Standard versus Custom calibration of a Fluke 5720A
|
|
- Clyde Summers
- 5 years ago
- Views:
Transcription
1 Standard versus Custom calibration of a Fluke 5720A How many calibration points suffice to be sure your instrument is completely within specifications? Preface In this document I would like to share an experience I had with the calibration of our Fluke 5720A Multifunction Calibrator. It will reveal a case in which the so called Standard calibration showed the unit to be within specifications, while a Custom calibration showed the unit to be out of specifications. Disclaimer It is not my intention to criticize either the calibration provider, their calibration methods and policies, or the performance of the Fluke 5720A Multifunction Calibrator. Introduction A Multifunction Calibrator like the Fluke 5720A is meant for calibrating measuring devices. To be sure that the values generated by the Multifunction Calibrator are traceable and accurate within their published specifications, the Multifunction Calibrator itself has to be calibrated too. During this calibration, the generated output signals of the Multifunction Calibrator are being measured and eventually the measured values compared to the published specifications. In this way it is possible to determine if the Multifunction Calibrator is within specs or out of specs. Remember however, that this statement only holds true for the points that have actually been measured! Hence, the more points a calibration provider measures, the more coverage you have, up to the point that you could state that the specific unit is completely (= at all values in all functions and ranges) within specifications. Since it is impossible to measure an infinitely number of points, there s always the dilemma of how many, and which calibration points suffice to be sure that the Multifunction Calibrator is completely within specifications. Of course there s a tradeoff between cost (effort) and confidence in this case. Here is where the so called Standard calibration comes in sight. A Standard calibration tries to cover all generally used functions and ranges of an instrument, while keeping costs down by reducing the number of calibration points as much as possible. One of the methods to accomplish this is to measure only one value halfway each range and function (the mid-range point). While this definitely reduces costs, it raises the question: Does it give the user enough confidence that the Multifunction Calibrator can be considered to be completely within specifications? TCEM 1208 Standard versus Custom calibration of a Fluke 5720A Page 1 of 6
2 Custom calibration Things become even more stringent when the Multifunction Calibrator is being used as a reference standard on which the Calibration and Measurement Capability (CMC) of an ISO accredited calibration laboratory is based. In this case the calibration of the Multifunction Calibrator must have enough coverage to calculate the CMC, and comply to the general rule that interpolation is allowed in the calculation of the CMC, while extrapolation is not allowed or only to a modest extent. This implies that you must have at least two widely spread calibration points on each function and range that are part of the CMC. This often disqualifies a Standard calibration for the purpose of deducing the CMC of an ISO accredited calibration laboratory. Some laboratories rely on the assumption that a Multifunction Calibrator is completely within specification when a Standard calibration is performed and all measured points are found to be within specs. But is this a safe assumption? Calibration of the Fluke 5720A We asked a calibration provider to perform a Custom calibration on our Fluke 5720A Multifunction Calibrator and supplied them with a list of calibration points we would like to have measured: two points per range (mostly around 10% and 90% of range) on all ranges and functions we are planning to use. The calibration provider apparently started with a Standard calibration and soon reported that our unit was found to be within specs, without any adjustments being performed. So far, so good. After about a week we got another, different message, telling us that during the Custom calibration it turned out that our unit was out of specs at four measured points and passed conditionally at another four measured points. [NB: Conditional Pass being that the measurement result is within the specification limit by a margin less than the measurement uncertainty: it is