Instruction Manual. (English-English)

Size: px
Start display at page:

Download "Instruction Manual. (English-English)"

Transcription

1 Instruction Manual (English-English)

2 Copyright Information 2014 Cynosure, Inc. Cynosure, Ellman and RadioLase are registered trademarks of Cynosure, Inc. Caution Federal law restricts this device to sale by or on the order of a Physician

3 CONTENTS I. INTRODUCTION Page 4 II. SAFETY Page 4 III. TECHNICAL INFORMATION Page 7 IV. DESCRIPTION OF CONTROL ELEMENTS Page 7 V. PREPARATION FOR USE Page 12 VI. SURGICAL DESCRIPTION AND USE Page 12 VII. POWER OUTPUT CHARACTERISTICS Page 14 VIII. TECHNICAL SPECIFICATIONS Page 17

4 I. INTRODUCTION 1.0 INTRODUCTION TO RADIOSURGERY Radiosurgery with the Radiolase 3 High Frequency Radio surgical Generator (Generator) is an excellent method of cutting and coagulating soft tissue. With four waveforms (3 monopolar and 1 bipolar) and a wide selection of electrode tips, the user can easily configure the system for a wide array of applications. 1.1 INTENDED USE The intended use for the Radiolase 3 Generator is: Monopolar Cutting: Skin and Mucosal Incisions, Biopsy, Cysts, Abscesses, Tumors, Cosmetic Repairs, Development of Skin Flaps, Skin Tags, Nevi, Keratosis, Oculoplastic Procedures, Blepharoplasty, Aponeurotic Repair, Levator Resection Monopolar Blended Cutting & Coagulation: Skin Tags, Papilloma, Keloids, Keratosis, Verrucae, Basal Cell Carcinoma, Nevi, Fistulas, Epithelioma, Cosmetic Repair, Cysts, Abscesses, Development of Skin Flaps, Oculoplastic Procedures Monopolar Hemostasis: Control of Bleeding, Epilation, Telangiectasia Bipolar Hemostasis: Pinpoint Hemostasis in any field (Wet or Dry). Note: The Radiolase 3 Generator is not indicated for non-ablative treatment of wrinkles. Furthermore, the Radiolase 3 Generator is neither designed to provide the required power levels, nor withstand the demands of lengthy non-ablative wrinkle treatment procedures. The Radiolase 3 Generator will overheat and possibly fail if used for non-ablative treatment of wrinkles. Such failures will not be covered by warranty. II. SAFETY Improper use of this, or any high frequency surgical instrument, can result in a patient injury. Read and become familiar with all instructions and directions before using this product. 2.0 WARNINGS High frequency surgical equipment should be used only by persons having adequate training and familiarity with high frequency surgical equipment and related techniques. A thorough understanding of the use of high frequency surgical equipment is essential to avoid shock and burn hazards to both patient and medical personnel and damage to the device or other medical equipment. Consult medical literature relative to related techniques, complications, and hazards prior to use of high frequency surgical equipment Never increase the power settings without first checking both the active electrode and the Neutral Pad Non-conductive Corneal Eye Shields (Plastic) must be used for any radiofrequency surgical procedure involving the eyelid and the immediate surrounding areas Use only with Ellman accessories including Ellman handpieces, Neutral Pads, cables, and electrodes. 4

5 2.1 PRECAUTIONS General Interference caused by the Generator could adversely affect the operation of other electronic equipment Electrical shock hazard. Do not remove cover. Refer to authorized personnel for service Generator may present a hazard to patients with pacemakers. Consult qualified medical personnel If higher than normal power settings are required, examine all cables, handpieces and the Neutral Pad for damage and confirm that cables are not wrapped around metal objects or touching the floor. In the event that higher power is still required, replace the Neutral Pads and handpieces Do not cut or modify any cables The patient should not come into contact with metal parts which are earthed or which have appreciable capacitance to earth. The use of antistatic sheeting is recommended for this purpose Physician to patient skin to skin contact should be avoided. Use of insulating gloves is recommended When the Generator is used simultaneously with physiological monitoring equipment on the same patient, the monitoring electrodes should be placed as far as possible from the surgical electrodes. Needle monitoring electrodes are not recommended. Monitoring systems incorporating high-frequency current-limiting devices are recommended The cables to the surgical electrodes should be positioned in such a way that contact with the patient or other leads is avoided. Temporarily unused active electrodes should be stored segregated from the patient The output power selected should be as low as possible for the intended purpose Apparent low output or failure of the surgical equipment to function correctly at the normal operating settings may indicate faulty application of the Neutral Pad or poor contact in its connection Before each use inspect the accessories, electrode cables and probes for possible physical damage Do not obstruct air circulation vents For surgical procedures where the high frequency current could flow through parts of the body having a relatively small cross-sectional area, the use of bipolar techniques may be desirable to avoid unwanted coagulation Similar to all energy sources, by-products such as tissue smoke plume and aerosols, create a health concern about their carcinogenic and infectious potential. Appropriate protective measures including effective smoke evacuation should always be utilized Neutral Pad The use and proper placement of the Neutral Pad is essential to the safe and effective use of this device in order to prevent burns from the Neutral Pad. The Neutral Pad should be reliably attached with its entire area attached to the patient s body as close as possible to the operating field. Direct skin contact is necessary for a conductive Neutral Pad. Direct skin contact is not necessary for a capacitive Neutral Pad The Neutral Pad should not be placed on the patients head or other areas where it does not make complete skin contact If patient is repositioned during treatment, check and verify the Neutral Pad is still making complete contact to the patient. Check the Neutral Pad for signs of heating during procedures. 5

6 2.1.3 Flammable Environment Some materials, for example cotton, wool, and gauze when saturated with oxygen may be ignited by sparks produced in normal use of the equipment Do not use the Generator in the presence of flammable anesthetics or other flammable gases, flammable liquids, or flammable objects Do not use this device in oxygen enriched atmospheres or in the presence of nitrous oxide (N20) Use only non-flammable agents for cleaning Flammable agents used as solvents for removing adhesives should be allowed to evaporate before powering on the Generator. Any flammable fluid pooled under the patient or in the body depressions should be mopped up prior to using the Generator. Attention should be called to the danger of ignition of endogenous gases. NOTE: To ensure safe and reliable operation, the Radiolase 3 Generator unit and relevant accessories should be checked for wear and tear before operation. A maintenance contract with the manufacturer or one of its authorized service agents is recommended after the warranty period. 3.0 SUPPLIER S INFORMATION Cynosure, Inc. dba Ellman 400 Karin Lane Hicksville, New York U.S.A. (516) (800) III. TECHNICAL INFORMATION Supplier will make the following available upon request; circuit diagrams, component parts list, description, calibration instructions or other information necessary to assist technical personnel. ATTENTION: Alcohol must never be used to clean and disinfect the front panel. 3.2 CLEANING, DISINFECTING AND STERILIZING THE UNIT AND ACCESSORIES Use only nonflammable and nonexplosive agents. No moisture or detergents should enter the unit Handpiece and Electrodes. Always clean and disinfect the electrodes prior to sterilization. Disinfect and sterilize handpieces and electrodes according to the directions supplied with them. All autoclavable products should be routinely inspected for any wear, breaks, or deterioration of the insulation, and should be discarded and replaced if any of these occurrences exist The footswitch is a waterproof unit; it can be cleaned with regular detergent. 6

