User manual Bruker DPX200 NMR spectrometer

Size: px
Start display at page:

Download "User manual Bruker DPX200 NMR spectrometer"

Transcription

1 User manual Bruker DPX200 NMR spectrometer Insert the NMR tube in the spinner in such a way that the bottom of the tube reaches the grey disc at the bottom of the spinnerholder. Make sure that the NMR tubes are clean on the outside in order to prevent the inside of the probe from getting filthy. Spinner On/Off Lift On/Off Lock Gain Spinning Rate Done Z Z 2 Z 3 Fine A Draainknop Figure 1 BSMS keyboard Activate the lift with the Lift on/off button (see Figure 1) and wait until you can hear a strong airflow. Insert the spinner with the sample into the shaft of the magnet and press Lift on/off again, the tube will now be transported into the magnet. Wait until the sign down switches from red to green and then turn on the spinner by pressing Spinner on/off. By pressing spinning rate (Figure 1) the spinning frequency can be checked; it should read 18Hz. In order to open a new dataset you type edc in the command line of the xwinnmr window (Figure 3). A dialoguebox appears in which you can enter the name and number of the measurement (Figure 2). Choose an appropriate name (e.g. your login name, which is a unique name), and then enter a sequential number in the field expno. In the field procno it should generally read 1. After the edc command has been given by clicking save all experimental parameters of the old dataset, in which you gave the edc command, will be taken over in the new measurement. Most convenient is to always use the same name and have the expno increase by 1 for every new experiment. In this way back-ups are easily made and spectra are easy to retrieve. 1

2 Figure 2 edc window Now the spectrometer has to be locked on the deuterium signal of the solvent. Enter Lock (or l) in the command line of the xwinnmr window (Figure 3). Figure 3 Xwinnmr window A list with solvents appears in which you have to select the solvent you are using. The spectrometer will now fully automatically lock onto the solvent deuterium signal. Below in the command line after a while the message check lock shift finished appears which means the spectrometer has succesfully locked. Next the spectrometer needs to be shimmed. First you need to activate the lock window by selecting lock [lockdisp] in the drop-down menu windows. (See Figure 4) 2

3 Figure 4 Lock window Press the Z button (Figure 1) and turn the turning knob A until the lock signal (in the lock display) ceases to move upward. If the lock signal runs out of the window press lock gain and turn back knob A until the lock signal is back in the middle of the lock display. Now press Z 2 and again turn knob A until the lock signal ceases to to move upward. The same procedure needs to be carried out for Z 3 and then iteratively Z, Z 2 and Z 3 need to be adjusted until no more improvement of the lock signal is observed. If the signal does not respond or only a little you could try switching of te fine mode to see if the response improves (for Z 3 this is usually necessary). It s important to thoroughly shim the spectrometer in order to get an optimal measurement. If the lock signal shows a lot of noise and does not respond to changing Z, Z 2 and Z 3 a standard shimfile with standard shimming parameters can be loaded. This is done by entering rsh inverseprobe in the xwinnmr command line. After having loaded the shimfile the spectrometer can be further shimmed as described above. 3

4 Figure 5 edpar window In order to manually alter the acquisition parameters the dialogue box for the acquisition parameters (Figure 5) needs to be activated by typing eda in the command line. In this dialogue box all the acquisition parameters can be entered. Explanation of the most important parameters; PULSPROG In this field the NMR pulse program used is indicated. Usually it should read zg. The pulse program will be activated when zg is entered in the command line. AQ_mod Shows the acquisition mode (usually qsim, which is a quadrature detection method). TD Here the amount of data points for the FID (free induction decay) is entered. Typically 16k is entered here. NS Number of scans. This should be dividable by 4. DS Number of dummy scans. These are used to reach an equilibrium in the sample before starting the actual measurement. D array Array containing the values for d1-d31. These are delays the pulse program uses. D1 is the relaxation delay. P array Array containing the values for p1-p31. These are pulse lengths the pulse program uses. P1 is the 90 o pulse. SW, SWH Sweepwidth of the spectrum. SW is given in ppm. SWH is the sweepwidth in Hertz. AQ Acquisition time (seconds). NUC1 O1 PROSOL In this field the nucleus to be measured is indicated. (Usually H). This is the offset frequency in Hz (O1) or in ppm (O1P). O1 is in the middle of the spectrum. Example: If O1P is 5 ppm and the Sweepwidth (SW) is 12 ppm the spectrum will range from 1 to 11 ppm. This should be set to true. The software will adjust the pulse length and power to the specifications of the probe. 4

5 Close the acquisition parameter dialogue box by clicking save. Now the receiver gain has to be determined. The receiver gain function adjusts the amplitude of the FID before it is sent to the digitizer. An optimal amplitude must be found as to have a maximal FID amplitude without overloading the digitizer. (Overloading the digitizer could lead to artefacts in the spectrum). The most convenient way to determine the receiver gain is by typing rga (receiver gain automatic) in the command line. The receiver gain is adjusted automatically and after a few pulses the message rga finished appears. Now the measurement can be started by typing zg in the command line. The measurement is started and the expected measuring time is displayed below the command line. To see the progress of the measurement the FID and the experiment counter can be viewed in the acquisition mode. This acquisition mode can be activated by selecting acquire at the top of the xwinnmr window or by typing a in the command line. After the measurement is completed the size si of the processed dataset has to be checked. This is done by typing si in the command line. The value for si should be half of TD; si=1/2*td. The spectrum can now be viewed by typing ft (Fourier transform) in the command line. The spectrum is now shown but it usually is still out of phase. A Figure 6 Phase window In order to correct the phase of the spectrum you have to activate the phase window (see Figure 6, click phase ). In the phase window first you need to click biggest. A dotted line will appear in the spectrum which indicates the biggest peak. (Figure 6A) Then by clicking PH0 and by dragging the mouse across the PH0 button the phase of the biggest peak can be adjusted. Subsequently with PH1 the phase of peaks far away 5

