1 Nuclear Instruments and Methods in Physics Research A () 9 Contents lists available at ScienceDirect Nuclear Instruments and Methods in Physics Research A journal homepage: Development of a large-area monolithic MPPC array for a future PET scanner employing pixelized Ce:LYSO and Pr:LuAG crystals T. Kato a,, J. Kataoka a, T. Nakamori a, T. Miura a, H. Matsuda a, K. Sato b, Y. Ishikawa b, K. Yamamura b, N. Kawabata b, H. Ikeda c, G. Sato c, K. Kamada d a Research Institute for Science and Engineering, Waseda University, --, Ohkubo, Shinjuku, Tokyo, Japan b Solid State Division, Hamamatsu Photonics K. K., -, Ichino-cho, Hamamatsu, Shizuoka, Japan c ISAS/JAXA, --, Yoshinodai, Chuo-ku, Sagamihara-shi, Kanagawa, Japan d Materials Research Laboratory, Furukawa Co., Ltd., --, Kannondai, Tsukuba, Ibaraki -, Japan article info Article history: Received December Received in revised form February Accepted February Available online February Keywords: Multi-Pixel Photon Counter Gamma-rays Positron emission tomography abstract We have developed a new type of large-area monolithic Multi-Pixel Photon Counter (MPPC) array consisting of a matrix of mm pixels. Each pixel comprises Geiger mode avalanche photodiodes (APDs) that achieve an average gain of 9. at 7.9 V at C with variations of only 77.% over pixels. Excellent uniformity was also obtained for photon detection efficiencies (PDE) of 7.%, whilst dark count rates at the single photoelectron ( p.e.) level amounted to C Mcps=pixel, measured at C. As the first step toward using the device in scintillation photon detectors, we fabricated a prototype gamma-ray camera consisting of an MPPC array optically coupled with a scintillator matrix, namely a array of mm crystals. Specifically, we tested the performance with Ce-doped (Lu, Y) (SiO )O (Ce:LYSO), Pr-doped Lu Al O (Pr:LuAG) and surface coated Pr:LuAG (Pr:LuAG (WLS)) matrices whereby the emission peak of Pr:LuAG was shifted from to nm via a wavelength shifter (WLS). Average energy resolutions of.%,.7% and.9% (FWHM) were obtained for kev gamma-rays, as measured at C with Ce:LYSO, Pr:LuAG and Pr:LuAG (WLS) scintillator matrices, respectively. We confirmed that the effective PDE for Pr:LuAG (WLS) had improved by more than % compared to original, non-coated Pr:LuAG matrix. These results suggest that a large-area monolithic MPPC array developed here could be promising for future medical imaging, particularly in positron emission tomography (PET). & Elsevier B.V. All rights reserved.. Introduction Positron emission tomography (PET) imaging is a well-established method for detecting cancers and diagnosing Alzheimer s in its early stages []. Currently, the PET combined with CT (CT-PET) has become more common as a multimodality imaging device [], as it provides an improved insight into the spatial and temporal interrelations between functional and anatomical images. However, CT imaging suffers from poor soft-tissue contrast with patients also subjected to a significant radiation dose that exceeds that received from the PET itself. Magnetic Resonance Imaging (MRI), however, offers excellent soft-tissue contrast and anatomical detail without the additional radiation. Many advantageous aspects of MRI-PET are now being proposed, and testing of their prototypes is underway [ ]. Unfortunately, the Photo-Multiplier Tube (PMT) incorporated in conventional PET scanners is difficult to use within the MRI high magnetic field. Corresponding author. address: katou.frme.@asagi.waseda.jp (T. Kato). Moreover, the spatial resolution attainable with a PMT-based PET is far from the theoretical limit of PET resolution. Various PET modules utilizing an avalanche photodiode (APD) have successfully demonstrated the potential for simultaneous MRI-PET imaging [] as well as ultimate sub-millimeter spatial resolutions [7]. One disadvantage of using APDs is, however, the relatively low Table Specification of the MPPC array at + C. Parameters Specification Number of elements [ch] ( ) Active area/channel (mm ) Pixel size of a Geiger-mode APD ðmmþ Typical photon detection efficiency ðl ¼ nmþ (%) Terminal capacitance/channel (pf) Operation voltage (V) 7.7. Gain 7. Typical dark current / channel ðmaþ Including cross-talk and after-pulse contributions. -9/$ - see front matter & Elsevier B.V. All rights reserved. doi:./j.nima...9
2 T. Kato et al. / Nuclear Instruments and Methods in Physics Research A () 9 Fig.. Pictures of the MPPC array developed in this paper. Fig.. Diagram of the readout system used for measuring the gain (top) and the proportionality (bottom). avalanche gain (typically r), meaning it is easily affected by electric noise contamination. MPPC, also known as a Silicon Photo-Multiplier (SiPM), is a compact, high performance semiconductor photodetector consisting of multiple Geiger-mode APD pixels. Its operating principles and basic performance are summarized in Ref. []. The MPPC has many advantages like APDs described above, such as insensitivity to magnetic fields, effective time resolution and compactness. In addition, it is operated in Geiger-mode, meaning its gain may be almost comparable to that of PMTs at up to the level. Despite its superior advantages, however, it also has several weak points compared to traditional PMTs and APDs. First, the
3 T. Kato et al. / Nuclear Instruments and Methods in Physics Research A () 9 dynamic range of the MPPC is limited by the number of Geigermode APD pixels comprised in the device. Each of the latter is subject to dead time (typically measured in tens of ns) once the Geiger-discharge has triggered, during which multiple photons entering a single pixel cannot be counted, resulting in a nonlinear response to the number of incident photons. Second, thermal electrons also trigger a Geiger-discharge, resulting in substantial contamination of dark counts, typically amounting to Mcps for mm MPPCs (whose Geiger-mode APDs are arranged with a pitch of mm) measured at + C [9]. Nevertheless, its great advantages make MPPC one of the ideal photosensors for MRI-PET and also for Time Of Flight (TOF) applications [ ]. There is also a wide-ranging choice of scintillators to be used in the PET scanner, such as Bi Ge O (BGO), Ce-doped Gd (SiO )O (Ce:GSO), Ce-doped Lu (SiO )O (Ce:LSO) and Ce-doped (Lu, Y) (SiO )O (Ce:LYSO). In modern PET scanners, BGO is no longer used due to its low light yield (% of Tl:NaI) and long scintillation Gain ( ) A- B- C A- B- C- D- Bias voltage [V] D A- B- C- D A- B- C- D Gain ( ) A- B- C- D-. A- B- C- D- 9. A- B- C- D A- B- C- D Fig.. Top: gain variation as a function of bias voltage for all pixels from 7.9 to 7.9 V, measured at C. In the top panel, circles and lines represent the actual measured value and the fitting line, respectively. The bottom panel shows the residual of the fitting line. Bottom: gain distribution at 7.9 V at C.
