Evaluation of MWT Materials Accusorb MRI Shield. Dr. E. Kanal, Department of Radiology, UPMC. Wednesday, April 21, 2010

Size: px
Start display at page:

Download "Evaluation of MWT Materials Accusorb MRI Shield. Dr. E. Kanal, Department of Radiology, UPMC. Wednesday, April 21, 2010"

Transcription

1 Evaluation of MWT Materials Accusorb MRI Shield Dr. E. Kanal, Department of Radiology, UPMC Wednesday, April 21, 2010 Background: In almost all x-ray and ultrasonographic imaging examinations, in order to diagnostically examine patient anatomy and/or physiology, external energies are transmitted into patients in a manner that is restricted to the anatomy being examined. Receivers then detect modifications of this transmitted energy induced by the varying anatomic structures the energy passes through. These modifications are then analyzed and utilized to provide anatomic and/or physiologic information regarding the anatomy that had been probed by these transmitted energies. Due to the resonant manner in which magnetic resonance imaging (MRI) is executed, large volumes of patient anatomy are irradiated with radiofrequency (RF) energies even if the anatomy being studied is restricted to a very small percentage thereof. Spatially localized anatomic and physiologic information is then gleaned, via resonance phenomenology, from what is typically only a small fraction of the tissue that had initially undergone RF irradiation and its related power deposition. Thus, the risks associated with the potential thermal sequelae of RF irradiation are experienced by ALL the tissue of the patient that was irradiated, even if the volume for which diagnostic information was acquired was actually only much smaller than that which had been irradiated for the study. This holds true for all aspects and types of MRI examinations, including but not limited to magnetic resonance angiography, magnetic resonance spectroscopy, functional MRI, diffusion weighted MRI, perfusion weighted MRI, etc. It also applies regardless of the MR imaging sequence utilized, be it gradient echo or spin echo or inversion recovery variants or fast/turbo spin echo or echo planar sequences, etc. Practically speaking, since RF power deposition generally falls off very rapidly as one leaves the volume contained by the RF transmitter coil(s), in MRI we are essentially RF irradiating the entire volume contained within the RF transmitter coils even if we are only studying a very localized anatomic region of interest contained within it. The rationale for this is relatively straightforward: in ultrasound, the ultrasonographic beam energy can be directed and/or focused specifically to the anatomic region to be examined. In CT the X-ray beam can be restricted or collimated to the precise anatomy undergoing diagnostic evaluation. An X-ray beam can be nominally narrowed to 1 mm thickness to produce diagnostic information regarding a 1 mm thick anatomic slice in the human being studied. Such is not the case with MR-related imaging, where resonance phenomena within the volume that had been irradiated produce diagnostic information about the slice or sub-volume to be examined. This is because there have been no effective RF collimators available to restrict the transmitted RF energies to the specific anatomy to be evaluated and no more. Therefore, since the total amount of RF power absorbed by the patient is itself proportional to the total amount of tissue exposed to these

2 RF energies, the patients find themselves in the unusual situation of absorbing far more RF exposure than is actually necessary for the requested diagnostic evaluation. Especially in this day in which the Image Gently campaign is so strongly encouraged among all radiologists for patients of all ages, this seems to be quite inefficient and wasteful, exposing all MR patients to more or far more RF energies and their related (especially thermal) risks than is actually needed for their requested study. In order to successfully restrict transmitted RF energies to just the anatomic region to be examined, the equivalent of a safe RF collimator needs to be developed. This material would need to markedly and efficiently restrict RF energies from passing through them (such as by reflection and/or absorption) while at the same time successfully dissipating any absorbed RF energies in a manner that does not in and of itself pose any threat to the patient or RF transmitting (or MRI-related) hardware. Early in the development of the MRI industry there had been several early attempts at developing such RF shields, but their efficacy and/or safety was not felt to be sufficient or successful and they are not used in clinical or research MR environments today. To this end, we researched material that was initially designed to be used to help absorb radar energies to assist the military in enabling objects to produce a much attenuated radar signature so-called stealth technology. This technology works primarily by absorbing, attenuating, and safely dissipating microwave (and RF) energies to which it is exposed. After discussions with the manufacturer of this material confirmed that these characteristics should work with RF energies of the frequencies utilized in 1.5 Tesla MR scanning environments, we elected to pursue evaluation of custom designed RF shields for human application to be used in MR imaging environments to attempt to restrict RF power deposition to the anatomic regions of interest diagnostically desired - and no more. This report summarizes our experience with thermal testing of the RF Shield material provided by MWT. In vitro testing: Heating measurements were obtained and recorded using a Luxtron m3300 fluoroptic thermography unit with True Temp software. Testing was performed utilizing an acrylic head/torso phantom filled with polyacrylic acid mixture, 2.7 grams per liter of distilled filtered water. The test device used for heating testing was as per the FDA design. This is a insulated 0.5 mm thick wire with a 1 cm length of insulation stripped from each end. This implant was suspended within the gel-filled torso phantom oriented parallel to the long axis of the phantom and RF bore and situated at isocenter but far left laterally within the torso portion of the phantom. Fiber optic probes were placed at the tip of the device, in the middle of the exposed metal at one end of the wire/device, and at the juncture of the exposed wire/insulation at one end of the device. The fourth probe was placed within the gel itself at a position remote from the device but within the RF irradiated volume of the transmitting body RF coil (Figure 1).

