Shimadzu RF-5301 Fluorimeter operation guide for students
|
|
- Karen Watson
- 5 years ago
- Views:
Transcription
1 Department of Chemistry Teaching Laboratories Shimadzu RF-5301 Fluorimeter operation guide for students General directions Detailed instructions for use of the fluorimeter may be given in the lab script, consult a demonstrator for assistance where necessary. The fluorimeter cell holder has two positions; Right angle and Front surface, see photographs; Right Angle Cell holder positions Front surface Handle the fluorimeter cells with care; Always handle on the corners, or at the top and bottom Ensure that the cell is washed thoroughly after each solution and then rinsed out with the new solution before taking its spectrum Ensure that the same cell face is used for each solution Ensure that the cell windows are clean, dry and not contaminated Page 1 of 5 Fluorimeter Last updated: 21/06/2017
2 How to record fluorescence emission spectra 1. Ensure the sampling geometry is right-angle illumination by carefully rotating the cell holder (if necessary). Please check with a demonstrator if you are unsure. 2. In the software click on Configure and select Parameters; the Spectrum parameters dialogue box will be displayed. 3. To run an emission spectrum, click on Spectrum type: Emission. 4. Use the UV absorption maximum, λmax, of your sample as the excitation wavelength; click on the EX wavelength box and type in this value. 5. Click on the EM wavelength boxes and type in (for e.g.) Start 300 nm, end 600 nm. This selects the range of wavelength over which the emission spectrum will be scanned. Details may be in your script. 6. Click on sensitivity: select Low. 7. Click Scan speed: Medium, and Slit width: EX 5 nm, EM 5 nm; dots will appear against the selected parameters. 8. Click on the Recording range boxes, adjust the spectrum y-axis to a size which fills the screen by typing values in, e.g. Low: 0 and High: 400. You can change the range later if necessary if the spectrum goes off scale. 9. Click on Ok to return to the main application window. 10. Click on Start at the bottom of the window; the spectrum will be scanned and displayed on the screen. 11. Save the spectrum obtained, type a filename into the box in the filename entry window (e.g. abcd1) when it appears at the end of the scan, and choose the Save option. Record this filename. 12. The main application window should now be displayed. Click on Manipulate and select Peakpick. A window showing the channels containing the emission spectrum will appear. Click on the channel you wish to analyze and then select Ok. The spectrum will be redrawn with the peaks labelled 13. Click on the cross in the top right hand corner to close the peak window. The main application window should now be displayed. Page 2 of 5 Fluorimeter Last updated: 21/06/2017
3 How to record fluorescence excitation spectra 1. Click on Configure and select Parameters. The Spectrum parameters dialogue box will be displayed. 2. To run an excitation spectrum, click on Spectrum type: Excitation. 3. Take the λem from the emission spectrum and use this value as the excitation wavelength. Click on the EM wavelength box and type in this value. This selects the wavelength of the light used to observe emission from the sample. 4. Click on the EX wavelength boxes and type in (for e.g.) Start 300 nm, End 600 nm. This selects the range of wavelength over which the excitation spectrum will be scanned. 5. All the other spectrum parameters should be the same as those used for obtaining the emission spectrum. 6. Click on OK to return to the main application window. 7. Click on Start at the bottom of the window; the spectrum will be scanned and displayed on the screen. 8. Save the spectrum using a different filename, use Manipulate and Peakpick as for the emission spectrum. Record the wavelength of excitation (λex) and intensity 9. Click on the cross in the top right hand corner to close the peak window. The main application window should now be displayed with both the emission and excitation spectrum shown. How to print out spectra 1. To print out the spectra it is useful to present both scans on the same plot. To select the files you wish to display, go to the main application window, click on Presentation and select Channel status. 2. The display option for each channel can be toggled between Y and N. Select Y for both of your emission and excitation spectra, N for other files. Click on OK 3. If you wish to change the x- and y-ranges for your graph on the screen, click on Presentation and select Set limit. The Set limits dialogue box will appear. Change any values by clicking on the appropriate box and entering new values. Click on OK to return to the main applications window. The graph will now be displayed with your new ranges. Page 3 of 5 Fluorimeter Last updated: 21/06/2017
