1 Internal Medicine Imaging Core Emory University Department of Medicine 1

2 OPERATION OF THE ZEISS LSM 510 META YOU MUST SIGN UP TO USE THE MICROSCOPE OR COMPUTER EVERY TIME NO EXCEPTIONS Before attempting to use the microscope: you should read information on fixatives for biological material, on photobleaching and anti-fade agents, and on mounting media so that your samples are well-prepared and artifacts are minimized. See Confocal Notebook under the file Preparation of Sample. It is best to grow specimens on glass slides or coverslips. you should also know the excitation and emission spectra for your particular fluorochromes so that you can select the proper configurations available in the software. See Confocal Notebook under the file Selection of Fluorochromes and Selection of Filters. The Zeiss LSM 510 META has three lasers with the following emission lines for excitation: (1) a BLUE DIODE LASER: 405 nm; (2) an ARGON LASER: 458, 477, 488 and 514 nm; (3) a GREEN HeNe LASER: 543 nm; and (4) a RED HeNe LASER: 633 nm. you should familiarize yourself with the choices of objectives on the microscope and the meanings of the markings on the objectives (e.g., the magnification, the numerical aperture (NA), the coverslip thickness requirements). See Confocal Notebook under the file Objectives to learn these markings and their meaning. A Zeiss LSM 510 META OPERATING MANUAL is available on the book shelf to read for more detailed discussions of the microscope operation. Also available are the revised edition of HANDBOOK OF BIOLOGICAL CONFOCAL MICROSCOPY edited by James B. Pawley for reading on the theory of confocal imaging systems (a newer 2 nd edition is available from the Emory Woodruff Library) and CONFOCAL MICROSCOPY FOR BIOLOGISTS by Alan Hibbs. A good introduction to light microscopy is OPTIMIZING LIGHT MICROSCOPY FOR BIOLOGICAL AND CLINICAL LABORATORIES by Barbara Foster available on book shelf (Hilenski personal copy, so please return). A good introduction to image acquisition is North, A.J Seeing is believing? A beginners guide to practical pitfalls in image acquisition. J Cell Biol 172:9-18. Copies are also available in the Imaging Core. What users should consider purchasing: Mounting Media Fluoromount G # (Electron Microscopy Sciences, P.O. Box 251, 321 Morris Road, Ft. Washington, PA 19034; Phone ; Vectashield #H-1000 (Vector Laboratories, Inc., 30 Ingold Road, Burlingame, CA 94010; Phone (10 ml=$45.00 as of 8/2001) 2

3 SlowFade Antifade Kit # S-2828 (Molecular Probes, Eugene, OR) Glass Coverslips (# 1.5) Glass Slides Blank CDs and DVDs (preferably Taiyo Yuden brand) Clear Fingernail Polish (Wet n Wild Clear Nail Protector is recommended other brands contain ethyl acetate/acetone which will lead to loss of fluorescence signal) Glass Bottom 35 mm Dishes (MatTek Corporation, 200 Homer Ave., Ashland, MA 01721; Phone ; FAX ) (used for viewing live cells/living tissue in a fluid) Chambered coverglasses (available from Applied Scientific at or for part # s AS-4851 one-chamber; AS two- chambers; AS-4854 four-chambers; AS-4858 eight-chambers) (used for viewing live cells) NOTE: Petri dishes are NOT optically clear and can only be used for low magnification (10X, 20X) work. Standard chamber slides will NOT work as they have glass slide bottoms which are too thick to be imaged through. IMPORTANT NOTE: Please read the contents of the folder on Digital Image Manipulation containing information from Nature (Guide for Digital Images), Journal of Cell Biology and the Microscopy Society of America on the acquisition, storage and image processing of digital images taken on the Zeiss LSM 510 META system. We advise you to save your original digital images exactly as they were acquired and to record your instrument settings. BEFORE USING THE SYSTEM, YOU MUST CONTACT: LU HILENSKI (WMB 329, , lhilens@emory.edu) OR DEBBY MARTINSON (WMB 303, , dmart05@emory.edu) FOR A TRAINING SESSION, YOU WILL BE ASSIGNED A USERNAME AND PASSWORD FOR LOGIN. 3

4 The following instructions are taken from the Zeiss LSM 510 META Operating Manual and from the Wright Cell Imaging Facility ( STARTING THE HARDWARE: 1. Turn on the EXFO X-Cite lamp power switch. The lamp indicator will illuminate and the display will flash for the warm up period of 90 seconds. NOTE: The unit should not be turned off unless the lamp has been on for a minimum of 5 minutes. 2. Sign in the login sheet with your starting time and the hours shown on the X-Cite lamp. 3. Using the REMOTE CONTROL, switch on System/PC to provide power to the microscope and computer. To switch on the system completely, switch on Components to initialize the LSM software. REMOTE CONTROL SWITCH 4

5 STARTING THE SOFTWARE 1. Log on to the computer using the log on and password provided to you by the Imaging Core staff. 2. Double click the LSM 510 icon on the desktop to start the LSM software. The LSM 510 Switchboard will appear on the screen.. 3. Click Scan New Images in the LSM 510 Switchboard window. For analysis of existing images, please use the confocal workstation in WMB 306 to keep the confocal microscope free for acquiring images. 4. Click on Start Expert Mode in the LSM 510 Switchboard window. Hardware Initialization window will appear briefly. 5

