass filter feeds a noninvertingamplifier stage that has a gain of 32. The total gain is 25 X 32 = 800. When ma 776 op amps were used, the circuit was found to have a CMRR of 86 db at 100 Hz and a noise level of 40 mv peak to peak at the output. The frequency response was 0.04 to 150 Hz for ±3 db and was flat over 4 to 40 Hz. A single op amp chip, the LM 324, that contains four individual op amps could also be used in this circuit reducing the total parts count. BME 420 Department of Biomedical Systems and Informatics Engineering 34 17

18 Cardiotachometer BME 420 Department of Biomedical Systems and Informatics Engineering 35 Beat- to-beat Cardiotachometer 36 BME 420 Department of Biomedical Systems and Informatics Engineering 36 18

19 37 BME 420 Department of Biomedical Systems and Informatics Engineering 37 Figure 6.22 Typical fetal ECG obtained from the maternal abdomen. F represents fetal QRS complexes; M represents maternal QRS complexes. Maternal ECG and fetal ECG (recorded directly from the fetus) are included for comparison. (From "Monitoring of Intrapartum Phenomena," by J. F. Roux, M. R. Neuman, and R. C. Goodlin, in CRC Critical Reviews in Bioengineering, 2, pp , January 1975, CRC Press. Used by permission of CRC Press, Inc.) BME 420 Department of Biomedical Systems and Informatics Engineering 38 19

20 Figure 6.23 Block diagram of a scheme for isolating fetal ECG from an abdominal signal that contains both fetal and maternal ECGs. (From "Monitoring of Intrapartum Phenomena," by J. F. Roux, M. R. Neuman, and R. C. Goodlin, in CRC Critical Reviews in Bioengineering, 2, pp , January 1975, CRC Press. Used by permission of CRC Press, Inc.) BME 420 Department of Biomedical Systems and Informatics Engineering 39 BME 420 Department of Biomedical Systems and Informatics Engineering 40 20

21 Figure 6.24 Block diagram of a cardiac monitor. BME 420 Department of Biomedical Systems and Informatics Engineering 41 BME 420 Department of Biomedical Systems and Informatics Engineering 42 21

22 Figure 6.25 Block diagram of a system used with cardiac monitors to detect increased electrode impedance, lead wire failure, or electrode fall-off. BME 420 Department of Biomedical Systems and Informatics Engineering 43 BME 420 Department of Biomedical Systems and Informatics Engineering 44 22

23 BME 420 Department of Biomedical Systems and Informatics Engineering 45 23

Electrocardiogram (ECG)

Electrocardiogram (ECG) Vectors and ECG s Vectors and ECG s 2 Electrocardiogram (ECG) Depolarization wave passes through the heart and the electrical currents pass into surrounding tissues. Small part of the extracellular current

More information

Bio-Potential Amplifiers

Bio-Potential Amplifiers Bio-Potential Amplifiers Biomedical Models for Diagnosis Body Signal Sensor Signal Processing Output Diagnosis Body signals and sensors were covered in EE470 The signal processing part is in EE471 Bio-Potential

More information

Lecture 4 Biopotential Amplifiers

Lecture 4 Biopotential Amplifiers Bioinstrument Sahand University of Technology Lecture 4 Biopotential Amplifiers Dr. Shamekhi Summer 2016 OpAmp and Rules 1- A = (gain is infinity) 2- Vo = 0, when v1 = v2 (no offset voltage) 3- Rd = (input

More information

Amplificador de Biopotencial

Amplificador de Biopotencial Amplificador de Biopotencial Prof. Sérgio F. Pichorim Baseado no cap 6 do Webster e cap 17 do Kutz & Towe From J. G. Webster (ed.), Medical instrumentation: application and design. 3 rd ed. New York: John

More information

Biomedical. Measurement and Design ELEC4623. Lectures 9 and 10 Practical biopotential amplifier design and multilead ECG systems

Biomedical. Measurement and Design ELEC4623. Lectures 9 and 10 Practical biopotential amplifier design and multilead ECG systems Biomedical Instrumentation, Measurement and Design ELEC4623 Lectures 9 and 10 Practical biopotential amplifier design and multilead ECG systems Feedback and stability A negative feedback system with closed

More information

BIOMEDICAL INSTRUMENTATION PROBLEM SHEET 1

BIOMEDICAL INSTRUMENTATION PROBLEM SHEET 1 BIOMEDICAL INSTRUMENTATION PROBLEM SHEET 1 Dr. Gari Clifford Hilary Term 2013 1. (Exemplar Finals Question) a) List the five vital signs which are most commonly recorded from patient monitors in high-risk

More information

GBM8320 Dispositifs Médicaux Intelligents

GBM8320 Dispositifs Médicaux Intelligents GBM8320 Dispositifs Médicaux Intelligents Biopotential amplifiers Part 1 Mohamad Sawan et al. Laboratoire de neurotechnologies Polystim http://www.cours.polymtl.ca/gbm8320/ mohamad.sawan@polymtl.ca M5418

More information

Kanchan S. Shrikhande. Department of Instrumentation Engineering, Vivekanand Education Society s Institute of.

