Sefelec. MG SERIES Revision : R MANUAL FOR MG SERIES DIELECTRIC TEST UNITS, GROUND CONTINUITY TESTERS AND LEAKAGE CURRENT

Size: px
Start display at page:

Download "Sefelec. MG SERIES Revision : R MANUAL FOR MG SERIES DIELECTRIC TEST UNITS, GROUND CONTINUITY TESTERS AND LEAKAGE CURRENT"

Transcription

1 MG SERIES Revision : R MANUAL FOR MG SERIES DIELECTRIC TEST UNITS, GROUND CONTINUITY TESTERS AND LEAKAGE CURRENT Sefelec Parc d'activités du Mandinet - 19, rue des Campanules LOGNES - FRANCE Phone : Téléfax : Sefelec@wanadoo.fr http :

2 THIS MANUAL IS WRITTEN FOR THE UNITS OF THE MG SERIES: CMG30: EARTH CONTUITY TESTING DMG50, DMG500, DMG500F: DIELECTRIMETERS MMG500: MEGOHMMETERS RMG50, RMG500, RMG500F, RMG15AC: DIELECTRIC STRENGTH TESTERS SMG50, SMG500, SMG500F: ELECTRICAL SAFETY TESTERS FMG500, FMG501: LEAKAGE CURRENT TESTER PLEASE REFER ONLY TO THE SECTIONS OF THIS MANUAL DESCRIBING YOUR OWN UNIT. WARRANTY: SEFELEC warrants that units are free from defects in material and workmanship. SEFELEC warrants also that, when properly used, that units will perform in accordance with specifications of this manual. If within one year after original delivery it is found not to meet this standard, it will be repaired at no charge in SEFELEC service facility in Lognes. Changes in the unit not approved by SEFELEC will cancel this warranty. SEFELEC will not be liable for any indirect damages resulting of the use of the unit. This warranty is in lieu of all other warranties. Page : 2 MG SERIE Revision : R

3 RESTRICTIONS DUE TO THE ACCESSORIES OR TO THE OPTIONS A) GROUND CONTINUITY RÉSISTANCE - CO184/3 to CO184/10 : Maximum current regulation 10A for 6V, 20A for 12V. - CO183/3 to CO183/10 : Maximum current regulation 10A for 6V, 20A for 12V. - TE66/3 to TE66/10 : Maximum current regulation 10A for 6V, 20A for 12V. - TE80/3 to TE80/10 : Maximum current regulation 10A for 6V, 20A for 12V. - TE81/3 to TE81/10 : Maximum current regulation 10A for 6V, 20A for 12V. - CS1 : Maximum current regulation 10A for 6V, 20A for 12V. B) DIELECTRIC STRENGTH - CO174 : 4000V limited max. voltage. - CO185 : 4000V limited max. voltage. - CO192 : 4000V limited max. Voltage. - CO193 : 4000V limited max. voltage. - CO200 to CO209 : 4000V limited max. voltage. - FMG501 : 4200V limited max. voltage. C) INSULATION MEASUREMENT - DMG500F and SMG500F (Floating output) : 2GΩ measurement limited. - CO178 : 2GΩ measurement limited. - TE59 : 2GΩ measurement limited. - TE67 : 2GΩ measurement limited. - TE69 : 2GΩ measurement limited. - FMG501 and MG-55 Or MG-57option (Triphased leakage) : 2GΩ measurement limited. MG SERIE Revision : R Page : 3

4 CONTENTS SPECIFICATIONS...7 A. Dielectric strength operation 50VA (RMG50, DMG50, SMG50)...8 B. Dielectric strength operation 500VA (RMG500, DMG500, SMG500, RMG500F, DMG500F, SMG500F)...10 C. Dielectric strength operation 500VA (RMG15AC)...12 D. Megohmmeter function...14 E. Ground continuity résistance function...16 SECTION 1: INTRODUCTION - OPERATING INSTRUCTIONS MEANING OF THE DIFFERENT SYMBOLS ON THE INSTRUMENT PRODUCT OVERVIEW FRONT PANEL DESCRIPTION Keyboard definition Liquid Crystal display description REAR PANEL DESCRIPTION : ACCESSORIES INCLUDED ACCESSORIES / OPTIONS AVAILABLE : INSTALLATION Preliminary instructions Power on Safety advice Connection to a specimen...26 SECTION 2 : SETUP LANGUAGE BEEP ON FAIL FILTER PARAM ACCESS DISPLAY REMOTE TRIGGER MODE INTERFACE SELECTION SEQUENCE (OPTION 04)...34 SECTION 3 : INSULATION RESISTANCE MEASUREMENT (DMG50, DMG500, MMG500, SMG50, SMG500) PARAMETERS Measurement, voltage selection Threshold selection Timer Parameters memorization INSULATION RESISTANCE MEASUREMENT ERROR MESSAGES: UNIT EQUIPPED WITH ANALOG METER (21 option)...41 SECTION 4 : DIELECTRIC STRENGTH TEST (DMG50, DMG500, RMG50, RMG500, SMG50, SMG500) Test voltage selection Current threshold selection Breakdown detection mode selection Timer Parameters memorization DIELECTRIC STRENGTH TEST...49 Page : 4 MG SERIE Revision : R

5 4.3 ERROR MESSAGES...50 SECTION 5 : GROUND CONTINUITY RESISTANCE MEASUREMENT (CMG30, SMG50, SMG500) Parameters setting Test current selection Test voltage selection Thresholds selection Timer Parameters memorizations Ground continuity resistance measurement Error messages...58 SECTION 6 : LEAKAGE CURRENT MEASUREMENT...59 SECTION 7 : TEST SEQUENCING (SMG50, SMG500) Setting of measurement functions Sequence setup Parameters memorization Function selection Function memory selection SERIES OF TESTS WITH MANUAL CONTROL SIMPLE Sequence performing Sequence performing with manual control REPEAT FUNCTION OF THE GROUND CONTINUITY Proceeding to multiple ground continuity Multiple ground continuity followed by a strength test Error messages Sequence error messages Function error messages...68 SECTION 8 : INPUT-OUTPUT INTERFACE Input-Output for Programmable Logic Controller Interface (PLC) - 02 option Electrical characteristics Description of the logical states Connections: Description of the input output signals Measurement / Discharge cycles VOLTS ANALOG INPUT-OUTPUTS - 03 OPTION Electrical specifications Possible setups according to the units Change of the board setup Operating the 0 to 10 volts feature Megohmmeter mode, 2 outputs setup Megohmmeter mode, 1 output setup Dielectric strength test mode, 2 outputs setup Dielectric strength test mode, 1 output + 1 input setup Operating instructions RS232 INTERFACE (01 OPTION) Syntax rules List of the RS232C HINTS AND PROGRAM EXAMPLES TROUBLE SHOOTING THE RS232C INTERFACE QUICK BASIC PROGRAM EXAMPLE MG SERIE Revision : R Page : 5

6 8.4 Printing output (92 option) Parameters setup Operation Configuration SECTION 9 : APPLICATION NOTES WHY DIELECTRIC TESTS? GLOSSARY OF TERMS ENVIRONMENTAL INFLUENCES INSULATION RESISTANCE MEASUREMENT PRECAUTIONS TO BE OBSERVED MEASUREMENTS OF CAPACITORS MEASUREMENTS ON CABLES MEASUREMENT VOLTAGE SELECTION DIELECTRIC STRENGTH TESTS SELECTION OF THE TEST VOLTAGE BREAKDOWN DETECTION MODE SELECTION Protective earth continuity measurement Current selection Voltage selection Test time Precautions to be observed SECTION 10: THEORETICAL PRINCIPLES UNIT ARCHITECTURE MICROPROCESSOR BOARD DESCRIPTION INSULATION RESISTANCE MEASUREMENT FUNCTION DIELECTRIC STRENGTH TEST FUNCTION Ground continuity resistance measurement function SECTION 11: MAINTENANCE AND CALIBRATION PRELIMINARY INSTRUMENT RETURN MAINTENANCE CLEANING CALIBRATION SECTION 12: DRAWINGS General Drawing DMG General Drawing CMG General Drawing SMG Mother board Microprocessor board Dielectric strength board Megohmmeter board Ground continuity board Page : 6 MG SERIE Revision : R

7 SPECIFICATIONS Power supply : Mains : 115/230V~ ±10% single phase 47 to 63 Hz. (CMG30, DMG50, MMG500, RMG50, SMG50). Mains : 230V~ ±10% single phase 47 to 63 Hz. (DMG500, RMG500, SMG500, DMG500F, RMG500F, SMG500F, FMG500, FMG501). Power consumption : 40VA without load, 550 VA maximum. Switching from 115 to 230 volts by switch in the line input plug. Temporized fuse protective in rear panel - RMG50, DMG50,SMG50, MMG500 : 2AT for 230V 4AT for 110V. - FMG500 et FMG501 : 2AT for 230V - RMG500, DMG500, SMG500, RMG500F, DMG500F, SMG500F, RMG15AC : 3.15AT for 230V. Conditions of use : The instrument must be used inside, in horizontal position or on tripode. Temperature : In storage : -10 C to +60 C. In operation : 0 C to +45 C. Accuracy is rated after half an hour of warm up and for a relative humidity < 50%. Altitude : up to 2000 meters Max. humidity rate : 80% for a temperature of 31 C. Weight and dimensions : MMG500 CMG30 DMG50 RMG50 DMG500 - RMG500 DMG500F - RMG500F FMG500 - FMG501 SMG50 SMG500 SMG500F RMG15AC Weight : 9Kg 15Kg 21Kg 27Kg 28Kg Height : Width : Depth : 131mm 344mm 332mm 177mm 427mm 465mm Note : As SMG50, SMG500, SMG500F and RMG15AC are over 25Kg, 2 persons are necessary for manipulations. Over voltage category : CAT II. Rate of pollution : Pollution 2 : Occasional conductive pollution only by condensation. Security range : Range I instrument : Earth protection by mains connection. MG SERIE Revision : R Page : 7

8 A. Dielectric strength operation 50VA (RMG50, DMG50, SMG50) A.1 Output voltage Alternative 50 or 60Hz ( DC voltage on option). From 0 to 5 kvac in one range (0 to 6 KVDC). Limited at 4.2 kvac with FMG501 (leakage current) Stability < ±1% for V mains of ±10%. Positive pole to earth ( for the DC option). Residual < 1% for Is=100 µa (for the DC option). Internal resistance : 0.5 MΩ. Accuracy of the output voltage : ±(2%+50volts) in relation with the set value for voltages between 100 and 5000 volts (6000 volts for DC option) and for a current < 100 µa in the fault detection modes : I, IMAX or I+IMAX. Discharge of the tested specimen and of the internal capacity through a 1.5 MΩ resistance (for the DC option). A.2 Voltage reading By kilovoltmeter directly connected to the output terminals. Accuracy : ±(1.5%+20volts) of the read value. Display by 600 points digital indicator. A.3 Short circuit current 10 ma +0%/ -20% for the adjustment of the maximum voltage (time limited to 5 seconds). A.4 Fault detection Either by variation of I current : The TEST detector automatically carries out the substraction between the current which flows normally in the device under test (I=U/Z) and the one which occurs at the time of a fault (I'=I+Ifault). Set amplitude : 1 ma ±10% Width of current pulse : 10µS ±20% By IMAX adjustable from 0.01mA up to 9.99mA by 0.01mA step By combination of both previous modes : I+IMAX By FIMAX adjustable from 0.01mA up to 9.99mA by 0.01mA step By combination of both previous modes : I+FIMAX Possibility to inhibit the detector in the OFF mode: the high voltage remains on the output in case of breakdown. Warning : in that mode, the hold time is limited to 5 seconds (otherwise possibility of thermal breakdown) and there is no automatic adjustment of the output voltage in accordance with the load of the specimen under test. Possibility in the above detection modes to set a minimum value of current flowing through the specimen under test : IMIN value adjustable from 0.00mA up to 9.99mA. A.5 Leakage current control Direct reading of the current on a shunt resistance in the test circuit. Display of the value on a 999 points digital indicator. Page : 8 MG SERIE Revision : R

9 Accuracy : ±(2.5% + 2 U) of the read value (1U=0.01mA). In VDC for resistance loads > 1 MΩ. Memory storage of the leakage current in IMAX detection mode. A.6 Fault indication By message on the LCD display, red and green LED, and audible signal (possibility to inhibit). Memory storage of the breakdown voltage. Memory storage of the leakage current in IMAX mode. Cut off of the high voltage at the first zero crossing of the control sinusoidal signal of the HV transformer, thus without over voltage. A.7 Timing MANUAL mode (without timing) and AUTO mode (with timing). Application of the test voltage during rise, hold, fall times adjustable from 000 up to 999 seconds. The time for each voltage step is 1 second (i.e. 1000V/ 5 seconds = 5 steps of 200 volts) MG SERIE Revision : R Page : 9

10 B. Dielectric strength operation 500VA (RMG500, DMG500, SMG500, RMG500F, DMG500F, SMG500F) B.1 Output voltage Alternative 50 or 60Hz according to the mains frequency ( DC voltage on option). From 0 to 5 kvac in one range (0 to 6 KVDC). Limited at 4.2kVAC with FMG501 (leakage current) Stability < ±3% for V mains of ±10%. Positive pole to earth ( for the DC option). Accuracy of the output voltage : ±(3%+50volts) in relation with the set value for voltages between 100 and 5000 volts (6000 volts for DC option) and for a current < 1mA in the fault detection modes : I, IMAX or I+IMAX. In fault detection mode : OFF, the output voltage changes according to the variation of the load connected on the HV output. Discharge of the tested specimen and of the internal capacity through a 1.5 MΩ resistance (for the DC option). B.2 Voltage reading By kilovoltmeter directly connected to the output terminals. Accuracy : ±(1.5%+20volts) of the read value. Display by 600 points digital indicator. B.3 Short circuit current 200 ma for the adjustment of the maximum AC voltage. 20 ma for the adjustment of the maximum DC voltage. B.4 Fault detection Either by variation of I current : The TEST detector automatically carries out the subtraction between the current which flows normally in the sample under test (I=U/Z) and the one which occurs at the time of a fault (I'=I+Ifault). Set amplitude : 10 ma ±10% Width of current pulse : 10µS ±20% By IMAX adjustable from 0.1mA up to 99.9mA by 0.1mA step By combination of both previous modes : I+IMAX By FIMAX adjustable from 0.1mA up to 99.9mA by 0.1mA step By combination of both previous modes : I+FIMAX Possibility to inhibit the detector in the OFF mode: the high voltage remains on the output in case of breakdown. Warning : in that mode, the hold time is limited to 5 seconds (otherwise possibility of thermal breakdown) and there is no automatic adjustment of the output voltage in accordance with the load of the specimen under test. * For RMG500F, DMG500 and SMG500F the TEST detector characteristics are not valid. Page : 10 MG SERIE Revision : R

11 Possibility in the above detection modes to set a minimum value of current flowing through the specimen under test : IMIN value adjustable from 00.0mA up to 99.9mA. From 2.44X Version : IMAX valueadjustable from 0.1mA up to 110mA by 0.1mA step. IMIN value adjustable from 00.0mA up to 110mA. B.5 Leakage current control Direct reading of the current on a shunt resistance in the test circuit. Display of the value on a 999 points digital indicator. Accuracy : ±(2.5% + 2 U) of the read value (1U=0.1mA). Memory storage of the leakage current in IMAX detection mode. Display of «STRONG CURRENT» message as well as «---MA» when the current is over 100mA (AC current) The current indicated on the display in DC option is the real efficient current. That is to say Iaff = I ac 2 + Idc 2 Display of «STRONG CURRENT» message as well as «---MA» when current is over 20mA (DC current). From 2.44X Version : Display of «STRONG CURRENT» message as well as «---MA» when the current is over 110mA (AC current) B.6 Fault indication By message on the LCD display, red and green LED, and audible signal (possibility to inhibit). Memory storage of the breakdown voltage. Memory storage of the leakage current in IMAX mode. Cut off of the high voltage at the first zero crossing of the control sinusoidal signal of the HV transformer, thus without over voltage. B.7 Timing MANUAL mode (without timing) and AUTO mode (with timing). Application of the test voltage during rise, hold, fall times adjustable from 000 up to 999 seconds. The time for each voltage step is 1 second (i.e. 1000V/ 5 seconds = 5 steps of 200 volts) B.8 Insulation impedance for SMG500F, RMG500F and DMG500F High voltage outputs insulation impedance in relation to the ground > 200 MΩ (VDE0104). 100V : 5GΩ, 500V : 4 GΩ Typical MG SERIE Revision : R Page : 11

12 C. Dielectric strength operation 500VA (RMG15AC) C.1 Output voltage Alternative 50 or 60Hz according to the mains frequency. From 0 to 15 kvac in one range Stability < ±3% for V mains of ±10%. Accuracy of the output voltage : ±(3%+50volts) in relation with the set value for voltages between 100 and volts and for a current < 1mA in the fault detection modes : I, IMAX, FIMAX, I+FIMAX or I+IMAX. In fault detection mode : OFF, the output voltage changes according to the variation of the load connected on the HV output. C.2 Voltage reading By kilovoltmeter directly connected to the output terminals. Accuracy : ±(1.5%+20volts) of the read value. Display by 1500 points digital indicator. C.3 Short circuit current 65 ma for the adjustment of the maximum AC voltage. C.4 Nominal current 35 ma for the adjustment of the maximum AC voltage. C.5 Fault detection Either by variation of I current : The TEST detector automatically carries out the subtraction between the current which flows normally in the sample under test (I=U/Z) and the one which occurs at the time of a fault (I'=I+Ifault). Set amplitude : 10 ma ±10% Width of current pulse : 10µS ±20% By IMAX adjustable from 0.1mA up to 40.0mA by 0.1mA step By combination of both previous modes : I+IMAX By FIMAX adjustable from 0.1mA up to 40.0mA by 0.1mA step By combination of both previous modes : I+FIMAX Possibility to inhibit the detector in the OFF mode: the high voltage remains on the output in case of breakdown. Warning : in that mode, the hold time is limited to 7 seconds (otherwise possibility of thermal breakdown) and there is no automatic adjustment of the output voltage in accordance with the load of the specimen under test. Possibility in the above detection modes to set a minimum value of current flowing through the specimen under test : IMIN value adjustable from 00.0mA up to 40.0mA. Page : 12 MG SERIE Revision : R

13 C.6 Leakage current control MG series OPERATING MANUAL - SEFELEC Direct reading of the current on a shunt resistance in the test circuit. Display of the value on a 999 points digital indicator. Accuracy : ±(2.5% + 3 U) of the read value (1U=0.1mA). Memory storage of the leakage current in IMAX detection mode. Display of «STRONG CURRENT» when the current is over 40mA C.7 Fault indication By message on the LCD display, red and green LED, and audible signal (possibility to inhibit). Memory storage of the breakdown voltage. Memory storage of the leakage current in IMAX mode. Cut off of the high voltage at the first zero crossing of the control sinusoidal signal of the HV transformer, thus without over voltage. C.8 Timing MANUAL mode (without timing) and AUTO mode (with timing). Application of the test voltage during rise, hold, fall times adjustable from 000 up to 999 seconds. The time for each voltage step is 1 second (i.e. 1000V/ 5 seconds = 5 steps of 200 volts) MG SERIE Revision : R Page : 13

14 D. Megohmmeter function 1) 500V Insulation card D.1 Measurement voltage 50, 100, 250, 500 DC volts. Accuracy of the voltage : ±(1%+2V). Positive pole grounded. Dynamic stability for V mains = ±10% : >1% Maximum intensity in the measurement circuit : 2 ma ±20%. Discharge current limited by a 2.2 kω resistance. D.2 Ranges of measurement Basic 200GΩ Version Option 2TΩ Version Voltages Ranges of measurement* Voltages Ranges of measurement 50V 50kΩ to 20GΩ 50V 50kΩ to 200GΩ 100V 100kΩ to 40GΩ 100V 100kΩ to 400GΩ 250V 250kΩ to 100GΩ 250V 250kΩ to 1TΩ 500V 500kΩ to 200GΩ 500V 500kΩ to 2TΩ * Ranges of measurement for DMG500F and SMG500F limited to 2GΩ. D.3 Accuracy of the resistance measurement Display on 2000 points digital indicator with unity indications ( KΩ, MΩ,...) Accuracy: (in % of the reading. 1U= 1 display count) : Megohmmeter (MMG500) Dielectrimeter (DMG50, DMG500) Security Tester (SMG50, SMG50) 200 GΩ Version : ±(1.5% + 1U) 200 GΩ Version : ±(1.5% + 1U) 2 TΩ Version : ±(2% + 1U) 2 TΩ Version* (Uessai=50 or 100V) : ±(2% + 1U) 2 TΩ Version* (Uessai=250 or 500V) : ±(1% x Uessai/ U) *Only on DMG50 and DMG500 CAPACITANCE mode: from 1.00 MΩ to 200 GΩ (2 TΩ for option 20) With accuracy = (NORMAL mode accuracy) ±100 kω. Input impedance = 10 MΩ ±1% D.4 Measurement thresholds Adjustment of the digital value of 2 resistance thresholds from 50 kω up to 200 GΩ (2 TΩ for option 20). Comparison result indication on 2 LED and audible signal (possibility to inhibit). D.5 Timing Permanent measurement or during a time between 1 and 999 seconds (adjustable by 1 second step). Page : 14 MG SERIE Revision : R

15 1) 1000V Insulation card (MG-26 option) Those specifications replace the ones from the 500V insulation card, given on page 6. D.1 Measurement voltage 100, 250, 500, 1000 DC volts. Accuracy of the voltage : ±(1%+2V). Positive pole grounded. Dynamic stability for V mains = ±10% : >1% Maximum intensity in the measurement circuit : 2 ma ±20%. Discharge current limited by a 2.2 kω resistance. D.2 Ranges of measurement Basic 200GΩ Version Option 2TΩ Version Voltages Ranges of measurement* Voltages Ranges of measurement 100V 100kΩ to 20GΩ 100V 100kΩ to 200GΩ 250V 250kΩ to 50GΩ 250V 250kΩ to 500GΩ 500V 500kΩ to 100GΩ 500V 500kΩ to 1TΩ 1000V 1MΩ to 200GΩ 1000V 1ΜΩ to 2TΩ * Ranges of measurement for DMG500F and SMG500F limited to 2GΩ. D.3 Accuracy of the resistance measurement Display on 2000 points digital indicator with unity indications ( KΩ, MΩ,...) Accuracy: (in % of the reading. 1U= 1 display count) : Megohmmeter (MMG500) Dielectrimeter (DMG50, DMG500) Security Tester (SMG50, SMG50) 200 GΩ Version : ±(1.5% + 1U) 200 GΩ Version : ±(1.5% + 1U) 2 TΩ Version : ±(2% + 1U) 2 TΩ Version* (Uessai=50 or 100V) : ±(2% + 1U) 2 TΩ Version* (Uessai=250 or 500V) : ±(1% x Uessai/ U) *Only on DMG50 and DMG500 CAPACITANCE mode: from 1.00 MΩ to 200 GΩ (2 TΩ for option 20) With accuracy = (NORMAL mode accuracy) ±100 kω. Input impedance = 10 MΩ ±1% D.4 Measurement thresholds Adjustment of the digital value of 2 resistance thresholds from 100 kω up to 200 GΩ (2 TΩ for option 20). Comparison result indication on 2 LED and audible signal (possibility to inhibit). D.5 Timing Permanent measurement or during a time between 1 and 999 seconds (adjustable by 1 second step). MG SERIE Revision : R Page : 15

16 E. Ground continuity résistance function E.1 Measurement current From 5 to 30A AC by step of 0.5 AAC. Accuracy : ±(1% + 500mA). E.2 Measurement voltage < 6 VAC ou < 12 VAC. Sinus wave. Frequency identical to the mains. E.3 Accuracy of the ground continuity resistance measurement Display on 1500 points digital indicator with unity indications (mω). Accuracy (in % of the reading. 1U = 1 display count 1mΩ) : ±(2.5%+10U) C U R R E N T A SHUNT mohms 6V 12 V Page : 16 MG SERIE Revision : R