6 from the biggest peak can be adjusted in a similar manner. After having adjusted the phase press return. In the dialogue box that now appears choose save&return. Further manipulations as well as printing the spectra can be done behind a workstation at the department. In order to take the sample out of the spectrometer first switch off the lock function and the spinner on the console and then activate the lift. After having taken out the sample switch off the lift and place the spinner back in the spinnerholder. The commands in a nutshell: -press lift on/off -Insert the spinner with the NMR tube in the shaft of the magnet and press lift on/off again. -type l or lock in the command line -select the solvent from the solvent list -shim the spectrometer by iteratively adjusting Z, Z 2 and Z 3. -type eda and insert or adjust the acquisition parameters -type rga to adjust the receiver gain. -type zg to start the measurement -after having checked si (command: si) enter ft to obtain the spectrum -Adjust the phase by click/dragging the PH0 and PH1 buttons in the phase window. -switch off the lock function and the spinner and then activate the lift to take out the sample. -switch off the lift and put the spinner back. Standard parameters If you want to start from scratch with the standard parameters you can use the rpar command to load a standard parameter file. To do this type rpar anorstd1h ( 1 H measurement) or rpar anorstd13c ( 13 C measurement) in the command line. 6

Step by step procedure for NMR data acquisition

Step by step procedure for NMR data acquisition Step by step procedure for NMR data acquisition Spectrometers The UTHSCSA 500, 600, and 700 MHz spectrometers are each equipped with 4 independent RF channels and are each operated by a Red Hat Linux workstation

More information

H Micro-Imaging. Tuning and Matching. i. Open any 1H data set and type wobb.

H Micro-Imaging. Tuning and Matching. i. Open any 1H data set and type wobb. - 1-1 H Micro-Imaging The NMR-specific properties of the objects are visualized as multidimensional images. Translational motion can be observed and spectroscopic information can be spatially resolved.

More information

HMBC 17. Goto. Introduction AVANCE User s Guide Bruker 185

HMBC 17. Goto. Introduction AVANCE User s Guide Bruker 185 Chapter HMBC 17 Introduction 17.1 Goto Heteronuclear Multiple Bond Correlation spectroscopy is a modified version of HMQC suitable for determining long-range 1 H- 13 C connectivity. This is useful in determining

More information

Two Dimensional Homonuclear Correlation Spectroscopy

Two Dimensional Homonuclear Correlation Spectroscopy Two Dimensional Homonuclear Correlation Spectroscopy Gradient COSY William D. Wheeler, Ph.D. Department of Chemistry University of Wyoming April 16, 1999 Revised September 22, 1999 2 INTRODUCTION Correlation

More information

Two Dimensional Homonuclear Correlation Spectroscopy

Two Dimensional Homonuclear Correlation Spectroscopy Two Dimensional Homonuclear Correlation Spectroscopy DQF-COSY William D. Wheeler, Ph.D. Department of Chemistry University of Wyoming September 23, 1999 2 INTRODUCTION Correlation Spectroscopy Correlation

More information

NMR Spectrometer Operation: xwinnmr

NMR Spectrometer Operation: xwinnmr NMR Spectrometer Operation: xwinnmr Dr. Robert Peterson Facility Manager NMR Technology Center UCLA-DOE Institute for Genomics and Proteomics UCLA Dept. of Chemistry and Biochemistry Overview This is a

More information

NMR spectrometer usage at the BioNMR facility ETH Zürich

NMR spectrometer usage at the BioNMR facility ETH Zürich NMR spectrometer usage at the BioNMR facility ETH Zürich Accounts 2 Safety precautions: strong magnetic fields 2 Parts of an NMR spectrometer 3 NMR data storage 3 Start topspin software 3 Initial steps

More information

1D Transient NOE on the Bruker DRX-500 and DRX-600

1D Transient NOE on the Bruker DRX-500 and DRX-600 1D Transient NOE on the Bruker DRX-500 and DRX-600 Reference: Stott, K., Stonehouse, J., Keeler, T.L. and Shaka, A.J., J. Amer. Chem. Soc. 1995, 117 (14), pp. 4199-4200. At thermal equilibrium in a strong

More information

Two Dimensional Heteronuclear Correlation Spectroscopy

Two Dimensional Heteronuclear Correlation Spectroscopy Two Dimensional Heteronuclear Correlation Spectroscopy Gradient HMQC William D. Wheeler, Ph.D. Department of Chemistry University of Wyoming Revised September 7, 2006 2 INTRODUCTION Correlation Spectroscopy

More information

400 MHz spectrometer user manual

400 MHz spectrometer user manual 400 MHz spectrometer user manual january 2017 Sandrine Denis-Quanquin 1. THE NMR SPECTROMETER... 3 2. MANUAL MODE / AUTOMATION... 4 2.1 SAMPLE CHANGER... 4 2.2 MANUAL MODE... 4 2.3 AUTOMATION... 4 3. PRELIMINARY

More information

Your first NMR measurement

Your first NMR measurement Your first NMR measurement Introduction Select 10mM water in D2O as NMR sample. The NMR spectrum of such sample consists of only two signals: the water signal and the peak of the reference (TSP). Follow

More information

NMR Hardware 06/06/2017. Outline. Instrumentation: Magnet. Increasing magnetic field increases Sensitivity, by power of 3/2 Dispersion, linearly

NMR Hardware 06/06/2017. Outline. Instrumentation: Magnet. Increasing magnetic field increases Sensitivity, by power of 3/2 Dispersion, linearly NMR Hardware Outline Magnet Lock Shims Gradient Probe Signal generation and transmitters Receiver and digitizer Variable temperature system Solids hardware Instrumentation: Magnet Often the most impressive

