Temperature transmitter itemp RTD DIN rail TMT 127

Similar documents
Temperature head transmitter itemp TC TMT 188

Temperature Transmitter itemp RTD DIN rail TMT 127

â itemp Pt100 TMT127 Technical information Pt100 transmitter for DIN rail mounting for high accuracy temperature monitoring

â Ó itemp TC TMT128 Technical information DIN rail mounted Temperature Transmitter for Thermocouples (TC)

Temperature transmitter itemp PCP DIN rail TMT 121

Temperature transmitter with local display RIT 261

itemp HART DIN rail TMT112

0 2 1 â Ó. itemp HART TMT182. Technical Information

0 2 1 â Ó. itemp HART TMT182. Technical Information

Process display RIA 261

Technical Information itemp TMT180

Temperature head transmitter itemp PCP TMT 181

Temperature head transmitter itemp PCP TMT 181

itemp PA TMT184 Technical information

Temperature Transmitters - DIN Rail Mounted

TT2780. Temperature head transmitter For resistance thermometers Pt100, settable using a PC, for installation in a sensor head Form B, BUZ, BUZ-H

Level Limit Switch nivotester FTL 325 P

Technical Information Barrier RB223

Temperature Transmitters - Head Mounted

RTD Temperature Sensor omnigrad T -TSM 480, TR 480. Hygienic RTD sensor with Pt 100, class A TSM 480 with electronics programmable via PC

Temperature Transmitters TMT 136, TMT 137

omnigrad T - TSM 480, TR 480. Technical Information. RTD Temperature Sensor Hygienic design TSM 480 with electronics programmable via PC

3 Ex i Isolators. Temperature Transmitter without Output 0/ ma, with Limit Value Contact (Field Circuit Ex i) Type 9182/.

3 Ex i Isolators. Temperature Transmitter with Output 0/ ma, without Limit Value Contact (Field Circuit Ex i) Type 9182/.0-5.

Analogue Temperature Transmitter Model T24.10, PC configurable, Head Mounting

RMA42. Technical Information

PROGRAMMABLE FIELD TEMPERATURE TRANSMITTER

TR 104, TR 104-Ex Rail mounted Temperature Transmitters

Technical Information RMA42

Advanced Paperless Recorder eco-graph a

Digital Temperature Transmitters Model T12.10, Universally Programmable, Head Mounting Model T12.30, Universally Programmable, Rail Mounting

Temperature measurement TMD 832 head transmitter

Remote Switching Unit nivotester FTL 320

TF 02 / TF 02-Ex. Head mounted temperature transmitters, FOUNDATION Fieldbus (H1), Pt 100 (RTD), thermocouples, 1 or 2 independent channels

Isolators A3/1. Transmitter Supply Unit Field Circuit Ex i Series 9160

Assembly. Front view. Power Rail

Digital temperature transmitter Model T15.H, head mounting version Model T15.R, rail mounting version

Digital temperature transmitter for resistance sensors Model T15.H, head mounting version Model T15.R, rail mounting version

TMT122 DIN rail itemp HART

09746E00. ATEX / IECEx NEC 505 NEC 506 NEC 500 Class I Class I Class II Class III Zone Zone Division Ex i

Isolators A3/1. Transmitter Supply Unit Field Circuit Ex i Series 9160 Rev. F

FieldIT Temperature Transmitter, rail mounted TH102/TH102-Ex

Miniature resistance thermometer Explosion-protected version Model TR34, thread-mounted

Analog Input Module HART Ex i / I.S. Inputs, Channels Type 9461/

Head mounted Temperature Transmitter TF02/TF02-Ex

06528E00. ATEX / IECEx NEC 505 NEC 506 NEC 500

Isolators A3/1.

Conductivity Measuring Cells for High-Temperature Applications ConduMax H CLS 16

OEM miniature resistance thermometer Models TR31-3 and TR31-K, thread-mounted

Isolators A3/1.

itemp HART TMT142 Technical Information

Paperless Recorder PR20

Transmitter. Configuration Code TM01 Series 440 Programmable RTD Temperature Transmitter. Application Areas. Features and Benefits

Isolators A3/1. Transmitter Supply Unit Field Circuit Ex i Series

SINEAX V 624 Programmable Temperature Transmitter for RTD and TC inputs

Rail mounted Temperature Transmitter TH102/TH102-Ex

OPTITEMP TT 20 Technical Datasheet

JUMO dtrans T01 HART / T01T HART / T01 Junior / T01 Ex / T01 HART Ex Programmable 2-wire transmitter

3 Ex i Isolators. Switching Repeater with Power Relay (250 V / 4 A) Type 9170/ Type 9170/

Subject to changes 11/2003 Page 1/7

Rosemount 148 Temperature Transmitter

C330 / C330X PC-Programmable Universal, 2-wire Transmitter

Temperature Measurement

Radar Level Measurement micropilot FMR 130 for stilling wells

Binary Output Series 9175

Binary Output without Power Supply Series 9176

Pressure transmitter AKS 32 and AKS 33

SINEAX TI 807 Passive DC signal isolator. SINEAX TI 807 Passive DC signal isolator

