The era of noninvasive blood pressure (NIBP) Oscillometric Blood Pressure Measurement: The Methodology, Some Observations, and Suggestions

Size: px
Start display at page:

Download "The era of noninvasive blood pressure (NIBP) Oscillometric Blood Pressure Measurement: The Methodology, Some Observations, and Suggestions"

Transcription

1 Oscillometric Blood Pressure Measurement: The Methodology, Some Observations, and Suggestions Jiri Jilek, MS; Teiichiro Fukushima, MD Most noninvasive blood pressure (NIBP) devices use the oscillometric method. Published studies of oscillometric methodology introduced varied algorithmic approaches for determination of systolic (SBP), diastolic (DBP), and mean arterial (MAP) blood pressures. While there is a general agreement about MAP determination, controversy exists about the determination of SBP and DBP. Accuracy of oscillometric devices has been questioned and validation studies have revealed problems. Several validation protocols have been developed but they are expensive and time consuming to conduct and they have their own limitations. Instruments for bench testing of NIBP devices are useful for some device functions, but they cannot perform dynamic accuracy tests. The issue of accuracy is becoming very important as health care professionals increasingly rely on electronic NIBP devices. The authors developed a compact system for acquisition of NIBP waveforms. Some representative oscillometric waveforms are introduced here to demonstrate the oscillometric method and its shortcomings. A finger photoplethysmograph (PPG) was used to demonstrate a potential improvement of SBP determination. The concept and significance of an oscillometric blood pressure waveform database is introduced and its applications are discussed. (Biomedical Instrumentation & Technology 2005; 39: ). The era of noninvasive blood pressure (NIBP) measurement started with the introduction of the modern sphygmomanometer by Riva- Rocci 1 in The Korotkoff 2 method for determining systolic and diastolic pressure in 1905 marked the beginning of the auscultatory method of blood pressure (BP) determination. This method has been used by health care professionals without substantive changes to the present time. The manual auscultatory method uses a sphygmomanometer (inflatable cuff and a manometer) and a stethoscope. The cuff is usually wrapped around an upper arm and inflated to about 30 mm Hg above the expected systolic blood pressure (SBP). With the stethoscope placed over the brachial artery, the cuff is deflated at a rate of 3 to 5 mm Hg per second. The SBP is determined as the point at which the phase 1 Korotkoff sound From the Department of Obstetrics and Gynecology, King/Drew Medical Center, Los Angeles, CA. Address correspondance and reprint requests to Mr. Jiri Jilek, Department of Obstetrics and Gynecology, Room 2015, King/Drew Medical Center Los Angeles, CA 90059; phone (310) is first heard. The diastolic blood pressure (DBP) is the cuff pressure at which the phase 4 (muffling) or phase 5 (disappearance) of Korotkoff sounds occurs. Disagreements over the use of phase 4 and phase 5 Korotkoff sounds have existed, but the latest guidelines from the Joint National Committee on Prevention, Detection, Evaluation, and Treatment of High Blood Pressure recommends phase 5 (the point before sound disappearance) for detection of DBP. 3 Automatic, electronic NIBP devices started appearing shortly after the introduction of microprocessors. In automated auscultatory NIBP measurement, at least one microphone is used in place of the stethoscope to detect Korotkoff sounds. Another method used in automated NIBP devices is the ultrasound method, which uses ultrasonic detection of arterial wall motion during cuff deflation. The auscultatory or ultrasonic electronic NIBP devices need external sensors in addition to the cuff. The sensors must be placed correctly over the artery for proper function of the device. The need for external sensors is eliminated by the oscillometric Biomedical Instrumentation &Technology 237

2 Oscillometric Blood-Pressure Measurement Figure 1. Oscillometric waveforms (OMW) derived from cuff pressure (CP) curve. Systolic (SBP) and diastolic (DBP) blood pressure reference points were determined by auscultation. Figure 2. Oscillometric waveforms (OMW) contain irregular heart beats. method, which requires only a cuff and a pressure sensor for its operation. Oscillometry Oscillometry was introduced by Marey in 1876, who placed the forearm and the hand in a water-filled chamber to which a variable counterpressure was applied. The counterpressure for maximum amplitude of oscillations determined that the vessel walls were maximally relieved of tension during the cardiac cycle and BP was communicated to the water in the chamber. A similar spectrum of oscillations in cuff pressure was seen when an airinflated cuff was used. For some time, it was thought that the lowest cuff pressure for maximal oscillations was equal to diastolic pressure. This assumption was proven to be incorrect by Posey and Geddes. 4 They showed that the point of maximal oscillations corresponded to true mean-arterial pressure. When pressure in the cuff was increased above mean pressure and then decreased below mean pressure, the amplitude of oscillations decreased. Electronic oscillometric instruments capable of determining SBP, DBP, and mean arterial pressure (MAP) started appearing on the market in the 1970s. Microprocessors facilitated algorithmic methods for the determination of SBP and DBP. One of the first descriptions of a microprocessor-based oscillometric device appeared in Unfortunately, disagreements about oscillometric methodology started appearing in the literature at that time and have continued to appear today. For example, a publication on hypertension 6 makes an assertion that the oscillations in the cuff appear at approximately systolic pressure. This assertion contradicts Geddes work. 7 Geddes showed that the artery pulsates under the upper edge of the cuff at pressures well above SBP and the pulsations are communicated to the cuff via the intervening tissues. Oscillometric waveforms in Figure 1 support the assertion of Geddes. The upper trace represents cuff pressure (CP) decreasing from supersystolic to supradiastolic pressure. The lower trace represents the oscillometric waveforms (OMWs) that were derived from the CP by low-pass filtering and amplification. The oscillometric waveforms appear well above the reference SBP and their amplitudes continue to rise until MAP is reached. Published methods for oscillometric determination of SBP and DBP present differing approaches. Geddes 7 makes certain assumptions about algorithmic determination of SBP and DBP. These algorithms are based on the ratio of oscillometric waveform amplitudes. According to Geddes, 7 SBP corresponds to the point of 50% of maximal amplitude; for DBP, the ratio is 80%. Another author 8 claims that SBP should be at 40% and DBP at 5% of maximal amplitude. The amplitude ratio methods are not the only ones used in oscillometric determination of SBP and DBP. Methods based on the change in the slope of amplitude envelope have been described. The investigators of slope methods differ in their claims as well. An article describing the function of an oscillometric BP device 9 claims that the device determines SBP as the point of the initial increase of the cuff pulsations. Another author 10 puts SBP on the minimal ascending slope of the amplitude envelope and DBP on the maximum slope of the descending amplitude envelope. The above algorithmic approaches result in differing SBP and DBP values. The published reports on algorithmic determination of SBP and DBP do not offer physiological explanation for their assertions. The only commonly recognized and physiologically verified oscillometric variable is MAP. 238 May/June 2005

