A Novel Computerized Surgeon Machine Interface for Robot-Assisted Laser Phonomicrosurgery

Size: px
Start display at page:

Download "A Novel Computerized Surgeon Machine Interface for Robot-Assisted Laser Phonomicrosurgery"

Transcription

1 The Laryngoscope VC 2014 The American Laryngological, Rhinological and Otological Society, Inc. A Novel Computerized Surgeon Machine Interface for Robot-Assisted Laser Phonomicrosurgery Leonardo S. Mattos, PhD; Nikhil Deshpande, PhD; Giacinto Barresi, MS; Luca Guastini, MD; Giorgio Peretti, MD Objectives/Hypothesis: To introduce a novel computerized surgical system for improved usability, intuitiveness, accuracy, and controllability in robot-assisted laser phonomicrosurgery. Study Design: Pilot technology assessment. Methods: The novel system was developed involving a newly designed motorized laser micromanipulator, a touchscreen display, and a graphics stylus. The system allows the control of a CO 2 laser through interaction between the stylus and the live video of the surgical area. This empowers the stylus with the ability to have actual effect on the surgical site. Surgical enhancements afforded by this system were established through a pilot technology assessment using randomized trials comparing its performance with a state-of-the-art laser microsurgery system. Resident surgeons and medical students were chosen as subjects in performing sets of trajectory-following exercises. Image processing based techniques were used for an objective performance assessment. A System Usability Scale based questionnaire was used for the qualitative assessment. Results: The computerized interface demonstrated superiority in usability, accuracy, and controllability over the stateof-the-art system. Significant ease of use and learning experienced by the subjects were demonstrated by the usability score assigned to the two compared interfaces: computerized interface % versus state-of-the-art %. The objective analysis showed a significant enhancement in accuracy and controllability: computerized interface % versus state-ofthe-art %. Conclusions: The novel system significantly enhances the accuracy, usability, and controllability in laser phonomicrosurgery. The design provides an opportunity to improve the ergonomics and safety of current surgical setups. Key Words: Robotics, CO 2 laser, microlaryngeal surgery, surgeon interface. Level of Evidence: N/A Laryngoscope, 124: , 2014 INTRODUCTION Lasers increasingly occupy a dominant position in the treatment of pathological conditions in the larynx. They are often used as a precision tool to perform delicate ablation or cutting procedures. Figure 1 provides an overview of the surgeon interfaces in laser-based microsurgeries. One such example is the use of CO 2 lasers in laser phonomicrosurgery (LP), which involve a suite of complex otolaryngological surgical techniques for the treatment of minute abnormalities in the voice box. 1 The From the Department of Advanced Robotics, Istituto Italiano di Tecnologia (L.S.M., N.D., G.B.); and Department of Otorhinolaryngology, University of Genoa (L.G., G.P.), Genoa, Italy Editor s Note: This Manuscript was accepted for publication December 12, The research was carried out at the Italian Institute of Technology in Genoa. The experimental trials were conducted at the Otorhinolaryngology Department of San Martino Hospital in Genoa. This research has received funding from the European Union Seventh Framework Program FP7/ , Challenge 2 Cognitive Systems, Interaction, Robotics, under grant agreement lralp No The authors have no other funding, financial relationships, or conflicts of interest to disclose. Send correspondence to Leonardo S. Mattos, Department of Advanced Robotics, Istituto Italiano di Tecnologia, Via Morego 30, 16163, Genova, Italy. leonardo.demattos@iit.it DOI: /lary CO 2 surgical laser, coupled with a surgical microscope, is one of the main tools in LP. 2 Here, a mechanical micromanipulator is used to manually aim the laser beam from a typical operating distance of 400 mm. This requires the surgeon to have high psychomotor skills to overcome challenges exacerbated by poor operating ergonomics, suboptimal visualization, and difficult surgical site access. 3 Recent improvements to the traditional surgical setup include the addition of motorized laser scanning mirrors able to execute preprogrammed patterns. This functionality enables higher quality incisions and ablations in comparison with results achieved using a manually guided laser beam. 4,5 At least three devices of this type are commercially available: the Lumenis (Tel Aviv, Israel) AcuBlade, the KLS Martin (Jacksonville, FL) SoftScan, and the DEKA (Manchester, NH) HiScan. However, these systems do not actually enhance the surgical user interface, as they still require the surgeon to operate while looking through the microscope and position the scan patterns on the desired targets using the same traditional micromanipulators. Figure 2 shows the state-ofthe-art surgical setup in LP, using the AcuBlade system. It is recognized that interfaces (and human factors) play a major role in the success and quality of laser surgery. 6 This article describes the research toward improving the precision, controllability, safety, and ergonomics 1887

2 Fig. 1. User interfaces for laser microsurgery: (a) Diomax handpiece, a handheld optical scalpel commercialized by KLS Martin; (b) UniMax 2000 by Reliant Laser (Agawam, MA), a passive laser micromanipulator for surgical microscopes; (c) da Vinci System by Intuitive Surgical (Sunnyvale, CA), which has been experimentally outfitted with an optical fiber for laser surgeries; (d) the KU Leuven (Leuven, Belgium) writing interface for robot-aided laser surgery based on a graphics tablet; and (e) the new Robot-Assisted Laser Phonomicrosurgery interface based on a tablet personal computer. [Color figure can be viewed in the online issue, which is available at for laser microsurgery. This is done with a redesign of the laser micromanipulator and the implementation of a new computerized surgeon machine interface, the Robot-Assisted Laser Phonomicrosurgery (RALP) system. The concept allows surgeons to perform operations using a graphics stylus over live video of the surgical Fig. 2. The state-of-the-art surgical setup in laser phonomicrosurgery. [Color figure can be viewed in the online issue, which is available at

3 Fig. 3. The Robot-Assisted Laser Phonomicrosurgery concept. The graphics pen controls in real time the aiming and activation of the surgical laser by touching the live microscope video shown on the interactive display. [Color figure can be viewed in the online issue, which is available at site. This system eliminates hand eye coordination problems and makes full use of surgeons skills without requiring operational training on the equipment. In addition, it greatly improves laser aiming precision and consistency with a custom-designed motorized laser micromanipulator. This provides for a 4-lm positioning accuracy and a mm range at the typical 400-mm operating distance, allowing the physician to realize long cuts with unprecedented levels of precision and uniformity. This article introduces the new RALP system and presents a pilot technology assessment through its comparison with a state-of-the-art laser microsurgery system, the Lumenis AcuBlade. A combination of objective (quantitative performance) and subjective (self-reported) assessment methods is used with randomized trials for the analysis. MATERIALS AND METHODS The RALP Concept The RALP concept 7 is shown in Figure 3. Here, the stylus controls the aiming and activation of the surgical laser in real time. This is done through interaction with the touch-screen monitor, where live video of the surgical site, captured with the microscope camera, is displayed. A Zeiss (Oberkochen, Germany) microscope (OPMI Sensera /S7 with xenon lamp) is used in the setup with a Karl Storz (Tuttlingen, Germany) HD camera attached. Software processing and control allow precise positioning of the laser spot with the stylus tip when it touches the screen. Visual feedback to the surgeon is augmented with cues and safety fixtures, as seen in Figure 3. Furthermore, the system is configured to keep the surgeon in full control of the surgical operations, so CO 2 laser activation requires an enabling signal from a footswitch. This system eliminates hand eye coordination problems and augments surgeons fine manual skills through gesture scaling and magnification. Safety and controllability are also improved, in this case via assistive teleoperation and automation techniques. The new system allows for the definition of virtual safety regions wherein the laser is automatically switched off. Finally, it also offers the option of programming scan patterns for the laser beam, which is also done using the stylus over the live video display. These are automatically executed by the control software at user-definable speeds, ensuring precise repeatability and high consistency on operations that require a number of laser passes over the same area. Motorized Laser Micromanipulator The motorized laser micromanipulator was designed to be attached to the state-of-the-art LP equipment and is installed on its laser entry port. 8 Figure 4 shows the working principle. The final laser micromanipulator device is fully computercontrolled and able to realize accurate scanning motions at frequencies up to 200 Hz. The scanning hardware is mounted into a custom-designed case (Fig. 4) that guarantees the necessary optical alignment. Validation trials with this motorized micromanipulator demonstrated an average trajectory-following error of only mm, a 70% reduction compared to the traditional manual micromanipulator. Experiment Design for Pilot Technology Assessment To validate the novel surgeon machine interface, the laser manipulation in LP with RALP was compared with the state-ofthe-art digital AcuBlade laser micromanipulator in a pilot technology assessment. The experiments in this research were designed following well-established norms for ergonomic evaluation and device usability analysis. The experimental tasks included sets of trajectory-following exercises, that is, surgical maneuvers to follow preset random shapes, including straight lines, C-curves, and S-curves, which require dexterity and good laser controllability. They were stamped on small plaster blocks with randomized sequences of shapes and shape orientations. Figure 5 shows the experimental setup. The trials were performed at San Martino Hospital (Genoa, Italy). At this stage, the assessment did not consider efficacy or morbidity using the two interfaces and did not use tissue samples (ex vivo or in vitro) for the trials. The trials do not involve real surgical tasks, and considered only the initial learning steps with emulated maneuvers. These aspects are part of future work in our research. 1889