4 4. To plot this data, click on Presentation and select Plot. The Plot layout dialogue box will be displayed. In row A double click on graph; a cross will be shown in the box and the Plot data dialogue box will appear. Type a title, if you wish, in the Graph title box. The y-axis range may be altered, if you wish. Click on OK. 5. Ensure that Screen is now selected in the Channel/f.name area of the Plot layout dialogue box; this will allow a plot of the contents of the graph as they appear on the screen. Click on the Quadrant boxes to ensure that only 1 and 2 are active (showing crosses); clicking on the boxes toggles between active and inactive. Click on Print to obtain the plot. How to record spectra of donor-acceptor mixtures 1. Set the sampling geometry to front-surface angle illumination by rotating the cell holder clockwise until it clicks (see page 1) 2. In the energy transfer study you need only obtain emission spectra; therefore Spectrum type = Emission. 3. Select the absorption maximum λmax of the donor as the excitation wavelength and type this into the EX wavelength box. 4. Run the fluorescence emission spectra of all solutions between (for e.g.) 300 nm and 600 nm; type Start 300 nm, End 600 nm in the EM wavelength boxes. 5. Click on sensitivity: select Low. Click Scan speed: Medium, Slit width: EX 5nm, EM 5 nm. A Recording range of nm is likely to be appropriate. 6. Commence with the donor only solution and ensure that the peak is on scale, then repeat the measurement for the remaining solutions and do not adjust any instrument settings or the filter in the excitation beam for the D A mixtures. 7. Ensure that the cell is washed thoroughly after each solution and then rinsed out with the new solution before taking its spectrum. 8. Obtain the emission intensity, i, at the peak of the donor emission for each of the solutions 9. When you have obtained all the spectra, make a plot with all of these overlaid on one chart. Page 4 of 5 Fluorimeter Last updated: 21/06/2017
5 Instrument set-up guide for technicians / demonstrators Turn the instrument on: see photo Power switch Turning on: XE Lamp switch to ON and Main switch to I listen to lamp striking; ticking noise should stop after a few seconds Turning off: XE Lamp switch to OFF, wait 5-10 min to cool, turn main switch to O Turn on the pc and start the software Open the RFPC software Select Configure from the header menu and then PC configuration. In the dialogue box choose the Serial port setting box and connect to the number which is suggested, then click Ok. Note that the PC sometimes connects to the correct com port automatically Select Configure from the header menu and then Instrument. In the dialogue box select the On button against the spectrometer. A new dialogue box will open with the progress of the instrument ready checks displayed Once all checks are complete (each box will go green against the tested parameters) click Ok The instrument is now ready for use and the PC communications established. Page 5 of 5 Fluorimeter Last updated: 21/06/2017
Renishaw InVia Raman microscope
Laser Spectroscopy Labs Renishaw InVia Raman microscope Operation instructions 1. Turn On the power switch, system power switch is located towards the back of the system on the right hand side. Wait ~10
More informationHoriba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer
Page 1 of 11 Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer The Aramis Raman system is a software selectable multi-wavelength Raman system with mapping capabilities with a 400mm monochromator and
More informationEDUCATION EXPERIMENT. Determining Color Difference (Delta E) Introduction
EDUCATION EXPERIMENT Determining Color Difference (Delta E) Introduction www.oceanoptics.com info@oceanoptics.com Color matching and color consistency have become important quality control parameters in
More informationPhysicsAndMathsTutor.com 1
PhysicsAndMathsTutor.com 1 Q1. Just over two hundred years ago Thomas Young demonstrated the interference of light by illuminating two closely spaced narrow slits with light from a single light source.
More informationUniversity of Wisconsin Chemistry 524 Spectroscopic Components *
University of Wisconsin Chemistry 524 Spectroscopic Components * In journal articles, presentations, and textbooks, chemical instruments are often represented as block diagrams. These block diagrams highlight
More informationSOP-P051. Scanning of Optical Filters With USB2000. Objective: To determine the spectral transmittance properties of an optical filter.
Purdue University Cytometry Laboratories SOP-P051 Scanning of Optical Filters With USB2000 Objective: To determine the spectral transmittance properties of an optical filter. Procedure: 1. Ensure that
More informationConfocal Raman Microscopy (WITec Alpha 300R)
Confocal Raman Microscopy (WITec Alpha 300R) Please refer to Witec Alpha300R Confocal Raman Microscope User Manual for the details of the operating procedure. Sample preparation 1. Attach your sample on
More informationThe equipment used share any common features regardless of the! being measured. Electronic detection was not always available.
