Radionuclide Imaging MII 3073 RADIONUCLIDE IMAGING SYSTEM
|
|
- Theodore Beasley
- 5 years ago
- Views:
Transcription
1 Radionuclide Imaging MII 3073 RADIONUCLIDE IMAGING SYSTEM
2 Preamplifiers and amplifiers The current from PMT must be further amplified before it can be processed and counted (the number of electrons yielded is still small). Typically, this amplification is divided into: 1. Preamplifier: located close to PMT, increases the number of charges sufficiently to allow the current to be transmitted to amplifier 2. Amplifier: main amplifier that increases the current of electrical pulses as much as a thousand-fold.
3 Pulse-height analyzer The amplitude of each electrical pulse from amplifiers is measured in the electrical circuits of pulse-height analyzer (PHA). Generates a plot of the number of pulses against their energy, called pulse-height spectrum. The pulse-height analyzer is often used to select Z pulses that correspond to a range of the acceptable energies, known as energy window. Eg: A window setting of 20% for 140keV 99m Tc means a range of kev energy will be accepted and counted.
4
5 Computers in nuclear medicine Used for acquisition, storage and processing of data. For each Z pulse that is accepted by the pulse-height analyzer, one count is added to the storage location that corresponds to its x,y location determined by the positioning circuit. The data storage can be visualized as a matrix. Each position within the matrix corresponds to a pixel within the image and is assigned a unique address composed of the row and column of its location. A 64 x 64 matrix has 4096 pixels; a 128 x 128 has pixels. The greater the number of pixels, the smaller the size of a pixel for a given FOV, and the better the resolution of the image.
6
7 Rectilinear Scanner Rectilinear scanner is one of the early scintillation imaging devices. Can be divided into two: 1. Single detector assembly: these automatic scanners, called rectilinear scanners, make a map of the location of radioactive materials in a patient s body. 2. Multiple detector assemblies: consists of two opposing detector heads, each head consisting of multiple NaI crystals and multiple associated photomultiplier tubes, each crystal with a focused collimator. Rectilinear: Refers to images with no geometric distortion in which the scales in the horizontal and vertical directions are identical.
8 In a rectilinear scanner, detector head is moved over the organ of interest in a straight line with the help of an electric motor. After travel at a specific distance, the detector head either steps up or down on a small distance and then continues the straight line motion in the opposite direction. The detector head assembly consists of a collimator, NaI(Tl) crystal and PMT with preamplifier. The information gathered by the detector head is continuously relayed to a photographic system using 14 x 17 inch x-ray film. This system is moved in a light-tight box in synchronization with the detector head. Dark areas on the film represent high areas of radioactivity in the organ, and vice versa.
9
10
11 Scintillation Scanner/Gamma Camera In gamma camera, the whole organ or a large part of the body is imaged simultaneously. This feature makes the study of rapid dynamic processes possible. A scintillation camera can physically be divided into two parts: 1. Detector head contains the collimator and NaI(Tl) crystal with PMT and associated electronics and is mounted on a stand where it can be easily moved up or down or rotated in any desired position with hand-held controls or computerdriven automatic. 2. Console houses the power supplies and operational controls including the display system.
12 The outputs from the PMTs are converted into three signals, two of which (X and Y) give the spatial location of the scintillation while the third (Z) represents the energy deposited in the crystal by the gamma ray. To improve their quality these signals then pass through correction circuits. The Z signal goes to a PHA, which tests whether the energy of the gamma ray is within the window. The ability of a gamma camera to accurately image sources of activity depends on several factors such as: 1. Intrinsic resolution of the camera (the ability of the camera to distinguish between two closely spaced interactions in the crystal). 2. The gamma-ray energy. 3. The choice of collimator. 4. The distance from the collimator face to the source.
13
14 The following steps are taken to obtain an image with a scintillation camera: 1. Select the study to be performed. 2. Select the radiopharmaceutical used and its dose. 3. Select the PHA parameters (peak energy and % window). 4. Select an appropriate collimator (with respect to energy and spatial resolution). 5. Select the mode: accumulation of a certain number of counts or exposure for fixed amount of time. 6. Select the appropriate intensity of the display system (cathode ray tube). 7. Position the patient under the camera. 8. Start and finish of the exposure. 9. Development of the film. If more than one view is to be displayed on the same film, then the development of the film takes place at the end of the study.