Kanchan S. Shrikhande. Department of Instrumentation Engineering, Vivekanand Education Society s Institute of. ISOLATED ECG AMPLIFIER WITH RIGHT LEG DRIVE Kanchan S. Shrikhande Department of Instrumentation Engineering, Vivekanand Education Society s Institute of Technology(VESIT),kanchans90@gmail.com Abstract

More information

Biomedical Sensor Systems Laboratory. Institute for Neural Engineering Graz University of Technology

Biomedical Sensor Systems Laboratory. Institute for Neural Engineering Graz University of Technology Biomedical Sensor Systems Laboratory Institute for Neural Engineering Graz University of Technology 2017 Bioinstrumentation Measurement of physiological variables Invasive or non-invasive Minimize disturbance

More information

ECE 480 Design Team 6 Electrocardiography and Design

ECE 480 Design Team 6 Electrocardiography and Design ECE 480 Design Team 6 Electrocardiography and Design Alex Volinski November 16 th, 2012 Executive Summary Recently there has been a large increase in consumer demand for a new and functional ECG (Electrocardiograph)

More information

BME 405 BIOMEDICAL ENGINEERING SENIOR DESIGN 1 Fall 2005 BME Design Mini-Project Project Title

BME 405 BIOMEDICAL ENGINEERING SENIOR DESIGN 1 Fall 2005 BME Design Mini-Project Project Title BME 405 BIOMEDICAL ENGINEERING SENIOR DESIGN 1 Fall 2005 BME Design Mini-Project Project Title Basic system for Electrocardiography Customer/Clinical need A recent health care analysis have demonstrated

More information

Special-Purpose Operational Amplifier Circuits

Special-Purpose Operational Amplifier Circuits Special-Purpose Operational Amplifier Circuits Instrumentation Amplifier An instrumentation amplifier (IA) is a differential voltagegain device that amplifies the difference between the voltages existing

More information

Indigenous Design of Electronic Circuit for Electrocardiograph

Indigenous Design of Electronic Circuit for Electrocardiograph Indigenous Design of Electronic Circuit for Electrocardiograph Raman Gupta 1, Sandeep Singh 2, Kashish Garg 3, Shruti Jain 4 U.G student, Department of Electronics and Communication Engineering,Jaypee

More information

Biopotential Amplifiers. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan

Biopotential Amplifiers. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan Biopotential Ampliiers Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan chanhl@mail.cgu.edu.tw Operational ampliier Practical Ideal Biopotential ampliiers Ideal vs. practical

More information

BME 3512 Bioelectronics Laboratory Five - Operational Amplifiers

BME 3512 Bioelectronics Laboratory Five - Operational Amplifiers BME 351 Bioelectronics Laboratory Five - Operational Amplifiers Learning Objectives: Be familiar with the operation of a basic op-amp circuit. Be familiar with the characteristics of both ideal and real

More information

EECE Circuits and Signals: Biomedical Applications. Lab ECG I The Instrumentation Amplifier

EECE Circuits and Signals: Biomedical Applications. Lab ECG I The Instrumentation Amplifier EECE 150 - Circuits and Signals: Biomedical Applications Lab ECG I The Instrumentation Amplifier Introduction: As discussed in class, instrumentation amplifiers are often used to reject common-mode signals

More information

ECG Project. Raphal Blanchet, Axel Boland, Thomas Donnay, Mario Jose Teles Varandas, University of Liege

ECG Project. Raphal Blanchet, Axel Boland, Thomas Donnay, Mario Jose Teles Varandas, University of Liege ECG Project Raphal Blanchet, Axel Boland, Thomas Donnay, Mario Jose Teles Varandas, University of Liege Abstract We were asked to design our own Electrocardiogram. Obviously, recording heart beats without

More information

IMPROVEMENTS IN ELECTROCARDIOGRAPHY SMOOTHENING AND AMPLIFICATION

IMPROVEMENTS IN ELECTROCARDIOGRAPHY SMOOTHENING AND AMPLIFICATION IMPROVEMENTS IN ELECTROCARDIOGRAPHY SMOOTHENING AND AMPLIFICATION Manan Joshi, Sarosh Patel, Dr. Lawrence Hmurcik Electrical Engineering Department University of Bridgeport Bridgeport, CT 06604 Abstract

More information

TRANSDUCER INTERFACE APPLICATIONS

TRANSDUCER INTERFACE APPLICATIONS TRANSDUCER INTERFACE APPLICATIONS Instrumentation amplifiers have long been used as preamplifiers in transducer applications. High quality transducers typically provide a highly linear output, but at a

More information

AD8232 EVALUATION BOARD DOCUMENTATION

AD8232 EVALUATION BOARD DOCUMENTATION One Technology Way P.O. Box 9106 Norwood, MA 02062-9106 Tel: 781.329.4700 Fax: 781.461.3113 www.analog.com AD8232 EVALUATION BOARD DOCUMENTATION FEATURES Ready to use Heart Rate Monitor (HRM) Front end

More information

Instrumentation amplifier

Instrumentation amplifier Instrumentationamplifieris a closed-loop gainblock that has a differential input and an output that is single-ended with respect to a reference terminal. Application: are intended to be used whenever acquisition

More information

ECG Measurement and Analysis

ECG Measurement and Analysis ECG Measurement and Analysis Rob MacLeod and Brian Birchler February 24, 2014 Contents 1 Purpose and Background 1 2 Lab Procedure 2 2.1 Organization........................................ 2 2.2 Limb Leads.........................................