3 Jiri Jilek, MS; Teiichiro Fukushima, MD INSTRUMENTATION RESEARCH Advantages and Pitfalls of the Oscillometric Method There are many oscillometric NIBP devices on the market today. They range from expensive, rugged instruments for clinical use to miniaturized units that fit on a wrist. Most of these devices use continuous linear or nonlinear deflation methods. Some devices use a stepdeflation method, and some recent monitors use an inflation method. Ease of use has made oscillometric devices very popular. The elimination of external sensors contributed to their popularity at the price of increased uncertainty in the determination of SBP and DBP. Examination of oscillometric waveforms in Figures 1 and 2 reveals the challenge that designers of oscillometric BP devices face in attempting to determine SBP and DBP algorithmically. There are no easily identifiable SBP and DBP points on the waveform envelope. Oscillometric BP determination is complicated by other factors, such as movement artifacts, deep breathing, tremors, and arrhythmias. Arrhythmias introduce especially difficult problems. Their nature and frequency of occurrence are not always apparent. The waveform envelope in Figure 2 contains irregular beats. The uneven sequence of waveform amplitudes results in unpredictable BP determination. Reference SBP and DBP could not be determined reliably by auscultation. An external sensor located distally from the cuff diminishes the uncertainty of oscillometric SBP determination. Figure 2 illustrates this point. The upper trace represents pressure waveforms obtained with a finger photoplethysmograph (PPG) and the lower trace shows OMWs. When the cuff is gradually deflated from supersystolic pressure, the oscillometric waveforms are present but the PPG baseline is flat. When the CP decreases below SBP, turbulent blood flow passes past the cuff. Blood volume and pressure in the finger begin to rise due to blocked venous blood return. The PPG trace reflects the rising blood volume and arterial pulsations. The availability of a PPG signal makes the determination of SBP easier than reliance on just the oscillometric waveforms. The waveforms shown in Figures 1 through 3 were acquired with an experimental data-acquisition system. The system consists of a BP cuff; a module containing pneumatic components, pressure circuits, and data acquisition electronics; and a notebook computer (Figure 4). Cuff pressure is converted to analog voltage by a pressure sensor. Figure 3. Oscillometric waveforms (OMW; bottom trace) and finger photoplethysmogragh (PPG; upper trace) waveforms. Systolic blood pressure (SBP) is the reference value determined by auscultation. The pressure-sensor output is amplified by an instrumentation amplifier and filtered by a low-pass antialiasing filter. A high-pass filter is used to eliminate the cuff deflation component of the signal and an amplifier is used to obtain waveforms of sufficient amplitude for digitization. A commercially available finger PPG was used to acquire the PPG waveforms. A PPG uses a light source to illuminate a circumscribed portion of the tissue. The intensity picked up by a photoelectric sensor is then related to the blood volume in the respective area. 11 Cuff pressure voltage, amplified oscillometric waveforms, and finger PPG signals are digitized by a 4-channel, 12-bit analog-to-digital converter. The sampling rate is 11.8 milliseconds. The digitized data are transmitted to the notebook computer via a serial interface. The computer controls acquisition and processing of pressure data. The data used in this report were acquired from volunteers in informal settings. Reference BPs were measured using the standard auscultation method. Oscillometric BP Device Performance Testing and Validations All oscillometric devices use algorithms for BP determination, but one can only speculate what algorithms they use. Published algorithms show disagreements, and device manufacturers consider their algorithms proprietary and keep them secret. 12 This makes verification of accuracy difficult. There are several NIBP test instruments on the market today. They can perform useful tests, such as static-pressure accuracy, cuff deflation rates, leak tests, and overpressure tests. They cannot, however, perform dynamic algorithmic NIBP accuracy tests. The operation manual of the popular NIBP ana- Biomedical Instrumentation &Technology 239

4 Oscillometric Blood-Pressure Measurement Figure 4. Oscillometric and photoplethysmograph (PPG) waveform acquisition system. lyzer CuffLink (Fluke Biomedical; Reno, Nevada) contains a chapter Introduction to NIBP Monitoring. Oscillometric techniques are briefly described in this chapter and the following conclusions are drawn: Various interpretations have emerged from the manufacturers with varying degrees of agreement. At present, no regulatory agency has put forth a standard as to how oscillometric pulse amplitudes should be interpreted to determine blood pressure. Therefore, the accuracy and repeatability of these monitors is difficult to determine. Should any public standards emerge, CuffLink may be programmed to implement them to test blood pressure monitor accuracy. Absolute dynamic accuracy cannot be assigned to MAP, diastolic, and systolic target values at this time because no standard exists. The issue of accuracy is becoming increasingly important as many health care institutions rely heavily on automatic NIBP devices. Because there are no reliable instruments for testing of algorithmic accuracy, performance-testing protocols for device validations have been developed. The Association for the Advancement of Medical Instrumentation, 13 the British Hypertension Society, 14 and the European Society of Hypertension 15 recommend validation of NIBP devices against auscultation or against intra-arterial methods. Many validation studies have been conducted and some reviews of validation results have been published. 16,17 Their findings indicate that the accuracy of BP determination is problematic for many NIBP devices. Validation protocols are not without problems either. A recently published study 18 exposed limitations of current validation protocols. The study concludes that the existing protocols are likely to pass devices that can be systematically inaccurate for some patients. Disappointing validation results, lack of information from device manufacturers, and errors observed in health care institutions have led to warnings issued by experts in the field of BP measurement. The American Heart Association issued an advisory statement from the Council for High Blood Pressure Research. 19 The Council cautioned health care professionals not to abandon mercury sphygmomanometers until adequate replacement instruments are available. A recent report by a group of leading experts stressed the importance of accurate BP measurements. 20 The report called for additional research to assess accuracy of NIBP devices and concluded that mercury sphygmomanometer remains the gold standard for noninvasive BP measurement. The Case for a Public Database of Oscillometric BP Waveforms Oscillometric BP waveforms are indispensable for noninvasive determination of BPs and they may contain other useful information. An investigator or a device developer who wants to study oscillometric waveforms needs a reasonably large database of waveforms and reference BP measurements. Manufacturers of oscillometric BP devices must have such databases in order to conduct their development efficiently. These databases are, however, proprietary. There are no publicly accessible databases of oscillometric waveforms at the present time. On the other hand, public databases for some physiologic waveforms do exist, mainly for interpretation of electrocardiograms. The National Center for Research Resources (NCRR) operates a website, PhysioNet. PhysioNet offers free access to a collection of recorded physiologic signals. Oscillometric BP waveforms are not included in the present PhysioNet collection. General principles of acquisition and use of physiological waveforms are described in the Association for the Advancement of Medical Instrumentation Technical Information Report. 21 The report stresses the necessity to test algorithmic functions of digital devices with real physiologic data. Properly documented databases are needed for such testing. The waveforms can then be used to test devices repeatedly and reproducibly. A wide-ranging, publicly available database of oscillometric BP waveforms could advance the field of oscillometric BP measurement in the following ways: New research into the largely unknown physiological basis of oscillometric BP measurement. The research could result in the development of a generic algorithmic method for the determination of SBP and DBP. 240 May/June 2005