15
Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT
Quality control of Gamma Camera By Dr/ Ibrahim Elsayed Saad 242 NMT WHAT IS QUALITY? The quality of a practice is to fulfill the expectations and demands from: Patient Clinicain Your self Quality assurance
More informationPhotomultiplier Tube
Nuclear Medicine Uses a device known as a Gamma Camera. Also known as a Scintillation or Anger Camera. Detects the release of gamma rays from Radionuclide. The radionuclide can be injected, inhaled or
More informationNM Module Section 2 6 th Edition Christian, Ch. 3
NM 4303 Module Section 2 6 th Edition Christian, Ch. 3 Gas Filled Chamber Voltage Gas filled chamber uses Hand held detectors cutie pie Geiger counter Dose calibrators Cutie pie Chamber voltage in Ionization
More informationElectronic Instrumentation for Radiation Detection Systems
Electronic Instrumentation for Radiation Detection Systems January 23, 2018 Joshua W. Cates, Ph.D. and Craig S. Levin, Ph.D. Course Outline Lecture Overview Brief Review of Radiation Detectors Detector
More informationNuclear Associates , &
Nuclear Associates 76-810, 76-814 76-815 & 76-818 Bar Phantoms and Test Patterns Operators Manual March 2005 Manual No. 76-810-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A.
More informationCHAPTER 8 GENERIC PERFORMANCE MEASURES
GENERIC PERFORMANCE MEASURES M.E. DAUBE-WITHERSPOON Department of Radiology, University of Pennsylvania, Philadelphia, Pennsylvania, United States of America 8.1. INTRINSIC AND EXTRINSIC MEASURES 8.1.1.
More informationCZT Technology: Fundamentals and Applications
GE Healthcare CZT Technology: Fundamentals and Applications White Paper Abstract Nuclear Medicine traces its technology roots to the 1950 s, and while it has continued to evolve since the invention of
More informationRobert Pagnanelli BSRT(R)(N), CNMT, NCT, FASNC Chief Technologist, Nuclear Imaging Duke University Medical Center. Thursday September 8, 2011
Robert Pagnanelli BSRT(R)(N), CNMT, NCT, FASNC Chief Technologist, Nuclear Imaging Duke University Medical Center Thursday September 8, 2011 Quality Control Quality control should be performed because:
More informationCHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES
CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES The current multiplication mechanism offered by dynodes makes photomultiplier tubes ideal for low-light-level measurement. As explained earlier, there
More informationInitial Certification
Initial Certification Nuclear Medical Physics (NMP) Study Guide Part 2 Content Guide and Sample Questions The content of all ABR exams is determined by a panel of experts who select the items based on
More informationFirst Applications of the YAPPET Small Animal Scanner
First Applications of the YAPPET Small Animal Scanner Guido Zavattini Università di Ferrara CALOR2 Congress, Annecy - FRANCE YAP-PET scanner Scintillator: YAP:Ce Size: matrix of 2x2 match like crystals
More informationGS Introduction to Medical Physics IV Laboratory 5 Gamma Camera Characteristics
GS02 0193 Introduction to Medical Physics IV Laboratory 5 Gamma Camera Characteristics Purpose: To introduce some of the basic characteristics of a gamma camera. This lab will introduce gamma camera QC
More informationDevelopment of a simplified readout for a compact gamma camera based on 2 2 H8500 multi-anode PSPMT array
University of Wollongong Research Online Faculty of Engineering and Information Sciences - Papers: Part A Faculty of Engineering and Information Sciences 2010 Development of a simplified readout for a
More informationRadiation Detection Instrumentation
Radiation Detection Instrumentation Principles of Detection and Gas-filled Ionization Chambers Neutron Sensitive Ionization Chambers Detection of radiation is a consequence of radiation interaction with
More informationGamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors
Nuclear Physics #1 Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors Introduction: In this experiment you will use both scintillation and semiconductor detectors to study γ- ray energy spectra. The
More informationSECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS
RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT
More informationGamma-ray spectral imaging using a single-shutter radiation camera
Nuclear Instruments and Methods in Physics Research A299 (1990) 495-500 North-Holland 495 Gamma-ray spectral imaging using a single-shutter radiation camera T.A. DeVol, D.K. Wehe and G.F. Knoll The University
More informationIntrinsic and Tomographic Evaluation of Siemens e.cam SPECT System at the Korle-Bu Teaching Hospital (Ghana)
Research Journal of Applied Sciences, Engineering and Technology 3(10): 1152-1158, 2011 ISSN: 2040-7467 Maxwell Scientific Organization, 2011 Submitted: July 17, 2011 Accepted: September 05, 2011 Published:
More informationDEVELOPMENT OF HIGH STABLE MONITOR FOR MEASURERING ENVIRONMENTAL RADIATION
DEVELOPMENT OF HIGH STABLE MONITOR FOR MEASURERING ENVIRONMENTAL RADIATION Ken ichiro Moriai.,Hiroshi Kawaguchi,Shohei Matsubara, Naoki Tateishi(ALOKA CO.,LTD.) Masatoshi Egawa,Hideaki Kakihana(THE KANSAI
More informationPHYSICS ADVANCED LABORATORY I COMPTON SCATTERING Spring 2002
PHYSICS 334 - ADVANCED LABORATORY I COMPTON SCATTERING Spring 00 Purposes: Demonstrate the phenomena associated with Compton scattering and the Klein-Nishina formula. Determine the mass of the electron.
More informationIntroduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image
Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology
More informationDesign of a Preamplifier Card for the Photomultiplier Tubes of a Gamma Camera
Journal of Physical Science and Application 8 (1) (2018) 46-54 doi: 10.17265/2159-5348/2018.01.006 D DAVID PUBLISHING Design of a Preamplifier Card for the Photomultiplier Tubes of a Gamma Camera Jorge
More informationA high energy gamma camera using a multiple hole collimator
ELSEVIER Nuclear Instruments and Methods in Physics Research A 353 (1994) 328-333 A high energy gamma camera using a multiple hole collimator and PSPMT SV Guru *, Z He, JC Ferreria, DK Wehe, G F Knoll
More informationGamma Spectrometer Initial Project Proposal
Gamma Spectrometer Initial Project Proposal Group 9 Aman Kataria Johnny Klarenbeek Dean Sullivan David Valentine Introduction There are currently two main types of gamma radiation detectors used for gamma
More informationParameters Affecting on Intrinsic Uniformity Test For MEDISO
ISPUB.COM The Internet Journal of Nuclear Medicine Volume 5 Number 2 Parameters Affecting on Intrinsic Uniformity Test For MEDISO S Zobly, A Osman Citation S Zobly, A Osman. Parameters Affecting on Intrinsic
More informationLaBr 3 :Ce, the latest crystal for nuclear medicine
10th Topical Seminar on Innovative Particle and Radiation Detectors 1-5 October 2006 Siena, Italy LaBr 3 :Ce, the latest crystal for nuclear medicine Roberto Pani On behalf of SCINTIRAD Collaboration INFN
More informationPinhole collimator design for nuclear survey system
Annals of Nuclear Energy 29 (2002) 2029 2040 www.elsevier.com/locate/anucene Pinhole collimator design for nuclear survey system Wanno Lee*, Gyuseong Cho Department of Nuclear Engineering, Korea Advanced
More information2010 Philips BrightView XCT SPECT/CT
2010 Philips BrightView XCT SPECT/CT Unit was purchased from Philips training center in 2015. Installed but never been used by the current facility. (Scroll for pictures) BrightView XCT Camera with PinPoint
More informationDevelopment of an innovative LSO-SiPM detector module for high-performance Positron Emission Tomography
Development of an innovative LSO-SiPM detector module for high-performance Positron Emission Tomography Maria Leonor Trigo Franco Frazão leonorfrazao@ist.utl.pt Instituto Superior Técnico, Lisboa, Portugal
More informationQUALITY CONTROL OF IMAGING DEVICES. P.S. Soni