17 E.4 Measurement thresholds Adjustment of the digital value of 2 resistance thresholds from Ω up to Ω. Comparison result indication on 2 LED and audible signal (possibility to inhibit). E.5 Measurement thresholds If the current generated is high ( 25A), it can lead to a current overheating of the transformer and start up its protection (See Section 5.3). The hereunder board indicates the maximum use times of the ground continuity resistance measurement according to the use cycle (1 ½ cycle represents use of the current during one second for a total duration of the cycle of 2 seconds) Note : All these duration concern a 30A current. For lower currents, these duration are higher. Cycle 1/1 1/2 1/3 Use time 25 minutes 5 hours Infiniti MG SERIE Revision : R Page : 17

18 SECTION 1: INTRODUCTION - OPERATING INSTRUCTIONS Warning : This unit must be used by qualified people. Every precautions for the use of units connected to the main must be taken during its use. In particular, this unit must be connected to earth. The specifications of this manual, the correct operation of the unit as well as the operator s security are guaranteed only when the supplied accessories (TE54, TE56, CO175, CO183, CO184,...) are used. The measurement probes can include limitation or protective elements, therefore it is forbidden to modify without written agreement from SEFELEC company. In case of use under other conditions than the one specified in this manual, the security of the user will be in danger. Except for the earth controller CMG30, this unit can supply voltages and currents which could be lethal. Comply with the safety regulations related to the use of high voltage devices. ALWAYS MAKE SURE THE HIGH VOLTAGE INDICATOR IS NOT ON WHEN CONNECTING OR DISCONNECTING THE SPECIMENS. 1.1 MEANING OF THE DIFFERENT SYMBOLS ON THE INSTRUMENT Warning (See document attached) Warning, risk of electric chock. DC voltage. AC and DC voltages. AC voltage. Earth connection. Page : 18 MG SERIE Revision : R

19 1.2 PRODUCT OVERVIEW The MG series is a range of units performing very easily insulation resistance measurements, dielectric strength tests and ground continuity tests. The MMG500 is a megohmmeter, the RMG50, RMG15AC and RMG500 are dielectric strength testers under 50VA or 500VA and the DMG50 and DMG500 are dielectric strength testers under 50VA or 500VA including a megohmmeter, the CMG30 is a ground continuity controller and the SMG50 and SMG500 include all the above mentionned functions. The MG series makes possible to perform insulation resistance measurement, ground continuity resistance measurement and dielectric strength test on a various number of insulating material as resin, porcelain, oil, plastic as well as on final products such as capacitors, transformers, switches, cables, connectors or on devices connected to the main or supplied by batteries. The instruments allow ground continuity measurement fitting most part of the existing standards regarding measuring and medical fields, office devices, machinery,... The MG series is fitted with a high resolution liquid crystal display (LCD) and a keyboard allowing an easy use. A single output for the dielectric strength tests and the insulation resistance measurements makes easy the connection on the specimens.red and green LED lamps indicate visually and without any confusion the tests results. The dielectric strength tests are performed under adjustable voltages from 100 VAC up to 5000 VAC ( up to 6000 VDC in option ) with a short circuit current of 10 ma for 50VA models and 200 ma for 500VA models. The breakdown voltages and currents are memorized on the display after breakdown detection and voltage cut off. The insulation resistance measurements are performed under voltages of 50, 100, 250 or 500 VDC. The MG unit indicates in a direct reading with unity display, the resistance values from 50 kohms up to 200 Gohms (2 Tohms in option ). The continuity resistance measurement are performed under voltages of 5 to 30A AC with open circuit voltages of 6 or 12V AC. The MG unit indicates in a direct reading with unity display, the resistance values from 1mΩ to 1500mΩ. In option, the unit can be equipped with PLC input/output interface or with a serial RS232C interface enabling a full remote control by a PC computer. MG SERIE Revision : R Page : 19

20 1.3 FRONT PANEL DESCRIPTION The main functions of the MG units have been arranged in the following areas : Z1 : Power on and off Z2 : Trigger/stop of the measurement and HV red indicator Z3 : LED lamps for test results Z4 : 4 functions keys with function name on the LCD display Z5 : 5 keys to change and enter parameters Z6 : Input/output connectors Z7 : Graphic liquid crystal display (LCD) to display test parameters and test results Keyboard definition The unit includes 10 keys arranged in 3 function areas. Each action on a key is followed by an audible signal, informing that the unit has decoded the key position. The Z4 area includes 4 keys whose function changes according to the unit mode. In each mode, the right side of the LCD screen is used to indicate the meaning of the keys. If a key has no label, this one is not active. The Z5 area includes 4 keys to change parameter values and 1 key to enter the inputs ( at the center ). The UP/DOWN keys allow to increase or decrease the parameters values and the LEFT/RIGHT keys allow to move the cursor on all the digits of the numbers. The key at the center enters the new value of the parameter. The Z2 area includes a mechanical push button to trigger or to stop the MEASUREMENT mode. A red lamp, inside the push button, indicates when the MEASUREMENT mode is on, and therefore the possibility to have high voltage on the test probes. ALWAYS MAKE SURE THE HIGH VOLTAGE INDICATOR IS NOT ON WHEN CONNECTING OR DISCONNECTING THE SPECIMENS. Page : 20 MG SERIE Revision : R

21 1.3.2 Liquid Crystal display description MG series OPERATING MANUAL - SEFELEC The unit is fitted with a liquid crystal display (LCD) of 64 points per 240 points used in a graphic mode. According to the position of the user related to the unit screen and the light conditions, it is possible to adjust the screen contrast by turning the potentiometer located on the rear panel (Z9). The screen is LED backlighted, this allowing the use of the unit even the light conditions are poor. The screen is divided in 5 areas arranged as follows : Area Z1 : Help messages (in normal video mode) or error messages ( in reverse video mode) Area Z2 : Display of the measurement results with big characters ( 18 mm x 12 mm ). In the input parameters mode, display of test parameters for possible changes. Area Z3 : Recall of the measurement parameters Area Z4 : Symbol recalling the selected test ( MO, KV,...) Area Z5 : Indicates the name of the function keys on the right side of the LCD display. 1.4 REAR PANEL DESCRIPTION : The following areas are arranged on the rear panel : Z8 : Power input connector with voltage selector (115V/230V) Z9 : LCD contrast adjustment potentiometer Z10 : 25 points female sub-d connector for remote control of the unit ( see pin out in section 10.1) Z11 : 9 points female sub-d connector for RS232C interface (option) Z12 : Area for the output of the cables in the REAR PANEL OUTPUT option. Z13 : 25 points female sub D connector for printout type CENTRONICS MG SERIE Revision : R Page : 21

22 1.5 ACCESSORIES INCLUDED Operating manual Power cord (SE1) Measurement return cable 50 or 500VA (CO175) 25 points male sub-d connector (FD25) 1.6 ACCESSORIES / OPTIONS AVAILABLE : TE54 : HV measurement probe ( to hold in hand) for DMG50, RMG50, SMG50 TE56 : Measurement probe for MMG500 TE58 : Remote control probe for DMG50 / RMG50 / SMG50 TE65 : HV measurement probe ( to hold in hand) for RMG500, DMG500, SMG500 TE66 : Ground continuity set including 1 x CO183 and 1 x CO184 for CMG30, SMG50 and SMG500 TE67 : Pistol with remote control for RMG500, DMG500, SMG500 CO160 : Red/green lamp to indicate when HV is on CO174 : External box equipped with female socket for RMG50, DMG50, SMG50 CO177 : HV measurement probe for automatic test systems (for 50VA) CO179 : RS232C Cable CO180 : HV measurement probe for automatic test systems (FOR 500VA) CO183 : 2 wires with crocodile clip cable (CMG30, SMG50, SMG500) CO184 : 2 wires with retractable tip probe(cmg30, SMG50, SMG500) CO185 : External box equipped with female socket for RMG500, DMG500, SMG500 CO192 : External box equipped with 6 X female socket for RMG50, DMG50, SMG50 CO193 : External box equipped with 6 X female socket for RMG500, DMG500, SMG500 CO200 to CO209 : External box equipped with international female socket for 50 and 500VA. AO10 : Trigger/stop of the measurement by remote control using 2 hands AO11 : Trigger/stop of the measurement by foot switch AO14 : Sub-D connector extension (x4) KRMG : Rack mount kit OPT01 : RS232C interface. Talker/listener OPT02 : Remote control interface by PLC OPT03 : 0-10 volts analog input/output (DMG50, MMG50, RMG50) OPT04 : Test sequence (DMG50, DMG500) OPT05 : Measurement input/output on the rear panel OPT10 : 5000 VAC and 6000 VDC for 50VA OPT13 : 5000 VAC and 6000 VDC for 500VA OPT20 : Measurement up to 2 Tohms (DMG50, DMG500, MMG500, SMG50, SMG500) OPT21 : Display of the insulation resistance by analog meter (MMG500) OPT22 : Display of the insulation resistance in Mohm per km (DMG50, MMG500) OPT23 : Built-in calibration resistance (MMG500) OPT26 : Measurement with voltage up to 1000V. MG-90 : Maintenance and calibration manual MG-91 : Calibration kit for 50VA MG-91-30A : Calibration kit for ground continuity function. MG VA : Calibration kit for 500VA MG-91-FUITE : calibration kit for leakage current. MG-91-15AC : calibration kit for RMG15AC. MG-92 : Printer output (SMG50, SMG500) Page : 22 MG SERIE Revision : R

23 MG-93 : Remote box. MG-96 : (SmgPro) Software for MG series. MG SERIE Revision : R Page : 23

24 1.7 INSTALLATION Preliminary instructions WARNING : This unit must be used by qualified people. Every precautions for the use of units connected to the main must be taken during its use. In particular, this unit must be connected to earth. The specifications of this manual, the correct operation of the unit as well as the operator s security are guaranteed only when the supplied accessories (TE54, TE56, CO175, CO183, CO184,...) are used. The measurement probes can include limitation or protective elements, therefore it is forbidden to modify without written agreement from SEFELEC company. In case of use under other conditions than the one specified in this manual, the security of the user will be in danger. Except for the earth controller CMG30, this unit can supply voltages and currents which could be lethal. Comply with the safety regulations related to the use of high voltage devices. ALWAYS MAKE SURE THE HIGH VOLTAGE INDICATOR IS NOT ON WHEN CONNECTING OR DISCONNECTING THE SPECIMENS. The MG unit can be operated from either 115 or 230 volts ±10%, 47 to 63 Hz power line. Before connecting the 3 wires power cord (SE1) between the unit and AC power source, make sure the voltage indicated on the inlet module (Z8) is in accordance with the power source. For a change of voltage selection proceed as follows: Front panel ON/OFF switch (Z1) in the OFF position Remove the power cord SE1 from the inlet module WAIT AT LEAST 5 MINUTES BEFORE PROCEEDING With a small screwdriver, remove the selection sub-assembly from the inlet module Remove the fuses Take off the pale gray plastic piece Select voltage : 110 for a 115 volts main and 220 for a 230 volts main ( the 240 position is not used). Replace the gray plastic piece to display in the window of the selection sub-assembly the required voltage. Put the right value for the fuses according to the main voltage Replace the selection sub-assembly in the inlet module Connect the SE1 power cord The protective earth connection is COMPULSORY before using the 4U units. Use a screwed connection on the gudgeon located on the rear panel of units (minimum section 6 mm 2 ). Page : 24 MG SERIE Revision : R

25 1.7.2 Power on Make sure the voltage indicated on the inlet module (Z8) is in accordance with the power source (see 1.7.1). Connect the unit and AC source ( with earth connection) with the SE1 power cord then press the power button (Z1) in the ON position. After a few seconds the LCD screen must display the instrument s designation (CMG30, DMG50, RMG50, MMG500, SMG500,...) : According to the position of the user related to the unit screen and the light conditions, it is possible to adjust the screen contrast by turning the potentiometer located on the rear panel (Z9) Safety advice WARNING : NEVER TOUCH THE METALLIC TIP OF THE HV TEST PROBES (TE54, CO177,...) WHEN THE TEST PROBES ARE CONNECTED ON THE UNIT AND THE RED HV LAMP IS ON (Z2). The guard connection on the rear panel is at the measurement voltage (Umax = 500 VDC - 2MA). The measurement probes can include limitation or protective elements, therefore it is forbidden to modify without written agreement from SEFELEC company. The instrument must be placed in a position where the light ON/OFF can be easily seen. Check the good working of cables before use. Take every precaution necessary to avoid inadvertently touching the sample under test when there is a voltage passing through it (Red lamp in the push button Z2 on ) Do not open the cover Do not place the unit close to a wall so that the air passes through the ventilation ears. The unit is supplied with a «SAFETY LOOP» preventing generation of voltage when pins 1 and 14 of the 25 points connector (Z10) located on the rear panel are not connected. The test can be carried out only when connected. In case of remote control test pins 13 to 25 must be connected too. Note : It is therefore advisable to put into operation in this connection, dry contacts in condition of safety ( closed door, secured cover lowered,...) It is possible to connect a RED/GREEN lamp ( CO160) on the rear connector Z10 in order to indicate in a visible way if the HV is ON or OFF on the outputs of the unit. MG SERIE Revision : R Page : 25

26 1.7.4 Connection to a specimen MG series OPERATING MANUAL - SEFELEC Connection for insulation resistance measurement and dielectric strength tests (DMG50, DMG500, MMG500, RMG50, RMG500) Connect the measurement probe (TE54, TE56,...) on the B1 connector Secure the probe by tightening the knurled ring Proceed the same way for the return probe (CO175) on the B2 connector. Connect the specimen to be tested as follows: Page : 26 MG SERIE Revision : R

27 Connection for ground continuity measurement (CMG30) Connect the measurement probe (CO183 or CO184) on the B1, B2, B3 and B5 connectors Connector = CURRENT, GREY Connector = VOLTAGE). Connect the specimen to be tested as follows: (GREEN WARNING : Measurement on a specimen connected to earth. In the case of ground continuity measurement on a specimen connected to earth, the grounded side must be connected to the terminals U and I, B1 and B2. MG SERIE Revision : R Page : 27

28 Connection for insulation resistance measurement, dielectric strength tests and ground continuity measurement (SMG50, SMG500) Connect the measurement probe (CO174, CO185,...) on the B1, B2,W1, W2, W3 and W4 connectors. (GREEN Connector = CURRENT, GREY Connector = Voltage) Connect the specimen to be tested as follows: WARNING : Measurement on a specimen connected to earth. In the case of ground continuity measurement on a specimen connected to earth, the grounded side must be connected to the terminals I and U, W1 and W3. Page : 28 MG SERIE Revision : R

29 SECTION 2 : SETUP After switching on or returning to the menu of the figure 01 (1.7.2), press the [SETUP] function key of the Z4 area. Then the LCD screen displays as follows : With the UP and DOWN arrow keys, move the reverse video line and go into modification mode by pressing on the RIGHT arrow key or on the ENTER key (key at the center of the selection keyboard). Scroll the possible selections with the UP and DOWN arrow keys, then enter the choice with the validation key. 2.1 LANGUAGE All the displayed messages on the LCD screen can be in several languages (French, English,...). Display the required language with the UP/DOWN arrows, then enter. The unit goes back to its initialization menu. MG SERIE Revision : R Page : 29

30 2.2 BEEP ON FAIL When a test result is failed, a permanent audible signal is emitted by the unit until the removal by pressing on the [MEASUREMENT-DISCHARGE] key. This function is active when YES is selected and not active when NO is selected. The choice is memorized after switching off the unit. 2.3 FILTER Allows to select a special functioning mode of the unit when making tests on capacitors or on specimens having a certain among of capacitance. Display the required mode with the UP/DOWN arrows : NORMAL, CAPACITOR, R.H. TIME..., then enter. The CAPACITOR mode allows to perform stable insulation resistance measurements on capacitive specimens (reel of cable, capacitors,...). During dielectric strength test with a DC voltage, the CAPACITOR mode inhibits the voltage adjustment to avoid overvoltage, switches the 1.5 Mohm discharge resistor at the beginning of the fall time and controls the residual discharge voltage until this one is lower than 100 volts. Operating this mode is recalled by the drawing of a capacitor over the unity display in the measurement menus. The R.H. TIME (Real Hold Time) mode improves the operation of voltage ramp during the dielectric strength tests. It is suitable for tests using a DC high voltage on capacitive specimens. At the end of the rise time, the unit controls the output voltage and triggers the hold time only when the voltage has reached the pre-set value. During the fall time, the unit stops the measurement cycle when the output voltage is lower than 30 Volts. Page : 30 MG SERIE Revision : R

31 2.4 PARAM ACCESS The access to the parameter modification can be secured by a password. This limitation is made of three different levels : - FREE : Let you modify all the parameters - READING : Let you only recall the memories but you can not modify them - FORBIDDEN : You can not modify anything ; security is total You can set up the password access in the configuration menu [CONFIG] and in the line [ACCES PARAM]. Follow the following instructions to modify the password and the access level. Switch no the unit, and from the main menu, select the [CONFIG] menu. The display is then indicating the following information. Then, with the up and down arrows go onto the «ACCES PARAM.» field and select it with enter. The display is then indicating the following information. Select the figure to modify with the right and left arrows. Increase or decrease the figures ( ) to select your code. Repeat this action for each figure. Once each figure is entered, validate with Enter key. WARNING THE FIRST TIME YOU SWITCH ON THE UNIT, THE PASSWORD IS VOID. A PASSWORD IS A REAL NUMBER. MG SERIE Revision : R Page : 31

32 If the code is not a real number, you will get an error and the following message "INPUT ERROR". In this example, you would get an error, because of the dot. Then to correct you have to : - Type a new password - Abandon the code by clicking on end key If the password is not good, the following message will be displayed : "PASSWORD UNKNOW" Then to correct you have to : - Type a new password - Abandon the code by clicking on end key In the case of a correct password, the following message will be displayed : Then enter the access level you wish, as described in the chapter II from the MG manual. PLEASE CONTACT OUR AFTER SALES SERVICE IN CASE YOU LOOSE THE PASSWORD Tel : (33) Page : 32 MG SERIE Revision : R

33 2.5 DISPLAY Press on the UP or DOWN arrows to select the next setup menu. Allows to inhibit the display of the digital measurement results, only the PASS-FAIL result is valid. This mode is specially interesting when controlling the unit by an external system (RS232C or PLC) because it allows to reduce the test times. When operating this mode, a fixed drawing is displayed in the measurement result window: 2.6 REMOTE TRIGGER MODE Allows to use a remote control system to trigger the measurement mode. Operating this mode is recalled by the following drawing in the measurement windows : WARNING : Please, follow carefully the procedure below to use the REMOTE TRIGGER mode : Remove the FD25 connector from the rear panel Connect on the 25 points rear panel connector a provided accessory for remote control (i.e. AO10) or connect the 2 wires provided for your own remote control system between the 13 and 25 points. Select YES on the input line REMOTE TRIG. Escape the setup menu with the [ESC] key Select measurement type with [MEGOHM] or [HIPOT] Verify the display of the remote control drawing The front panel [MEASUREMENT-DISCHARGE] push-button is inoperative Close the contact between 13 and 25 Keep the contact closed during the test time Release the contact at the end of the test time When operating this mode, the opening of the safety loop (points 1 and 14) inhibits to go in the measurement mode, but the unit will not indicate the INTERLOCK DISABLE error message. IMPORTANT : If you have selected the REMOTE TRIG. mode and if the connections 1-14 and are already done when selecting a measurement function (Megohm or Hipot), the following message is displayed : MG SERIE Revision : R Page : 33

34 INPUT ERROR and the REMOTE TRIG. mode is automatically removed. Please refer to the above procedure to select correctly the REMOTE TRIG. mode. 2.7 INTERFACE SELECTION Allows to select the remote control mode. Display on the LCD screen the various interface types with the UP and DOWN arrows : OFF, PLC, RS232,... When using a remote control mode it is necessary to select it in this line. When pressing on ENTER key, if the option is not installed in the unit, the following message is displayed : MISSING OPTION For more details regarding the remote control modes, please refer to the 8.3 section of this manual. WARNING : RS232 speed : 9600 bauds for MG series Bauds for "MG+" series bauds for MG series with MG-70 option. 2.8 SEQUENCE (OPTION 04) If the option is valid and only for a DMG50, this feature allows to perform automatic cycling of the megohm and dielectric strength functions. Display on the LCD screen the various sequence modes with the UP and DOWN arrows : OFF,M+H,H+M,M+H+M,... The M letter shows an insulation resistance measurement and the H letter shows a dielectric strength test. Operating the sequence mode is recalled in the measurement windows by the following drawing : In a M+H sequence the unit performs first an insulation test and then a dielectric test. It is necessary to set the parameters specially the test times before running a sequence. In case of a fault during the test, the sequence is stopped. When operating in sequence mode, at power on, the first sequence function type is automatically selected. When attempting to trigger the measurement mode in a function type which is not the first one of the sequence, the following message is displayed : SEQUENCE ERROR Escape of the function and select either another function or another sequence mode. Example of a sequence setup : An operator wants to perform an insulation control on a specimen, then a dielectric strength test and finally a new insulation control to verify that the dielectric strength test hasn't degraded the specimen. SETUP menu Page : 34 MG SERIE Revision : R

35 <page2>, line SEQUENCE Select the sequence mode : M+H+M Escape with [ESC] key The unit switches automatically in the insulation test Check that parameters are correct (Necessary to select times) Press on the [MEASUREMENT-DISCHARGE] push button After the test time the unit switches automatically to the next test if the test result was PASS. When an insulation test has to be perform after a dielectric strength test, the unit controls the discharge voltage of the specimen and follows its sequence only if the voltage is lower than 100 volts with the display of : DISCHARGE CONTROL In case of a fault during the insulation test, the number 1 or 2 of the test is displayed. At the end of a sequence, after release of the result by pressing the [MEASUREMENT-DISCHARGE] push button the unit switches in the first function of the sequence. To escape of the SETUP menu, press on the function key [ESC]. MG SERIE Revision : R Page : 35

36 SECTION 3 : INSULATION RESISTANCE MEASUREMENT (DMG50, DMG500, MMG500, SMG50, SMG500) Access to the insulation resistance measurement function [MEGOHM] for a SMG50 or SMG500, dielectric strength test function [HIPOT] and ground continuity tester [GROUND]. From the initialization menu press on the [MEGOHM] key, display indicates : 3.1 PARAMETERS From the initialization menu press on the [MEGOHM] function key to get the menu shown in figure 10. The main test parameters are recalled on the bottom line of the LCD screen. If the message [ACCESS DENIED] is displayed, please refer to section 2.4. To modify these parameters press on the [PARAM] function key of this menu. Display shows : Note : For DMG50, DMG500, SMG50, SMG500, parameters setup is made on a separate display in which you must select the required function. Page : 36 MG SERIE Revision : R

37 3.1.1 Measurement, voltage selection MG series OPERATING MANUAL - SEFELEC The unit offers the possibility to select preset measurement voltages ( VDC). The voltage selection is done according to the standard or the specifications used for the test. Without any specifications, select 100 VDC. Move the reverse video line in front of the VOLTAGE line with the UP/DOWN arrows keys. Press on the RIGHT arrow or ENTER key Display shows : With the UP/DOWN arrow keys scroll all the available voltages (50,100,250,500) Enter the selected value with the ENTER key Threshold selection The unit includes 2 comparison thresholds making possible to check if the specimen under test is good or bad. The HI-LIMIT defines the maximum insulation resistance value for the specimen in order to detect a possible bad connection of the test probes on the specimen. The LO-LIMIT defines the minimum insulation resistance value that must be reached by the specimen under test. A specimen is good (PASS) if its insulation resistance value is < HI LIMIT and > LO LIMIT, otherwise the specimen is declared bad (FAIL). In the basic version of the unit the thresholds are adjustable from 0 kω up to GΩ. A HI LIMIT set to GΩ cancels the comparison to the high threshold, in that case the value of the high threshold is not displayed on the parameters line of the figure 10. When a value above 200GΩ (or 2TΩ according to the option) is entered, the message LIMIT ERROR is displayed. Move the reverse video line in front of the HI LIMIT line. Press on the RIGHT arrow or ENTER key Display shows : MG SERIE Revision : R Page : 37