5 Jiri Jilek, MS; Teiichiro Fukushima, MD Device developers would enjoy the advantage of not having to develop their own proprietary databases, as the past and present manufacturers had to do. Costs of development and time to market could be decreased. A standardized, public database would serve as a common knowledge base and it should produce devices performing in a similar, predictable manner. Repeatable, reproducible performance testing of oscillometric BP devices could become possible. The expensive, time-consuming testing as performed today could eventually be eliminated. Determination of hemodynamic variables. It may be possible to derive cardiac output (CO), total peripheral resistance, and arterial compliance from oscillometric waveform contours and blood pressures. Several contour methods for CO determination already exist. 22,23 A computerized, compact system similar to the system developed by the authors could be used to develop the database. The oscillometric and reference BP data would be acquired from a varied population and a wide range of blood pressures. The acquired, carefully annotated database could be made available to the public via PhysioNet or other media. Conclusion Oscillometric method for NIBP measurement as used in the state-of-the-art devices is attractive for the simplicity of cuff application and device operation. On the other hand, the empirical nature of existing algorithmic methods, the disagreements over methodology, the proprietary nature of commercial algorithms, and the resulting problems with accuracy have led to warnings and calls for more research. The proposed oscillometric BP waveform database could facilitate such studies. The widespread use of oscillometric BP devices and the importance of accurate BP measurements justify increased efforts devoted to the improvement of oscillometric methodology. References 1. Riva-Rocci S. Un sfigmomanometro Nuevo. Gaz Med Trino. 1896: Korotkov NS. A contribution to the problem of methods for the determination of the blood pressure. In: Ruskin A, ed. Classics of Arterial Hypertension. Thomas, Springfield, Ill., 1956: Chobanian AV, Bakris GL, Black HR, et al. The seventh report of the Joint National Committee on Prevention, Detection, Evaluation, and Treatment of High Blood Pressure. JAMA. 2003;289: Posey JA, Geddes LA. The meaning of the point of maximum oscillations in cuff pressure in the indirect measurement of blood pressure. Cardiovasc Res Bul. 1969;8: Looney J. Blood pressure by oscillometry. Med Electron. April 1978: Oparil S, Weber MA, ed. Hypertension: A Companion to Brenner and Rector s The Kidney. WB Saunders; Geddes LA. Characterization of the oscillometric method for measuring indirect blood pressure. Ann Biomed Eng. 1982;10: Sapinsky A. Standard algorithm of blood pressure measurement by oscillographic method. Med Biol Eng Comput. 1992;30: Borow KM, Newburger JW. Noninvasive estimation of central aortic pressure using the oscillometric method for analyzing systemic artery pulsatile blood flow: comparative study of indirect systolic, diastolic and mean brachial artery pressure with simultaneous direct ascending aortic pressure measurements. Am Heart J. 1982;103: NG KG. Blood pressure measurement. Med Electron. February 1999: Fronek A. Noninvasive Diagnostics in Vascular Disease. McGraw-Hill; New York, NY, Jilek J. Electronic sphygmomanometers: the problems and some suggestions. Biomed Instrument Technol. 2003;37: ANSI/AAMI. SP10:2002, Manual, Electronic or Automated Sphygmomanometers. Arlington, Va: ANSI/AAMI; O Brien E, Petrie J, Littler W, et al. An outline of the revised British Hypertension Society protocol for the evaluation of blood pressure monitoring devices. Blood Press Monit. 2002;7: O Brien E, Pickering T, Asmar R, Myers M, Parati G, Staessen J. Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults. Blood Press Monit. 2002;7: O Brien E, Waeber B, Parati G, Staessen J, Myers MG. Blood pressure measuring devices: recommendations of the European Society of Hypertension. BMJ. 2001;322: Van Egmond J, Lenders JW, Weernink E, Thien T. Accuracy and reproducibility of 30 devices for self-measurement of arterial blood pressure. Am J Hypertens. 1993; 6: Gerin W, Schwartz AR, Pickering TG, Davidson KW, Bress J, O Brien E. Limitations of current validation protocols for home blood pressure monitors for individual patients. Blood Press Monit. 2002;7(6): Jones DW, Frohlich ED, Grim CM, Grim CE, Taubert KA. Mercury sphygmomanometers should not be abandoned: an advisory statement from the Council for High Blood Pressure Research, American Heart Association. Hypertension. 2001;37: Jones DW, Appel LJ, Sheps SG, Roccella EJ, Lenfant C. Measuring blood pressure accurately. JAMA. 2003;289: AAMI. Acquisition and Use of Physiologic Waveform Databases for Testing of Medical Devices (technical information report, AAMI TIR no. 24), Arlington, Va: AAMI; Romano SM, Pistolesi M. Assessment of cardiac output from systemic arterial pressure in humans. Crit Care Med. 2002;30: Linton NWF, Linton RAF. Estimation of changes in cardiac output from the arterial blood pressure waveform in the upper limb. Br J Aneasth. 2001;86(4): Biomedical Instrumentation &Technology 241

Laboratory Kit for Oscillometry Measurement of Blood Pressure

Laboratory Kit for Oscillometry Measurement of Blood Pressure Laboratory Kit for Oscillometry Measurement of Blood Pressure Jan Dvořák, Jan Havlík Department of Circuit Theory Faculty of Electrical Engineering, Czech Technical University in Prague Technická 2, CZ-16627

More information

Lab: Blood Pressure. Goal: Design and test a bandpass filter that can isolate a blood pressure signal.

Lab: Blood Pressure. Goal: Design and test a bandpass filter that can isolate a blood pressure signal. Page /10 1 Lab: Blood Pressure Goal: Design and test a bandpass filter that can isolate a blood pressure signal. This week you will design and build a system to estimate your mean arterial pressure (blood

More information

Towards inexpensive home Ambulatory BP Monitors [Work in Progress]

Towards inexpensive home Ambulatory BP Monitors [Work in Progress] Towards inexpensive home Ambulatory BP Monitors [Work in Progress] 27 July 2009 Larry Beaty labeaty@ieee.org Phoenix Project, Twin Cities IEEE See http://www.phoenix.tc.ieee.org/ then sign up as a volunteer

More information

WRIST BAND PULSE OXIMETER

WRIST BAND PULSE OXIMETER WRIST BAND PULSE OXIMETER Vinay Kadam 1, Shahrukh Shaikh 2 1,2- Department of Biomedical Engineering, D.Y. Patil School of Biotechnology and Bioinformatics, C.B.D Belapur, Navi Mumbai (India) ABSTRACT

More information

THIS SPEC IS OBSOLETE

THIS SPEC IS OBSOLETE THIS SPEC IS OBSOLETE Spec No: 001-62651 Spec Title: PSOC(R) 3 / PSOC 5 - BLOOD PRESSURE MONITOR ANALOG FRONT END - AN62651 Sunset Owner:Praveen Sekar (PFZ) Replaced by: None PSoC 3 / PSoC 5 - Blood Pressure

More information

V850ES/Jx3L USB and 78K0R/Lx3

V850ES/Jx3L USB and 78K0R/Lx3 Introduction APPLICATION NOTE R01AN0713EU0100 Rev.1.00 The measurement of the arterial blood pressure is of great clinical significance mainly for the detection and follow-up of hypertension which affects

More information

Research Article Monitoring Personalized Trait Using Oscillometric Arterial Blood Pressure Measurements

Research Article Monitoring Personalized Trait Using Oscillometric Arterial Blood Pressure Measurements Applied Mathematics Volume 2012, Article ID 591252, 12 pages doi:10.1155/2012/591252 Research Article Monitoring Personalized Trait Using Oscillometric Arterial Blood Pressure Measurements Young-Suk Shin

More information

* Notebook is excluded. Features KL-720 contains nine modules, including Electrocardiogram Measurement, E lectromyogram Measurement,

* Notebook is excluded. Features KL-720 contains nine modules, including Electrocardiogram Measurement, E lectromyogram Measurement, KL-720 Biomedical Measurement System Supplied by: 011 683 4365 This equipment is intended for students to learn how to design specific measuring circuits and detect the basic physiological signals with