XA9847605 Chapter 8 P.S. Soni Quality Assurance and Quality Control in Nuclear Medicine Quality assurance in nuclear medicine refers collectively to all aspects of a nuclear medicine service. It would
More informationOn the initiation of lightning in thunderclouds (Instrumentation, Supplementary information)
On the initiation of lightning in thunderclouds (Instrumentation, Supplementary information) Ashot Chilingarian 1,2, Suren Chilingaryan 1, Tigran Karapetyan 1, Lev Kozliner 1, Yeghia Khanikyants 1, Gagik
More informationDetecting and Suppressing Background Signal
Detecting and Suppressing Background Signal Valerie Gray St. Norbert College Advisors: Dr. Michael Wiescher Freimann Professor Nuclear Physics University of Notre Dame Dr. Ed Stech Associate Professional
More informationPD233: Design of Biomedical Devices and Systems
PD233: Design of Biomedical Devices and Systems (Lecture-8 Medical Imaging Systems) (Imaging Systems Basics, X-ray and CT) Dr. Manish Arora CPDM, IISc Course Website: http://cpdm.iisc.ac.in/utsaah/courses/
More informationNew Technology in Nuclear Medicine
New Technology in Nuclear Medicine Reed G. Selwyn, PhD, DABR Vice Chair of Research & Imaging Sciences Associate Professor and Chief, Medical Physics Dept. of Radiology, University of New Mexico Objectives
More informationORTEC. Research Applications. Pulse-Height, Charge, or Energy Spectroscopy. Detectors. Processing Electronics
ORTEC Spectroscopy systems for ORTEC instrumentation produce pulse height distributions of gamma ray or alpha energies. MAESTRO-32 (model A65-B32) is the software included with most spectroscopy systems
More information(12) Patent Application Publication (10) Pub. No.: US 2004/ A1
(19) United States US 2004.0036026A1 (12) Patent Application Publication (10) Pub. No.: US 2004/0036026 A1 Engdahl et al. (43) Pub. Date: Feb. 26, 2004 (54) SYSTEM AND METHOD FOR CALIBRATING AND TUNING
More informationFast first practical help -- detailed instructions will follow- preliminary Experiment F80
Fast first practical help -- detailed instructions will follow- preliminary Experiment F80 Measurement Methods of Nuclear and Particle Physics Introduction: This experiment is going to introduce you to
More informationReal Time Linear Array Imaging. Brian Caccamise
Real Time Linear Array Imaging Brian Caccamise 1 Real Time Linear Array Imaging What is Real Time Linear Array Imaging? Or Real Time Radiography (RTR)? 2 Real Time Linear Array Imaging It s Not This! Shoe
More informationCOMPUTED TOMOGRAPHY 1
COMPUTED TOMOGRAPHY 1 Why CT? Conventional X ray picture of a chest 2 Introduction Why CT? In a normal X-ray picture, most soft tissue doesn't show up clearly. To focus in on organs, or to examine the
More informationFRCR Nuclear Medicine
FRCR Nuclear Medicine FRCR LECTURES Lecture I 20/09/2016: Nuclear Medicine and Image Formation Lecture II 22/09/2016: Positron Emission Tomography & QA Lecture III 27/09/2016: Radiation Detectors - Radiation
More informationPhysics Laboratory Scattering of Photons from Electrons: Compton Scattering
RR Oct 2001 SS Dec 2001 MJ Oct 2009 Physics 34000 Laboratory Scattering of Photons from Electrons: Compton Scattering Objective: To measure the energy of high energy photons scattered from electrons in
More informationPerformance evaluation of two CZT gamma ray imaging systems
Louisiana State University LSU Digital Commons LSU Master's Theses Graduate School 2005 Performance evaluation of two CZT gamma ray imaging systems Laurie Kelly Louisiana State University and Agricultural
More informationOverview 256 channel Silicon Photomultiplier large area using matrix readout system The SensL Matrix detector () is the largest area, highest channel
技股份有限公司 wwwrteo 公司 wwwrteo.com Page 1 Overview 256 channel Silicon Photomultiplier large area using matrix readout system The SensL Matrix detector () is the largest area, highest channel count, Silicon
More information... In vivo imaging in Nuclear Medicine. 1957: Anger camera (X;Y) X Y