7 4.0 INTERFACE ACCESORIES IV. DESCRIPTION OF CONTROL ELEMENTS Two Pedal Footswitch Footswitch for CUT (yellow) and COAG (blue) modes is supplied. The footswitch may be used with any handpiece compatible with the RF Generator Handpieces A variety of handpieces are available. When using a fingerswitch handpiece, the depressed button will activate monopolar energy in the mode and power level selected on the front panel IEC Neutral Pad The IEC Neutral Pad must be used with the RF Generator during activation. To use, insert the molded plug of the Neutral Pad into the RF Generator as shown in Figure Power Cord The power cord connects RF Generator to power supply. 7

8 4.1 REAR PANEL CONTROL ELEMENTS Figure 4.1 Rear Panel 3 2 Cynosure, Inc. dba Ellman HF RADIO SURGICAL GENERATOR MODEL P/N Operating Frequency 4 MHz POWER OUTPUT MONOPOLAR BIPOLAR 500 OHM LOAD 200 OHM LOAD CUT 50 WATTS CUT/COAG 50 WATTS HEMO 35 WATTS 35 WATTS RATED DUTY CYCLE 20 SEC ON, 30 SEC OFF SUPPLY RATINGS at 50-60Hz VAC 0.80A 220 VAC 0.90A 120 VAC 1.40A 100 VAC 1.70A FUSE (2) VAC : T1.25A, 250V 100/120 VAC : T2.0A, 250V POWER ENTRY VOLUME FOOT SWITCH Power Cord 2 Power Socket 3 Line Fuses 4 Audio Volume Control 5 Footswitch Receptacle 6 7 Footswitch Pedal Speaker 8

9 4.2 FRONT PANEL CONTROL ELEMENTS Figure 4.2 Front Panel 6a 6 4a 9 5a a 7 8a 8 92 NEUTRAL FINGERSWITCH/HANDPIECE BIPOLAR HANDPIECE! ON / OFF NEUTRAL PAD MONOPOLAR FINGERSWITCH HANDPIECE BIPOLAR HANDPIECE OR CABLE 1 Power Switch: 1 = ON 0 = OFF 9

10 2 Power On Indicator: 3 Neutral Pad Receptacle: 4 Bipolar Cord Receptacle: 5 Fingerswitch/Handpiece Receptacle: 6 Yellow Light (Monopolar Cut Energized) Indicator: The indicator illuminates the Monopolar when Cut function is selected and has been selected and energized. 7 Yellow Light (Monopolar Cut/Coag Energized) Indicator: The indicator illuminates when the Monopolar Cut/Coag function is selected and has been energized. 8 Blue Light (Monopolar Hemo Energized) Indicator: The indicator illuminates when Monopolar Hemo function is selected and has been energized. 9 Monopolar Mode Select Knob 10 Monopolar Power Intensity Setting Knob: The selected output power will be delivered to the patient for true output power. See figures in section Blue Light (Bipolar Hemo Energized) Indicator: The indicator illuminates when Bipolar Hemo function is selected and has been energized. 12 Bipolar Power Intensity Setting Knob: The selected power will be delivered to the patient for true output power. See figures in section SYMBOL DEFINITIONS 10

11 4.3.1 Type BF defibrillator protection! Refer to manual Protective earth connection Neutral electrode floating at high frequency Fingerswitch control Footswitch control Neutral Pad Non-Ionization radiation Equipotential Symbol 11

12 V. PREPARATION FOR USE 5.0 PREPARATION FOR USE On the Generator check to make sure that the power switch is in the OFF position Connect the power cord to a 3-wire ground AC power receptacle Insert the Neutral Pad plug into the receptacle on the front panel marked Neutral The Neutral Pad should be reliably attached as per Section Plug the handpiece into the appropriate receptacle on front panel Make sure that the electrode is seated fully so that no conductive shaft is exposed. 5.1 WAVEFORM SELECTION Reference Section 7 Power output characteristics for more details about waveforms CUT - For micro-smooth cutting CUT/COAG - For cutting with coagulation HEMO - For pure Hemostasis-Coagulation to control most forms of bleeding. 5.2 SMOKE HAZARD By-products of high frequency surgical procedures are an occupational exposure hazard. Standards based policies and procedures must be developed in all practice settings to protect personnel. VI. SURGICAL DESCRIPTION AND USE 6.0 LEARNING TO USE RADIOSURGERY Before making contact with the tissue, the power intensity must be selected and the foot pedal or fingerswitch must be activated. During the actual cutting, it is important to use a smooth uninterrupted motion with light, even pressure. The movement should not be too slow. If it is, the build-up of lateral heat in the tissue may cause charring, followed by necrosis and sloughing (see Preoperative Cutting Practice Section 6.1). When performing a second or third cut in the same surgical site, allow approximately ten seconds for the tissue to cool between applications of the electrode to the site. Unlike a steel scalpel or standard electrosurgery, Radiosurgery does not use manually applied pressure to cut. Therefore, a light, smooth, continuous, and brush-like stroke should be developed. Tissue damage will occur if heat is allowed to accumulate to the point where excessive dehydration occurs. The spread of lateral heat in tissue depends upon various factors as indicated in the following formula: Time x Power Lateral Heat+? Frequency Additionally the electrode type and selected waveform impact the thermal build up. A. Time of Electrode Contact B. Power Level C. Frequency Tissue provides lower resistance to higher frequencies, reducing thermal build up. D. Electrode Type A broader electrode tip requires more power and creates greater thermal spread. E. Nature of the Waveform - See Section 5.1 for details 12

13 6.1 PRE-OPERATIVE CUTTING PRACTICE Prepare the Generator for operations as described in Section 5- PREPARATION FOR USE, then follow the steps below. A. Select a piece of fresh, lean beef containing very little fat. Allow meat to reach room temperature. Note: Do not use veal because it does not change color when cut with an electrode. B. Place the meat on a Neutral Pad. C. Insert the electrode of choice (Empire, Loop, Diamond, Vari-Tip, etc.) into the handpiece. D. Select "MONOPOLAR CUT". E. Set output power intensity to 5. F. Depress the MONOPOLAR Footswitch pedal or Fingerswitch button. G. Using a smooth, brush-like motion, make several incisions of various lengths and depths. If you notice that the electrode will not cut, cuts only with pulling and dragging, or if tissue shreds adhere to the electrode while cutting, then increase the power gradually until a smooth cut occurs. H. If the setting is too high then you will notice sparking and discoloration along the electrode path. Reduce power and repeat the procedure as outlined above. I. Repeat the above procedure until such a point when no discoloration occurs and there is no visible sparking. The Radiosurgical tip should not encounter resistance. The cut should be smooth without sparking and without any drag. 6.2 PRE-OPERATIVE COAGULATION PRACTICE Intermittent gentle contact with the tissue is performed until bleeding stops. Effective coagulation is achieved when the treated area appears on the meat as blanched spot. Feather-light touching of the ball electrode is recommended to coagulate bleeders efficiently. A. Select a piece of fresh, lean beef containing very little fat. Allow meat to reach room temperature. Note: Do not use veal because it does not change color when cut with an electrode. B. Place the meat on a Neutral Pad. C. Insert a ball electrode into the Handpiece. D. Select MONOPOLAR HEMO. E. Set the output power intensity to 5. F. Position the ball electrode in light contact with the beef. G. Depress the MONOPOLAR Footswitch pedal or the Fingerswitch button. H. Adjust power setting for desired results. Bleeding should be stopped by some form of direct manually applied pressure. When bleeding has momentarily stopped, final sealing of the vessels can be accomplished by brief application of the partially rectified modulated waveform. Bipolar can also be used for coagulation. 13