4 T. Kato et al. / Nuclear Instruments and Methods in Physics Research A () 9 decay time ( ns). The most popular scintillator at present is Ce:LSO or Ce:LYSO [], both of which feature high light yield (7% of Tl:NaI), short scintillation decay time ( ns) and high density (7. g/cm ) greater than BGO (7. g/cm ). Alternatively, brand-new scintillators with high light output and/or fast timing properties are being tested, especially for future applications in TOF-PET scanners. Pr-doped Lu Al O (Pr:LuAG) is one such scintillator characterized by a very fast decay time ( ns) and good light yield (% of Tl:NaI) [,]. Since its light emission is sharply concentrated at around nm, UV-enhanced APD arrays have been specifically developed to read out the Pr:LuAG scintillators [7]. Currently, the development of a high-resolution MRI-PET/TOF- PET technique utilizing the newly designed MPPC array is underway. As the first step, we report in this paper on the development of a large-area, monolithic MPPC array and its performance as a gamma-ray detector coupled with Ce:LYSO and Pr:LuAG matrices. The advantage of using MPPC monolithic arrays, instead of single MPPC devices assembled together [], is that we can easily achieve good uniformity of gain, PDE and dark counts over the pixels. Moreover, the gap between each pixel can be minimized, thereby improving the effective area of the MPPC arrays as a whole. Since the MPPC developed here is expected to be less sensitive to scintillation light at around nm, we also fabricated a sample of a Pr:LuAG matrix, the surface of which was coated with an appropriate wavelength shifter (WLS). This paper is organized as follows. In Section, we present the basic characteristics of a large-area monolithic MPPC array developed in this paper. In Section, performances with various pixelized scintillator arrays will be given as a probe of the compact gamma-ray imager. The final conclusions are presented in Section.. Basic properties of the MPPC array.. MPPC array and setup The large-area monolithic MPPC array described here was specifically designed and developed for future applications in A- A- A- A- B- B- B- B- C- C- C- C- D- D- D- D- PDE [%] Fig.. PDE distribution (at an LED light wavelength of nm) for all pixels at 7.9 V at C. 9 7 Fired Geiger-mode APD pixels ( ) A- A- A- A- B- B- B- B- C- Number of incident photons to MPPC ( ) C- C- C- D- D- D- D- Fig.. Fired Geiger-mode APD pixels as a function of the number of incident photons to the MPPC, as described in the text. All data were measured at 7.9 V at C. Circles and dashed lines represent the actual measured value and the fitting line in linear region (i.e. r fired APD pixels), respectively.
5 T. Kato et al. / Nuclear Instruments and Methods in Physics Research A () 9 7 nuclear medicine (e.g., PET scanners) by Hamamatsu Photonics K. K. The MPPC array consists of a array of individual mm pixels and a. mm gap between them. Each pixel comprises Geiger-mode APDs arranged with a pitch of mm. The dark current of each MPPC pixel ( mm ) is typically ma with a gain of 7. measured at C. Other basic characteristics of the MPPC array are summarized in Table. The MPPC array tested here is placed on a printed wiring board (PWB) package measuring. by. mm and.7 mm thick, and using conventional epoxy resin as an entrance window to the MPPC array. Also the bonding pads are gathered at the bottom side of the device, as shown in Fig. on the left. All pixels have a common cathode, namely, the same positive voltage is biased at the same time, whilst signals from individual anodes can be read through the signal pins gathered at the bottom-side of the PWB (Fig. right). First, the gain characteristic of each MPPC pixel was measured as a function of bias voltage. All the measurements were conducted at C to reduce the contamination of dark counts. The diagram of setup used for measuring gain is shown in Fig. (top). The MPPC array was irradiated by the weak blue light ( nm; close to the peak wavelength of Ce:LYSO) of a light emitting diode (LED), and the output charges from the MPPC were estimated by the single photoelectron ( p.e.) peak, which was individually taken with charge-sensitive ADC (HOSHIN V; hereafter CSADC). The input gate of the CSADC, ns in width, was activated simultaneously when triggering the LED. The output signals from the MPPC array were amplified by a factor of using the fast current amplifier (Phillips MODEL 9). Second, the proportionality of the output signal relative to the number of incident photons was measured for each MPPC pixel at a bias voltage of 7.9 V. The diagram of setup used for measuring the proportionality is shown in Fig. (bottom). To calibrate the absolute number of incident LED photons entering the MPPC window, a PMT (Hamamatsu R 799-MOD (EG)) was set beside the MPPC array. The output signals from both the PMT and MPPC were measured directly by the CSADC. Subsequently, the number of incident photons to the MPPC array was estimated by taking account of the various effective areas of the PMT and MPPC array as well as minor distance corrections between the LED and each of the photo-detectors. Also we assumed the quantum efficiency of the PMT to be % at an LED wavelength of nm. Finally, the variation of dark counts for individual MPPC pixels was measured as a function of the pulse height of output signals. The MPPC was operated with a bias voltage of 7.9 V. Output signals from the individual MPPC pixel were fed into the nonupdate discriminator (Technoland N-TM ) and directly counted by visual scalers (Technoland N-OR )... Result A gain characteristic as a function of bias voltage is shown in Fig. (top). The curves generally showed good linearity between 7.9 and 7.9 V, corresponding to the measured gain of CðÞ. Note that the gain curves fitted well with a slope of. gain/v, and distribution (among pixels) of only 7.%. The uniformity of the gain measured at a bias voltage of Table Basic characteristics of the Ce:LYSO and Pr:LuAG scintillators. Ce:LYSO Pr:LuAG Pixel size (mm ) Density (g/cm ) 7..7 Light yield (photons/mev),, Decay time (ns) Peak wavelength (nm) A- B- C- D- C- A- B- D- Dark count rates [kcps] A- B- C- D- A- B- C- D- Threshold [-mv] Fig.. Dark count rates as a function of a threshold voltage for all pixels measured at 7.9 V at C.