3 Figure 1 After acquiring 3-plane scout scans, an RF aggressive pulse sequence was prescribed resulting in estimated specific absorption rates (SAR) of 2 W/kg average total body weight with peak SAR values of 4 W/kg. Specific scan parameters utilized were as follows: Scan parameters were as follows: Axial, fast spin echo TE=14ms, TR=900ms, ETL=64 FOV=48cmx48cm 36 slices, 10mm with a 2.5mm gap NEX=20 Matrix=128x128 Scan Time=24:11 EST SAR=1.9986, Peak SAR= The first test consisted of scanning the device with the RF shield completely surrounding the torso section of the phantom containing the embedded device, as illustrated in Figure 2:

4 Figure 2 This was compared to the temperature changes observed during imaging of the same phantom and device with the identical scan parameters and scan time duration with no RF shield present. With no RF shield present significant, roughly 17 degree Celsius increases in surface temperatures were observed during MR imaging at the tip of the implanted device as seen in Figure 3:

5 Figure 3 (We attribute the abrupt modification of the temperature reading for the tip probe to variation in probe tip positioning that initially occurred during scanning) Compare this to the results observed with the RF shield completely surrounding the device within the scanned phantom, where no significant temperature modification was seen in any probe/lead, as shown in Figure 4:

6 Figure 4 The RF shield was then superiorly displaced so that it only covered half the superoinferior extent of the scanned device/phantom, as shown in figure 5: Figure 5 Scanning at this time demonstrated device heating to as high as 14 degrees Celsius, which was between that which was seen with no RF shield present versus that which was seen when the device was fully RF shielded, as seen in Figure 6:

7 Figure 6 We then studied the effectiveness of the RF shield when only draped over the entire phantom and device, not circumferentially entirely surrounding it, as demonstrated in Figure 7:

8 Figure 7 Device heating was far more restricted here (figure 8), with maximal tip heating observed to only increase by roughly 6 degrees Celsius. Figure 8 We then tested the imaging impact of the presence of the RF shield. Recognizing that the MR imaging process requires precise magnitudes of RF energies to reach the tissue to be examined, we examined the impact on the images acquired of the tissue phantom with versus without the presence of the surrounding RF shield when positioned superoinferiorly half way down the SI length of the torso phantom. These results are illustrated in images 9 and 10, where the RF attenuating properties of the RF shield are clearly evident:

9 Figure 9 No RF Shield present

10 Figure 10 Same MR imaging sequence, RF shield present surrounding the superior half of the torso phantom. As a result of these in vitro testing results we elected to proceed with testing thermal and imaging effects of the utilization of the RF shield on human imaging. Since one of the the targeted theoretical objectives of such RF shielding material might be to shield cardiac pacemakers and pacemaker leads from transmitted RF energies to diminish arhythmogenic and thermal possible risks of such exposures, it was elected to perform our first human imaging tests on MR imaging of the cervical spine region. We wanted to ensure that diagnostic MR images of the cervical spine could still be obtained while severely attenuating RF energies that would reach the thoracic cage. We therefore designed our next research protocol to focus on targeted MR imaging of the human

11 cervical spine with and without the presence of the RF shielding material surrounding the thorax of these research subjects. Purpose: To assess the RF shielding capabilities of the MWT RF shield while assessing thermal impact on the patient and shielding material. We also assessed the impact on the diagnostic content of the MR images of the cervical spine as well as the rapidity of signal attenuation once reaching the (thoracic) volume encircled by the RF shield. Methods: A total of seven research subjects were studied with IRB approval. For each research subject studied the patient underwent the typical, clinically requested MR imaging examination of the cervical spine in one of our 1.5 Tesla General Electric MR scanners as per the routine clinical imaging protocol. This included a positioning scout sequence followed by sagittal spin echo T1 weighted imaging, sagittal fast spin echo T2 weighted imaging, axial spoiled gradient echo imaging, and three dimensional gradient echo axial imaging. For each research subject, four Luxtron fiber optic temperature probes were utilized for all temperature measurements, with measurements taken at each probe tip at a frequency of 1 Hz. During the portion of the exam where the patient/research subject was not wearing the RF shield, probes 1 and 2 were placed directly on patient s skin, with probe 1 over the manubrium and probe 2 positioned on the left side of neck just inferior to angle of the mandible. Probes 3 and 4 were placed on patient s gown with lead 3 at the level of xyphoid and lead 4 inferior to lead 3 at a location midway between the xyphoid and the umbilicus. The times correlating to the beginning (of the pre-scan phase) and end of each MR imaging sequence were recorded. Upon completion of the clinical portion of the examination the probes were removed and the Accusorb RF shield was placed around the patient s chest in such a way as to ensure that the so anterior and posterior edge of the vest was positioned as superiorly as possible without covering/obscuring the anatomic structures of the cervical spine. The subject was then re-positioned in a manner matching as precisely as possible that of the prior, clinically indicated study just completed. The fiber optic probes were then repositioned on the subject with probes 1 and 2 placed in the identical locations as before and again contacting the patient s skin directly. Probe 1 was again placed at level of the manubrium, ensuring that the superiormost extent of the RF shield covered the probe tip. Probe 2 was again positioned on left side of the neck in such a way that the probe tip was NOT covered by the RF shield even in part. Probe 3 was positioned directly on the anterosuperiormost edge of the shield, located slightly superior to probe tip 1 s position. Probe 4 was placed directly on the left anterior collar surface of the RF shield, somewhat inferior to probe tip 2. The MR examination (and all associated imaging parameters) executed with the presence of the RF shield was in every way identical to that performed without the presence of the RF shield in place. Scan start/stop time data for each sequence was again recorded in same manner as it had been for the clinical scan portion of the examination.