6 5. The LSM 510-Expert Mode Main menu appears on the screen. CREATING A DATABASE FOR ACQUIRED IMAGES 1. Click on the File button in the Main menu toolbar. 2. Click on the New button in the File subordinate toolbar. The Create New Database window appears. 3. Select drive D: from pull down menu. Create a new folder with your name. If you have a folder already, create new file if needed. TURNING ON THE LASERS 1. Click on the Acquire button on the Main menu to open the Acquire subordinate toolbar. 2. Click the Laser button to open the Laser Control button. 3. Click the appropriate Laser Unit. Click on Standby button for the argon and 405 nm lasers and the On button for the HeNe s. WAIT 5 MINUTES for the argon laser to become ready. 4. When status is Ready, click On button for the argon and 405 nm lasers. 5. For the argon laser, set Output (%) so the Tube Current is ~6 A (~45% of the output power). 6

7 SETTING THE MICROSCOPE 1. The VIS, TV and LSM buttons switch the beam path and indicate which beam path has been set. 2. Click the VIS button on the Acquire subordinate toolbar to set the microscope for direct observation via the eyepieces of the binocular tube. 3. Click on the Micro button in the Acquire subordinate toolbar to open the Microscope Control window. 4. Select an objective by clicking on the pop-up menu. The selected objective will automatically move into the beam path. NOTE: Do not use air-objectives after an immersion objective without wiping the immersion liquid (water or oil) from the specimen and objectives. 5. Select a fluorescence filter block by clicking on the Reflector or by clicking on the Microscope Settings buttons. 7

8 6. The X-Cite lamp can be shuttered on and off by the FL button on the body of the microscope on the right hand-side. 7. Once you have found and focused your specimen, redirect the light path by clicking on the LSM button in the LSM software. The microscope will automatically change the filter-position and shutter the light sources. CONFIGURING THE BEAM PATH AND LASERS 1. Click on the LSM button in the Acquire subordinate toolbar for laser scanning. 2. Click on the Config button on the Acquire subordinate toolbar. 3. Click on either Single or Multi Track. Use Single for single, double and triple labeling, simultaneous scanning only. Advantage: faster image acquisition. Disadvantage: cross talk between channels. Use Multi Track for double and triple labeling, sequential scanning, line by line or frame or frame. Advantage: when one track is active, only one detector and one laser are switched on, reducing cross-talk. Disadvantage: slower image acquisition. 4. Click Config button in the Configuration Control window to use one of the pre-programmed track configurations. 8

9 5. Load the configuration from the Channel Mode Configurations that matches your fluorophores (referred to as tracks ). Click on the dropdown box and select a desired configuration, click Apply, then Close. IF YOU CANNOT SEE THE APPROPRIATE CONFIGURATION FOR YOUR FLUOROPHORES, CONTACT THE IMAGING CORE STAFF. SCANNING AN IMAGE 1. Click on the Scan button in the Acquire subordinate toolbar to open the Scan Control window. 2. Set the parameters for scanning by selecting Mode. 9

10 3. Select Frame, Speed (scan speed 8 usually produces good results), dynamic range of 8 or 12 bit in the Pixel Depth, Scan Direction & Scan Average. NOTE: Publication quality images should be acquired using 12 bit data depth (4096 gray levels). 12 bit is also recommended when doing quantitative measurements or when imaging low fluorescence intensities. HOWEVER, 12 bit files are LARGE. 4. Set the scan averaging to increase the signal/noise ratio. IMAGE ACQUISITION 1. Select Channels in the Scan Control window. For each channel, click on the colored channel button, and then click 1 (for 1 Airy Unit). 2. For each channel, click on the colored channel button, and then click 1. This sets the Pinhole size to 1 (Airy unit) for best results. Pinhole adjustment changes the Optical Slice. NOTE: When collecting multi channel images, adjust the pinholes so that each channel has the same Optical Slice. Increase the pinholes of the shorter wavelength channels to match the longest wavelength. Do not go below 1 Airy unit. This is important for colocalization studies. 3. Start scanning using one of the Find, Fast XY, Single or Cont buttons. 4. Click Stop button to stop the current scan. IMAGE OPTIMIZATION 1. If using Multi Track, it is easier to set the imaging parameters for each Track individually. Switch off each track with the checkbox along side it. Highlight the chosen track. 10

11 2. Optimizing the confocal settings is best done with the image pseudocolored to highlight saturated and zero value pixels. Click Palette on the Image window tool bar and the Color Palette window will open. 3. Click Range Indicator and close the window. If the scanned image is too bright, it appears red=saturation (maximum). If the image is not bright enough, it appears blue=zero (minimum). 4. While acquiring an image with Fast Scan, adjust Ampl Offset (setting the min signal) so that only a few of the pixels in the background are blue. 5. Set Detector Gain and Excitation power (setting the max signal). Low laser power causes less bleaching but requires the detector gain to be set high, which introduces noise. You may have to adjust the Ampl Offset until all blue pixels disappear, and then make it slightly positive. 6. Reduce the Detector Gain until the red pixels only just disappear. 7. Leave Ampl Gain as 1 unless you have a very dim signal and nothing else works. This will amplify noise as well as signal. REMEMBER: A few red speckles; a few blue speckles in each channel. 8. After you have optimized the first channel, stop scanning by clicking Stop. 9. Turn off the optimized channel in the Configuration Control window. Turn on the next channel and optimize these settings for this channel. 10. Once you are satisfied that each channel detector is set optimally, ensure that all Tracks are checked. 11. Go to Scan Control window and click the Mode button. 11

12 12. Change the Frame size by clicking on the Optimal button. Noise can be reduced by averaging a number of frames and slowing the scan speed. Try the following: Mode: Line Method: Mean Number: 4 Scan Speed: Change the image Palette back to No Palette. 14. Click Start. 15. Save your final image to a database. Click Save as and save image to an existing database (Open MDB button) or create a new database (New MDB button). Do not select Compress file option. 12