More information

BME/ISE 3512 Bioelectronics. Laboratory Five - Operational Amplifiers

BME/ISE 3512 Bioelectronics. Laboratory Five - Operational Amplifiers BME/ISE 3512 Bioelectronics Laboratory Five - Operational Amplifiers Learning Objectives: Be familiar with the operation of a basic op-amp circuit. Be familiar with the characteristics of both ideal and

More information

Instrumentation Amplifier and Filter Design for Biopotential Acquisition System CHANG-HAO CHEN

Instrumentation Amplifier and Filter Design for Biopotential Acquisition System CHANG-HAO CHEN Instrumentation Amplifier and Filter Design for Biopotential Acquisition System by CHANG-HAO CHEN Master of Science in Electrical and Electronics Engineering 2010 Faculty of Science and Technology University

More information

E40M. Instrumentation Amps and Noise. M. Horowitz, J. Plummer, R. Howe 1

E40M. Instrumentation Amps and Noise. M. Horowitz, J. Plummer, R. Howe 1 E40M Instrumentation Amps and Noise M. Horowitz, J. Plummer, R. Howe 1 ECG Lab - Electrical Picture Signal amplitude 1 mv Noise level will be significant will need to amplify and filter We ll use filtering

More information

Bioelectric Signal Analog Front-End Module Electrocardiograph

Bioelectric Signal Analog Front-End Module Electrocardiograph ***LOGO*** Bioelectric Signal Analog Front-End Module Electrocardiograph Features Single or Dual Supply Operation Quiescent Current: 220µA at 3.3v Internal Reference Generator with External Override Option

More information

Operational Amplifier BME 360 Lecture Notes Ying Sun

Operational Amplifier BME 360 Lecture Notes Ying Sun Operational Amplifier BME 360 Lecture Notes Ying Sun Characteristics of Op-Amp An operational amplifier (op-amp) is an analog integrated circuit that consists of several stages of transistor amplification

More information

Biopotential Electrodes

Biopotential Electrodes Biomedical Instrumentation Prof. Dr. Nizamettin AYDIN naydin@yildiz.edu.tr naydin@ieee.org http://www.yildiz.edu.tr/~naydin Biopotential Electrodes 1 2 Electrode electrolyte interface The current crosses

More information

Development of Electrocardiograph Monitoring System

Development of Electrocardiograph Monitoring System Development of Electrocardiograph Monitoring System Khairul Affendi Rosli 1*, Mohd. Hafizi Omar 1, Ahmad Fariz Hasan 1, Khairil Syahmi Musa 1, Mohd Fairuz Muhamad Fadzil 1, and Shu Hwei Neu 1 1 Department

More information

Chapter 9: Operational Amplifiers

Chapter 9: Operational Amplifiers Chapter 9: Operational Amplifiers The Operational Amplifier (or op-amp) is the ideal, simple amplifier. It is an integrated circuit (IC). An IC contains many discrete components (resistors, capacitors,

More information

BENG 186B Winter 2013 Final

BENG 186B Winter 2013 Final Name (Last, First): BENG 186B Winter 2013 Final This exam is closed book, closed note, calculators are OK. Circle and put your final answers in the space provided; show your work only on the pages provided.

More information

EE 230 Experiment 10 ECG Measurements Spring 2010

EE 230 Experiment 10 ECG Measurements Spring 2010 EE 230 Experiment 10 ECG Measurements Spring 2010 Note: If for any reason the students are uncomfortable with doing this experiment, please talk to the instructor for the course and an alternative experiment

More information

2. The. op-amp in and 10K. (a) 0 Ω. (c) 0.2% (d) (a) 0.02K. (b) 4. The. 5 V, then. 0V (virtual. (a) (c) Fall V. (d) V.

2. The. op-amp in and 10K. (a) 0 Ω. (c) 0.2% (d) (a) 0.02K. (b) 4. The. 5 V, then. 0V (virtual. (a) (c) Fall V. (d) V. Homework Assignment 04 Question 1 (2 points each unless noted otherwise) 1. A 9-V dc power supply generates 10 W in a resistor. What peak-to-peak amplitude should an ac source have to generate the same

More information

Ques on (2): [18 Marks] a) Draw the atrial synchronous Pacemaker block diagram and explain its operation. Benha University June 2013

Ques on (2): [18 Marks] a) Draw the atrial synchronous Pacemaker block diagram and explain its operation. Benha University June 2013 Benha University June 2013 Benha Faculty of Engineering Electrical Department Hospital Instrumentations (E472) 4 Th year (control) Dr.Waleed Abdel Aziz Salem Time: 3 Hrs Answer the following questions.