therefore not possible to state compliance based on the stated level of confidence. However the results indicate that compliance is more probable than non-compliance with the specification limit.] Our unit was subsequently adjusted and returned to us and we received the 5720A with no less than 4 calibration certificates: Calibration certificate with the results of the Standard calibration before adjustment Calibration certificate with the results of the Custom calibration before adjustment Calibration certificate with the results of the Standard calibration after adjustment Calibration certificate with the results of the Custom calibration after adjustment On the next pages the most interesting results of the Standard calibration and the Custom calibration are shown. TCEM 1208 Standard versus Custom calibration of a Fluke 5720A Page 2 of 6
3 Calibration results of the Standard calibration Before Adjustment, Calibration Date = DC Voltage 1 22 V V V V ± 0.5 ppm 28% Pass 2 22 V V V V ± 0.5 ppm 8% Pass V V V V ± 0.6 ppm 40% Pass V V V V ± 0.6 ppm 38% Pass AC Current pa No points measured on this range 6 22 ma ma 55 Hz ma ma ± 21 ppm 1% Pass 7 22 ma ma 400 Hz ma ma ± 21 ppm 1% Pass 8 22 ma ma 1 khz ma ma ± 21 ppm 1% Pass 9 22 ma ma 5 khz ma ma ± 21 ppm 3% Pass ma ma 10 khz ma ma ± 21 ppm 7% Pass TCEM 1208 Standard versus Custom calibration of a Fluke 5720A Page 3 of 6
4 Calibration results of the Custom calibration Before Adjustment; Calibration Date = DC Voltage V V V V ± 0.5 ppm 28% Pass V V V V ± 1.8 ppm 167% Fail V V V V ± 0.5 ppm 8% Pass V V V V ± 1.8 ppm 174% Fail V V V V ± 1.8 ppm 107% Fail V V V V ± 2.1 ppm 61% Pass V V V V ± 1.8 ppm 98% Pass V V V V ± 2.1 ppm 69% Pass AC Current pa pa 20 Hz pa pa ± 200 ppm 98% Conditional Pass pa pa 40 Hz pa pa ±200 ppm 91% Pass pa pa 1 khz pa pa ± 200 ppm 82% Pass pa pa 5 khz pa pa ± 200 ppm 42% Pass pa pa 10 khz pa pa ± 200 ppm 7% Pass pa pa 20 Hz pa pa ± 150 ppm 98% Conditional Pass pa pa 40 Hz pa pa ± 150 ppm 90% Conditional Pass pa pa 1 khz pa pa ± 150 ppm 18% Pass pa pa 5 khz pa pa ± 220 ppm 2% Pass pa pa 10 khz pa pa ± 600 ppm 2% Pass ma ma 20 Hz ma ma ± 56 ppm 112% Fail ma ma 40 Hz ma ma ± 36 ppm 46% Pass TCEM 1208 Standard versus Custom calibration of a Fluke 5720A Page 4 of 6
5 31 22 ma ma 1 khz ma ma ± 36 ppm 28% Pass ma ma 5 khz ma ma ± 36 ppm 9% Pass ma ma 10 khz ma ma ± 36 ppm 1% Pass ma ma 20 Hz ma ma ± 56 ppm 99% Conditional Pass ma ma 40 Hz ma ma ± 36 ppm 61% Pass ma ma 1 khz ma ma ± 36 ppm 33% Pass ma ma 5 khz ma ma ± 36 ppm 17% Pass ma ma 10 khz ma ma ± 36 ppm 4% Pass Within specifications out of specifications It is interesting to determine why the Multifunction Calibrator was judged to be within specs according to the results of the Standard calibration (see page 3) while judged to be out of specs according to the results of the Custom calibration (see page 4). Therefore we compared the measurement results from both calibrations. It is obvious that in the 22 ma AC Current range the Fail result at 2.2 ma@20 Hz (Row 29) wasn t detected because the lowest measured point of the Standard calibration was 10 ma@55 Hz (Row 6). Notice that even 2.2 ma@40 Hz (Row 30) measured well within specs. The two Fail results in the 22 V DC Voltage range are a completely different story. This range has been measured, but only at the mid-range points of 10 V and -10 V. [NB: In fact 10 V and -10 V are the low-range points of this range, because these values would normally be generated in the 11 V DC Voltage range.] Looking at the measured values of these points (Row 1 and 2), there is no hint whatsoever that would indicate the Multifunction Calibrator to be out of specifications on this range. However, when measured at the high-range points of +20 V and -20 V (Rows 12 and 14) there is no doubt that the unit is out of specifications by quite a margin. To a lesser extent the same applies to the Fail result on the 220 V DC Voltage range. Measured only at the mid-range points of 100 V and -100 V (Rows 3 and 4) the measurement results look okay and again there is no reason to believe that the Multifunction Calibrator would be out of specifications on this range. But then again, when measured at the low-range points it shows the unit to be just out of specifications at 20 V (Row 15) and very close to its spec limits at -20 V (Row 17). Regarding the three Conditional Pass results on the 220 pa range it is clear that they weren t detected during the Standard calibration because this range hasn t been measured at all. Evaluation of the Standard calibration The 5720A got a Pass result on all ca. 150 measured points There were no measured points with a Fail or even a Conditional Pass result TCEM 1208 Standard versus Custom calibration of a Fluke 5720A Page 5 of 6