More information

Instruction for Operating the Bruker Avance III 800 MHz NMR Spectrometers in UTMB

Instruction for Operating the Bruker Avance III 800 MHz NMR Spectrometers in UTMB Instruction for Operating the Bruker Avance III 800 MHz NMR Spectrometers in UTMB Written by Tianzhi Wang, date: February 8, 2013. No food, no drink in NMR room and no internet in NMR host computer except

More information

Implementing ultrafast 2D NMR experiments on a Bruker Avance Spectrometer

Implementing ultrafast 2D NMR experiments on a Bruker Avance Spectrometer Implementing ultrafast 2D NMR experiments on a Bruker Avance Spectrometer Laetitia Rouger, Benoît Charrier, Serge Akoka, Patrick Giraudeau EBSI group CEISAM laboratory http://www.sciences.univ-nantes.fr/ceisam/en_ebsi1.php

More information

8 COSY. 8.1 Introduction. 8.2 Magnitude COSY

8 COSY. 8.1 Introduction. 8.2 Magnitude COSY 8 COSY 8.1 Introduction COSY (COrrelation SpectroscopY) is a homonuclear 2D technique that is used to correlate the chemical shifts of 1 H nuclei which are J-coupled to one another. In this chapter, two

More information

NMR Basics. Lecture 2

NMR Basics. Lecture 2 NMR Basics Lecture 2 Continuous wave (CW) vs. FT NMR There are two ways of tuning a piano: - key by key and recording each sound (or frequency). - or, kind of brutal, is to hit with a sledgehammer and

More information

Instructions for 1 H-, 13 C-, 19 F-, and 31 P-Spectra on the Varian Mercury-Vx-300

Instructions for 1 H-, 13 C-, 19 F-, and 31 P-Spectra on the Varian Mercury-Vx-300 1 Instructions for 1 H-, 13 C-, 19 F-, and 31 P-Spectra on the Varian Mercury-Vx-300 Please note: Under no circumstances move the magnet or the automatic sampler table. Do not attempt to use the auto sampler

More information

Student Name: Date Completed: Supervisor:

Student Name: Date Completed: Supervisor: 2 NMR Training for the 600 MHz NMR with Chempack INOVA 600 Tests and Assignment Certification Student Name: 600-Test #1: The student will be given a written test administered by Dr. Lee. This test will

More information

Lab 8 6.S02 Spring 2013 MRI Projection Imaging

Lab 8 6.S02 Spring 2013 MRI Projection Imaging 1. Spin Echos 1.1 Find f0, TX amplitudes, and shim settings In order to acquire spin echos, we first need to find the appropriate scanner settings using the FID GUI. This was all done last week, but these

More information

Open acqi window if the button has been lost. autolocking routine, alock= y for autolocking, alock= n for typical manual locking

Open acqi window if the button has been lost. autolocking routine, alock= y for autolocking, alock= n for typical manual locking Glossary of Common NMR Commands and Terms aa acqi ai alock aph array at points (np) axis='p' axis= pd BPsvf bc bs cd directory abort acquisition, hard stop Open acqi window if the button has been lost

More information

TopSpin Guide Book. Basic NMR Experiments User Manual. Innovation with Integrity. Version 002 NMR

TopSpin Guide Book. Basic NMR Experiments User Manual. Innovation with Integrity. Version 002 NMR TopSpin Guide Book Basic NMR Experiments User Manual Version 002 Innovation with Integrity NMR Copyright by Bruker Corporation All rights reserved. No part of this publication may be reproduced, stored

More information

2015 Spin echoes and projection imaging

2015 Spin echoes and projection imaging 1. Spin Echoes 1.1 Find f0, transmit amplitudes, and shim settings In order to acquire spin echoes, we first need to find the appropriate scanner settings using the FID GUI. This was all done last week,

More information

Chapter 1. 1 The NMR Spectrometer. 1.1 Components of an NMR Spectrometer The Magnet

Chapter 1. 1 The NMR Spectrometer. 1.1 Components of an NMR Spectrometer The Magnet Chapter 1 1 The NMR Spectrometer 1.1 Components of an NMR Spectrometer 1.1.1 The Magnet In most current NMR spectrometers the magnetic field is generated by a superconducting magnet (Fig. 1.1). The first

More information

Exercise at the Spectrometer

Exercise at the Spectrometer Exercise at the Spectrometer General The script is written for Bruker spectrometers. The color code means, red for all commands and green for parameters in the TopSpin commando line. The script gives additional

More information

If the magnetic field is larger, more energy is required to excite a given nucleus.

If the magnetic field is larger, more energy is required to excite a given nucleus. 1 2 If an NMR-active nucleus such as 1 H or 13 C is put into a magnet field, then it will come into resonance if it is irradiated with rf at the correct frequency. The correct frequency depends mainly

More information

An NMR Caveman s Guide to Quickly Acquiring Spectroscopic Data By Brian Sparling

An NMR Caveman s Guide to Quickly Acquiring Spectroscopic Data By Brian Sparling An NMR Caveman s Guide to Quickly Acquiring Spectroscopic Data By Brian Sparling Disclaimer: this guide is meant to be a quick, routine means of obtaining characterization data for unknown compounds in

More information

Laboratory Experiments for Nuclear Magnetic Resonance Spectroscopy May 6, 2004

Laboratory Experiments for Nuclear Magnetic Resonance Spectroscopy May 6, 2004 CONTENTS 1 Contents Laboratory Experiments for Nuclear Magnetic Resonance Spectroscopy May 6, 2004 1 Introduction 3 2 Safety 3 2.1 High Magnetic Fields......................................... 3 2.1.1