13 SCANNING A Z STACK 1. Select Z Stack in the Scan Control window. 2. Using Fast scan, focus up and down with the microscope focus wheel to ensure your specimen is in the center of the field of view and the microscope is focused in the middle of the specimen. 3. Click Stop acquisition. 4. Click Mark First/Last on the Z Settings panel. Click on the XY cont button. 5. Use the focusing drive of the microscope to focus on the upper position of the specimen area where the Z stack is to start. 6. Click on the Mark First button to set the upper position of the Z stack. 7. Focus on the lower specimen area where the recording of the Z stack is to end. 13

14 8. Click on the Mark Last button to set this lower position. 9. Click on the X:Y:Z button to set the Z interval in such a way that the voxel has identical dimensions in the X-, Y-, and Z- directions (cube). 10. Click on Z Slice in the Scan Control window to bring up the Optical Slice dialog. 11. Click the Optimal Interval button. Check that the optical section for each channel is the same. If your optical sections are different for each wavelength, go to Mode/Channels and adjust the Pinhole so that each channel has the same optical section around 1 Airy Unit, but one channel will have to be slightly larger. 12. Click on the Start button to start the recording of the Z stack. You can check progress by selecting Gallery button on the image window sidebar. 14

15 Z ATTENUATION COMPENSATION 1. In the Scan Control: Z Settings window, click on Auto Z Corr. 2. While in Fast XY scan mode, focus to near the top of the sample where it is brightest. You should have already set the detector and laser settings for this slice. If not, go to the Scan Control: Channels window and set the detector gains and laser intensity to get adequate signal at this focus position. 3. In the Scan Control: Z Settings, click Set A. 4. While in the Fast XY scan, manually focus to near the bottom of the Z series. 5. Go to the Scan Control: Channels window and set the detector gains and laser intensity to get adequate signal. 6. Click Set B button in the Scan Control: Z Settings window. 7. Click Start to begin the Z series acquisition. 15

16 SWITCHING OFF THE SYSTEM 1. Click on File button in the Main menu and then click on the Exit button to exit the LSM software program. 2. To turn off the lasers, click Acquire, then click Laser button in the subtoolbar. 3. In the Laser Control window, click Off for each laser (or to Standby if someone else is using the system within the next two hours). 4. Remove specimen and clean microscope. Wipe off oil from objective using the LENS PAPER provided. 5. Move to a low power objective (10X or 20X). 6. Cover the microscope with the blue Zeiss cover avoiding the hot lamp housing. 7. Tidy up the workspace! 8. Burn your data to CD or DVD. 9. If you are the last one to use the system for the day, wait until the argon laser fan has switched off. 10. Switch off the Components and System/PC on the Remote Control. 11. Switch off the EXFO X-Cite lamp. Record the end time and the lamp hours on the sign-up sheet. OPENING YOUR IMAGES OFFLINE 1. You may use the free Zeiss Image Browser available from: Follow link in lower right hand corner: Free LSM Image Browser 2. You may use the full version of the software LSM Image Examiner on the computer in WMB

LSM 510 Meta Training Notes

LSM 510 Meta Training Notes LSM 510 Meta Training Notes Turning on the system Turn on X-Cite power supply. This supplies light for epifluorescence for viewing your samples through the microscope. Turn on the remote control switch.

More information

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling Zeiss LSM 510 Confocor III Training Notes Center for Cell Analysis & Modeling Confocor 3 Start Up Go to System Module Turn on Main Switch, System/ PC, and Components Switches Do you need the arc lamp?

More information

LSM 510 Training Notes

LSM 510 Training Notes LSM 510 Training Notes Turning on the system Turn on the arc lamp, found on the bench top left of the microscope. This supplies light for epifluorescence for viewing your samples through the microscope.

More information

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s LSM 5 MP, LSM 510 and LSM 510 META M i c r o s c o p y f r o m C a r l Z e i s s Quick Guide Laser Scanning Microscopes LSM Software ZEN 2007 August 2007 We make it visible. Contents Page Contents... 1

More information

Training Guide for Carl Zeiss LSM 510 META Confocal Microscope

Training Guide for Carl Zeiss LSM 510 META Confocal Microscope Training Guide for Carl Zeiss LSM 510 META Confocal Microscope AIM 4.2 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn ON Components and System/PC switches

More information

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope AIM 4.2 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Verify that main power switches on the

More information

ZEISS LSM510META confocal manual

ZEISS LSM510META confocal manual ZEISS LSM510META confocal manual Switching on the system 1) Switch on the Remote Control button located on the table to the right of the microscope. This is the main switch for the whole system including

More information

Operating Instructions for Zeiss LSM 510

Operating Instructions for Zeiss LSM 510 Operating Instructions for Zeiss LSM 510 Location: GNL 6.312q (BSL3) Questions? Contact: Maxim Ivannikov, maivanni@utmb.edu 1 Attend A Complementary Training Before Using The Microscope All future users

More information

LSM 780 Confocal Microscope Standard Operation Protocol

LSM 780 Confocal Microscope Standard Operation Protocol LSM 780 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Sign on log sheet according to Actual start time 2. Check Compressed Air supply for the air table 3. Switch

More information

Zeiss 880 Training Notes Zen 2.3

Zeiss 880 Training Notes Zen 2.3 Zeiss 880 Training Notes Zen 2.3 1 Turn on the HXP 120V Lamp 2 Turn on Main Power Switch Turn on the Systems PC Switch Turn on the Components Switch. 3 4 5 Turn on the PC and log into your account. Start

More information

LSM 710 Confocal Microscope Standard Operation Protocol

LSM 710 Confocal Microscope Standard Operation Protocol LSM 710 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Switch on Main power switch 2. Switch on System / PC power button 3. Switch on Components power button 4.