More information

The Removal Of Motion Artifacts From Noninvasive Blood Pressure Measurements

The Removal Of Motion Artifacts From Noninvasive Blood Pressure Measurements University of Central Florida Electronic Theses and Dissertations Masters Thesis (Open Access) The Removal Of Motion Artifacts From Noninvasive Blood Pressure Measurements 2004 Paresh Thakkar University

More information

PHOTOPLETHYSMOGRAPHIC DETECTOR FOR PERIPHERAL PULSE REGISTRATION

PHOTOPLETHYSMOGRAPHIC DETECTOR FOR PERIPHERAL PULSE REGISTRATION PHOTOPLETHYSMOGRAPHIC DETECTOR FOR PERIPHERAL PULSE REGISTRATION Tatyana Dimitrova Neycheva, Dobromir Petkov Dobrev Centre of Biomedical Engineering Ivan Daskalov Bulgarian Academy of Sciences, Bl. 105

More information

JorVet. Hand-Held Doppler. Operating Manual. Jorgensen Laboratories, Inc. Loveland, Colorado 80538

JorVet. Hand-Held Doppler. Operating Manual. Jorgensen Laboratories, Inc. Loveland, Colorado 80538 JorVet Hand-Held Doppler Operating Manual Jorgensen Laboratories, Inc. Loveland, Colorado 80538 Thank You! for choosing the JorVet Hand-Held Doppler! This choice acknowledges the many advantages of the

More information

Arterial pulse waves measured with EMFi and PPG sensors and comparison of the pulse waveform spectral and decomposition analysis in healthy subjects

Arterial pulse waves measured with EMFi and PPG sensors and comparison of the pulse waveform spectral and decomposition analysis in healthy subjects Arterial pulse waves measured with EMFi and PPG sensors and comparison of the pulse waveform spectral and decomposition analysis in healthy subjects Matti Huotari 1, Antti Vehkaoja 2, Kari Määttä 1, Juha

More information

Design of Virtual Sphygmomanometer Based on LABVIEWComparison, Reflection, Biological assets, Accounting standard.

Design of Virtual Sphygmomanometer Based on LABVIEWComparison, Reflection, Biological assets, Accounting standard. Design of Virtual Sphygmomanometer Based on LABVIEWComparison, Reflection, Biological assets, Accounting standard. Li Su a, Boxin Zhang b School of electronic engineering, Xi'an Aeronautical University,

More information

*Notebook is excluded

*Notebook is excluded Biomedical Measurement Training System This equipment is designed for students to learn how to design specific measuring circuits and detect the basic physiological signals with practical operation. Moreover,

More information

Design and Validation of an. Arterial Pulse Wave Analysis

Design and Validation of an. Arterial Pulse Wave Analysis Design and Validation of an Arterial Pulse Wave Analysis Device Geoffrey Douglas Salter A dissertation submitted to the Faculty of Engineering and the Built Environment, University of the Witwatersrand,

More information

This work is licensed under a Creative Commons Attribution 3.0 License. For more information, see

This work is licensed under a Creative Commons Attribution 3.0 License. For more information, see Pan F, He P, Liu C, Li T, Murray A, Zheng D. Variation of the Korotkoff stethoscope sounds during blood pressure measurement: Analysis using a convolutional neural network. IEEE Journal of 2017, (99) DOI

More information

Doppler in Obstetrics: book by K Nicolaides, G Rizzo, K Hecher. Chapter on Doppler ultrasound: principles and practice by Colin Deane

Doppler in Obstetrics: book by K Nicolaides, G Rizzo, K Hecher. Chapter on Doppler ultrasound: principles and practice by Colin Deane Doppler in Obstetrics: book by K Nicolaides, G Rizzo, K Hecher Chapter on Doppler ultrasound: principles and practice by Colin Deane INTRODUCTION Competent use of Doppler ultrasound techniques requires

More information

Central Blood Pressure Meter Model cbp301. Operating Manual

Central Blood Pressure Meter Model cbp301. Operating Manual Central Blood Pressure Meter Model cbp301 Operating Manual Document cbp301-009 Issue 4 September 2011 Contents Introduction... 3 Package Contents... 5 Warnings and Cautions... 6 Contraindications... 7

More information

GE Healthcare. Dash 2500 The standard of excellence for sub-acuity monitoring

GE Healthcare. Dash 2500 The standard of excellence for sub-acuity monitoring GE Healthcare Dash 2500 The standard of excellence for sub-acuity monitoring The clinical With monitored patients, even small measurement variations can indicate significant changes in your patient s condition.

More information

Biosignal Data Acquisition and its Post-processing

Biosignal Data Acquisition and its Post-processing Biosignal Data Acquisition and its Post-processing MILAN CHMELAR 1, RADIM CIZ 2, ONDREJ KRAJSA 2, JIRI KOURIL 2 Brno University of Technology 1 Department of Biomedical Engineering Kolejni 4, 612 00 Brno

More information

Design of Arterial Blood Pressure, Heart Rate Variability, and Breathing Rate Monitoring Device. Mastan Singh Kalsi

Design of Arterial Blood Pressure, Heart Rate Variability, and Breathing Rate Monitoring Device. Mastan Singh Kalsi Design of Arterial Blood Pressure, Heart Rate Variability, and Breathing Rate Monitoring Device by Mastan Singh Kalsi Electrical and Biomedical Engineering Design Project (4BI6) Department of Electrical

More information

Development of an Active Sensor for non-invasive Arterial Blood Pressure Monitoring

Development of an Active Sensor for non-invasive Arterial Blood Pressure Monitoring Development of an Active Sensor for non-invasive Arterial Blood Pressure Monitoring Viacheslav Antsiperov, Gennady Mansurov Laboratory of Diagnostic Systems Kotel nikov Institute of REE of RAS Moscow,

More information

Computer Evaluation of Exercise Based on Blood Volume Pulse (BVP) Waveform Changes

Computer Evaluation of Exercise Based on Blood Volume Pulse (BVP) Waveform Changes Computer Evaluation of Exercise Based on Blood Volume Pulse (BVP) Waveform Changes ARMANDO BARRETO 1,2, CHAO LI 1 and JING ZHAI 1 1 Electrical & Computer Engineering Department 2 Biomedical Engineering

More information

MULTIPLE PULSE WAVE MEASUREMENT TOWARD ESTIMATING CONDITION OF HUMAN ARTERIES

MULTIPLE PULSE WAVE MEASUREMENT TOWARD ESTIMATING CONDITION OF HUMAN ARTERIES IADIS International Journal on WWW/Internet Vol. 11, No. 3, pp. 116-125 ISSN: 1645-7641 MULTIPLE PULSE WAVE MEASUREMENT TOWARD ESTIMATING CONDITION OF HUMAN Shusaku Nomura. Nagaoka University of Technology.