More information

KJM D-SELECTIVE NMR Experiments on the AVIIIHD-800. Version 1.0. Topspin 3.5 Windows 7

KJM D-SELECTIVE NMR Experiments on the AVIIIHD-800. Version 1.0. Topspin 3.5 Windows 7 KJM 9250 1D-SELECTIVE NMR Experiments on the AVIIIHD-800 Version 1.0 Topspin 3.5 Windows 7 Professor Emeritus Alistair Lawrence Wilkins, University of Waikato, New Zealand. January 2018 1D-SELECTIVE NMR

More information

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010 (N)MR Imaging Lab Course Script FMP PhD Autumn School Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder Date: November 3rd, 2010 1 Purpose: Understanding the basic principles of MR imaging

More information

Development of an Enantioselective Route towards the. Lycopodium Alkaloids: Total Synthesis of Lycopodine.

Development of an Enantioselective Route towards the. Lycopodium Alkaloids: Total Synthesis of Lycopodine. Development of an Enantioselective Route towards the Lycopodium Alkaloids: Total Synthesis of Lycopodine. Hua Yang and Rich G. Carter* Department of Chemistry, Oregon State University, Corvallis, OR 97331.

More information

KJM D-SELECTIVE NMR Experiments on the AVI-600 and AVII-600. Version 1.0. Topspin 3.5 Windows 7 Topspin 1.3 Windows XP

KJM D-SELECTIVE NMR Experiments on the AVI-600 and AVII-600. Version 1.0. Topspin 3.5 Windows 7 Topspin 1.3 Windows XP KJM 9250 1D-SELECTIVE NMR Experiments on the AVI-600 and AVII-600 Version 1.0 Topspin 3.5 Windows 7 Topspin 1.3 Windows XP Professor Emeritus Alistair Lawrence Wilkins, University of Waikato, New Zealand.

More information

PINMRF. Varian 300 MHz NMR Spectrometers User Guide for Advanced 1D and Basic 2D NMR Experiments

PINMRF. Varian 300 MHz NMR Spectrometers User Guide for Advanced 1D and Basic 2D NMR Experiments PINMRF Varian 300 MHz NMR Spectrometers User Guide for Advanced 1D and Basic 2D NMR Experiments INCLUDING: Inova-300-1 w/ 5mm 4-nucleus probe 365 WTHR Inova-300-2 w/ 5mm 4-nucleus probe 4100 BRWN Table

More information

Recording EPR Spectra using ER 4102ST Resonator

Recording EPR Spectra using ER 4102ST Resonator Recording EPR Spectra using ER 4102ST Resonator This protocol gives step-by-step instructions for recording an EPR spectrum using the high sensitivity Bruker SHQE cavity (assuming the SHQE is already in

More information

MT-540. D-Ribavirin, [ 3 H]- Lot A DG

MT-540. D-Ribavirin, [ 3 H]- Lot A DG MT-54 D-Ribavirin, [ 3 H]- Lot 165-145-225-A-2835-DG A) All chromatograms were run using the HPLC method described on the Product Data Sheet. Concentrations and volumes: Standard solution concentration

More information

Program. short intro the hardware locking and shimming 1D proton setup presaturation spectra

Program. short intro the hardware locking and shimming 1D proton setup presaturation spectra Program short intro the hardware locking and shimming 1D proton setup presaturation spectra 13 C spectra, DEPT processing of 1D file transfer and backup 2D homonuclear + 2D processing 2D heteronuclear

More information

Arenium Acid-Catalyzed Deuteration of Aromatic Hydrocarbons

Arenium Acid-Catalyzed Deuteration of Aromatic Hydrocarbons Supporting Information for Arenium Acid-Catalyzed euteration of Aromatic Hydrocarbons Simon uttwyler 1, Anna Butterfield 2, and Jay S. Siegel 2 * 1 epartment of Chemistry, Yale University, 350 Edwards

More information

MT Gemcitabine, [5-3 H(N)]- Lot A NK

MT Gemcitabine, [5-3 H(N)]- Lot A NK MT-1572 Gemcitabine, [5-3 H(N)]- Lot 194-14-234-A-2514-NK A) All chromatograms were run using the HPLC method described on the Product Data Sheet. Concentrations and volumes: Standard solution concentration

More information

KJM Version 1.0. Topspin 3.5 Windows 7 Topspin 1.3 Windows XP

KJM Version 1.0. Topspin 3.5 Windows 7 Topspin 1.3 Windows XP KJM 9250 1 H NMR spectra on the AVI-600 and AVII-600 Version 1.0 Topspin 3.5 Windows 7 Topspin 1.3 Windows XP Professor Emeritus Alistair Lawrence Wilkins, University of Waikato, New Zealand. January 2018

More information

Chapter 11 Coherence Editing: Pulse-field Gradients and Phase Cycling

Chapter 11 Coherence Editing: Pulse-field Gradients and Phase Cycling Chapter 11 Coherence Editing: Pulse-field Gradients and Phase Cycling Coherence editing is used to remove unwanted signals from NMR spectra. For example, in the double quantum filtered COSY experiment,

More information

Eclipse+ NMR Training Guide

Eclipse+ NMR Training Guide Eclipse+ NMR Training Guide Version 4.3.4 ECLIPSE+ NMR Training Guide Revision 20050617 Copyright 2005 by JEOL USA, Inc. Analytical Instruments Division 11 Dearborn Road Peabody, MA 01960 (978) 535-5900

More information

Chem 203. Organic Spectroscopy. Midterm Examination, Part II (60 points total) Problem 1 of 4 (three out of four required, 20 points)

Chem 203. Organic Spectroscopy. Midterm Examination, Part II (60 points total) Problem 1 of 4 (three out of four required, 20 points) NAME Chem 203 Organic Spectroscopy Midterm Examination, Part II (60 points total) Problem 1 of 4 (three out of four required, 20 points) Saturday, November 15, 2014, 9 am -??? SUBMIT THREE OF THE FOUR

More information

Module 1 Experimental Procedures. Experiment 1. What s in the Spectrometer?