More information

Development of an alternative method for the calibration of ECG simulators

Development of an alternative method for the calibration of ECG simulators Development of an alternative method for the calibration of ECG simulators Roberto Benitez 1,, Romi Uresti 2,, and Carolina Solorzano 3, 1 ETALONS, S.A. de C.V., 64640 Monterrey, Mexico Abstract. In Mexico,

More information

Lecture #4 Special-purpose Op-amp Circuits

Lecture #4 Special-purpose Op-amp Circuits Spring 2015 Benha University Faculty of Engineering at Shoubra ECE-322 Electronic Circuits (B) Lecture #4 Special-purpose Op-amp Circuits Instructor: Dr. Ahmad El-Banna Agenda Instrumentation Amplifiers

More information

Operational Amplifiers. Boylestad Chapter 10

Operational Amplifiers. Boylestad Chapter 10 Operational Amplifiers Boylestad Chapter 10 DC-Offset Parameters Even when the input voltage is zero, an op-amp can have an output offset. The following can cause this offset: Input offset voltage Input

More information

Concepts to be Reviewed

Concepts to be Reviewed Introductory Medical Device Prototyping Analog Circuits Part 3 Operational Amplifiers, http://saliterman.umn.edu/ Department of Biomedical Engineering, University of Minnesota Concepts to be Reviewed Operational

More information

PHYS 536 The Golden Rules of Op Amps. Characteristics of an Ideal Op Amp

PHYS 536 The Golden Rules of Op Amps. Characteristics of an Ideal Op Amp PHYS 536 The Golden Rules of Op Amps Introduction The purpose of this experiment is to illustrate the golden rules of negative feedback for a variety of circuits. These concepts permit you to create and

More information

Testing and Stabilizing Feedback Loops in Today s Power Supplies

Testing and Stabilizing Feedback Loops in Today s Power Supplies Keywords Venable, frequency response analyzer, impedance, injection transformer, oscillator, feedback loop, Bode Plot, power supply design, open loop transfer function, voltage loop gain, error amplifier,

More information

AC-Coupled Front-End for Biopotential Measurements

AC-Coupled Front-End for Biopotential Measurements IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, VOL. 50, NO. 3, MARCH 2003 391 AC-Coupled Front-End for Biopotential Measurements Enrique Mario Spinelli 3, Student Member, IEEE, Ramon Pallàs-Areny, Fellow,

More information

Lab E5: Filters and Complex Impedance

Lab E5: Filters and Complex Impedance E5.1 Lab E5: Filters and Complex Impedance Note: It is strongly recommended that you complete lab E4: Capacitors and the RC Circuit before performing this experiment. Introduction Ohm s law, a well known

More information

CHAPTER 3. Instrumentation Amplifier (IA) Background. 3.1 Introduction. 3.2 Instrumentation Amplifier Architecture and Configurations

CHAPTER 3. Instrumentation Amplifier (IA) Background. 3.1 Introduction. 3.2 Instrumentation Amplifier Architecture and Configurations CHAPTER 3 Instrumentation Amplifier (IA) Background 3.1 Introduction The IAs are key circuits in many sensor readout systems where, there is a need to amplify small differential signals in the presence

More information

Biomedical Engineering Electrophysiology

Biomedical Engineering Electrophysiology Biomedical Engineering Electrophysiology Dr. rer. nat. Andreas Neubauer Sources of biological potentials and how to record them 1. How are signals transmitted along nerves? Transmit velocity Direction

More information

ANALYSIS AND DESIGN OF HIGH CMRR INSTRUMENTATION AMPLIFIER FOR ECG SIGNAL ACQUISITION SYSTEM USING 180nm CMOS TECHNOLOGY

ANALYSIS AND DESIGN OF HIGH CMRR INSTRUMENTATION AMPLIFIER FOR ECG SIGNAL ACQUISITION SYSTEM USING 180nm CMOS TECHNOLOGY International Journal of Electronics and Communication Engineering (IJECE) ISSN 2278-9901 Vol. 2, Issue 4, Sep 2013, 67-74 IASET ANALYSIS AND DESIGN OF HIGH CMRR INSTRUMENTATION AMPLIFIER FOR ECG SIGNAL

More information

Homework Assignment True or false. For both the inverting and noninverting op-amp configurations, V OS results in

Homework Assignment True or false. For both the inverting and noninverting op-amp configurations, V OS results in Question 1 (Short Takes), 2 points each. Homework Assignment 02 1. An op-amp has input bias current I B = 1 μa. Make an estimate for the input offset current I OS. Answer. I OS is normally an order of

More information

Homework Assignment 03

Homework Assignment 03 Homework Assignment 03 Question 1 (Short Takes), 2 points each unless otherwise noted. 1. Two 0.68 μf capacitors are connected in series across a 10 khz sine wave signal source. The total capacitive reactance

More information

ELEC4623 / ELEC9734 BIOMEDICAL ENGINEERING LABORATORY 3: DESIGN, TESTING AND ANALYSIS OF A HIGH QUALITY ISOLATED BIOPOTENTIAL AMPLIFIERS