6 There are no hints whatsoever that would indicate the unit to be out of specifications on any measured function or range Based on the results of the Standard calibration, the calibration provider could (would) decide not to adjust the Multifunction Calibrator because no out of tolerance values were found Based on the results of the Standard calibration, the user of the Multifunction Calibrator would be tempted to judge the 5720A to be completely (= at all values in all functions and ranges) within specifications Conclusions A Standard calibration not necessarily covers all functions and ranges During a Standard calibration often only mid-range calibration points are being measured Sometimes mid-range calibration points fail to show offset and gain errors to their full extent A Standard calibration may not suffice to judge if a certain instrument is completely (= at all values in all functions and ranges) within specifications Guidelines Offset errors are best detected by choosing a calibration point as low as practical in each range, for example 0% - 10% of range Gain errors are best detected by choosing a calibration point as high as practical in each range, for example 90% - 100% of range Amplitude errors over the frequency range are best detected by choosing calibration points at strategically spread frequencies, especially in the frequency extremes Recommendations Define your own Custom calibration points for the calibration of your Multifunction Calibrator: o Include calibration points on all ranges that you are (probably) going to use o Take at least one low-range and one high-range calibration point and eventually an additional mid-range point in all these ranges o For AC ranges choose strategically spread calibration frequencies, especially at the frequency extremes o Have a linearity check performed on (at least) one range These recommendations also apply to other instruments like Digital Multi Meters R.J. Timmerman MiPlaza, Philips Innovation Services TCEM 1208 Standard versus Custom calibration of a Fluke 5720A Page 6 of 6
Calibration Laboratory Assessment Service CLAS Certificate Number Page 1 of 10
Calibration Laboratory Assessment Service CLAS Certificate Number 95-02 Page 1 of 10 400 Britannia Road East, Unit #1 Mississauga, Ontario L4Z 1X9 Contact: Mr. Vince Casali Tel (905) 890-7600, (800) 36FLUKE
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. ZETEC, INCORPORATED 875, boul. Charest ouest, Suite 100 Quebec, QC, G1N 2C9 CANADA
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 ZETEC, INCORPORATED 875, boul. Charest ouest, Suite 100 Quebec, QC, G1N 2C9 CANADA Ghislain Cournoyer, Systems Manufacturer & Services Manager Phone: 418 263
More informationAn Evaluation of Artifact Calibration in the 5700A Multifunction Calibrator
An Evaluation of Artifact Calibration in the 57A Multifunction Calibrator Application Note Artifact Calibration, as implemented in the Fluke Calibration 57A Multifunction Calibrator, was a revolutionary
More informationSchedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK
DC RESISTANCE Unit 4 Riverside Business Centre Walnut Tree Close Guildford Surrey GU1 4UG Contact: Ian Norman Tel: +44 (0)1483 302700 Fax: +44 (0)1483 300562 E-Mail: inorman@caltest.co.uk Website: www.caltest.co.uk
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSLI Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSLI Z540-1-1994 TEKTRONIX, INC. 13725 SW Karl Braun Drive Beaverton, OR 97077 Andrew Mumford Phone: 503 627 2476 CALIBRATION Valid to: August 31, 2018
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 BRUEL AND KJAER North American Service Center Calibration Laboratory 3079 Premiere Parkway Suite 120 Duluth, GA 30097 William Shipman
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSLI Z & ANSI/NCSLI Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSLI Z540-1-1994 & ANSI/NCSLI Z540.3-2006 NORTHROP GRUMMAN SYSTEMS CORPORATION, AEROSPACE SYSTEMS One Space Park, S/2746 Redondo Beach, CA 90278 Robert
More informationCertificate of Accreditation
PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: (Hereinafter called the Organization) and hereby declares