József Varga, PhD EMISSION IMAGING BASICS OF QUANTIFICATION Imaging devices Aims of image processing Reconstruction University of Debrecen Department of Nuclear Medicine. In vivo imaging in Nuclear Medicine
More informationSYLLABUS. 1. Identification of Subject:
SYLLABUS Date/ Revision : 30 January 2017/1 Faculty : Life Sciences Approval : Dean, Faculty of Life Sciences SUBJECT : Biophysics 1. Identification of Subject: Name of Subject : Biophysics Code of Subject
More informationPET Detectors. William W. Moses Lawrence Berkeley National Laboratory March 26, 2002
PET Detectors William W. Moses Lawrence Berkeley National Laboratory March 26, 2002 Step 1: Inject Patient with Radioactive Drug Drug is labeled with positron (β + ) emitting radionuclide. Drug localizes
More informationPositron Emission Tomography
Positron Emission Tomography UBC Physics & Astronomy / PHYS 409 1 Introduction Positron emission tomography (PET) is a non-invasive way to produce the functional 1 image of a patient. It works by injecting
More informationCHAPTER 7 MEDICINE IMAGING DEVICES
CHAPTER 7 ELECTRONICS RELATED TO NUCLEAR MEDICINE IMAGING DEVICES R.J. OTT Joint Department of Physics, Royal Marsden Hospital and Institute of Cancer Research, Surrey R. STEPHENSON Rutherford Appleton
More informationHow Gamma Camera s Head-Tilts Affect Image Quality of a Nuclear Scintigram?
November 2014, Volume 1, Number 4 How Gamma Camera s Head-Tilts Affect Image Quality of a Nuclear Scintigram? Hojjat Mahani 1,2, Alireza Kamali-Asl 3, *, Mohammad Reza Ay 2, 4 1. Radiation Application
More informationDevelopment of Personal Dosimeter Using Electronic Dose Conversion Method
Proceedings of the Korean Nuclear Spring Meeting Gyeong ju, Korea, May 2003 Development of Personal Dosimeter Using Electronic Dose Conversion Method Wanno Lee, Bong Jae Lee, and Chang Woo Lee Korea Atomic
More informationMammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over
Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over a lifetime Breast cancer screening programs rely on
More informationRadionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT)
Radionuclide Imaging MII 3073 Single Photon Emission Computed Tomography (SPECT) Single Photon Emission Computed Tomography (SPECT) The successful application of computer algorithms to x-ray imaging in
More informationErgo TM Imaging System
Ergo TM Imaging System Unparalleled Clinical Flexibility and Imaging Quality The Ergo Imaging System is Digirad s advanced solid-state large field-of-view (LFOV) general purpose nuclear medicine camera.
More informationTHE increasing interest on pinhole collimation of gamma
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 54, NO. 3, JUNE 2007 469 CsI(Tl) Micro-Pixel Scintillation Array for Ultra-high Resolution Gamma-ray Imaging M. N. Cinti, R. Scafè, R. Pellegrini, C. Trotta,
More informationAcquisition, Processing and Display
Acquisition, Processing and Display Terri L. Fauber, R.T. (R)(M) Department of Radiation Sciences School of Allied Health Professions Virginia Commonwealth University Topics Image Characteristics Image
More informationDetector And Front-End Electronics Of A Fissile Mass Flow Monitoring System
Detector And Front-End Electronics Of A Fissile Mass Flow Monitoring System M. J. Paulus, T. Uckan, R. Lenarduzzi, J. A. Mullens, K. N. Castleberry, D. E. McMillan, J. T. Mihalczo Instrumentation and Controls
More informationInstructions for gg Coincidence with 22 Na. Overview of the Experiment
Overview of the Experiment Instructions for gg Coincidence with 22 Na 22 Na is a radioactive element that decays by converting a proton into a neutron: about 90% of the time through β + decay and about
More informationFinal Report Data Acquisition Box
Final Report Data Acquisition Box By Gene Bender DeSmet Jesuit High School July 25, 2003 Contents Overview...2 A Hybrid LBNL Cosmic Ray Detector...2 The Detectors...6 Changing PMT Voltage...7 Comparator
More informatione t Development of Low Cost γ - Ray Energy Spectrometer