4 Fig. 4. The new motorized laser micromanipulator (Fast Steering Mirror [FSM]) system: (a) working principle; (b) photograph of the final FSM device. [Color figure can be viewed in the online issue, which is available at Subjects and Groups Resident surgeons and medical students were chosen as subjects for the assessment. After classifying the subjects based on gender and educational level, they were divided into two groups; the AcuBlade group was composed of 16 subjects (average age years; 6 male, 10 female); the RALP group was Fig. 5. Experimental setup, video snapshot from a trial, and details of the precision targets stamped on a plaster block. RALP 5 Robot- Assisted Laser Phonomicrosurgery. [Color figure can be viewed in the online issue, which is available at

5 TABLE I. SUS Questionnaire Items. 9 TABLE II. Imaging-Based Metrics and Unified Rating. 11 Item Question Metric Description 1 I think that I would like to use this system frequently. 2 I found the system unnecessarily complex. 3 I felt very confident using the system. 4 I found the various functions in this system were well integrated. 5 I thought the system was easy to use. 6 I thought there was too much inconsistency in this system. 7 I found the system very cumbersome to use. 8 I would imagine that most people would learn to use this system very quickly. 9 I think that I would need the support of a technical person to be able to use this system. 10 I needed to learn a lot of things before I could get going with this system. SUS 5 System Usability Scale. composed of 16 subjects (average age years; 6 male, 10 female). Subjects were randomly chosen to be part of one group or the other. The groups were balanced for gender, and as far as possible for educational level. The two groups were made to obtain data about the effects of the two different interfaces on the subjects. There was no subjective bias in the choice of the subjects. The only criteria for selection were that the subjects should have the background and motivation to learn to use the surgeon machine interfaces (medical students) as well as good knowledge of surgical procedures (resident surgeons). By only including subjects without experience on the specific devices tested, we avoided biases in their performances using a particular system over another. All the subjects being nonexperts allowed us to understand the effects that the two interfaces had on the usability, performance, and mental workload of the users. This allowed an unbiased comparison of the interfaces. In the AcuBlade group, the experimental setup with the surgical equipment had the same configuration as in the operating room (Fig. 5, left). The subjects used their nondominant hand for laser manipulation, as per current medical practice. In the RALP condition, the subjects employed the novel user interface, as shown in Figure 5 (top, right). The subjects performed the trials in two sessions of two target blocks each, with a break of at least 10 minutes between the two. A within-group experimental design was not considered at this stage; the goal was to focus the study on the comparison between the two interfaces. The implemented experimental design avoided the effect of the prior experience of the subjects with the other interface. AR Ratio of the area covered by the laser-traced and desired shapes. PR Ratio of the boundary lengths of the laser-traced and desired shapes. ARM Ratio of the aspect ratios of the laser-traced and desired shapes. OM Absolute difference in orientations of laser-traced and desired shapes. RMSE Root mean square error method for distance between the laser-traced and desired shapes. Error max Maximum value of the distance between the lasertraced and desired shapes. f f AR 1 f PR 1 f ARM 1 f OM 1 f RMSE 1 f Errormax. Execution Time taken in laser-tracing the given shape. time AR 5 area ratio; ARM 5 aspect ratio measure; Error max 5 maximum path error; f 5 unified rating; OM 5 orientation measure; PR 5 perimeter ratio; RMSE 5 path following error by root mean square error. Execution time. The correlation of the subjective selfevaluation and quantitative measurement of total time spent in performing the trials can indicate the effect the interfaces have on the mental workload of the subjects. For this purpose, a question regarding the perceived time spent performing the trial was included in the questionnaire. The actual value of the time spent was obtained from the objective measures. RESULTS Statistical tests were used for the analysis of significance in the data. Figure 6 shows representative trials. Subjective Analysis The detailed list of the comparisons of SUS scores is presented in Table III. The average SUS global scores demonstrated that the subjects assigned a much greater overall usability for the RALP at 83.96% (standard Analysis Methodology Subjective measures. User-centered interface evaluations allow assessment of usability, ergonomics, and mental workload. For the subjective evaluations, the usability analysis questionnaire was used, 9 based on the System Usability Scale (SUS). 10 SUS provided a global view of usability through the assessment of aspects such as ease of use, ease of learning, et cetera, 9 as listed in Table I. Objective measures. To establish a quantitative basis for the evaluation of the two conditions, imaging-based metrics were used, as introduced. 11 The metrics are listed in Table II. A weighted combination of the metrics, a unified rating, was used to classify the performance of each trial by each subject. 11 Fig. 6. Representative sample from the trials. Top: The desired and laser-traced shapes for the AcuBlade condition. Bottom: The desired and laser-traced shapes for the Robot-Assisted Laser Phonomicrosurgery (RALP) condition. [Color figure can be viewed in the online issue, which is available at

6 TABLE III. Comparison of Average SUS Scores Over the Complete Set of Trials, 16 Subjects per Interface, 48 Trials per Subject. Subscale RALP Condition, Mean, % AcuBlade Condition, Mean, % Improvement for RALP Over AcuBlade, % Global score Subscale Subscale Subscale Subscale Subscale Subscale Subscale Subscale Subscale Subscale The 10 subscales correspond to the 10 questions that form the SUS. RALP 5 Robot-Assisted Laser Phonomicrosurgery; SUS 5 System Usability Scale. deviation [SD] %) over the AcuBlade at 68.02% (SD %). Student t test shows P 5.004, affirming the statistical significance of the improvement. The subscales in the SUS further illustrate the different aspects of usability for the interfaces. Because the subscale scores are human-generated, they do not allow the assumptions of normality and homoscedasticity, precluding the use of Student t test. The following observations are derived from an explorative comparison of the subscales: 1. The users of RALP feel more confident (subscale 3) than the users of AcuBlade during the tasks. 2. The RALP interface is easier to use (subscale 5) than the AcuBlade interface. 3. The RALP interface is easier to learn (subscale 8) than the AcuBlade interface. TABLE IV. Comparison of Average Values of the Metrics and the Unified Rating Over the Complete Set of Trials, 16 Subjects per Interface, 48 Trials per Subject. Metric RALP Condition AcuBlade Condition Raw Rating Raw Rating t for Rating P for Rating AR PR* e-6 ARM* e-4 OM* e-4 RMSE* e-6 Error max * e-9 Unified rating, %* e-8 *Statistically significant subscales. AR 5 area ratio; ARM 5 aspect ratio measure; Error max 5 maximum path error; OM 5 orientation measure; PR 5 perimeter ratio; RALP 5 Robot- Assisted Laser Phonomicrosurgery; RMSE 5 path following error by root mean square error. 4. The users of RALP would require less support by an expert (subscale 9) than the users of AcuBlade. 5. The users of RALP would require learning fewer processes (subscale 10) than the users of AcuBlade. Conversely, the RALP system also performs slightly worse than the AcuBlade with respect to preferred frequency of usage (subscale 1) and inconsistency in the system operations (subscale 10). This is taken as important feedback toward simplifying the user interface presented to the surgeons during surgery. Objective Analysis Each individual laser-traced shape contributes to the quantitative analysis, giving 768 data points for each interface. Table IV summarizes the raw values of TABLE V. Observations Derived From Values of Specific Metrics for the Two Conditions. RALP Condition AcuBlade Condition Metric Mean SD Mean SD Observations PR The laser-traced shapes under the RALP condition conform to the length and thickness of the desired shapes significantly better than in the AcuBlade condition. ARM The RALP condition allows better shape conformance and positioning with the laser than the AcuBlade condition. OM, degrees The laser-traced shapes under the RALP condition are better aligned with the desired shapes than in the AcuBlade condition. RMSE, mm The RALP condition results in more precise laser control than the AcuBlade condition. Error max, mm The RALP condition permits safer control of the laser than the AcuBlade condition. It minimizes deviations from the desired trajectories. ARM 5 aspect ratio measure; Error max 5 maximum path error; OM 5 orientation measure; PR 5 perimeter ratio; RALP 5 Robot-Assisted Laser Phonomicrosurgery; RMSE 5 path following error by root mean square error; SD 5 standard deviation. 1892