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005, ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005, ANSI/NCSL Z540-1-1994 MEGGITT (ORANGE COUNTY), INC. 14600 Myford Road Irvine, CA 92606 Gil Gonzalez Phone: 949 493 8181 CALIBRATION Valid To: September 30,
More informationAmerican Association for Laboratory Accreditation
Page 1 of 6 5700A/5720A/3458A Purpose of Study This Measurement Advisory Committee (MAC) working group was formed to focus on the performance enhancement offered by artifact calibration for the Fluke 5700A,
More informationCertificate of Accreditation
PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: (Hereinafter called the Organization) and hereby declares
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 ENDRESS+HAUSER, INC. 10057 Porter Road Suite 100 LaPorte, TX 77571 Helmut Grohnert Phone: 317 535 1350 CALIBRATION Valid To: September
More information0.017 % % %
Parameter/Range Frequency CMC 2, 3 ( ) Comments AC Voltage Measure 10 mv 10 Hz 100 Hz 1 khz 10 khz 100 khz 1 MHz 97 µv/v 88 µv/v 88 µv/v 88 µv/v 0.017 % 0.037 % Fluke-792, AC standard 100 mv 10 Hz 100
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 ENDRESS+HAUSER, INC. 2350 Endress Place Greenwood, IN 46143 Helmut Grohnert Phone: 317 535 7138 CALIBRATION Valid To: January 31, 2017
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 THE MODAL SHOP, INC. A PCB GROUP COMPANY 3149 E. Kemper Road Cincinnati OH 45241 Lisa Moore Phone: 513 351-9919 CALIBRATION Valid To:
More information5790B/5/AF AC Measurement Standard
/5/AF AC Measurement Standard General Description The /5/AF specifications in Table 1 supersede those published in the Operator and Service Manuals. The specification differences (between the and the /5/AF)
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 SIMCO ELECTRONICS 7730B Trinity Road Suite 130 Cordova, TN 38018 Jeff Coddington Phone: 901-624-7448 CALIBRATION Valid to: September
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 KIMBALL ELECTRONIC LABORATORY, INC. 8081 W 21st Lane Hialeah, FL 33016 Javier Balceiro Phone: 305 822 5792 CALIBRATION Valid To: June
More informationUnderstanding GO/NO GO Gauges (Fixed Limit Gauging)
Understanding GO/NO GO Gauges (Fixed Limit Gauging) How do I choose a plug gauge for my measurement application. Therefore I put together this document to help everyone understand the concept of fixed
More informationCERTIFICATE OF ACCREDITATION
CERTIFICATE OF ACCREDITATION ANSI-ASQ National Accreditation Board 500 Montgomery Street, Suite 625, Alexandria, VA 22314, 877-344-3044 This is to certify that Wilmington Instrument Company, Inc. 332 North
More informationSchedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK
72 Manchester Road Kearsley Bolton BL4 8NZ Contact: Mr A Leonard Tel: +44 (0) 1204 571499 Fax: +44 (0) 1204 571734 E-Mail: enquiries@airflowmeasurements.com Website: www.airflowmeasurements.com AIR VELOCITY
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 ZETEC, INCORPORATED 8226 Bracken Place SE, Suite 100 Snoqualmie, WA 98065 Scott Taylor, After Market Service Manager Phone: 425 974 2766 CALIBRATION Valid To:
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 ANSI/NCSL Z & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 ANSI/NCSL Z540-1-1994 & ANSI/NCSL Z540.3-2006 ACUCAL INC. 600 South 56th Street Chandler, AZ 85226 Thomas Efaw Phone: 480 753 7000 CALIBRATION Valid To: September
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSLI Z & ANSI/NCSLI Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSLI Z540-1-1994 & ANSI/NCSLI Z540.3-2006 NORTHROP GRUMMAN SYSTEMS CORPORATION, AEROSPACE SYSTEMS One Space Park, S/2746 Redondo Beach, CA 90278 Rafael
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 NEW JERSEY INDUSTRIAL CONTROLS, LLC 28 River Street Dover, NJ 07801 Stephen Jamison Phone: 973 328 1745 / 201 306 2970 CALIBRATION Valid To: December 31, 2018
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 ELECTRO LAB SERVICES, INC. 517 N. Fulton Avenue Evansville, IN 47710 Robert J. Dreier Phone: 812 423 5211 CALIBRATION Valid To: April 30, 2019 Certificate Number:
More informationCertificate of Accreditation
PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: (Hereinafter called the Organization) and hereby declares
More informationCERTIFICATE of CALIBRATION
Primary Laboratory within AC Electricity Danish Standards Laboratory AC and RF Quantities DPLE Partner (Danish Primary Laboratory for Electricity) Page no. Appendix 1/5 1 CERTIFICATE of CALIBRATION for
More informationINAB Policy Statement on Scope Format for Calibration Laboratories PS27 1. INTRODUCTION