e t International Journal on Emerging Technologies (Special Issue on NCRIET-2015) 6(2): 315-319(2015) ISSN No. (Print) : 0975-8364 ISSN No. (Online) : 2249-3255 Development of Low Cost γ - Ray Energy Spectrometer
More informationINDEX. Firmware for DPP (Digital Pulse Processing) DPP-PSD Digital Pulse Processing for Pulse Shape Discrimination
Firmware for DPP (Digital Pulse Processing) Thanks to the powerful FPGAs available nowadays, it is possible to implement Digital Pulse Processing (DPP) algorithms directly on the acquisition boards and
More informationEnsuring Shielding adequacy in Lead shielded spent fuel transportation casks using gamma scanning
Ensuring Shielding adequacy in Lead shielded spent fuel transportation casks using gamma scanning More info about this article: http://www.ndt.net/?id=21208 M.Ravichandra 1, P.Raghavendra 1, Dhiren Kothari
More informationThe Light Amplifier Concept
The Light Amplifier Concept Daniel Ferenc 1 Eckart Lorenz 1,2 Daniel Kranich 1 Alvin Laille 1 (1) Physics Department, University of California Davis (2) Max Planck Institute, Munich Work supported partly
More informationECC419 IMAGE PROCESSING
ECC419 IMAGE PROCESSING INTRODUCTION Image Processing Image processing is a subclass of signal processing concerned specifically with pictures. Digital Image Processing, process digital images by means
More informationResults of the Measurement of the Collimator Hole Angulation for Different Collimators of SPECT with Adaptive Quality Control Phantom
Modern Instrumentation, 2012, 1, 4953 http://dx.doi.org/10.4236/mi.2012.14007 Published Online October 2012 (http://www.scirp.org/journal/mi) Results of the Measurement of the Collimator Hole Angulation
More informationPrimer on molecular imaging technology
Primer on molecular imaging technology Craig S. Levin Division of Nuclear Medicine, Department of Radiology and Molecular Imaging Program at Stanford (MIPS), Stanford University School of Medicine, 300
More information60 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 54, NO. 1, FEBRUARY /$ IEEE
60 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 54, NO. 1, FEBRUARY 2007 Prototype Parallel Readout System for Position Sensitive PMT Based Gamma Ray Imaging Systems Frezghi Habte, Member, IEEE, Peter D.
More informationORTEC Experiment 3. Gamma-Ray Spectroscopy Using NaI(Tl) Equipment Required. Purpose. Gamma Emission
ORTEC Experiment 3 Equipment Required Electronic Instrumentation o SPA38 Integral Assembly consisting of a 38 mm x 38 mm NaI(Tl) Scintillator, Photomultiplier Tube, and PMT Base with Stand o 4001A/4002D
More informationDesign and development of compact readout electronics with silicon photomultiplier array for a compact imaging detector *
CPC(HEP & NP), 2012, 36(10): 973 978 Chinese Physics C Vol. 36, No. 10, Oct., 2012 Design and development of compact readout electronics with silicon photomultiplier array for a compact imaging detector
More informationCOMPTON SCATTERING. Purpose. Introduction. Fundamentals of Experiment
COMPTON SCATTERING Purpose The purpose of this experiment is to verify the energy dependence of gamma radiation upon scattering angle and to compare the differential cross section obtained from the data
More informationDigital Radiographic Inspection replacing traditional RT and 3D RT Development
Digital Radiographic Inspection replacing traditional RT and 3D RT Development Iploca Novel Construction Meeting 27&28 March 2014 Geneva By Jan van der Ent Technical Authority International Contents Introduction
More informationExprerimental Evaluation of a Dedicated Pinhole SPECT System for Small Animal Imaging and Scintimammography
ETASR - Engineering, Technology & Applied Science Research Vol. 1, o. 1, 211, 17-22 17 Exprerimental Evaluation of a Dedicated Pinhole SPECT System for Small Animal Imaging and Scintimammography S. David
More information2/14/2019. Nuclear Medicine Artifacts. Symmetric energy windows