14 VII. POWER OUTPUT CHARACTERISTICS Figure 7.0 Typical Nominal Plot Power Output vs. Dial Setting Monopolar: 500 Ohm load Bipolar: 200 Ohm Load Figure 7.1 Typical Nominal Plot Monopolar Cut Power Output vs. Resistive Load 14

15 Figure 7.2 Typical Nominal Plot Monopolar Cut/Coag Power Output vs. Resistive Load Figure 7.3 Typical Nominal Plot Monopolar Hemo Power Output vs. Resistive Load 15

16 Figure 7.4 Typical Nominal Plot Bipolar Hemo Power Output vs. Resistive Load Figure 7.5 Typical Nominal Plot Voltage Output vs. Dial Setting (Open Circuit) 16

17 VIII. TECHNICAL SPECIFICATIONS 8.0 OUTPUT CHARACTERISTICS Mode/ Output Nominal Power Output Power Output Activation Character Waveform Output Power vs. Setting vs. Load Resistance Cut 4.0 MHz 50W See Figure 7.0 See Figure 7.1 via footswitch or 500 Ω fingerswitch Cut/Coag 4.0 MHz with 50W See Figure 7.0 See Figure 7.2 via footswitch or fully 500 Ω fingerswitch envelope Hemo 4.0 MHz with 35W See Figure 7.0 See Figure 7.3 via footswitch or partially 500 Ω fingerswitch envelope Bipolar 4.0 MHz with 35W See Figure 7.0 See Figure 7.4 via footswitch Hemo partially 200 Ω envelope RF Power-out is within 20% of nominal above (when the dial setting is greater than 1). RF Power-out is stable within 10% within 10 seconds after activation. RF output frequency is within 10% of nominal. NOTE: The open-circuit output peak-to-peak voltage for all modes is approximately 1200 Volts. 8.1 PREVENTIVE INSPECTION AND MAINTENANCE Regular preventive inspection should be carried out to prevent reduced safety of the unit due to aging, wear and tear, etc. The cable and electrode must be checked before each use. The Generator must be inspected for technical safety at least once a year. The required technical safety inspection is detailed in the Maintenance Manual. The manufacturer assumes no responsibility for improper changes or repairs carried out on the unit or its accessories by unauthorized persons, and the warranty of the unit expires immediately. 17

18 8.2 SAFETY FEATURES Standard Compliance IEC IEC IEC Electrical Safety Electric Shock: Defibrillator Protection Class I Type BF Neutral Pad Earthed at HF Defined duty cycle protection 20 sec. ON / 30 sec. OFF Ingress of water Ordinary equipment. 8.3 INPUT CHARACTERISTICS Nominal line voltage 100V, 120V / 220V, 240V ~ + 10% Line frequency 60 Hz./50 Hz Input current at max output power 0.80A; 230/240VAC 0.90A; 220VAC 1.40A; 120VAC 1.70A; 100VAC Rating of fuses in the main supply T 1.25A, 250V, 240V, 220V power T 2.0A, 250V, 120V, 100VAC 8.4 SIZE AND WEIGHT W x H x D 10 x 4.5 x Weight 10 lbs. 8.5 OPERATING CONDITIONS Ambient Temperature: +10 C C Relative Humidity: 30% - 75% 8.6 TRANSPORTATION & STORAGE CONDITIONS Ambient Temperature: -10 C C Relative Humidity: 10% - 95% Atmospheric Pressure: 500 hpa hpa 18

19

20 Cynosure, Inc. dba Ellman, 400 Karin Lane, Hicksville, N.Y U.S.A. Tel: Fax: DIM REV E 8 December 2015

J-470 ELECTROSURGE. Radio Frequency Electrosurgery Unit. Operation Instructions

J-470 ELECTROSURGE. Radio Frequency Electrosurgery Unit. Operation Instructions J-470 ELECTROSURGE Radio Frequency Electrosurgery Unit Operation Instructions Power Foot Switch Electrodes ELECTROSURGE J-470 Waveform Mode Intensity Level RF Light Hand Piece Jack Indifferent Plate Jack

More information

USER S GUIDE. User s Guide Bovie IDS-400

USER S GUIDE. User s Guide Bovie IDS-400 USER S GUIDE User s Guide Bovie IDS-400 i ii Bovie Medical Corporation This manual and the equipment it describes are for use only by qualified medical professionals trained in the particular technique

More information

ELECTROSURGICAL GENERATOR USER S GUIDE

ELECTROSURGICAL GENERATOR USER S GUIDE IDS-300 ELECTROSURGICAL GENERATOR ELECTROSURGICAL GENERATOR USER S GUIDE USER S GUIDE User s Guide Bovie IDS-300 i ii Bovie Medical Corporation This manual and the equipment it describes are for use only

More information

Electrosurgery Units (ESU s) and their variants

Electrosurgery Units (ESU s) and their variants Electrosurgery Units (ESU s) and their variants Key Points Electrosurgical Units: - typical output frequency of modern ESU s (450-500 khz, but can be up to 5 MHz) higher frequency makes finer cut with

More information

User s Guide Aaron 950

User s Guide Aaron 950 USER S GUIDE User s Guide Aaron 950 i ii Bovie / Aaron Medical This manual and the equipment it describes are for use only by qualified medical professionals trained in the particular technique and surgical

More information

User s Guide Aaron 940

User s Guide Aaron 940 USER S GUIDE User s Guide Aaron 940 1 TABLE OF CONTENTS Introduction...3 Operating Principle...3 Safety...3 Introduction...3 Warnings and Cautions...3 Contraindications...6 Application Specification...6

More information

Crossfire Footswitch. User Guide

Crossfire Footswitch. User Guide Crossfire Footswitch 0475000100 User Guide Contents Warnings and Cautions... 3 Symbol Glossary... 4 Product Description and Intended Use... 6 System Overview...6 Indications/Contraindications...8 Package

More information

BOVIE SURGI-CENTER PRO. Electrosurgical Generator. User's Guide

BOVIE SURGI-CENTER PRO. Electrosurgical Generator. User's Guide BOVIE SURGI-CENTER PRO Electrosurgical Generator User's Guide USER S GUIDE User s Guide Bovie Surgi-Center PRO i ii Bovie Medical Corporation This manual and the equipment it describes are for use only