38 Select the number to be modified by moving the cursor with the RIGHT/LEFT arrow keys Increase or decrease the value with respectively the UP and the DOWN arrow keys ( etc) Repeat operation for all the numbers if necessary With the RIGHT arrow key, move the cursor up to the unity zone Scroll the unity: KΩ (1 000 ohms), MΩ ( ohms), GΩ ( ohms) and TΩ ( ohms). Enter the threshold with the ENTER key WARNING : the HI LIMIT must be always higher than the LO LIMIT, otherwise an error message is displayed : HI LIMIT < LO LIMIT To set the LO LIMIT value proceed the same way as for the HI LIMIT. WARNING: the LO LIMIT must be always lower than the HI LIMIT, otherwise an error message is displayed: LO LIMIT > HI LIMIT Timer The unit is equipped with a timer to define the insulation resistance measurement time (from 1 up to 999 seconds). This feature is particularly interesting when measuring on capacitive specimens of which the insulation resistance increases as a function of the measurement time. At the end of the test time, the unit automatically stops the test and memorizes the last measured value on the LCD screen. In case where a test time is set to 0, the unit remains in the measurement mode permanently until the [MEASUREMENT/DISCHARGE] key is pressed. Move the reverse video line in front of the TIMER line Press on the RIGHT arrow or ENTER key Display shows : Select the number to be modified by moving the cursor with the RIGHT/LEFT arrow keys Increase or decrease the value with respectively the UP and the DOWN arrow keys ( etc). Repeat operation for all the numbers if necessary Enter the time with the ENTER key Page : 38 MG SERIE Revision : R

39 3.1.4 Parameters memorization MG series OPERATING MANUAL - SEFELEC This feature allows to store in 10 memories (numbers 0 to 9) measurement parameters (voltage, threshold, time,...). From the measurement menu, to modify the memory number : Press on the [MEM:x] function key Display shows : With the UP/DOWN arrow keys, increase or decrease the memory number (from 0 to 9). The parameters recall line (area 3 of the LCD screen) indicates the content of each memory. Enter the selected memory either by pressing again on the [MEM:x] function key or by pressing the ENTER key. From the parameter menu, to modify a memory number : Press on the [MEM:x] function key With the UP/DOWN arrow keys, increase or decrease the memory number (from 0 to 9). The input parameters lines indicate the content of each memory. Enter the selected memory either by pressing again on the [MEM:x] function key or by pressing the ENTER key. WARNING : ANY PARAMETER MODIFICATION IS AUTOMATICALLY STORED IN THE MEMORY. See section 2.4 PARAM ACCESS to control the access to the parameter modifications. MG SERIE Revision : R Page : 39

40 3.2 INSULATION RESISTANCE MEASUREMENT Select the voltage, threshold and time parameters as described in section to NOTE : these parameters are stored even after the unit has been switched off. WARNING : ALWAYS MAKE SURE THE HIGH VOLTAGE INDICATOR IS NOT ON WHEN CONNECTING OR DISCONNECTING THE SPECIMENS. Connect the specimen to be measured as described in section Press on the [MEASUREMENT/DISCHARGE] button Indicator inside the button is illuminated and the display shows : In case where a test time is set to 0, the unit remains in the measurement mode permanently until the [MEASUREMENT/DISCHARGE] key is pressed. If a test time value has been selected, every second the value of the test time counts down one unit. When the displayed time reaches 0 the output voltage is cut off automatically and the last measured value is stored on the LCD screen. According to the insulation resistance value in comparison with the HIGH and LOW limits, the red LED (FAIL) or the green LED (PASS) is illuminated. Press on the [MEASUREMENT/DISCHARGE] key to cancel the memorization on the LCD screen. (The HV indicator goes out.) Display shows figure 11 Press on the [ESC] key to escape the function Page : 40 MG SERIE Revision : R

41 3.3 ERROR MESSAGES: The following error messages may appear during the measurement : INTERLOCK DISABLE : the safety loop is not closed. There is no connection between the points 1-14 and of the 25 points Z10 connector of the rear panel (See Section 1.7.3). Make the connection, press again on the [MEASUREMENT/DISCHARGE] key to trigger the measurement mode. OVER-RANGE : the insulation resistance of the specimen under test exceeds the measurement specifications of the unit (refer to specification section of this manual). UNDER-RANGE : the insulation resistance of the specimen under test is lower than the measurement specifications of the unit (refer to specification section of this manual). CHARGING : the measurement voltage has not reached its final value. On capacitive specimen, wait for the end of the charge, on resistive specimen check with the current specifications of the power supply. BOARD NOT READY : the microprocessor board cannot communicate with the insulation resistance board. You cannot perform a measurement, get in contact with our Service department. 3.4 UNIT EQUIPPED WITH ANALOG METER (21 option) In this case the insulation resistance is displayed as a scientific number format For example : 18 x 10 5 Ohms (1.8 Mohms) with 18 displayed by the pointer of the analog meter on a linear scale from 0 to 20 and the power of ten on the LCD display. Accuracy of the measurement : Megohmmeter - version 200 Gohms : ±(1.5%+CG*) - version 2 Tohms : ±( 2%+CG*) *CG = Class of the analog meter = 1.5 % of the full scale MG SERIE Revision : R Page : 41

42 SECTION 4 : DIELECTRIC STRENGTH TEST (DMG50, DMG500, RMG50, RMG500, SMG50, SMG500) From the initialization menu press on the [HIPOT] function key to obtain the menu shown in figure 12. The main test parameters are recalled on the bottom line of the LCD screen. 4.1 SETTING To modify these parameters, press on the [PARAM] function key of this menu. If the message [ACCESS DENIED] is displayed, please refer to section 2.4. The display shows then : In the basic version of the unit, the measurement parameters are stored in a battery backup memory and are automatically recalled at the power on. With the option OPT00 it is possible to store 10 various setups in the memory (see 2.3.5). Note : For DMG50, DMG500, SMG50, SMG500, parameters setup is made on a separate display in which you must select the required function. Page : 42 MG SERIE Revision : R

43 4.1.1 Test voltage selection MG series OPERATING MANUAL - SEFELEC The unit offers the possibility to select dielectric strength test voltages between 0.10 and 5.00 KVAC or 0.10 and 6.00 KVDC (option). The test voltage selection is done according to the standards. Move the reverse video line in front of the VOLTAGE line Press on the RIGHT arrow or ENTER key Display shows : Select the number to be modified by moving the cursor with the RIGHT/LEFT arrow keys Increase or decrease the value with respectively the UP and the DOWN arrow keys ( etc) Repeat operation for all the numbers if necessary With the RIGHT arrow key, move the cursor up to the unity zone Scroll AC, DC (if option OPT10 is present) with the UP/DOWN arrow keys Enter the voltage with the ENTER key. If the value is higher than the maximum limits (5.00 in AC and 6.00 in DC), following error message appears : LIMIT ERROR Enter a correct value or [ESC] MG SERIE Revision : R Page : 43

44 4.1.2 Current threshold selection The unit includes 2 comparison thresholds making possible to check if the specimen under test is good or bad. The IMIN LIMIT defines the minimum current value which must flow through the specimen in order to detect a possible bad connection of the test probes on the specimen. The IMAX LIMIT defines the maximum current value which is allowed to flow through the specimen under test. According to the selected breakdown detection mode (see 4.1.3), a specimen is good (PASS) if the supplied current is < IMAX LIMIT and > IMIN LIMIT, otherwise the specimen is declared bad (FAIL). The thresholds are adjustable from 0.00 ma up to 9.99 ma for 50VA and from 0.00 ma up to 99.9 ma for 500VA. A IMIN LIMIT set to 0.00 ma cancels the comparison to the minimum current threshold. Move the reverse video line in front of the IMAX LIMIT line Press on the RIGHT arrow or ENTER key Display shows : Select the number to be modified by moving the cursor with the RIGHT/LEFT arrow keys Increase or decrease the value with respectively the UP and the DOWN arrow keys ( etc) Repeat operation for all the numbers if necessary Enter the threshold with the ENTER key WARNING : the IMAX LIMIT must be always higher than the IMIN LIMIT, otherwise an error message is displayed : IMAX LIMIT < IMIN LIMIT To set the IMIN LIMIT value proceed the same way as for the IMAX LIMIT. WARNING: the IMIN LIMIT must be always lower than the IMAX LIMIT, otherwise an error message is displayed : IMIN LIMIT > IMAX LIMIT Page : 44 MG SERIE Revision : R

45 4.1.3 Breakdown detection mode selection The unit offers the possibility to select several breakdown detection modes : IMAX mode : makes possible to monitor the permanent leakage current flowing through the specimen under test and compares this current in regard to an adjustable limit (see IMAX LIMIT). This mode does not perform fast arc detection and its relatively slow reaction time can be destructive for the specimen under test. In addition the value of the IMAX threshold has to be adjusted regarding the normal current of each specimen (AC leakage current under test voltage). FIMAX mode (Fast IMAX) : Same as IMAX but detection is made more rapidly. -I mode : arc detection mode which monitors arcs having a minimum width of 10uS and an amplitude of 1 ma. This mode does not take care for the normal leakage current and therefore does not need any adjustment. Its fast reaction time makes possible the limitation of the damages caused to the specimen under test. On the other hand, the -I mode does not find a fault in case of specimen already in short circuit at the beginning of the test. FIMAX+ I mode : combination of the two last mode. This mode allows you to realise hipot tests, in an easy and accurate way. IMAX detection is made more rapidly. IMAX+ -I mode : combination of the two previous detection modes. This mode makes possible to perform reliable dielectric strength tests very easily. OFF mode : disable the detectors. This mode makes possible to locate a dielectric strength fault by burning. WARNING : this mode can be used only for a limited time (5 seconds). In case of over heating a built-in thermal switch cuts off the high voltage generator. Wait for 3-5 minutes before going on. In this mode there is no automatic adjustment of the high voltage regarding the impedance of the load. NOTE : the minimum leakage current control mode (IMIN LIMIT) is independent of the above breakdown detection modes. Move the reverse video line in front of the DETECTION line with the UP/DOWN arrows keys. Press on the RIGHT arrow or ENTER key Display shows: With the UP/DOWN arrow keys scroll the various detection modes available : -I, IMAX+ -I, IMAX, OFF, FIMAX+ I, FIMAX... Enter the selected mode with the ENTER key. MG SERIE Revision : R Page : 45

46 4.1.4 Timer The sudden application of the test voltage on a specimen could stressed it more than required. Therefore the unit is fitted with a voltage rise time system. The same phenomena existing during the cut off of the high voltage, the unit can perform the following test cycle : There are four possible ways to use the unit, let = AUTO, MANUAL, FAIL and U:2. The AUTO mode is described here above. The U:2 runs like the auto mode, but the test starts with the programmed voltage divide by 2. The FAIL mode (only with 500VA model), let the test be permanent to a break détection, without rising time. The MANUAL mode, the user can increase or decrease the output test voltage as he wants by pressing on respectively the UP or the DOWN arrow keys. It is necessary to press on the keys for each step. From the parameters menu (FIG. 20), press on the DOWN or UP arrow keys as many times necessary to display the following menu : Page : 46 MG SERIE Revision : R

47 Move the reverse video line in front of the TIME line with the UP/DOWN arrows keys. Press on the RIGHT arrow or ENTER key Display shows : With the UP/DOWN arrow keys scroll the various available timer modes : AUTO, MANUAL Enter the selected mode with the ENTER key. In the AUTO mode, it is possible to set the RISE, HOLD and FALL times from 0 up to 999 seconds. Move the reverse video line in front of the HOLD line Press on the RIGHT arrow or ENTER key Display shows : Select the number to be modified by moving the cursor with the RIGHT/LEFT arrow keys Increase or decrease the value with respectively the UP and the DOWN arrow keys ( etc) Repeat operation for all the numbers if necessary Enter the time with the ENTER key Proceed the same way for the RISE and FALL times. MG SERIE Revision : R Page : 47

48 4.1.5 Parameters memorization MG series OPERATING MANUAL - SEFELEC This feature allows to store in 10 memories (numbers 0 to 9) measurement parameters (voltage, threshold, time,...). From the measurement menu, to modify the memory number: Press on the [MEM:x] function key Display shows : With the UP/DOWN arrow keys, increase or decrease the memory number (from 0 to 9). The parameters recall line (area 3 of the LCD screen) indicates the content of each memory. Enter the selected memory either by pressing again on the [MEM:x] function key or by pressing the ENTER key. From the parameter menu, to modify memory number : Press on the [MEM:x] function key With the UP/DOWN arrow keys, increase or decrease the memory number (from 0 to 9). The input parameters lines indicate the content of each memory. Enter the selected memory either by pressing again on the [MEM:x] function key or by pressing the ENTER key. WARNING : ANY PARAMETER MODIFICATION IS AUTOMATICALLY STORED IN THE MEMORY. See section 2.4 «PARAM ACCESS» to control the access to the parameter modifications. Page : 48 MG SERIE Revision : R

49 4.2 DIELECTRIC STRENGTH TEST Select the voltage, current thresholds, breakdown detection mode and time parameters as described in section to NOTE : these parameters are stored even after the unit has been switched off. WARNING : ALWAYS MAKE SURE THE HIGH VOLTAGE INDICATOR IS NOT ON WHEN CONNECTING OR DISCONNECTING THE SPECIMENS. Connect the specimen to be measured as described in section Press on the [MEASUREMENT/DISCHARGE] button Indicator inside the button is illuminated and the display shows : In case where the MANUAL test time mode has been selected, the unit remains in the measurement mode permanently until the [MEASUREMENT/DISCHARGE] key is pressed. Use the UP arrow key to increase the output voltage and the DOWN arrow key to decrease the output voltage. The maximum voltage which can be reached is those which is displayed on the parameter line : VOLTAGE:x.xxKVAC However according to the impedance of the specimen under test, the maximum reached voltage could be lower. If the AUTO test time has been selected, every second the value of the RISE time counts down one unit, then the value of the HOLD time and at least the value of the FALL time. When the FALL time reaches 0 the output voltage is cut off automatically. According to the leakage current value during the test time in comparison with the IMAX and IMIN limits, the red LED (FAIL) or the green LED (PASS) is illuminated. MG SERIE Revision : R Page : 49

50 In both modes (AUTO and MANUAL) if the leakage current flowing through the specimen under test overruns the set breakdown detection values (in -I or IMAX modes) the unit declares breakdown with cut off of the high voltage (at the next zero crossing). The display shows : The memorized voltage is the breakdown voltage and the leakage current is also memorized and is corresponding to a IMAX value. Press on the [MEASUREMENT/DISCHARGE] key to cancel the memorization of the breakdown data on the LCD screen (the HV indicator goes out). Display shows figure 11 Press on the [ESC] key to escape the function 4.3 ERROR MESSAGES The following error messages may appear during the measurement : INTERLOCK DISABLE : the safety loop is not closed. There is no connection between the points 1-14 and of the 25 points Z10 connector of the rear panel (See Section 1.7.3). Make the connection, press again on the [MEASUREMENT/DISCHARGE] key to trigger the measurement mode. In case of over heating a builtin thermal switch cuts off the high voltage generator. Wait for 3-5 minutes before going on. In this mode there is no automatic adjustment of the high voltage regarding the impedance of the load. VOLTAGE ERROR : the test voltage has not reached its final value (impedance of the specimen under test too low), in DC voltage wait for the end of the charge, in AC voltage check with the current specifications of the power supply. In case of over heating a built-in thermal switch cuts off the high voltage generator. Wait for 3-5 minutes before going on. I<IMIN : during the HOLD time, the leakage current flowing through the specimen under test has not reached the minimum value set by the IMIN parameter (possible bad connection of the test probe). This fault gives a bad (FAIL) test result even there is no breakdown. BOARD NOT READY : the microprocessor board cannot communicate with the dielectric strength test board. You cannot perform a measurement, get in contact with our Service department. SYNCHRO ERROR : the dielectric strength test board is not able to generate the sinusoidal signal which drives the HV transformer. You cannot perform a test, get in contact with our Service department. Page : 50 MG SERIE Revision : R

51 SECTION 5 : GROUND CONTINUITY RESISTANCE MEASUREMENT (CMG30, SMG50, SMG500) From the initialization menu press on the [GROUND] function key to obtain the figure 12 for a CMG30 or on the [NEXT] key and then [GROUND] for a SMG50 or a SMG500. The main test parameters are recalled on the bottom line of the LCD screen. 5.1 Parameters setting To set parameters, press on the [PARAM] function key of this menu. If the message [ACCESS DENIED] is displayed, please refer to section 2.4. The display shows then : Note : For DMG50, DMG500, SMG50, SMG500, parameters setup is made on a separate display in which you must select the required function. MG SERIE Revision : R Page : 51

52 5.1.1 Test current selection MG series OPERATING MANUAL - SEFELEC The unit offers the possibility to select ground continuity test current (step of 0.5 A). The test current selection is done according to the standards. Move the reverse video line in front of the CURRENT line. Press on the RIGHT arrow or ENTER key. Display shows : Select the number to be modified by moving the cursor with the RIGHT/LEFT arrow keys. Increase or decrease the value with respectively the UP and the DOWN arrow keys. Repeat operation for all the numbers if necessary. Enter the current with the ENTER key. If the value is higher than the maximum limits (5.0A-30.0A), following error message appears : LIMIT ERROR. Enter a correct value or [ESC] Test voltage selection The unit offers the possibility to select ground continuity test voltage (6 and 12 Volts). The test voltage selection is done according to the standards. Move the reverse video line in front of the VOLTAGE line. Press on the RIGHT arrow or ENTER key. Display shows : With the UP/DOWN arrow keys scroll all the available voltages (6, 12). Enter the selected value with the ENTER key. Page : 52 MG SERIE Revision : R

53 5.1.3 Thresholds selection MG series OPERATING MANUAL - SEFELEC The unit includes two comparison thresholds making possible to check if the specimen under test is good or bad. These thresholds can be a resistance or a voltage (according to the requirements of standard EN ). The H LIMIT defines the maximum resistance value which is allowed for the specimen under test. The L LIMIT defines the minimum resistance value in Ohms or Volts that can be reached by the specimen under test. A specimen is good (PASS) if the resistance is under H LIMIT and over L LIMIT, otherwise the specimen is declared bad (FAIL). Those thresholds are adjustable from 0 to 1500mΩ or 0.01V to 12.0V. Move the reverse video line in front of the H LIMIT line. Press on the RIGHT arrow or ENTER key. Display shows : Select the number to be modified by moving the cursor with the RIGHT/LEFT arrow keys. Increase or decrease the value with respectively the UP and the DOWN arrow keys ( ). Enter the threshold with the ENTER key. Warning, the H LIMIT must always be higher than the L LIMIT, otherwise an error message is displayed : H LIMIT < L LIMIT. To set the L LIMIT value, proceed the same way as for the H LIMIT. Warning, the L LIMIT must always be lower than the H LIMIT, otherwise an error message is displayed IMIN LIMIT > IMAX LIMIT. If you whish to modify the unit (for instance switch from Ohms to Volts), use the right arrow as much times as you need to place the specimen on reverse video. Then change it with UP and DOWN arrows. Warning : when changing the unit, you must enter again the two thresholds. MG SERIE Revision : R Page : 53

54 5.1.4 Timer The sudden application of the test current on a specimen could stressed it more than required. Therefore the unit is fitted with a current rise time system. The same phenomena existing during the cut off of the current the unit can perform the following test cycle: The three possible ways to use the unit are the AUTO mode (described hereabove), the DEFAULT mode and the MANUAL mode. In that last mode, the current is permanently applied to the tested specimen. The DEFAULT mode or stop on default is similar to the AUTO mode except that the test is stopped when the first FAIL measurement regarding threshold appears. From the parameters menu (figure 30), press on the DOWN or UP arrow keys as many times necessary to display the following menu : Page : 54 MG SERIE Revision : R

55 Move the reverse video line in front of the TIME line with the UP/DOWN arrows keys. Press on the RIGHT arrow or ENTER key. With the UP/DOWN arrow keys scroll the various available timer modes (AUTO, MANUAL). Enter the selected mode with the ENTER key. In the AUTO mode, it is possible to set the RISE, HOLD and FALL times from 0 up to 999 seconds. Move the reverse video line in front of the HOLD line. Press on the RIGHT arrow or ENTER key. Select the number to be modified by moving the cursor with the RIGHT/LEFT arrow keys. Increase or decrease the value with respectively the UP and the DOWN arrow. Repeat operation for all the numbers if necessary. Enter the time with the ENTER key Proceed the same way for the RISE and FALL times. MG SERIE Revision : R Page : 55

56 5.1.5 Parameters memorizations MG series OPERATING MANUAL - SEFELEC The memorization feature allows to store in 10 memories (numbers 0 to 9) measurement parameters (voltage, threshold, time,...). From the measurement menu, to modify the memory number : Press on the [MEM:x] function key. With the UP/DOWN arrow keys, increase or decrease the memory number (from 0 to 9). The parameters recall line (area 3 of the LCD screen) indicates the content of each memory. Enter the selected memory either by pressing again on the [MEM:x] function key or by pressing the ENTER key. From the parameter menu, to modify memory number : Press on the [MEM:x] function key. With the UP/DOWN arrow keys, increase or decrease the memory number (from 0 to 9). The input parameters lines indicate the content of each memory. Enter the selected memory either by pressing again on the [MEM:x] function key or by pressing the ENTER key. Warning : Any parameter modification is automatically stored in the memory. See section 2.4 to control the access to the parameter modifications. Page : 56 MG SERIE Revision : R

57 5.2 Ground continuity resistance measurement Select the voltage, current, thresholds and time parameters as described in paragraphs to Note : those parameters are stored even after the unit has been switched off. They can be stored in various memories numbers. Connect the specimen to be measured as described in paragraph Press on the [MEASUREMENT/DISCHARGE] button. The push button must switch on red light and the LCD display shows : in this example the unit of the thresholds is the Ohm, the result is given in Ohm and the voltage is reminded in the upper right corner. If the unit of the thresholds was the volt the given result is in Volt, and reminder is in Ohm. The indicator inside the button is illuminated and the display shows : If the AUTO test time has been selected, every second the value of the RISE time counts down one unit, then the value of the HOLD time and at least the value of the FALL time. When the FALL time reaches 0 the output current is cut off automatically. According to the resistance value during the test time in comparison with the H LIMIT and L LIMIT limits, the red LED (FAIL) or the green LED (PASS) is illuminated. At the end of the test, the unit cut off the current and END OF TEST appears at top of the screen. Press on the [MEASUREMENT-DISCHARGE]. Press on the [ESC] key to escape the function. MG SERIE Revision : R Page : 57

58 5.3 Error messages The following error messages may appear during the measurement : INTERLOCK DISABLE : The safety loop is not closed. There is no connection between the points 1-14 and of the 25 points Z10 connector of the rear panel (See paragraph 1.7.3). Make the connection, press again on the [MEASUREMENT/DISCHARGE] key to trigger the measurement mode. OVER-RANGE : The continuity resistance of the specimen under test exceeds the measurement specification of the unit. OVERHEATING : The unit is equiped with a heating security checked to 80. Wait half an hour before go on measure. CONTINUITY ERROR : Cables are not correctly connected or the continuity resistance of the specimen under test is too importante (some ohms) for the unit. BOARD NOT READY : The microprocessor board cannot communicate with the ground continuity test board. You cannot perform a measurement, get in contact with our Service department. WARNING : Measurement on a specimen connected to earth. In the case of ground continuity measurement on a specimen connected to earth, the grounded side must be connected to the terminals U and I, W1 and W3. Page : 58 MG SERIE Revision : R

59 SECTION 6 : LEAKAGE CURRENT MEASUREMENT SEE FMG 500 FMG501 MANUAL MG SERIE Revision : R Page : 59

60 SECTION 7 : TEST SEQUENCING (SMG50, SMG500) SMG50 and SMG500 are able to perform a measurement sequence including up to eight different test steps. Moreover, they can be used as independent measurement unit. Different measurement parameters and conditions can be defined for each test in the sequence. 7.1 Setting of measurement functions To access to measurement functions, press the [FUNCT] key from the initialization menu, then press on the selected function key. Function not defined on the first screen, can be accessible with the [NEXT] key. See section 3, 4 and 5 for details on the setting and the operation of the different functions. 7.2 Sequence setup Press on the [SEQ.] key from the initialization menu to obtain the next screen. Eight programme lines of the selected memory are recalled on the bottom line of the LCD screen. The contents of each line is described by 2 digits : for a function, the first digit sets for the type of test (M for Megohmmeter, H for Hipot, G for Ground continuity), the second digit refering to the number of the corresponding parameter set. The 2 digits ".." feature an empty line, the word "OK" a break between 2 functions and the operator "x" a multiple ground continuity test. Page : 60 MG SERIE Revision : R