More information

Things to check before start-up.

Things to check before start-up. Byeong Cha Page 1 11/24/2009 Manual for Leica SP2 Confocal Microscope Enter you name, the date, the time, and the account number in the user log book. Things to check before start-up. Make sure that your

More information

Zeiss LSM 880 Protocol

Zeiss LSM 880 Protocol Zeiss LSM 880 Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

Microscopy from Carl Zeiss

Microscopy from Carl Zeiss Microscopy from Carl Zeiss Contents Page Contents... 1 Introduction... 1 Starting the System... 2 Introduction to ZEN Efficient Navigation... 5 Setting up the microscope... 10 Configuring the beam path

More information

CONFOCAL MICROSCOPE (Zeiss LSM 510 META v4.2)

CONFOCAL MICROSCOPE (Zeiss LSM 510 META v4.2) Wellcome Trust Centre for Human Genetics Molecular Cytogenetics and Microscopy Core CONFOCAL MICROSCOPE (Zeiss LSM 510 META v4.2) 1) STARTING THE SYSTEM Abridged INSTRUCTIONS Switch on the mercury bulb

More information

Zeiss LSM 780 Protocol

Zeiss LSM 780 Protocol Zeiss LSM 780 Protocol 1) System Startup F Please note the sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge for unused time.

More information

Zeiss 780 Training Notes

Zeiss 780 Training Notes Zeiss 780 Training Notes Turn on Main Switch, System PC and Components Switches 780 Start up sequence Do you need the argon laser (458, 488, 514 nm lines)? Yes Turn on the laser s main power switch and

More information

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL MIF ZEISS LSM510 CONFOCAL USER PROTOCOL START-UP Turn on the Mercury Bulb Power Supply (if needed). Power-on the Control Box. Turn on the computer. Open the LSM 510 software. Choose Scan New Images and

More information

Contents. Introduction

Contents. Introduction Contents Page Contents... 1 Introduction... 1 Starting the System... 2 Introduction to ZEN Efficient Navigation... 5 Setting up the microscope... 10 Configuring the beam path and lasers... 12 Scanning

More information

Microscope Confocal LSM510 META

Microscope Confocal LSM510 META Microscope Confocal LSM510 META Welcome to the Zeiss LSM 510 Meta Confocal tutorial. Before using the LSM 510 META, Log off any other computer that is open with your personal login. You will need to put

More information

OPERATING INSTRUCTIONS

OPERATING INSTRUCTIONS Zeiss LSM 510 M eta Confocal M icroscope OPERATING INSTRUCTIONS Starting the System: 1. Turn the black knob on the laser box one-quarter turn from Off to On. You will hear the laser cooling mechanisms

More information

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014 CIAN LSM1 or LSM2 short instructions, version 1.4, September 2014 page 1 of 6 SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014 Before starting To work with LSM1

More information

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO TRAINING MANUAL Multiphoton Microscopy LSM 510 META-NLO September 2010 Multiphoton Microscopy Training Manual Multiphoton microscopy is only available on the LSM 510 META-NLO system. This system is equipped

More information

Practical work no. 3: Confocal Live Cell Microscopy

Practical work no. 3: Confocal Live Cell Microscopy Practical work no. 3: Confocal Live Cell Microscopy Course Instructor: Mikko Liljeström (MIU) 1 Background Confocal microscopy: The main idea behind confocality is that it suppresses the signal outside

More information

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center For any questions or concerns, please contact: Linda Nieman lnieman@mgh.harvard.edu Office: (617) 643-9684 Cell: (512) 565-8076 Chenyue

More information

Zeiss LSM 510 Multiphoton Confocal Microscope

Zeiss LSM 510 Multiphoton Confocal Microscope Zeiss LSM 510 Multiphoton Confocal Microscope Quick Start User Guide LSU Health Sciences Research Core Facility Table of Contents 1 Safety... Page 3 2 Turn On the System... Page 4 3 Start Up the ZEN Software.

More information

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging Bi/BE 227 Winter 2016 Assignment #3 Adding the third dimension: 3D Confocal Imaging Schedule: Jan 20: Assignment Jan 20-Feb 8: Work on assignment Feb 10: Student PowerPoint presentations. Goals for this

More information

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009 Introduction of Fluoresence Confocal Microscopy The first confocal microscope was invented by Princeton

More information

Guide to Confocal 5. Starting session

Guide to Confocal 5. Starting session Guide to Confocal 5 Remember that when booking and before starting session you can check for any problems at https://www.bris.ac.uk/biochemistry/uobonly/cif/index.html Starting session Switch on microscope

More information

Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope

Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope ZEN 2009 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn Chameleon TiS laser key from Standby

More information

Training Guide for Leica SP8 Confocal/Multiphoton Microscope

Training Guide for Leica SP8 Confocal/Multiphoton Microscope Training Guide for Leica SP8 Confocal/Multiphoton Microscope LAS AF v3.3 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Turn ON power switch for epifluorescence

More information

Zeiss LSM 510 Multiphoton Confocal Microscope

Zeiss LSM 510 Multiphoton Confocal Microscope Zeiss LSM 510 Multiphoton Confocal Microscope User Guide LSU Health Sciences Center-Shreveport Research Core Facility Table of Contents 1 Safety... Page 3 2 Turn On the System... Page 4 3 Start Up the

More information

TRAINING MANUAL. Olympus FV1000

TRAINING MANUAL. Olympus FV1000 TRAINING MANUAL Olympus FV1000 September 2014 TABLE OF CONTENTS A. Start-Up Procedure... 1 B. Visual Observation under the Microscope... 1 C. Image Acquisition... 4 A brief Overview of the Settings...