More information

AN INTELLIGENT SYSTEM FOR CONTINUOUS BLOOD PRESSURE MONITORING ON REMOTE MULTI-PATIENTS IN REAL TIME

AN INTELLIGENT SYSTEM FOR CONTINUOUS BLOOD PRESSURE MONITORING ON REMOTE MULTI-PATIENTS IN REAL TIME AN INTELLIGENT SYSTEM FOR CONTINUOUS BLOOD PRESSURE MONITORING ON REMOTE MULTI-PATIENTS IN REAL TIME Roberto Marani and Anna Gina Perri Electrical and Information Engineering Department, Electronic Devices

More information

Low-cost photoplethysmograph solutions using the Raspberry Pi

Low-cost photoplethysmograph solutions using the Raspberry Pi Low-cost photoplethysmograph solutions using the Raspberry Pi Tamás Nagy *, Zoltan Gingl * * Department of Technical Informatics, University of Szeged, Hungary nag.tams@gmail.com, gingl@inf.u-szeged.hu

More information

City, University of London Institutional Repository

City, University of London Institutional Repository City Research Online City, University of London Institutional Repository Citation: Rybynok, V., May, J.M., Budidha, K. and Kyriacou, P. A. (2013). Design and Development of a novel Multi-channel Photoplethysmographic

More information

Journal of Applied Research and Technology ISSN: Centro de Ciencias Aplicadas y Desarrollo Tecnológico.

Journal of Applied Research and Technology ISSN: Centro de Ciencias Aplicadas y Desarrollo Tecnológico. Journal of Applied Research and Technology ISSN: 1665-6423 jart@aleph.cinstrum.unam.mx Centro de Ciencias Aplicadas y Desarrollo Tecnológico México Fuentes González, Rosendo; Bañuelos Saucedo, Miguel Angel

More information

Signal Extraction Technology

Signal Extraction Technology Signal Extraction Technology Technical bulletin Introduction Masimo SET pulse oximetry is a new and fundamentally distinct method of acquiring, processing and reporting arterial oxygen saturation and pulse

More information

Blood Flow Analysis Using Ultrasound Doppler Technique

Blood Flow Analysis Using Ultrasound Doppler Technique IOSR Journal of Electronics and Communication Engineering (IOSR-JECE) e-issn: 2278-2834,p- ISSN: 2278-8735.Volume 10, Issue 5, Ver. I (Sep - Oct.2015), PP 31-35 www.iosrjournals.org Blood Flow Analysis

More information

DOPPLEX VASCULAR DOPPLERS. ...with people in mind

DOPPLEX VASCULAR DOPPLERS. ...with people in mind DOPPLEX VASCULAR DOPPLERS...with people in mind VASCULAR DOPPLEX DOPPLERS Based on over 20 years experience in this field, the latest generation of the world renowned Dopplex handheld doppler range offers

More information

HIGH FREQUENCY FILTERING OF 24-HOUR HEART RATE DATA

HIGH FREQUENCY FILTERING OF 24-HOUR HEART RATE DATA HIGH FREQUENCY FILTERING OF 24-HOUR HEART RATE DATA Albinas Stankus, Assistant Prof. Mechatronics Science Institute, Klaipeda University, Klaipeda, Lithuania Institute of Behavioral Medicine, Lithuanian

More information

Making sense of electrical signals

Making sense of electrical signals Making sense of electrical signals Our thanks to Fluke for allowing us to reprint the following. vertical (Y) access represents the voltage measurement and the horizontal (X) axis represents time. Most

More information

MASIMO RADICAL 7 Signal Extraction Pulse CO-Oximeter

MASIMO RADICAL 7 Signal Extraction Pulse CO-Oximeter MASIMO RADICAL 7 Signal Extraction Pulse CO-Oximeter Women s Health Manual MCH Only Policy Group: Cardiovascular Approved by: Heather Crosland Director, Women s Health, Covenant Health, GNH/MCH Site Lead

More information

Design and Implementation of Digital Stethoscope using TFT Module and Matlab Visualisation Tool

Design and Implementation of Digital Stethoscope using TFT Module and Matlab Visualisation Tool World Journal of Technology, Engineering and Research, Volume 3, Issue 1 (2018) 297-304 Contents available at WJTER World Journal of Technology, Engineering and Research Journal Homepage: www.wjter.com

More information

HUMAN BODY MONITORING SYSTEM USING WSN WITH GSM AND GPS

HUMAN BODY MONITORING SYSTEM USING WSN WITH GSM AND GPS HUMAN BODY MONITORING SYSTEM USING WSN WITH GSM AND GPS Mr. Sunil L. Rahane Department of E & TC Amrutvahini College of Engineering Sangmaner, India Prof. Ramesh S. Pawase Department of E & TC Amrutvahini

More information

Statistical Pulse Measurements using USB Power Sensors

Statistical Pulse Measurements using USB Power Sensors Statistical Pulse Measurements using USB Power Sensors Today s modern USB Power Sensors are capable of many advanced power measurements. These Power Sensors are capable of demodulating the signal and processing

More information

Implementation of wireless ECG measurement system in ubiquitous health-care environment

Implementation of wireless ECG measurement system in ubiquitous health-care environment Implementation of wireless ECG measurement system in ubiquitous health-care environment M. C. KIM 1, J. Y. YOO 1, S. Y. YE 2, D. K. JUNG 3, J. H. RO 4, G. R. JEON 4 1 Department of Interdisciplinary Program

More information

Principle of Pulse Oximeter. SpO2 = HbO2/ (HbO2+ Hb)*100% (1)

Principle of Pulse Oximeter. SpO2 = HbO2/ (HbO2+ Hb)*100% (1) Design of Pulse Oximeter Simulator Calibration Equipment Pu Zhang, Jing Chen, Yuandi Yang National Institute of Metrology, East of North Third Ring Road, Beijing, China,100013 Abstract -Saturation of peripheral

More information

Robust Wrist-Type Multiple Photo-Interrupter Pulse Sensor

Robust Wrist-Type Multiple Photo-Interrupter Pulse Sensor Robust Wrist-Type Multiple Photo-Interrupter Pulse Sensor TOSHINORI KAGAWA, NOBUO NAKAJIMA Graduate School of Informatics and Engineering The University of Electro-Communications Chofugaoka 1-5-1, Chofu-shi,

More information

Printed in Japan E318

Printed in Japan E318 Diagnostic Ultrasound System MODEL PROSOUND 6 The specifications, shape and color of this product are subject to change without notice. The standard components and optional items vary depending on the

More information

Masimo Corporation 40 Parker Irvine, California Tel Fax

Masimo Corporation 40 Parker Irvine, California Tel Fax Instruments and sensors containing Masimo SET technology are identified with the Masimo SET logo. Look for the Masimo SET designation on both the sensors and monitors to ensure accurate pulse oximetry

More information

Lesson 12: Doppler Principles. This lesson contains 50 slides plus 26 multiple-choice questions.