The equipment used share any common features regardless of the! being measured. Each will have a light source sample cell! selector We ll now look at various equipment types. Electronic detection was not
More informationChemistry 524--"Hour Exam"--Keiderling Mar. 19, pm SES
Chemistry 524--"Hour Exam"--Keiderling Mar. 19, 2013 -- 2-4 pm -- 170 SES Please answer all questions in the answer book provided. Calculators, rulers, pens and pencils permitted. No open books allowed.
More informationSodium Thiosulphate and hydrochloric acid
Sodium Thiosulphate and hydrochloric acid Introduction The rate of reaction can be altered by varying temperature or concentration, or by changing the surface area of a solid reactant, or by adding a catalyst.
More informationThe Shoebox spectrograph construction and lab investigations. By Timothy Grove
The Shoebox spectrograph construction and lab investigations By Timothy Grove 1 Part 1. Build your own spectrograph from flat cardboard Tools and materials: Necessary items Scrap cardboard (You will need
More informationMeasurement Method of High Absorbance (Low Transmittance) Samples by UH4150 INTRODUCTION
INTRODUCTION With UH4150, a detector can be selected depending on the analysis purpose. When analyzing a solid sample which doesn t contain any diffuse components, by selecting the direct light detector,
More informationChemistry workshops and investigations
Chemistry workshops and investigations Make a DVD smartphone spectrometer CfE Advanced Higher Inorganic and Physical Chemistry Electromagnetic radiation and atomic spectra Background Spectroscopy is an
More informationSensors and Scatterplots Activity Excel Worksheet
Name: Date: Sensors and Scatterplots Activity Excel Worksheet Directions Using our class datasheets, we will analyze additional scatterplots, using Microsoft Excel to make those plots. To get started,
More informationLSM 780 Confocal Microscope Standard Operation Protocol
LSM 780 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Sign on log sheet according to Actual start time 2. Check Compressed Air supply for the air table 3. Switch
More informationSPECTRAL SCANNER. Recycling
SPECTRAL SCANNER The Spectral Scanner, produced on an original project of DV s.r.l., is an instrument to acquire with extreme simplicity the spectral distribution of the different wavelengths (spectral
More informationVS7550 VUV/UV Mini Spectrograph Operating Manual
Document RD 15 11 No: VS7550 VUV/UV Mini Spectrograph Operating Manual VS7550 Operating Manual 1 Table of Contents Table of Contents Overview Specifications Vacuum Interface Software and Drivers Packing
More informationExcel Lab 2: Plots of Data Sets
Excel Lab 2: Plots of Data Sets Excel makes it very easy for the scientist to visualize a data set. In this assignment, we learn how to produce various plots of data sets. Open a new Excel workbook, and
More informationExperiment 2B Integrated Laboratory Experiment DETERMINATION OF RIBOFLAVIN: A COMPARISON OF TECHNIQUES PART B. MOLECULAR FLUORESCENCE SPECTROSCOPY
CH 461 & CH 461H F 14 Name OREGON STATE UNIVERSITY DEPARTMENT OF CHEMISTRY Experiment 2B Integrated Laboratory Experiment DETERMINATION OF RIBOFLAVIN: A COMPARISON OF TECHNIQUES PART B. MOLECULAR FLUORESCENCE
More informationExcel Tool: Plots of Data Sets
Excel Tool: Plots of Data Sets Excel makes it very easy for the scientist to visualize a data set. In this assignment, we learn how to produce various plots of data sets. Open a new Excel workbook, and
More informationPractical work no. 3: Confocal Live Cell Microscopy
Practical work no. 3: Confocal Live Cell Microscopy Course Instructor: Mikko Liljeström (MIU) 1 Background Confocal microscopy: The main idea behind confocality is that it suppresses the signal outside
More informationUV-Vis-NIR Spectrophotometer Quick Start Guide
UV-Vis-NIR Spectrophotometer Quick Start Guide The following instructions are provided as a Quick Start Guide for powering up, running measurements, and shutting down the Lambda 950 UV-Vis Spectrophotometer.