More information

Electrosurgery Units (ESU s) and their variants. D. J. McMahon rev cewood

Electrosurgery Units (ESU s) and their variants. D. J. McMahon rev cewood Electrosurgery Units (ESU s) and their variants D. J. McMahon 150107 rev cewood 2017-12-23 Key Points Electrosurgical Units: - typical output frequency of modern ESU s (450-500 khz, but can be up to 5

More information

PULSAR Generator Operator s Manual

PULSAR Generator Operator s Manual PULSAR Generator Operator s Manual Copyright 2012 Medtronic, Inc. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, stored in retrieval systems, or translated

More information

ERBE. ERBOTOM ICC 80 / 50 Instruction manual 08.00

ERBE. ERBOTOM ICC 80 / 50 Instruction manual 08.00 ERBE ERBOTOM ICC 80 / 50 Instruction manual 08.00 ERBOTOM ICC 80 / 50 V 1.X 10122-103, 10122-101 10122-003, 10122-041 Instruction manual 08.00 ISO 9001 EN 46001 All rights to this instruction manual,

More information

ERBE ERBOTOM ICC 300. Instruction manual 08.00

ERBE ERBOTOM ICC 300. Instruction manual 08.00 ERBE ERBOTOM ICC 300 Instruction manual 08.00 ERBOTOM ICC 300-H V 4.X 10128-070, 10128-077, 10128-306, 10128-075 ERBOTOM ICC 300-E V 4.X 10128-071, 10128-078, 10128-076, 10128-307, 10128-401 Instruction

More information

DERM 941 DERM 942 High-Frequency Desiccators

DERM 941 DERM 942 High-Frequency Desiccators DERM 941 DERM 942 High-Frequency Desiccators User s Guide USER S GUIDE User s Guide DERM 941 / 942 1 TABLE OF CONTENTS Introduction...3 Operating Principle...3 Safety...3 Introduction...3 Warnings and

More information

D C 01/2019 3

D C 01/2019 3 D-0117968-C 01/2019 3 4 D-0117968-C 01/2019 Screw Driver Screw Driver Unplug both the Red & Blue connectors. (see above) Place a small flat head screw driver on the small orange tabs and push down while

More information

Aquamantys 3 Pump Generator User Guide. Revision F

Aquamantys 3 Pump Generator User Guide. Revision F User Guide Aquamantys 3 Pump Generator User Guide Revision F Foreword The Aquamantys 3 Pump Generator is for use only by qualified medical personnel properly trained in the use of electrosurgical equipment,

More information

INSTRUCTIONS ESG-400

INSTRUCTIONS ESG-400 INSTRUCTIONS ELECTROSURGICAL GENERATOR ESG-400 USA: CAUTION: Federal Law restricts this device to use by, or on the order of, a physician. Contents i Contents Labels and Symbols... 4 Important Information

More information

RAISING THE LEVEL OF PROTECTION IN YOUR ORS

RAISING THE LEVEL OF PROTECTION IN YOUR ORS Outpatient Supplement to www outpatientsurgery.net Surgery R October 2009 Magazine 2009-10 Manager s Guide to Patient Safety RAISING THE LEVEL OF PROTECTION IN YOUR ORS E L E C T R O S U R G E R Y S A

More information

BM VET 80, 120, 140, 180 W

BM VET 80, 120, 140, 180 W Special Medical Technology, Ltd. VETERINARY RADIOFREQUENCY Electrosurgical Units SMT BM VET 80, 120, 140, 180 W 1014 Operating Instructions Special Medical Technology, Ltd. CONTENT: 1. INTRODUCTION...4

More information

INSTRUCTION MANUAL LKG 601 Electrical Safety Analyzer

INSTRUCTION MANUAL LKG 601 Electrical Safety Analyzer INSTRUCTION MANUAL LKG 601 Electrical Safety Analyzer 110 Toledo Street Farmingdale, NY 11735 USA http://www.netech.org 510-USER-Manual Rev3 10/29/2007 Dear User, We appreciate your purchase of the LKG

More information

SAVI SCOUT Surgical Guidance System. Console Operation Manual

SAVI SCOUT Surgical Guidance System. Console Operation Manual SAVI SCOUT Surgical Guidance System Console Operation Manual 2 Copyrights and Trademarks 2016 Cianna Medical, Inc. All rights reserved. Patents pending. Cianna Medical and SAVI are registered trademarks

More information

Neo Ultrasound Module Manual

Neo Ultrasound Module Manual Neo Ultrasound Module Manual Installation Instructions For complete User Operating Instructions, including Cautions, Warnings, Dangers, Indications, and Contraindications, refer to the User s Manuals.

More information

Model 5100F. Advanced Test Equipment Rentals ATEC (2832) OWNER S MANUAL RF POWER AMPLIFIER

Model 5100F. Advanced Test Equipment Rentals ATEC (2832) OWNER S MANUAL RF POWER AMPLIFIER Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) OWNER S MANUAL Model 5100F RF POWER AMPLIFIER 0.8 2.5 GHz, 25 Watts Ophir RF 5300 Beethoven Street Los Angeles, CA 90066

More information

Surgitron Dual Frequency 4.0 MHz Radiowave Technology

Surgitron Dual Frequency 4.0 MHz Radiowave Technology Surgitron Dual Frequency 4.0 MHz Radiowave Technology With over 50 years of experience, over 70 patents and more than 200 journal articles, Ellman is your trusted worldwide partner for surgical products

More information

ELECTROSURGICAL UNIT ANALYZER

ELECTROSURGICAL UNIT ANALYZER ELECTROSURGICAL UNIT ANALYZER ESU-2000A USER MANUAL BC BIOMEDICAL ESU-2000A TABLE OF CONTENTS WARNINGS, CAUTIONS, NOTICES... ii DESCRIPTION... 1 OVERVIEW... 2 OPERATING INSTRUCTIONS... 3 MANUAL REVISIONS...

More information

LEEP System 1000 Table of Contents

LEEP System 1000 Table of Contents 38281 Rev. A 10/14 LEEP System 1000 Table of Contents Section Content Page 1. Professional Use Guide................................................... 1 2. System Features........................................................