6 T. Kato et al. / Nuclear Instruments and Methods in Physics Research A () V is shown in Fig. (bottom). The average gain was 9. and its variation was only 77.% over MPPC pixels. Note that the uniformity of the gain at a bias voltage of 7.9 V is slightly worse than that expected from the variation of the gain slope (7%). This is possibly due to fluctuations in the breakdown voltage caused by the uniformity of the avalanche gain within the pixels. As shown in Fig., the total number of fired Geiger-mode APD pixels does not directly correspond to the number of detected photons; each of the MPPC pixels respond almost linearly up to the fired Geiger-mode APD pixels, even though the total number of pixels is. To quantify the PDE, the slope of the curves was determined within this linear range (i.e. r fired APD pixels). The resultant PDE map is shown in Fig.. The average PDE is 7.% (at an LED light wavelength of nm), and its variation is 7.% among MPPC pixels. The variations of the dark count rates as a function of threshold voltage for each MPPC pixel are shown in Fig.. It is clear that dark count rates due to p.e. clearly appear as a plateau between and mv, and its count rate is about Mcps/pixel. Fig. 7. Pictures of the scintillator matrices to be coupled with the MPPC array. (From left to right: Ce:LYSO, Pr:LuAG, Pr:LuAG (WLS)).. Performances with pixelized scintillators.. Scintillators and setup In our experiments, Ce:LYSO, Pr:LuAG and Pr:LuAG (WLS) were chosen to be coupled with the MPPC array described above. The basic parameters of Ce:LYSO and Pr:LuAG are summarized in Table. We note that the peak wavelength of Ce:LYSO favors MPPC because the latter is sensitive within the range nm []. In this sense, a UV light emission of Pr:LuAG, peaking at nm, is not favorable but its short decay time ( ns) is particularly noteworthy as implemented in future TOF-PET scanners (see Section ) and also beneficial for MPPC to effectively reduce dark counts within a narrower coincidence window. We therefore fabricated another sample, Pr:LuAG (WLS), which was coated with a WLS layer of mm that converts the nm scintillation light of the Pr:LuAG to nm light. The WLS comprises a solution including polyvinyl-toluene and plastic scintillators such as bis-msb (p-bis(o-methylstyryl) benzene) and TV-PVD (-(- terbutylphenyl)--(-biphenyl)-oxadiazol-,,) [9]. These scintillator matrices were fabricated to have geometries precisely matching the MPPC array, namely, a arrayof mm pixels, and a. mm gap between them. Each scintillator pixel is divided with a reflective BaSO layer of. mm thickness. Fig. 7 shows a picture of the matrices consisting of Ce:LYSO (right), Pr:LuAG (center) and Pr:LuAG (WLS) (right), respectively. Silicon optical grease (OKEN NA) was used for optical coupling between the scintillator matrices and the surface of the MPPC array. The performance of the MPPC array coupled with these pixelized scintillators was evaluated by taking the energy spectra of a 7 Cs source. All data were taken at 7.9 V at C. The diagram of setup used for measuring energy spectra is shown in Fig.. The output signals from each MPPC pixel were divided by a Quad Linear FAN IN/OUT (Phillips MODEL 9) and then sent to the CSADC after a ns delay, to ensure appropriate timing against the input gate signal (also activated from the same anode signal). We set gate widths of ns for the Ce:LYSO and ns for the Pr:LuAG and Pr:LuAG (WLS), respectively. The threshold voltage of the signals was set at around p.e... Result When the MPPC array is coupled with the Ce:LYSO matrix, the output charge for kev gamma-rays is approximately pc, CSADC channel [ch] Fig.. Diagram of the readout system used for measuring energy spectra. 7 Energy [kev] Fig. 9. An example calibration curve of the CSADC channel versus gamma-ray energy for a pixel A- with Ce:LYSO. Circles and line represent the actual measured value and fitting function, respectively. From left to right circle, Am, 7 Co, Ba, Na, 7 Cs, and Na sources were used. which corresponds to C fired Geiger-mode APD pixels. Apparently such abundant incident photons will cause substantial non-linearity in the MPPC response, as already seen in Fig.. To calibrate the energy and output ADC channel (i.e. charge) relation of the CSADC, the proper functional dependence was initially measured for the Ce:LYSO crystal using various gamma-ray
7 T. Kato et al. / Nuclear Instruments and Methods in Physics Research A () 9 9 Normalized counts - A- A- A- B- - B- C- D- - C- D- - B- C- D A- B- C- D Energy [kev] Fig.. Energy spectra of a 7 Cs source coupled with the Ce:LYSO matrices on the MPPC array for all pixels at 7.9 V at C. Normalized counts A- A- A- A- B- B- B- B- C- Energy [kev] C- D- D- C- D- C- D- Fig.. Energy spectra of a 7 Cs source coupled with the Pr:LuAG matrices on the MPPC array for all pixels at 7.9 V at C. sources, Na, 7 Co, Ba, and Am for each MPPC pixel. The results are shown in Fig. 9. As discussed in detail in literature (e.g., Ref. []), the relation is well represented by a simple function of the form ADC½chŠ¼a f expð beðkevþ=aþg ðþ
8 9 T. Kato et al. / Nuclear Instruments and Methods in Physics Research A () 9 Normalized counts A- - A- A- B- B- B- C- C- C- D- D- D- A- B- C- D- Energy [kev] Fig.. Energy spectra of a 7 Cs source coupled with the Pr:LuAG (WLS) matrices on the MPPC array for all pixels at 7.9 V at C. A- Energy resolution (FWHM) [%] B- C- D- A- Energy resolution (FWHM) [%] B- C- D-. A- B- C- D-. A- B- C- D-. A- A- B- B- C- C- D- D-... A- A- B- B- C- C- D- D-... A- A- A- A- Energy resolution (FWHM) [%] B- C- D-.. B- C- D-.. B- C- D-... B- C- D-.. Fig.. Variation of energy resolution for kev gamma-rays measured with the MPPC array and the various scintillators, the Ce:LYSO (top left), the Pr:LuAG (top right) and the Pr:LuAG (WLS) (bottom).