More informationUsing DIAGNOSE.EXE with a ColorQuest 45/0
Insight on Color November, 1999, Vol. 11, No. 11 (Rev. 5/02) Using DIAGNOSE.EXE with a ColorQuest 45/0 The DIAGNOSE.EXE program, which may be used to assess instrument performance and diagnose potential
More informationECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the
ECEN 4606 Lab 8 Spectroscopy SUMMARY: ROBLEM 1: Pedrotti 3 12-10. In this lab, you will design, build and test an optical spectrum analyzer and use it for both absorption and emission spectroscopy. The
More informationAtomic and nuclear physics
Atomic and nuclear physics X-ray physics Physics of the atomic shell LEYBOLD Physics Leaflets Investigating the energy spectrum of an x-ray tube as a function of the high voltage and the emission current
More informationAqualog. Water Quality Measurements Made Easy PARTICLE CHARACTERIZATION ELEMENTAL ANALYSIS FLUORESCENCE
Aqualog Water Quality Measurements Made Easy ELEMENTAL ANALYSIS FLUORESCENCE GRATINGS & OEM SPECTROMETERS OPTICAL COMPONENTS PARTICLE CHARACTERIZATION RAMAN SPECTROSCOPIC ELLIPSOMETRY SPR IMAGING Water
More informationGuide to Confocal 5. Starting session
Guide to Confocal 5 Remember that when booking and before starting session you can check for any problems at https://www.bris.ac.uk/biochemistry/uobonly/cif/index.html Starting session Switch on microscope
More informationLSM 710 Confocal Microscope Standard Operation Protocol
LSM 710 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Switch on Main power switch 2. Switch on System / PC power button 3. Switch on Components power button 4.
More informationExperiment 10. Color. Observe the transmission properties of the three additive primary color filters and the three subtractive primary color filters.
Experiment 10 Color 10.1 Objectives Observe the transmission properties of the three additive primary color filters and the three subtractive primary color filters. Observe and interpret the color sensations
More informationThings to check before start-up.
Byeong Cha Page 1 11/24/2009 Manual for Leica SP2 Confocal Microscope Enter you name, the date, the time, and the account number in the user log book. Things to check before start-up. Make sure that your
More informationCell Biology and Bioimaging Core
Cell Biology and Bioimaging Core Leica TCS SP5 Operating Instructions Starting up the instrument 1. First, log in the log book located on the confocal desk. Include your name, your lab s PI, an account
More informationFaculty of Electrical & Electronics Engineering BEE4233 Antenna and Propagation. LAB 1: Introduction to Antenna Measurement
Faculty of Electrical & Electronics Engineering BEE4233 Antenna and Propagation LAB 1: Introduction to Antenna Measurement Mapping CO, PO, Domain, KI : CO2,PO3,P5,CTPS5 CO1: Characterize the fundamentals
More informationIn the Heat of the Light
The Electromagnetic Spectrum Laboratory Investigation TEACHER NOTES In the Heat of the Light Key Concept Fluorescent and incandescent lights work in different ways. Skills Focus observing, measuring, interpreting
More informationMicroLab 500-series Getting Started
MicroLab 500-series Getting Started 2 Contents CHAPTER 1: Getting Started Connecting the Hardware....6 Installing the USB driver......6 Installing the Software.....8 Starting a new Experiment...8 CHAPTER
More informationHow to make a list sweep measurement
How to make a list sweep measurement This material shows how to perform a list sweep measurement through an example of the Photovoltaic Cell IV measurement. Figure 1 illustrates the connection and condition
More informationNCSS Statistical Software
Chapter 147 Introduction A mosaic plot is a graphical display of the cell frequencies of a contingency table in which the area of boxes of the plot are proportional to the cell frequencies of the contingency
More informationHow to Make a Run Chart in Excel
How to Make a Run Chart in Excel While there are some statistical programs that you can use to make a run chart, it is simple to make in Excel, using Excel s built-in chart functions. The following are
More informationRENISHAW INVIA RAMAN SPECTROMETER
STANDARD OPERATING PROCEDURE: RENISHAW INVIA RAMAN SPECTROMETER Purpose of this Instrument: The Renishaw invia Raman Spectrometer is an instrument used to analyze the Raman scattered light from samples
More informationUnit 3.11 Introduction to Absolute & Polar Coordinates in AutoCAD Ardrey Kell High School Charlotte, NC Date: 11/18/13 (Revision II)
Unit 3.11 Introduction to Absolute & Polar Coordinates in AutoCAD X,Y,Z @3
More informationApplications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region
Feature Article JY Division I nformation Optical Spectroscopy Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region Raymond Pini, Salvatore Atzeni Abstract Multichannel
More informationWITec Alpha 300R Quick Operation Summary October 2018
WITec Alpha 300R Quick Operation Summary October 2018 This document is frequently updated if you feel information should be added, please indicate that to the facility manager (currently Philip Carubia,
More informationOperating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.