More information

Meg1 (with Ligasmart)

Meg1 (with Ligasmart) In the name of God OPERATING INSTRUCTIONS Meg1 (with Ligasmart) Electrosurgical Generator Certifications: Thired Edition Electrosurgical Generator Made in IRAN Printed in IRAN Date of issue: JULY 2013

More information

Andy Falconer. Principles of surgical diathermy

Andy Falconer. Principles of surgical diathermy Andy Falconer Principles of surgical diathermy 1 Andy Falconer Principles of surgical diathermy From an Engineer s perspective 2 Surgical Diathermy Uses RF Current to cut tissue or cause coagulation Modes

More information

Operator s Manual E L E C T R O S U R G I C A L U N I T

Operator s Manual E L E C T R O S U R G I C A L U N I T Operator s Manual TM E L E C T R O S U R G I C A L U N I T LIMITED WARRANTY For a period of two years following the date of delivery, CONMED Corporation warrants the CONMED System 5000 Electrosurgical

More information

returned to your distributor for attention. 14. Do not operate the function switch while the instrument is connected to a circuit.

returned to your distributor for attention. 14. Do not operate the function switch while the instrument is connected to a circuit. 1. This instrument must only be used by a competent and trained person and operated in strict accordance with the instructions. KYORITSU will not accept liability for any damage or injury caused by misuse

More information

R-Series R235LS 2-Channel Power Amplifier with Local Source Switching

R-Series R235LS 2-Channel Power Amplifier with Local Source Switching R-Series R235LS 2-Channel Power Amplifier with Local Source Switching User s Manual On Off R235LS POWER A MPLIFIER IMPORTANT SAFEGUARDS WARNING TO REDUCE THE RISK OF FIRE OR ELECTRIC SHOCK, DO NOT EXPOSE

More information

This manual is valid for the TM. In TENSity 5000 TENS Stimulator. This user manual is published by Current Solutions, LLC

This manual is valid for the TM. In TENSity 5000 TENS Stimulator. This user manual is published by Current Solutions, LLC INSTRUCTION MANUAL This manual is valid for the TM In TENSity 5000 TENS Stimulator This user manual is published by Current Solutions, LLC Current Solutions, LLC does not guarantee its contents and reserves

More information

ENGLISH (Original instructions) INSTRUCTION MANUAL. Drill DS4012 DOUBLE INSULATION. IMPORTANT: Read Before Using.

ENGLISH (Original instructions) INSTRUCTION MANUAL. Drill DS4012 DOUBLE INSULATION. IMPORTANT: Read Before Using. ENGLISH (Original instructions) INSTRUCTION MANUAL Drill DS402 05402 DOUBLE INSULATION IMPORTANT: Read Before Using. ENGLISH (Original instructions) SPECIFICATIONS Model DS402 Capacities Steel 3 mm Wood

More information

INSTRUCTION MANUAL LKG

INSTRUCTION MANUAL LKG INSTRUCTION MANUAL LKG 610 Electrical Safety Analyzer With 10 ECG Connectors 110 Toledo Street Farmingdale, NY 11735 USA Homepage: www.netech.org Dear User, We appreciate your purchase of the LKG 610 Electrical

More information

perioperativecpd.com The basics of diathermy

perioperativecpd.com The basics of diathermy perioperativecpd.com continuing professional development The basics of diathermy By the perioperativecpd team Introduction Electrosurgery or diathermy is the passage of a high-frequency alternating current

More information

Application Notes. Electrosurgical Analyzer Primer

Application Notes. Electrosurgical Analyzer Primer Electrosurgical Analyzer Primer Purpose Scope Overview The purpose is to outline the characteristics of electrosurgical analyzers, describe the challenges of measuring RF power and review the inspection

More information

1/4in. SHEET PALM SANDER

1/4in. SHEET PALM SANDER 1/4in. SHEET PALM SANDER OPERATING INSTRUCTIONS SPECIFICATIONS ITEM# 151608 Voltage: Max. output: No load speed: Cord: 120V ~ 60Hz 180W 14000 RPM 6 feet (2m) Dust-Sealed On/off Switch Dust Extraction Adaptor

More information

WRM-10 TM TRANSFORMER WINDING RESISTANCE METER

WRM-10 TM TRANSFORMER WINDING RESISTANCE METER WRM-10 TM TRANSFORMER WINDING RESISTANCE METER USER S MANUAL Vanguard Instruments Company, Inc. 1520 S. Hellman Ave. Ontario, California 91761, USA TEL: (909) 923-9390 FAX: (909) 923-9391 June 2009 Revision

More information

V50D ELECTRIC GUITAR AMPLIFIER WITH DIGITAL DELAY AND CHORUS. User s Manual

V50D ELECTRIC GUITAR AMPLIFIER WITH DIGITAL DELAY AND CHORUS. User s Manual V50D ELECTRIC GUITAR AMPLIFIER WITH DIGITAL DELAY AND CHORUS User s Manual R Velocity is a registered trademark of GHS Corporation Battle Creek MI, USA May be covered by one or more of the following: U.S.

More information

EXCELL NHP/T. A new step AheAd to the future ELECTROSURGICAL UNITS FOR MONOPOLAR AND BIPOLAR USE ISO ALSA IT ISO

EXCELL NHP/T. A new step AheAd to the future ELECTROSURGICAL UNITS FOR MONOPOLAR AND BIPOLAR USE ISO ALSA IT ISO EXCELL NHP/T ELECTROSURGICAL UNITS FOR MONOPOLAR AND BIPOLAR USE A new step AheAd to the future ISO 9001 9120.ALSA IT - 1231 ISO 13485 9124.ALS2 0051 The units EXCELL NHP/T series are provided with all

More information

What tests should be done?

What tests should be done? Valleylab Force FX periodic testing Electrosurgery The passage of high frequency electrical current through tissue to create a desired clinical effect. 2/7/2009 1 What tests should be done? 1. Output tests

More information

ATBG280/6 Bench Grinder Bench Grinder ATBG280/6 230V-50Hz 280 Watt 150mm x 25mm Wheel size

ATBG280/6 Bench Grinder Bench Grinder ATBG280/6 230V-50Hz 280 Watt 150mm x 25mm Wheel size Bench Grinder ATBG280/6 230V-50Hz 280 Watt 150mm x 25mm Wheel size SPECIFICATIONS Model Number : ATBG280/6 Nominal Voltage Power Consumption No load speed Wheel size Weight 230Volt 50Hz 280 Watts 2880

More information

MICROWAVE DIATHERMY AND SURGICAL DIATHERMY DIATHERMICS

MICROWAVE DIATHERMY AND SURGICAL DIATHERMY DIATHERMICS MICROWAVE DIATHERMY AND SURGICAL DIATHERMY 1 Microwave diathermy Microwave diathermy uses microwaves to generate heat in the body. It can be used to evenly warm deep tissues without heating the skin. Microwave

More information

Operator s Manual. PP022 Passive Probe

Operator s Manual. PP022 Passive Probe Operator s Manual PP022 Passive Probe 700 Chestnut Ridge Road Chestnut Ridge, NY, 10977-6499 Tel: (845) 425-2000, Fax: (845) 578 5985 teledynelecroy.com PP022 Passive Probe Instruction Manual 2017 Teledyne

More information

ELEKTROTOM 610. Operating Instructions. Valid from version 1130

ELEKTROTOM 610. Operating Instructions. Valid from version 1130 ELEKTROTOM 60 Operating Instructions Valid from version 30 Contens Page. Introduction 4 2. Routine checks following delivery 5 3. Initial installation 5 4. First usage 5 5. Important notes for safe usage