7 the individual metrics and their contribution to the unified rating scores. Student t test analysis was performed on the unified rating. The RALP condition presents a significantly better performance at 90.02% (SD ) than the AcuBlade condition at 75.59% (SD ; P 5 9.3e-8). An analysis of the raw values of the individual metrics presents a clearer difference in the performance impact of the two interfaces, as summarized in Table V. Welch t test was used for the individual metrics due to the lack of homoscedasticity in the variances of their distributions. 12 The metrics in Table V are significant because they are directly related to the safety and efficiency of the interfaces. A high value for the orientation measure indicates that the majority of the laser-traced shape is not aligned with the desired shape. A high value for the root mean square error indicates imprecise laser control. Similar inferences are drawn from the perimeter and aspect ratio metrics. In the case of real surgery, this is highly undesired, impacting the safety and quality of procedures. On these parameters, the RALP system consistently shows closer to ideal values, much more so than the AcuBlade interface. Execution Time Analysis To understand the effect of the two interfaces on the perceived and actual times for trial completion, the Mann-Whitney U test was used, due to the lack of normality in the distributions. In subjective evaluations, there is a significant difference for the trial times in the two conditions (P 5 2e-4). The subjects in the AcuBlade condition estimated a longer time taken to complete the trials (mean seconds, SD ) than the subjects in the RALP condition (mean seconds, SD ). The quantitative analysis is contrary to the subjective evaluation. It shows a significant difference (P 5.002), but the RALP condition (mean seconds, SD ) takes a longer time to complete than the AcuBlade interface (mean seconds, SD ). This is an interesting result, as it allows inferring the level of mental workload necessitated in each condition. The inverse correlation between the perceived and actual times indicates the subjects in the RALP condition perceive themselves to have spent less time performing the trials than the subjects in the AcuBlade condition. The RALP condition seems to induce less mental workload than the AcuBlade condition. This is advantageous for the acceptance and usability of the new RALP system, indicating it can also help reduce mental demands imposed on the physician. CONCLUSION This article presented a novel medical robotic system, the RALP system, created to provide better precision, controllability, safety, and ergonomics for LP procedures. These objectives were achieved with the implementation of the RALP concept and creation of a custom motorized laser micromanipulator. The novel interface was compared Fig. 7. The unified rating and the System Usability Scale (SUS) global scores for the AcuBlade and the Robot-Assisted Laser Phonomicrosurgery (RALP) conditions. The means and the standard deviations are shown as percentages. [Color figure can be viewed in the online issue, which is available at www. laryngoscope.com.] to the current, state-of-the-art surgeon interface in LP. The combination of the subjective SUS-based evaluation and the quantitative metrics-based evaluation provide for a clear classification of the two examined surgical interfaces from the usability and performance perspectives. The results are summarized as follows: 1. In the RALP system, delicate maneuvers can be accurately performed from a computer monitor using a graphics stylus directly over the live video captured from the surgical microscope. This was shown to eliminate typical hand eye coordination problems and make better use of surgeons manual dexterity. 2. The subjective evaluation shows a superior overall usability for the RALP system over the AcuBlade interface. The global SUS score of 83.96% for the RALP system is significantly higher, by about 23%, than that for the AcuBlade interface at 68.02%. The advantages of the RALP interface are demonstrated through five different subscales of the SUS scale, which indicate that it is an easy to use and easy to learn interface. 3. The objective evaluation points to the clear advantage of the RALP interface over the AcuBlade interface. The analysis showed the significant superiority of RALP in terms of the efficiency and safety aspects of its performance. With a unified rating of 90.02%, it compared advantageously, about 19% better, against the AcuBlade interface with a unified rating of 75.59%. Figure 7 captures the values for the quantitative and qualitative analysis in a histogram plot. It is evident from the plot that the RALP system shows an overall higher performance and usability rating than the Acu- Blade condition. In the extension of this research, further studies are planned to better understand the usability of surgical interfaces through an integrated framework for evaluation. The technology advances presented here are now in the process of being evaluated jointly by surgeons and engineers using ex vivo pig larynxes. This will provide the groundwork for further evaluation with animal trials and future clinical application. Continuing interactions with the surgeons and other experts in the ergonomics 1893

8 field shall further help adapt the system architecture for the operating room. BIBLIOGRAPHY 1. Rosen J, Solazzo M, Hannaford B, Sinanan MN. Task decomposition of laparoscopic surgery for objective evaluation of surgical residents learning curve using hidden Markov model. Comput Aided Surg 2002;7: Solares CA, Strome M. Transoral robot-assisted CO2 laser supraglottic laryngectomy: experimental and clinical data. Laryngoscope 2007;117: Dagnino G, Mattos LS, Caldwell DG. New Software Tools for Enhanced Precision in Robot-Assisted Laser Phonomicrosurgery. In Proceedings of 34th International Conference of IEEE Engineering in Medicine and Biology Society. 2012: , IEEE, San Diego, CA, USA. 4. Remacle M, Hassan F, Cohen D, Lawson G, Delos M. New computerguided scanner for improving CO2 laser-assisted microincision. Eur Arch Otorhinolaryngol 2005;262: Remacle M, Lawson G, Nollevaux M, Delos M. Current state of scanning micromanipulator applications with the carbon dioxide laser. Ann Otol Rhinol Laryngol 2008;117: Wong Y-T, Finley CC, Giallo JF Jr, Buckmire RA. Novel CO2 laser robotic controller outperforms experienced laser operators in tasks of accuracy and performance repeatability. Laryngoscope 2011;121: Mattos LS, Dagnino G, Becattini G, Dellepiane M, Caldwell DG. A virtual scalpel system for computer-assisted laser microsurgery. In: Proceedings of IEEE/RSJ International Conference on Intelligent Robots and Systems. 2011: , IEEE, San Francisco, CA, USA. 8. Mattos LS, Dellepiane M, Caldwell DG. Next generation micromanipulator for computer-assisted laser phonomicrosurgery. In: Proceedings of 33rd International Conference of IEEE Engineering in Medicine and Biology Society. 2011: , IEEE, Boston, MA, USA. 9. Barresi G, Deshpande N, Mattos LS, et al. Comparative usability and performance evaluation of surgeon interfaces in laser phonomicrosurgery. In: Proceedings of IEEE/RSJ International Conference on Intelligent Robots and Systems. 2013: , IEEE, Tokyo, Japan. 10. Brooke J. SUS: a quick and dirty usability scale. In: Jordan PW, Thomas B, Weerdmeester BA, McClelland AL, eds. Usability Evaluation in Industry. 1996: , Taylor & Francis, London, UK. 11. Deshpande N, Mattos LS, Barresi G, et al. Imaging based metrics for performance assessment in laser phonomicrosurgery. In: Proceedings of International Conference on Robotics and Automation. 2013: , IEEE, Karlsruhe, Germany. 12. Welch BL. The generalization of Student s problem when several different population variances are involved. Biometrika 1947;34:

Conflict of Interest: None. ABSTRACT

Conflict of Interest: None. ABSTRACT Title: A Novel Computerized Surgeon-Machine Interface for Robot-Assisted Laser Phonomicrosurgery Authors: Leonardo S. Mattos 1, PhD; Nikhil Deshpande 1, PhD; Giacinto Barresi 1, MS; Luca Guastini 2, MD;

More information

Measurements of the Level of Surgical Expertise Using Flight Path Analysis from da Vinci Robotic Surgical System

Measurements of the Level of Surgical Expertise Using Flight Path Analysis from da Vinci Robotic Surgical System Measurements of the Level of Surgical Expertise Using Flight Path Analysis from da Vinci Robotic Surgical System Lawton Verner 1, Dmitry Oleynikov, MD 1, Stephen Holtmann 1, Hani Haider, Ph D 1, Leonid

More information

CO 2 Laser. AcuPulse. with SurgiTouch. Automation System. An Evolution in Automated Laser Surgery. Surgical: AcuPulse CO 2 Laser

CO 2 Laser. AcuPulse. with SurgiTouch. Automation System. An Evolution in Automated Laser Surgery. Surgical: AcuPulse CO 2 Laser AcuPulse CO 2 Laser with SurgiTouch Automation System An Evolution in Automated Laser Surgery www.lumenis.com www.surgical.lumenis.com SurgiTouch Automation System Automated and Robotic Laser Surgery The

More information

Evaluation of Haptic Virtual Fixtures in Psychomotor Skill Development for Robotic Surgical Training

Evaluation of Haptic Virtual Fixtures in Psychomotor Skill Development for Robotic Surgical Training Department of Electronics, Information and Bioengineering Neuroengineering and medical robotics Lab Evaluation of Haptic Virtual Fixtures in Psychomotor Skill Development for Robotic Surgical Training

More information

MECHANICAL DESIGN LEARNING ENVIRONMENTS BASED ON VIRTUAL REALITY TECHNOLOGIES

MECHANICAL DESIGN LEARNING ENVIRONMENTS BASED ON VIRTUAL REALITY TECHNOLOGIES INTERNATIONAL CONFERENCE ON ENGINEERING AND PRODUCT DESIGN EDUCATION 4 & 5 SEPTEMBER 2008, UNIVERSITAT POLITECNICA DE CATALUNYA, BARCELONA, SPAIN MECHANICAL DESIGN LEARNING ENVIRONMENTS BASED ON VIRTUAL

More information

Using Simulation to Design Control Strategies for Robotic No-Scar Surgery

Using Simulation to Design Control Strategies for Robotic No-Scar Surgery Using Simulation to Design Control Strategies for Robotic No-Scar Surgery Antonio DE DONNO 1, Florent NAGEOTTE, Philippe ZANNE, Laurent GOFFIN and Michel de MATHELIN LSIIT, University of Strasbourg/CNRS,

More information

da Vinci Skills Simulator

da Vinci Skills Simulator da Vinci Skills Simulator Introducing Simulation for the da Vinci Surgical System Skills Practice in an Immersive Virtual Environment Portable. Practical. Powerful. The da Vinci Skills Simulator contains

More information

Surgical robot simulation with BBZ console

Surgical robot simulation with BBZ console Review Article on Thoracic Surgery Surgical robot simulation with BBZ console Francesco Bovo 1, Giacomo De Rossi 2, Francesco Visentin 2,3 1 BBZ srl, Verona, Italy; 2 Department of Computer Science, Università

More information

Robots in the Field of Medicine

Robots in the Field of Medicine Robots in the Field of Medicine Austin Gillis and Peter Demirdjian Malden Catholic High School 1 Pioneers Robots in the Field of Medicine The use of robots in medicine is where it is today because of four

More information

Integre Pro Scan combines pattern scanning and multi-color photocoagulation in our unique all-in-one laser/slit lamp design.

Integre Pro Scan combines pattern scanning and multi-color photocoagulation in our unique all-in-one laser/slit lamp design. Integre Pro Scan combines pattern scanning and multi-color photocoagulation in our unique all-in-one laser/slit lamp design. Multi-color scanning photocoagulation takes on a new look. Integre Pro Scan

More information

Novel machine interface for scaled telesurgery

Novel machine interface for scaled telesurgery Novel machine interface for scaled telesurgery S. Clanton, D. Wang, Y. Matsuoka, D. Shelton, G. Stetten SPIE Medical Imaging, vol. 5367, pp. 697-704. San Diego, Feb. 2004. A Novel Machine Interface for

More information

VisuaLax: Visually Relaxing Augmented Reality Application Using Music and Visual Therapy

VisuaLax: Visually Relaxing Augmented Reality Application Using Music and Visual Therapy DOI: 10.7763/IPEDR. 2013. V63. 5 VisuaLax: Visually Relaxing Augmented Reality Application Using Music and Visual Therapy Jeremiah Francisco +, Benilda Eleonor Comendador, Angelito Concepcion Jr., Ron

More information

Effects of Geared Motor Characteristics on Tactile Perception of Tissue Stiffness

Effects of Geared Motor Characteristics on Tactile Perception of Tissue Stiffness Effects of Geared Motor Characteristics on Tactile Perception of Tissue Stiffness Jeff Longnion +, Jacob Rosen+, PhD, Mika Sinanan++, MD, PhD, Blake Hannaford+, PhD, ++ Department of Electrical Engineering,

More information

2. Publishable summary

2. Publishable summary 2. Publishable summary CogLaboration (Successful real World Human-Robot Collaboration: from the cognition of human-human collaboration to fluent human-robot collaboration) is a specific targeted research

More information

Medical robotics and Image Guided Therapy (IGT) Bogdan M. Maris, PhD Temporary Assistant Professor

Medical robotics and Image Guided Therapy (IGT) Bogdan M. Maris, PhD Temporary Assistant Professor Medical robotics and Image Guided Therapy (IGT) Bogdan M. Maris, PhD Temporary Assistant Professor E-mail bogdan.maris@univr.it Medical Robotics History, current and future applications Robots are Accurate

More information

Evaluation of Operative Imaging Techniques in Surgical Education

Evaluation of Operative Imaging Techniques in Surgical Education SCIENTIFIC PAPER Evaluation of Operative Imaging Techniques in Surgical Education Shanu N. Kothari, MD, Timothy J. Broderick, MD, Eric J. DeMaria, MD, Ronald C. Merrell, MD ABSTRACT Background: Certain

More information

Usability Evaluation of Multi- Touch-Displays for TMA Controller Working Positions

Usability Evaluation of Multi- Touch-Displays for TMA Controller Working Positions Sesar Innovation Days 2014 Usability Evaluation of Multi- Touch-Displays for TMA Controller Working Positions DLR German Aerospace Center, DFS German Air Navigation Services Maria Uebbing-Rumke, DLR Hejar

More information

577 nm: The Preferred Wavelength

577 nm: The Preferred Wavelength 577 nm: The Preferred Wavelength The confidence of precise performance Less light scattering for a confined beam and concise lesion which minimizes spot size and reduces thermal spread Peak absorption