Smith College August 2010 Operating Checklist for using the Laser Scanning Confocal Microscope Leica TCS SP5. CONTENT, page no. Startup, 1 Initial set-up, 1 Software, 2 Microscope Specimen observation
More informationSolid-state physics. Bragg reflection: determining the lattice constants of monocrystals. LEYBOLD Physics Leaflets P
Solid-state physics Properties of crystals X-ray structural analysis LEYBOLD Physics Leaflets Bragg reflection: determining the lattice constants of monocrystals P7.1.2.1 Objects of the experiment Investigating
More informationStep 1: Set up the variables AB Design. Use the top cells to label the variables that will be displayed on the X and Y axes of the graph
Step 1: Set up the variables AB Design Use the top cells to label the variables that will be displayed on the X and Y axes of the graph Step 1: Set up the variables X axis for AB Design Enter X axis label
More informationAqualog. Water Quality Measurements Made Easy FLUORESCENCE
Aqualog Water Quality Measurements Made Easy FLUORESCENCE Water quality measurements made easy The only simultaneous absorbance and fluorescence system for water quality analysis! The new Aqualog is the
More informationBasic Hyperspectral Analysis Tutorial
Basic Hyperspectral Analysis Tutorial This tutorial introduces you to visualization and interactive analysis tools for working with hyperspectral data. In this tutorial, you will: Analyze spectral profiles
More informationSlideBook 5. FRAP Imaging Module
SlideBook 5 FRAP Imaging Module for Windows XP User s Manual (Latest Revision 7.9.09) 1 Table of Contents TABLE OF CONTENTS...2 CONTACTING INTELLIGENT IMAGING INNOVATIONS, INC...3 MANUAL CONVENTIONS...4
More informationLeica Sp5 II Confocal User Guide
Leica Sp5 II Confocal User Guide Turning on the Confocal System (instructions are posted in the room) 1. Turn on Laser Power Button 2. Turn Key to On position 3. Turn on Scanner Power Button 4. Turn on
More informationTraining Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope
Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope AIM 4.2 Optical Imaging & Vital Microscopy Core Baylor College of Medicine (2017) Power ON Routine 1 2 Verify that main power switches on the
More informationSpreadsheets 3: Charts and Graphs
Spreadsheets 3: Charts and Graphs Name: Main: When you have finished this handout, you should have the following skills: Setting up data correctly Labeling axes, legend, scale, title Editing symbols, colors,
More informationSoftware Manual for the Economy LBIC Demo
printed organic photovoltaics solar testing euipment specialty materials Software Manual for the Economy LBIC Demo Introduction This demonstration program is meant to show the capabilities of the infinitypv
More informationMath Labs. Activity 1: Rectangles and Rectangular Prisms Using Coordinates. Procedure
Math Labs Activity 1: Rectangles and Rectangular Prisms Using Coordinates Problem Statement Use the Cartesian coordinate system to draw rectangle ABCD. Use an x-y-z coordinate system to draw a rectangular
More informationPlotting scientific data in MS Excel 2003/2004
Plotting scientific data in MS Excel 2003/2004 The screen grab above shows MS Excel with all the toolbars switched on - remember that some options only become visible when others are activated. We only
More informationQuick Start MMRC Nicolet 6700 Revised 10/24/2018
Do Not Remove Quick Start MMRC Nicolet 6700 Revised 10/24/2018 1. Fill out the logbook before you begin. 2. Before you start the Nicolet should be on with purge gas flowing. 3. Check that the purge gas
More informationLeica SPEII confocal microscope. Short Manual
Leica SPEII confocal microscope Short Manual Switching ON sequence: 1. Turn on the Workstation under the bench (top, far right). 2. Turn on the Supply Unit - Laser box (big green switch first and then
More informationSpark Spectral Sensor Offers Advantages
04/08/2015 Spark Spectral Sensor Offers Advantages Spark is a small spectral sensor from Ocean Optics that bridges the spectral measurement gap between filter-based devices such as RGB color sensors and
More informationTHIS IS A NEW SPECIFICATION
THIS IS A NEW SPECIFICATION ADVANCED SUBSIDIARY GCE PHYSICS A Electrons, Waves and Photons G482 *OCE/23017* Candidates answer on the Question Paper OCR Supplied Materials: Data, Formulae and Relationships
More informationIntroduction to Spectrometry
Experiment 11 Introduction to Spectrometry Portions adapted by Ross S. Nord and Colleagues, Chemistry Department, Eastern Michigan University from ANAL 043, written by Donald F. Clemens and Warren A. McAllister,
More informationExperiment P55: Light Intensity vs. Position (Light Sensor, Motion Sensor)
PASCO scientific Vol. 2 Physics Lab Manual: P55-1 Experiment P55: (Light Sensor, Motion Sensor) Concept Time SW Interface Macintosh file Windows file illuminance 30 m 500/700 P55 Light vs. Position P55_LTVM.SWS
More informationMicroscopy. The dichroic mirror is an important component of the fluorescent scope: it reflects blue light while transmitting green light.