ELEC4623 / ELEC9734 BIOMEDICAL ENGINEERING LABORATORY 3: DESIGN, TESTING AND ANALYSIS OF A HIGH QUALITY ISOLATED BIOPOTENTIAL AMPLIFIERS UNIVERSITY OF N.S.W. SCHOOL OF ELECTRICAL ENGINEERING AND TELECOMMUNICATIONS ELEC4623 / ELEC9734 BIOMEDICAL ENGINEERING LABORATORY 3: DESIGN, TESTING AND ANALYSIS OF A HIGH QUALITY ISOLATED BIOPOTENTIAL

More information

Crossover Distortion (hole) Amplifier Classification. Amplifiers: Class A, B, AB, D Op-amp Application: ECG. Why is [b] better? - LF356 - LF

Crossover Distortion (hole) Amplifier Classification. Amplifiers: Class A, B, AB, D Op-amp Application: ECG. Why is [b] better? - LF356 - LF Crossover Distortion (hole) 10kΩ Amplifiers: Class A, B, AB, D Op-amp Application: ECG v in 10kΩ 2 3 10kΩ 0.1µF - LF356 + 0.1µF?kΩ [a] +15 7 6 4-15 2N3906 +15 10kΩ 2N3904 v out R L v in -15 2 3 - LF356

More information

Introduction to Op Amps

Introduction to Op Amps Introduction to Op Amps ENGI 242 ELEC 222 Basic Op-Amp The op-amp is a differential amplifier with a very high open loop gain 25k AVOL 500k (much higher for FET inputs) high input impedance 500kΩ ZIN 10MΩ

More information

DEPARTMENT OF ELECTRICAL ENGINEERING AND COMPUTER SCIENCE MASSACHUSETTS INSTITUTE OF TECHNOLOGY CAMBRIDGE, MASSACHUSETTS 02139

DEPARTMENT OF ELECTRICAL ENGINEERING AND COMPUTER SCIENCE MASSACHUSETTS INSTITUTE OF TECHNOLOGY CAMBRIDGE, MASSACHUSETTS 02139 DEPARTMENT OF ELECTRICAL ENGINEERING AND COMPUTER SCIENCE MASSACHUSETTS INSTITUTE OF TECHNOLOGY CAMBRIDGE, MASSACHUSETTS 019.101 Introductory Analog Electronics Laboratory Laboratory No. READING ASSIGNMENT

More information

ELEC207 LINEAR INTEGRATED CIRCUITS

ELEC207 LINEAR INTEGRATED CIRCUITS Concept of VIRTUAL SHORT For feedback amplifiers constructed with op-amps, the two op-amp terminals will always be approximately equal (V + = V - ) This condition in op-amp feedback amplifiers is known

More information

Wireless Transmission of Real Time Electrocardiogram (ECG) Signals through Radio Frequency (RF) Waves

Wireless Transmission of Real Time Electrocardiogram (ECG) Signals through Radio Frequency (RF) Waves Wireless Transmission of Real Time Electrocardiogram (ECG) Signals through Radio Frequency (RF) Waves D.Sridhar raja Asst. Professor, Bharath University, Chennai-600073, India ABSTRACT:-In this project

More information

Lecture 4. Integrated Electronics

Lecture 4. Integrated Electronics Lecture 4 Integrated Electronics P, N is the doping of silicon to carry P (+) or N (-) charge) DIODES -> Recitifier I P N If V > V ON of diode, V V ON I = R Forward bias, conducting I Von ~ 0.6 V Example:

More information

Florida Atlantic University Biomedical Signal Processing Lab Experiment 2 Signal Transduction: Building an analog Electrocardiogram (ECG)

Florida Atlantic University Biomedical Signal Processing Lab Experiment 2 Signal Transduction: Building an analog Electrocardiogram (ECG) Florida Atlantic University Biomedical Signal Processing Lab Experiment 2 Signal Transduction: Building an analog Electrocardiogram (ECG) 1. Introduction: The Electrocardiogram (ECG) is a technique of

More information

Chapter 9: Operational Amplifiers

Chapter 9: Operational Amplifiers Chapter 9: Operational Amplifiers The Operational Amplifier (or op-amp) is the ideal, simple amplifier. It is an integrated circuit (IC). An IC contains many discrete components (resistors, capacitors,

More information

CHAPTER 7 INTERFERENCE CANCELLATION IN EMG SIGNAL

CHAPTER 7 INTERFERENCE CANCELLATION IN EMG SIGNAL 131 CHAPTER 7 INTERFERENCE CANCELLATION IN EMG SIGNAL 7.1 INTRODUCTION Electromyogram (EMG) is the electrical activity of the activated motor units in muscle. The EMG signal resembles a zero mean random

More information

*Notebook is excluded

*Notebook is excluded Biomedical Measurement Training System This equipment is designed for students to learn how to design specific measuring circuits and detect the basic physiological signals with practical operation. Moreover,

More information

* Notebook is excluded. Features KL-720 contains nine modules, including Electrocardiogram Measurement, E lectromyogram Measurement,

* Notebook is excluded. Features KL-720 contains nine modules, including Electrocardiogram Measurement, E lectromyogram Measurement, KL-720 Biomedical Measurement System Supplied by: 011 683 4365 This equipment is intended for students to learn how to design specific measuring circuits and detect the basic physiological signals with