Lesson 12: Doppler Principles. This lesson contains 50 slides plus 26 multiple-choice questions. Lesson 12: Doppler Principles This lesson contains 50 slides plus 26 multiple-choice questions. Accompanying text for the slides in this lesson can be found on pages 59 through 80 in the textbook: DOPPLER

More information

ECONOMICAL HEART RATE MEASUREMENT DEVICE WITH REMOTE MONITORING USING FINGERTIP

ECONOMICAL HEART RATE MEASUREMENT DEVICE WITH REMOTE MONITORING USING FINGERTIP ECONOMICAL HEART RATE MEASUREMENT DEVICE WITH REMOTE MONITORING USING FINGERTIP PROJECT REFERENCE NO. : 37S1390 COLLEGE : SRI SIDDHARTHA INSTITUTE OF TECHNOLOGY, TUMKUR. BRANCH : TELECOMMUNICATION ENGINEERING

More information

An Automated Algorithm for Fast Pulse Wave Detection

An Automated Algorithm for Fast Pulse Wave Detection An Automated Algorithm for Fast Pulse Wave Detection Bistra Nenova, Ivo Iliev * Technical University Sofia 8 Kliment Ohridski Blvd., 1 Sofia, Bulgaria E-mail: izi@tu-sofia.bg * Corresponding author Received:

More information

MCU-Based Multi Parameter Patient Monitor

MCU-Based Multi Parameter Patient Monitor International Journal of Sciences: Basic and Applied Research (IJSBAR) ISSN 2307-4531 (Print & Online) http://gssrr.org/index.php?journal=journalofbasicandapplied ---------------------------------------------------------------------------------------------------------------------------

More information

1. Blood pressure (BP) module and accessories

1. Blood pressure (BP) module and accessories I. Introduction The VetSpecs BP-3 is developed specifically for surgical monitoring and blood pressure screening in cats, dogs, and other similar sized animals. The BP-3 Plus incorporates a VetSpecs proprietary

More information

Chapter 2. Design and development of blood volume pulse sensor and heart rate meter. Abstract

Chapter 2. Design and development of blood volume pulse sensor and heart rate meter. Abstract Chapter 2 Design and development of blood volume pulse sensor and heart rate meter Abstract A low power, low cost sensor has been developed for sensing the blood volume pulse using transmission mode photoplethysmography

More information

NIBP2010 with SpO2. Non Invasive Blood Pressure OEM board NIBP with. PULSE OXIMETRY SpO 2. Hardware-Version : C Firmware-Version : 3.

NIBP2010 with SpO2. Non Invasive Blood Pressure OEM board NIBP with. PULSE OXIMETRY SpO 2. Hardware-Version : C Firmware-Version : 3. Non Invasive Blood Pressure OEM board NIBP 2010 with PULSE OXIMETRY SpO 2 Hardware-Version : C Firmware-Version : 3.44 Issued: L. Engel Approved: B. Tek Released: B. Tek Date: 24.01.17 Date : 24.01.17

More information

School of Electronic Science and Engineering, Nanjing University of Posts and Telecommunications, Nanjing, , China

School of Electronic Science and Engineering, Nanjing University of Posts and Telecommunications, Nanjing, , China 4th International Conference on Mechatronics, Materials, Chemistry and Computer Engineering (ICMMCCE 2015) A design and implementation of Pulse-Measure instrument based on Microcontroller Zhu Siqing1,

More information

Excellent Performance; Ease of Use

Excellent Performance; Ease of Use Excellent Performance; Ease of Use High performance with a broad range of applications - the compact ProSound α6 is at your service. Our ProSound series has a well-established reputation in hospitals and

More information

Module 1: Introduction to Experimental Techniques Lecture 2: Sources of error. The Lecture Contains: Sources of Error in Measurement

Module 1: Introduction to Experimental Techniques Lecture 2: Sources of error. The Lecture Contains: Sources of Error in Measurement The Lecture Contains: Sources of Error in Measurement Signal-To-Noise Ratio Analog-to-Digital Conversion of Measurement Data A/D Conversion Digitalization Errors due to A/D Conversion file:///g /optical_measurement/lecture2/2_1.htm[5/7/2012

More information

DESIGN OF A PHOTOPLETHYSMOGRAPHY BASED PULSE RATE DETECTOR

DESIGN OF A PHOTOPLETHYSMOGRAPHY BASED PULSE RATE DETECTOR DESIGN OF A PHOTOPLETHYSMOGRAPHY BASED PULSE RATE DETECTOR Srijan Banerjee 1, Subhajit Roy 2 1 Department of Electrical Engineering, Siliguri Institute of Technology, 2 Department of Electrical Engineering,

More information

Application Notes. Electrosurgical Analyzer Primer

Application Notes. Electrosurgical Analyzer Primer Electrosurgical Analyzer Primer Purpose Scope Overview The purpose is to outline the characteristics of electrosurgical analyzers, describe the challenges of measuring RF power and review the inspection

More information

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity)

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity) Vascular Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO Medical Systems Division, Shimadzu Corporation Yoshiaki Miura 1. Introduction In recent years, digital cardiovascular

More information

Design of Wearable Pulse Oximeter Sensor Module for Capturing PPG Signals

Design of Wearable Pulse Oximeter Sensor Module for Capturing PPG Signals Design of Wearable Pulse Oximeter Sensor Module for Capturing PPG Signals Mr. Vishwas Nagekar 1, Mrs Veena S Murthy 2 and Mr Vishweshwara Mundkur 3 1 Department of ECE, BNMIT, Bangalore 2 Assoc. Professor,

More information

NIBP2O}O UP. with PULSE OXIMETRY. spot. Hardware-Version : A Firmware-Version : 6.0. Signature, ffi

NIBP2O}O UP. with PULSE OXIMETRY. spot. Hardware-Version : A Firmware-Version : 6.0. Signature, ffi parmedizinrechnik ^*g;tli*"iji1$ä, 1.0 "oc.-rev. I Non Invasive Blood Pressure OEM board NIBP2O}O UP with PULSE OXIMETRY spot Hardware-Version : A Firmware-Version : 6.0 Erstellt: L. Engel Geprüft: A.

More information

City, University of London Institutional Repository

City, University of London Institutional Repository City Research Online City, University of London Institutional Repository Citation: Zaman, T., Kyriacou, P. A. & Pal, S. (2013). Free flap pulse oximetry utilizing reflectance photoplethysmography. 35th

More information

understand compatibility of photoplethysmographic pulse rate variability with electrocardiogramic heart rate variability

understand compatibility of photoplethysmographic pulse rate variability with electrocardiogramic heart rate variability Loughborough University Institutional Repository A preliminary attempt to understand compatibility of photoplethysmographic pulse rate variability with electrocardiogramic heart rate variability This item

More information

Lab #9: Compound Action Potentials in the Toad Sciatic Nerve

Lab #9: Compound Action Potentials in the Toad Sciatic Nerve Lab #9: Compound Action Potentials in the Toad Sciatic Nerve In this experiment, you will measure compound action potentials (CAPs) from an isolated toad sciatic nerve to illustrate the basic physiological

More information

Noninvasive Radial Pressure Waveform Estimation by Transfer Functions Using Particle Swarm Optimization

Noninvasive Radial Pressure Waveform Estimation by Transfer Functions Using Particle Swarm Optimization Proceedings of the 7th WSEAS International Conference on Systems Theory and Scientific Computation, Athens, Greece, August 4-6, 7 63 Noninvasive Radial Pressure Waveform Estimation by Transfer Functions

More information

An algorithm to estimate the transient ST segment level during 24-hour ambulatory monitoring

An algorithm to estimate the transient ST segment level during 24-hour ambulatory monitoring ELEKTROTEHNIŠKI VESTNIK 78(3): 128 135, 211 ENGLISH EDITION An algorithm to estimate the transient ST segment level during 24-hour ambulatory monitoring Aleš Smrdel Faculty of Computer and Information

More information

PULSE OXIMETRY MODULE TO IMPLEMENT IN TEAM MONITOR OF VITAL SIGNS

PULSE OXIMETRY MODULE TO IMPLEMENT IN TEAM MONITOR OF VITAL SIGNS PULSE OXIMETRY MODULE TO IMPLEMENT IN TEAM MONITOR OF VITAL SIGNS A. Soto Otalora 1, L. A. Guzman Trujilloy 2 and A. DiazDiaz 3 1 Industrial Control Engineering, Universidad Surcolombiana Neiva, Avenida