Microscopy I. Before coming to lab Read this handout and the background. II. Learning Objectives In this lab, you'll investigate the physics of microscopes. The main idea is to understand the limitations
More information1 Placing particles on the slide
Aerosols Transport Particle Removal Experiment E.S. Kenney, J.A. Taylor, and G. Ahmadi 1 Placing particles on the slide 1.1 Swing the light beneath the test section down and to the left. Figure 1: Light
More informationUltraviolet Visible Infrared Instrumentation
Ultraviolet Visible Infrared Instrumentation Focus our attention on measurements in the UV-vis region of the EM spectrum Good instrumentation available Very widely used techniques Longstanding and proven
More informationIBIL setup operation manual for SynerJY software version
IBIL setup operation manual for SynerJY software version 1.8.5.0 Manual version 1.0, 31/10/2008 Author: Carlos Marques Equipment Managers: Carlos Marques, +351219946084, cmarques@itn.pt Luís Alves, +351219946112,
More informationSwept-Field User Guide
Swept-Field User Guide Note: for more details see the Prairie user manual at http://www.prairietechnologies.com/resources/software/prairieview.html Please report any problems to Julie Last (jalast@wisc.edu)
More informationMicroscopy from Carl Zeiss
Microscopy from Carl Zeiss Contents Page Contents... 1 Introduction... 1 Starting the System... 2 Introduction to ZEN Efficient Navigation... 5 Setting up the microscope... 10 Configuring the beam path
More informationTRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO
TRAINING MANUAL Multiphoton Microscopy LSM 510 META-NLO September 2010 Multiphoton Microscopy Training Manual Multiphoton microscopy is only available on the LSM 510 META-NLO system. This system is equipped
More informationSimplified Instructions: Olympus Widefield Microscope S1230
Contents General Microscope Operation Simple Image Capture Multi-Wavelength Capture Z-Series Timelapse Combining Capture Modes Synopsis of Other Functions Pages 2-23 24-40 41-47 48-56 57-59 60-68 69-83
More informationLSM 510 Training Notes
LSM 510 Training Notes Turning on the system Turn on the arc lamp, found on the bench top left of the microscope. This supplies light for epifluorescence for viewing your samples through the microscope.
More informationStandard Operating Procedure
Standard Operating Procedure Nanosurf Atomic Force Microscopy Operation Facility NCCRD Nanotechnology Center for Collaborative Research and Development Department of Chemistry and Engineering Physics The
More informationSpectroscopy Lab 2. Reading Your text books. Look under spectra, spectrometer, diffraction.
1 Spectroscopy Lab 2 Reading Your text books. Look under spectra, spectrometer, diffraction. Consult Sargent Welch Spectrum Charts on wall of lab. Note that only the most prominent wavelengths are displayed
More informationSpectroscopy of Ruby Fluorescence Physics Advanced Physics Lab - Summer 2018 Don Heiman, Northeastern University, 1/12/2018
1 Spectroscopy of Ruby Fluorescence Physics 3600 - Advanced Physics Lab - Summer 2018 Don Heiman, Northeastern University, 1/12/2018 I. INTRODUCTION The laser was invented in May 1960 by Theodor Maiman.
More information[4] (b) Fig. 6.1 shows a loudspeaker fixed near the end of a tube of length 0.6 m. tube m 0.4 m 0.6 m. Fig. 6.