INAB Policy Statement on Scope Format for Calibration Laboratories PS27 1. INTRODUCTION 1.1. The definitive statement of the accreditation status of a calibration laboratory is the accreditation certificate
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 SdM AEROSPACE PYROMETRY CALIBRATION SERVICES Avenida Tecnologico # 5754 Colonia del Real Cd Juarez Chihuahua, Mexico CP 32690 Victor
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 YOKOGAWA CORPORATION OF AMERICA 2 Dart Road Newnan, GA 30265 Van Puymbrouck Phone: 770 254 0400 x 5021 CALIBRATION Valid To: December
More informationSchedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK
2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK Unit 8 New Vision Business Park Glascoed Road St Asaph LL17 0LP Contact: Mr Alan Horner Tel: +44 (0)1492 550 398 E-Mail: alan.horner@mcs-testequipment.com
More informationApplication Note. Calibration of 7000 Series Precision LCR Meters
Calibration of 7000 Series Precision LCR Meters Ever consider just how a precision impedance meter is calibrated? How the accuracy values are attained or what is meant by the uncertainty? What is the method
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 TRADEPORT ELECTRONICS CALIBRATION LABORATORY 1750 Steeles Ave. West, Suite 5 Concord, Ontario L4K 2L7 CANADA Barry Conway Phone: 905 660 3797 CALIBRATION Valid
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 INSTRUMENT CALIBRATION SOLUTIONS, INC. 413 Hopewell Avenue Aliquippa, PA 15001 Mr. David Beier Phone: 724 846 0136 CALIBRATION Valid
More informationCERTIFICATE OF ACCREDITATION
CERTIFICATE OF ACCREDITATION ANSI National Accreditation Board 11617 Coldwater Road, Fort Wayne, IN 46845 USA This is to certify that Kanawha Scales and Systems, Inc. 26 Whitney Drive Milford, OH 45150
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 WEST CALDWELL CALIBRATION LABORATORIES, INC. 1575 State Route 96 Victor, NY 14564 Felix Christopher Phone: 585 586 3900 CALIBRATION
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 RS CALIBRATION SERVICES, INC. 1047 Serpentine Lane Pleasanton, CA 94566 Mr. Ralph Sabiel Phone: 925 462 4217 CALIBRATION Valid To: April
More informationSchedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK
2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK Standards & Calibration Laboratory Contact: Mr Martin Matthews Archcliffe Road Tel: +44 (0)1304 502121 Dover Fax: +44 (0)1304 502268 Kent
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 ANSI/NCSL Z & ANSI/NCSLI Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 ANSI/NCSL Z540-1-1994 & ANSI/NCSLI Z540.3-2006 ANRITSU COMPANY MORGAN HILL CALIBRATION SERVICES 490 Jarvis Drive Morgan Hill, CA 95037 Yeou-Song (Brian) Lee
More informationSchedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK
2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK Measurement House Contact: Mr David Gray Kingsway Tel: +44 (0)1792 588722 Fforestfach Fax: +44 (0)1792 582624 Swansea E-Mail: mail@haven.co.uk
More informationCERTIFICATE OF ACCREDITATION
CERTIFICATE OF ACCREDITATION ANSI-ASQ National Accreditation Board 500 Montgomery Street, Suite 625, Alexandria, VA 22314, 877-344-3044 This is to certify that Instrument Company, Inc. 332 North Fries
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 &ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 &ANSI/NCSL Z540-1-1994 CIH EQUIPMENT COMPANY, INC. 1806 South Highland Ave Clearwater, FL 33756-1762 Dr. David Silver Phone: 727 584 5063 Email: dsilver@cihequipment.com
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 TESEQ INC., A UNIT OF AMETEK COMPLIANCE TEST SOLUTIONS 52 Mayfield Ave Edison, NJ 08837 Mary Jane Salvador Phone: 732 417 0501 E-mail:
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 ANSI/NCSL Z & ANSI/NCSLI Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 ANSI/NCSL Z540-1-1994 & ANSI/NCSLI Z540.3-2006 ANRITSU COMPANY MORGAN HILL CALIBRATION SERVICES 490 Jarvis Drive Morgan Hill, CA 95037 Yeou-Song (Brian) Lee
More informationIn late 2011, The International Standards
CISPR 32: New International Standard on Electromagnetic Emissions from Multimedia Equipment DAN HOOLIHAN Hoolihan EMC Consulting Lindstrom, Minnesota USA In late 2011, The International Standards Commission's
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 & ANSI/NCSL Z540.3-2006 PCB PIEZOTRONICS INC. 3425 Walden Avenue Depew, NY 14043 David J. Dulanski Phone: 716 684 0002 ext 2617 CALIBRATION