More information

Uncertainty factors in time-interval measurements in ballistocardiography

Uncertainty factors in time-interval measurements in ballistocardiography Uncertainty factors in time-interval measurements in ballistocardiography Joan Gomez-Clapers 1, Albert Serra-Rocamora 1, Ramon Casanella 1, Ramon Pallas-Areny 1 1 Instrumentation, Sensors and Interfaces

More information

Internet Based Artificial Neural Networks for the Interpretation of Medical Images

Internet Based Artificial Neural Networks for the Interpretation of Medical Images Internet Based Artificial Neural Networks for the Interpretation of Medical Images Andreas Järund, Lars Edenbrandt Department of Clinical Physiology, Lund University, Lund, Sweden andreas.järund@klinfys.lu.se

More information

common type of cardiac diseases and may indicate an increased risk of stroke or sudden cardiac death. ECG is the most

common type of cardiac diseases and may indicate an increased risk of stroke or sudden cardiac death. ECG is the most ISSN: 0975-766X CODEN: IJPTFI Available Online through Research Article www.ijptonline.com DESIGNING OF ELECTRONIC CARDIAC EVENTS RECORDER *Dr. R. Jagannathan, K.Venkatraman, R. Vasuki and Sundaresan Department

More information

EG medlab. Three Lead ECG OEM board. Version Technical Manual. Medlab GmbH Three Lead ECG OEM Module EG01010 User Manual

EG medlab. Three Lead ECG OEM board. Version Technical Manual. Medlab GmbH Three Lead ECG OEM Module EG01010 User Manual Medlab GmbH Three Lead ECG OEM Module EG01010 User Manual medlab Three Lead ECG OEM board EG01010 Technical Manual Copyright Medlab 2008-2016 Version 1.03 1 Version 1.03 28.04.2016 Medlab GmbH Three Lead

More information

Measurement of Blood Pressure Using an Arterial Pulsimeter Equipped with a Hall Device

Measurement of Blood Pressure Using an Arterial Pulsimeter Equipped with a Hall Device Sensors 2011, 11, 1784-1793; doi:10.3390/s110201784 OPEN ACCESS sensors ISSN 1424-8220 www.mdpi.com/journal/sensors Article Measurement of Blood Pressure Using an Arterial Pulsimeter Equipped with a Hall

More information

monitoring device ought to find wide clinical application. available, while section three describes the Doppler interface with the B-scan machine.

monitoring device ought to find wide clinical application. available, while section three describes the Doppler interface with the B-scan machine. THE YALE JOURNAL OF BIOLOGY AND MEDICINE 50 (1977), 367-373 Pulse-Doppler Ultrasound and Its Clinical Application PETER ATKINSON AND PETER N.T. WELLS Yale University School of Medicine, New Haven, Connecticut

More information

OPERATING MANUAL MINIDOP ES-100VX POCKET DOPPLER

OPERATING MANUAL MINIDOP ES-100VX POCKET DOPPLER OPERATING MANUAL MINIDOP ES-100VX POCKET DOPPLER CONTENTS * Features.......................... 1 * Cautions.......................... 2 * Clinical applications................. 4 * Operating controls..................

More information

Medical Imaging (EL582/BE620/GA4426)

Medical Imaging (EL582/BE620/GA4426) Medical Imaging (EL582/BE620/GA4426) Jonathan Mamou, PhD Riverside Research Lizzi Center for Biomedical Engineering New York, NY jmamou@riversideresearch.org On behalf of Prof. Daniel Turnbull Outline

More information

NeuVision 500. Abundant and friendly display interface, multifold ECG display screen:

NeuVision 500. Abundant and friendly display interface, multifold ECG display screen: NeuVision 500 Features This monitoring system may be used to monitor patient s 6 physiological parameters: ECG, respiratory rate, body temperature, non-invasive blood pressure (NIBP), pulse oxygen saturation

More information

ARTICLE IN PRESS. Computers in Biology and Medicine

ARTICLE IN PRESS. Computers in Biology and Medicine Computers in Biology and Medicine 39 (2009) 1145 1152 Contents lists available at ScienceDirect Computers in Biology and Medicine journal homepage: www.elsevier.com/locate/cbm Adaptive threshold method

More information

ELR 4202C Project: Finger Pulse Display Module

ELR 4202C Project: Finger Pulse Display Module EEE 4202 Project: Finger Pulse Display Module Page 1 ELR 4202C Project: Finger Pulse Display Module Overview: The project will use an LED light source and a phototransistor light receiver to create an

More information

GTIN:

GTIN: GTIN: 00894912002920 TABLE OF CONTENTS Introduction...1 Features...1 Indications for Use...1 Contraindications...1 Warnings & Precautions...2 Controls...3 Operation...4 Recharging the Battery...5 Replacing

More information

Security and Privacy for Health Care Applications

Security and Privacy for Health Care Applications Security and Privacy for Health Care Applications Yih-Chun Hu University of Illinois at Urbana-Champaign May 7, 203 Story Time Who is the adversary? NSFNet The power grid Maps provided by geni.org and

More information

Doppler Ultrasound Blood Flow Measurement System for Assessing Coronary Revascularization

Doppler Ultrasound Blood Flow Measurement System for Assessing Coronary Revascularization Doppler Ultrasound Blood Flow Measurement System for Assessing Coronary Revascularization J. Solano 1, M. Fuentes 1, A. Villar 2 J. Prohias 2, F. García-Nocetti 1 1 Universidad Nacional Autónoma de México,

More information

Bio Medical Instrumentation's

Bio Medical Instrumentation's Bio Medical Instrumentation's DAE First Year Biomedical Technology Table of Contents S. No Topic Name Page. No 01 Introduction to biomedical technology 01 02 Blood Pressure apparatus 01 03 Function of

More information

Validation of the Happify Breather Biofeedback Exercise to Track Heart Rate Variability Using an Optical Sensor

Validation of the Happify Breather Biofeedback Exercise to Track Heart Rate Variability Using an Optical Sensor Phyllis K. Stein, PhD Associate Professor of Medicine, Director, Heart Rate Variability Laboratory Department of Medicine Cardiovascular Division Validation of the Happify Breather Biofeedback Exercise

More information

Three-dimensional display of calculated velocity profiles for physiological flow waveforms

Three-dimensional display of calculated velocity profiles for physiological flow waveforms Three-dimensional display of calculated velocity profiles for physiological flow waveforms Ramez E. N. Shehada, MSc, Richard S. C. Cobbold, Phi), K. Wayne Johnston, MD, FRCS(C), and Rene Aarnink, MSc,

More information

Device Equivalence Evaluation Form

Device Equivalence Evaluation Form Device Equivalence Evaluation Form Comparison of the PAR Medizintechnik PHYSIO-PORT with the PAR Medizintechnik TONOPORT VI Devices Item 9 PHYSIO-PORT TONOPORT VI Pictures Display Image The complete screen

More information

INTRODUCTION. 4 SAFETY INSTRUCTIONS. 5 ABOUT THIS DEVICE. 6 FIRST OPERATION. 7 THE DISPLAY. 8 ATTACHING THE WRIST SLEEVE. 9 CORRECT MEASUREMENT.