9 T. Kato et al. / Nuclear Instruments and Methods in Physics Research A () 9 9 Normalized counts where ADC [ch] is a measured channel of the CSADC, E (kev) is the energy of the gamma-ray sources, and a,b are fitting constants. Similarly, for the case of the Pr:LuAG or Pr:LuAG (WLS) matrix, the kev photoelectric peak corresponds to about or 7 fired Geiger-mode APD pixels, respectively. Therefore, the MPPC array with a Pr:LuAG matrix is operated well within the linear range, whilst the data taken with the Pr:LuAG (WLS) are marginally affected by the non-linear response. Therefore, a calibration equivalent to that applied for the Ce:LYSO matrix was also used to fix the response of the Pr:LuAG (WLS) matrix. The energy spectra measured with 7 Cs corrected for nonlinearity are shown in Figs. (Ce:LYSO), (Pr:LuAG) and (Pr:LuAG (WLS)). The average values of energy resolutions for the kev photoelectric peak measured with Ce:LYSO, Pr:LuAG, and Pr:LuAG (WLS) matrices were.%,.7%, and.9% (FWHM), respectively. The energy resolution variations among pixels were 7.%, 7.%, and 7.%, respectively, as shown in Fig.. Finally, a direct comparison of output charges from the MPPC array was made for different scintillators in Fig., corrected for non-linearity. It was clear that the output charge from the MPPC array with the Pr:LuAG (WLS) matrix was about % larger than that of the Pr:LuAG matrix, as expected from the improvement of PDE between and nm.. Conclusion Pr:LuAG Pr:LuAG (WLS) Ce:LYSO Charge [pc] Fig.. Comparison of the charge amount from one of the MPPC array pixels (pixel A-) coupled with the three scintillator matrices (Ce:LYSO, Pr:LuAG and Pr:LuAG (WLS)). The solid, thick solid and dashed lines represent the spectra of a 7 Cs source with Ce:LYSO, Pr:LuAG and Pr:LuAG (WLS), respectively. In this paper, the initial performances of the large-area monolithic MPPC array of mm pixels newly developed by Hamamatsu Photonics K. K. were reported. Excellent uniformities of 77.% and 7.% were obtained for gain variations and PDE distribution, respectively, as measured at C. The dark count rates were also measured to Mcps. Next, Ce:LYSO, Pr:LuAG, and Pr:LuAG (WLS) matrices of individual mm scintillator crystals were fabricated as a sensor head of a compact gamma-ray camera. An average energy resolution of.% (FWHM) was obtained with the Ce:LYSO matrix, as opposed to.7% (FWHM) and.9% (FWHM) with the Pr:LuAG and Pr:LuAG (WLS) matrices, respectively. The output charges for kev gamma-rays, corrected for non-linearity, were, 77 and pc, respectively, as measured with one of the MPPC pixels (pixel A-) coupled with a Ce:LYSO, Pr:LuAG, and Pr:LuAG (WLS) pixel. These results indicate that a large-area monolithic MPPC array coupled with a Ce:LYSO or Pr:LuAG matrix could be a useful device for medical imaging. Acknowledgment We thank an anonymous referee for his/her useful comments and suggestions to improve the manuscript. References [] W.W. Moses, Nucl. Instr. and Meth. A 7 () 9. [] J. Malamitsi, et al., Nucl. Instr. and Meth. A 9 () 9. [] R.R. Raylman, et al., Nucl. Instr. and Meth. A 9 (). [] J.E. Mackewn, et al., in: IEEE Trans. Nucl. Sci. Conference Record, vol. M9-, 9, p. 7. [] H. Peng, et al., Nucl. Instr. and Meth. A (). [] C. Woody, et al., Nucl. Instr. and Meth. A 7 (7). [7] J. Kataoka, et al., IEEE Trans. Nucl. Sci. NS-7 () (). [] K. Yamamoto, et al., in: IEEE Trans. Nucl. Sci. Conference Record, vol. N-,, p. 9. [9] K. Yamamoto, et al., in: IEEE Trans. Nucl. Sci. Conference Record, vol. N-9, 7, p.. [] A. Braem, et al., Nucl. Instr. and Meth. A (). [] C.L. Kim, et al., IEEE Trans. Nucl. Sci. NS- () (9). [] S. España, et al., Nucl. Instr. and Meth. A (). [] A. Nassalski, et al., IEEE Trans. Nucl. Sci. NS-7 () (). [] H. Alva-Sánchez, et al., Nucl. Instr. and Meth. A (9). [] M. Conti, et al., IEEE Trans. Nucl. Sci. NS- () (9) 9. [] K. Kamada, et al., IEEE Trans. Nucl. Sci. NS- () (9) 7. [7] M. Yoshino, J. Kataoka, T. Nakamori, H. Matsuda, T. Miura, Y. Ishikawa, N. Kawabata, Y. Matsunaga, K. kamada, Y. Usuki, A. Yoshikawa, T. Yanagida, Development and performance of UV enhanced APD-arrays for high resolution PET imaging coupled with pixelized Pr:LuAG crystal, Nucl. Instr. and Meth. A, submitted for publication. [] C.L. Kim, et al., in: IEEE Trans. Nucl. Sci. Conference Record, vol. M-7, 9, p.. [9] K. Kamada, T. Yanagida, T. Endo, Y. Fujimoto, Y. Usuki, A. Yoshikawa, Read out test of inorganic-organic hybrid scintillator; Pr:LuAG single crystal covered with plastic scintillator, IEEE Trans. Nucl. Conference Record, A, submitted for publication. [] M. Szawlowski, et al., in: IEEE Trans. Nucl. Sci. Conference Record, vol. NM-, 7, p. 9.
Nuclear Instruments and Methods in Physics Research A
Nuclear Instruments and Methods in Physics Research A 699 () Contents lists available at SciVerse ScienceDirect Nuclear Instruments and Methods in Physics Research A journal homepage: www.elsevier.com/locate/nima
More informationStudy of Silicon Photomultipliers for Positron Emission Tomography (PET) Application
Study of Silicon Photomultipliers for Positron Emission Tomography (PET) Application Eric Oberla 5 June 29 Abstract A relatively new photodetector, the silicon photomultiplier (SiPM), is well suited for
More informationLaBr 3 :Ce, the latest crystal for nuclear medicine
10th Topical Seminar on Innovative Particle and Radiation Detectors 1-5 October 2006 Siena, Italy LaBr 3 :Ce, the latest crystal for nuclear medicine Roberto Pani On behalf of SCINTIRAD Collaboration INFN
More informationARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A
Nuclear Instruments and Methods in Physics Research A 614 (2010) 308 312 Contents lists available at ScienceDirect Nuclear Instruments and Methods in Physics Research A journal homepage: www.elsevier.com/locate/nima
More informationTutors Dominik Dannheim, Thibault Frisson (CERN, Geneva, Switzerland)
Danube School on Instrumentation in Elementary Particle & Nuclear Physics University of Novi Sad, Serbia, September 8 th 13 th, 2014 Lab Experiment: Characterization of Silicon Photomultipliers Dominik
More informationSilicon Photomultiplier
Silicon Photomultiplier Operation, Performance & Possible Applications Slawomir Piatek Technical Consultant, Hamamatsu Corp. Introduction Very high intrinsic gain together with minimal excess noise make
More informationLaBr 3 :Ce scintillation gamma camera prototype for X and gamma ray imaging