More informationTechnical Services Department, BHEL, Piplani, Bhopal, Madhya Pradesh. Discipline Electro Technical Calibration Issue Date
Last Amended on 29.02.2016 Page 1 of 6 SOURCE 1. DC VOLTAGE $ 100mV to 220 mv 220 mv to 220 V 220 V to 1000 V 2. AC VOLTAGE $ 50 Hz to 1 khz 3 mv to 220 mv 220 mv to 220 V 220 V to 1000 V 3. DC CURRENT
More informationCertificate of Calibration
Certificate of Calibration ISO/IEC 17025:2005 Certificate Number 1-5607144213-1 Model Number Manufacturer Description Serial Number 34401A Keysight Technologies Inc Digital multimeter, 6.5 digit MY41007563
More informationCERTIFICATE OF ACCREDITATION ISO/IEC 17025:2005 ANSI/NCSL Z (R2002)
CERTIFICATE OF ACCREDITATION ANSI-ASQ National Accreditation Board 500 Montgomery Street, Suite 625, Alexandria, VA 22314, 877-344-3044 This is to certify that Rothe Development, Inc. Metrology Services
More informationCERTIFICATE OF ACCREDITATION ISO/IEC 17025:2005 ANSI/NCSL Z
CERTIFICATE OF ACCREDITATION ANSI-ASQ National Accreditation Board 500 Montgomery Street, Suite 625, Alexandria, VA 22314, 877-344-3044 This is to certify that Advanced Technical Services NW, Inc. 8612
More informationMPS Electrical Test Laboratory, Larsen & Toubro Limited, Mysore Campus, KIADB Industrial Area, Hebbal Hootagalli, Mysore, Karnataka
Last Amended on - Page 1 of 6 SOURCE 1. DC VOLTAGE $ 1mV to 10 mv 0.13 % to 0.014% Using Multifunction Calibrator 10 mv to 1 V 0.014% to 0.0016% By Direct Method 1 v to 10 V 0.0016% to 0.0017% 10 V to
More informationElectrical Load Report
ELECTRICAL LOAD REPORT REPORT NO.: IANZ LR 1004 Date of Issue:- 3 October 2017 CUSTOMER: Name Schréder Australia (Pty) Ltd Address 6 Jayelem Crescent, Padstow, NSW 2211, AUSTRALIA OBJECT OF TEST: To produce
More information2-3 Calibration of Standard Voltage and Current Generator
2-3 Calibration of Standard Voltage and Current Generator Katsumi FUJII, Kojiro SAKAI, Tsutomu SUGIYAMA, Kouichi SEBATA, and Iwao NISHIYAMA This paper describes the calibration method of standard voltage
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 & ANSI/NCSL Z540.3-2006 SIMCO ELECTRONICS 8203-J Piedmont Triad Parkway Greensboro, NC 27409 Gene Denmark Phone: 336 605 1665 Fax: 336
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 ENDRESS+HAUSER, INC. 10057 Porter Road Suite 100 LaPorte, TX 77571 Helmut Grohnert Phone: 317 535 1350 CALIBRATION Valid To: September
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 EAGLE EYE CALIBRATION LABS, INC. 21358 Nordhoff Street, Suite 101 Chatsworth, CA 91311 Kirpal Chhokar Phone: 818 993 0055 CALIBRATION
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 GEFEN ELECTRONIC SERVICES LTD 30 Hasivim Street Petah-Tikva 4951169 Israel Arcady Gefen Phone: 97239266044 CALIBRATION Valid To: June 30, 2019 Certificate Number:
More informationCertificate of Accreditation
PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: Doko Bldg., 3-3-4,Minamisuna, Koto-ku, Tokyo 136-8580
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 JM TEST SYSTEMS, INC. Calibration Laboratory Division 1600 Watterberg Way Alexandria, LA 71303 Jerry Kraus Phone: 225 925 2029 x 3028
More informationCertificate of Accreditation
PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: Instrumentación Analítica y Metrología S.A. de C.V
More informationUsing Reference Multimeters for Precision Measurements
Using Reference Multimeters for Precision Measurements Advanced techniques for improved confidence in metrology Teleconference: US & Canada Toll Free Dial-In Number: 1-(866) 230-5936 International Dial-In
More informationSchedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK
Unit 3, Watt House Innovation Centre Pensnett Estate Kingswinford West Midlands DY6 7YD Contact: Mr A P Walker Tel: +44 (0)1384 401132 Fax: +44 (0)1384 400754 E-Mail: mail@quasartronics.com Website: www.quasartronics.com
More informationFiscal 2007 Environmental Technology Verification Pilot Program Implementation Guidelines
Fifth Edition Fiscal 2007 Environmental Technology Verification Pilot Program Implementation Guidelines April 2007 Ministry of the Environment, Japan First Edition: June 2003 Second Edition: May 2004 Third
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 INSTRUMENTATION DYNATHERM INC. 4127 Rue Cousens Ville St-Laurent, QC Canada H4S 1V6 Graziano Chila Phone: 514 636 7800 CALIBRATION Valid To: May 31, 2018 Certificate
More information5790A/AF AC Measurement Standard
AC Measurement Standard General Description The 5790A/AF is similar to the standard 5790A AC Measurement Standard except for different performance specifications and added accessories as noted below. The