61 7.2.1 Parameters memorization MG series OPERATING MANUAL - SEFELEC The unit contains 10 measurement sequences (numbered from 0 to 9) including up to eight different test steps. Each step corresponds to a test function associated to a parameter set, a pause between two tests or to the operating of several test points (at least 2) of the function defined on the previous line (multiple ground continuity). On the above drawing, a and b represent the numbers of parameters sets associated to functions x and y chosen between those available (HIPOT, MEGOHM, GROUND, LEAK, POINTS, and PAUSE) or EMPTY in case the test line is not used. To modify the memory number from the measurement screen (figure 41), press on the [MEM:x] key. Display shows : With the up and down arrow keys, increase or decrease the memory number (from 0 to 9). The parameters recall line resumes the content of each memory. Enter the selected memory either by pressing again on the [MEM:x] function key or by pressing the enter key. WARNING : ANY PARAMETER MODIFICATION IS AUTOMATICALLY STORED IN THE MEMORY. See section 2.4 PARAM ACCESS to control the access to the parameter modifications. MG SERIE Revision : R Page : 61

62 7.2.2 Function selection MG series OPERATING MANUAL - SEFELEC From the initialization menu, press on the [PARAM] key. Display shows : Select the test to be modified and enter it with the enter key. Display shows : Scroll the possible selection (HIPOT, MEGOHM, RIGID) with up and down arrow keys. Page : 62 MG SERIE Revision : R

63 7.2.3 Function memory selection MG series OPERATING MANUAL - SEFELEC Select the memory number with the right arrow. With the up and down arrow keys, increase or decrease the memory number (from 0 to 9). Validate the function with its number memory by pressing on the enter key. 7.3 SERIES OF TESTS WITH MANUAL CONTROL By default, the proceeding of a sequence is automatic. For instance, in the case of a sequence including the following lines : LINE 1 : MEGOHM 3 LINE 2 : STRENGTH 4 The strength test will be carried out immediately after the insulation measurement without intervention of the user. It is possible to control the step from one function to the other by inserting a pause between the lines. The sequence will be as follows : MG SERIE Revision : R Page : 63

64 7.4 SIMPLE Sequence performing From the measurement screen (figure 41), press on [MEASURE-DISCHARGE] button. The indicator inside the button is illuminated. Warning : Always make sure the high voltage indicator is not on when connecting or disconnecting the specimens. Tests go on and the result of each is written as soon as it is finished. When option 92 (Printer output) is installed in the unit, and when the result ticket print out has been activated (see section 8.4), the sequencing will be ended by the result ticket print out. At the end of the sequence, the unit indicated END OF TEST. It possible to read the result of each step of the sequence. Press on [MEASURE-DISCHARGE] and on [ESC]. Warning : Test results will be lost once the MEASURE-DISCHARGE key is pressed. Page : 64 MG SERIE Revision : R

65 7.4.1 Sequence performing with manual control Compared to the example on drawing n 47, the proceeding of a sequence is slightly different. After the first test performing, i.e. insulation test, the screen displays as follows : The test in progress is stopped until the user presses the [ENTER] key. 7.5 REPEAT FUNCTION OF THE GROUND CONTINUITY According to the device to be tested, it may be necessary to proceed to several tests on different points. This function is called multiple ground continuity and is set as follows : In the above example, the user needs to test in the same conditions and successively two "Ground" points. The number of test points is 2 min. and 99 max. MG SERIE Revision : R Page : 65

66 7.5.1 Proceeding to multiple ground continuity If we refer to the above example where the user has to carry out two successive ground continuity tests in two different points of an electrical device, When starting the test, the screen displays as follows : The first ground continuity test (STEP : 1) starts only when the user presses the [ENTER] key (or presses the button on the test probe if he has one). The result of the measurement appears on line 1. Then the screen displays as follows : As mentioned above, the device stops until the user presses the [ENTER] key. This operation is repeated as much time as steps have been programmed under the condition that the test result is PASS. If not, the user has the opportunity either to repeat the failed test or to stop the sequence at this step (press END). Page : 66 MG SERIE Revision : R

67 7.5.2 Multiple ground continuity followed by a strength test The case of a multiple continuity immediately followed by a strength test requires a precise care. The multiple ground continuity function means that the user sometimes enters the security area or is in direct contact with the DUT. For that reason, in order to protect the user's safety, the SMG stops the proceeding of the test before starting the strength test. At this stage, test can be released by : 1. With the keyboard : Press [ENTER] and then [LOW ARROW] within 0,5 s and keep this key pressed until it is taken into account. 2. Using the safety loop : If the safety loop is open during or at the end of the multiple continuity test, closing of the safety loop will start the strength test. MG SERIE Revision : R Page : 67

68 7.6 Error messages Sequence error messages INTERLOCK DISABLE : The safety loop is not closed. There is no connection between the points 1-14 and of the 25 points Z10 connector of the rear panel. Make the connection, press again on the [MEASUREMENT-DISCHARGE] key to trigger the measurement mode. BOARD NOT READY : The microprocessor board cannot communicate with one of the test board. You cannot perform a measurement, get in contact with our Service department. PARAMETER ERROR : Some function parameter are not correct. Verify parameters of the functions of your sequence. SEQUENCE ERROR : The programmed sequence includes at least one error such as : 1. Less than 2 points for a multiple ground continuity 2. No PAUSE between 2 tests 3. Several pauses have been set one after the other CONFIGURATION ERROR : In the general configuration function of the device, on the INTERFACE line, option "WITHOUT" must be selected Function error messages See paragraph 3 section 3, 4 and 5 for the signification of the different error messages. Page : 68 MG SERIE Revision : R

69 SECTION 8 : INPUT-OUTPUT INTERFACE 8.1 Input-Output for Programmable Logic Controller Interface (PLC) - 02 option This option gives a functional control of the measurement unit by a Programmable Logic Controller system Electrical characteristics INPUTS : OUTPUTS : - Number : 7 - Type : Optoelectronic - Input resistance : 1.5 kohms - Minimum voltage : 11 VDC or 8 VAC - Maximum voltage : 43 VDC or 30 VAC - Number : 5 - Type : Dry contacts - Maximum voltage : 70 VDC - Switching rating : 30W - Maximum current : 0.15 ADC Description of the logical states Input : Logical state HIGH : DC or AC voltage with an amplitude between Umin and Umax. Logical state LOW : no voltage. Output : Logical state HIGH : closed contact. Logical state LOW : opened contact. MG SERIE Revision : R Page : 69

70 8.1.3 Connections: The previously described set of input/output signals are available on the rear panel of the unit on a 25 points Sub-D connector. 1 Safety loop: connect to 14 2 RED light (test voltage generated) : refer to connection drawing 3 25 VDC protected by internal fuse 4 I/O of the PLC option (opt02) = CTRLIN 5 I/O of the PLC option (opt02) = N1 6 I/O of the PLC option (opt02) = N3 7 I/O of the PLC option (opt02) = MES_DCH 8 I/O of the PLC option (opt02) = COM_OUT 9 I/O of the PLC option (opt02) = FAIL 10 I/O of the PLC option (opt02) = PASS 11 Input 0-10 volts (opt03) 12 Ground 13 Remote Control: connect to Safety loop: connect to 1 15 GREEN light (no test voltage) : refer to connection drawing 16 I/O of the PLC option (opt02) = COM_IN 17 I/O of the PLC option(opt02) = N0 18 I/O of the PLC option (opt02) = N2 19 I/O of the PLC option (opt02) = TYPE 20 Not used 21 I/O of the PLC option (opt02) = CTRLOUT 22 I/O of the PLC option (opt02) = EOT 23 I/O of the PLC option (opt02) = ERROR 24 Output 0-10 volts (opt03) 25 Remote Control: connect to 13 RED GREEN Red and green lights connection drawing In case of connection with a PLC system or any other relay control system, make sure to use shielded wire with braid connected to the metallic cover of Sub-D connector (connection at 360, shielding connected at each side of the cable). Page : 70 MG SERIE Revision : R

71 8.1.4 Description of the input output signals * COM_IN : electrical common between input signals. * COM_OUT : electrical common between output signals. * CTRLIN (input) : request for remote control of the measurement unit. * TYPE (input) : selection of the measurement function : Logical state HIGH : Megohmmeter. Logical state LOW : Strength Test. * MES_DCH (input) : selection of the measurement or discharge state : Logical state HIGH : Measurement Logical state LOW : Discharge * N0,N1,N2,N3 (input) : binary coded data lines for the number of the parameters set : N3 N2 N1 N0 Memory * CTRLOUT (output) : acknowledgment of the CTRLIN signal. * EOT (output) : the contact of the relay is closed at the end of the test. * PASS (output) : the contact of the relay is closed when the test pass. * FAIL (output) : the contact of the relay is closed when the test fail. * ERROR (output) : the contact of the relay is closed when an error occurs ('CHARGE', 'OVERLOAD', 'OVERVOLTAGE', 'BOARD NOT READY', 'SYNCHRO ERROR', 'INTERLOCK DISABLE', 'VOLTAGE ERROR').. MG SERIE Revision : R Page : 71

72 8.1.5 Measurement / Discharge cycles MG series OPERATING MANUAL - SEFELEC IMPORTANT : PLC function must be selected in the configuration menu (line INTERFACE : PLC). To take the control of the unit, you must be on the first screen. If not, you will have the DIALOG ERROR : <2> message displayed. Note : For DMG50 and DMG500 in sequence mode,the control is taken from the measurement screen of the first sequence function. To do a measurement : CTRLIN must be on high logical state, Except for CMG30, SMG50 and SMG500 where the state of the TYPE Line is not taken into account, the programmable logic controller must select a measurement function (Megohmmeter or dielectric strength test). For MMG or RMG, the choice of the measurement function must be the measurement function of the unit. If it is not the measurement function, the unit gives an error. To avoid this, it is advised to put TYPE line on a high logical state for a MMG and for the RMG, it advises to let it not connected. Thus, measurement function used will be measurement function of the unit. For CMG30, the TYPE line is not used. The programmable logic controller must choice a number of parameters which will be binary coded with N0 to N3 bits (N0 is the least significant bit and N3 is the mot significant bit). To enter in measurement mode, a rising edge must be put on the MES_DCH line. The high logical state must be set all during the measurement. Note : input states are enabled only on a rising edge of MES_DCH. Then the unit will do a measurement or, if parameters aren't correct, will give an error. The CTRLOUT signal will appear at the first measurement and stay until a low logical state is set on the CTRLIN. During measurement, the unit will send various signals according to the test running and the selected parameters. Those parameters can be an error signal (ERROR), an end of test signal (EOT), or a signal showing the test result (PASS or FAIL). Those signals are active on a high logical state. For CMG30, DMG50, DMG500, RMG510 and RMG500, each measurement is made with aone measurement function. In the xase of SMG50, SMG500, DMG50 and DMG500 used in sequence mode, the measurement includes all functions of the sequence. Warning : For SMG50 and SMG500, only the sequence mode can be driven by the PLC option. To force the unit to go in the discharge state, the programmable logic controller must set a low logical state on the MES_DCH input. Then, the programmable logic controller can reset the CTRLOUT signal by resetting the CTRLIN, set the MES_DCH and reset it. Drawing n 1 : CTRLOUT signal off Page : 72 MG SERIE Revision : R

73 Minimum test time in HIPOT : MG series OPERATING MANUAL - SEFELEC Measurement parameters : - Software revision = 1.57 or higher for DMG & RMG or higher for SMG50 - MEASUREMENT DISPLAY = OFF (second page of the SETUP menu) - INTERFACE = PLC - Rise time = 0 - Fall time = 0 - Hold time = 1 - Time = AUTO - Other test parameters = independent Test time = time between change of the MES_DCH (start test) signal and change of EOT (end of test) signal = 980 ms Measurement parameters : same as above but with hold time = 0 Test time = 700 ms. In this case there is no output voltage control. Measurement parameters : memory number change between each measurement Test time = basic minimum test time S Minimum test time in MEGOHMMETER : Measurement parameters : - Software revision = 1.57 or higher for DMG & RMG or higher for SMG50 - MEASUREMENT DISPLAY = OFF (second page of the SETUP menu) - INTERFACE = PLC - Hold time = 1 - Other test parameters = independent Test time = time between change of the MES_DCH (start test) signal and change of EOT (end of test) signal = 1.58 S Measurement parameters : memory number change between each measurement Test time = basic minimum test time S Minimum test time in GROUND CONTINUITY : Measurement parameters : - Software revision = 1.01 or higher - MEASUREMENT DISPLAY = OFF (second page of the SETUP menu) - INTERFACE = PLC - Rise time = 0 - Fall time = 0 - Hold time = 1 - Other test parameters = independent Test time = time between change of the MES_DCH (start test) signal and change of EOT (end of test) signal = 1.19 S Measurement parameters : memory number change between each measurement Test time = basic minimum test time S MG SERIE Revision : R Page : 73

74 Drawing n 2 : Example of test with megohmmeter function on DMG50, DMG500, MMG500 Memory 5 - Test failed. Page : 74 MG SERIE Revision : R

75 Drawing n 3 : Example of test with strength test function on SMG50 or SMG500 Memory 3 - Test passed. MG SERIE Revision : R Page : 75

76 Drawing n 4 : Connection of PLC option with external supply. PLC SYSTEM MG UNIT Page : 76 MG SERIE Revision : R

77 Drawing n 5 : Connection of PLC option with internal supply. PLC SYSTEM MG UNIT MG SERIE Revision : R Page : 77

78 VOLTS ANALOG INPUT-OUTPUTS - 03 OPTION This option allows the unit to deliver a proportional analog voltage to the displayed measurements on the LCD screen (insulation resistance, test voltages, leakage currents) and allows to control the dielectric strength test high voltage by an analog voltage between 0 and 10 volts. Jumpers for the setup of the board allow to select different features Electrical specifications Output voltage Input voltage : from 0 to 10 VDC by 2.44 mv steps : from 0 to 10 VDC Output impedance : 1 kohm ±5% Input impedance : 10 kohms ±5% Insulation Connection : Not insulated, the ground is earthed : 25 points, sub-d connector on the rear panel 11 0 to 10 VDC input or output - option Ground 24 0 to 10 VDC output - option Possible setups according to the units The board includes some jumpers described on the schematic underneath. These jumpers allow to setup the board in a 2 analog outputs mode or in a 1 analog output + 1 analog input mode Page : 78 MG SERIE Revision : R

79 LIST OF THE SETUPS : MG series OPERATING MANUAL - SEFELEC MEGOHMMETER : 1 output for the linear value of the resistance in a range 1 output for the value of the range J1 OFF (factory setup) J J J45 ON MEGOHMMETER : 1 output for the logarithmic value of the resistance 1 output not used J1 ON J J or 2-3 (not used in megohmmeter) J45 ON DIELECTRIC STRENGTH TEST : 1 output for the high voltage value 1 output for the leakage current J1 OFF (factory setup) J J J45 ON DIELECTRIC STRENGTH TEST : 1 output for the high voltage value 1 input for the high voltage Change of the board setup J1 ON J J J45 ON IMPORTANT : the change of the board setup must be performed by skill people. Remove the power cord Remove the side handles screws Push the gray cover backward to access to the second board (remove the earth wire if necessary) Unscrew the board locking nut with a 7mm box spanner Pull the board out Set the jumpers as described here above Plug the board Lock the nut Put the cover in its original position (with the earth connection) Screw the side handles Connect the power cord Put the power on MG SERIE Revision : R Page : 79

80 8.2.4 Operating the 0 to 10 volts feature MG series OPERATING MANUAL - SEFELEC Go in the [SETUP] menu Go in <page 2> On the INTERFACE line, select the PLC mode (Programmable Logic Controller) If the [MISSING OPTION] message is displayed, please check that the board has been correctly installed. IMPORTANT : if the PLC mode isn't selected, the Analog input-output function isn't operating Megohmmeter mode, 2 outputs setup Refer to section for the board jumpers setup. This setup is the factory default setup. The pin 24 of the rear panel connector gives the value of the resistance in a range (V1). The pin 11 of the rear connector gives the value of the range (V2). V1= value in a range, from 0 to 10 volts compared to the displayed value V2= value of the range : kohm = 0 volt kohm = 1 volt Mohm = 2 volts Mohm = 3 volts Mohm = 4 volts Gohm = 5 volts Gohm = 6 volts Gohm = 7 volts Tohm = 8 volts The resistor value is given by the following formula : R = 2 x V1 x 10 V2 kohms For example : if V1=5.2 V and V2=3.0 V then R=10.4 Mohms Accuracy of the 0-10 volts output voltage compared to the display : ±(0.1% + 5 mv) Page : 80 MG SERIE Revision : R

81 8.2.6 Megohmmeter mode, 1 output setup MG series OPERATING MANUAL - SEFELEC Refer to section for the board jumpers setup. The pin 24 of the rear panel connector isn't used The pin 11 of the rear connector gives the logarithmic value of the measured resistor (V1) The resistor value is given by the following formula : For example: if V1=3.7V then R=10.0 Mohms R = 2 x 10 V1 kohms Accuracy of the 0-10 volts output voltage compared to the display : ±(0.1% + 5 mv) Accuracy of the resistance computed with the output voltage : ±1% Dielectric strength test mode, 2 outputs setup Refer to section for the board jumpers setup. This setup is the factory default setup. The pin 24 of the rear panel connector gives the value of the measured high voltage (V1) : from 0 to 10 volts for 0 to 5000 VAC from 0 to 10 volts for 0 to 6000 VDC The pin 11 of the rear connector gives the value of the leakage current (V2) : from 0 to 10 for 0.00mA to 9.99 ma Test voltages and leakage current are given by the following formulas : HTVAC = (V1/10) x 5000 volts HTVDC = (V1/10) x 6000 volts I = (V2/10) x Imax With Imax = 9.99mA for RMG50, DMG50 and SMG50 With Imax = 99.9mA for RMG500, DMG500 and SMG500 Accuracy of the 0-10 volts output voltage compared to the display : ±(0.1% + 5 mv) MG SERIE Revision : R Page : 81

82 8.2.8 Dielectric strength test mode, 1 output + 1 input setup Refer to section for the board jumpers setup. This setup is operating only in the MANUAL mode Select HIPOT function Select PARAM On <page2>, on the line TIMER : select the MANUAL mode The pin 24 of the rear panel connector gives the value of the measured high voltage (V1) : from 0 to 10 volts for 0 to 5000 VAC from 0 to 10 volts for 0 to 6000 VDC Test voltages are given by the following formulas : HTVAC = (V1/10) x 5000 volts HTVDC = (V1/10) x 6000 volts Accuracy of the 0-10 volts output voltage compared to the display : ±(0.1% + 5 mv) The pin 11 of the rear panel connector is an input which can receive voltages between 0 and 10 VDC for output high voltages from 0 to the maximum value displayed on the parameter line: VOLTAGE:x.xx VAC. If the input voltage is higher than 10 V +5%, the following message is displayed : [LIMIT ERROR] Accuracy of the output high voltage compared to the input voltage : ±(1% + 20 volts) for a leakage current < 100uA There is no automatic adjustment of the output voltage according to the load. Output high voltage response time for an input voltage variation : with display mode without display mode : < 1 second : < 0.5 second Operating instructions Use a shielding cable for the analog input and output lines. Connect the shielding to the pin 12 of the connector. Page : 82 MG SERIE Revision : R

83 8.3 RS232 INTERFACE (01 OPTION) IMPORTANT: the RS232C function must be selected in SETUP menu, <page2> line INTERFACE: RS232 (see section 2.8). The test units of the MG series can be equipped with a RS232C interface operating in talker and listener modes. This option makes possible the integration of any unit of the series in an automatic measurement or test system in manufacturing or incoming inspection department. A 9 pins connector is provided on the rear panel for the interface connection. The RS232C standard defines electrical specifications for the transmission of serial information. The use of the RS232C port requires five lines: Receive data Transmit data Data terminal ready Data set ready Signal ground (RXD) (TXD) (DTR) (DSR) (GND) This interface also requires a cable type CO179. Refer below for the cable wiring. FOLD THE BRAIDED WIRES AND TIGHTEN THEM INTO THE CLAMPING RING The communication parameters cannot be changed and are : Speed : bauds for "MG+" series and MG series with MG-70 option, 9600 bauds for other. Parity : no Format : 8 bits Stop bit : 1 MG SERIE Revision : R Page : 83

84 8.3.1 Syntax rules The end of a message must be the LF character (hexadecimal 0A, decimal 10). The separators inside a message are ; or : The commands can be sent either in small letters or in capital letters. The maximum number of characters in the message is 100 or 8 different commands The end of the execution of a complete message by the unit is indicated by the emission of the Xon (hexadecimal 11, decimal 17) character and allows to synchronize the communication with the computer. The computer must wait the Xon character before sending the next message. The events occurring during the measurement are indicated to the computer by the emission of the Z character (format error, end of test, interlock open,...). To activate this function, it is necessary to send the SRQ command after the initialization of the unit. When receiving this character, a special command allows the computer to get the event (commands *STB?, *ESR?) If the message is not recognized by the unit, the error message <DIALOG ERROR : 1> is displayed If the code is out of context, the error message <DIALOG ERROR: 2> is displayed (special code for a function when the function was not selected or numerical value out of range) List of the RS232C The syntax of the commands complies with the IEEE488-2 standard (1992 revision). IEEE488-2 numerical formats : Format NR1 : ±<digit>...<digit> Format NR3 : ±<digit>...<digit>.<digit>...<digit>e+/-<digit>..<digit> Note : Codes in brackets are expansed codes which can be understood by the unit General commands REM(REMote) GTL(GoToLocal) LLO(LLockOut) : Go to remote mode. WARNING : first command to be sent : Go back to the local mode. : Return to the local mode is locked Common commands *SRE <NR1> : "Service Request Enable Register". Enables the corresponding summary messages (bits) in the status byte register. Thus,the application programmer can select reasons for the device to issue a service request (Z character). See *STB? code. *ESE <NR1> : "Standard Event Enable Status". Select which event bits in the corresponding Event register will cause a TRUE summary message when set. By use of the enable bits the programmer can program the device to request for a single event or an inclusive OR of any group of events. *CLS : Sets all the standard registers in the state the programmer founds them after a power on. *RST : The Clear status command almost resets the apparatus as a power on. WARNING : the unit goes back to Local mode, send a REM command before any following commands. Page : 84 MG SERIE Revision : R

85 Common queries *STB? : return in hexadecimal format a <NR1> which is the value of "STB". (i.e: "#H80") b0 : 0 = Interlock opened 1 = Interlock closed b1 : 0 = No error 1 = Error (voltage not correct,...) b2 : 0 = End of test 1 = Test running b3 : 0 = Test failed 1 = Test pass b4 : Not used b5 : Logical OR of the ESR register bits b6 : Logical OR of the STB register bits b7 : Not used *SRE? : return in hexadecimal format a <NR1> which is the value of "SRE". (STB register mask) *ESR? : return in hexadecimal format a <NR1> which is the value of "ESR". b0 to b3 : Not used b4 : 1 = Dialog error type 2 (Numerical value out of range, out of context command) b5 : 1 = Dialog error type 1 (incorrect command) b6 : Not used b7 : 1 = Power on NOTE : the bits are reset after the reading of the byte by the *ESR? command *ESE? : return in hexadecimal format a <NR1> which is the value of "ESE". (ESR register mask) *LRN? : This device query allows the programmer to receive a response message units that informs the programmer on the current device state (function's running and current test parameters) and may be later used as program message unit elements to place the device in the state it was. *IDN? : Allows the identification of the unit. The message is as follows : <field1>,<field2>,<field3>,<field4> field1 : Manufacturer name = Sefelec field2 : Unit reference = CMG30, DMG50, DMG500, RMG50, MMG500, SMG50... field3 : Serial number = 0 (not used) field4 : Software revision = VERSION 1.60 NOTE : the command must be used before selecting a function (from the start display) *TST? : Allows to check that the unit is correctly working. The message is as follows : #H<NR1> with: b3 b2 b1 b0 Result Hipot not present X X X 1 #H01 Megohmmeter. not present X X 1 X #H02 Ground not present X 1 X X #H04 Leakage not present 1 X X X #H08 with X = 0 or 1 NOTE : this command must be used before selecting a function (from the start display) MG SERIE Revision : R Page : 85