More information

Lecture Notes Unit-III

Lecture Notes Unit-III Lecture Notes Unit-III FAQs Q1: An operational amplifier has a differential gain of 103 and CMRR of 100, input voltages are 120µV and 80µV, determine output voltage. 2 MARKS

More information

THE AMPLIFIER. A-B = C subtractor. INPUTS Figure 1

THE AMPLIFIER. A-B = C subtractor. INPUTS Figure 1 OBJECTIVES: THE AMPLIFIER 1) Explain the operation of the differential amplifier. 2) Determine the gain of each side of the differential amplifier. 3) Determine the gain of the differential amplifier as

More information

Portable EEG Signal Acquisition System

Portable EEG Signal Acquisition System Noor Ashraaf Noorazman, Nor Hidayati Aziz Faculty of Engineering and Technology, Multimedia University, Jalan Ayer Keroh Lama, 75450 Melaka, Malaysia Email: noor.ashraaf@gmail.com, hidayati.aziz@mmu.edu.my

More information

EE 3305 Lab I Revised July 18, 2003

EE 3305 Lab I Revised July 18, 2003 Operational Amplifiers Operational amplifiers are high-gain amplifiers with a similar general description typified by the most famous example, the LM741. The LM741 is used for many amplifier varieties

More information

Signal Conditioning Systems

Signal Conditioning Systems Note-13 1 Signal Conditioning Systems 2 Generalized Measurement System: The output signal from a sensor has generally to be processed or conditioned to make it suitable for the next stage Signal conditioning

More information

A Comprehensive Model for Power Line Interference in Biopotential Measurements

A Comprehensive Model for Power Line Interference in Biopotential Measurements IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, VOL. 49, NO. 3, JUNE 2000 535 A Comprehensive Model for Power Line Interference in Biopotential Measurements Mireya Fernandez Chimeno, Member, IEEE,

More information

InstrumentationTools.com

InstrumentationTools.com Author: Instrumentation Tools Categories: Multiple Choice Questions Measurement and Instrumentation Objective Questions Part 4 Measurement and Instrumentation Objective Questions 1. The decibel is a measure

More information

Operational Amplifiers

Operational Amplifiers Operational Amplifiers Table of contents 1. Design 1.1. The Differential Amplifier 1.2. Level Shifter 1.3. Power Amplifier 2. Characteristics 3. The Opamp without NFB 4. Linear Amplifiers 4.1. The Non-Inverting

More information

Detection of Abnormalities in Fetal by non invasive Fetal Heart Rate Monitoring System

Detection of Abnormalities in Fetal by non invasive Fetal Heart Rate Monitoring System IOSR Journal of Electronics and Communication Engineering (IOSR-JECE) e-issn: 2278-2834,p- ISSN: 2278-8735.Volume 11, Issue 3, Ver. III (May-Jun.2016), PP 35-41 www.iosrjournals.org Detection of Abnormalities

More information

Applied Electronics II

Applied Electronics II Applied Electronics II Chapter 3: Operational Amplifier Part 1- Op Amp Basics School of Electrical and Computer Engineering Addis Ababa Institute of Technology Addis Ababa University Daniel D./Getachew

More information

Development of a Low Cost ECG Data Acquisition Module

Development of a Low Cost ECG Data Acquisition Module Development of a Low Cost ECG Data Acquisition Module Deboleena Sadhukhan 1, Rohit Mitra 2, Avik Kundu 2, Madhuchhanda Mitra 3 Research Scholar, Department of Applied Physics, University of Calcutta, Kolkata,

More information

Design on Electrocardiosignal Detection Sensor

Design on Electrocardiosignal Detection Sensor Sensors & Transducers 203 by IFSA http://www.sensorsportal.com Design on Electrocardiosignal Detection Sensor Hao ZHANG School of Mathematics and Computer Science, Tongling University, 24406, China E-mail:

More information

Chapter 10: Operational Amplifiers

Chapter 10: Operational Amplifiers Chapter 10: Operational Amplifiers Differential Amplifier Differential amplifier has two identical transistors with two inputs and two outputs. 2 Differential Amplifier Differential amplifier has two identical

More information

Introduction to Biomedical Engineering

Introduction to Biomedical Engineering Introduction to Biomedical Engineering Biomedical Instrumentation Kung-Bin Sung 5/8/007 Outline Chapter 8 and chapter 5 of st edition: Bioinstrumentation Bridge circuit Operational amplifiers, instrumentation

More information

This is an outline of the subjects that will be touched upon during this presentation.

This is an outline of the subjects that will be touched upon during this presentation. Welcome to A Precision Low-Level DAS/ECG Cardio tachometer Demo board presentation. The presentation will focus on an interesting application of analog circuits where they are utilized to amplify and condition

More information

Section 6 Chapter 2: Operational Amplifiers

Section 6 Chapter 2: Operational Amplifiers 03 Section 6 Chapter : Operational Amplifiers eference : Microelectronic circuits Sedra sixth edition 4//03 4//03 Contents: - DC imperfections A. Offset voltage B. Solution of offset voltage C. Input bias

More information

WHALETEQ Common Mode Rejection Ratio Tester (CMRR 3.0) User Manual (Revision )