8th International Workshop on Radiation Imaging Detectors Pisa 2-6 July 2006 LaBr 3 :Ce scintillation gamma camera prototype for X and gamma ray imaging Roberto Pani On behalf of SCINTIRAD Collaboration
More informationPerformance Evaluation of SiPM Detectors for PET Imaging in the Presence of Magnetic Fields
2008 IEEE Nuclear Science Symposium Conference Record M02-4 Performance Evaluation of SiPM Detectors for PET Imaging in the Presence of Magnetic Fields Samuel España, Student Member, IEEE, Gustavo Tapias,
More informationTiming Resolution Performance Comparison for Fast and Standard Outputs of SensL SiPM
Timing Resolution Performance Comparison for Fast and Standard Outputs of SensL SiPM Sergei Dolinsky, Geng Fu, and Adrian Ivan Abstract A new silicon photomultiplier (SiPM) with a unique fast output signal
More informationSilicon Carbide Solid-State Photomultiplier for UV Light Detection
Silicon Carbide Solid-State Photomultiplier for UV Light Detection Sergei Dolinsky, Stanislav Soloviev, Peter Sandvik, and Sabarni Palit GE Global Research 1 Why Solid-State? PMTs are sensitive to magnetic
More informationRecent Development and Study of Silicon Solid State Photomultiplier (MRS Avalanche Photodetector)
Recent Development and Study of Silicon Solid State Photomultiplier (MRS Avalanche Photodetector) Valeri Saveliev University of Obninsk, Russia Vienna Conference on Instrumentation Vienna, 20 February
More informationAn innovative detector concept for hybrid 4D-PET/MRI Imaging
Piergiorgio Cerello (INFN - Torino) on behalf of the 4D-MPET* project *4 Dimensions Magnetic compatible module for Positron Emission Tomography INFN Perugia, Pisa, Torino; Polytechnic of Bari; University
More informationAn active gain-control system for Avalanche photo-diodes under moderate temperature variations
Nuclear Instruments and Methods in Physics Research A 564 (2006) 300 307 www.elsevier.com/locate/nima An active gain-control system for Avalanche photo-diodes under moderate temperature variations J. Kataoka
More informationHighlights of Poster Session I: SiPMs
Highlights of Poster Session I: SiPMs Yuri Musienko* FNAL(USA)/INR(Moscow) NDIP 2011, Lyon, 5.07.2011 Y. Musienko (Iouri.Musienko@cern.ch) 1 Poster Session I 21 contributions on SiPM characterization and
More informationSilicon Photo Multiplier SiPM. Lecture 13
Silicon Photo Multiplier SiPM Lecture 13 Photo detectors Purpose: The PMTs that are usually employed for the light detection of scintillators are large, consume high power and are sensitive to the magnetic
More informationInvestigation of low noise, low cost readout electronics for high sensitivity PET systems based on Avalanche Photodiode arrays
Investigation of low noise, low cost readout electronics for high sensitivity PET systems based on Avalanche Photodiode arrays Frezghi Habte, Member, IEEE and Craig S.Levin, Member, IEEE Abstract A compact,
More informationInitial results on Sipm array based on a symmetric resistive voltage division readout
Initial results on Sipm array based on a symmetric resistive voltage division readout S. David, M. Georgiou, E. Fysikopoulos, N. Efthimiou, T. Paipais, L. Kefalidis and G. Loudos Abstract The aim of this
More informationTime-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud
University of Groningen Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you
More informationInvestigation of Solid-State Photomultipliers for Positron Emission Tomography Scanners
Journal of the Korean Physical Society, Vol. 50, No. 5, May 2007, pp. 1332 1339 Investigation of Solid-State Photomultipliers for Positron Emission Tomography Scanners Jae Sung Lee Department of Nuclear
More informationPoS(PhotoDet 2012)016
SiPM Photodetectors for Highest Time Resolution in PET, E. Auffray, B. Frisch, T. Meyer, P. Jarron, P. Lecoq European Organization for Nuclear Research (CERN), 1211 Geneva 23, Switzerland E-mail: stefan.gundacker@cern.ch
More informationarxiv: v2 [physics.ins-det] 17 Oct 2015
arxiv:55.9v2 [physics.ins-det] 7 Oct 25 Performance of VUV-sensitive MPPC for Liquid Argon Scintillation Light T.Igarashi, S.Naka, M.Tanaka, T.Washimi, K.Yorita Waseda University, Tokyo, Japan E-mail:
More informationTotal Absorption Dual Readout Calorimetry R&D
Available online at www.sciencedirect.com Physics Procedia 37 (2012 ) 309 316 TIPP 2011 - Technology and Instrumentation for Particle Physics 2011 Total Absorption Dual Readout Calorimetry R&D B. Bilki
More informationCHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES
CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES The current multiplication mechanism offered by dynodes makes photomultiplier tubes ideal for low-light-level measurement. As explained earlier, there
More informationAdvanced Materials Research Vol
Advanced Materials Research Vol. 1084 (2015) pp 162-167 Submitted: 22.08.2014 (2015) Trans Tech Publications, Switzerland Revised: 13.10.2014 doi:10.4028/www.scientific.net/amr.1084.162 Accepted: 22.10.2014
More informationCharacterization of a 64 Channel PET Detector Using Photodiodes for Crystal Identification *
Characterization of a 64 Channel PET Detector Using Photodiodes for Crystal Identification * J. S. Huber, Member, IEEE, W.W. Moses, Senior Member, IEEE, S.E. Derenzo, Senior Member, IEEE, M.H. Ho, M.S.
More informationCharacterisation of SiPM Index :
Characterisation of SiPM --------------------------------------------------------------------------------------------Index : 1. Basics of SiPM* 2. SiPM module 3. Working principle 4. Experimental setup
More informationPET Detectors. William W. Moses Lawrence Berkeley National Laboratory March 26, 2002
PET Detectors William W. Moses Lawrence Berkeley National Laboratory March 26, 2002 Step 1: Inject Patient with Radioactive Drug Drug is labeled with positron (β + ) emitting radionuclide. Drug localizes
More informationDevelopment of the Pixelated Photon Detector. Using Silicon on Insulator Technology. for TOF-PET
July 24, 2015 Development of the Pixelated Photon Detector Using Silicon on Insulator Technology for TOF-PET A.Koyama 1, K.Shimazoe 1, H.Takahashi 1, T. Orita 2, Y.Arai 3, I.Kurachi 3, T.Miyoshi 3, D.Nio
More informationDesign and Simulation of a Silicon Photomultiplier Array for Space Experiments
Journal of the Korean Physical Society, Vol. 52, No. 2, February 2008, pp. 487491 Design and Simulation of a Silicon Photomultiplier Array for Space Experiments H. Y. Lee, J. Lee, J. E. Kim, S. Nam, I.