Module 1 Experimental Procedures. Experiment 1. What s in the Spectrometer? Module 1 Experimental Procedures Experiment 1. What s in the Spectrometer? Identify elements of the spectrometer shown in Fig. 9 in the open spectrometer unit and identify the path of the light. Experiment

More information

Chem 203. Organic Spectroscopy. Midterm Examination, Part II (60 points total) Problem 4 of 4 (three out of four required, 20 points)

Chem 203. Organic Spectroscopy. Midterm Examination, Part II (60 points total) Problem 4 of 4 (three out of four required, 20 points) NAME Chem 203 Organic Spectroscopy Midterm Examination, Part II (60 points total) Problem 4 of 4 (three out of four required, 20 points) Saturday, November 9, 2013, 9 am -??? SUBMIT THREE OF THE FOUR PROBLEMS

More information

Arrayed Acquisition in VNMR and on the Gemini 1

Arrayed Acquisition in VNMR and on the Gemini 1 Arrayed Acquisition in VNMR and on the Gemini 1 I. Arrays in VNMR Vnmr allows you to quickly define experiments in which a series of spectra can be obtained as a function of any NMR parameter. For example,

More information

Pulsed-Field Gradient NMR Diffusion Measurement Protocol

Pulsed-Field Gradient NMR Diffusion Measurement Protocol Pulsed-Field Gradient NMR Diffusion Measurement Protocol Advanced Magnetic Resonance Imaging and Spectroscopy Facility National High Magnetic Field Laboratory University of Florida, Gainesville, FL. Contents

More information

Acceptance Solid State NMR Test Procedure. for Avance NMR Systems

Acceptance Solid State NMR Test Procedure. for Avance NMR Systems Acceptance Solid State NMR Test Procedure for Avance NMR Systems Manual P/N B92999 Acceptance Solid State NMR Test Procedures for AVANCE systems Index: 04 Number: B92999 Page: 1 (42) 1. Purpose 2. Area

More information

Chem 203. Organic Spectroscopy. Midterm Examination, Part II (60 points total) Problem 4 of 4 (three out of four required, 20 points)

Chem 203. Organic Spectroscopy. Midterm Examination, Part II (60 points total) Problem 4 of 4 (three out of four required, 20 points) NAME Chem 203 Organic Spectroscopy Midterm Examination, Part II (60 points total) Problem 4 of 4 (three out of four required, 20 points) Saturday, November 15, 2014, 9 am -??? SUBMIT THREE OF THE FOUR

More information

Chem 203 December 15, Final Exam Part II Problem 3 of 3 (30 points)

Chem 203 December 15, Final Exam Part II Problem 3 of 3 (30 points) Name: Chem 203 December 15, 2012 Final Exam Part II Problem 3 of 3 (30 points) Select and submit TWO OUT OF THE THREE PROBLEMS FROM PART II for grading. Do not submit three problems. If you wish to unstaple

More information

Ultrahigh-resolution Total Correlation NMR Spectroscopy

Ultrahigh-resolution Total Correlation NMR Spectroscopy Ultrahigh-resolution Total Correlation NMR Spectroscopy Supporting Information Mohammadali Foroozandeh, Ralph W. Adams, Mathias Nilsson and Gareth A. Morris* All experimental spectra were recorded at a

More information

RF Amplifier. User Manual EDACB BRUKER. Version

RF Amplifier. User Manual EDACB BRUKER. Version RF Amplifier User Manual EDACB Version 001 BRUKER The information in this manual may be altered without notice. BRUKER accepts no responsibility for actions taken as a result of use of this manual. BRUKER

More information

SUPPORTING INFORMATION

SUPPORTING INFORMATION Eur. J. Org. Chem. 2008 WILEY-VCH Verlag GmbH & Co. KGaA, 69451 Weinheim, 2008 ISSN 1434 193X SUPPORTING INFORMATION Title: Structural Elucidation with NMR Spectroscopy: Practical Strategies for Organic

More information

6.S02 MRI Lab Acquire MR signals. 2.1 Free Induction decay (FID)

6.S02 MRI Lab Acquire MR signals. 2.1 Free Induction decay (FID) 6.S02 MRI Lab 1 2. Acquire MR signals Connecting to the scanner Connect to VMware on the Lab Macs. Download and extract the following zip file in the MRI Lab dropbox folder: https://www.dropbox.com/s/ga8ga4a0sxwe62e/mit_download.zip

More information

Temps can range -130 to +120 C. See instructions below (part I). Such data can also be acquired on Athena and the AVANCE-360.

Temps can range -130 to +120 C. See instructions below (part I). Such data can also be acquired on Athena and the AVANCE-360. Flourine-19 Experiments on Varian Spectrometers updated: 2010 May 12 The difficulty with 19 F experiments involving decoupling (or 2D heterocorrelation) is the close proximity of the 19 F resonant frequency

More information

Recording EPR spectra using the Loop Gap Resonator (LGR)

Recording EPR spectra using the Loop Gap Resonator (LGR) Recording EPR spectra using the Loop Gap Resonator (LGR) This protocol gives step-by-step instructions for recording EPR spectra of spin labeled proteins (Nitroxide label like MTSSL) using the LGR assuming

More information

ISONIC PA AUT Spiral Scan Inspection of Tubular Parts Operating Manual and Inspection Procedure Rev 1.00 Sonotron NDT

ISONIC PA AUT Spiral Scan Inspection of Tubular Parts Operating Manual and Inspection Procedure Rev 1.00 Sonotron NDT ISONIC PA AUT Spiral Scan Inspection of Tubular Parts Operating Manual and Inspection Procedure Rev 1.00 Sonotron NDT General ISONIC PA AUT Spiral Scan Inspection Application was designed on the platform

More information

Avance GRASP. Installation/User Manual BRUKER. Version

Avance GRASP. Installation/User Manual BRUKER. Version Avance GRASP Installation/User Manual Version 002 BRUKER The information in this manual may be altered without notice. BRUKER accepts no responsibility for actions taken as a result of use of this manual.