Nuclear Medicine Artifacts SCPMG Medical Imaging Technology & Informatics Medical Physics Group Brian Helbig, MS, DABR 1 2 Symmetric energy windows 3 1 Dynamic clinical study Energy peak shift Electrical
More informationInternational Journal of Scientific & Engineering Research, Volume 4, Issue 9, September ISSN
International Journal of Scientific & Engineering Research, Volume 4, Issue 9, September-013 06 Evaluating the effect of acquisition parameters on image quality and acquisition time with SPECT using collimator
More informationMethod for digital particle spectrometry Khryachkov Vitaly
Method for digital particle spectrometry Khryachkov Vitaly Institute for physics and power engineering (IPPE) Obninsk, Russia The goals of Analog Signal Processing Signal amplification Signal filtering
More informationI. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA
FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA EDWARD L. NICKOLOFF DEPARTMENT OF RADIOLOGY COLUMBIA UNIVERSITY NEW YORK, NY ACCEPTANCE TESTING GOALS PRIOR TO 1st CLINICAL USAGE
More informationHardware for High Energy Applications 30 October 2009
Paper No. 003 09 Hardware for High Energy Applications 30 October 2009 This document was created by the Federal Working Group on Industrial Digital Radiography. Reproduction is authorized. Federal Working
More informationGAMMA-GAMMA CORRELATION Latest Revision: August 21, 2007
C1-1 GAMMA-GAMMA CORRELATION Latest Revision: August 21, 2007 QUESTION TO BE INVESTIGATED: decay event? What is the angular correlation between two gamma rays emitted by a single INTRODUCTION & THEORY:
More informationORTEC Experiment 13. Gamma-Gamma Coincidence with Angular Correlation. Equipment Required
ORTEC Experiment 13 Equipment Required Two 905-3 2-in. x 2-in. NaI(Tl) Scintillation Detector Assemblies. Two 266 Photomultiplier Tube Bases. Two 113 Scintillation Preamplifiers. Two 556 High Voltage Power
More informationX-RAY BACKSCATTER IMAGING: PHOTOGRAPHY THROUGH BARRIERS
Copyright JCPDS-International Centre for Diffraction Data 2006 ISSN 1097-0002 X-RAY BACKSCATTER IMAGING: PHOTOGRAPHY THROUGH BARRIERS 13 Joseph Callerame American Science & Engineering, Inc. 829 Middlesex
More informationPh 3324 The Scintillation Detector and Gamma Ray Spectroscopy
Ph 3324 The Scintillation Detector and Gamma Ray Spectroscopy Required background reading Attached are several pages from an appendix on the web for Tipler-Llewellyn Modern Physics. Read the section on
More informationGet more from your images with Symphony Image Processing
DIRECT RADIOGRAPHY The user-friendly DelWorks image acquisition and processing software provides a wide range of tools for a variety of image enhancements. Its user interface simplifies every step of the
More informationVoltage Dividers & Electronics Scintillation detectors usually employ a Voltage Divider (VD) network to operate the PMT. This sometimes called "bleeder network" defines a potential (voltage) difference
More informationDigital Imaging CT & MR
Digital Imaging CT & MR January 22, 2008 Digital Radiography, CT and MRI generate images in a digital format What is a Digital Image? A digital image is made up of picture elements, pixels row by column
More information1 Purpose of This Lab Exercise:
Physics 4796 - Experimental Physics Temple University, Spring 2010-11 C. J. Martoff, Instructor J. Tatarowicz, TA Physics 4796 Lab Writeup Hunting for Antimatter with NaI Spectroscopy 1 Purpose of This
More informationA new operative gamma camera for Sentinel Lymph Node procedure
A new operative gamma camera for Sentinel Lymph Node procedure A physicist device for physicians Samuel Salvador, Virgile Bekaert, Carole Mathelin and Jean-Louis Guyonnet 12/06/2007 e-mail: samuel.salvador@ires.in2p3.fr
More informationDESIGN AND OPERATION OF A WIDE RANGE SEGMENTED GAMMA RAY SCANNING ASSAY INSTRUMENT FOR THE MEASUREMENT OF BOTH LOW AND INTERMEDIATE LEVEL WASTE
11-A-424-INMM DESIGN AND OPERATION OF A WIDE RANGE SEGMENTED GAMMA RAY SCANNING ASSAY INSTRUMENT FOR THE MEASUREMENT OF BOTH LOW AND INTERMEDIATE LEVEL WASTE John A. Mason, Marc R. Looman, Robert A. Price