INTRODUCTION. 4 SAFETY INSTRUCTIONS. 5 ABOUT THIS DEVICE. 6 FIRST OPERATION. 7 THE DISPLAY. 8 ATTACHING THE WRIST SLEEVE. 9 CORRECT MEASUREMENT. Table of contents INTRODUCTION.............................. 4 SAFETY INSTRUCTIONS........................ 5 ABOUT THIS DEVICE.......................... 6 FIRST OPERATION............................ 7

More information

Biomedical Signal Processing and Applications

Biomedical Signal Processing and Applications Proceedings of the 2010 International Conference on Industrial Engineering and Operations Management Dhaka, Bangladesh, January 9 10, 2010 Biomedical Signal Processing and Applications Muhammad Ibn Ibrahimy

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD IEC 62092 First edition 2001-08 Utrasonics Hydrophones Characteristics and calibration in the frequency range from 15 MHz to 40 MHz Ultrasons Hydrophones Caractéristiques et étalonnage

More information

(51) Int Cl.: A61B 5/00 ( ) G06F 17/00 ( )

(51) Int Cl.: A61B 5/00 ( ) G06F 17/00 ( ) (19) (11) EP 1 424 934 B1 (12) EUROPEAN PATENT SPECIFICATION (4) Date of publication and mention of the grant of the patent: 06.08.08 Bulletin 08/32 (21) Application number: 01981641.2 (22) Date of filing:

More information

User instructions for the ABI-system 100

User instructions for the ABI-system 100 User instructions for the ABI-system 100 0124 User instructions for the ABI-system 100 1 ABI-system 100 2 CA04 cuffs (for upper arms) 2 CL04 cuffs (for legs) 1 ABI software CD 1 USB cable 1 boso mains

More information

Operation Guide Model: 405-HT V3.0

Operation Guide Model: 405-HT V3.0 MODEL 405-HT Operation Guide Model: 405-HT V3.0 Fully Automatic Wrist Cuff Blood Pressure Monitor (ELECTRONIC SPHYGMOMANOMETER) OPERATION GUIDE INDEX IMPORTANT INFORMATION... 2 CONTENTS AND DISPLAY INDICATORS...

More information

Biomechanical Instrumentation Considerations in Data Acquisition ÉCOLE DES SCIENCES DE L ACTIVITÉ PHYSIQUE SCHOOL OF HUMAN KINETICS

Biomechanical Instrumentation Considerations in Data Acquisition ÉCOLE DES SCIENCES DE L ACTIVITÉ PHYSIQUE SCHOOL OF HUMAN KINETICS Biomechanical Instrumentation Considerations in Data Acquisition Data Acquisition in Biomechanics Why??? Describe and Understand a Phenomena Test a Theory Evaluate a condition/situation Data Acquisition

More information

Bayesian Estimation of Tumours in Breasts Using Microwave Imaging

Bayesian Estimation of Tumours in Breasts Using Microwave Imaging Bayesian Estimation of Tumours in Breasts Using Microwave Imaging Aleksandar Jeremic 1, Elham Khosrowshahli 2 1 Department of Electrical & Computer Engineering McMaster University, Hamilton, ON, Canada

More information

Dash 3000, 4000 & 5000

Dash 3000, 4000 & 5000 GE Healthcare Dash 3000, 4000 & 5000 Flexible acuity monitoring Bedside flexibility and adaptability The Dash monitoring family is a portable monitoring system that is flexible and easy to use. The adaptability

More information

Design Considerations for Wrist- Wearable Heart Rate Monitors

Design Considerations for Wrist- Wearable Heart Rate Monitors Design Considerations for Wrist- Wearable Heart Rate Monitors Wrist-wearable fitness bands and smart watches are moving from basic accelerometer-based smart pedometers to include biometric sensing such

More information

3880 Technical Specifications_ DRAFT_A Page 1 of 11

3880 Technical Specifications_ DRAFT_A Page 1 of 11 3880 Technical Specifications_ DRAFT_A Page 1 of 11 3880 3880 Technical Specifications_ DRAFT_A Page 2 of 11 Type: Color TFT resistive touchscreen Screen Size: 25.7 cm (10.1 inches) diagonal Pixels: 800

More information

Hardware. MRI System. MRI system Multicoil Microstrip. Part1

Hardware. MRI System. MRI system Multicoil Microstrip. Part1 Hardware MRI system Multicoil Microstrip MRI System Part1 1 The MRI system is made up of a variety of subsystems. the Operator Workspace Gradient Driver subsystem The Physiological Acquisition Controller

More information

UGEO H60. Performance in Style. Features

UGEO H60. Performance in Style. Features UGEO H60 Performance in Style The UGEO H60 implements superior performance with new design principles of simplicity and lightness. Its 10.1" touchscreen improves usability while its 18.5" LED monitor enhances

More information

Biomedical Equipment Technician

Biomedical Equipment Technician Biomedical Equipment Technician Occupational Skill Stards Texas Skill Stards Board Recognized Critical Work Function 1. Install Biomedical Equipment 1.1 Receive, inspect inventory upon delivery 1.2 Deliver

More information

HAL S3010. A Neonate at 40 Weeks Gestational Age

HAL S3010. A Neonate at 40 Weeks Gestational Age HAL S3010 A Neonate at 40 Weeks Gestational Age Simulation Made Easy Proven HAL technology Tetherless with wireless communication Fully responsive even while being carried Modeling and trending Comprehensive

More information

Measurement of Spatial Pulse Wave Velocity by Using a Clip-Type Pulsimeter Equipped with a Hall Sensor and Photoplethysmography

Measurement of Spatial Pulse Wave Velocity by Using a Clip-Type Pulsimeter Equipped with a Hall Sensor and Photoplethysmography Sensors 2013, 13, 4714-4723; doi:10.3390/s130404714 Article OPEN ACCESS sensors ISSN 1424-8220 www.mdpi.com/journal/sensors Measurement of Spatial Pulse Wave Velocity by Using a Clip-Type Pulsimeter Equipped

More information

Ambulatory Blood Pressure Monitor Via Pulse Wave Transit Time. Final Report

Ambulatory Blood Pressure Monitor Via Pulse Wave Transit Time. Final Report Ambulatory Blood Pressure Monitor Via Pulse Wave Transit Time Final Report Michael Adamson, David Hoff, Jason Malisheski, Daniel Yilma and Jasmin Zukic Faculty Advisor: Murti Salapaka, UMN Industry Sponsor:

More information

Introduction to Oscilloscopes Instructor s Guide

Introduction to Oscilloscopes Instructor s Guide Introduction to Oscilloscopes A collection of lab exercises to introduce you to the basic controls of a digital oscilloscope in order to make common electronic measurements. Revision 1.0 Page 1 of 25 Copyright

More information

An Advanced Architecture & Instrumentation for Developing the System of Monitoring a Vital Sign (Oxygen Saturation) of a Patient.

An Advanced Architecture & Instrumentation for Developing the System of Monitoring a Vital Sign (Oxygen Saturation) of a Patient. An Advanced Architecture & Instrumentation for Developing the System of Monitoring a Vital Sign (Oxygen Saturation) of a Patient. 1 Md.Mokarrom Hossain, 2 A.S.M.Mohsin*, 3 Md.Nasimul Islam Maruf, 4 Md.

More information