More information

CRL MASS SPECTROMETRY FACILITY INSTRUMENT USER MANUAL MALDI MICRO. Operating Instructions. Basement Spec Lab:

CRL MASS SPECTROMETRY FACILITY INSTRUMENT USER MANUAL MALDI MICRO. Operating Instructions. Basement Spec Lab: CRL MASS SPECTROMETRY FACILITY INSTRUMENT USER MANUAL MALDI MICRO Operating Instructions Basement Spec Lab: 00.097 crs Page of 2 This is a guide to using the MALDI Micro for those who have received training.

More information

BRUKER BSMS. BSMS User Manual. Version 002

BRUKER BSMS. BSMS User Manual. Version 002 BRUKER BSMS BSMS User Manual Version 002 The information in this manual may be altered without notice. Bruker accepts no responsibility for actions taken as a result of use of this manual. Bruker accepts

More information

EPR spectrometer & components

EPR spectrometer & components EPR spectrometer & components Water lines µ wave bridge 293.2 VC41 Gas Flow Controller Temperature Controller Vacuum pump for cryostat on left hand side water chiller HASKRIS EMX EPR Spectrometer EMX Magnet

More information

NMR Spectrometer Hardware

NMR Spectrometer Hardware NMR Spectrometer Hardware Manoj Naik NMR Facility, TIFR Workshop on NMR & its applications TIFR, Mumbai November 23, 2009 Principle of NMR ω = γ B Nuclear Magnetic Resonance A Simplified 6 0 MHz NMR Spectrometer

More information

Principios Básicos de RMN en sólidos destinado a usuarios. Gustavo Monti. Fa.M.A.F. Universidad Nacional de Córdoba Argentina

Principios Básicos de RMN en sólidos destinado a usuarios. Gustavo Monti. Fa.M.A.F. Universidad Nacional de Córdoba Argentina Principios Básicos de RMN en sólidos destinado a usuarios Gustavo Monti Fa.M.A.F. Universidad Nacional de Córdoba Argentina magnet 1 2 4 5 6 computer 3 Block diagrama of a traditional NMR spectrometer.

More information

Chem 203 December 15, Final Exam Part II Problem 2 of 3 (30 points)

Chem 203 December 15, Final Exam Part II Problem 2 of 3 (30 points) Name: Chem 203 December 15, 2012 Final Exam Part II Problem 2 of 3 (30 points) Select and submit TWO OUT OF THE THREE PROBLEMS FROM PART II for grading. Do not submit three problems. If you wish to unstaple

More information

Introduction to NeuroScript MovAlyzeR Handwriting Movement Software (Draft 14 August 2015)

Introduction to NeuroScript MovAlyzeR Handwriting Movement Software (Draft 14 August 2015) Introduction to NeuroScript MovAlyzeR Page 1 of 20 Introduction to NeuroScript MovAlyzeR Handwriting Movement Software (Draft 14 August 2015) Our mission: Facilitate discoveries and applications with handwriting

More information

VnmrJ Liquids NMR User Guide. Varian NMR Spectrometer Systems with VnmrJ Software Pub. No , Rev. A0604

VnmrJ Liquids NMR User Guide. Varian NMR Spectrometer Systems with VnmrJ Software Pub. No , Rev. A0604 VnmrJ Liquids NMR User Guide Varian NMR Spectrometer Systems with VnmrJ Software Pub. No. 01-999250-00, Rev. A0604 VnmrJ Liquids NMR User Guide Varian NMR Spectrometer Systems with VnmrJ Software Pub.

More information

COPYRIGHT NATIONAL DESIGN ACADEMY

COPYRIGHT NATIONAL DESIGN ACADEMY National Design Academy How 2 Guide Use SketchUp with LayOut COPYRIGHT NATIONAL DESIGN ACADEMY Use SketchUp with LayOut In order to be able to produce accurate scale drawings in SketchUp, you must use

More information

Guide to Operating the Bruker FT-IR Microscope I. Basic Spectrum Acquisition with Vertex 70

Guide to Operating the Bruker FT-IR Microscope I. Basic Spectrum Acquisition with Vertex 70 Guide to Operating the Bruker FT-IR Microscope I. Basic Spectrum Acquisition with Vertex 70 Susheng Tan, Ph.D. NanoScale Fabrication and Characterization Facility, University of Pittsburgh SB60 Benedum

More information

10. Phase Cycling and Pulsed Field Gradients Introduction to Phase Cycling - Quadrature images

10. Phase Cycling and Pulsed Field Gradients Introduction to Phase Cycling - Quadrature images 10. Phase Cycling and Pulsed Field Gradients 10.1 Introduction to Phase Cycling - Quadrature images The selection of coherence transfer pathways (CTP) by phase cycling or PFGs is the tool that allows the

More information

JEOL 6500 User Manual

JEOL 6500 User Manual LOG IN to your session on the computer to the left of the microscope. Starting Conditions 1. Press Ctrl-Alt-Del and log on to the microscope computer. Click on JEOL PC SEM 6500 icon. Click yes if message