1 (a) Describe, in terms of vibrations, the difference between a longitudinal and a transverse wave. Give one example of each wave.................... [4] (b) Fig. 6.1 shows a loudspeaker fixed near the
More informationComparing Across Categories Part of a Series of Tutorials on using Google Sheets to work with data for making charts in Venngage
Comparing Across Categories Part of a Series of Tutorials on using Google Sheets to work with data for making charts in Venngage These materials are based upon work supported by the National Science Foundation
More informationLA-T LED ANALYSER EVALUATION KIT INSTRUCTION MANUAL. rev
LA-T LED ANALYSER EVALUATION KIT INSTRUCTION MANUAL rev. 300117 TABLE OF CONTENTS General Information 3 Application 3 Design 3 Features 3 Operation conditions 3 Operation instructions 4-7 2 GENERAL INFORMATION
More informationThis procedure assumes the user is already familiar with basic operation of the SEM and the MiraTC interface.
Tescan MIRA3 SEM: EDS using EDAX TEAM Nicholas G. Rudawski ngr@ufl.edu Cell: (805) 252-4916 Office: (352) 392-3077 Last updated: 12/04/17 This procedure assumes the user is already familiar with basic
More informationThe Photoelectric Effect
The Photoelectric Effect 1 The Photoelectric Effect Overview: The photoelectric effect is the light-induced emission of electrons from an object, in this case from a metal electrode inside a vacuum tube.
More informationQuick Guide. NucleoCounter NC-3000
Quick Guide NucleoCounter NC-3000 Table of contents Setting up the FlexiCyte Protocol 2 Editing Image Capture and Analysis Parameters 3 Optimizing Exposure Time 4 Compensation for Spectral Overlap 6 Creating
More informationGraphing with Excel. Data Table
Graphing with Excel Copyright L. S. Quimby There are many spreadsheet programs and graphing programs that you can use to produce very nice graphs for your laboratory reports and homework papers, but Excel
More informationExperiment #12 X-Ray Diffraction Laboratory
Physics 360/460 Experiment #12 X-Ray Diffraction Laboratory Introduction: To determine crystal lattice spacings, as well as identify unknown substances, a x-ray diffractometer is used to replace the traditional
More informationOPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop
OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop 1. Log in usage using the SMIF web site 2. Turn power
More informationISET Selecting a Color Conversion Matrix
ISET Selecting a Color Conversion Matrix Contents How to Calculate a CCM...1 Applying the CCM in the Processor Window...6 This document gives a step-by-step description of using ISET to calculate a color
More informationDual-FL. World's Fastest Fluorometer. Measure absorbance spectra and fluorescence simultaneously FLUORESCENCE
Dual-FL World's Fastest Fluorometer Measure absorbance spectra and fluorescence simultaneously FLUORESCENCE 100 Times Faster Data Collection The only simultaneous absorbance and fluorescence system available
More informationVolume of Revolution Investigation
Student Investigation S2 Volume of Revolution Investigation Student Worksheet Name: Setting up your Page In order to take full advantage of Autograph s unique 3D world, we first need to set up our page
More informationPh 3455 The Photoelectric Effect
Ph 3455 The Photoelectric Effect Required background reading Tipler, Llewellyn, section 3-3 Prelab Questions 1. In this experiment you will be using a mercury lamp as the source of photons. At the yellow
More informationAqualog. CDOM Measurements Made Easy PARTICLE CHARACTERIZATION ELEMENTAL ANALYSIS FLUORESCENCE GRATINGS & OEM SPECTROMETERS OPTICAL COMPONENTS RAMAN
Aqualog CDOM Measurements Made Easy ELEMENTAL ANALYSIS FLUORESCENCE GRATINGS & OEM SPECTROMETERS OPTICAL COMPONENTS PARTICLE CHARACTERIZATION RAMAN SPECTROSCOPIC ELLIPSOMETRY SPR IMAGING CDOM measurements
More informationjust below the screen. Data collection will begin, and a graph will show your data being plotted in real time.