12 Thermal data were analyzed to compare temperatures measured at the probe tips with the RF shield in place relative to that experienced without the presence of the RF shield. The RF shield serves as another layer of clothing, as it wrapped around the chest as would be a vest, and thus some minimal elevation of temperature underneath the shield was expected relative to sampling without the shield present. The study was performed with the vest off and with the vest on in all instances (the order of which was decided by for patient, technologist, and study convenience and efficiency). This research project was performed with the approval of the Institutional Review Board of the University of Pittsburgh. Results: A total of seven patients agreed to serve as research subjects for this study. Of these, there was inconsistent probe tip positioning/dislodgement during the study in one patient (the first studied in this protocol) in whom there was also inconsistent positioning of the study before and after RF shield placement. As such we discarded all data from this research subject. Thus, there is complete data available for six research subjects. Of all the studies performed on each research subject, the greatest degree of temperature elevation observed in the RF shield material itself was in a single research subject (# ) where the temperature at the RF shield edge increased by 3.21 degrees Celsius over the course of a 5 minutes and six second study. In this sequence, 3 degrees Celsius of temperature elevation had already been reached by 3 minutes and fifteen seconds into the study after which the measured temperature essentially leveled off despite continued RF power deposition at the same levels and rates throughout the entire study. This suggested that the RF shield material was able to absorb and dissipate this level of RF imaging power without undue heating of the surfaces or edges of the RF shield. In all other cases the temperature elevations observed on the patient s skin was considered trivial and consistent with physiologic temperature changes that one might observe when lying quietly while wearing a sweater or similar such outer garment. No patient/research subject thermal complaints were reported and no focal hot spots were palpated or observed on the skin or the RF shield.

13

14

15

16

17

18 Imaging of the cervical spine demonstrated very sharp cut-off of signal from the superior edge of the RF shield when this was included in the displayed field of view (100% of all cases in the sagittal plane). This provided for diagnostic MR imaging in the anatomic region of interest not covered by the RF shield while rendering the anatomy uninterpretable in the regions (intentionally) enveloped by the RF shielding material. Signal drop off to half of initial signal intensities were within one to two centimeters in all cases (superoinferiorly).

19

20

21

22 Conclusions: The data have confirmed that the RF thermal loads produced by routine clinical MR imaging of the cervical spine were capably absorbed and dissipated by the RF shielding material tested in a manner that posed no safety threat to the patient. Further, imaging signal intensities were tightly restricted to the volume not enveloped by the RF shield material, thus providing further evidence of the successful RF shielding capabilities of this material. It should therefore be possible to significantly decrease total patient RF energy deposition from routine MR imaging studies by encasing the anatomy which the diagnostician does NOT want to examine (yet might still be located be within the RF excitation/transmission coil) in such RF shielding material, thus effectively collimating the RF energies to the anatomic regions of interest to be examined. Further, in the case of implanted leads, wires, or electrically conductive devices, it might well be possible to significantly reduce thermal concerns and the potential for focusing of RF energies via an antenna effect at the tips of these electrically conductive materials by shielding these implants with such RF shielding material if they are located in or near anatomy that is not designated for study yet might still otherwise be exposed to significant RF excitation energies in the course or routine MR imaging.

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging Medical Imaging X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging From: Physics for the IB Diploma Coursebook 6th Edition by Tsokos, Hoeben and Headlee And Higher Level Physics 2 nd Edition

More information

High Field MRI: Technology, Applications, Safety, and Limitations

High Field MRI: Technology, Applications, Safety, and Limitations High Field MRI: Technology, Applications, Safety, and Limitations R. Jason Stafford, Ph.D. The University of Texas M. D. Anderson Cancer Center, Houston, TX Introduction The amount of available signal

More information

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia M R I Physics Course Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia M R I Physics Course Magnetic Resonance Imaging Spatial

More information

Beam-Restricting Devices

Beam-Restricting Devices Beam-Restricting Devices Three factors contribute to an increase in scatter radiation: Increased kvp Increased Field Size Increased Patient or Body Part Size. X-ray Interactions a some interact with the

More information

Potential Risks of MRI in Device Patients

Potential Risks of MRI in Device Patients Outline Potential Risks of MRI in Device Patients Redha Boubertakh r.boubertakh@qmul.ac.uk MRI and cardiac implantable electronic devices (CIED) Components of an MRI scanner MRI implant and device safety

More information

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions TOPICS: Computed Tomography Quick Overview CT Dosimetry Effects of CT

More information

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010 (N)MR Imaging Lab Course Script FMP PhD Autumn School Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder Date: November 3rd, 2010 1 Purpose: Understanding the basic principles of MR imaging

More information

Advanced MSK MRI Protocols at 3.0T. Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University

Advanced MSK MRI Protocols at 3.0T. Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University Advanced MSK MRI Protocols at 3.0T Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University Outline Why High Field for MSK? SNR and Relaxation Times Technical Issues Example

More information

Health Issues. Introduction. Ionizing vs. Non-Ionizing Radiation. Health Issues 18.1

Health Issues. Introduction. Ionizing vs. Non-Ionizing Radiation. Health Issues 18.1 Health Issues 18.1 Health Issues Introduction Let s face it - radio waves are mysterious things. Especially when referred to as electromagnetic radiation the concept makes many people nervous. In this

More information

Lesson 06: Pulse-echo Imaging and Display Modes. These lessons contain 26 slides plus 15 multiple-choice questions.

Lesson 06: Pulse-echo Imaging and Display Modes. These lessons contain 26 slides plus 15 multiple-choice questions. Lesson 06: Pulse-echo Imaging and Display Modes These lessons contain 26 slides plus 15 multiple-choice questions. These lesson were derived from pages 26 through 32 in the textbook: ULTRASOUND IMAGING

More information

Radio Frequency Field

Radio Frequency Field Radio Frequency Field Radio Frequency Coils and RF Power Distribution RF Coil Maps Distribution of RF Power GE 750W RF maps courtesy of Tobias Gilk Siemens Prisma (courtesy Siemens) Radio Frequency Field

More information

Analysis of spatial dependence of acoustic noise transfer function in magnetic resonance imaging

Analysis of spatial dependence of acoustic noise transfer function in magnetic resonance imaging Analysis of spatial dependence of acoustic noise transfer function in magnetic resonance imaging Award: Magna Cum Laude Poster No.: C-1988 Congress: ECR 2014 Type: Scientific Exhibit Authors: T. Hamaguchi,