More information

STEM Spectrum Imaging Tutorial

STEM Spectrum Imaging Tutorial STEM Spectrum Imaging Tutorial Gatan, Inc. 5933 Coronado Lane, Pleasanton, CA 94588 Tel: (925) 463-0200 Fax: (925) 463-0204 April 2001 Contents 1 Introduction 1.1 What is Spectrum Imaging? 2 Hardware 3

More information

Experiment 2: Electronic Enhancement of S/N and Boxcar Filtering

Experiment 2: Electronic Enhancement of S/N and Boxcar Filtering Experiment 2: Electronic Enhancement of S/N and Boxcar Filtering Synopsis: A simple waveform generator will apply a triangular voltage ramp through an R/C circuit. A storage digital oscilloscope, or an

More information

NMR FACILITY NEWSLETTER

NMR FACILITY NEWSLETTER NMR FACILITY NEWSLETTER Department of Chemistry and Biochemistry Matt Revington-Facility Coordinator mrevingt@uwindsor.ca Ext 3997 Workshop Announcement : Advanced Topics in NMR There will be an Advanced

More information

SigCal32 User s Guide Version 3.0

SigCal32 User s Guide Version 3.0 SigCal User s Guide . . SigCal32 User s Guide Version 3.0 Copyright 1999 TDT. All rights reserved. No part of this manual may be reproduced or transmitted in any form or by any means, electronic or mechanical,

More information

Guide to Varian Spectrometers running VnmrJ

Guide to Varian Spectrometers running VnmrJ Guide to Varian Spectrometers running VnmrJ David N.M. Jones Department of Pharmacology and Program in Biomolecular Structure University of Colorado Denver and Health Sciences Center All original material

More information

SpinCore RadioProcessor LabVIEW Extensions

SpinCore RadioProcessor LabVIEW Extensions NMR Interface User's Manual SpinCore Technologies, Inc. http:// Congratulations and thank you for choosing a design from SpinCore Technologies, Inc. We appreciate your business! At SpinCore we try to fully

More information

Agilent 33522A Function Arbitrary Waveform Generator. Tektronix TDS 3012B Oscilloscope

Agilent 33522A Function Arbitrary Waveform Generator. Tektronix TDS 3012B Oscilloscope Agilent 33522A Function/Arbitrary Waveform Generator and Tektronix TDS 3012B Oscilloscope Agilent 33522A Function Arbitrary Waveform Generator The signal source for this lab is the Agilent 33522A Function

More information

Lock and Shim Crudely on your Sample

Lock and Shim Crudely on your Sample 1 Practicum 1, Spring 2004 Anne-Frances Miller, 2004 Collecting a basic 1d spectrum. We will go through basic steps in maintaining your account, loading a sample, locking, shimming, collecting a simple

More information

Applications Guide. Spectral Editing with SVS. (Works-in-Progress) MAGNETOM TaTs and Verio Systems (3T)

Applications Guide. Spectral Editing with SVS. (Works-in-Progress) MAGNETOM TaTs and Verio Systems (3T) Applications Guide Spectral Editing with SVS (Works-in-Progress) MAGNETOM TaTs and Verio Systems (3T) syngo MR Numaris 4 VB17A June 2009 Version 1.1 WIP #529 Important Note This document provides a description

More information

7 Equipments. Spectrometers

7 Equipments. Spectrometers 7 Equipments Spectrometers There are three spectrometers located in the NMR laboratory: Varian UNITYplus 500 MHz (NMR500), Varian UNITYplus 500 MHz (Nightmare) and Varian INOVA 600 MHz. All spectrometers

More information

The operation manual of spotlight 300 IR microscope

The operation manual of spotlight 300 IR microscope The operation manual of spotlight 300 IR microscope Make sure there is no sample under the microscope and then click spotlight on the desktop to open the software. You can do imaging with the image mode

More information

SOP-P051. Scanning of Optical Filters With USB2000. Objective: To determine the spectral transmittance properties of an optical filter.

SOP-P051. Scanning of Optical Filters With USB2000. Objective: To determine the spectral transmittance properties of an optical filter. Purdue University Cytometry Laboratories SOP-P051 Scanning of Optical Filters With USB2000 Objective: To determine the spectral transmittance properties of an optical filter. Procedure: 1. Ensure that

More information

Laboratory Experiment #2 Frequency Response Measurements

Laboratory Experiment #2 Frequency Response Measurements J.B.Francis College of Engineering Mechanical Engineering Department 22-403 Laboratory Experiment #2 Frequency Response Measurements Introduction It is known from dynamic systems that a structure temporarily

More information

Renishaw InVia Raman microscope

Renishaw InVia Raman microscope Laser Spectroscopy Labs Renishaw InVia Raman microscope Operation instructions 1. Turn On the power switch, system power switch is located towards the back of the system on the right hand side. Wait ~10

More information

Bruker Dimension Icon AFM Quick User s Guide

Bruker Dimension Icon AFM Quick User s Guide Bruker Dimension Icon AFM Quick User s Guide August 8 2014 GLA Contacts Jingjing Jiang (jjiang2@caltech.edu 626-616-6357) Xinghao Zhou (xzzhou@caltech.edu 626-375-0855) Bruker Tech Support (AFMSupport@bruker-nano.com

More information

Workshop on Rapid Scan EPR. University of Denver EPR Center and Bruker BioSpin July 28, 2013

Workshop on Rapid Scan EPR. University of Denver EPR Center and Bruker BioSpin July 28, 2013 Workshop on Rapid Scan EPR University of Denver EPR Center and Bruker BioSpin July 28, 2013 Direct detection Direct detected magnetic resonance that is, without modulation and phase-sensitive detection