More information

Lumenis Array LaserLink Pattern Scanning Laser Technology RETINA

Lumenis Array LaserLink Pattern Scanning Laser Technology RETINA Lumenis Array LaserLink Pattern Scanning Laser Technology RETINA Array LaserLink Pattern Scanning Laser Technology Pattern Scanning Laser can reduce photocoagulation treatment time by as much as 60% Pattern

More information

A Kinect-based 3D hand-gesture interface for 3D databases

A Kinect-based 3D hand-gesture interface for 3D databases A Kinect-based 3D hand-gesture interface for 3D databases Abstract. The use of natural interfaces improves significantly aspects related to human-computer interaction and consequently the productivity

More information

Medical Robotics. Part II: SURGICAL ROBOTICS

Medical Robotics. Part II: SURGICAL ROBOTICS 5 Medical Robotics Part II: SURGICAL ROBOTICS In the last decade, surgery and robotics have reached a maturity that has allowed them to be safely assimilated to create a new kind of operating room. This

More information

Objective Data Analysis for a PDA-Based Human-Robotic Interface*

Objective Data Analysis for a PDA-Based Human-Robotic Interface* Objective Data Analysis for a PDA-Based Human-Robotic Interface* Hande Kaymaz Keskinpala EECS Department Vanderbilt University Nashville, TN USA hande.kaymaz@vanderbilt.edu Abstract - This paper describes

More information

ERC: Engineering Research Center for Computer- Integrated Surgical Systems and Technology (NSF Grant # )

ERC: Engineering Research Center for Computer- Integrated Surgical Systems and Technology (NSF Grant # ) ERC: Engineering Research Center for Computer- Integrated Surgical Systems and Technology (NSF Grant #9731748) MARCIN BALICKI 1, and TIAN XIA 2 1,2 Johns Hopkins University, 3400 Charles St., Baltimore,

More information

Methods for Haptic Feedback in Teleoperated Robotic Surgery

Methods for Haptic Feedback in Teleoperated Robotic Surgery Young Group 5 1 Methods for Haptic Feedback in Teleoperated Robotic Surgery Paper Review Jessie Young Group 5: Haptic Interface for Surgical Manipulator System March 12, 2012 Paper Selection: A. M. Okamura.

More information

Telemanipulation and Telestration for Microsurgery Summary

Telemanipulation and Telestration for Microsurgery Summary Telemanipulation and Telestration for Microsurgery Summary Microsurgery presents an array of problems. For instance, current methodologies of Eye Surgery requires freehand manipulation of delicate structures

More information

Computer Assisted Medical Interventions

Computer Assisted Medical Interventions Outline Computer Assisted Medical Interventions Force control, collaborative manipulation and telemanipulation Bernard BAYLE Joint course University of Strasbourg, University of Houston, Telecom Paris

More information

Imagine your future lab. Designed using Virtual Reality and Computer Simulation

Imagine your future lab. Designed using Virtual Reality and Computer Simulation Imagine your future lab Designed using Virtual Reality and Computer Simulation Bio At Roche Healthcare Consulting our talented professionals are committed to optimising patient care. Our diverse range

More information

Design and control of a novel robotic microsurgical forceps for Transoral Laser Microsurgery

Design and control of a novel robotic microsurgical forceps for Transoral Laser Microsurgery Design and control of a novel robotic microsurgical forceps for Transoral Laser Microsurgery Manish Chauhan, Nikhil Deshpande, Giacinto Barresi, Claudio Pacchierotti, Domenico Prattichizzo, Darwin Caldwell,

More information

Concerning the Potential of Using Game-Based Virtual Environment in Children Therapy

Concerning the Potential of Using Game-Based Virtual Environment in Children Therapy Concerning the Potential of Using Game-Based Virtual Environment in Children Therapy Andrada David Ovidius University of Constanta Faculty of Mathematics and Informatics 124 Mamaia Bd., Constanta, 900527,

More information

Kinesthetic and Vibrotactile Haptic Feedback Improves the Performance of Laser Microsurgery

Kinesthetic and Vibrotactile Haptic Feedback Improves the Performance of Laser Microsurgery Kinesthetic and Vibrotactile Haptic Feedback Improves the Performance of Laser Microsurgery Loris Fichera 1, Claudio Pacchierotti 1, Emidio Olivieri 1, Domenico Prattichizzo 1,2, and Leonardo S. Mattos

More information

Markerless 3D Gesture-based Interaction for Handheld Augmented Reality Interfaces

Markerless 3D Gesture-based Interaction for Handheld Augmented Reality Interfaces Markerless 3D Gesture-based Interaction for Handheld Augmented Reality Interfaces Huidong Bai The HIT Lab NZ, University of Canterbury, Christchurch, 8041 New Zealand huidong.bai@pg.canterbury.ac.nz Lei

More information

Green Scan Laser PhotocoagulatorGYC-500 Vixi. Green Laser PhotocoagulatorGYC-500 US EDITION

Green Scan Laser PhotocoagulatorGYC-500 Vixi. Green Laser PhotocoagulatorGYC-500 US EDITION Green Scan Laser PhotocoagulatorGYC-500 Vixi Green Laser PhotocoagulatorGYC-500 US EDITION The Small, Incredibly Versatile Green Laser Photocoagulator The GYC-500 Vixi / GYC-500 is a solid state green

More information

3D Modelling Is Not For WIMPs Part II: Stylus/Mouse Clicks

3D Modelling Is Not For WIMPs Part II: Stylus/Mouse Clicks 3D Modelling Is Not For WIMPs Part II: Stylus/Mouse Clicks David Gauldie 1, Mark Wright 2, Ann Marie Shillito 3 1,3 Edinburgh College of Art 79 Grassmarket, Edinburgh EH1 2HJ d.gauldie@eca.ac.uk, a.m.shillito@eca.ac.uk

More information

L100 The ultimate CMM laser scanner combining productivity and accuracy NIKON METROLOGY I VISION BEYOND PRECISION

L100 The ultimate CMM laser scanner combining productivity and accuracy NIKON METROLOGY I VISION BEYOND PRECISION L100 The ultimate CMM laser scanner combining productivity and accuracy NIKON METROLOGY I VISION BEYOND PRECISION It s a Nikon... The L100 CMM laser scanner offers the best possible combination of speed,

More information

Surgeon-Tool Force/Torque Signatures - Evaluation of Surgical Skills in Minimally Invasive Surgery

Surgeon-Tool Force/Torque Signatures - Evaluation of Surgical Skills in Minimally Invasive Surgery # J. Rosen et al. Surgeon-Tool Force/Torque Signatures Surgeon-Tool Force/Torque Signatures - Evaluation of Surgical Skills in Minimally Invasive Surgery Jacob Rosen +, Ph.D., Mark MacFarlane *, M.D.,

More information

Discrimination of Virtual Haptic Textures Rendered with Different Update Rates

Discrimination of Virtual Haptic Textures Rendered with Different Update Rates Discrimination of Virtual Haptic Textures Rendered with Different Update Rates Seungmoon Choi and Hong Z. Tan Haptic Interface Research Laboratory Purdue University 465 Northwestern Avenue West Lafayette,

More information

Comparison of Wrap Around Screens and HMDs on a Driver s Response to an Unexpected Pedestrian Crossing Using Simulator Vehicle Parameters

Comparison of Wrap Around Screens and HMDs on a Driver s Response to an Unexpected Pedestrian Crossing Using Simulator Vehicle Parameters University of Iowa Iowa Research Online Driving Assessment Conference 2017 Driving Assessment Conference Jun 28th, 12:00 AM Comparison of Wrap Around Screens and HMDs on a Driver s Response to an Unexpected

More information

Distributed Robotics: Building an environment for digital cooperation. Artificial Intelligence series

Distributed Robotics: Building an environment for digital cooperation. Artificial Intelligence series Distributed Robotics: Building an environment for digital cooperation Artificial Intelligence series Distributed Robotics March 2018 02 From programmable machines to intelligent agents Robots, from the