More informationImage Processing (EA C443)
Image Processing (EA C443) OBJECTIVES: To study components of the Image (Digital Image) To Know how the image quality can be improved How efficiently the image data can be stored and transmitted How the
More informationLunar Technology Advantages
Lunar Technology Advantages DXA stands for Dual-Energy X-ray Absorptiometry. It is a measurement method that uses the differences in the absorption of high energy and low energy X-ray photons by different
More informationX-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto
X-ray Imaging PHYS Lecture Carlos Vinhais Departamento de Física Instituto Superior de Engenharia do Porto cav@isep.ipp.pt Overview Projection Radiography Anode Angle Focal Spot Magnification Blurring
More informationLaBr 3 :Ce scintillation gamma camera prototype for X and gamma ray imaging
8th International Workshop on Radiation Imaging Detectors Pisa 2-6 July 2006 LaBr 3 :Ce scintillation gamma camera prototype for X and gamma ray imaging Roberto Pani On behalf of SCINTIRAD Collaboration
More informationCAEN. Electronic Instrumentation. CAEN Silicon Photomultiplier Kit
CAEN Tools for Discovery Electronic Instrumentation CAEN Silicon Photomultiplier Kit CAEN realized a modular development kit dedicated to Silicon Photomultipliers, representing the state-of-the art in
More informationPulse Shape Analysis for a New Pixel Readout Chip
Abstract Pulse Shape Analysis for a New Pixel Readout Chip James Kingston University of California, Berkeley Supervisors: Daniel Pitzl and Paul Schuetze September 7, 2017 1 Table of Contents 1 Introduction...
More informationChemistry 985. Some constants: q e 1.602x10 19 Coul, ɛ x10 12 F/m h 6.626x10 34 J-s, c m/s, 1 atm = 760 Torr = 101,325 Pa
Chemistry 985 Fall, 2o17 Distributed: Mon., 17 Oct. 17, 8:30AM Exam # 1 OPEN BOOK Due: 17 Oct. 17, 10:00AM Some constants: q e 1.602x10 19 Coul, ɛ 0 8.854x10 12 F/m h 6.626x10 34 J-s, c 299 792 458 m/s,
More informationApplication of the HICAM Camera for Imaging of Prompt Gamma Rays in Measurements of Proton Beam Range
Application of the HICAM Camera for Imaging of Prompt Gamma Rays in Measurements of Proton Beam Range R. Peloso 1, P. Busca 1, A.Celani 2, C. Fiorini 1, I.Perali 1, M. Basilavecchia 2, T. Frizzi 2, J.
More informationAdvanced Materials Research Vol
Advanced Materials Research Vol. 1084 (2015) pp 162-167 Submitted: 22.08.2014 (2015) Trans Tech Publications, Switzerland Revised: 13.10.2014 doi:10.4028/www.scientific.net/amr.1084.162 Accepted: 22.10.2014
More informationCharacterization of a 64 Channel PET Detector Using Photodiodes for Crystal Identification *
Characterization of a 64 Channel PET Detector Using Photodiodes for Crystal Identification * J. S. Huber, Member, IEEE, W.W. Moses, Senior Member, IEEE, S.E. Derenzo, Senior Member, IEEE, M.H. Ho, M.S.
More informationTime-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud
University of Groningen Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you
More informationReal Time Pulse Pile-up Recovery in a High Throughput Digital Pulse Processor
Real Time Pulse Pile-up Recovery in a High Throughput Digital Pulse Processor Paul A. B. Scoullar a, Chris C. McLean a and Rob J. Evans b a Southern Innovation, Melbourne, Australia b Department of Electrical
More informationApplication Note. ipix A Gamma imager to support various applications. Introduction. An easy to carry and deploy instrument
Application Note ipix A Gamma imager to support various applications Introduction ipix is a unique gamma imager that quickly locates low level radioactive sources from a distance and estimates the dose
More informationChromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC
Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC R. Bellazzini a,b, G. Spandre a*, A. Brez a, M. Minuti a, M. Pinchera a and P. Mozzo b a INFN Pisa
More information