More information

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center For any questions or concerns, please contact: Linda Nieman lnieman@mgh.harvard.edu Office: (617) 643-9684 Cell: (512) 565-8076 Chenyue

More information

Leica SP8 TCS Users Manual

Leica SP8 TCS Users Manual Version : 07/08/0 Leica SP8 TCS Users Manual Start up:. Turn the PC Microscope, Scanner Power, Laser Power, and the Laser Emission key to on (bottom right of desk).. Turn on the fluorescent lamp (top left

More information

Leica SP8 TCS Users Manual

Leica SP8 TCS Users Manual Leica SP8 TCS Users Manual Follow the procedure for start up and log on as posted in the lab. Please log on with your account only and do not share your password with anyone. We track and confirm usage

More information

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5. Smith College August 2010 Operating Checklist for using the Laser Scanning Confocal Microscope Leica TCS SP5. CONTENT, page no. Startup, 1 Initial set-up, 1 Software, 2 Microscope Specimen observation

More information

Training Guide for Carl Zeiss LSM 880 with AiryScan FAST

Training Guide for Carl Zeiss LSM 880 with AiryScan FAST Training Guide for Carl Zeiss LSM 880 with AiryScan FAST ZEN 2.3 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2018) Power ON Routine 1 2 Turn ON Main Switch from the remote control

More information

b. Turn the power switch and key to on position for blue laser.

b. Turn the power switch and key to on position for blue laser. OLYMPUS FLUOVIEW 300 CONFOCAL MICOSCOPE OPERATION PROCEDURE 1. Turn ON microscope in this order: 1) Turn on mercury lamp (Note: once the mercury lamp is turned off, DO NOT turn it back on for at least

More information

ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL

ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL START THE SYSTEM... 2 START ZEN SOFTWARE... 3 SET THE TEMPERATURE AND THE CO2 CONTROLLERS... OBSERVATION AT OCULARS... 5 STATIF PRESENTATION... 6 ACQUIRE ONE

More information

LSM 800 Confocal Microscope Standard Operation Protocol

LSM 800 Confocal Microscope Standard Operation Protocol LSM 800 Confocal Microscope Standard Operation Protocol Turning on the system 1. Switch on the Main switch (labeled 1 and 2 ) mounted on the wall. 2. Turn the Laser Key (labeled 3 ) 90 clockwise for power

More information

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software:

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software: Confocal imaging on the Leica TCS SP8 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software: 4) Do not touch the microscope while the software is initializing. Choose your options: Turn

More information

Microscope Confocal Sp2 Upright.

Microscope Confocal Sp2 Upright. Microscope Confocal Sp2 Upright. Welcome to the Leica Sp2 Confocal Upright tutorial. Before using the Sp2 Invert, You will need to put down your name on the reservation system = http://svintranet.epfl.ch/index.php?optio

More information

Overview. About other software. Administrator password. 58. UltraVIEW VoX Getting Started Guide

Overview. About other software. Administrator password. 58. UltraVIEW VoX Getting Started Guide Operation 58. UltraVIEW VoX Getting Started Guide Overview This chapter outlines the basic methods used to operate the UltraVIEW VoX system. About other software Volocity places great demands on the computer

More information

Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev. Microscopy course, Michmoret Dec 2005

Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev. Microscopy course, Michmoret Dec 2005 Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev Why use confocal microscopy? Principles of the laser scanning confocal microscope. Image resolution. Manipulating the

More information

Leica SPEII confocal microscope. Short Manual

Leica SPEII confocal microscope. Short Manual Leica SPEII confocal microscope Short Manual Switching ON sequence: 1. Turn on the Workstation under the bench (top, far right). 2. Turn on the Supply Unit - Laser box (big green switch first and then

More information

Title: Leica SP5 Confocal User Manual

Title: Leica SP5 Confocal User Manual Title: Leica SP5 Confocal User Manual Date of first issue: 23/10/2015 Date of review: Version: Admin For assistance or to report an issue Office: CG07 or 05 Email: Igmm-imaginghelpdesk@igmm.ed.ac.uk Website:

More information

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017.

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017. Nikon AZ100 Laser Scanning Macro Confocal Microscope Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin May 2017 Contents 1 Introduction 2 2 Hardware - Startup 2 3 Software/Operation 4 3.1 Multidimensional

More information

Leica SP8 Resonant Confocal. Quick-Start Guide

Leica SP8 Resonant Confocal. Quick-Start Guide Leica SP8 Resonant Confocal Quick-Start Guide Contents Start-up Preparing for Imaging Part 1 On the scope Part 2 Software interface Part 3 Heat & CO2 incubation Part 4 Other hardware options Shut-down

More information

Boulevard du Temple Daguerrotype (Paris,1838) a busy street? Nyquist sampling for movement

Boulevard du Temple Daguerrotype (Paris,1838) a busy street? Nyquist sampling for movement Boulevard du Temple Daguerrotype (Paris,1838) a busy street? Nyquist sampling for movement CONFOCAL MICROSCOPY BioVis Uppsala, 2017 Jeremy Adler Matyas Molnar Dirk Pacholsky Widefield & Confocal Microscopy

More information

Leica Sp5 II Confocal User Guide

Leica Sp5 II Confocal User Guide Leica Sp5 II Confocal User Guide Turning on the Confocal System (instructions are posted in the room) 1. Turn on Laser Power Button 2. Turn Key to On position 3. Turn on Scanner Power Button 4. Turn on

More information

Quick Start Guide. Leica SP5 X

Quick Start Guide. Leica SP5 X Quick Start Guide Leica SP5 X Please note: Some of the information in this guide was taken from Leica Microsystems Leica TCS SP5 LAS AF Guide for New Users. This work is licensed under the Creative Commons

More information

REMEMBER: You have 5GB of disk space on this microscope. Check before you start if you have room for your experiment. If not delete your old data.