More information

VIRTUAL ASSISTIVE ROBOTS FOR PLAY, LEARNING, AND COGNITIVE DEVELOPMENT

VIRTUAL ASSISTIVE ROBOTS FOR PLAY, LEARNING, AND COGNITIVE DEVELOPMENT 3-59 Corbett Hall University of Alberta Edmonton, AB T6G 2G4 Ph: (780) 492-5422 Fx: (780) 492-1696 Email: atlab@ualberta.ca VIRTUAL ASSISTIVE ROBOTS FOR PLAY, LEARNING, AND COGNITIVE DEVELOPMENT Mengliao

More information

Evaluation of an Enhanced Human-Robot Interface

Evaluation of an Enhanced Human-Robot Interface Evaluation of an Enhanced Human-Robot Carlotta A. Johnson Julie A. Adams Kazuhiko Kawamura Center for Intelligent Systems Center for Intelligent Systems Center for Intelligent Systems Vanderbilt University

More information

Introduction to HCI. CS4HC3 / SE4HC3/ SE6DO3 Fall Instructor: Kevin Browne

Introduction to HCI. CS4HC3 / SE4HC3/ SE6DO3 Fall Instructor: Kevin Browne Introduction to HCI CS4HC3 / SE4HC3/ SE6DO3 Fall 2011 Instructor: Kevin Browne brownek@mcmaster.ca Slide content is based heavily on Chapter 1 of the textbook: Designing the User Interface: Strategies

More information

CSE 165: 3D User Interaction. Lecture #14: 3D UI Design

CSE 165: 3D User Interaction. Lecture #14: 3D UI Design CSE 165: 3D User Interaction Lecture #14: 3D UI Design 2 Announcements Homework 3 due tomorrow 2pm Monday: midterm discussion Next Thursday: midterm exam 3D UI Design Strategies 3 4 Thus far 3DUI hardware

More information

MEL 80 Excimer Laser. When you want to see better performance

MEL 80 Excimer Laser. When you want to see better performance MEL 80 Excimer Laser When you want to see better performance Reward your practice Invest in the very latest refractive excimer technology! The MEL 80 makes vision correction even safer, more patient-friendly

More information

Robots for Medicine and Personal Assistance. Guest lecturer: Ron Alterovitz

Robots for Medicine and Personal Assistance. Guest lecturer: Ron Alterovitz Robots for Medicine and Personal Assistance Guest lecturer: Ron Alterovitz Growth of Robotics Industry Worldwide $70 $56 Market Size (Billions) $42 $28 $14 $0 1995 2000 2005 2010 2015 2020 2025 Source:

More information

Nonuniform multi level crossing for signal reconstruction

Nonuniform multi level crossing for signal reconstruction 6 Nonuniform multi level crossing for signal reconstruction 6.1 Introduction In recent years, there has been considerable interest in level crossing algorithms for sampling continuous time signals. Driven

More information

Physical Affordances of Check-in Stations for Museum Exhibits

Physical Affordances of Check-in Stations for Museum Exhibits Physical Affordances of Check-in Stations for Museum Exhibits Tilman Dingler tilman.dingler@vis.unistuttgart.de Benjamin Steeb benjamin@jsteeb.de Stefan Schneegass stefan.schneegass@vis.unistuttgart.de

More information

UGEO H60. Performance in Style. Features

UGEO H60. Performance in Style. Features UGEO H60 Performance in Style The UGEO H60 implements superior performance with new design principles of simplicity and lightness. Its 10.1" touchscreen improves usability while its 18.5" LED monitor enhances

More information

CS 889 Advanced Topics in Human- Computer Interaction. Experimental Methods in HCI

CS 889 Advanced Topics in Human- Computer Interaction. Experimental Methods in HCI CS 889 Advanced Topics in Human- Computer Interaction Experimental Methods in HCI Overview A brief overview of HCI Experimental Methods overview Goals of this course Syllabus and course details HCI at

More information

Cutaneous Feedback of Fingertip Deformation and Vibration for Palpation in Robotic Surgery

Cutaneous Feedback of Fingertip Deformation and Vibration for Palpation in Robotic Surgery Cutaneous Feedback of Fingertip Deformation and Vibration for Palpation in Robotic Surgery Claudio Pacchierotti Domenico Prattichizzo Katherine J. Kuchenbecker Motivation Despite its expected clinical

More information

Virtual Grasping Using a Data Glove

Virtual Grasping Using a Data Glove Virtual Grasping Using a Data Glove By: Rachel Smith Supervised By: Dr. Kay Robbins 3/25/2005 University of Texas at San Antonio Motivation Navigation in 3D worlds is awkward using traditional mouse Direct

More information

CUSUM Analysis of Learning Curves for the Head-Mounted Microscope in Phonomicrosurgery

CUSUM Analysis of Learning Curves for the Head-Mounted Microscope in Phonomicrosurgery The Laryngoscope VC 2015 The American Laryngological, Rhinological and Otological Society, Inc. CUSUM Analysis of Learning Curves for the Head-Mounted Microscope in Phonomicrosurgery Ting Chen, MD, PhD;

More information

CSE 190: 3D User Interaction. Lecture #17: 3D UI Evaluation Jürgen P. Schulze, Ph.D.

CSE 190: 3D User Interaction. Lecture #17: 3D UI Evaluation Jürgen P. Schulze, Ph.D. CSE 190: 3D User Interaction Lecture #17: 3D UI Evaluation Jürgen P. Schulze, Ph.D. 2 Announcements Final Exam Tuesday, March 19 th, 11:30am-2:30pm, CSE 2154 Sid s office hours in lab 260 this week CAPE

More information

Generic noise criterion curves for sensitive equipment

Generic noise criterion curves for sensitive equipment Generic noise criterion curves for sensitive equipment M. L Gendreau Colin Gordon & Associates, P. O. Box 39, San Bruno, CA 966, USA michael.gendreau@colingordon.com Electron beam-based instruments are

More information

Development and Validation of Virtual Driving Simulator for the Spinal Injury Patient

Development and Validation of Virtual Driving Simulator for the Spinal Injury Patient CYBERPSYCHOLOGY & BEHAVIOR Volume 5, Number 2, 2002 Mary Ann Liebert, Inc. Development and Validation of Virtual Driving Simulator for the Spinal Injury Patient JEONG H. KU, M.S., 1 DONG P. JANG, Ph.D.,

More information

Wearable Haptic Feedback Actuators for Training in Robotic Surgery

Wearable Haptic Feedback Actuators for Training in Robotic Surgery Wearable Haptic Feedback Actuators for Training in Robotic Surgery NSF Summer Undergraduate Fellowship in Sensor Technologies Joshua Fernandez (Mechanical Eng.) University of Maryland Baltimore County

More information

Simendo laparoscopy. product information

Simendo laparoscopy. product information Simendo laparoscopy product information Simendo laparoscopy The Simendo laparoscopy simulator is designed for all laparoscopic specialties, such as general surgery, gynaecology en urology. The simulator

More information

Optical Marionette: Graphical Manipulation of Human s Walking Direction

Optical Marionette: Graphical Manipulation of Human s Walking Direction Optical Marionette: Graphical Manipulation of Human s Walking Direction Akira Ishii, Ippei Suzuki, Shinji Sakamoto, Keita Kanai Kazuki Takazawa, Hiraku Doi, Yoichi Ochiai (Digital Nature Group, University

More information

Evaluation of Visuo-haptic Feedback in a 3D Touch Panel Interface

Evaluation of Visuo-haptic Feedback in a 3D Touch Panel Interface Evaluation of Visuo-haptic Feedback in a 3D Touch Panel Interface Xu Zhao Saitama University 255 Shimo-Okubo, Sakura-ku, Saitama City, Japan sheldonzhaox@is.ics.saitamau.ac.jp Takehiro Niikura The University