More informationCertificate of Accreditation
PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: 241 Riverview Avenue. Newton MA 02466 505 Boul.du
More informationMeasurement of RF Emissions from a Caterpillar Inc. MSS3s RF ID Key Fob
Measurement of RF Emissions from a Caterpillar Inc. MSS3s RF ID Key Fob For Caterpillar Inc. 330 S.W. Adams Street Peoria, IL 61630 P.O. Number JBL 11260 Date Tested May 11, 2016 Test Personnel Mark Longinotti
More informationCalibration Laboratory Assessment Service CLAS Certificate Number Page 1 of 19
Calibration Laboratory Assessment Service CLAS Certificate Number 2000-03 Page 1 of 19 Transmation Canada nc. 916 Gateway Canada L7L 5K7 Contact: Robert Whittaker Tel (905) 632-5869 Toll Free (800) 897-0067
More informationCERTIFICATE OF ACCREDITATION
CERTIFICATE OF ACCREDITATION ANSI-ASQ National Accreditation Board 500 Montgomery Street, Suite 625, Alexandria, VA 22314, 877-344-3044 This is to certify that TesCom 2600 Longhorn Blvd., Ste. 112 Austin,
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 JM TEST SYSTEMS, INC. 3947 Lincoln Ave., Ste. B Groves, TX 77619 Jerry Kraus Phone: 432 580 7040 CALIBRATION Valid To: February 29,
More informationNew Standards for Test and Calibration of Phasor Measurement Units
New Standards for Test and Calibration of Phasor Measurement Units Jack Somppi Fluke Calibration NCSLI Conference Sacramento, CA August 2, 2012 2012 Fluke Corporation NCSLI PMU 20120802 1 Stability of
More informationSchedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK
CET Structures Ltd. trading as CET Calibration M1 Commerce Park Markham Lane Duckmanton Chesterfield S44 5HS Contact: Mr Chris Locke Tel: +44 (0)1246 828318 Fax: +44 (0)1246 828319 E-Mail: chris.locke@cetcalibration.com
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2017 & ANSI/NCSL Z VITREK LLC Kirkham Road Poway, CA Kevin Clark Phone:
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017 & ANSI/NCSL Z540-1-1994 VITREK LLC 12169 Kirkham Road Poway, CA 92064 Kevin Clark Phone: 858-689-2755 CALIBRATION Valid To: October 31, 2020 Certificate Number:
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 & ANSI/NCSL Z540.3-2006 EXPHIL CALIBRATION LABS, INC. 415 Central Avenue Bohemia, NY 11716 Steven Pidd Phone: 631 563 3520 CALIBRATION
More informationSchedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK
1 Wilson Street Contact: Mr Ibrahim Ibrahim Thornaby Tel: +44 (0) 1642 626148 Stockton-On-Tees Fax: +44 (0) 870 143 1869 TS17 7AR E-Mail: ibrahim@calibrate.co.uk United Kingdom Website: www.calibrate.co.uk
More informationSCOPE OF ACCREDITATION TO ISO 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO 17025:2005 & ANSI/NCSL Z540-1-1994 IET LABS, INC. 1202 VFW Parkway West Roxbury, MA 02132 David Sheena Phone: 800 899 8438 CALIBRATION Valid To: April 30, 2020 Certificate
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 & ANSI/NCSL Z540.3-2006 TESTEQUITY LLC 6100 Condor Drive Moorpark, CA 93021 Ed Yankajtis Phone: 805 498 9933 CALIBRATION Valid To: October
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005, ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005, ANSI/NCSL Z540-1-1994 NATIONAL TECHNICAL SYSTEMS (NTS) 7800 Highway 20 West Huntsville, AL 35806 Rick Davis Phone: 256 716 4483 CALIBRATION Valid To: December
More informationFCC Certification Test Report for the MEI Cashflow RFID Reader Base FCC ID: QP8EASITRAXRB
for the FCC ID: QP8EASITRAXRB WLL JOB# 9915 September 21, 2007 Prepared for: 1301 Wilson Drive West Chester, PA 19380 Prepared By: Washington Laboratories, Ltd. 7560 Lindbergh Drive Gaithersburg, Maryland
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2017 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017 & ANSI/NCSL Z540-1-1994 MTS METROLOGY AND CALIBRATION LABORATORY 14000 Technology Drive Eden Prairie, MN 55344 Kevin Rust Phone: 952 937 4790 www.mts.com CALIBRATION
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. PYLON ELECTRONICS INC. 147 Colonnade Road Ottawa, Ontario Canada K2E 7L9 Jim Mullins Phone:
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 PYLON ELECTRONICS INC. 147 Colonnade Road Ottawa, Ontario Canada K2E 7L9 Jim Mullins Phone: 613 226 7920 CALIBRATION Valid To: September 30, 2018 Certificate
More informationSchedule. This laboratory has demonstrated its technical competence to operate in accordance with MS ISO/IEC 17025:2005 (ISO/IEC 17025:2005).