REMEMBER: You have 5GB of disk space on this microscope. Check before you start if you have room for your experiment. If not delete your old data. 1 Use of the Zeiss LSM 510 Inverted Firstly please be aware that this microscope should be treated with respect and care at all times. Rules of use: This Microscope can only be used by Masters by Research

More information

Cell Biology and Bioimaging Core

Cell Biology and Bioimaging Core Cell Biology and Bioimaging Core Leica TCS SP5 Operating Instructions Starting up the instrument 1. First, log in the log book located on the confocal desk. Include your name, your lab s PI, an account

More information

3. are adherent cells (ie. cells in suspension are too far away from the coverslip)

3. are adherent cells (ie. cells in suspension are too far away from the coverslip) Before you begin, make sure your sample... 1. is seeded on #1.5 coverglass (thickness = 0.17) 2. is an aqueous solution (ie. fixed samples mounted on a slide will not work - not enough difference in refractive

More information

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope Nikon Eclipse Ti A1-A Confocal Operating Manual Start-up 1. Turn on Excite Fluorescent light power supply- metal halide. a. Cool down as for mercury bulb b. Wheel closed liquid light guide 2. Turn on power

More information

Zeiss LSM880 Operating Instructions. UTMB Optical Microscopy Core Jan. 16, 2018

Zeiss LSM880 Operating Instructions. UTMB Optical Microscopy Core Jan. 16, 2018 Zeiss LSM880 Operating Instructions UTMB Optical Microscopy Core Jan. 16, 2018 1 1. Power up the microscope Sing the LOGBOOK Steps below will provide power to the computer and all of the microscope components.

More information

Nikon. King s College London. Imaging Centre. N-SIM guide NIKON IMAGING KING S COLLEGE LONDON

Nikon. King s College London. Imaging Centre. N-SIM guide NIKON IMAGING KING S COLLEGE LONDON N-SIM guide NIKON IMAGING CENTRE @ KING S COLLEGE LONDON Starting-up / Shut-down The NSIM hardware is calibrated after system warm-up occurs. It is recommended that you turn-on the system for at least

More information

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each lens with a 1 Airy unit pinhole setting. Many water lenses

More information

LEICA TCS SP5 AOBS TANDEM USER MANUAL

LEICA TCS SP5 AOBS TANDEM USER MANUAL LEICA TCS SP5 AOBS TANDEM USER MANUAL STARTING THE SYSTEM...2 THE LAS AF SOFTWARE...3 THE «ACQUIRE» MENU...5 CHOOSE AND CREATE A SETTING...6 THE CONTROL PANEL...8 THE DMI6000B MICROSCOPE...10 ACQUIRE ONE

More information

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL START-UP On the Switchbox, turn both black switches to the ON position. Wait for the microscope to boot up completely (watch the screen on the side of the microscope).

More information

3 Choose the Channels button and set the Channel Settings. Set the Pinhole to 1 Airy unit.

3 Choose the Channels button and set the Channel Settings. Set the Pinhole to 1 Airy unit. 1 Set up the confocal light path for imaging a green dye (e.g. Alexa488-EGFP). For example, under the Configuration Control window the light path could be set up as shown here using the 488 nm LASER (found

More information

1 Co Localization and Working flow with the lsm700

1 Co Localization and Working flow with the lsm700 1 Co Localization and Working flow with the lsm700 Samples -1 slide = mousse intestine, Dapi / Ki 67 with Cy3/ BrDU with alexa 488. -1 slide = mousse intestine, Dapi / Ki 67 with Cy3/ no BrDU (but with

More information

Nikon C1si Spectral Laser Scanning Confocal Microscope. User Guide

Nikon C1si Spectral Laser Scanning Confocal Microscope. User Guide Nikon C1si Spectral Laser Scanning Confocal Microscope User Guide Contents: C1Si Turn-On/ShutDown Procedures... 2 Overview... 4 Setup for epi-illumination to view through the eyepieces:... 5 Setup for

More information

Nikon SIM-E & A1-R System

Nikon SIM-E & A1-R System Nikon SIM-E & A1-R System USER GUIDE LSU Health Sciences Center Shreveport Research Core Facility June 01 2017 Chaowei Shang 1 Table of Content 1. Start Up the System... Page 3 Hardware and microscope

More information

ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide

ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide Matyas Molnar, Biovis 2016 Starting the microscpe 1. Check the microscope if everything looks clean and normal. If not, report it in the logbook.