More informationCharacterization of Silicon Photomultipliers and their Application to Positron Emission Tomography. Zhiwei Yang. Abstract
DESY Summer Student Program 2009 Report No. Characterization of Silicon Photomultipliers and their Application to Positron Emission Tomography Zhiwei Yang V. N. Karazin Kharkiv National University E-mail:
More informationA high energy gamma camera using a multiple hole collimator
ELSEVIER Nuclear Instruments and Methods in Physics Research A 353 (1994) 328-333 A high energy gamma camera using a multiple hole collimator and PSPMT SV Guru *, Z He, JC Ferreria, DK Wehe, G F Knoll
More informationPerformance measurements of a depth-encoding PET detector module based on positionsensitive
Home Search Collections Journals About Contact us My IOPscience Performance measurements of a depth-encoding PET detector module based on positionsensitive avalanche photodiode read-out This article has
More informationScintillation Counters
PHY311/312 Detectors for Nuclear and Particle Physics Dr. C.N. Booth Scintillation Counters Unlike many other particle detectors, which exploit the ionisation produced by the passage of a charged particle,
More informationTime-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud
University of Groningen Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you
More informationDevelopment of an innovative LSO-SiPM detector module for high-performance Positron Emission Tomography
Development of an innovative LSO-SiPM detector module for high-performance Positron Emission Tomography Maria Leonor Trigo Franco Frazão leonorfrazao@ist.utl.pt Instituto Superior Técnico, Lisboa, Portugal
More informationA flexible compact readout circuit for SPAD arrays ABSTRACT Keywords: 1. INTRODUCTION 2. THE SPAD 2.1 Operation 7780C - 55
A flexible compact readout circuit for SPAD arrays Danial Chitnis * and Steve Collins Department of Engineering Science University of Oxford Oxford England OX13PJ ABSTRACT A compact readout circuit that
More informationTime-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud
University of Groningen Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you
More informationReview of Solidstate Photomultiplier. Developments by CPTA & Photonique SA
Review of Solidstate Photomultiplier Developments by CPTA & Photonique SA Victor Golovin Center for Prospective Technologies & Apparatus (CPTA) & David McNally - Photonique SA 1 Overview CPTA & Photonique
More informationSolid-State Photomultiplier in CMOS Technology for Gamma-Ray Detection and Imaging Applications
Solid-State Photomultiplier in CMOS Technology for Gamma-Ray Detection and Imaging Applications Christopher Stapels, Member, IEEE, William G. Lawrence, James Christian, Member, IEEE, Michael R. Squillante,
More informationHigh collection efficiency MCPs for photon counting detectors
High collection efficiency MCPs for photon counting detectors D. A. Orlov, * T. Ruardij, S. Duarte Pinto, R. Glazenborg and E. Kernen PHOTONIS Netherlands BV, Dwazziewegen 2, 9301 ZR Roden, The Netherlands
More informationTest results on hybrid photodiodes
Nuclear Instruments and Methods in Physics Research A 421 (1999) 512 521 Test results on hybrid photodiodes N. Kanaya*, Y. Fujii, K. Hara, T. Ishizaki, F. Kajino, K. Kawagoe, A. Nakagawa, M. Nozaki, T.Ota,
More informationSmall Prototype Gamma Spectrometer Using CsI(Tl) Scintillator Coupled to a Solid-State Photomultiplier
10P-58 1 Small Prototype Gamma Spectrometer Using CsI(Tl) Scintillator Coupled to a Solid-State Photomultiplier Eric M. Becker, Member IEEE, Abdollah T. Farsoni, Member, IEEE, Abdulsalam M. Alhawsawi,
More informationDesigning an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare
GE Healthcare Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare There is excitement across the industry regarding the clinical potential of a hybrid
More informationSolid State Photomultiplier: Noise Parameters of Photodetectors with Internal Discrete Amplification
Solid State Photomultiplier: Noise Parameters of Photodetectors with Internal Discrete Amplification K. Linga, E. Godik, J. Krutov, D. Shushakov, L. Shubin, S.L. Vinogradov, and E.V. Levin Amplification
More informationDesign and development of compact readout electronics with silicon photomultiplier array for a compact imaging detector *
CPC(HEP & NP), 2012, 36(10): 973 978 Chinese Physics C Vol. 36, No. 10, Oct., 2012 Design and development of compact readout electronics with silicon photomultiplier array for a compact imaging detector
More informationInstitute for Particle and Nuclear Studies, High Energy Accelerator Research Organization 1-1 Oho, Tsukuba, Ibaraki , Japan
1, Hiroaki Aihara, Masako Iwasaki University of Tokyo 7-3-1 Hongo, Bunkyo-ku, Tokyo 113-0033, Japan E-mail: chojyuro@gmail.com Manobu Tanaka Institute for Particle and Nuclear Studies, High Energy Accelerator
More informationX-ray Detectors: What are the Needs?
X-ray Detectors: What are the Needs? Sol M. Gruner Physics Dept. & Cornell High Energy Synchrotron Source (CHESS) Ithaca, NY 14853 smg26@cornell.edu 1 simplified view of the Evolution of Imaging Synchrotron
More informationThomas Frach, Member, IEEE, Walter Ruetten, Member, IEEE, Klaus Fiedler, Gunnar Maehlum, Member, IEEE, Torsten Solf, and Andreas Thon
Assessment of Photodiodes as a Light Detector for PET Scanners Thomas Frach, Member, IEEE, Walter Ruetten, Member, IEEE, Klaus Fiedler, Gunnar Maehlum, Member, IEEE, Torsten Solf, and Andreas Thon Abstract
More informationA Measurement of the Photon Detection Efficiency of Silicon Photomultipliers
A Measurement of the Photon Detection Efficiency of Silicon Photomultipliers A. N. Otte a,, J. Hose a,r.mirzoyan a, A. Romaszkiewicz a, M. Teshima a, A. Thea a,b a Max Planck Institute for Physics, Föhringer
More informationSiPMs for solar neutrino detector? J. Kaspar, 6/10/14
SiPMs for solar neutrino detector? J. Kaspar, 6/0/4 SiPM is photodiode APD Geiger Mode APD V APD full depletion take a photo-diode reverse-bias it above breakdown voltage (Geiger mode avalanche photo diode)
More informationPerformance Assessment of Pixelated LaBr 3 Detector Modules for TOF PET
Performance Assessment of Pixelated LaBr 3 Detector Modules for TOF PET A. Kuhn, S. Surti, Member, IEEE, J. S. Karp, Senior Member, IEEE, G. Muehllehner, Fellow, IEEE, F.M. Newcomer, R. VanBerg Abstract--
More informationAPD Quantum Efficiency
APD Quantum Efficiency Development of a 64-channel APD Detector Module with Individual Pixel Readout for Submillimeter Spatial Resolution in PET Philippe Bérard a, Mélanie Bergeron a, Catherine M. Pepin
More informationWhite Paper. Zecotek MAPD (Multi- pixel Avalanche Photo Diode) Enabling the future of imaging and detection
White Paper Zecotek MAPD (Multi- pixel Avalanche Photo Diode) Enabling the future of imaging and detection Zecotek Photonics Inc. (TSX- V: ZMS; Frankfurt: W1I) www.zecotek.com is a Canadian photonics technology
More informationP ILC A. Calcaterra (Resp.), L. Daniello (Tecn.), R. de Sangro, G. Finocchiaro, P. Patteri, M. Piccolo, M. Rama
P ILC A. Calcaterra (Resp.), L. Daniello (Tecn.), R. de Sangro, G. Finocchiaro, P. Patteri, M. Piccolo, M. Rama Introduction and motivation for this study Silicon photomultipliers ), often called SiPM
More informationPoS(PhotoDet 2012)022
SensL New Fast Timing Silicon Photomultiplier Kevin O`Neill 1 SensL Technologies Limited 6800 Airport Business Park, Cork, Ireland E-mail: koneill@sensl.com Nikolai Pavlov SensL Technologies Limited 6800
More informationScintillators as an external trigger for cathode strip chambers
Scintillators as an external trigger for cathode strip chambers J. A. Muñoz Department of Physics, Princeton University, Princeton, NJ 08544 An external trigger was set up to test cathode strip chambers
More informationarxiv:astro-ph/ v1 17 Feb 2006
An Active Gain-control System for Avalanche Photo-Diodes under Moderate Temperature Variations arxiv:astro-ph/0602392v1 17 Feb 2006 J. Kataoka a, R. Sato a, T. Ikagawa a, J. Kotoku a, Y. Kuramoto a, Y.