LABORATORY LOCATION: (PERMANENT LABORATORY) Page: 1 of 3 LABORATORY OF NIHON DENKEI (M) SDN. BHD. UNIT NO. A-11-1, NORTH POINT OFFICE MID VALLEY CITY NO. 1, MEDAN SYED PUTRA UTARA 59200 KUALA LUMPUR FIELD
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 CALIBRATION LABORATORY, ELECTRONICS TESTING CENTER, TAIWAN No. 8, Lane 29, Wen Ming Rd., Kui Shan District Taoyuan City, Taiwan (R.O.C.) Wan-Chau Mou Phone:
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 ILLIANA INSTRUMENTATION SERVICE LLC 1831 Govert Dr Schererville, IN 46375 Paul Grolla Phone: 219 227 8788 CALIBRATION Valid To: September
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 AMETEK CO., LTD. NAGOYA OFFICE. (Main Laboratory) Onna, Atsugi-shi, Kanagawa-ken, Japan
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 AMETEK CO., LTD. NAGOYA OFFICE (Main Laboratory) 5-1-1 Onna, Atsugi-shi, Kanagawa-ken, 243-0032 Japan (Commercial Office) 1-329 Kifune, Meito-ku Nagoya-shi,
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 EXPHIL CALIBRATION LABS, INC. 415 Central Avenue Bohemia, NY 11716 Steven Pidd Phone: 631 563 3520 CALIBRATION Valid To: October 31,
More informationCalibration Certificate
ÌÇa8SÈ-EÇÂ!ÈfeFÎ Calibration Certificate Date of issue 2015-12-08 Calibrated by Gert Schnipper Approved by Ole F. Carlsen 1 (5) CUSTOMER: GN RESOUND A/S GTO LAUTRUPBJERG 7-9 2750 BALLERUP DEVICE UNDER
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. PYLON ELECTRONICS INC. 147 Colonnade Road Ottawa, Ontario Canada K2E 7L9 Jim Mullins Phone:
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 PYLON ELECTRONICS INC. 147 Colonnade Road Ottawa, Ontario Canada K2E 7L9 Jim Mullins Phone: 613 226 7920 CALIBRATION Valid To: September 30, 2020 Certificate
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 TEKTRONIX CANADA 400 Britannia Road E, Unit 2 Mississauga, Ontario L4Z 1X9 Canada Andrew Sweeny Phone: 905 241 3021 CALIBRATION Valid
More informationCERTIFICATE OF ACCREDITATION
CERTIFICATE OF ACCREDITATION ANSI-ASQ National Accreditation Board 500 Montgomery Street, Suite 625, Alexandria, VA 22314, 877-344-3044 This is to certify that Tra-Cal, LLC 7901 Beechcraft Avenue, Suite
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2017 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017 & ANSI/NCSL Z540-1-1994 G.T. MICHELLI COMPANY, INC. 130 Brookhollow Esplanade Harahan, LA 70123 Patrick Jester Phone: 504 733 9822 CALIBRATION Valid To: May
More informationCertificate of Accreditation
PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: (Hereinafter called the Organization) and hereby declares
More informationFCC Certification Test Report For the Mars Electronics (MEI) easitrax FCC ID: QP8EASITRAX
For the FCC ID: QP8EASITRAX WLL JOB# 9554 April 19, 2007 Prepared for: 1301 Wilson Drive West Chester, PA 19380 Prepared By: Washington Laboratories, Ltd. 7560 Lindbergh Drive Gaithersburg, Maryland 20879
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 CONNECTICUT CALIBRATION LABS 327 S. Main Street Newtown, CT 06470 Joseph Zeolla Phone: 203 270 1263 CALIBRATION Valid To: October 31,
More informationCertificate of Accreditation
PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: (Hereinafter called the Organization) and hereby declares
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 TRANE LA CROSSE METROLOGY LAB 2313 20th Street South La Crosse, WI 54601 Rick McKenzie Phone: 608 787 4032 CALIBRATION Valid To: October
More informationTR CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF ELECTRICAL DCLF METROLOGY. Approved By: Senior Manager: Mpho Phaloane Revised By:
CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF ELECTRICAL LF METROLOGY Approved By: Senior Manager: Mpho Phaloane Revised By: Specialist Technical Committee Date of Approval: 2015-08-26 Date of
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 & ANSI/NCSL Z540.3-2006 SIMCO ELECTRONICS 6295 Ferris Square San Diego, CA 92121 Jeremy Cooper Phone: 858 200 7027 CALIBRATION Valid
More informationKeysight Technologies Phase Noise, Amplitude and TOI Measurement Errors
Keysight Technologies Phase Noise, mplitude and TOI Measurement Errors rticle Reprint Reprinted with permission of MICROWVE JOURNL from the May 2015 issue. 2015 Horizon House Publications, Inc. Product
More information