WHALETEQ Common Mode Rejection Ratio Tester (CMRR 3.0) User Manual (Revision ) WHALETEQ Common Mode Rejection Ratio Tester (CMRR 3.0) User Manual (Revision 2017-10-06) Copyright (c) 2013-2017, All Rights Reserved. WhaleTeq Co. LTD No part of this publication may be reproduced, transmitted,

More information

DUAL ULTRA MICROPOWER RAIL-TO-RAIL CMOS OPERATIONAL AMPLIFIER

DUAL ULTRA MICROPOWER RAIL-TO-RAIL CMOS OPERATIONAL AMPLIFIER ADVANCED LINEAR DEVICES, INC. ALD276A/ALD276B ALD276 DUAL ULTRA MICROPOWER RAILTORAIL CMOS OPERATIONAL AMPLIFIER GENERAL DESCRIPTION The ALD276 is a dual monolithic CMOS micropower high slewrate operational

More information

Chapter 5. Operational Amplifiers and Source Followers. 5.1 Operational Amplifier

Chapter 5. Operational Amplifiers and Source Followers. 5.1 Operational Amplifier Chapter 5 Operational Amplifiers and Source Followers 5.1 Operational Amplifier In single ended operation the output is measured with respect to a fixed potential, usually ground, whereas in double-ended

More information

Changing the sampling rate

Changing the sampling rate Noise Lecture 3 Finally you should be aware of the Nyquist rate when you re designing systems. First of all you must know your system and the limitations, e.g. decreasing sampling rate in the speech transfer

More information

55:041 Electronic Circuits The University of Iowa Fall Exam 3. Question 1 Unless stated otherwise, each question below is 1 point.

55:041 Electronic Circuits The University of Iowa Fall Exam 3. Question 1 Unless stated otherwise, each question below is 1 point. Exam 3 Name: Score /65 Question 1 Unless stated otherwise, each question below is 1 point. 1. An engineer designs a class-ab amplifier to deliver 2 W (sinusoidal) signal power to an resistive load. Ignoring

More information

Voltage-to-Frequency and Frequency-to-Voltage Converter ADVFC32

Voltage-to-Frequency and Frequency-to-Voltage Converter ADVFC32 a FEATURES High Linearity 0.01% max at 10 khz FS 0.05% max at 100 khz FS 0.2% max at 500 khz FS Output TTL/CMOS Compatible V/F or F/V Conversion 6 Decade Dynamic Range Voltage or Current Input Reliable

More information

IC Preamplifier Challenges Choppers on Drift

IC Preamplifier Challenges Choppers on Drift IC Preamplifier Challenges Choppers on Drift Since the introduction of monolithic IC amplifiers there has been a continual improvement in DC accuracy. Bias currents have been decreased by 5 orders of magnitude

More information

Chapter 4 4. Optoelectronic Acquisition System Design

Chapter 4 4. Optoelectronic Acquisition System Design 4. Optoelectronic Acquisition System Design The present chapter deals with the design of the optoelectronic (OE) system required to translate the obtained optical modulated signal with the photonic acquisition

More information

Circuit Applications of Multiplying CMOS D to A Converters

Circuit Applications of Multiplying CMOS D to A Converters Circuit Applications of Multiplying CMOS D to A Converters The 4-quadrant multiplying CMOS D to A converter (DAC) is among the most useful components available to the circuit designer Because CMOS DACs

More information

AN-1106 Custom Instrumentation Amplifier Design Author: Craig Cary Date: January 16, 2017

AN-1106 Custom Instrumentation Amplifier Design Author: Craig Cary Date: January 16, 2017 AN-1106 Custom Instrumentation Author: Craig Cary Date: January 16, 2017 Abstract This application note describes some of the fine points of designing an instrumentation amplifier with op-amps. We will

More information

Laboratory Project 1B: Electromyogram Circuit

Laboratory Project 1B: Electromyogram Circuit 2240 Laboratory Project 1B: Electromyogram Circuit N. E. Cotter, D. Christensen, and K. Furse Electrical and Computer Engineering Department University of Utah Salt Lake City, UT 84112 Abstract-You will

More information

EDL Group #3 Final Report - Surface Electromyograph System

EDL Group #3 Final Report - Surface Electromyograph System EDL Group #3 Final Report - Surface Electromyograph System Group Members: Aakash Patil (07D07021), Jay Parikh (07D07019) INTRODUCTION The EMG signal measures electrical currents generated in muscles during

More information

Introduction to Analog Interfacing. ECE/CS 5780/6780: Embedded System Design. Various Op Amps. Ideal Op Amps

Introduction to Analog Interfacing. ECE/CS 5780/6780: Embedded System Design. Various Op Amps. Ideal Op Amps Introduction to Analog Interfacing ECE/CS 5780/6780: Embedded System Design Scott R. Little Lecture 19: Operational Amplifiers Most embedded systems include components that measure and/or control real-world

More information

EXPERIMENT 7 The Amplifier

EXPERIMENT 7 The Amplifier Objectives EXPERIMENT 7 The Amplifier 1) Understand the operation of the differential amplifier. 2) Determine the gain of each side of the differential amplifier. 3) Determine the gain of the differential