86 Device commands MEG(MEGohmmeter) : Selects the megohmmeter. Must be sent from the initialization display. HIP(HIPot) : Selecs the hipot tester. Must be sent from the initialization display. LEAK(LEAKage) : Selecs the leakage function. Must be sent from the initialization display. GND(GrouND) : Selects ground continuity function. Must be sent from the initialization display. SEQ(SEQuence) : Selects the sequence function. Must be sent from the initialization display. CONF(CONFig) : Selects the configuration function of the unit. Must be sent from the initialization display. PAR(PARameter) <NR1> : Selects the parameter set for the running function. For instance : MEG : PAR 1 DCV(DCVoltage) <NR1> : Sets a new value in volt to the current DC voltage parameter to the parameter set of the running function. For instance : DCV 500 ACV(ACVoltage) <NR1> : Sets a new value in volt to the current AC voltage parameter to the parameter set of the running function. - Hipot tester : Min. : 10 Max. : 5000 limited 4200 with FMG501 - RMG15AC : Min : 10 Max : Leakage : Min. : 200 Max. : 270 For instance : hipot : ACV 5000 leakage : ACV 244 ACC (ACCurrent) <NR3> : Sets a new value in ampere to the AC current parameter to the parameter set of the ground continuity function. OHM (OHMmeter) : Sets the main thresholds unit (Ohm) and the ground continuity display function. Caution : The action on this command will erase all previous recorded thresholds (Reset to zero). VOLT (VOLTmeter) : Sets the main thresholds unit (Volt) and the ground continuity display function. Caution : The action on this command will erase all previous recorded thresholds (Reset to zero). HTIM(HTIMe) <NR1> : Sets a new value in second to the current hold time parameter to the parameter set of the running function. Min. to the leakage current function : 2 Seconds For instance : HTIM 3 RTIM(RTIMe) <NR1> : Sets a new value in second to the current rise time parameter to the parameter set of the running function For instance : RTIM 10 FTIM(FTIMe) <NR1> : Sets a new value to the current fall time parameter to the parameter set of the running function. For instance : FTIM 5 UHLIM(UHLIMit) <NR1> : Sets a new value to the high limit current AC voltage of the leakage current function. Min : 200 Max : 270 For instance : UHLIM 244 Page : 86 MG SERIE Revision : R

87 ULLIM(ULLIMit) <NR1> : Sets a new value to the low limit current AC voltage of the leakage current function. Min : 200 Max : 270 For instance : ULLIM 230 HLIM(HLIMit) <NR3> : Sets a new value to the high limit parameter to the parameter set of the running function. For instance : MEG:HLIM 2.0E+6 (new value in Ohms) RIG:HLIM 1.45E-4 (new value in Amps) GND:HLIM 1.00E-1 (new value in milliohm) LEAK:HLIM 3.5E-3 (new value in Amps) The unity is automatic, given in Volt, Ohm or Amp and function dependent. LLIM(LLIMit) <NR3> : Sets a new value to the low limit parameter to the parameter set of the running function. For instance : MEG:LLIM 1.0E+6 (new value in Ohms) RIG:LLIM 3.50E-6 (new value in Amps) GND:LLIM 5.02E-2 (new value in milliohm) LEAK:LLIM 1.5E-3 (new value in Amps) The unity is automatic, given in Volt, Ohm or Amp and function dependent. WAY A1,A2,A3...A1A2A3A4PE : Sets the test mode to the leakage measure. Available modes are the followings : A1, A2, A3, A4, B, A1PE, A2PE, A3PE, A4PE, BPE, A1A2, A1A2A3, A1A2A3A4, A1A2PE, A1A2A3PE, A1A2A3A4PE, A1A4PE, A2A4PE, A3A4PE. NORM , 60950, , , 60065, 60990, : Sets the standard to use regarding the type of the device under test to leakage current function. BREAK ON / OFF / END / ALL : Sets the possibility to start time counting before / after operator intervention to leakage current function. - OFF : No pause. - ON : Pause before measurement. - END : Pause after measurement. - ALL : Pause before and after measurement. CAP ON/OFF : Sets the capacitor mode to leakage current function. POWER EXT/INT : Set INTernal or EXTernal power line to the leakage current function. CORR ON/OFF : Sets the correction mode to the leakage current function. CONT : In the leakage current function, when the mode PAUSE is actived, the time counting is waiting for an operation to start. This command is must be used to perform this. MEDI : Only concerns the leakage measurement function. Allow to receive the measured values continuously without that the PC asks them for. This code is sent 3 seconds after the MEAS code. VAL AC / CR / A / COS : Only concerns the leakage measurement function. Allow to send the function parameters that will be displayed on the LCD central section (in big characters). - AC : true RMS leakage current value - CR : peak leakage current value (depends on the leakage measurement options) - A : phase current value (only with power measurment option) - COS : cosinus Phi value (only with power measurment option) MG SERIE Revision : R Page : 87

88 UNITR V / W / VA : Only concerns the leakage measurement function. Allow to send the function parameters that will be displayed on the LCD top right section (in small characters). - V : true RMS voltage value between phase and neutral. - W : apparent power value (only with power measurement option) - VA : active power value (only with power measurement option) SCH G / H / F3 / F4 / F5 / A / B / A1 / A2 / A3 : Diagram setup (SCHema) according the standard. Only for Leakage function. - G / H : Diagram for EN standard (MG-56) before June F4 / F5 : Diagram for EN standard (MG-56) after June F3 / F4 / F5 : Diagram for CEI990 (MG-60). - A / B : Diagram for EN60065 standard (MG-59). - A1 / A2 / A3 : Diagram for EN standard (MG-61). SWI -- / ID / AP / APID : Switching setup (SWItching) used according the standards. Only for leakage function. (Aa, Bb, Cc, Dd, Ee, Ff, Gg, Hh) : Parameters setting a sequence of 8 tests of the selected memory only for SMG50 and SMG500. Each test is defined by 2 letters : one for the function (M for megohmmeter, H (HIPOT) for strength, G (GROUND) for earth continuity, X(points=) for test number in case of multiple earth continuity test, P(Pause) to place a waiting phase between two consecutive tests and L (LEAKAGE) for leakage current), and one digit to indicate the memory number (cases M, H, G, L, V). V0 is an empty test. The case X is special, because it can be followed by one or 2 digits. These one indicate the step number of the multiple continuity. This number can be set from 2 to 99.Caution : All 8 tests must be carried out (Enter V0 for an empty test). For instance : SEQ : PAR 0 : (G1,M2,H1,M3,L1,V0,V0,V0) DISP(DISPlay) ON/OFF : This command is attached to the configuration function and controls the display mode. For instance : CONF:DISP OFF TIM(TIMe) AUT/FAIL/UDIV2 : Select the temporization mode. AUT for automatic, FAIL for default and UDIV2 for U divided by 2. The UDIV2 mode concerns the hipot function. The FAIL mode concerns the ground continuity function and the hipot 500VA function only. MOD (MODe) AUT/MAN : This command is attached to the setup function. It switches the mode AUTO and MANUAL of the sequence function (only with the family units of SMG and FMG). SBS (Step By Step) ON/OFF : This command takes part of sequence function of the SMG and FMG. It fixes on the way of the test results are send back to the controller (RS232, IEEE). In ON mode, the result of every steps of test is transmitted as soon as available,without any request before (no MEAS?). In OFF mode, all the results are transmitted together at the end of the sequence in answer to the command MEAS? IMPORTANT : After power on the mode SBS is automatically set with OFF value. FILT(FILTer) NOR/CAP/RHT : Enables the normal or the capacitor or the Real Hold Time measurement modes. For instance : CONF:FILT CAP DET(DETection) OFF / I / FI / I+DELTA / FI+DELTA / DELTA : Hipot function only sets the mode of the breakdown. MEAS(MEASure) : Runs the current function. Page : 88 MG SERIE Revision : R

89 STOP : Stops the current function. QUIT : Exits the current function. SRQ : Similar to the Service Request feature of the IEEE488-2 bus, this code allows the emission of the Z character to inform the computer about the events (end of test, format error,interlock open,...). This command has to be sent at the program start, after the REM command. MG SERIE Revision : R Page : 89

90 Device queries MEAS? : Return the current measurement value(s) regarding the running function. A) Following caracters in the function Insulation Dielectric strength Continuity Continuity Leakage : OHM 4.700E+06 : VOLT 9.900E+02 AMP 7.000E-05 : OHM 3.210E-1 VOLT 2.810E+00 (main unit =OHM) : VOLT 2.830E+00 OHM 3.230E-1 (main unit =VOLT) : - when the measure is running : XX VOLT <NR1> AMP <NR3> where XX is the name of the current step (A1,A2,A3,A4,B) - when the measure is finished (Discharge), it returnes all the values delimited by a comma «,». For instance : Between - A2 VOLT 234 AMP 1.02E-3 After - A1 VOLT 232 AMP 1.02E-3,A2 VOLT 234 AMP 0.02E-3 B) Following characters in sequence mode For instance : L1 C0: 0.15mΩ 0.00V,L2 R0: 1.50KV 0.02mA,L3 M0: 41.7 GΩ,L4 F0:q 0.01mA 223V A2,L5..:,L6..:,L7..:,L8..: The syntax rules are the following : the end of the message is : CR ( Carriage Return) each test step starts with a «L» and ends with «,» ( except for the last one). the «L» letter is followed by a number between 1 and 8, giving the test step number and then by a «space». then comes a letter according to the type of the test step : Function French English German Insulation M M M Dielectric strength R H H Continuity C G E Leakage F F A Void Paused OK. OK. OK the measurement type is followed by the memory number which contains the measurement parameters, then by a separator character «:» if the test step is void, the measurement will be replaced by a «space» otherwise, the next character gives the test step result : «space» indicates that the test is good «q» indicates that the test is bad for the insulation and dielectric strength tests the previous character is followed by 1 space and for the ground continuity test by 2 space. then the numerical values are given. When the value cannot be measured ( no continuity or overload ) the numerical indication is replaced by «----» To check if a tested specimen is good or bad at the end of a sequence, it is necessary to check the STB register ( bit 3).To get the result of each test step, check if the result character is «space» or «q». Page : 90 MG SERIE Revision : R

91 RS232 commands summary CODE Startup Cnf. mω kv MΩ ma Seq. REM X X X X X X X *CLS X X X X X X X *ESE X X X X X X X *ESE? X X X X X X X *ESR? X X X X X X X *IDN? X *LRN? X X X X *RST X X X X X X X *SRE X X X X X X X *SRE? X X X X X X X *STB? X X X X X X X *TST? X GTL X X X X X X X LLO X X X X X X X ACC X ACV X X BREAK X CAP X CONF X X CONT X CORR X DCV X X DET X MOD X SBS X CODE Startup Cnf. mω kv MΩ ma Seq. MG SERIE Revision : R Page : 91

92 CODE Startup Cnf. mω kv MΩ ma Seq. DISP X FILT X FTIM X X GND X X HIP X X HLIM X X X X HTIM X X X X LEAK X X LLIM X X X X MEAS X X X X X MEAS? X X X X X MEG X X NORM X OHM X PAR X X X X X POWER X QUIT X X X X X X X RTIM X X SEQ X X SRQ X X X X X X X STOP X X X X X TIM X X UHLIM X ULLIM X VOLT X WAY X VAL X UNITR X CODE Startup Cnf. mω kv MΩ ma Seq. Page : 92 MG SERIE Revision : R

93 8.3.3 HINTS AND PROGRAM EXAMPLES The information hereunder are from a control program for a MG unit, written in Quick Basic. These are tutorial information to help you to write your own software. You will find at the end of this section the complete demo program. Serial port RS232 initialization OPEN "COM1:9600,N,8,1" FOR RANDOM AS #1 ON COM(1) GOSUB INTERUPTIONRS COM(1) ON enable of interrupts on COM1. Serial port COM1 open with following setup : - speed : 9600 bauds, - N : no parity control, - data bits : 8, - stop bit : 1. If a character arrives on the serial port, the program will go automatically to the subroutine INTERUPTIONRS RS232 interruptions subroutine INTERUPTIONRS : COM(1) OFF : nochr = 0 disable of RS interruptions char$ = INPUT$(1, #1) 'Reading of the first character (8bits) on COM1 IF char$ = CHR$(17) THEN varxon = 1: nochr = 1 'received character is XON IF char$ = "Z" THEN received character is Z,and is equivalent to a GOSUB GESTIONSRQ service request from the MG unit (end of test, nochr = 1 error...) END IF IF nochr = 0 THEN LINE INPUT #1, msg$ 'if other character, reading of the complete message msg$ = char$ + msg$ Rebuilt of the message varxon = 1 'Enable of next command emission (no XON wait) eom = 1 'End of the message (eom=1) END IF COM(1) ON RETURN The nochr variable makes the difference between the XON et Z characters (nochr=1) and the others characters (nochr=0). MG SERIE Revision : R Page : 93

94 Emission of a command on the RS232 port Before sending a command on the RS232 port, the program must wait for the Xon reception of the previous command in order to synchronize the communication. Warning, for the first REM command, the program doesn t have to wait for the Xon reception of a non existing previous command. A$ holds the code to be sent with the LF character to indicate the end of the command bloc. ENVOI ATTENTEXON varxon = 0 PRINT #1, A$ + CHR$(10) RETURN 'waiting for the Xon of the previous command Emission of the command bloc + LF character Waiting for XON The subroutine makes a loop until the MG unit send the Xon character (see Xon reception in the INTERUPTIONRS sub-routine) ATTENTEXON : WHILE varxon = 0 : WEND RETURN Waiting for a message The routine makes a loop until the MG unit send a message (see the message reception in the INTERUPTIONRS subroutine) ATTENTEMSG: WHILE eom = 0 : WEND eom = 0 RETURN Page : 94 MG SERIE Revision : R

95 Analyze and emission of a command to the MG unit This part of the program allows to input through the keyboard commands and to send these commands to the MG unit. The commands can be divided in 3 groups : 1) General commands, common commands, device commands, i.e. : REM, *RST, MEG, HIP 2) Common queries i.e. : *TST?, *LRN? 3) Device queries i.e. : MEAS? According to the command, the program will wait or not for a message coming from the MG unit. For a command which isn t finishing by?, the code is sent to the MG unit and the program makes a loop on the input of the next code. For a command starting with * and finishing by?, the program sends the command and waits only for the message coming from the MG unit. For the MEAS? command, the program sends the code to the MG unit and then waits for Xon character and finally for the message. COMMAND : COM(1) ON PRINT "Enter the command to send to MG unit (RETURN to exit)" LOOP1: PRINT "A$=";: LINE INPUT A$ IF LEN(A$) = 0 THEN GOTO SORTIE ELSE GOSUB ENVOI Sending the code if A$ isn t empty IF RIGHT$(A$, 1) = "?" THEN IF LEFT$(A$, 1) <> "*" THEN MEAS? command: GOSUB ATTENTEXON - waiting for Xon GOSUB ATTENTEMSG - waiting for the message l ELSE common query GOSUB ATTENTEMSG - waiting for the message END IF PRINT "Message: "; msg$ display of the message END IF GOTO LOOP1 SORTIE : RETURN exit of the function MG SERIE Revision : R Page : 95

96 Interruptions handle Warning : To operate in the SRQ mode (the MG unit sends the Z character to indicate an end of test or a possible error), it is necessary to initialize this mode by sending the SRQ command. The call of the GESTIONSRQ routine is done from the RS232 interruption routine (INTERUPTIONRS) GESTIONSRQ : COM(1) ON enable interruption on the COM1 port A$ = "*STB?": GOSUB ENVOI Sending the *STB? command GOSUB ATTENTEMSG Waiting for the answer IF LEN(msg$) = 4 THEN errdec = VAL("&H" + RIGHT$(msg$, 2)) IF LEN(msg$) = 3 THEN errdec = VAL("&H" + RIGHT$(msg$, 1)) hexa to decimal conversion FOR i = 0 TO 7 decimal to binary conversion IF (errdec AND (2 ^ i)) <> 0 THEN bitstb(i) = 1 ELSE bitstb(i) = 0 NEXT i IF bitstb(5) = 1 THEN If ESR bit is set to 1: A$ = "*ESR?": GOSUB ENVOI Sending of *ESR? GOSUB ATTENTEMSG Waiting for the answer IF LEN(msg$) = 4 THEN errdec = VAL("&H" + RIGHT$(msg$, 2)) IF LEN(msg$) = 3 THEN errdec = VAL("&H" + RIGHT$(msg$, 1)) hexa to decimal conversion FOR i = 0 TO 7 decimal to binary conversion IF (errdec AND (2 ^ i)) <> 0 THEN bitesr(i) = 1 ELSE bitesr(i) = 0 NEXT i errors definition according to the ESR bits set at 1 IF bitesr(4) = 1 THEN PRINT "DIALOGUE 2 ERROR" IF bitesr(5) = 1 THEN PRINT "DIALOGUE 1 ERROR" ELSE message definition according to STB bits set at 1 IF bitstb(0) = 0 THEN PRINT "INTERLOCK OPEN!" IF bitstb(1) = 1 THEN PRINT "ERROR!" IF bitstb(2) = 0 THEN PRINT "END OF TEST" IF bitstb(3) = 0 THEN PRINT "TEST FAIL." ELSE PRINT "TEST PASS" END IF RETURN Page : 96 MG SERIE Revision : R

97 Functions cycling example a. Example for DMG50 or DMG500 This program automatically sequences the commands allowing to setup the MG unit, to set the insulation resistance measurement and dielectric strength test parameters and then to run tests and get back the tests results as well as possible error messages. CYCLE: COM(1) ON A$ = "REM:SRQ": GOSUB ENVOI 'remote mode and SRQ mode A$ = "CONF:DISP ON:FILT NOR:QUIT": GOSUB ENVOI 'Setup of the unit A$ = "MEG:PAR 0:DCV 500:HLIM 1.0E+9:LLIM 10.0E+3:HTIM 5:QUIT": GOSUB ENVOI 'megohmmeter parameters A$ = "HIP:PAR 0:TIME AUT:HTIM 5:RTIM 1:FTIM 2": GOSUB ENVOI A$ = "ACV 1000:HLIM 1.0E-3:LLIM 1.0E-5:DET I:QUIT": GOSUB ENVOI 'dielectric parameters A$ = "HIP:MEAS": GOSUB ENVOI 'measurement in dielectric test WHILE srq = 0: WEND: srq = 0 'waiting for end of test Srq A$ = "MEAS?": GOSUB ENVOI 'request for measurement results GOSUB ATTENTEXON 'waiting for XON GOSUB ATTENTEMSG 'waiting for measurement result message PRINT "Measurement result:"; msg$ 'display of the measurement results A$ = "STOP:QUIT": GOSUB ENVOI 'discharge and exit of the function A$ = "MEG:MEAS": GOSUB ENVOI 'megohmmeter and measurement WHILE srq = 0: WEND: srq = 0 ''waiting for end of test Srq A$ = "MEAS?": GOSUB ENVOI 'request for measurement results GOSUB ATTENTEXON 'waiting for XON GOSUB ATTENTEMSG 'waiting for measurement result message PRINT "Measurement result:"; msg$ 'display of the measurement results A$ = "STOP:QUIT": GOSUB ENVOI 'discharge and exit of the function RETURN MG SERIE Revision : R Page : 97

98 b. Example for CMG30 This program automatically sequences the commands allowing to setup the MG unit, to set the ground continuity resistance measurement parameters and then to run tests and get back the tests results as well as possible error messages. CYCLECMG: COM(1) ON A$ = "REM:SRQ": GOSUB ENVOI 'remote mode and SRQ mode A$ = "CONF:DISP ON:FILT NOR:QUIT": GOSUB ENVOI 'Setup of the unit A$ = "GND:PAR 0:DCV 6:LLIM 50.0E-3:HLIM 1.0E-1:DCC 5.0E+0": GOSUB ENVOI A$ = "TIME AUT:HTIM 5:RTIM 1:FTIM 2": GOSUB ENVOI memory 0 parameters A$ = "PAR 1:DCV 12:LLIM 0.0E+0:HLIM 5.0E-1:DCC 1.0E+1": GOSUB ENVOI A$ = "TIME AUT:HTIM 3:RTIM 2:FTIM 1": GOSUB ENVOI memory 1 parameters A$ = "PAR 0:MEAS": GOSUB ENVOI 'measurement in parameter 0 WHILE srq = 0: WEND: srq = 0 'waiting for end of test Srq A$ = "MEAS?": GOSUB ENVOI 'request for measurement results GOSUB ATTENTEXON 'waiting for XON GOSUB ATTENTEMSG 'waiting for measurement result message PRINT "Measurement result:"; msg$ 'display of the measurement results A$ = "STOP": GOSUB ENVOI 'discharge A$ = "PAR 1:MEAS": GOSUB ENVOI 'parameters 1 and measurement WHILE srq = 0: WEND: srq = 0 ''waiting for end of test Srq A$ = "MEAS?": GOSUB ENVOI 'request for measurement results GOSUB ATTENTEXON 'waiting for XON GOSUB ATTENTEMSG 'waiting for measurement result message PRINT "Measurement result:"; msg$ 'display of the measurement results A$ = "STOP:QUIT": GOSUB ENVOI 'discharge and exit of the function RETURN Page : 98 MG SERIE Revision : R

99 c. Example for SMG50 and SMG500 MG series OPERATING MANUAL - SEFELEC This program automatically sequences the commands allowing to setup the MG unit, to set the ground continuity resistance measurement parameters, insulation resistance measurement, dielectric strength and leakage current measurement and then to run tests and get back the tests results as well as possible error messages. CYCLESEQUENCE: COM(1) ON A$ = "REM:SRQ": GOSUB ENVOI Remote mode and SRQ mode A$ = "CONF:DISP ON:FILT NOR:QUIT": GOSUB ENVOI Setup of the unit A$ = "MEG:PAR 0:DCV 500:HLIM 1.0E+9:LLIM 10.0E+3:HTIM 5:QUIT": GOSUB ENVOI Parameters of insulation of memory 0 A$ = "HIP:PAR 2:TIME AUT:HTIM 5:RTIM 1:FTIM 2": GOSUB ENVOI A$ = "ACV 1000:HLIM 1.0E-3:LLIM 1.0E-5:DET I:QUIT": GOSUB ENVOI Parameters of hipot tester of memory 1 A$ = "GND:PAR 1:DCV 6:LLIM 50.0E-3:HLIM 1.0E-1:DCC 5.0E+0": GOSUB ENVOI A$ = "TIME AUT:HTIM 1:RTIM 0:FTIM 0:QUIT": GOSUB ENVOI Parameters of ground continuity of memory 1 A$ = "LEAK:PAR 0:ACV 244:UHLIM 250:ULLIM 230:WAY A1A2PE": GOSUB ENVOI A$ = "LEAK:PAR 0:HLIM 3.5E-3:LLIM 1.5E-3:HTIM 3:NORM ": GOSUB ENVOI A$ = "LEAK:PAR 0:CORR OFF:BREAK OFF:CAP ON:POWER EXT:QUIT": GOSUB ENVOI Parameters of leakage current of memory 0 A$ = "SEQ:PAR 0:(G1,M0,H2,M0,L0,VO,VO,VO):QUIT" Parameters of sequence of memory 0 A$ = "SEQ:PAR 0:MEAS": GOSUB ENVOI measurement in parameter 0 of sequence WHILE srq = 0: WEND: srq = 0 Waiting for end A$ = "MEAS?": GOSUB ENVOI Demande de résultat de mesure GOSUB ATTENTEXON Attente du caractère XON GOSUB ATTENTEMSG Attente du message de l'appareil MG PRINT "Résultat de la mesure :"; msg$ Affichage du résultat de la séquence A$ = "STOP:QUIT": GOSUB ENVOI Passage en décharge et sortie de la séquence RETURN MG SERIE Revision : R Page : 99

100 8.3.4 TROUBLE SHOOTING THE RS232C INTERFACE When the RS232C interface is not operating as described in this manual, please check the following points: 1 - No reaction from the MG unit when sending commands : The cable between the MG unit and the PC computer must be correctly connected at both ends. The cable is a special wiring cable : check that the cable is a Sefelec CO179 model or check that the wiring has been done according to the instructions of section The MG series units have several possible interface types : check that in the SETUP menu <page2>, the RS232 mode has been selected on the INTERFACE line The end of message must be the LF (hexa 0A,decimal 10) character. If this character is not sent, the MG unit won t handle the message : check that this character is added to the command The first command must be the REM command, which displays the REMLOC message on the LCD display, the unit being ready to receive all the others commands. WARNING : the first REM command should not wait for the Xon character before being sent. 2 - The unit goes in REMOTE mode and then hang up : Before sending a new command, it is mandatory to wait that the MG unit indicates its non busy state by sending the Xon character (hexa 11,decimal 17). If a code arrives during the handle of the previous one, it can produce an erratic operating mode of the unit or stop it. Page : 100 MG SERIE Revision : R