More information

2015 Spin echoes and projection imaging

2015 Spin echoes and projection imaging 1. Spin Echoes 1.1 Find f0, transmit amplitudes, and shim settings In order to acquire spin echoes, we first need to find the appropriate scanner settings using the FID GUI. This was all done last week,

More information

Analysis of magnetic and electromagnetic field emissions produced by a MRI device

Analysis of magnetic and electromagnetic field emissions produced by a MRI device Sept. 8-1, 21, Kosice, Slovakia Analysis of magnetic and electromagnetic field emissions produced by a MRI device D. Giordano, M. Borsero, G. Crotti, M. ucca INRIM Istituto Nazionale di Ricerca Metrologica,

More information

Pulse Sequence Design and Image Procedures

Pulse Sequence Design and Image Procedures Pulse Sequence Design and Image Procedures 1 Gregory L. Wheeler, BSRT(R)(MR) MRI Consultant 2 A pulse sequence is a timing diagram designed with a series of RF pulses, gradients switching, and signal readout

More information

Experience in implementing continuous arterial spin labeling on a commercial MR scanner

Experience in implementing continuous arterial spin labeling on a commercial MR scanner JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 6, NUMBER 1, WINTER 2005 Experience in implementing continuous arterial spin labeling on a commercial MR scanner Theodore R. Steger and Edward F. Jackson

More information

MR Basics: Module 8 Image Quality

MR Basics: Module 8 Image Quality Module 8 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.

More information

MRI SYSTEM COMPONENTS Module One

MRI SYSTEM COMPONENTS Module One MRI SYSTEM COMPONENTS Module One 1 MAIN COMPONENTS Magnet Gradient Coils RF Coils Host Computer / Electronic Support System Operator Console and Display Systems 2 3 4 5 Magnet Components 6 The magnet The

More information

MICROWAVE & RF RADIATION: (RFR Information - Technology Newsletter, Full Version)

MICROWAVE & RF RADIATION: (RFR Information - Technology Newsletter, Full Version) MICROWAVE & RF RADIATION: (RFR Information - Technology Newsletter, Full Version) George M. Harris, P.E. (February, 2011) Questions: -What is Microwave & Radiofrequency, (RF), Radiation? -What are its

More information

Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare

Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare GE Healthcare Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare There is excitement across the industry regarding the clinical potential of a hybrid

More information

Numerical Assessment of Specific Absorption Rate in the Human Body Caused by NFC Devices

Numerical Assessment of Specific Absorption Rate in the Human Body Caused by NFC Devices Second International Workshop on Near Field Communication Numerical Assessment of Specific Absorption Rate in the Human Body Caused by NFC Devices S. Cecil, G. Schmid, K. Lamedschwandner EMC&Optics Seibersdorf

More information

Cardiac MR. Dr John Ridgway. Leeds Teaching Hospitals NHS Trust, UK

Cardiac MR. Dr John Ridgway. Leeds Teaching Hospitals NHS Trust, UK Cardiac MR Dr John Ridgway Leeds Teaching Hospitals NHS Trust, UK Cardiac MR Physics for clinicians: Part I Journal of Cardiovascular Magnetic Resonance 2010, 12:71 http://jcmr-online.com/content/12/1/71

More information

TITLE: Prostate Cancer Detection Using High-Spatial Resolution MRI at 7.0 Tesla: Correlation with Histopathologic Findings at Radical Prostatectomy

TITLE: Prostate Cancer Detection Using High-Spatial Resolution MRI at 7.0 Tesla: Correlation with Histopathologic Findings at Radical Prostatectomy Award Number: W81XWH-11-1-0253 TITLE: Prostate Cancer Detection Using High-Spatial Resolution MRI at 7.0 Tesla: Correlation with Histopathologic Findings at Radical Prostatectomy PRINCIPAL INVESTIGATOR:

More information

Measurement of Implant Electrode Leads using Time-Domain Reflectometry to Predict the Resonant Length for MRI Heating

Measurement of Implant Electrode Leads using Time-Domain Reflectometry to Predict the Resonant Length for MRI Heating Measurement of Implant Electrode Leads using Time-Domain Reflectometry to Predict the Resonant Length for MRI Heating Steven McCabe Department of Engineering The University of Waikato Hamilton, New Zealand

More information

H 2 O and fat imaging

H 2 O and fat imaging H 2 O and fat imaging Xu Feng Outline Introduction benefit from the separation of water and fat imaging Chemical Shift definition of chemical shift origin of chemical shift equations of chemical shift

More information

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of radiation to the population due to Medical Imaging

More information

MRI Metal Artifact Reduction

MRI Metal Artifact Reduction MRI Metal Artifact Reduction PD Dr. med. Reto Sutter University Hospital Balgrist Zurich University of Zurich OUTLINE Is this Patient suitable for MR Imaging? Metal artifact reduction Is this Patient suitable

More information

1.5T HIGH FIELD SMALL ANIMAL MRI

1.5T HIGH FIELD SMALL ANIMAL MRI 1.5T HIGH FIELD SMALL ANIMAL MRI Designed Specifically for Veterinarians TECHNICAL GUIDE ADVANCING THE ART AND SCIENCE OF VETERINARY MRI The PetVet is the only high-field MRI system designed specifically

More information

Nuclear Associates , , CT Head and Body Dose Phantom

Nuclear Associates , , CT Head and Body Dose Phantom Nuclear Associates 76-414,76-414-4150,76-415 CT Head and Body Dose Phantom Users Manual March 2005 Manual No. 76-414-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product