To Collect Additional Data To start a second data collection run, tap the file should now see Run 2 displayed with a blank graph. cabinet in the upper right corner. You just below the screen. Data collection
More informationCREATING (AB) SINGLE- SUBJECT DESIGN GRAPHS IN MICROSOFT EXCEL Lets try to graph this data
CREATING (AB) SINGLE- SUBJECT DESIGN GRAPHS IN MICROSOFT EXCEL 2003 Lets try to graph this data Date Baseline Data Date NCR (intervention) 11/10 11/11 11/12 11/13 2 3 3 1 11/15 11/16 11/17 11/18 3 3 2
More informationAbstract No. 32. Arne Bengtson and Tania Irebo. Swerea KIMAB AB, Isafjordsgatan 28A, SE Kista, Sweden
Abstract No. 32 Ultraviolet Fluorescence using a deep UV LED source and multiple optical filters new possibilities for advanced on-line surface inspection Arne Bengtson and Tania Irebo Swerea KIMAB AB,
More informationThe operation manual of spotlight 300 IR microscope
The operation manual of spotlight 300 IR microscope Make sure there is no sample under the microscope and then click spotlight on the desktop to open the software. You can do imaging with the image mode
More informationThe Spectrometer manual
The Spectrometer manual Data Harvest Group Ltd. 1 Eden Court Leighton Buzzard Bedfordshire LU7 4FY Tel: +44 (0)1525 373666 Fax: +44 (0)1525 851638 e-mail: sales@data-harvest.co.uk support@data-harvest.co.uk
More informationCONFIGURING. Your Spectroscopy System For PEAK PERFORMANCE. A guide to selecting the best Spectrometers, Sources, and Detectors for your application
CONFIGURING Your Spectroscopy System For PEAK PERFORMANCE A guide to selecting the best Spectrometers, s, and s for your application Spectral Measurement System Spectral Measurement System Spectrograph
More informationCreating a Color Compensation file for Roche LightCycler 480 vers.i & vers. II
Creating a Color Compensation file for Roche LightCycler 480 vers.i & vers. II Due to ROX and Cy5 dye emmision spectra overlap Roche LightCycler 480 can pick up signals from a dye measured by another channel.
More informationBasic P-XRD instructions for Operating the Instrument
Basic P-XRD instructions for Operating the Instrument Instrument Parts Incident Beam Optics (left arm) 1) X-ray source (Cu) i. Rest settings: 45 kv, 20mA ii. Run settings: 45 kv, 40mA 2) Monochromator
More informationAnalog Discovery Arbitrary Function Generator for Windows 7 by Mr. David Fritz and Ms. Ellen Robertson
Analog Discovery Arbitrary Function Generator for Windows 7 by Mr. David Fritz and Ms. Ellen Robertson Financial support to develop this tutorial was provided by the Bradley Department of Electrical and
More informationUV / VIS Spectrophotometer EMCLAB Instruments GmbH
A Professional Manufacturer UV / VIS Spectrophotometer EMCLAB Instruments GmbH EMCLAB Instruments GmbH Bismarckstrasse 120 47057 Duisburg Germany Phone: +49 203 3064042 Fax: +49 203 3064044 E-mail: info@emc-lab.de
More informationPhysics 253 Fundamental Physics Mechanic, September 9, Lab #2 Plotting with Excel: The Air Slide
1 NORTHERN ILLINOIS UNIVERSITY PHYSICS DEPARTMENT Physics 253 Fundamental Physics Mechanic, September 9, 2010 Lab #2 Plotting with Excel: The Air Slide Lab Write-up Due: Thurs., September 16, 2010 Place
More informationPage 1 BIOIMAGER. Spectrophotometers. Toll Free: (855) BIO- IMAG / (855) , E: S:
Page 1 Comparison of all models Page 2 Model BK- V1000 BK- UV1000 BK- V1200 BK- UV1200 BK- V1600 BK- UV1600 BK- V1800 BK- UV1800 BK- V1900 BK- UV1900 BK- S360 BK- S380 BK- S390 BK- D560 BK- D580 BK- D590
More informationLSM 510 Meta Training Notes
LSM 510 Meta Training Notes Turning on the system Turn on X-Cite power supply. This supplies light for epifluorescence for viewing your samples through the microscope. Turn on the remote control switch.
More informationSupporting Materials for. A Novel Color Modulation Analysis Strategy. through Tunable Multiband Laser for. Nanoparticle Identification and Evaluation
Supporting Materials for A Novel Color Modulation Analysis Strategy through Tunable Multiband Laser for Nanoparticle Identification and Evaluation Xuan Cao,, Gang Lei, Jingjing Feng, Qi Pan, Xiaodong Wen,
More informationGentec-EO USA. T-RAD-USB Users Manual. T-Rad-USB Operating Instructions /15/2010 Page 1 of 24
Gentec-EO USA T-RAD-USB Users Manual Gentec-EO USA 5825 Jean Road Center Lake Oswego, Oregon, 97035 503-697-1870 voice 503-697-0633 fax 121-201795 11/15/2010 Page 1 of 24 System Overview Welcome to the
More information