More information

Zeiss LSM 510 META Guide

Zeiss LSM 510 META Guide Zeiss LSM 50 META Guide Ge#ng Started. Ensure the main power is switched on (black knob on the far le; of the laser module box).. Turn on the System/PC and Components switches (located on the grey switch

More information

Using the Nikon TE2000 Inverted Microscope

Using the Nikon TE2000 Inverted Microscope Wellcome Trust Centre for Human Genetics Molecular Cytogenetics and Microscopy Core Using the Nikon TE2000 Inverted Microscope Fluorescence image acquisition using Scanalytic s IPLab software and the B&W

More information

Nikon A1R. Multi-Photon & Laser Scanning Confocal Microscope. Kyle Marchuk Adam Fries Jordan Briscoe Kaitlin Corbin. April 2017.

Nikon A1R. Multi-Photon & Laser Scanning Confocal Microscope. Kyle Marchuk Adam Fries Jordan Briscoe Kaitlin Corbin. April 2017. Nikon A1R Multi-Photon & Laser Scanning Confocal Microscope Kyle Marchuk Adam Fries Jordan Briscoe Kaitlin Corbin April 2017 Contents 1 Introduction 2 2 Start-Up 2 3 Imaging 4 3.1 Sample Alignment...........................................

More information

Zeiss Axio Imager.A1 manual

Zeiss Axio Imager.A1 manual Zeiss Axio Imager.A1 manual Power-up protocol 1. Mercury lamp 2. Power strip on shelf 3. Computer The Mercury lamp should always be first-on and last-off. This prevents any electrical surges caused by

More information

Leica TCS SL Confocal Training. Neuroscience Imaging Core Staff. Core Director. Facility Manager

Leica TCS SL Confocal Training. Neuroscience Imaging Core Staff. Core Director. Facility Manager Leica TCS SL Confocal Training Neuroscience Imaging Facility The Ohio State University Rightmire Hall 614-292-1367 Staff Core Director Anthony Brown, Ph. D. 060 Rightmire Hall 614-292-1205 brown.2302@osu.edu

More information

ScanArray Overview. Principle of Operation. Instrument Components

ScanArray Overview. Principle of Operation. Instrument Components ScanArray Overview The GSI Lumonics ScanArrayÒ Microarray Analysis System is a scanning laser confocal fluorescence microscope that is used to determine the fluorescence intensity of a two-dimensional

More information

Diskovery Spinning Disk Guide

Diskovery Spinning Disk Guide Diskovery Spinning Disk Guide qbi.microscopy@uq.edu.au Getting started The microscope and its peripherals (Fig. 1a) should always be turned on, but if they are not, turn them on in the following way: 1.

More information

Zeiss Axiovert 135 Fluorescence Microscope Quick Guide / Operations Manual (v. 1.0 February 09)

Zeiss Axiovert 135 Fluorescence Microscope Quick Guide / Operations Manual (v. 1.0 February 09) University of Chicago Integrated Light Microscopy Core Dr. Vytas Bindokas, Director http://digital.bsd.uchicago.edu By: Christine Labno, Assistant Director Room: AB-129 Phone: 4-9040 Zeiss Axiovert 135

More information

Nikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol

Nikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol Nikon Eclipse Ti-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol Please sign on the log sheet before switching on system. Turn on system Turn on A only if confocal mode or laser

More information

Leica TCS SP8 Quick Start Guide

Leica TCS SP8 Quick Start Guide Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, Fluorescent Light for the microscope stand. 2. Turn on the Scanner Power (1) on the front

More information

Leica TCS SP2 User Manual 1.1. Markus Friedrich and James Lee. Department of Biological Sciences Wayne State University

Leica TCS SP2 User Manual 1.1. Markus Friedrich and James Lee. Department of Biological Sciences Wayne State University Leica TCS SP2 User Manual 1.1 Markus Friedrich and James Lee Department of Biological Sciences Wayne State University Confocal Manual 1 Helpful hints The confocal room is cold. Bring a sweater. Confocaling

More information

Title: Nikon A1R Confocal User Manual

Title: Nikon A1R Confocal User Manual Title: Nikon A1R Confocal User Manual Date of first issue: 23/10/2015 Date of review: Version: Admin For assistance or to report an issue Office: CG.07 or CG.05 Email: Igmm-imaginghelpdesk@igmm.ed.ac.uk

More information

AxioVision 4.5 Brightfield Image Capture Procedure

AxioVision 4.5 Brightfield Image Capture Procedure AxioVision 4.5 Brightfield Image Capture Procedure 1. STARTING-UP PROCEDURE: Remove blue dust cover and place on shelf under microscope. Turn on the halogen lamp by pushing the switch at the back right

More information

Supplemental Method Information Zeiss LSM710

Supplemental Method Information Zeiss LSM710 Supplemental Method Information Zeiss LSM710 1 Under the Light Path window set up the confocal for imaging a green dye (Alexa488-EGFP). For example, set up the light path as shown here using the 488 nm

More information

The Zeiss AiryScan System, Confocal Four.

The Zeiss AiryScan System, Confocal Four. The Zeiss AiryScan System, Confocal Four. Overview. The Zeiss AiryScan module is a segmented, radially stacked GaASP detector and collector system designed to subsample the airy disk of a point emission

More information

Leica Confocal - 2. Instructions for Leica SP2 Confocal Equipped with Visible Laser Lines

Leica Confocal - 2. Instructions for Leica SP2 Confocal Equipped with Visible Laser Lines Leica Confocal - 2 Instructions for Leica SP2 Confocal Equipped with Visible Laser Lines Imaging Technology Group Revised as of 08/2011 by Dianwen Zhang (Email: zhangdw@illinois.edu) Revision History Authors

More information

Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope

Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope Improved Optics More Lasers 405 diode 440 diode 488 Argon 515 Argon 559