More information

2014 M.S. Cohen all rights reserved

2014 M.S. Cohen all rights reserved 2014 M.S. Cohen all rights reserved mscohen@g.ucla.edu IMAGE QUALITY / ARTIFACTS SYRINGOMYELIA Source http://gait.aidi.udel.edu/res695/homepage/pd_ortho/educate/clincase/syrsco.htm Surgery is usually recommended

More information

X-RAYS - NO UNAUTHORISED ENTRY

X-RAYS - NO UNAUTHORISED ENTRY Licencing of premises Premises Refer Guidelines A radiation warning sign and warning notice, X-RAYS - NO UNAUTHORISED ENTRY must be displayed at all entrances leading to the rooms where x-ray units are

More information

Ultrasound Bioinstrumentation. Topic 2 (lecture 3) Beamforming

Ultrasound Bioinstrumentation. Topic 2 (lecture 3) Beamforming Ultrasound Bioinstrumentation Topic 2 (lecture 3) Beamforming Angular Spectrum 2D Fourier transform of aperture Angular spectrum Propagation of Angular Spectrum Propagation as a Linear Spatial Filter Free

More information

Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI

Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI Val M. Runge, MD Institute for Diagnostic and Interventional Radiology Clinics for Neuroradiology and Nuclear Medicine University Hospital Zurich

More information

MARP. MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005

MARP. MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005 ACR MRI accreditation program MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005 Carl R. Keener, Ph.D., DABMP, DABR keener@marpinc.com MARP Medical & Radiation Physics,

More information

MRI with the VNS Therapy System October 2017

MRI with the VNS Therapy System October 2017 MRI with the VNS Therapy System October 2017 The information contained in this document is one part of the full labeling for the implanted portions of the VNS Therapy System. It is not intended to serve

More information

Standards for Imaging Endpoints in Clinical Trials: Standardization and Optimization of Image Acquisitions: Magnetic Resonance

Standards for Imaging Endpoints in Clinical Trials: Standardization and Optimization of Image Acquisitions: Magnetic Resonance FDA Workshop April 13, 2010 Standards for Imaging Endpoints in Clinical Trials: Standardization and Optimization of Image Acquisitions: Magnetic Resonance Edward F. Jackson, PhD Professor and Chief, Section

More information

10. Phase Cycling and Pulsed Field Gradients Introduction to Phase Cycling - Quadrature images

10. Phase Cycling and Pulsed Field Gradients Introduction to Phase Cycling - Quadrature images 10. Phase Cycling and Pulsed Field Gradients 10.1 Introduction to Phase Cycling - Quadrature images The selection of coherence transfer pathways (CTP) by phase cycling or PFGs is the tool that allows the

More information

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS Veterinary Science Preparatory Training for the Veterinary Assistant Floron C. Faries, Jr., DVM, MS Radiology Floron C. Faries, Jr., DVM, MS Objectives Determine the appropriate machine settings for making

More information

High-Field Surface-Coil MR Imaging of Localized Anatomy

High-Field Surface-Coil MR Imaging of Localized Anatomy 181 High-Field Surface-Coil MR Imaging of Localized Anatomy John F. Schenck,' Thomas H. Foster,' John l. Henkes,' William J. Adams,' Cecil Hayes,2 Howard R. Hart, Jr.,' William A. Edelstein,' Paul A. Bottomley,'

More information

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah X-RAY IMAGING EE 472 F2017 Prof. Yasser Mostafa Kadah www.k-space.org Recommended Textbook Stewart C. Bushong, Radiologic Science for Technologists: Physics, Biology, and Protection, 10 th ed., Mosby,

More information

MR in RTP. MR Data for Treatment Planning: Spatial Accuracy Issues, Protocol Optimization, and Applications (Preview of TG117 Report) Acknowledgements

MR in RTP. MR Data for Treatment Planning: Spatial Accuracy Issues, Protocol Optimization, and Applications (Preview of TG117 Report) Acknowledgements MR Data for Treatment Planning: Issues, Protocol Optimization, and s (Preview of TG117 Report) Debra H. Brinkmann Mayo Clinic, Rochester MN Acknowledgements TG-117 Use of MRI Data in Treatment Planning

More information

Image Quality/Artifacts Frequency (MHz)

Image Quality/Artifacts Frequency (MHz) The Larmor Relation 84 Image Quality/Artifacts (MHz) 42 ω = γ X B = 2πf 84 0.0 1.0 2.0 Magnetic Field (Tesla) 1 A 1D Image Magnetic Field Gradients Magnet Field Strength Field Strength / Gradient Coil

More information

MR Advance Techniques. Flow Phenomena. Class II

MR Advance Techniques. Flow Phenomena. Class II MR Advance Techniques Flow Phenomena Class II Flow Phenomena In this class we will explore different phenomenona produced from nuclei that move during the acquisition of data. Flowing nuclei exhibit different

More information

Photomultiplier Tube

Photomultiplier Tube Nuclear Medicine Uses a device known as a Gamma Camera. Also known as a Scintillation or Anger Camera. Detects the release of gamma rays from Radionuclide. The radionuclide can be injected, inhaled or

More information

GafChromic EBT2 and EBT3 Films for Ball Cube II Phantom

GafChromic EBT2 and EBT3 Films for Ball Cube II Phantom GafChromic EBT2 and EBT3 Films for Ball Cube II Phantom Introduction: These EBT2/EBT3 films, shown in Figure 1a-c, are specially sized and formatted to uniquely fit the Accuray Ball Cube II Phantom. Each

More information

3T Unlimited. ipat on MAGNETOM Allegra The Importance of ipat at 3T. medical

3T Unlimited. ipat on MAGNETOM Allegra The Importance of ipat at 3T. medical 3T Unlimited ipat on MAGNETOM Allegra The Importance of ipat at 3T s medical ipat on MAGNETOM Allegra The Importance of ipat at 3T The rise of 3T MR imaging Ultra High Field MR (3T) has flourished during