More information764 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 51, NO. 3, JUNE 2004
764 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 51, NO. 3, JUNE 2004 Study of Low Noise Multichannel Readout Electronics for High Sensitivity PET Systems Based on Avalanche Photodiode Arrays Frezghi Habte,
More informationPerformance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET
2005 IEEE Nuclear Science Symposium Conference Record M11-126 Performance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET Jin Zhang, Member,
More informationChromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC
Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC R. Bellazzini a,b, G. Spandre a*, A. Brez a, M. Minuti a, M. Pinchera a and P. Mozzo b a INFN Pisa
More informationLarge area silicon photomultipliers: Performance and applications
Nuclear Instruments and Methods in Physics Research A 567 (26) 78 82 www.elsevier.com/locate/nima Large area silicon photomultipliers: Performance and applications P. Buzhan a, B. Dolgoshein a,, L. Filatov
More informationHF Upgrade Studies: Characterization of Photo-Multiplier Tubes
HF Upgrade Studies: Characterization of Photo-Multiplier Tubes 1. Introduction Photomultiplier tubes (PMTs) are very sensitive light detectors which are commonly used in high energy physics experiments.
More informationAn Introduction to the Silicon Photomultiplier
An Introduction to the Silicon Photomultiplier The Silicon Photomultiplier (SPM) addresses the challenge of detecting, timing and quantifying low-light signals down to the single-photon level. Traditionally
More informationGas scintillation Glass GEM detector for high-resolution X-ray imaging and CT
Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT Takeshi Fujiwara 1, Yuki Mitsuya 2, Hiroyuki Takahashi 2, and Hiroyuki Toyokawa 2 1 National Institute of Advanced Industrial
More informationFundamentals of CMOS Image Sensors
CHAPTER 2 Fundamentals of CMOS Image Sensors Mixed-Signal IC Design for Image Sensor 2-1 Outline Photoelectric Effect Photodetectors CMOS Image Sensor(CIS) Array Architecture CIS Peripherals Design Considerations
More informationA Study of Silicon Photomultiplier Sensor Prototypes for Readout of a Scintillating Fiber / Lead Sheet Barrel Calorimeter
2007 IEEE Nuclear Science Symposium Conference Record N41-6 A Study of Silicon Photomultiplier Sensor Prototypes for Readout of a Scintillating Fiber / Lead Sheet Barrel Calorimeter Carl J. Zorn Abstract:
More informationDevelopment of the LBNL Positron Emission Mammography Camera
Development of the LBNL Positron Emission Mammography Camera J.S. Huber, Member, IEEE, W.S. Choong, Member, IEEE, J. Wang, Member, IEEE, J.S. Maltz, Member, IEEE, J. Qi, Member, IEEE, E. Mandelli, Member,
More informationPoS(PhotoDet 2012)058
Absolute Photo Detection Efficiency measurement of Silicon PhotoMultipliers Vincent CHAUMAT 1, Cyril Bazin, Nicoleta Dinu, Véronique PUILL 1, Jean-François Vagnucci Laboratoire de l accélérateur Linéaire,
More informationHow to Evaluate and Compare Silicon Photomultiplier Sensors. October 2015
The Silicon Photomultiplier (SiPM) is a single-photon sensitive light sensor that combines performance characteristics that exceed those of a PMT, with the practical advantages of a solid state sensor.
More informationR & D for Aerogel RICH
1 R & D for Aerogel RICH Ichiro Adachi KEK Proto-Collaboration Meeting March 20, 2008 2 1 st Cherenkov Image detected by 3 hybrid avalanche photon detectors from a beam test About 3:00 AM TODAY Clear image
More information2448 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 57, NO. 5, OCTOBER 2010
2448 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 57, NO. 5, OCTOBER 2010 Development of an APD-Based PET Module and Preliminary Resolution Performance of an Experimental Prototype Gantry Jun Kataoka, Hidenori
More informationCHAPTER 11 HPD (Hybrid Photo-Detector)
CHAPTER 11 HPD (Hybrid Photo-Detector) HPD (Hybrid Photo-Detector) is a completely new photomultiplier tube that incorporates a semiconductor element in an evacuated electron tube. In HPD operation, photoelectrons
More informationDetectors for microscopy - CCDs, APDs and PMTs. Antonia Göhler. Nov 2014
Detectors for microscopy - CCDs, APDs and PMTs Antonia Göhler Nov 2014 Detectors/Sensors in general are devices that detect events or changes in quantities (intensities) and provide a corresponding output,
More informationAFBR-S4N44C013-DS100. Data Sheet. NUV-HD Silicon Photo Multiplier. Features. Description. Applications
Data Sheet AFBR-S4N44C013 Description The AFBR-S4N44C013 is a silicon photo multiplier (SiPM) used for ultra-sensitive precision measurement of single photons. The active area is 3.72 x 3.72 mm 2. High
More informationRed, Green, Blue (RGB) SiPMs
Silicon photomultipliers (SiPMs) from First Sensor are innovative solid-state silicon detectors with single photon sensitivity. SiPMs are a valid alternative to photomultiplier tubes. The main benefits
More informationApplication of Silicon Photomultipliers to Positron Emission Tomography
Annals of Biomedical Engineering, Vol. 39, No. 4, April 2011 (Ó 2011) pp. 1358 1377 DOI: 10.1007/s10439-011-0266-9 Application of Silicon Photomultipliers to Positron Emission Tomography EMILIE RONCALI
More informationCAEN. Electronic Instrumentation. CAEN Silicon Photomultiplier Kit
CAEN Tools for Discovery Electronic Instrumentation CAEN Silicon Photomultiplier Kit CAEN realized a modular development kit dedicated to Silicon Photomultipliers, representing the state-of-the art in
More informationDesign and Simulation of N-Substrate Reverse Type Ingaasp/Inp Avalanche Photodiode
International Refereed Journal of Engineering and Science (IRJES) ISSN (Online) 2319-183X, (Print) 2319-1821 Volume 2, Issue 8 (August 2013), PP.34-39 Design and Simulation of N-Substrate Reverse Type
More informationarxiv: v3 [astro-ph.im] 17 Jan 2017
A novel analog power supply for gain control of the Multi-Pixel Photon Counter (MPPC) Zhengwei Li a,, Congzhan Liu a, Yupeng Xu a, Bo Yan a,b, Yanguo Li a, Xuefeng Lu a, Xufang Li a, Shuo Zhang a,b, Zhi