101 QUICK BASIC PROGRAM EXAMPLE General initialization : SCREEN 0, 0, 0: COLOR 14, 1: CLS varxon = 1 srq = 0 eom = 0 msg$ = "" 'Serial port RS232 initialization : OPEN "COM1:9600,N,8,1" FOR RANDOM AS #1 ON COM(1) GOSUB INTERUPTIONRS COM(1) ON 'Program main menu : KEY 1, "Codes": KEY 2, "Cycle" ON KEY(1) GOSUB COMMAND ON KEY(2) GOSUB CYCLE KEY(1) ON: KEY(2) ON KEY ON 10 GOTO 10 'RS232 interruptions subroutine : INTERUPTIONRS: COM(1) OFF: nochr = 0 char$ = INPUT$(1, #1) IF char$ = CHR$(17) THEN varxon = 1: nochr = 1 IF char$ = "Z" THEN GOSUB GESTIONSRQ nochr = 1 END IF IF nochr = 0 THEN LINE INPUT #1, msg$ msg$ = char$ + msg$ varxon = 1 eom = 1 END IF COM(1) ON RETURN Emission of a command on the RS232 port : ENVOI: GOSUB ATTENTEXON varxon = 0 PRINT #1, A$ + CHR$(10) RETURN MG SERIE Revision : R Page : 101

102 'Waiting for XON : ATTENTEXON: WHILE varxon = 0: WEND RETURN MG series OPERATING MANUAL - SEFELEC 'Analyse and emission of a command to the MG unit : COMMAND: COM(1) ON PRINT "Enter codes to send to the unit MG (RETURN pour escape)" LOOP1: PRINT "A$=";: LINE INPUT A$ IF LEN(A$) = 0 THEN GOTO SORTIE ELSE GOSUB ENVOI IF RIGHT$(A$, 1) = "?" THEN IF LEFT$(A$, 1) <> "*" THEN GOSUB ATTENTEXON GOSUB ATTENTEMSG ELSE GOSUB ATTENTEMSG END IF PRINT "Message coming from the unit MG: "; msg$ END IF GOSUB ATTENTEXON GOTO LOOP1 SORTIE: CLS RETURN Waiting for a message : ATTENTEMSG: 'srq = 0 WHILE eom = 0: WEND eom = 0 RETURN Interruptions handle : GESTIONSRQ: TEMP$ = A$ COM(1) ON A$ = "*STB?": GOSUB ENVOI GOSUB ATTENTEMSG IF LEN(msg$) = 4 THEN errdec = VAL("&H" + RIGHT$(msg$, 2)) IF LEN(msg$) = 3 THEN errdec = VAL("&H" + RIGHT$(msg$, 1)) FOR i = 0 TO 7 IF (errdec AND (2 ^ i)) <> 0 THEN bitstb(i) = 1 ELSE bitstb(i) = 0 NEXT i Page : 102 MG SERIE Revision : R

103 IF bitstb(5) = 1 THEN A$ = "*ESR?": GOSUB ENVOI GOSUB ATTENTEMSG IF LEN(msg$) = 4 THEN errdec = VAL("&H" + RIGHT$(msg$, 2)) IF LEN(msg$) = 3 THEN errdec = VAL("&H" + RIGHT$(msg$, 1)) FOR i = 0 TO 7 IF (errdec AND (2 ^ i)) <> 0 THEN bitesr(i) = 1 ELSE bitesr(i) = 0 NEXT i IF bitesr(4) = 1 THEN PRINT "DIALOG ERROR 2" IF bitesr(5) = 1 THEN PRINT "DIALOG ERROR 1" ELSE IF bitstb(0) = 0 THEN PRINT "INTERLOCK DISABLE" IF bitstb(1) = 1 THEN PRINT "ERROR" IF bitstb(2) = 0 THEN PRINT "End of Test." IF bitstb(3) = 0 THEN PRINT "Test: Fail." ELSE PRINT "Test: Pass." END IF srq = 1 A$ = TEMP$ RETURN Function cycling example for DMG50 or DMG500 : CYCLE: PRINT "Number of cycle? ";: INPUT nb COM(1) ON FOR j = 1 TO nb PRINT: PRINT "Cycle nø"; j A$ = "REM:SRQ": GOSUB ENVOI A$ = "CONF:DISP ON:FILT NOR:QUIT": GOSUB ENVOI A$ = "MEG:PAR 0:DCV 500:HLIM 1.0E+9:LLIM 10.0E+3:HTIM 5:QUIT": GOSUB ENVOI A$ = "HIP:PAR 0:TIME AUT:HTIM 5:RTIM 1:FTIM 2": GOSUB ENVOI A$ = "ACV 1000:HLIM 1.0E-3:LLIM 1.0E-5:DET I:QUIT": GOSUB ENVOI A$ = "HIP:MEAS": GOSUB ENVOI WHILE srq = 0: WEND: srq = 0 A$ = "MEAS?": GOSUB ENVOI GOSUB ATTENTEXON GOSUB ATTENTEMSG PRINT "Measure result: "; msg$ A$ = "STOP:QUIT": GOSUB ENVOI A$ = "MEG:MEAS": GOSUB ENVOI WHILE srq = 0: WEND: srq = 0 A$ = "MEAS?": GOSUB ENVOI GOSUB ATTENTEXON GOSUB ATTENTEMSG PRINT "Measure result: "; msg$ A$ = "STOP:QUIT": GOSUB ENVOI NEXT j WHILE INKEY$ = "": WEND CLS RETURN MG SERIE Revision : R Page : 103

104 'Function cycling example for CMG30 : MG series OPERATING MANUAL - SEFELEC CYCLECMG: PRINT "Number of cycle? "; : INPUT nb COM(1) ON FOR j = 1 TO nb PRINT : PRINT "Cycle n "; j A$ = "REM:SRQ": GOSUB ENVOI A$ = "CONF:DISP ON:FILT NOR:QUIT": GOSUB ENVOI A$ = "GND:PAR 0:DCV 6:LLIM 50.0E-3:HLIM 1.0E-1:DCC 5.0E+0": GOSUB ENVOI A$ = "TIME AUT:HTIM 5:RTIM 1:FTIM 2": GOSUB ENVOI A$ = "PAR 1:DCV 12:LLIM 0.0E+0:HLIM 5.0E-1:DCC 1.0E+1": GOSUB ENVOI A$ = "TIME AUT:HTIM 3:RTIM 2:FTIM 1": GOSUB ENVOI A$ = "PAR 0:MEAS": GOSUB ENVOI WHILE srq = 0: WEND: srq = 0 A$ = "MEAS?": GOSUB ENVOI GOSUB ATTENTEXON GOSUB ATTENTEMSG PRINT "Measure result:"; msg$ A$ = "STOP": GOSUB ENVOI A$ = "PAR 1:MEAS": GOSUB ENVOI WHILE srq = 0: WEND: srq = 0 A$ = "MEAS?": GOSUB ENVOI GOSUB ATTENTEXON GOSUB ATTENTEMSG PRINT "Measure result:"; msg$ A$ = "STOP": GOSUB ENVOI NEXT j A$ = "QUIT": GOSUB ENVOI WHILE INKEY$ = "": WEND CLS RETURN Page : 104 MG SERIE Revision : R

105 8.4 Printing output (92 option) Parameters setup This option allows the unit to print out a result ticket at the end of a test sequence (SMG50 and SMG500). The print out is done through a Centronics interface (parallel printer) with its connector located on the rear panel. Any model of parallel printer can be used. Connect the printer to the unit using a standard Centronics cable. The code page used is 437, allowing to print semi-graphic and symbol characters. Some parameters can be set through a menu located on the second page of the functions (FUNCT then NEXT) as follows : When pressing the «PRINT» key, the first parameters page is displayed as follows : The first line allows to enable/disable the printing at the end of the sequence. A «YES» means that the print out will be performed. In case of «NO», the test sequence will end without printing. The year can be set on the second line. The SMG unit is fitted with a real time keeper to provide Date and Hour. The third line allows to set the day and the month. The fourth line allows to set the hours and the minutes. MG SERIE Revision : R Page : 105

106 The setup of these parameters is done in factory, the timer will update automatically the data. In order to limit unwanted actions, the access is protected by a switch inside the device (INT2-refer to section 8.4.3). The following drawing shows the second page of the print out parameters. This second page of parameters can display on the ticket free information, like company name, tested product name, serial number... These data are set in 8 fields (4 lines of 2 fields) of alphanumerical characters ( A.. Z, 0.. 9, ) Operation At the end of a test sequence, the ticket is printed as follows : SEFELEC SMG :34 SN : PRODUCT : W123 : : INSULATION 15.2GΩ PASS DIELECTRIC 1.50kV 2.0mA PASS CONTINUITY 1.348mΩ PASS SEQUENCE PASS The first line gives the unit model (SMG50 or SMG500), its serial number, the date and the test hour. The second and third lines show the free information mentionned hereabove. In this example, only 4 fields have been used. The lines number 4 to 11 show each performed test step with the name of the measurement (INSULATION, HIPOT, GROUND), the measured values, and the PASS/FAIL result according to the test parameters. The last line gives the complete test result. Page : 106 MG SERIE Revision : R

107 8.4.3 Configuration The configuration of the board is done by a set of 2 jumpers located on the printer board. The printer board is a daughter board of the MPU board as describes on the following drawing. The J1 jumper, when set, will automatically increase a serial number in the free field of the first line. After each test, the numerical value will be increased : this is a factory setup. When J1 jump is removed, the free field of the line 1 is like the other lines. The J2 jumper, when set, will send automatically a Form Feed to the printer at the end of the printout. This gives only one sequence result by page. When the jumper is removed, the printouts are separated by a blank line : this is a factory setup. The following board is a summary of available configurations. Jumper Position Default Action J1 ON * Incremental serial number OFF Free field J2 ON Form Feed OFF * Line Feed To modify the date and hour, you must access to the INT2 switch. Switch INT2 in the other position, modifies the data and then sets the switch in its original position. MG SERIE Revision : R Page : 107

108 SECTION 9 : APPLICATION NOTES 9.1. WHY DIELECTRIC TESTS? The dielectric tests are performed in order : to detect manufacturing fault on electrical equipment to verify the quality of the insulating material of an electrical equipment to verify that an electrical installation has been correctly done to control the insulation resistance of an equipment or an installation to trace its changes during the years. Dielectric tests consist of both insulation resistance measurements and dielectric strength tests GLOSSARY OF TERMS LEAKAGE DISTANCE : The smallest required distance, measured on the surface of the insulating material, between 2 conductive parts, to avoid breakdown. LEAKAGE CURRENT : Steady current flowing through an insulating material subject to high voltage. BREAKDOWN : The immediate break of the dielectric property of an insulating material. Every breakdown creates more or less damages to the insulating material. The breakdown tests can be therefore destructive or not. INSULATION RESISTANCE : Characteristic of an insulating material that being subject to a voltage, shows a resistance such as the value of the leakage current which flows through it stays within acceptable limits. DIELECTRIC STRENGTH : Ratio between the voltage at which a dielectric break of the insulating material occurs and the distance between the two points subject to the voltage (generally given in kv/cm). Regarding the insulating material type (solid, liquid or gas) a dielectric break can be : a perforation, a flashover or an arc. Page : 108 MG SERIE Revision : R

109 9.3 ENVIRONMENTAL INFLUENCES Temperature, pressure and humidity conditions have an influence on the dielectric tests results. TEMPERATURE : The temperature having an influence on the gas density, this one is altering the performances of the liquid or gaseous insulating material. The oils, often used as insulation are never pure, the dissolved quantity of gas increases with the temperature and is decreasing the insulation quality of the oil. The large variety of materials used as solid insulation doesn t allow to deduce a general rule on their behavior with the temperature (the insulation specifications having a tendency to dissipate when the temperature increases). PRESSURE : The withstand voltage in gas changes with the pressure following the Paschen s law. This law shows a minimum of the breakdown voltage for a particular value of the pressure by distance product, otherwise the more the pressure increases, the more the breakdown voltage is high. The liquids used as dielectric insulation are influenced by the pressure, the dielectric strength increasing with the pressure. In theory, the solid insulations are a little influenced by the pressure because this one doesn t modify a lot their thickness and their internal composition. HUMIDITY : The withstand voltage of gas changes with humidity. In example for the air and for values of relative humidity < 80%,the dielectric strength increases a little with the humidity increasing (the water molecules,more dense than the gas, slows down the avalanche phenomena). The water presence in an insulating liquid such as oil degrades the dielectric strength by water electrolysis, development of gas producing partial discharges bringing to breakdown. Under combined effect of the humidity (>95%) and the temperature (>100 C) most of the polymers dissipate. The water can produce inflating in the insulation and create cracks which will facilitate the electric arcs advance. 9.4 INSULATION RESISTANCE MEASUREMENT The insulation resistance measurement is intended to verify that the various components and sub-assemblies of an electrical equipment have an insulation resistance such as the leakage currents don t reached inadmissible values. The principle is to apply a DC voltage, stable and specified (selected among the standard values) between defined points and after a prescribed time, to measure the current flowing through the tested material. By using the Ohm s law (resistance = voltage / current), the result is given by the value of the insulation resistance. Then this value is compared to the minimum threshold specified by the standard used for the test. MG SERIE Revision : R Page : 109

110 9.4.1 PRECAUTIONS TO BE OBSERVED It is important to connect the specimen to be measured in taking care of the parasite leakages which could be created by the measurement operating procedure. The supplied accessories have a shielding which is connected to a guard potential, this insures a good immunity of the measurement regarding the parasitic leakage currents and the AC residuals. When using extendings of the basic probes, take the necessary cares to avoid to introduce measurement errors (short leads, leads not touching any metallic or insulating parts,...) During insulation resistance measurements having high values (> 100 Gohms), the proximity of the operator putting his hand close to the specimen under test, can alter or make unstable the measurement. It is important to beware of Nylon blouses or of insulating material things capable of generating by static electricity, high electrical fields which can alter the measurement of high value resistances (100 Gohms under 100 VDC = 1 na of measurement current) MEASUREMENTS OF CAPACITORS We remind that a lot of recent electrical units are fitted with main line filters including capacitors for the electromagnetic compatibility. When measuring on capacitors it is advised to select the filter mode CAPACITOR of the SETUP menu in order to stabilize the measured values. A) Indeed, on capacitors, the variations of the measurement power supply, even small as well as the interferences are entirely transmitted to the input of the current measurement system which have a very high gain, and therefore will amplify these variations. The CAPACITOR filter switches on circuits which will limit the instability of the measurement. B) Never perform insulation resistance measurements on capacitive specimens in reducing the measurement voltage between each test, but always in increasing the voltage.the hysteresis and polarization phenomenas of the dielectric material will alter the results. In that case the unit indicates its maximum value and takes a long time to come back to the real measured value. C) The insulation resistance value of a capacitor being a function following a time exponential law, it is important to make sense to the measured value, to indicate the duration of the measurement. The units of the MG series allow to comply with this requirement with the built-in timer, able to measure times going from 1 second up to 16 minutes. D) Never disconnect a capacitive specimen before switching into DISCHARGE mode and waiting the necessary time to discharge its capacitance through the 2.2 kohms built-in resistance of the discharge circuit (about 1 second per 100 uf). Page : 110 MG SERIE Revision : R

111 9.4.3 MEASUREMENTS ON CABLES MG series OPERATING MANUAL - SEFELEC The measurement on cables is similar to the measurement on capacitors, please refer to the section for the basic advice. The measurement configurations on cables are very varied. The measurements have to be performed either between each conductors for multi-wires cables, either between main conductor and shielding for shielded cables, either between the cable and its environment for mono-wire cables. A) In that last case, the generally used way is to immerse the cable reel in a water tank (called SWIMMING POOL), to wait for the water penetration in to the cable reel center, and then to perform the insulation resistance between the cable and the water. For safety and construction reasons the water tank is grounded. The insulation resistance measurement unit must be able to measure a specimen with one grounded end. The units of the MG series allow to perform easily this type of measurement, because the hot point of the high voltage generator is already grounded. You just have to connect the measurement input of the unit (with the HV probe) on the cable to be measured and to trigger the measurement. B) Another specific point, when measuring on cables, is that the specifications of the cable manufacturers give resistance values for a standard length of cable equal to 1 km (1000 meters). When testing the reels of cable, those are never equal to the standard length, that forces the operators to calculate the resistance as a function of the cable length and the number of wires in parallel for the multi-wires cables. Consequently the built-in comparators of the measurement units can not be used, because they compare regarding to the total insulation value and not regarding the standard value. The units of the MG series allow with the option 23 to display insulation resistance measurements reduced to 1 km and 1 wire, and therefore allow the use of the built-in comparators. The operator can enter in a specific menu of the unit the length of the cable under test as well as the number of wires. The result is given in Mohm per km. i.e. : the unit measures a value of 10 Mohms for a 10 km long mono-wire cable. Therefore, the value reduced to 1 km will be : (Rtotale / 1 km) x Length = 100 Mohm.km For the same cable with 10 wires, the value for 1 wire will be : 100 Mohm.km x 10 = 1000 Mohm.km C) The insulation resistance value of a cable being a function following a time exponential law, it is important to make sense to the measured value, to indicate the duration of the measurement. The units of the MG series allow to comply with this requirement with the built-in timer, able to measure times going from 1 second up to 16 minutes MEASUREMENT VOLTAGE SELECTION The insulation resistance measurements intending to verify that materials or equipments comply with standard requirements, it is important to refer to these standards to select the voltage. The standard voltages are generally : 50, 100, 250, 500 VDC. In case of no standard, select a 100 VDC value. When measuring on capacitive specimens and when studying the voltage influence on the insulation resistance values, it is important to start always with the lowest voltage and then to follow the measurements in increasing the voltage. A procedure in the reverse way could give incoherent results. MG SERIE Revision : R Page : 111

112 9.5 DIELECTRIC STRENGTH TESTS The dielectric strength test is intended to stress components and sub-assemblies of electrical equipments and to check that the leakage lines either between points or between points and ground are correctly designed according to the used technology. The principle of a dielectric strength test is to apply a voltage (DC or AC) between defined points and after voltage stabilization, to check that the leakage current, created by breakdown phenomenas or breakdown discharges (in the air or in the insulating materials), is not greater than the nominal acceptable value. The default sanction is determined by the analyze of the shape, the amplitude and the holding time of the current supplied by the generator to the specimen under test and by comparison with a preset limit SELECTION OF THE TEST VOLTAGE The dielectric strength tests intending to verify that materials or equipments comply with standard requirements, it is important to refer to these standards to select the voltage. In case of no indication regarding the test voltage, a common rule is to apply the following formula : Utest = 2 x Unominal volts Most of the standards specify the type of the test voltage : AC (50-60Hz) or DC. A common rule is to test the specimen with a test voltage of the same type that the voltage which will be apply during the final use. However, it is existing a certain among of technical difficulties which force to depart from this common rule DIELECTRIC STRENGTH TESTS WITH AN AC VOLTAGE BENEFITS : The specimen is stressed with the both voltage polarities. The specimen being not charged, there is no need for a discharge system DISADVANTAGES : Most of the tested specimens having a certain amount of capacitance, the HV source have to supply the leakage current as well as the reactive current, this involves an over-sized generator with an increase in prices, in weight and a decreasing of the operator safety who is exposed to higher currents. The reactive current can be evaluated with the following formula : Impedance = Voltage / Current (Ohm s law : Z=U/I) for capacitances : Z = 1 / Cw with w = 2*Pi*F reactive current : Ir = U*C*2*Pi*F i.e. U=3000 volts C=1nF (~ 10 meters of shielded cable) Ir=3000*1E-9*2*3.14*50=0.942 ma Require to adjust the permanent leakage current threshold (IMAX) regarding the capacitance of each specimen. When testing a specimen which will be used with a DC voltage, the AC voltage test can result in a decreasing of the life time because in particular of the heating and the CORONA effect. Under the effect of an electrical field, the orientation of molecules is done with friction which will occur with an AC voltage at each cycle (every 20 or 16 ms). Consequently the AC test is more severe than the DC voltage test. Page : 112 MG SERIE Revision : R

113 DIELECTRIC STRENGTH TESTS WITH A DC VOLTAGE BENEFITS: The power of the HV source can be lower than the one necessary in AC voltage (less weight and more safety for the operator). The current flows through the specimen only during the charging phase. DISADVANTAGES: The charging current can trigger the breakdown detection The specimen being charged, it must be discharged through the built-in discharge resistance (1.5 Mohm). WARNING : Wait enough for the discharge of the specimen capacitance before disconnection from the unit (about 8 seconds per uf). The specimen is tested in only 1 polarity The test voltage must be higher than the one provided with AC test voltage. A common rule is to use a 1.4 correction factor between the DC and the AC voltages (= square root of 2 = ratio between the rms value of a sinusoidal wave and its crest value) : Udc = Uac * BREAKDOWN DETECTION MODE SELECTION The most common and simple leakage current control mode is the threshold current control mode or IMAX mode. This mode allows to set a maximum limit of current flowing through the specimen under test above which the unit detects a breakdown and stops the test by cutting off the HV generation and memorizing the voltage value on the LCD screen. As described in section 3.5.1, the HV source has to supply the leakage current as well as the reactive current coming from its capacitance. Therefore this requires to adjust the breakdown threshold regarding the reactive current of each specimen, and to follow the procedure : Make a test on a good specimen Collect the total current flowing Adjust the current threshold to a value greater than the total current. The MG series units offer the above described detection mode, combined with the -I detection mode (called sometimes ARC detection mode). The -I mode allows to release from the reactive current flowing through the capacitive specimens. To detect a breakdown, the -I mode (ARC detection) monitors only the fast current variations (t>10us and amplitude>1ma). This doesn t require any adjustment regarding the specimen capacitance. However this mode can not detect a dielectric strength default in case of a specimen in short-circuit since the high voltage application. That is why the MG series units allow to combine the IMAX and the -I modes in order to make reliable and without adjustment of dielectric strength tests. The IMAX value is set to a value close to the unit short-circuit current under the test voltage. The MG series units allow to inhibit the detection systems (OFF mode) in order to locate visually where is the dielectric strength fault. WARNING : this mode doesn t cut off the high voltage and therefore it is possible to destroy or burn the specimen under test. The power of the HV source being limited, a continued use of the OFF detection mode can trigger the built-in thermal safety switch. Under those circumstances (display of the message : INTERLOCK DISABLE) wait between 3 and 5 minutes before proceeding to the tests. MG SERIE Revision : R Page : 113

114 9.6 Protective earth continuity measurement On a unit or an electrical device using or generating hazardous voltages, ground continuity measurements insure that all the accessible protective parts are correctly connected to the protective earth connection wire. This measurement is almost a low resistance measurement except that it has to be performed with a high current, mostly in AC. The principle is to flow a current between each metallic accessible parts and the protective earth connection, to measure the voltage drop between the 2 parts and to check with the Ohm s law that the equivalent resistance is lower than the standard required value Current selection The measurement current is defined by the safety standards which have to be used for each product. According to the main standards, this high current is comprised between 10 and 25 A AC or equal to 2 times the nominal operating current of the device. The choice is justified by the fact that the protective earth connections have to be able to flow the fault current for the maximum value of the unit operating current and this during the reaction time of the other protective devices (fuses, breakers, etc...) Voltage selection This is the open voltage of the current generator used for the measurement. This voltage, mostly given by each safety standards is mandatory a low value (lower than the threshold defining an hazardous voltage) but it must allow the measurement current flowing, taking care of the voltage drops between the generator and the measurement points. The open voltages are comprised between 6 and 12 AC volts Test time Unlike a simple resistance measurement, the time that the current is being flowing for the ground continuity measurement is important because of the Safety function of the tested connection. Beyond the ohmic value evaluation, it is mandatory to test the quality of the connection to the earth potential (diameter of the wire, solders quality, screwing quality,...). A manufacturing fault of one of these connections may in certain circumstances give an immediate correct ohmic value, but increasing quickly by heating causes by the high current value used for the test : this high current can produce a break of the wire (wounded wire, which diameter becomes too small). This is why some standards are requiring a minimum test time from 1 to 5 minutes for this measurement. Other standards don t give any time indication, but it is advised to apply for type test a minimum time of 1 minute and for series tests at least 10 seconds. Page : 114 MG SERIE Revision : R