More information

MRI Systems and Coil Technology

MRI Systems and Coil Technology MRI for Technologists MRI Systems and Coil Technology PROGRAM INFORMATION MRI for Technologists is a training program designed to meet the needs of radiologic technologists entering or working in the field

More information

Area Network Applications] Notice: This document has been prepared to assist the IEEE P It is

Area Network Applications] Notice: This document has been prepared to assist the IEEE P It is Project: IEEE P802.15 Working Group for Wireless Personal Area Networks N (WPANs) Submission Title: [RF Safety Considerations for Body Area Network Applications] Date Submitted: [] Source: [Kamya Yekeh

More information

Initial setup and subsequent temporal position monitoring using implanted RF transponders

Initial setup and subsequent temporal position monitoring using implanted RF transponders Initial setup and subsequent temporal position monitoring using implanted RF transponders James Balter, Ph.D. University of Michigan Has financial interest in Calypso Medical Technologies Acknowledgements

More information

Effect of RF Pulse Sequence on Temperature Elevation for a Given Time-Average SAR

Effect of RF Pulse Sequence on Temperature Elevation for a Given Time-Average SAR Effect of RF Pulse Sequence on Temperature Elevation for a Given Time-Average SAR ZHANGWEI WANG, 1 CHRISTOPHER M. COLLINS 2 1 GE Healthcare, Aurora, OH 44202 2 Department of Radiology and ioengineering,

More information

Background (~EE369B)

Background (~EE369B) Background (~EE369B) Magnetic Resonance Imaging D. Nishimura Overview of NMR Hardware Image formation and k-space Excitation k-space Signals and contrast Signal-to-Noise Ratio (SNR) Pulse Sequences 13

More information

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question.

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question. RA110 test 3 Multiple Choice Identify the letter of the choice that best completes the statement or answers the question. 1. An object 35 cm in width is radiographed at 100 cm SID and at a 50 cm SOD. What

More information

Multi-Access Biplane Lab

Multi-Access Biplane Lab Multi-Access Biplane Lab Advanced technolo gies deliver optimized biplane imaging Designed in concert with leading physicians, the Infinix VF-i/BP provides advanced, versatile patient access to meet the

More information

Chapter 4. Pulse Echo Imaging. where: d = distance v = velocity t = time

Chapter 4. Pulse Echo Imaging. where: d = distance v = velocity t = time Chapter 4 Pulse Echo Imaging Ultrasound imaging systems are based on the principle of pulse echo imaging. These systems require the use of short pulses of ultrasound to create two-dimensional, sectional

More information

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging Shawn D. Teague, MD DISCLOSURES 3DR- advisory committee CT PHYSICS WITH AN EMPHASIS ON APPLICATION IN THORACIC AND CARDIAC IMAGING

More information

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image Introduction Chapter 16 Diagnostic Radiology Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther In diagnostic radiology

More information

Liver imaging beyond expectations with Ingenia

Liver imaging beyond expectations with Ingenia Publication for the Philips MRI Community Issue 47 2012/3 Liver imaging beyond expectations with Ingenia Contributed by John Penatzer, RT, MR clinical product specialist, Cleveland, OH, USA Publication

More information

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London Automated dose control in multi-slice CT Nicholas Keat Formerly ImPACT, St George's Hospital, London Introduction to presentation CT contributes ~50+ % of all medical radiation dose Ideally all patients

More information

SYLLABUS. 1. Identification of Subject:

SYLLABUS. 1. Identification of Subject: SYLLABUS Date/ Revision : 30 January 2017/1 Faculty : Life Sciences Approval : Dean, Faculty of Life Sciences SUBJECT : Biophysics 1. Identification of Subject: Name of Subject : Biophysics Code of Subject

More information

MAGNETIC RESONANCE IMAGING

MAGNETIC RESONANCE IMAGING CSEE 4620 Homework 3 Fall 2018 MAGNETIC RESONANCE IMAGING 1. THE PRIMARY MAGNET Magnetic resonance imaging requires a very strong static magnetic field to align the nuclei. Modern MRI scanners require

More information

Acoustic noise reduction of MRI systems by means of magnetic shielding

Acoustic noise reduction of MRI systems by means of magnetic shielding Acoustic noise reduction of MRI systems by means of magnetic shielding D. Biloen, N.B. Roozen Philips Applied Technologies, P.O.Box 218/Bldg. SAQ 2121, 56MD Eindhoven, The Netherlands {david.biloen, n.b.roozen}@philips.com,

More information

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto X-ray Imaging PHYS Lecture Carlos Vinhais Departamento de Física Instituto Superior de Engenharia do Porto cav@isep.ipp.pt Overview Projection Radiography Anode Angle Focal Spot Magnification Blurring

More information

HETERONUCLEAR IMAGING. Topics to be Discussed:

HETERONUCLEAR IMAGING. Topics to be Discussed: HETERONUCLEAR IMAGING BioE-594 Advanced MRI By:- Rajitha Mullapudi 04/06/2006 Topics to be Discussed: What is heteronuclear imaging. Comparing the hardware of MRI and heteronuclear imaging. Clinical applications

More information

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp Dose Reduction and Image Preservation After the Introduction of a into the LODOX Statscan unit above 110 kvp Abstract: CJ Trauernicht 1, C Rall 1, T Perks 2, G Maree 1, E Hering 1, S Steiner 3 1) Division

More information

Delivering Better Patient Care with SIGNA Architect

Delivering Better Patient Care with SIGNA Architect Delivering etter Patient Care with SIGNA Architect As a regional leader in outpatient-based diagnostic imaging, the radiologists and staff at Inova Fairfax MRI Center are focused on one thing: the patient.