More informationSPMMicro. SPMMicro. Low Cost High Gain APD. Low Cost High Gain APD. Page 1
SPMMicro Page 1 Overview Silicon Photomultiplier (SPM) Technology SensL s SPMMicro series is a High Gain APD provided in a variety of miniature, easy to use, and low cost packages. The SPMMicro detector
More informationDesign and development of compact readout electronics with silicon photomultiplier array for a compact imaging detector
University of Wollongong Research Online Faculty of Engineering and Information Sciences - Papers: Part A Faculty of Engineering and Information Sciences 2012 Design and development of compact readout
More informationA multipixel silicon APD with ultralow dark count rate at liquid nitrogen temperature
A multipixel silicon APD with ultralow dark count rate at liquid nitrogen temperature M. Akiba 1, K. Tsujino 1, K. Sato 2, and M. Sasaki 1 1 National Institute of Information and Communications Technology,
More informationNear Ultraviolet (NUV) SiPMs
Silicon photomultipliers (SiPMs) from First Sensor are innovative solid-state silicon detectors with single photon sensitivity. SiPMs are a valid alternative to photomultiplier tubes. The main benefits
More informationDevelopment of Photon Detectors at UC Davis Daniel Ferenc Eckart Lorenz Alvin Laille Physics Department, University of California Davis
Development of Photon Detectors at UC Davis Daniel Ferenc Eckart Lorenz Alvin Laille Physics Department, University of California Davis Work supported partly by DOE, National Nuclear Security Administration
More informationA Survey of Power Supply Techniques for Silicon Photo-Multiplier Biasing
A Survey of Power Supply Techniques for Silicon Photo-Multiplier Biasing R. Shukla 1, P. Rakshe 2, S. Lokhandwala 1, S. Dugad 1, P. Khandekar 2, C. Garde 2, S. Gupta 1 1 Tata Institute of Fundamental Research,
More informationNuclear Instruments and Methods in Physics Research A
Nuclear Instruments and Methods in Physics Research A ] (]]]]) ]]] ]]] Contents lists available at SciVerse ScienceDirect Nuclear Instruments and Methods in Physics Research A journal homepage: www.elsevier.com/locate/nima
More informationImportance of Precise Timing for Medical Diagnostic Devices
Importance of Precise Timing for Medical Diagnostic Devices M.C.S. Williams a,b, A. Zichichi a,b,c and the CERN-Bologna group. a CERN Geneva, Switzerland b INFN and Dipartimento di Fisica e Astronomia,
More informationDetector technology challenges for nuclear medicine and PET
Nuclear Instruments and Methods in Physics Research A 513 (2003) 1 7 Detector technology challenges for nuclear medicine and PET Paul K. Marsden Guy s and St. Thomas Clinical PET Centre, King s College
More informationDevelopment of Personal Dosimeter Using Electronic Dose Conversion Method
Proceedings of the Korean Nuclear Spring Meeting Gyeong ju, Korea, May 2003 Development of Personal Dosimeter Using Electronic Dose Conversion Method Wanno Lee, Bong Jae Lee, and Chang Woo Lee Korea Atomic
More informationA BaF2 calorimeter for Mu2e-II
A BaF2 calorimeter for Mu2e-II I. Sarra, on behalf of LNF group Università degli studi Guglielmo Marconi Laboratori Nazionali di Frascati NEWS General Meeting 218 13 March 218 Proposal (1) q This technological
More informationA test of silicon photomultipliers as readout for PET
Nuclear Instruments and Methods in Physics Research A 545 (25) 75 715 www.elsevier.com/locate/nima A test of silicon photomultipliers as readout for PET A.N. Otte a,, J. Barral b, B. Dolgoshein c, J. Hose
More informationWeek 9: Chap.13 Other Semiconductor Material
Week 9: Chap.13 Other Semiconductor Material Exam Other Semiconductors and Geometries -- Why --- CZT properties -- Silicon Structures --- CCD s Gamma ray Backgrounds The MIT Semiconductor Subway (of links
More informationMPPC and Liquid Xenon technologies from particle physics to medical imaging
CANADA S NATIONAL LABORATORY FOR PARTICLE AND NUCLEAR PHYSICS Owned and operated as a joint venture by a consortium of Canadian universities via a contribution through the National Research Council Canada
More informationElectronic Instrumentation for Radiation Detection Systems
Electronic Instrumentation for Radiation Detection Systems January 23, 2018 Joshua W. Cates, Ph.D. and Craig S. Levin, Ph.D. Course Outline Lecture Overview Brief Review of Radiation Detectors Detector
More informationMICRO PIXEL AVALANCHE PHOTODIODE AS ALTERNATIVE TO VACUUM PHOTOMULTIPLIER TUBES
MICRO PIXEL AVALANCHE PHOTODIODE AS ALTERNATIVE TO VACUUM PHOTOMULTIPLIER TUBES G.S. Ahmadov, Z.Y. Sadygov, F.I. Ahmadov National Nuclear Research Centre, Baku, Azerbaijan G.S. Ahmadov, Z.Y. Sadygov, Yu.N.
More informationCharacterizing a single photon detector
Michigan Technological University Digital Commons @ Michigan Tech Dissertations, Master's Theses and Master's Reports - Open Dissertations, Master's Theses and Master's Reports 2011 Characterizing a single
More informationMPPC versus MRS APD in two-phase Cryogenic Avalanche Detectors
MPPC versus MRS APD in two-phase Cryogenic Avalanche Detectors A. Bondar, a,b A. Buzulutskov, a,b A. Dolgov, b E. Shemyakina, a,b,* A. Sokolov, a,b a Budker Institute of Nuclear Physics SB RAS, Lavrentiev
More informationA Continuous Crystal Detector for TOF PET
1 A Continuous Crystal Detector for TOF PET T. Szczęśniak, Member, IEEE, M. Moszyński, Fellow, IEEE, Ł. Świderski, Member, IEEE, A. Nassalski, Member, IEEE, A. Syntfeld-KaŜuch, Member, IEEE, P. Ojala,
More informationSimulation of Algorithms for Pulse Timing in FPGAs
2007 IEEE Nuclear Science Symposium Conference Record M13-369 Simulation of Algorithms for Pulse Timing in FPGAs Michael D. Haselman, Member IEEE, Scott Hauck, Senior Member IEEE, Thomas K. Lewellen, Senior
More informationThe Light Amplifier Concept
The Light Amplifier Concept Daniel Ferenc 1 Eckart Lorenz 1,2 Daniel Kranich 1 Alvin Laille 1 (1) Physics Department, University of California Davis (2) Max Planck Institute, Munich Work supported partly
More information