More information

Emprint Ablation System with Thermosphere Technology

Emprint Ablation System with Thermosphere Technology Emprint Ablation System with Thermosphere Technology Specification Guide + + + = FIELD CONTROL THERMAL CONTROL WAVELENGTH CONTROL THERMOSPHERE TM TECHNOLOGY Product Information REF / SKU Product Length

More information

GT-1050A 2 GHz to 50 GHz Microwave Power Amplifier

GT-1050A 2 GHz to 50 GHz Microwave Power Amplifier Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) Giga-tronics GT-1050A Microwave Power Amplifier GT-1050A 2 GHz to 50 GHz Microwave Power Amplifier Operation Manual

More information

USER'S MANUAL DMR-6700

USER'S MANUAL DMR-6700 USER'S MANUAL Multimeter True RMS DMR-6700 CIRCUIT-TEST ELECTRONICS www.circuittest.com Introduction This meter measures AC/DC Voltage, AC/DC Current, Resistance, Capacitance, Frequency (electrical & electronic),

More information

OPERATING MANUAL ELECTROSURGICAL UNIT

OPERATING MANUAL ELECTROSURGICAL UNIT OPERATING MANUAL ELECTROSURGICAL UNIT BOWA-IFU-11779-ARC400-V2.1.0-S0-EN Operating Manual ARC 400 3 Contents Contents 1. Using this operating manual... 9 1.1. Revision index... 9 1.2. Validity... 9 1.3.

More information

OPERATING MANUAL ELECTROSURGICAL UNIT

OPERATING MANUAL ELECTROSURGICAL UNIT OPERATING MANUAL ELECTROSURGICAL UNIT 2 Operating Manual ARC 350 900-351_IFU_V2.0.0_20221-S0-20140210-EN Contents Contents 1. Using this operating manual... 7 1.1. Revision index... 7 1.2. Validity...

More information

DM-45 Digital Multimeter

DM-45 Digital Multimeter INSTRUCTION MANUAL DM-45 Digital Multimeter Read and understand all of the instructions and safety information in this manual before operating or servicing this tool. Description The Greenlee DM-45 Digital

More information

CX-A6 Amplifier Installation & User Guide V9.0

CX-A6 Amplifier Installation & User Guide V9.0 CX-A6 Amplifier Installation & User Guide V9.0 Cloud Electronics Limited 140 Staniforth Road, Sheffield, S9 3HF England Tel + 44 (0) 114 244 7051 Fax + 44 (0) 114 242 5462 E-mail info@cloud.co.uk Web site

More information

Single Channel Radio Mic System USER MANUAL. WMU-116-H (Hand Held) WMU-116-B (Belt Pack) Single Channel Radio Mic System

Single Channel Radio Mic System USER MANUAL. WMU-116-H (Hand Held) WMU-116-B (Belt Pack) Single Channel Radio Mic System Single Channel Radio Mic System USER MANUAL WMU-116-H (Hand Held) WMU-116-B (Belt Pack) Single Channel Radio Mic System Welcome Thank you for choosing Hill Audio for your sound system. To make sure that

More information

Atomscope PORTABLE VETERINARY X-RAY EQUIPMENT INSTRUCTION MANUAL. Version 1.0

Atomscope PORTABLE VETERINARY X-RAY EQUIPMENT INSTRUCTION MANUAL. Version 1.0 Atomscope PORTABLE VETERINARY X-RAY EQUIPMENT INSTRUCTION MANUAL Version 1.0 Thank you for purchasing our HFX90V portable veterinary x-ray unit. We are confident that you will be pleased with the radiographs

More information

ENGLISH (Original instructions) INSTRUCTION MANUAL. Drill DOUBLE INSULATION. IMPORTANT: Read Before Using.

ENGLISH (Original instructions) INSTRUCTION MANUAL. Drill DOUBLE INSULATION. IMPORTANT: Read Before Using. ENGLISH (Original instructions) INSTRUCTION MANUAL Drill 64 642 643 007894 DOUBLE INSULATION IMPORTANT: Read Before Using. ENGLISH (Original instructions) SPECIFICATIONS Model 64 642 643 Capacities Steel

More information

Operator s Manual. PP016 Passive Probe

Operator s Manual. PP016 Passive Probe Operator s Manual PP016 Passive Probe 2017 Teledyne LeCroy, Inc. All rights reserved. Unauthorized duplication of Teledyne LeCroy documentation materials is strictly prohibited. Customers are permitted

More information

User s Manual Current Probe IM E. 8th Edition

User s Manual Current Probe IM E. 8th Edition User s Manual 701931 Current Probe 8th Edition Thank you for purchasing the Current Probe (Model 701931). This instruction manual contains useful information about the instrument s functions and operating

More information

DPA-1.2. Instruction Manual. 2 Channel Amplifier with Auto A/B Selector DPA-1.2 DPA-1.2 POWER SERIAL # LINE INPUT SENSING SPEAKER B OUT

DPA-1.2. Instruction Manual. 2 Channel Amplifier with Auto A/B Selector DPA-1.2 DPA-1.2 POWER SERIAL # LINE INPUT SENSING SPEAKER B OUT POWER Russound DPA-1.2 Instruction Manual 2 Channel Amplifier with Auto A/B Selector NEWMARKET, NH USA DPA-1.2 Russound 68835 Conforms to UL 6500 Certified to CSA C22.2 No1-94 DPA-1.2 Tested to Comply

More information

MODEL 3810/2 Line Impedance Stabilization Network

MODEL 3810/2 Line Impedance Stabilization Network EMC TEST SYSTEMS FEBRUARY 1996 REV C PN 399197 MODEL 3810/2 Line Impedance Stabilization Network OPERATION MANUAL USA P.O. Box 80589 Austin, Texas 78708-0589 2205 Kramer Lane, Austin, Texas 78758-4047

More information

Challenger 400. Instruction manual

Challenger 400. Instruction manual Challenger 400 Instruction manual 0349 301 097 041220 Valid for serial no. 448 DECLARATION OF CONFORMITY Murex Welding Products Ltd. Declare hereby that: Murex Challenger 400 Part No. 0349 308 110, 0349

More information

Drill INSTRUCTION MANUAL. WARNING: For your personal safety, READ and UNDERSTAND before using. SAVE THESE INSTRUCTIONS FOR FUTURE 1 REFERENCE.

Drill INSTRUCTION MANUAL. WARNING: For your personal safety, READ and UNDERSTAND before using. SAVE THESE INSTRUCTIONS FOR FUTURE 1 REFERENCE. ENGLISH (Original instructions) INSTRUCTION MANUAL Drill 6411 6412 6413 007894 DOUBLE INSULATION WARNING: For your personal safety, READ and UNDERSTAND before using. SAVE THESE INSTRUCTIONS FOR FUTURE

More information

OPERATION & SERVICE MANUAL FOR FC 110 AC POWER SOURCE

OPERATION & SERVICE MANUAL FOR FC 110 AC POWER SOURCE OPERATION & SERVICE MANUAL FOR FC 100 SERIES AC POWER SOURCE FC 110 AC POWER SOURCE VERSION 1.3, April 2001. copyright reserved. DWG No. FC00001 TABLE OF CONTENTS CHAPTER 1 INTRODUCTION... 1 1.1 GENERAL...