More information

Magnetic Resonance Imaging Principles, Methods, and Techniques

Magnetic Resonance Imaging Principles, Methods, and Techniques Magnetic Resonance Imaging Principles, Methods, and Techniques Perry Sprawls Jr., Emory University Publisher: Medical Physics Publishing Corporation Publication Place: Madison, Wisconsin Publication Date:

More information

CHAPTER 2 WIRELESS CHANNEL

CHAPTER 2 WIRELESS CHANNEL CHAPTER 2 WIRELESS CHANNEL 2.1 INTRODUCTION In mobile radio channel there is certain fundamental limitation on the performance of wireless communication system. There are many obstructions between transmitter

More information

Focused RF Hyperthermia Using Ultra-High Field MRI Joshua de Bever, PhD

Focused RF Hyperthermia Using Ultra-High Field MRI Joshua de Bever, PhD Focused RF Hyperthermia Using Ultra-High Field MRI Joshua de Bever, PhD Department of Radiology Stanford University STANFORD CANCER IMAGING TRAINEESHIP UHF Focused RF (FRF) Hyperthermia GOAL: Generate

More information

MR Basics: Module 6 Pulse Sequences

MR Basics: Module 6 Pulse Sequences Module 6 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.

More information

Detector technology in simultaneous spectral imaging

Detector technology in simultaneous spectral imaging Computed tomography Detector technology in simultaneous spectral imaging Philips IQon Spectral CT Z. Romman, I. Uman, Y. Yagil, D. Finzi, N. Wainer, D. Milstein; Philips Healthcare While CT has become

More information

Magnetic Resonance Imaging

Magnetic Resonance Imaging Magnetic Resonance Imaging Principles, Methods, and Techniques Perry Sprawls, Ph.D., FACR, FAAPM, FIOMP Distinguished Emeritus Professor Department of Radiology Emory University Atlanta, Georgia Medical

More information

160-slice CT SCANNER / New Standard for the Future

160-slice CT SCANNER / New Standard for the Future TECHNOLOGY HISTORY For over 130 years, Toshiba has been a world leader in developing technology to improve the quality of life. Our 50,000 global patents demonstrate a long, rich history of leading innovation.

More information

SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS

SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS 7:28-14.1 Scope (a) This subchapter covers therapeutic installations used in the healing arts. These therapeutic installations include x-ray, accelerator and teletherapy

More information

QC Testing for Computed Tomography (CT) Scanner

QC Testing for Computed Tomography (CT) Scanner QC Testing for Computed Tomography (CT) Scanner QA - Quality Assurance All planned and systematic actions needed to provide confidence on a structure, system or component. all-encompassing program, including

More information

RADIOFREQUENCY ELECTROMAGNETIC FIELDS

RADIOFREQUENCY ELECTROMAGNETIC FIELDS CHAPTER 19. RADIOFREQUENCY ELECTROMAGNETIC FIELDS 19.1 INTRODUCTION 19.1.1 CONTEXT The proposed buildings of the World Trade Center Memorial and Redevelopment Plan (Proposed Action) are being designed

More information

Fundamental and Clinical Studies for Effectiveness of Zero-filling Interpolation on k-space for Improvement of Sharpness in Magnetic Resonance Imaging

Fundamental and Clinical Studies for Effectiveness of Zero-filling Interpolation on k-space for Improvement of Sharpness in Magnetic Resonance Imaging Fundamental and Clinical Studies for Effectiveness of Zero-filling Interpolation on k-space for Improvement of Sharpness in Magnetic Resonance Imaging Poster No.: C-0709 Congress: ECR 2014 Type: Scientific

More information

1 Introduction. 2 The basic principles of NMR

1 Introduction. 2 The basic principles of NMR 1 Introduction Since 1977 when the first clinical MRI scanner was patented nuclear magnetic resonance imaging is increasingly being used for medical diagnosis and in scientific research and application

More information

SCINTILLATING FIBER DOSIMETER ARRAY

SCINTILLATING FIBER DOSIMETER ARRAY SCINTILLATING FIBER DOSIMETER ARRAY FIELD OF THE INVENTION [0001] This invention relates generally to the field of dosimetry and, more particularly, to rapid, high-resolution dosimeters for advanced treatment

More information

Design of an implanted compact antenna for an artificial cardiac pacemaker system

Design of an implanted compact antenna for an artificial cardiac pacemaker system Design of an implanted compact antenna for an artificial cardiac pacemaker system Soonyong Lee 1,WonbumSeo 1,KoichiIto 2, and Jaehoon Choi 1a) 1 Department of Electrical and Computer Engineering, Hanyang

More information

6 Appendix SCR Recording During FMRI Acquisition

6 Appendix SCR Recording During FMRI Acquisition 85 6 Appendix SCR Recording During FMRI Acquisition This work was conducted by Antoine Bruguier, R. McKell Carter, Christof Koch and Steven Quartz. Experiments were carried out at the Caltech Biological

More information

Focal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution

Focal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution Focal Spot Blooming in CT: We Didn t Know We Had a Problem Until We Had a Solution Cynthia H. McCollough, PhD, DABR, FAAPM, FACR Director, CT Clinical Innovation Center Professor of Medical Physics and

More information

Maximizing clinical outcomes

Maximizing clinical outcomes Maximizing clinical outcomes Digital Tomosynthesis Dual Energy Subtraction Automated Long Length Imaging Improved image quality at a low dose Xray Xray Patented ISS capture technology promotes high sensitivity

More information

Ques on (2): [18 Marks] a) Draw the atrial synchronous Pacemaker block diagram and explain its operation. Benha University June 2013

Ques on (2): [18 Marks] a) Draw the atrial synchronous Pacemaker block diagram and explain its operation. Benha University June 2013 Benha University June 2013 Benha Faculty of Engineering Electrical Department Hospital Instrumentations (E472) 4 Th year (control) Dr.Waleed Abdel Aziz Salem Time: 3 Hrs Answer the following questions.