115 9.6.4 Precautions to be observed The measured resistance values being very low (< 1 ohm), it is necessary to perform the measurement in using the 4 wire method to avoid the measurement lead parasitic resistance (interfaces, adapters,...) It is necessary to warrant this measurement principle when connecting the measurement unit to the test points. If specific connections have to be done, it is necessary to use correctly rated cables (at least 5A/mm 2 ) for the current and to connect voltage measurement cables as close as possible to the measured points. During the test time, it is advised not to move the probe to avoid breaking the current flowing which will produce sparks. This can modify the real total test time and can alter the contact surface of the specimen. MG SERIE Revision : R Page : 115

116 SECTION 10: THEORETICAL PRINCIPLES 10.1 UNIT ARCHITECTURE The unit has an architecture organized on a communication bus between the microprocessor board and the test and measurement boards. This communication bus is on a printed circuit called mother board. On the mother board there is also a power supply consisting in a two inputs voltages (115/230V) supplying AC voltages which are used to make the DC voltages for the measurement functions: +5VM : Logic power supply,red Led comes up when voltage on ±25VG : Relays and hipot power assembly supply VEE : LCD display contrast adjustment between -9 and -13volts on the rear panel. A double outputs winding (18VAC1/18VAC2) insulated from the others windings allows to supply the insulation function. The mother board consists also in : A 25 points female sub-d connector with the following pin out : 1 Safety interlock to connect with point 14 2 Red lamp(voltage ON) 3 25VDC protected by an internal fuse 4 I/O for the PLC option (OPT02) = CTRLIN 5 I/O for the PLC option (OPT02) = N1 6 I/O for the PLC option (OPT02) = N3 7 I/O for the PLC option (OPT02) = MES_DCH 8 I/O for the PLC option (OPT02) = COM_OUT 9 I/O for the PLC option (OPT02) = FAIL 10 I/O for the PLC option (OPT02) = PASS volts input (OPT03) 12 Ground 13 Trigger remote control,to connect to Safety interlock to connect to point 1 15 Green lamp(voltage OFF) 16 I/O for the PLC option (OPT02) = COM_IO 17 I/O for the PLC option (OPT02) = N0 18 I/O for the PLC option (OPT02) = N2 19 I/O for the PLC option (OPT02) = TYPE 20 Not used 21 I/O for the PLC option (OPT02) = CTRLOUT 22 I/O for the PLC option (OPT02) = EOT 23 I/O for the PLC option (OPT02) = ERROR 24 Output 0-10volts (OPT03) 25 Trigger remote control to connect to point 13 A 9 points female sub-d connector for the connection of the unit to a computer via a serial RS232C interface (OPT01). Page : 116 MG SERIE Revision : R

117 It is possible to connect the following boards on the mother board : 1 microprocessor board 2 test or measurement boards 1 option board 10.2 MICROPROCESSOR BOARD DESCRIPTION The core of the unit control board is based on a microcontroller from the INTEL s 80C188 family. The selected version, 80C188EB-8Mhz includes the time measurement feature, serial interfaces and address decoders. In combination with the microcontroller, there are the main elements of an embedded system : static RAM (64ko), EPROM (256ko), battery backup RAM (64ko) and programmable logic devices (EPLD) which complete the address decoding. The communication with the different boards is done through a handshake bus called «MG BUS», which consists of 8 data bits, 8 address bits and 6 decoding bits. A set of 8 jumpers allows to dedicate the unit control software. The LCD screen and the 10 keys keyboard are directly connected to the microprocessor board. A serial interface component makes the necessary voltage level shift to comply with the RS232C interface INSULATION RESISTANCE MEASUREMENT FUNCTION The insulation resistance measurement function consists of a DC voltage generator combined with an ammeter. The unit measures the current flowing through the resistance under test and the voltage across the resistance, by using the Ohm s law (U=R*I) the insulation resistance value is displayed. The measurement conditions involving for safety reasons (refer to section A) to earth the positive pole of the generator, the features of the board have to be optically or by a galvanic way insulated from the microprocessor board signals. The insulation resistance measurement board communicates with the microprocessor board through the MG bus. The signals coming from the MG bus interface having to be insulated to control the measurement board functions, a serial to parallel conversion allows to reduce the number of insulation components. A reference voltage is supplied by a 10 bits resolution digital to analogic converter to the voltage generator. This reference voltage controls a pulse width modulation circuit (PWM) which drives by switching a step-up transformer through a power transistors stage. The secondary voltage of the transformer is rectified by a high voltage diode bridge then filtered. A high voltage high impedance resistive divider allows to the PWM circuit to perform the output voltage regulation regarding the load while another divider allows to measure the value of the voltage. This value is multiplexed with the ground, the reference and the current signals before being connected on the input of a voltage to frequency converter working at a 100 khz maximum frequency. This frequency after optical insulation is sent on the MG bus and is measured by the built-in counter of the microcontroller. The whole current scale is made by switching different feedback resistance values on an operational amplifier used as a current to voltage converter. A low pass active filter and an integration network insure the measurement stability on capacitive specimens. MG SERIE Revision : R Page : 117

118 10.4 DIELECTRIC STRENGTH TEST FUNCTION The dielectric strength test function consists of a high voltage AC and DC (option) generator combined with a current measurement. The dielectric strength test board communicates with the microprocessor board through the MG bus. A clock signal coming from the microcontroller built-in frequency divider scans all the locations of a PROM memory which contains the digitalised Sinus function. The output of this memory is connected on a 8 bits resolution digital to analogic converter which gives at its output a microcontroller controlled frequency sinusoidal wave. This wave is modulated in amplitude by a digital to analogic converter used as a multiplier, and then controls a semiconductor power stage which drives the high voltage step-up transformer. A current measurement resistance in the return circuit of the dielectric strength test allows to measure the current flowing through the specimen under test and to control fast current variations. That is why the voltage created by the current across the resistance is differencied by a capacitive network and then amplitude compared by two fast comparators. If one of these comparators detects and if the unit is set in the -I detection mode, the primary control signal of the HV transformer is cut off at the next zero crossing of the sinusoidal wave and the microcontroler is kept informed of the breaking. A high value resistance located across the output terminals allows the high voltage measurement. The signal coming out from the resistance is rectified, filtered then multiplexed with the ground, reference and current signals before being connected on the input of a voltage to frequency converter working at a 100 khz maximum frequency. This frequency is sent on the MG bus and is measured by the built-in counter of the microcontroler Ground continuity resistance measurement function The ground continuity resistance measurement includes an AC current generator combined to a voltage measurement. The ground continuity board communicates with the microprocessor board through the MG bus. A reference voltage is supplied by a 8 bits resolution digital to analog converter (8 bits DAC) to the current amplifier. The current amplifier controls through a switching converter system a transformer with at secondary winding, voltages of 6 or 12 volts and a maximum current of 30A. The reading of the current is done on a 5 mohms shunt resistor in a 4 wire measurement method. The values of the current through the specimen and the voltage across the specimen are multiplexed and sent to the input of an AC to DC converter and then to a voltage to frequency converter. The frequency is sent on the MG bus and measured by the internal timer of the microcontroler. The microcontroler computes the U/I ratio according the Ohm s law to get and display the ground continuity resistance value Page : 118 MG SERIE Revision : R

119 SECTION 11: MAINTENANCE AND CALIBRATION 11.1 PRELIMINARY Our warranty (refer to the beginning of this manual) attests the quality of materials and workmanship in our products. If malfunction should be suspected or other information be desired call our technical assistance: (33) for FRANCE or contact your local distributor INSTRUMENT RETURN Before returning an instrument to our Service Department, please call them at the above phone number for shipment instructions. Use packaging that is adequate to protect it from damage MAINTENANCE Our units don t need particular maintenance except an annual calibration. If problems, please follow the brief check list here after. If the problem continues, call our service department at the above number. LCD SCREEN DOESN T COME UP: Check the correct connection of the main cord SE1 Check that the main voltage is in accordance with the value displayed on the main inlet on the rear panel of the unit (see 1.7.1) Check the fuse in the main inlet on the rear panel (see 1.7.1) DISPLAY OF THE MESSAGE : INTERLOCK OPEN Check that the 25 points connector has been correctly connected on the rear panel Check that the correct connection have been done in the 25 points connector (see 1.7.3) If using an external contact to close the safety interlock,check that the contact works as expected. If the loop is clased, wait 3 to 5 mn so that the thermic security included in the unit allows again voltage supply. The other possibilities for a bad functioning need an intervention inside the unit by qualified people. However we can supply a service manual including schematics of our units. Please get in contact with our Service department in order to know price and delivery time. MG SERIE Revision : R Page : 119

120 11.4 CLEANING Only clean the instrument with a mild rag or slighly soaked with water CALIBRATION We recommend to calibrate our units each year. The calibration must be performed by qualified people having the complete procedure as well as correctly checked standards. Our Maintenance department is at your service to perform the annual calibration. Nevertheless, if you wish to perform yourself the calibration, we can provide a calibration kit including a manual (MG90) and a calibration box (MG91). Please get in contact with our Maintenance department in order to know the price and the delivery time. Page : 120 MG SERIE Revision : R

121 SECTION 12: DRAWINGS 12.1 General Drawing DMG50 MG SERIE Revision : R Page : 121

122 12.1 General Drawing CMG30 Page : 122 MG SERIE Revision : R

123 12.3 General Drawing SMG50 MG SERIE Revision : R Page : 123

124 12.4 Mother board Page : 124 MG SERIE Revision : R

Electrical safety testers

Electrical safety testers Electrical safety testers ISO 9001 Certified Dielectric strength tester from 0 to 5 kvac and 0 to 6 kvdc (SXS56, SXS506 models) Insulation resistance from 50 kω up to 200 GΩ (2 TΩ option) Ground continuity

More information

Dielectric strength tester RMG15AC

Dielectric strength tester RMG15AC Dielectric strength tester RMG15AC The MG series electrical safety testers perform easily and simply all the electrical tests according to the VDE, UL, CSA standards and to the main EN European standards

More information

Portable Appliance Testers. OmegaPAT MI 2140 BetaPAT MI 2141 User Manual Ver Code No

Portable Appliance Testers. OmegaPAT MI 2140 BetaPAT MI 2141 User Manual Ver Code No Portable Appliance Testers OmegaPAT MI 2140 BetaPAT MI 2141 User Manual Ver. 1.2. Code No. 20 750 684 Distributor: Producer: METREL d.d. Ljubljanska 77 SI-1354 Horjul E-mail: metrel@metrel.si http://www.metrel.si

More information

Operating manual M1501 SERIES - English SEFELEC REFERENCE PENT0435

Operating manual M1501 SERIES - English SEFELEC REFERENCE PENT0435 M1501 SERIES - English SEFELEC REFERENCE PENT0435 WARRANTY SEFELEC warrants that units are free from defects in material and workmanship. SEFELEC warrants also that, when properly used, that units will

More information

USER'S MANUAL DMR-6700

USER'S MANUAL DMR-6700 USER'S MANUAL Multimeter True RMS DMR-6700 CIRCUIT-TEST ELECTRONICS www.circuittest.com Introduction This meter measures AC/DC Voltage, AC/DC Current, Resistance, Capacitance, Frequency (electrical & electronic),

More information

Fluke 123/124 Industrial ScopeMeter

Fluke 123/124 Industrial ScopeMeter Fluke 123/124 Industrial ScopeMeter Getting started GB Sep 2002 2002 Fluke Corporation, All rights reserved. Printed in The Netherlands All product names are trademarks of their respective companies. 1

More information

AC/DC Clamp Meter. Owner's Manual. Model No Safety Operation Maintenance Español

AC/DC Clamp Meter. Owner's Manual. Model No Safety Operation Maintenance Español Owner's Manual AC/DC Clamp Meter Model No. 82369 CAUTION: Read, understand and follow Safety Rules and Operating Instructions in this manual before using this product. Safety Operation Maintenance Español

More information

TOS5300 SERIES Hipot Tester/Hipot Tester with Insulation Resistance Test

TOS5300 SERIES Hipot Tester/Hipot Tester with Insulation Resistance Test A new standard for Hipot & Insulation resistance testing Applied to World-Wide input voltage TOS5301 TOS5300 TOS5302 TOS5300(ACW) TOS5301(ACW/DCW) TOS5302(ACW/IR) New low-cost standard model that provides

More information

TECHNICAL DATA SHEET. Safety and function tester GLP1-g. Standard model GLP1-g ELECTRICAL SPECIFICATION

TECHNICAL DATA SHEET. Safety and function tester GLP1-g. Standard model GLP1-g ELECTRICAL SPECIFICATION TECHNICAL DATA SHEET Safety and function tester GLP1-g Revision 2.7 / valid from July 2017 Standard model GLP1-g ELECTRICAL SPECIFICATION Supply voltage Mains frequency No load current consumption 110

More information

User s Manual. MiniTec TM Series. Model MN26 (Model MN26T includes temperature probe) Mini Autoranging MultiMeter

User s Manual. MiniTec TM Series. Model MN26 (Model MN26T includes temperature probe) Mini Autoranging MultiMeter User s Manual MiniTec TM Series Model MN26 (Model MN26T includes temperature probe) Mini Autoranging MultiMeter Introduction Congratulations on your purchase of Extech s MN26 Autoranging Multimeter. This

More information

Breakdown indication by visual signal (BUZ in option) breakdown voltage and current are stored on the LCD display

Breakdown indication by visual signal (BUZ in option) breakdown voltage and current are stored on the LCD display Programmable High Voltage Tester AC dielectric strength tester from 1, 5, 10, 15, 20, 30, 40, 50, 60, 80, 100, 120, 130, 150, 180, 200,... to 400 kvac (maximum power 500 kva) DC dielectric strength (option)

More information

User s Guide. 400A AC/DC Clamp Meter. Model MA220

User s Guide. 400A AC/DC Clamp Meter. Model MA220 User s Guide 400A AC/DC Clamp Meter Model MA220 Introduction Thank you for selecting the Extech MA200 AC/DC Clamp Meter. This meter measures AC/DC Current, AC/DC Voltage, Resistance, Capacitance, Frequency,

More information

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL:

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: 61-920 The Service Information provides the following information: Precautions and safety information Specifications Performance test procedure Calibration

More information

User s Guide. RP7000 Series Active Probe. Dec RIGOL Technologies, Inc.

User s Guide. RP7000 Series Active Probe. Dec RIGOL Technologies, Inc. User s Guide RP7000 Series Active Probe Dec. 2012 RIGOL Technologies, Inc. Guaranty and Declaration Copyright 2011 RIGOL Technologies, Inc. All Rights Reserved. Trademark Information RIGOL is a registered

More information

DIGITAL MULTIMETER OPERATOR'S INSTRUCTION MANUAL HOLD 10A COM LIGHT MS8265 ON/OFF. 200M KHz 2K 20K μ μ μ n.

DIGITAL MULTIMETER OPERATOR'S INSTRUCTION MANUAL HOLD 10A COM LIGHT MS8265 ON/OFF. 200M KHz 2K 20K μ μ μ n. MS8265 DIGITAL MULTIMETER OPERATOR'S INSTRUCTION MANUAL HOLD ON/OFF LIGHT 1000V CAT II 600V CAT III MS8265 200K 2M 20M 20K 200M KHz 2K 20 200 2 20 200μ 200 20μ 750 2μ 1000 200n F 20n 10A 2m 200m 10 10

More information

Testers: / Ohmmeters Insulation Measuring Instruments. n Compact and rugged for service calls and laboratory use. Scope of delivery:

Testers: / Ohmmeters Insulation Measuring Instruments. n Compact and rugged for service calls and laboratory use. Scope of delivery: Table of Contents Testers: / Ohmmeters Insulation Measuring Instruments 51 METRAOHM 413 Article: M630A Digital Low-Resistance Measuring Instrument for Measuring Protective Conductors, Equipotential Bonding

More information

Datasheet Electrical Safety Tester

Datasheet Electrical Safety Tester ENGLISH Datasheet Electrical Safety Tester Stock No. : 9176291 9176294 9176298 9176308 Model : IIT-5000 IIT-5010 IIT-2000 IIT-2010 A Solid Foundation for Reliable, Safe and Practical Safety Compliance

More information

USER'S MANUAL ACDC-100 TRMS ACDC-100. Versatile AC/DC Clamp-on Multimeter Series

USER'S MANUAL ACDC-100 TRMS ACDC-100. Versatile AC/DC Clamp-on Multimeter Series 99 Washington Street Melrose, MA 02176 Fax 781-665-0780 TestEquipmentDepot.com USER'S MANUAL ACDC-100 TRMS ACDC-100 Versatile AC/DC Clamp-on Multimeter Series 1 1) SAFETY This manual contains information

More information

AMM-1022 Digital Multimeter USER`S MANUAL

AMM-1022 Digital Multimeter USER`S MANUAL Digital Multimeter USER`S MANUAL www.tmatlantic.com CONTENTS 1. SAFETY INFORMATION.3 2. DESCRIPTION..6 3. SPECIFICATIONS.8 4. OPERATING INSTRUCTION..11 4.1 Voltage measurement...11 4.2 Current measurement

More information

DM-45 Digital Multimeter

DM-45 Digital Multimeter INSTRUCTION MANUAL DM-45 Digital Multimeter Read and understand all of the instructions and safety information in this manual before operating or servicing this tool. Description The Greenlee DM-45 Digital

More information

Technical Specification

Technical Specification LARGEST RANGE OF IMPULSE TESTERS UP TO 100KV/100KA Technical Specification E-MIG1203CWG revised: 22. June 2003 1 MIG Tester Type MIG1203CWG 1 MIG Tester Type MIG1203CWG 1 1.1 Introduction 1 2 General 2

More information

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL:

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: 61-352 The Service Information provides the following information: Precautions and safety information Specifications Basic maintenance (cleaning, replacing

More information

OPERATION & SERVICE MANUAL FOR FC 110 AC POWER SOURCE

OPERATION & SERVICE MANUAL FOR FC 110 AC POWER SOURCE OPERATION & SERVICE MANUAL FOR FC 100 SERIES AC POWER SOURCE FC 110 AC POWER SOURCE VERSION 1.3, April 2001. copyright reserved. DWG No. FC00001 TABLE OF CONTENTS CHAPTER 1 INTRODUCTION... 1 1.1 GENERAL...

More information

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: MODEL: Multimeter Service Information

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: MODEL: Multimeter Service Information IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: 61-340 MODEL: 61-342 Multimeter Service Information The Service Information provides the following information: Precautions and safety information Specifications

More information

DM-46 Instruction Manual

DM-46 Instruction Manual Auto Meter Products Inc. Test Equipment DM-46 Instruction Manual Automotive Multimeter and Inductive Amp Probe The DM-46 is the auto industry s answer to pocket portability in a 20 2650-1552-00 3/8/11

More information

MS8268 HANDHELD DIGITAL MULTIMETER OPERATOR S INSTRUCTION MANUAL

MS8268 HANDHELD DIGITAL MULTIMETER OPERATOR S INSTRUCTION MANUAL MS8268 HANDHELD DIGITAL MULTIMETER OPERATOR S INSTRUCTION MANUAL Table of Contents TITLE PAGE 1. GENERAL INSTRUCTIONS 1 1.1 Precaution safety measures 1 1.1.1 Preliminary 1 1.1.2 During use 2 1.1.3 Symbols

More information

DVM98. True RMS Digital Multimeter. 1 Safety information. 1.1 Preliminary. 1.2 During use

DVM98. True RMS Digital Multimeter. 1 Safety information. 1.1 Preliminary. 1.2 During use True RMS Digital Multimeter DVM98 1 Safety information This multimeter has been designed according to IEC - 1010 concerning electronic measuring instruments with an overvoltage category (CAT II) and pollution

More information

Insulation Testers and Earth Ground Testers

Insulation Testers and Earth Ground Testers Insulation Testers and Earth Ground Testers With a 5000 V MegOhmMeter for industrial-strength insulation testing and a range of compact hand-held instruments, we offer a solution for every troubleshooting

More information

MW3105 DIGITAL CLAMP MULTIMETER

MW3105 DIGITAL CLAMP MULTIMETER MW3105 DIGITAL CLAMP MULTIMETER 2 M MW3105 A 01 INTRODUCTION 1.1 - Unpacking and inspection Upon removing your new Digital Clamp Meter from its packing, you should have the following items: 1. Digital

More information

OPERATOR S INSTRUCTION MANUAL M-2625 AUTO RANGING DIGITAL MULTIMETER

OPERATOR S INSTRUCTION MANUAL M-2625 AUTO RANGING DIGITAL MULTIMETER OPERATOR S INSTRUCTION MANUAL M-2625 AUTO RANGING DIGITAL MULTIMETER with Temperature Probe Copyright 2007 Elenco Electronics, Inc. Contents 1. Safety Information 3,4 2. Safety Symbols 5 3. Front Plate

More information

PEN TYPE DIGITAL MULTIMETER OPERATION MANUAL T8211D

PEN TYPE DIGITAL MULTIMETER OPERATION MANUAL T8211D PEN TYPE DIGITAL MULTIMETER OPERATION MANUAL T8211D T8211D 1 1. SAFETY INFORMATION BE EXTREMELY CAREFUL IN THE USE OF THIS METER. Improper use of this device can result in electric shock or destroy of

More information

Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.com DIGITAL SUPER MEGOHMMETER DSM-8104, DSM-8542

Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.com DIGITAL SUPER MEGOHMMETER DSM-8104, DSM-8542 Test Equipment Depot - 800.517.8431-99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com DIGITAL SUPER MEGOHMMETER, DIGITAL SUPER MEGOHMMETERS (Single-Channel) (Four-Channel) Options Supporting

More information

True RMS Bench Multimeter 5492B

True RMS Bench Multimeter 5492B Data Sheet True RMS Bench Multimeter 5492B True RMS Bench Multimeter The B&K Precision model 5492B is a versatile 5½-digit, 120,000-count bench multimeter suitable for applications in education, service

More information

User Manual Digital Multimeter

User Manual Digital Multimeter User Manual Digital Multimeter model no.: MSR-R500 Questions or Concerns? support@etekcity.com visit etekcity.com for more products Safe and Proper Usage Thank you for purchasing the Etekcity MSR-R500

More information

Fluke 125. Getting Started. Industrial ScopeMeter

Fluke 125. Getting Started. Industrial ScopeMeter Fluke 125 Industrial ScopeMeter Getting Started GB Dec 2006, Rev. 1, 09/2009 2006, 2009 Fluke Corporation, All rights reserved. Printed in The Netherlands All product names are trademarks of their respective

More information

1. SAFETY 1.1. SAFETY INFORMATION 1.2. SAFETY SYMBOLS

1. SAFETY 1.1. SAFETY INFORMATION 1.2. SAFETY SYMBOLS To all residents of the European Union Important environmental information about this product This symbol on the device or the package indicates that disposal of the device after its lifecycle could harm

More information

MIT415 Insulation tester

MIT415 Insulation tester MIT415 Insulation tester General features o 10V & 100V ESD testing o 25V, 50V, 250V & 500V additional ranges o Insulation testing up to 50Gohms (500V) o Continuity testing at 200mA or 20mA from 0.01ohms

More information

Electrical Safety Tester

Electrical Safety Tester Electrical Safety Tester GPT-9600 Series QUICK START GUIDE GW INSTEK PART NO. 82PT-96030MA1 ISO-9001 CERTIFIED MANUFACTURER This manual contains proprietary information, which is protected by copyright.