More information

2015 RIGOL TECHNOLOGIES, INC.

2015 RIGOL TECHNOLOGIES, INC. Service Guide DG000 Series Dual-channel Function/Arbitrary Waveform Generator Oct. 205 TECHNOLOGIES, INC. Guaranty and Declaration Copyright 203 TECHNOLOGIES, INC. All Rights Reserved. Trademark Information

More information

Crossfire 2 Integrated Resection and Energy System

Crossfire 2 Integrated Resection and Energy System Crossfire 2 Integrated Resection and Energy System 0475-100-000 Table of Contents Warnings and Cautions...3 Product Description/Intended Use...6 Indications...7 Contraindications...7 Package Contents...7

More information

Your Rocktron Bass15 Bass Amplifier has been tested and complies with the following Standards and Directives as set forth by the European Union:

Your Rocktron Bass15 Bass Amplifier has been tested and complies with the following Standards and Directives as set forth by the European Union: Instruction Manual Your Rocktron Bass15 Bass Amplifier has been tested and complies with the following Standards and Directives as set forth by the European Union: Standard(s): IEC60065 Professional Audio/Video

More information

ENGLISH (Original instructions) INSTRUCTION MANUAL. Drill MT600 MT601 DOUBLE INSULATION. IMPORTANT: Read Before Using.

ENGLISH (Original instructions) INSTRUCTION MANUAL. Drill MT600 MT601 DOUBLE INSULATION. IMPORTANT: Read Before Using. ENGLISH (Original instructions) INSTRUCTION MANUAL Drill MT600 MT60 003635 DOUBLE INSULATION IMPORTANT: Read Before Using. ENGLISH (Original instructions) SPECIFICATIONS Model MT600 MT60 Capacities Steel

More information

CX-A6 Amplifier Installation & User Guide V8.0

CX-A6 Amplifier Installation & User Guide V8.0 CX-A6 Amplifier Installation & User Guide V8.0 Cloud Electronics Limited 140 Staniforth Road, Sheffield, S9 3HF England Tel + 44 (0) 114 244 7051 Fax + 44 (0) 114 242 5462 E-mail info@cloud.co.uk Web site

More information

Operator s Manual. PP017 and PP018 Passive Probes

Operator s Manual. PP017 and PP018 Passive Probes Operator s Manual PP017 and PP018 Passive Probes 700 Chestnut Ridge Road Chestnut Ridge, NY, 10977-6499 Tel: (845) 425-2000, Fax: (845) 578 5985 teledynelecroy.com PP017 and PP018 Passive Probes Operator

More information

User s Manual Current Probe. IM E 2nd Edition IM E

User s Manual Current Probe. IM E 2nd Edition IM E User s Manual 700937 Current Probe 2nd Edition Introduction Thank you for purchasing the 700937 Current Probe. This Instruction Manual contains useful information about the instrument s functions and operating

More information

RIGOL. User s Guide. RP1000D Series High Voltage Differential Probe. Feb RIGOL Technologies, Inc

RIGOL. User s Guide. RP1000D Series High Voltage Differential Probe. Feb RIGOL Technologies, Inc User s Guide RP1000D Series High Voltage Differential Probe Feb. 2013 RIGOL Technologies, Inc Guaranty and Declaration Copyright 2012 RIGOL Technologies, Inc. All Rights Reserved. Trademark Information

More information

Installation & Operation Manual SAGA1-K Series Industrial Radio Remote Control

Installation & Operation Manual SAGA1-K Series Industrial Radio Remote Control Installation & Operation Manual SAGA1-K Series Industrial Radio Remote Control Gain Electronic Co. Ltd. Table Of Contents Safety Considerations ------------------------------------------------------------2

More information

Model S-520 Coin Counter / Sorter Operating Manual

Model S-520 Coin Counter / Sorter Operating Manual Model S-520 Coin Counter / Sorter Operating Manual Table of Contents Using the Product Safely... 2 About the Warning Stickers and Points of Warning/... 2 Warning.... 3. 4 Names of the Main Parts......

More information

MAC-AFRIC USER S MANUAL MODEL: D.C.INVERTER WELDER 160/200 AMP

MAC-AFRIC USER S MANUAL MODEL: D.C.INVERTER WELDER 160/200 AMP MAC-AFRIC USER S MANUAL MODEL: D.C.INVERTER WELDER 160/200 AMP Safety precautions! The processes of welding and cutting, involve potential hazards, so please take the necessary precautions when performing

More information

Tube Facing Tool.

Tube Facing Tool. www.swagelok.com Tube Facing Tool This manual contains important information for the safe and effective operation of the Swagelok TF72 series tube facing tool. Users should read and understand its contents

More information

HF- GENERATORS.

HF- GENERATORS. HF- GENERATORS www.mgb-berlin.de 2 CONTENT n HF-4000 L Page 4-5 n HF-4000 B & HF-2000 B Page 6 n HF-1500 B & HF-800 B n Technics of HF-4000 B/HF-2000 B/ HF-1500 B/HF-800 B n Active monopolar electrodes,

More information

Opus 21 s80 Integrated Amplifier Owner's Manual

Opus 21 s80 Integrated Amplifier Owner's Manual Opus 21 s80 Integrated Amplifier Owner's Manual r e s o l u t i o n From all of us at Resolution Audio, thank you for choosing the Opus 21 s80 amplifier. We went to great lengths to design and produce

More information

On-Line Cardio Theater Wireless Digital Transmitter Installation and Instruction Manual

On-Line Cardio Theater Wireless Digital Transmitter Installation and Instruction Manual On-Line Cardio Theater Wireless Digital Transmitter Installation and Instruction Manual Full installation instructions accompany your Cardio Theater equipment order. This On-Line version of our Installation/Instruction

More information

Unity Tuner Instruction Manual

Unity Tuner Instruction Manual Unity Tuner Instruction Manual This pedal is: Compliance Your Unity Tuner pedal has been tested and complies with the following Standards and Directives as set forth by the European Union: Council Directive(s):

More information

plifier D-501 otion Am Tactile M

plifier D-501 otion Am Tactile M Tactile Motion Amplifier D-501 IMPORTANT SAFETY INSTRUCTIONS WARNING: 1. Read and keep these instructions for future reference. 2. Do not use this apparatus near water. 3. Clean only with a dry cloth.