More information

SMart wearable Robotic Teleoperated surgery

SMart wearable Robotic Teleoperated surgery SMart wearable Robotic Teleoperated surgery This project has received funding from the European Union s Horizon 2020 research and innovation programme under grant agreement No 732515 Context Minimally

More information

Here I present more details about the methods of the experiments which are. described in the main text, and describe two additional examinations which

Here I present more details about the methods of the experiments which are. described in the main text, and describe two additional examinations which Supplementary Note Here I present more details about the methods of the experiments which are described in the main text, and describe two additional examinations which assessed DF s proprioceptive performance

More information

Prospective Teleautonomy For EOD Operations

Prospective Teleautonomy For EOD Operations Perception and task guidance Perceived world model & intent Prospective Teleautonomy For EOD Operations Prof. Seth Teller Electrical Engineering and Computer Science Department Computer Science and Artificial

More information

Haptic control in a virtual environment

Haptic control in a virtual environment Haptic control in a virtual environment Gerard de Ruig (0555781) Lourens Visscher (0554498) Lydia van Well (0566644) September 10, 2010 Introduction With modern technological advancements it is entirely

More information

Second Announcement Call for Participation. (Evaluation Criteria added)

Second Announcement Call for Participation. (Evaluation Criteria added) Second Announcement Call for Participation 2 nd International Scanner Contest (ISC) (Evaluation Criteria added) P. Hufnagl 1, T. Schrader 1, 2, M.G. Rojo 3, A. Laurinavicius 4, G. Kayser 5, Y. Yagi 6 1

More information

Interacting within Virtual Worlds (based on talks by Greg Welch and Mark Mine)

Interacting within Virtual Worlds (based on talks by Greg Welch and Mark Mine) Interacting within Virtual Worlds (based on talks by Greg Welch and Mark Mine) Presentation Working in a virtual world Interaction principles Interaction examples Why VR in the First Place? Direct perception

More information

Tobii T60XL Eye Tracker. Widescreen eye tracking for efficient testing of large media

Tobii T60XL Eye Tracker. Widescreen eye tracking for efficient testing of large media Tobii T60XL Eye Tracker Tobii T60XL Eye Tracker Widescreen eye tracking for efficient testing of large media Present large and high resolution media: display double-page spreads, package design, TV, video

More information

Yellow Scan Laser Photocoagulator YLC-500 Vixi. Yellow Laser Photocoagulator YLC-500

Yellow Scan Laser Photocoagulator YLC-500 Vixi. Yellow Laser Photocoagulator YLC-500 Yellow Scan Laser Photocoagulator YLC-500 Vixi Yellow Laser Photocoagulator YLC-500 A 577 nm Yellow Laser with Multiple Scan Patterns The YLC-500 Vixi / YLC-500 is a yellow laser using the innovative OPSL

More information

3D Data Navigation via Natural User Interfaces

3D Data Navigation via Natural User Interfaces 3D Data Navigation via Natural User Interfaces Francisco R. Ortega PhD Candidate and GAANN Fellow Co-Advisors: Dr. Rishe and Dr. Barreto Committee Members: Dr. Raju, Dr. Clarke and Dr. Zeng GAANN Fellowship

More information

Haptic Camera Manipulation: Extending the Camera In Hand Metaphor

Haptic Camera Manipulation: Extending the Camera In Hand Metaphor Haptic Camera Manipulation: Extending the Camera In Hand Metaphor Joan De Boeck, Karin Coninx Expertise Center for Digital Media Limburgs Universitair Centrum Wetenschapspark 2, B-3590 Diepenbeek, Belgium

More information

Design Science Research Methods. Prof. Dr. Roel Wieringa University of Twente, The Netherlands

Design Science Research Methods. Prof. Dr. Roel Wieringa University of Twente, The Netherlands Design Science Research Methods Prof. Dr. Roel Wieringa University of Twente, The Netherlands www.cs.utwente.nl/~roelw UFPE 26 sept 2016 R.J. Wieringa 1 Research methodology accross the disciplines Do

More information

The Effect of Haptic Feedback on Basic Social Interaction within Shared Virtual Environments

The Effect of Haptic Feedback on Basic Social Interaction within Shared Virtual Environments The Effect of Haptic Feedback on Basic Social Interaction within Shared Virtual Environments Elias Giannopoulos 1, Victor Eslava 2, María Oyarzabal 2, Teresa Hierro 2, Laura González 2, Manuel Ferre 2,

More information

Green Scan Laser PhotocoagulatorGYC-500 Vixi. Green Laser PhotocoagulatorGYC-500

Green Scan Laser PhotocoagulatorGYC-500 Vixi. Green Laser PhotocoagulatorGYC-500 Green Scan Laser PhotocoagulatorGYC-500 Vixi Green Laser PhotocoagulatorGYC-500 The Small, Incredibly Versatile Green Laser Photocoagulator The GYC-500 Vixi / GYC-500 is a solid state green laser that

More information

Haptic Feedback in Laparoscopic and Robotic Surgery

Haptic Feedback in Laparoscopic and Robotic Surgery Haptic Feedback in Laparoscopic and Robotic Surgery Dr. Warren Grundfest Professor Bioengineering, Electrical Engineering & Surgery UCLA, Los Angeles, California Acknowledgment This Presentation & Research

More information

Technical Requirements of a Social Networking Platform for Senior Citizens

Technical Requirements of a Social Networking Platform for Senior Citizens Technical Requirements of a Social Networking Platform for Senior Citizens Hans Demski Helmholtz Zentrum München Institute for Biological and Medical Imaging WG MEDIS Medical Information Systems MIE2012

More information

Active microwave moisture sensors advanced solutions for the pulp and paper industry

Active microwave moisture sensors advanced solutions for the pulp and paper industry Active microwave moisture sensors advanced solutions for the pulp and paper industry AMS an integrated system of sensors installed in various phases of the pulp and paper manufacturing process to measure

More information

Differences in Fitts Law Task Performance Based on Environment Scaling

Differences in Fitts Law Task Performance Based on Environment Scaling Differences in Fitts Law Task Performance Based on Environment Scaling Gregory S. Lee and Bhavani Thuraisingham Department of Computer Science University of Texas at Dallas 800 West Campbell Road Richardson,

More information

Autonomous Surgical Robotics

Autonomous Surgical Robotics Nicolás Pérez de Olaguer Santamaría Autonomous Surgical Robotics 1 / 29 MIN Faculty Department of Informatics Autonomous Surgical Robotics Nicolás Pérez de Olaguer Santamaría University of Hamburg Faculty

More information

Optic Flow Based Skill Learning for A Humanoid to Trap, Approach to, and Pass a Ball

Optic Flow Based Skill Learning for A Humanoid to Trap, Approach to, and Pass a Ball Optic Flow Based Skill Learning for A Humanoid to Trap, Approach to, and Pass a Ball Masaki Ogino 1, Masaaki Kikuchi 1, Jun ichiro Ooga 1, Masahiro Aono 1 and Minoru Asada 1,2 1 Dept. of Adaptive Machine

More information

Current Status and Future of Medical Virtual Reality

Current Status and Future of Medical Virtual Reality 2011.08.16 Medical VR Current Status and Future of Medical Virtual Reality Naoto KUME, Ph.D. Assistant Professor of Kyoto University Hospital 1. History of Medical Virtual Reality Virtual reality (VR)

More information

Measurement of Microscopic Three-dimensional Profiles with High Accuracy and Simple Operation

Measurement of Microscopic Three-dimensional Profiles with High Accuracy and Simple Operation 238 Hitachi Review Vol. 65 (2016), No. 7 Featured Articles Measurement of Microscopic Three-dimensional Profiles with High Accuracy and Simple Operation AFM5500M Scanning Probe Microscope Satoshi Hasumura