More information

Pulse Sequence Design Made Easier

Pulse Sequence Design Made Easier Pulse Sequence Design Made Easier Gregory L. Wheeler, BSRT(R)(MR) MRI Consultant gurumri@gmail.com 1 2 Pulse Sequences generally have the following characteristics: An RF line characterizing RF Pulse applications

More information

Magnetic Resonance Imaging and Radio Frequency. Part 1. Produced on behalf of Mid Sussex Amateur Radio Society by M5BTB

Magnetic Resonance Imaging and Radio Frequency. Part 1. Produced on behalf of Mid Sussex Amateur Radio Society by M5BTB Magnetic Resonance Imaging and Radio Frequency Part 1 Produced on behalf of Mid Sussex Amateur Radio Society by M5BTB Why Now? During 2011 my physical health was deteriorating, and a brain tumour was diagnosed

More information

EE225E/BIOE265 Spring 2011 Principles of MRI. Assignment 6 Solutions. (y 0 + vt) dt. 2 y 0T + 3 )

EE225E/BIOE265 Spring 2011 Principles of MRI. Assignment 6 Solutions. (y 0 + vt) dt. 2 y 0T + 3 ) EE225E/BIOE265 Spring 211 Principles of MRI Miki Lustig Handout Assignment 6 Solutions 1. Nishimura 6.7 (Thanks Galen!) a) After the 9 y pulse, the spin is in the ˆx direction (using left-handed rotations).

More information

COMPUTED TOMOGRAPHY 1

COMPUTED TOMOGRAPHY 1 COMPUTED TOMOGRAPHY 1 Why CT? Conventional X ray picture of a chest 2 Introduction Why CT? In a normal X-ray picture, most soft tissue doesn't show up clearly. To focus in on organs, or to examine the

More information

Introduction. Parametric Imaging. The Ultrasound Research Interface: A New Tool for Biomedical Investigations

Introduction. Parametric Imaging. The Ultrasound Research Interface: A New Tool for Biomedical Investigations The Ultrasound Research Interface: A New Tool for Biomedical Investigations Shelby Brunke, Laurent Pelissier, Kris Dickie, Jim Zagzebski, Tim Hall, Thaddeus Wilson Siemens Medical Systems, Issaquah WA

More information

Philips Site Yearly Performance Evaluation Philips Openview 16-Jan-08. Table of Contents

Philips Site Yearly Performance Evaluation Philips Openview 16-Jan-08. Table of Contents Philips Site Yearly Performance Evaluation Philips Openview 6-Jan-8 Table of Contents Summary and Signature Page 2 Specific Comments 3 Site Information 4 Equipment Information 4 Table Position Accuracy

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 07-649 CDRH Fluoroscopic Phantom Users Manual March 2005 Manual No. 07-649-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks

More information

A miniature all-optical photoacoustic imaging probe

A miniature all-optical photoacoustic imaging probe A miniature all-optical photoacoustic imaging probe Edward Z. Zhang * and Paul C. Beard Department of Medical Physics and Bioengineering, University College London, Gower Street, London WC1E 6BT, UK http://www.medphys.ucl.ac.uk/research/mle/index.htm

More information

Field Simulation Software to Improve Magnetic Resonance Imaging

Field Simulation Software to Improve Magnetic Resonance Imaging Field Simulation Software to Improve Magnetic Resonance Imaging a joint project with the NRI in South Korea CST Usergroup Meeting 2010 Darmstadt Institute for Biometry and Medicine Informatics J. Mallow,

More information

Ultrasound Beamforming and Image Formation. Jeremy J. Dahl

Ultrasound Beamforming and Image Formation. Jeremy J. Dahl Ultrasound Beamforming and Image Formation Jeremy J. Dahl Overview Ultrasound Concepts Beamforming Image Formation Absorption and TGC Advanced Beamforming Techniques Synthetic Receive Aperture Parallel

More information

CR Basics and FAQ. Overview. Historical Perspective

CR Basics and FAQ. Overview. Historical Perspective Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique image receptors

More information

Magnetic field measurements, Helmholtz pairs, and magnetic induction.

Magnetic field measurements, Helmholtz pairs, and magnetic induction. Magnetic field measurements, Helmholtz pairs, and magnetic induction. Part 1: Measurement of constant magnetic field: 1. Connections and measurement of resistance: a. Pick up the entire magnet assembly

More information

Diffusion and Functional MRI of the Spinal Cord Methods and Clinical Applications

Diffusion and Functional MRI of the Spinal Cord Methods and Clinical Applications Diffusion and Functional MRI of the Spinal Cord Methods and Clinical Applications Susceptibility artifacts in DTI of the spinal cord J. Cohen-Adad Q-space imaging and axon diameter measurements Functional

More information

MR in Tx Planning. Acknowledgements. Outline. Overview MR in RTP

MR in Tx Planning. Acknowledgements. Outline. Overview MR in RTP MR Data for Treatment Planning and Stereotactic Procedures: Sources of Distortion, Protocol Optimization, and Assessment (Preview of TG117 Report) Debra H. Brinkmann Mayo Clinic, Rochester MN Acknowledgements

More information

The development of the RF-pulse for the low level SAR used by the MRI.

The development of the RF-pulse for the low level SAR used by the MRI. The development of the RF-pulse for the low level SAR used by the MRI. Kojiro Yamaguchi a*, Eizo Umezawa a, Sachiko Ueoku b, Kazuhiro Katada c a Faculty of radiological technology, School of Health Science,

More information

1.6 Beam Wander vs. Image Jitter

1.6 Beam Wander vs. Image Jitter 8 Chapter 1 1.6 Beam Wander vs. Image Jitter It is common at this point to look at beam wander and image jitter and ask what differentiates them. Consider a cooperative optical communication system that

More information

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT

More information