More information

Olympus IX71 Microscope and DP71 Camera Instructions

Olympus IX71 Microscope and DP71 Camera Instructions Olympus IX71 Microscope and DP71 Camera Instructions Microscopy in Medicine (MiM) Core Emory University Department of Medicine 1 Olympus IX71 Image Capture Procedure 2 3 1. STARTING-UP PROCEDURE: Remove

More information

Zeiss Deconvolution Microscope: A Quick Guide

Zeiss Deconvolution Microscope: A Quick Guide Zeiss Deconvolution Microscope: A Quick Guide Start-up Uncover microscope. Do not put dust cover on the floor. Plug in both cameras. The default camera is the AxioCam HRm (monochrome camera) for fluorescence

More information

Confocal 510 Tutorial

Confocal 510 Tutorial Confocal 510 Tutorial You will have to log on in order to use the microscope. You will be charged according the time you are logged in, so please don t forget to log out after you are done. If you don

More information

Simplified Instructions: Olympus Widefield Microscope S1230

Simplified Instructions: Olympus Widefield Microscope S1230 Contents General Microscope Operation Simple Image Capture Multi-Wavelength Capture Z-Series Timelapse Combining Capture Modes Synopsis of Other Functions Pages 2-23 24-40 41-47 48-56 57-59 60-68 69-83

More information

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING Operations Guide Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING Nikon Eclipse 90i Operations Guide STARTUP Startup Powering Up Fluorescence

More information

CMI STANDARD OPERATING PROCEDURE. Fluoview 300 laser scanning confocal microscope

CMI STANDARD OPERATING PROCEDURE. Fluoview 300 laser scanning confocal microscope CMI STANDARD OPERATING PROCEDURE Fluoview 300 laser scanning confocal microscope CMI documentid:sop001 CONTACT INFORMATION: Peter Owens: 091 494036 (office) Peter.owens@nuigalway.ie Kerry Thompson: 091

More information

Brief manual how to start and close the Leica sp2 Confocal. (TCS SP2 AOBS system mounted on a DM IRE2)

Brief manual how to start and close the Leica sp2 Confocal. (TCS SP2 AOBS system mounted on a DM IRE2) Brief manual how to start and close the Leica sp2 Confocal (TCS SP2 AOBS system mounted on a DM IRE2) A. Switching on hardware B. Acquiring and saving images C. Switching off the microscope D. Good working

More information

1 Set up the confocal light path for imaging a green dye (Alexa488-EGFP). For example, the

1 Set up the confocal light path for imaging a green dye (Alexa488-EGFP). For example, the 1 Set up the confocal light path for imaging a green dye (Alexa488-EGFP). For example, the light path as shown here using the 488 nm LASER (Laser Unit 1) reflecting off of the 405/488 nm Dichroic mirror

More information

Nikon A1Rsi Confocal Start-Up Sequence

Nikon A1Rsi Confocal Start-Up Sequence 1. Turn the key on the Nikon LUN-V Laser Launch. Nikon A1Rsi Confocal Start-Up Sequence 2. Press the button the left side of the A1Rsi Controller unit. 3. Turn on the power strip underneath the microscope.

More information

Zeiss AxioImager.Z2 Brightfield Protocol

Zeiss AxioImager.Z2 Brightfield Protocol Zeiss AxioImager.Z2 Brightfield Protocol 1) System Startup Please note put sign-up policy. You must inform the facility at least 24 hours beforehand if you can t come; otherwise, you will receive a charge

More information

Leica TCS SP8 Quick Start Guide

Leica TCS SP8 Quick Start Guide Leica TCS SP8 Quick Start Guide Leica TCS SP8 System Overview Start-Up Procedure 1. Turn on the CTR Control Box, EL6000 fluorescent light source for the microscope stand. 2. Turn on the Scanner Power

More information

Nikon Ti-E Microscope Manual. Rightmire Hall Ohio State University. Director: Tony Brown Rightmire

Nikon Ti-E Microscope Manual. Rightmire Hall Ohio State University. Director: Tony Brown Rightmire Nikon Ti-E Microscope Manual Rightmire Hall Ohio State University Director: Tony Brown Rightmire 060 292-1205 brown.2302@osu.edu Facility Manager: Paula Monsma Rightmire 062 293-0939 292-1367 monsma.1@osu.edu

More information

Nikon E800 Operating Instructions.

Nikon E800 Operating Instructions. Nikon E800 Operating Instructions. You can request electronic copies of this manual by contacting lshats@jhsph.edu Copies are also available on the JHU MMI Department web site. Please send your comments

More information

Multifluorescence The Crosstalk Problem and Its Solution

Multifluorescence The Crosstalk Problem and Its Solution Multifluorescence The Crosstalk Problem and Its Solution If a specimen is labeled with more than one fluorochrome, each image channel should only show the emission signal of one of them. If, in a specimen

More information

Nikon E800 Microscope. Operating Instructions

Nikon E800 Microscope. Operating Instructions Nikon E800 Microscope Operating Instructions B Watson 12/2005 Table of contents: 1. The Nikon E800 Microscope 2. Turning the system ON and OFF 3. Selecting the light path 4. Operating in transmitted light

More information

Confocal Laser Scanning Microscopy

Confocal Laser Scanning Microscopy Name of the Core Facility: Confocal Laser Scanning Microscopy CORE Forschungszentrum Immunologie Mainz Welcome to the CSLM Core Facility: The CLSM Core Facility enables working groups to incorporate high

More information

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer Page 1 of 11 Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer The Aramis Raman system is a software selectable multi-wavelength Raman system with mapping capabilities with a 400mm monochromator and

More information