More information

DIGITAL MULTIMETER CONTENTS DIGITAL MULTIMETER CONTENTS

DIGITAL MULTIMETER CONTENTS DIGITAL MULTIMETER CONTENTS CONTENTS CONTENTS CONTENTS 1. SAFETY INFORMATION...1 1.1 Preliminary...1 1.2 Dos and don ts...2 1.3 Symbols...3 1.4 Precautions...4 2. DESCRIPTION...5 2.1 Names of parts...6 2.2 Switches, buttons and input

More information

200Amp AC Clamp Meter + NCV Model MA250

200Amp AC Clamp Meter + NCV Model MA250 User's Guide 200Amp AC Clamp Meter + NCV Model MA250 Introduction Congratulations on your purchase of this Extech MA250 Clamp Meter. This meter measures AC Current, AC/DC Voltage, Resistance, Capacitance,

More information

DECLARATION OF CONFORMITY

DECLARATION OF CONFORMITY DECLARATION OF CONFORMITY Manufacturer's Name: Transmille Ltd. Manufacturer's Address: Unit 4, Select Business Centre Lodge Road Staplehurst TN12 0QW. United Kingdom. Declares, that the product Product

More information

USER S MANUAL REVEXplus REVEXplus USB

USER S MANUAL REVEXplus REVEXplus USB USER S MANUAL REVEXplus REVEXplus USB CONTENTS 1. INTRODUCTION...3 1.1. Safety... 3 1.2. Explanation of symbols on instrument... 3 1.3. General description... 4 1.4. Applied standards... 5 1.5. Symbols

More information

MS8211 DIGITAL MULTIMETER INSTRUCTION MANUAL

MS8211 DIGITAL MULTIMETER INSTRUCTION MANUAL MS8211 DIGITAL MULTIMETER INSTRUCTION MANUAL Ω CONTENTS CONTENTS 1. SAFETY INFORM...1 4.4 Range Transform...10 1.1 Preliminary...1 4.5 Auto Power Off...10 1.2 During use...2 4.6 Preparation For Measurement...11

More information

Technical Specification

Technical Specification LARGEST RANGE OF IMPULSE TESTERS UP TO 100KV/100KA Technical Specification No. E-MIG0603IN2.doc revised: 20.October 2004 1 MIG Tester Type MIG0603IN2 1 MIG Tester Type MIG0603IN2 1 1.1 Introduction 1 2

More information

Dawson DDM190. Digital Multimeter User s Manual

Dawson DDM190. Digital Multimeter User s Manual Dawson DDM190 Digital Multimeter User s Manual TABLE OF CONTENTS LIMITED WARRANTY AND LIMITATION OF LIABILITY... 3 Out of the Box... 3 Accessories.. Error! Bookmark not defined. Safety Information... 7

More information

MODEL 9041 EXTENDED SPECIFICATIONS. 25ppm TRANSPORTABLE CALIBRATOR

MODEL 9041 EXTENDED SPECIFICATIONS. 25ppm TRANSPORTABLE CALIBRATOR MODEL 9041 EXTENDED SPECIFICATIONS 25ppm TRANSPORTABLE CALIBRATOR 9041 EXTENDED SPECIFICATIONS General Specifications TRANSMILLE LTD Warm Up Time Double the time since last used up to 20 minutes maximum

More information

Model: Pro95 TRUE RMS MILLIAMP CLAMP METER

Model: Pro95 TRUE RMS MILLIAMP CLAMP METER Model: Pro95 TRUE RMS MILLIAMP CLAMP METER TABLE OF CONTENTS 1. SAFETY INFORMATION... 1 2. GENERAL SPECIFICATION... 1 3. ELECTRICAL SPECIFICATION... 2 3-1 ACMA MEASUREMENT... 2 3-2 ACA MEASUREMENT...

More information

INSTRUCTION MANUAL. Model Autoranging DMM ProbeMeter TM. Measures voltage, resistance, frequency, capacitance, temperature, and duty cycle.

INSTRUCTION MANUAL. Model Autoranging DMM ProbeMeter TM. Measures voltage, resistance, frequency, capacitance, temperature, and duty cycle. INSTRUCTION MANUAL Model 403380 Autoranging DMM ProbeMeter TM Measures voltage, resistance, frequency, capacitance, temperature, and duty cycle. Back lit LCD with Autorange and full function displays Audible

More information

Autoranging Multimeter Extech EX503

Autoranging Multimeter Extech EX503 User's Guide Autoranging Multimeter Extech EX503 Introduction Congratulations on your purchase of the Extech EX503 Autoranging Multimeter. This meter measures AC/DC Voltage, AC/DC Current, Resistance,

More information

Digital insulation tester

Digital insulation tester Digital insulation tester C.A 6505 > > Insulation at 5 kv > > Wide measurement range from 10 kω to 10 TΩ > > Large backlit LCD screen, with digital display and bargraph > > Fixed test voltages and programmable

More information

OPERATOR S INSTRUCTION MANUAL

OPERATOR S INSTRUCTION MANUAL OPERATOR S INSTRUCTION MANUAL AUTO-RANGE DUAL DISPLAY CONFORMED IEC1010 DIGITAL MULTIMETER CONTENTS PAGE SAFETY INFORMATION..... DESCRIPTION.. OPERATING INSTRUCTION.. SPECIFICATIONS.... ACCESSORIES. BATTERY

More information

MODEL 9050 EXTENDED SPECIFICATIONS. 50ppm TRANSPORTABLE CALIBRATOR

MODEL 9050 EXTENDED SPECIFICATIONS. 50ppm TRANSPORTABLE CALIBRATOR MODEL 9050 EXTENDED SPECIFICATIONS Ü 50ppm TRANSPORTABLE CALIBRATOR 9050A EXTENDED SPECIFICATIONS General Specifications TRANSMILLE LTD Warm Up Time Double the time since last used up to 20 minutes maximum

More information

DM-46 Instruction Manual

DM-46 Instruction Manual Test Equipment Auto Meter Products Inc. 413 West Elm Street Sycamore, IL 60178 Service (815) 899-0801 Toll Free (866) 883-TEST (8378) www.autometer.com/test DM-46 Instruction Manual Automotive Multimeter

More information

USER'S MANUAL. ACD-6 Pro. ACD-6 TRMS Pro. Versatile Clamp-on Multimeter Series

USER'S MANUAL. ACD-6 Pro. ACD-6 TRMS Pro. Versatile Clamp-on Multimeter Series USER'S MANUAL ACD-6 Pro ACD-6 TRMS Pro Versatile Clamp-on Multimeter Series 1 1) SAFETY This manual contains information and warnings that must be followed for operating the instrument safely and maintaining

More information

MD-5060 DIGITAL MEGOHMMETER

MD-5060 DIGITAL MEGOHMMETER DIGITAL MEGOHMMETER ( M98126501-03 / 04A ) TECHNICAL SPECIFICATIONS OPERATING INSTRUCTIONS DESCRIPTION The CIRCUTOR model MD-5060 digital electronic ohmmeter is a versatile, robust and easy-to-use equipment.

More information

HANDHELD DIGITAL MULTIMETER OPERATOR S INSTRUCTION MANUAL

HANDHELD DIGITAL MULTIMETER OPERATOR S INSTRUCTION MANUAL HANDHELD DIGITAL MULTIMETER OPERATOR S INSTRUCTION MANUAL GENERAL INSTRUCTIONS This instrument complies with IEC 1010-1 (61010-1@IEC: 2001), CAT. II 1000V and CAT. III 600V overvoltage standards. See Specifications.

More information

EA3012 TRANSCONDUCTANCE AMPLIFIER OPERATION MANUAL

EA3012 TRANSCONDUCTANCE AMPLIFIER OPERATION MANUAL EA3012 TRANSCONDUCTANCE AMPLIFIER Ü EA3012 Transconductance Amplifier Operation Manual Version 2.00 Dec 2011 All product names are trademarks of their respective companies Guarantee and service Transmille

More information

R/C SMD Tweezers Multimeter Model RC200

R/C SMD Tweezers Multimeter Model RC200 User's Guide R/C SMD Tweezers Multimeter Model RC200 Introduction Congratulations on your purchase of the Extech RC200 RC Tweezers Multimeter This meter, using the tweezers adaptor, measures SMD capacitors,

More information

1507/1503. Insulation Testers. Users Manual

1507/1503. Insulation Testers. Users Manual 1507/1503 Insulation Testers 1507/1503 Insulation Testers Introduction The Fluke model 1507 and model 1503 are battery-powered insulation testers (hereafter, "the Tester"). Although this manual describes

More information

TM1600. Circuit Breaker Analyzer System

TM1600. Circuit Breaker Analyzer System Circuit Breaker Analyzer System Circuit breaker analyzer The breaker analyzer measures a circuit breaker s timing cycle. The timing channels record closings and openings of main contacts, resistor contacts

More information

C.A 6115N. Safety at all costs. IEE 16th: Electrical installation IEC/EN I E C. Insulation RCD. Earth. Loop. Continuity.

C.A 6115N. Safety at all costs. IEE 16th: Electrical installation IEC/EN I E C. Insulation RCD. Earth. Loop. Continuity. IEE 16th: Safety at all costs C.A 6115N Electrical installation Insulation RCD Earth Loop Continuity Phase rotation I E C CONFIRMS TO CONFIRMS TO IEC/EN 61557 1010-1 III 300V Cat Cat The most complete

More information

8000 SERIES PRECISION MULTIMETER VERIFICATION AND ADJUSTMENT GUIDE

8000 SERIES PRECISION MULTIMETER VERIFICATION AND ADJUSTMENT GUIDE 8000 SERIES PRECISION MULTIMETER VERIFICATION AND ADJUSTMENT GUIDE TRANSMILLE LTD. Version 1.1 : Apr 2015 TABLE OF CONTENTS PREPARING FOR CALIBRATION... 4 INTRODUCTION... 4 CALIBRATION INTERVAL SELECTION...

More information

Safety. This symbol, adjacent to a terminal, indicates that, under normal use, hazardous voltages may be present.

Safety. This symbol, adjacent to a terminal, indicates that, under normal use, hazardous voltages may be present. 9305 Safety International Safety Symbols This symbol, adjacent to another symbol or terminal, indicates the user must refer to the manual for further information. This symbol, adjacent to a terminal, indicates

More information

Contents 1. General instructions. 1.1 Precautions safety measures Protection mechanisms. 2. Description. 2.1 Instrument Familiarization. 2.

Contents 1. General instructions. 1.1 Precautions safety measures Protection mechanisms. 2. Description. 2.1 Instrument Familiarization. 2. Contents 1. General instructions. 1.1 Precautions safety measures... 1.2 Protection mechanisms. 2. Description. 2.1 Instrument Familiarization. 2.2 LCD Display 2.3 Keypad. 3. Function description. 3.1

More information

GE Infrastructure Sensing. Druck DPI 841/842. Frequency calibrator and Frequency loop calibrator User manual - K395

GE Infrastructure Sensing. Druck DPI 841/842. Frequency calibrator and Frequency loop calibrator User manual - K395 GE Infrastructure Sensing Druck DPI 841/842 Frequency calibrator and Frequency loop calibrator User manual - K395 A1 B1 10 1 A 2 9 A 3 8 2 3 7 6 11 4 5 B1 12 A2 DPI 842 13 15 14 A3 19 18 17 16 27 20 21

More information

MS8250A/B OPERATION MANUAL MS8250A. Hz% FUNC REL RANGE REL HOLD OFF 10A. Hz% A NCV. Hz% COM. A ma 10A FUSED 600V CAT IV.

MS8250A/B OPERATION MANUAL MS8250A. Hz% FUNC REL RANGE REL HOLD OFF 10A. Hz% A NCV. Hz% COM. A ma 10A FUSED 600V CAT IV. MS8250A/B DIGITAL MULTIMETER OPERATION MANUAL AUTO DC AC REL hfe PCLINK % C F kmωkz nµmfav MS8250A DIGITAL MULTIMETER Auto Power Off RANGE REL HOLD FUNC NCV A ma OFF 10A A ma 10A FUSED 600V CAT IV COM

More information

Pen Multimeter. Model

Pen Multimeter. Model Pen Multimeter Model 381626 CAUTION: Read, understand and follow all Safety Rules and Operating Instructions in this manual before using this product. This instrument is a 3200 count pen style digital

More information

RIGOL. User s Guide. RP5600 Passive Probe. July 2010 RIGOL Technologies, Inc.

RIGOL. User s Guide. RP5600 Passive Probe. July 2010 RIGOL Technologies, Inc. User s Guide RP5600 Passive Probe July 2010 RIGOL Technologies, Inc. Guaranty and Declaration Copyright 2010 RIGOL Technologies, Inc. All Rights Reserved. Trademark Information RIGOL is a registered trademark

More information

multi-function + auto-range, CATIII 1000V ACV/DCV/OHMS/4 to 4000 M ohm, 100V, 250V, 500V, 1000V INSULATION TESTER

multi-function + auto-range, CATIII 1000V ACV/DCV/OHMS/4 to 4000 M ohm, 100V, 250V, 500V, 1000V INSULATION TESTER multi-function + auto-range, CATIII 1000V ACV/DCV/OHMS/4 to 4000 M ohm, 100V, 250V, 500V, 1000V INSULATION TESTER Model : DI-6400 Your purchase of this INSULATION TESTER marks a step forward for you into

More information

Battery Analyzers BA6010 Series

Battery Analyzers BA6010 Series Data Sheet Battery Analyzers RS232 The battery analyzers measure voltage and resistance of modern battery technologies with high accuracy, resolution, and speed. Additionally, these instruments provide

More information

TRANSMILLE 3010A PRECISION MULTIPRODUCT CALIBRATOR EXTENDED SPECIFICATIONS

TRANSMILLE 3010A PRECISION MULTIPRODUCT CALIBRATOR EXTENDED SPECIFICATIONS TRANSMILLE 3010A PRECISION MULTIPRODUCT CALIBRATOR EXTENDED SPECIFICATIONS www. transmille.com 3010A EXTENDED SPECIFICATIONS General Specifications TRANSMILLE LTD Warm Up Time Double the time since last

More information

INSTRUCTION MANUAL ACD-10 PRO & TRMS PRO

INSTRUCTION MANUAL ACD-10 PRO & TRMS PRO INSTRUCTION MANUAL ACD-10 PRO & TRMS PRO AMPROBE 1 Congratulations! You are now the owner of an AMPROBE Instrument. It has been quality crafted according to quality standards and contains quality components

More information

DC Resistance Meters 2840 Series

DC Resistance Meters 2840 Series Data Sheet DC Meters Touch screen to zoom, select, and enter values The DC resistance meters feature high accuracy and resolution measurements in the milliohm range. Both meters are ideally suited for

More information

N. TEST TEST DESCRIPTION

N. TEST TEST DESCRIPTION Multi function system for testing substation equipment such as: current, voltage and power transformers, all type of protection relays, energy meters and transducers Primary injection testing capabilities

More information

MIT400/2. CAT IV Insulation testers MIT430/2: CAT IV Insulation testers

MIT400/2. CAT IV Insulation testers MIT430/2: CAT IV Insulation testers MIT400/2 Designed for electrical and industrial testing Insulation testing up to 1000 V and 200 GΩ range Stabilized insulation test voltage (New) Single range, faster continuity testing from 0.01 Ω to

More information

CD770 DIGITAL MULTIMETER INSTRUCTION MANUAL

CD770 DIGITAL MULTIMETER INSTRUCTION MANUAL CD770 DIGITAL MULTIMETER INSTRUCTION MANUAL Table of Contents 1 SAFETY PRECAUTIONS Before use, read the following safety precautions.- 1-1 Explanation of Warning Symbols 001 1-2 Warning Messages for Safe

More information

GFL-1000 User Manual Ground Fault Locator

GFL-1000 User Manual Ground Fault Locator GFL-Series User Manual V1.1 GFL-1000 User Manual Ground Fault Locator Contents Contents... 1 1 Declaration of Conformity... 3 2 Introduction... 3 3 Equipment Information... 3 3.1 Safety Precautions...

More information

1.General instructions Specifications Description...7

1.General instructions Specifications Description...7 USER S Manual CONTENTS 1.General instructions...1 1.1 Precautions safety measures...1 1.1.1 Preliminary...1 1.1.2 During use...2 1.1.3 Symbols...4 1.1.4 Instructions...5 1.2 Protection mechanisms...6 2.

More information

CM605. User Manual AC/DC LOW CURRENT CLAMP-ON METER ENGLISH

CM605. User Manual AC/DC LOW CURRENT CLAMP-ON METER ENGLISH AC/DC LOW CURRENT CLAMP-ON METER CM605 ENGLISH User Manual Statement of Compliance Chauvin Arnoux, Inc. d.b.a. AEMC Instruments certifies that this instrument has been calibrated using standards and instruments

More information

User's Guide Mini Multimeter with Non-Contact Voltage Detector (NCV) Model EX330

User's Guide Mini Multimeter with Non-Contact Voltage Detector (NCV) Model EX330 User's Guide Mini Multimeter with Non-Contact Voltage Detector (NCV) Model EX330 Introduction Congratulations on your purchase of the Extech EX330 Meter. The EX330 offers AC/DC Voltage, AC/DC Current,

More information

EurotestCOMBO MI 3125 / MI 3125E Short instructions Version 2.2, Code no

EurotestCOMBO MI 3125 / MI 3125E Short instructions Version 2.2, Code no EurotestCOMBO MI 3125 / MI 3125E Short instructions Version 2.2, Code no. 20 751 519 Distributor: Manufacturer: METREL d.d. Ljubljanska cesta 77 1354 Horjul Slovenia web site: http://www.metrel.si e-mail:

More information

1. SAFETY INFORMATION.1 2. DESCRIPTION SPECIFICATIONS.6 4. OPERATING INSTRUCTION Voltage measurement Current measurement 10

1. SAFETY INFORMATION.1 2. DESCRIPTION SPECIFICATIONS.6 4. OPERATING INSTRUCTION Voltage measurement Current measurement 10 CONTENTS 1. SAFETY INFORMATION.1 2. DESCRIPTION..4 3. SPECIFICATIONS.6 4. OPERATING INSTRUCTION..9 4.1 Voltage measurement...10 4.2 Current measurement 10 4.3 Resistance measurement...12 4.4 Diode test.12

More information

DIGITAL ANALOG MEGOHMMETERS

DIGITAL ANALOG MEGOHMMETERS DIGITAL ANALOG MEGOHMMETERS 1035 1039 E N G L I S H User Manual Statement of Compliance Chauvin Arnoux, Inc. d.b.a. AEMC Instruments certifies that this instrument has been calibrated using standards and

More information

LEAK CURRENT HiTESTER ST5540/ST Washington Street Melrose, MA Phone Toll Free

LEAK CURRENT HiTESTER ST5540/ST Washington Street Melrose, MA Phone Toll Free LEAK CURRENT HiTESTER ST5540/ST5541 Safety measuring instruments 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Leak Current Measurement

More information

User s Guide. Model MA A AC Mini Clamp-on Meter

User s Guide. Model MA A AC Mini Clamp-on Meter User s Guide Model MA150 200A AC Mini Clamp-on Meter Introduction Congratulations on your purchase of Extech s MA150 AC Mini Clamp Meter. This meter is shipped fully tested and calibrated and, with proper

More information

External RF Driver. Electrical Driver for AOTF-crystal. Instruction Manual

External RF Driver. Electrical Driver for AOTF-crystal. Instruction Manual External RF Driver Electrical Driver for AOTF-crystal Instruction Manual Table of Contents 1 General... 3 2 Compliance... 3 3 Labeling... 4 4 Interface... 5 4.1 Front Panel... 5 4.2 Back Panel... 6 4.2.1

More information

INDEX. Micro-ohmmeter MH-10 Page 3. MI (5, 10, 15, 20 kv) Page 4 Mega-ohmmeter MD (5 kv) Page 4

INDEX. Micro-ohmmeter MH-10 Page 3. MI (5, 10, 15, 20 kv) Page 4 Mega-ohmmeter MD (5 kv) Page 4 Micro-ohmmeter MH-10 Page 3 MI (5, 10, 15, 20 kv) Page 4 Mega-ohmmeter MD (5 kv) Page 4 Earth resistance meter TL-5 Page 5 Indirect earth contact simulator MPC (5, 20, 50 A) Page 6 Trip time tester CR

More information

Multifunction AC/DC Voltage Tester

Multifunction AC/DC Voltage Tester User Manual Multifunction AC/DC Voltage Tester Model 20250-59 THE STANDARD IN PRECISION MEASUREMENT 1065DGMAN_20250-59 DS multifunctional voltage tester.indd 1 8/8/2017 1:37:47 PM Introduction The Digi-Sense

More information

# Digital Multimeter

# Digital Multimeter #61-797 Digital Multimeter - 1 - TABLE OF CONTENTS 1.Introduction Precautions and Safety Information Symbols Safety 2.Specifications General Specifications Electrical Specifications Required Equipment

More information

Agilent E3633A and E3634A DC Power Supplies

Agilent E3633A and E3634A DC Power Supplies Service Guide Part Number: E3634-90010 October 2007. For Warranty information, refer to the back of the manual. Copyright Agilent Technologies, Inc. 1998 2007 All Rights Reserved. Agilent E3633A and E3634A

More information

AX-C Introduction. 2. Safety Information

AX-C Introduction. 2. Safety Information AX-C708 1. Introduction Read Safety Information before using the meter. ProcessMeter ( referred to as the meter )is a handheld, battery-operated tool for measuring electrical parameters. It has all the

More information

HP 34401A Specifications 8

HP 34401A Specifications 8 8 HP 34401A Specifications 8 DC Characteristics DC Characteristics Accuracy Specifications ± ( % of reading + % of range ) [ 1 ] Function Range [ 3 ] Test Current or Burden Voltage 24 Hour [ 2 ] 23 C ±

More information

MIT400/2 CAT IV Insulation testers

MIT400/2 CAT IV Insulation testers MIT400/2 CAT IV Insulation testers MIT400/2 CAT IV Insulation testers Designed for Electrical and Industrial testing Insulation testing up to 1000 V and 200 GΩ range Stabilised insulation test voltage

More information

INSULATION TESTER SERIES

INSULATION TESTER SERIES INSULATION TESTER SERIES Field measuring instruments A Full Line-up of Insulation Testers to Suit Any Need From Basic Testing to High Performance Analysis 2 INSULATION TESTER SERIES Selection guide IR4056-20

More information

Harris IRT Enterprises Multi-Channel Digital Resistance Tester Model XR

Harris IRT Enterprises Multi-Channel Digital Resistance Tester Model XR Harris IRT Enterprises Multi-Channel Digital Resistance Tester Model 6012-06XR Specifications & Dimensions 2 Theory of Operation 3 System Block Diagram 4 Operator Controls & Connectors 5 Test Connections

More information

1. General Instructions 2 2. Safety 2 3. Lamp Starting Test Instrument LSTI 5 3

1. General Instructions 2 2. Safety 2 3. Lamp Starting Test Instrument LSTI 5 3 1. General Instructions 2 2. Safety 2 3. Lamp Starting Test Instrument LSTI 5 3 3.1. Components and Connections of the Front Panel (Fig. 1) 5 3.2. Connection of the Rear Panel (Fig. 2) 7 3.3. Operation

More information

The future of installation testing is here. The Fluke 1650 Series Multifunction Testers

The future of installation testing is here. The Fluke 1650 Series Multifunction Testers The future of installation testing is here The Fluke 50 Series Multifunction Testers Fluke 50 Series Safer, easier installation testing. The 50 Series testers verify the safety of electrical installations

More information

ProfiScale MULTI Multimeter

ProfiScale MULTI Multimeter 1,5 V 9V 200 mv 600 V 200 ma 1/10 A ProfiScale MULTI Multimeter en Operating instructions BURG-WÄCHTER KG Altenhofer Weg 15 58300 Wetter Germany Introduction Want the reassurance of knowing whether current

More information

EX350 Series USER GUIDE. True RMS Digital Multimeters. EX350 True RMS Digital Multimeter EX355 True RMS Digital Multimeter with Temperature

EX350 Series USER GUIDE. True RMS Digital Multimeters. EX350 True RMS Digital Multimeter EX355 True RMS Digital Multimeter with Temperature USER GUIDE True RMS Digital Multimeters EX350 Series EX350 True RMS Digital Multimeter EX355 True RMS Digital Multimeter with Temperature Table of Contents 1. INTRODUCTION 3 2. SAFETY INFORMATION 4 3.

More information

WRM-10 TM TRANSFORMER WINDING RESISTANCE METER

WRM-10 TM TRANSFORMER WINDING RESISTANCE METER WRM-10 TM TRANSFORMER WINDING RESISTANCE METER USER S MANUAL Vanguard Instruments Company, Inc. 1520 S. Hellman Ave. Ontario, California 91761, USA TEL: (909) 923-9390 FAX: (909) 923-9391 June 2009 Revision

More information