More information

Experiments #7. Operational Amplifier part 1

Experiments #7. Operational Amplifier part 1 Experiments #7 Operational Amplifier part 1 1) Objectives: The objective of this lab is to study operational amplifier (op amp) and its applications. We will be simulating and building some basic op-amp

More information

HUMAN DETECTION AND RESCUE USING BIO POTENTIAL SIGNALS

HUMAN DETECTION AND RESCUE USING BIO POTENTIAL SIGNALS ISET GOLDEN JUBILEE SYMPOSIUM Indian Society of Earthquake Technology Department of Earthquake Engineering Building IIT Roorkee, Roorkee October 20-21, 2012 Paper No. A007 HUMAN DETECTION AND RESCUE USING

More information

DESIGN OF OTA-C FILTER FOR BIOMEDICAL APPLICATIONS

DESIGN OF OTA-C FILTER FOR BIOMEDICAL APPLICATIONS DESIGN OF OTA-C FILTER FOR BIOMEDICAL APPLICATIONS Sreedhar Bongani 1, Dvija Mounika Chirumamilla 2 1 (ECE, MCIS, MANIPAL UNIVERSITY, INDIA) 2 (ECE, K L University, INDIA) ABSTRACT-This paper presents

More information

UNIVERSITY OF UTAH ELECTRICAL AND COMPUTER ENGINEERING DEPARTMENT ELECTROMYOGRAM (EMG) DETECTOR WITH AUDIOVISUAL OUTPUT

UNIVERSITY OF UTAH ELECTRICAL AND COMPUTER ENGINEERING DEPARTMENT ELECTROMYOGRAM (EMG) DETECTOR WITH AUDIOVISUAL OUTPUT UNIVESITY OF UTAH ELECTICAL AND COMPUTE ENGINEEING DEPATMENT ECE 3110 LABOATOY EXPEIMENT NO. 5 ELECTOMYOGAM (EMG) DETECTO WITH AUDIOVISUAL OUTPUT Pre-Lab Assignment: ead and review Sections 2.4, 2.8.2,

More information

UNIT - 1 OPERATIONAL AMPLIFIER FUNDAMENTALS

UNIT - 1 OPERATIONAL AMPLIFIER FUNDAMENTALS UNIT - 1 OPERATIONAL AMPLIFIER FUNDAMENTALS 1.1 Basic operational amplifier circuit- hte basic circuit of an operational amplifier is as shown in above fig. has a differential amplifier input stage and

More information

Nitish V. Thakor. "Bipotentials and Electrophysiology Measurement." Copyright 2000 CRC Press LLC. <http://www.engnetbase.com>.

Nitish V. Thakor. Bipotentials and Electrophysiology Measurement. Copyright 2000 CRC Press LLC. <http://www.engnetbase.com>. Nitish V. Thakor. "Bipotentials and Electrophysiology Measurement." Copyright 2000 CRC Press LLC. . Biopotentials and Electrophysiology Measurement Nitish V. Thakor Johns Hopkins

More information

DEPARTMENT OF ELECTRICAL ENGINEERING AND COMPUTER SCIENCE MASSACHUSETTS INSTITUTE OF TECHNOLOGY CAMBRIDGE, MASSACHUSETTS 02139

DEPARTMENT OF ELECTRICAL ENGINEERING AND COMPUTER SCIENCE MASSACHUSETTS INSTITUTE OF TECHNOLOGY CAMBRIDGE, MASSACHUSETTS 02139 DEPARTMENT OF ELECTRICAL ENGINEERING AND COMPUTER SCIENCE MASSACHUSETTS INSTITUTE OF TECHNOLOGY CAMBRIDGE, MASSACHUSETTS 019 Spring Term 00.101 Introductory Analog Electronics Laboratory Laboratory No.

More information

BME/ECE 463. Computers in Medicine. Answers to Selected Textbook Problems

BME/ECE 463. Computers in Medicine. Answers to Selected Textbook Problems BME/ECE 463 Computers in Medicine Answers to Selected Textbook Problems W. J. Tompkins ed. Biomedical Digital Signal Processing: C Language Examples and Laboratory Experiments for the IBM PC. Englewood

More information

Laboratory Project 1: Design of a Myogram Circuit

Laboratory Project 1: Design of a Myogram Circuit 1270 Laboratory Project 1: Design of a Myogram Circuit Abstract-You will design and build a circuit to measure the small voltages generated by your biceps muscle. Using your circuit and an oscilloscope,

More information

INTERFERENCE REDUCTION IN ECG RECORDINGS BY USING A DUAL GROUND ELECTRODE

INTERFERENCE REDUCTION IN ECG RECORDINGS BY USING A DUAL GROUND ELECTRODE XIX IMEKO World Congress Fundamental and Applied Metrology September 6 11, 29, Lisbon, Portugal INTERFERENCE REDUCTION IN ECG RECORDINGS BY USING A DUAL GROUND ELECTRODE Delia Díaz, Óscar Casas, Ramon

More information

Infrared Communications Lab

Infrared Communications Lab Infrared Communications Lab This lab assignment assumes that the student knows about: Ohm s Law oltage, Current and Resistance Operational Amplifiers (See Appendix I) The first part of the lab is to develop

More information