More information

Green Scan Laser PhotocoagulatorGYC-500 Vixi. Green Laser PhotocoagulatorGYC-500

Green Scan Laser PhotocoagulatorGYC-500 Vixi. Green Laser PhotocoagulatorGYC-500 Green Scan Laser PhotocoagulatorGYC-500 Vixi Green Laser PhotocoagulatorGYC-500 The Small, Incredibly Versatile Green Laser Photocoagulator The GYC-500 Vixi / GYC-500 is a solid state green laser that

More information

Application of 3D Terrain Representation System for Highway Landscape Design

Application of 3D Terrain Representation System for Highway Landscape Design Application of 3D Terrain Representation System for Highway Landscape Design Koji Makanae Miyagi University, Japan Nashwan Dawood Teesside University, UK Abstract In recent years, mixed or/and augmented

More information

Automatic Detection and Segmentation of Robot-Assisted Surgical Motions

Automatic Detection and Segmentation of Robot-Assisted Surgical Motions Automatic Detection and Segmentation of Robot-Assisted Surgical Motions Henry C. Lin 1, Izhak Shafran 2, Todd E. Murphy, Allison M. Okamura, David D. Yuh, and Gregory D. Hager 1 1 Department of Computer

More information

Resona 6 Premium Ultrasound System

Resona 6 Premium Ultrasound System The system Resona 6 is the newly developed, unique result of the mergence of leading companies Mindray Bio-medical Electronics Co. Ltd. and ZONARE Medical Systems, Inc.. By additions to the core competencies

More information

The Effect of Display Type and Video Game Type on Visual Fatigue and Mental Workload

The Effect of Display Type and Video Game Type on Visual Fatigue and Mental Workload Proceedings of the 2010 International Conference on Industrial Engineering and Operations Management Dhaka, Bangladesh, January 9 10, 2010 The Effect of Display Type and Video Game Type on Visual Fatigue

More information

Improving Depth Perception in Medical AR

Improving Depth Perception in Medical AR Improving Depth Perception in Medical AR A Virtual Vision Panel to the Inside of the Patient Christoph Bichlmeier 1, Tobias Sielhorst 1, Sandro M. Heining 2, Nassir Navab 1 1 Chair for Computer Aided Medical

More information

System Two Making your sugical outcomes brighter

System Two Making your sugical outcomes brighter System Two System Two Making your sugical outcomes brighter When you re performing surgery, you re guided by what you see. That s why it s absolutely critical that you can visualize your work with complete

More information

Development of an Intuitive Interface for PC Mouse Operation Based on Both Arms Gesture

Development of an Intuitive Interface for PC Mouse Operation Based on Both Arms Gesture Development of an Intuitive Interface for PC Mouse Operation Based on Both Arms Gesture Nobuaki Nakazawa 1*, Toshikazu Matsui 1, Yusaku Fujii 2 1 Faculty of Science and Technology, Gunma University, 29-1

More information

A Three-Dimensional Evaluation of Body Representation Change of Human Upper Limb Focused on Sense of Ownership and Sense of Agency

A Three-Dimensional Evaluation of Body Representation Change of Human Upper Limb Focused on Sense of Ownership and Sense of Agency A Three-Dimensional Evaluation of Body Representation Change of Human Upper Limb Focused on Sense of Ownership and Sense of Agency Shunsuke Hamasaki, Atsushi Yamashita and Hajime Asama Department of Precision

More information

Small Occupancy Robotic Mechanisms for Endoscopic Surgery

Small Occupancy Robotic Mechanisms for Endoscopic Surgery Small Occupancy Robotic Mechanisms for Endoscopic Surgery Yuki Kobayashi, Shingo Chiyoda, Kouichi Watabe, Masafumi Okada, and Yoshihiko Nakamura Department of Mechano-Informatics, The University of Tokyo,

More information

endoscope for observing vocal fold

endoscope for observing vocal fold NAOSITE: Nagasaki University's Ac Title Author(s) Citation High-speed digital imaging system w endoscope for observing vocal fold Kaneko, Kenichi; Watanabe, Takeshi; Takahashi, Haruo Acta medica Nagasakiensia,

More information

Operating Lights. marled with VariLUX. Outstanding LED light of the 2 nd generation

Operating Lights. marled with VariLUX. Outstanding LED light of the 2 nd generation Operating Lights marled with VariLUX Outstanding LED light of the 2 nd generation The operation starts. The light is turned on. It is easy to set. Color temperature can be adjusted in accordance with specific

More information

A TELE-INSTRUCTION SYSTEM FOR ULTRASOUND PROBE OPERATION BASED ON SHARED AR TECHNOLOGY

A TELE-INSTRUCTION SYSTEM FOR ULTRASOUND PROBE OPERATION BASED ON SHARED AR TECHNOLOGY A TELE-INSTRUCTION SYSTEM FOR ULTRASOUND PROBE OPERATION BASED ON SHARED AR TECHNOLOGY T. Suenaga 1, M. Nambu 1, T. Kuroda 2, O. Oshiro 2, T. Tamura 1, K. Chihara 2 1 National Institute for Longevity Sciences,

More information

The Perception of Optical Flow in Driving Simulators

The Perception of Optical Flow in Driving Simulators University of Iowa Iowa Research Online Driving Assessment Conference 2009 Driving Assessment Conference Jun 23rd, 12:00 AM The Perception of Optical Flow in Driving Simulators Zhishuai Yin Northeastern

More information

Haptic Virtual Fixtures for Robot-Assisted Manipulation

Haptic Virtual Fixtures for Robot-Assisted Manipulation Haptic Virtual Fixtures for Robot-Assisted Manipulation Jake J. Abbott, Panadda Marayong, and Allison M. Okamura Department of Mechanical Engineering, The Johns Hopkins University {jake.abbott, pmarayong,

More information

Digital Reality TM changes everything

Digital Reality TM changes everything F E B R U A R Y 2 0 1 8 Digital Reality TM changes everything Step into the future What are we talking about? Virtual Reality VR is an entirely digital world that completely immerses the user in an environment

More information

Leading in Desktop SEM Imaging and Analysis

Leading in Desktop SEM Imaging and Analysis Leading in Desktop SEM Imaging and Analysis Fast. Outstanding. Reliable SEM imaging and analysis. The Phenom: World s Fastest Scanning Electron Microscope With its market-leading Phenom desktop Scanning

More information

On The Correlation of Image Size to System Accuracy in Automatic Fingerprint Identification Systems

On The Correlation of Image Size to System Accuracy in Automatic Fingerprint Identification Systems On The Correlation of Image Size to System Accuracy in Automatic Fingerprint Identification Systems J.K. Schneider, C. E. Richardson, F.W. Kiefer, and Venu Govindaraju Ultra-Scan Corporation, 4240 Ridge

More information

Virtual and Augmented Reality techniques embedded and based on a Operative Microscope. Training for Neurosurgery.

Virtual and Augmented Reality techniques embedded and based on a Operative Microscope. Training for Neurosurgery. Virtual and Augmented Reality techniques embedded and based on a Operative Microscope. Training for Neurosurgery. 1 M. Aschke 1, M.Ciucci 1,J.Raczkowsky 1, R.Wirtz 2, H. Wörn 1 1 IPR, Institute for Process

More information

Job Description. Commitment: Must be available to work full-time hours, M-F for weeks beginning Summer of 2018.

Job Description. Commitment: Must be available to work full-time hours, M-F for weeks beginning Summer of 2018. Research Intern Director of Research We are seeking a summer intern to support the team to develop prototype 3D sensing systems based on state-of-the-art sensing technologies along with computer vision

More information

The pattern scanning multi-color photocoagulator that puts you in control

The pattern scanning multi-color photocoagulator that puts you in control The pattern scanning multi-color photocoagulator that puts you in control RETINAL PHOTOCOAGULATION LASER TRABECULOPLASTY LASER IRIDOTOMY Helping the world see clearly 2 INTEGRE PRO SCAN FROM ELLEX Transforming

More information