More information

ScanArray Overview. Principle of Operation. Instrument Components

ScanArray Overview. Principle of Operation. Instrument Components ScanArray Overview The GSI Lumonics ScanArrayÒ Microarray Analysis System is a scanning laser confocal fluorescence microscope that is used to determine the fluorescence intensity of a two-dimensional

More information

MicroLab 500-series Getting Started

MicroLab 500-series Getting Started MicroLab 500-series Getting Started 2 Contents CHAPTER 1: Getting Started Connecting the Hardware....6 Installing the USB driver......6 Installing the Software.....8 Starting a new Experiment...8 CHAPTER

More information

The Agilent OneNMR Probe

The Agilent OneNMR Probe The Agilent OneNMR Probe Technical Overview Introduction The Agilent OneNMR probe represents a new class of NMR probes. This technology is the most signifi cant advance in solution-state probes in over

More information

The University of Texas at Austin Electron Paramagnetic Resonance Facility. Bruker EMX+ CW EPR Spectrometer

The University of Texas at Austin Electron Paramagnetic Resonance Facility. Bruker EMX+ CW EPR Spectrometer The University of Texas at Austin Electron Paramagnetic Resonance Facility Bruker EMX+ CW EPR Spectrometer Contents: General Notes...2 Turning on the Instrument...2 Collecting a Spectrum...3 Turning off

More information

Measuring Kinetics of Luminescence with TDS 744 oscilloscope

Measuring Kinetics of Luminescence with TDS 744 oscilloscope Measuring Kinetics of Luminescence with TDS 744 oscilloscope Eex Nex Luminescence Photon E 0 Disclaimer Safety the first!!! This presentation is not manual. It is just brief set of rule to remind procedure

More information

Shimadzu RF-5301 Fluorimeter operation guide for students

Shimadzu RF-5301 Fluorimeter operation guide for students Department of Chemistry Teaching Laboratories Shimadzu RF-5301 Fluorimeter operation guide for students General directions Detailed instructions for use of the fluorimeter may be given in the lab script,

More information

II. LAB. * Open the LabVIEW program (Start > All Programs > National Instruments > LabVIEW 2012 > LabVIEW 2012)

II. LAB. * Open the LabVIEW program (Start > All Programs > National Instruments > LabVIEW 2012 > LabVIEW 2012) II. LAB Software Required: NI LabVIEW 2012, NI LabVIEW 4.3 Modulation Toolkit. Functions and VI (Virtual Instrument) from the LabVIEW software to be used in this lab: niusrp Open Tx Session (VI), niusrp

More information

Instruction manual for T3DS software. Tool for THz Time-Domain Spectroscopy. Release 4.0

Instruction manual for T3DS software. Tool for THz Time-Domain Spectroscopy. Release 4.0 Instruction manual for T3DS software Release 4.0 Table of contents 0. Setup... 3 1. Start-up... 5 2. Input parameters and delay line control... 6 3. Slow scan measurement... 8 4. Fast scan measurement...

More information

Bruker Dimension Icon AFM Quick User s Guide

Bruker Dimension Icon AFM Quick User s Guide Bruker Dimension Icon AFM Quick User s Guide March 3, 2015 GLA Contacts Jingjing Jiang (jjiang2@caltech.edu 626-616-6357) Xinghao Zhou (xzzhou@caltech.edu 626-375-0855) Bruker Tech Support (AFMSupport@bruker-nano.com

More information

Relaxation-encoded NMR experiments for mixture analysis: REST and beer. Electronic Supporting Information

Relaxation-encoded NMR experiments for mixture analysis: REST and beer. Electronic Supporting Information Electronic Supplementary Material (ESI) for ChemComm. This journal is The Royal Society of Chemistry 2017 Relaxation-encoded NMR experiments for mixture analysis: REST and beer Electronic Supporting Information

More information

IVI STEP TYPES. Contents

IVI STEP TYPES. Contents IVI STEP TYPES Contents This document describes the set of IVI step types that TestStand provides. First, the document discusses how to use the IVI step types and how to edit IVI steps. Next, the document

More information

PHY3902 PHY3904. Nuclear magnetic resonance Laboratory Protocol

PHY3902 PHY3904. Nuclear magnetic resonance Laboratory Protocol PHY3902 PHY3904 Nuclear magnetic resonance Laboratory Protocol PHY3902 PHY3904 Nuclear magnetic resonance Laboratory Protocol GETTING STARTED You might be tempted now to put a sample in the probe and try

More information

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software:

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software: Confocal imaging on the Leica TCS SP8 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software: 4) Do not touch the microscope while the software is initializing. Choose your options: Turn

More information

EPR2010 Puerto Rico. Rapid Scan EPR. Mark Tseitlin, Deborah G. Mitchell, Joshua A. Biller, Richard W. Quine, George A. Rinard, Sandra S.

EPR2010 Puerto Rico. Rapid Scan EPR. Mark Tseitlin, Deborah G. Mitchell, Joshua A. Biller, Richard W. Quine, George A. Rinard, Sandra S. EPR2010 Puerto Rico Rapid Scan EPR Mark Tseitlin, Deborah G. Mitchell, Joshua A. Biller, Richard W. Quine, George A. Rinard, Sandra S. Eaton, Gareth R. Eaton, and Ralph T. Weber University of Denver and

More information

ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide

ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide Matyas Molnar, Biovis 2016 Starting the microscpe 1. Check the microscope if everything looks clean and normal. If not, report it in the logbook.

More information

LAB II. INTRODUCTION TO LABVIEW

LAB II. INTRODUCTION TO LABVIEW 1. OBJECTIVE LAB II. INTRODUCTION TO LABVIEW In this lab, you are to gain a basic understanding of how LabView operates the lab equipment remotely. 2. OVERVIEW In the procedure of this lab, you will build

More information