More information

SAFETY AND OPERATING MANUAL

SAFETY AND OPERATING MANUAL SAFETY AND OPERATING MANUAL 2 General Power Tool Safety Warnings WARNING: Read all safety warnings and all instructions. Failure to follow the warnings and instructions may result in electric shock, fire

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 07-424 Digital Densitometer II Operators Manual March 2005 Manual No. 112111 Rev. 4 2003, 2005 Fluke Corporation, All rights reserved. Printed U.S.A. All product names are trademarks

More information

MICRO SERIES MONITORS

MICRO SERIES MONITORS LTO R OWNER'S MANUAL MICRO SERIES MONITORS PASSIVE & ACTIVE MODELS www.altoproaudio.com Version. SEPTEMBER 007 English IMPORTANT SAFETY INSTRUCTION CAUTION RISK OF ELECTRIC SHOCK DO NOT OPEN TO REDUCE

More information

TETRIS 1000 High Impedance Active Probe. Instruction Manual

TETRIS 1000 High Impedance Active Probe. Instruction Manual TETRIS 1000 High Impedance Active Probe Instruction Manual Copyright 2015 PMK GmbH All rights reserved. Information in this publication supersedes that in all previously published material. Specifications

More information

User s Manual Current Probe. IM E 1st Edition. Yokogawa Electric Corporation

User s Manual Current Probe. IM E 1st Edition. Yokogawa Electric Corporation User s Manual 701932 Current Probe Yokogawa Electric Corporation 1st Edition Foreword Revisions Thank you for purchasing the 701932 Current Probe. This user's manual contains useful information about the

More information

Content. User Instruction - Pointoselect digital. Intended use...3 Safe usage of the device...3 Description...4

Content. User Instruction - Pointoselect digital. Intended use...3 Safe usage of the device...3 Description...4 2 Content Intended use...3 Safe usage of the device...3 Description...4 Operating the device Setting up the device...4 Starting the device...4 Mode selection...5 Measurement...5 Automatic...5 Manual...6

More information

MDC4500-4B Amplifier/Combiner Data Sheet & User Manual. February Rev 0.95

MDC4500-4B Amplifier/Combiner Data Sheet & User Manual. February Rev 0.95 MDC4500-4B Amplifier/Combiner Data Sheet & User Manual February 2016 - Rev 0.95 MDC4500-4B Amplifier IMPORTANT SAFETY and USEAGE INFORMATION Please review the following safety precautions to avoid injury

More information

White Paper: Technology and Clinical Overview of the Genii TouchSoft Coagulator monopolar probe

White Paper: Technology and Clinical Overview of the Genii TouchSoft Coagulator monopolar probe White Paper: Technology and Clinical Overview of the Genii TouchSoft Coagulator monopolar probe Prepared by Genii, Inc. St. Paul, MN, January 2012. (Updated April, 2015) All rights reserved. Introduction

More information

TETRIS User's Guide. High Impedance Active Probe DO177-1

TETRIS User's Guide. High Impedance Active Probe DO177-1 TETRIS 1500 High Impedance Active Probe User's Guide DO177-1 TETRIS 1500 Copyright 2010 Ltd. All rights reserved. Information in this publication supersedes that in all previously published material. Specifications

More information

PROFESSIONAL STEREO AMPLIFIERS USER GUIDE

PROFESSIONAL STEREO AMPLIFIERS USER GUIDE PROFESSIONAL STEREO AMPLIFIERS USER GUIDE 967/9673 - September 00 - Version.0 ENGLISH Ti SERIES - Professional stereo amplifiers Page Ti SERIES - Professional stereo amplifiers ENGLISH - Safety information

More information

User Instruction Computer Assisted Local Analgesia. 337 Marion, Le Gardeur QC, Canada, J5Z 4W8

User Instruction Computer Assisted Local Analgesia. 337 Marion, Le Gardeur QC, Canada, J5Z 4W8 User Instruction Computer Assisted Local Analgesia 1-800-667-9622 337 Marion, Le Gardeur QC, Canada, J5Z 4W8 USER INSTRUCTION Congratulations on your new CALAJECT! Please read these instructions thoroughly

More information

MODEL W Power Amplifier

MODEL W Power Amplifier TEGAM, INC. MODEL 2348 18.75 W Power Amplifier This owner s manual was as current as possible when this product was manufactured. However, products are constantly being updated and improved. Because of

More information

Model 228 Current Leakage Tester INSTRUCTION MANUAL

Model 228 Current Leakage Tester INSTRUCTION MANUAL Model 228 Current Leakage Tester INSTRUCTION MANUAL 1 About this Manual To the best of our knowledge and at the time written, the information contained in this document is technically correct and the procedures

More information

CANARY AUDIO. Power Amplifier CA-309 OWNER S MANUAL. Handcrafted in California MADE IN USA

CANARY AUDIO. Power Amplifier CA-309 OWNER S MANUAL. Handcrafted in California MADE IN USA CANARY AUDIO 300B Push-Pull Parallel Power Amplifier Mono Block Handcrafted in California CA-309 OWNER S MANUAL MADE IN USA Dear Customer: Please allow us to take this opportunity to thank you for purchasing

More information

U S E R S M A N U A L

U S E R S M A N U A L U S E R S M A N U A L! Note! Read this manual thoroughly before operating this unit. Failure to heed any warnings contained herein could potentially result in damage to your equipment. May be covered by

More information

ROTARY HAMMER OWNER'S MANUAL

ROTARY HAMMER OWNER'S MANUAL ROTARY HAMMER OWNER'S MANUAL WARNING: Read carefully and understand all INSTRUCTIONS before operating. Failure to follow the safety rules and other basic safety precautions may result in serious personal

More information

Reaction Chromatic Tuner Instruction Manual

Reaction Chromatic Tuner Instruction Manual Reaction Chromatic Tuner Instruction Manual Compliance Your Reaction Chromatic Tuner pedal has been tested and complies with the following Standards and Directives as set forth by the European Union: Council

More information

Model: PR-55. High Voltage Oscilloscope Probe 10 kv, 40 MHz USER MANUAL

Model: PR-55. High Voltage Oscilloscope Probe 10 kv, 40 MHz USER MANUAL Model: PR-55 High Voltage Oscilloscope Probe 10 kv, 40 MHz USER MANUAL 1 2017 All rights reserved. B&K Precision products are covered by US and foreign patents, issued and pending. Information is this

More information

Current Probes. User Manual

Current Probes. User Manual Current Probes User Manual ETS-Lindgren Inc. reserves the right to make changes to any product described herein in order to improve function, design, or for any other reason. Nothing contained herein shall

More information

Directions for Use. Veinlite A. TransLite, LLC.

Directions for Use. Veinlite A. TransLite, LLC. Directions for Use Veinlite A TransLite, LLC www.veinlite.com IMPORTANT: Veinlite A is a device which allows the medical practitioner to better visualize superficial veins. This information should be incorporated

More information

Rolyan Splint Pan OPERATION MANUAL. Item # Small Item # Large

Rolyan Splint Pan OPERATION MANUAL. Item # Small Item # Large Rolyan Splint Pan OPERATION MANUAL Item #081544816 - Small Item #081544808 Large PLEASE READ THIS ENTIRE MANUAL BEFORE OPERATING YOUR NEW SPLINT PAN. Failure to follow these instructions could result in

More information

troubleshooting LEVEL 1 training

troubleshooting LEVEL 1 training troubleshooting LEVEL 1 training EQUIPMENT TROUBLESHOOTING GENERATOR ERRORS AND SOLUTIONS? FOOT PEDAL UNRESPONSIVE - Check foot pedal connection and check line for signs of bends and kinks. - Review temperature

More information