Isolators A3/1. Vibration Transducer Supply Unit Series

Soliphant T FTM20, FTM21

SINEAX TI 807 Passive DC signal isolator

itemp TMT82 Technical Information Dual-input temperature transmitter with HART protocol

itemp HART TMT142 Technical Information

Isolators A3/1. Temperature Transmitter Field Circuit Non-Ex i Series

ph / ORP Measurement mypro CPM 431

PRODUCT SPECIFICATION AND MODEL SELECTION GUIDE

Inductive sensor slot-type SI2-K08-AP7

Isolating Repeater Input Series 9163

JUMO dtrans T01 HART / T01T HART / T01 Junior / T01 Ex / T01 HART Ex Programmable 2-wire transmitter

Advanced Temperature Transmitter

Assembly. Front view. Parametrization interface. Power Bus

Temperature Measurement Transmitters for mounting in sensor head

Type CP-S, CP-C & CP-A Switch mode

08/07/2015

Resistance thermometer Model TR30, compact design

Digital Input Module NAMUR Ex i / I.S. Inputs, 16 Channels for Zone 1 / Div. 1 Series 9470/22

Memosens New technology for contactless, inductive, digital transmission of ph signals

Resistance thermometer Model TR30, compact design

3RF A A 0 2. Switching function. Control voltage Operating voltage with heat sink 10 = 10.5 A. A = Zero-point switching

Assembly. Front view. LED yellow: Relay output. LED red: LB/SC

Rosemount 148 Temperature Transmitter

E+H Model PT TMT184. Temperature head transmitter with PROFIBUS-PA -interface. Supply and digital communication using PROFIBUS-PA

Two-Wire Programmable Transmitter for:

FieldIT Temperature Transmitter, head mounted TH02/TH02-Ex

RMC621. Technical Information. Flow and Energy Manager Universal flow and energy computer for gases, liquids and steam

Universal Intelligent 2-wire DIN Rail Transmitters

Level Limit Switch nivotester FTL 370, FTL 372

OPTITEMP TT 10 C/R Technical Datasheet

Temperature Measurement Transmitters for mounting in sensor head

Model 144H NEW. PC-Programmable Temperature Transmitter. Product Data Sheet

Transcription:

Technical information TI 095R/24/ae Nr. 510 06165 Temperature transmitter itemp RTD DIN rail TMT 127 Pt100 transmitter for an economical, high accuracy temperature monitoring. Fixed range measurement, for installation on a DIN rail. Application Economical and technical alternative to direct wiring to DCS or PLC Fixed adjusted temperature Din-rail mounted transmitter for converting various input signals into a scalable 4 to 20 ma analog output signal 2-wire transmitter for a linear temperature proportional analog output Features and benefits High accuracy: 0.08 % of span Breakdown information in event of sensor break or sensor short-circuit enables a quick maintenance intervention Galvanic isolation 2 kv (from the sensor input to the output) Long term stability: < 0.05 %/year Electromagnetic compatibility to IEC 61326 for use in noisy environments Approvals: FM, CSA and ATEX for high safety standards UL recognized component to UL 3111-1 GL German Lloyd marine approval

Function and system design Measuring principle Measuring system Electronic monitoring and conversion of input signals in industrial temperature measurement. The itemp RTD DIN rail TMT 127 DIN rail temperature transmitter is a 2-wire transmitter with analog output, measuring input for Pt100 in 2, 3, or 4-wire connection. Input values Measured variable Measuring range Input type Temperature Depending on the application, different measuring ranges can be ordered (see Product structure ). Input Designation Measuring range limits Min. span Resistance thermometer (RTD) Pt100 as per IEC 751-328 to 1562 F (-200 to 850 C) 18 F (10 C) Type of connection: 2, 3 or 4-wire connection Cable resistance: sensor cable resistance of max. 40 per cable Sensor current: 0.6 ma Output values Output signal Analog 4 to 20 ma Breakdown information Breakdown information to NAMUR NE 43 Breakdown information is created when the measuring information is invalid or not present anymore and gives a complete listing of all errors occuring in the measuring system. Under ranging Standard 3.8 Over ranging Standard 20.5 Signal (ma) Sensor break; sensor short circuit low Sensor break; sensor short circuit high To NAMUR NE 43 3.6 To NAMUR NE 43 21.5 Source impedance Transmission behavior Galvanic isolation Induced current requirement Current limitation Switch-on delay Max. (V Power supply - 12V) / 0.022 A (current output) e. g. (24 V - 12 V)/0.022 A = 545.5 Temperature linear U = 2 kv AC (input/output) 3.5 ma 23 ma 4 s (during switch-on procedure I a 3.8 ma) 2

Power supply Electrical connection Temperature transmitter terminal assignment Supply voltage Residual ripple U b = 12 to 35 V, reverse polarity protection Permitted residual ripple U ss 3 V at U b 15 V, f max. = 1 khz Accuracy Response time Reference operating conditions Measuring error 1 s Calibration temperature: + 73.4 F ± 9 F (+ 23 C ± 5 C) Designation Accuracy 1 Resistance thermometer RTD Pt100 0.36 F (0.2 C) or 0.08% 1) % refer to the set span. The highest value is valid. Influence of supply voltage Influence of ambient temperature (temperature drift) Influence of load Long-term stability ±0.01%/V deviation from 24 V Percentages refer to the full scale value. Pt100 resistance thermometer: T d = ±(8.3 ppm/ F * (full scale value + 200) + 27.8 ppm/ F * of set measuring range) * = deviation of ambient temperature from the reference operating condition (+73.4 F ± 9 F). ± 0.02%/100 Percentages refer to the full scale value. 0.18 F/year ( 0.1 C/year) or 0.05%/year Values under reference operating conditions. % refer to the set span. The highest value is valid. 3

Installation conditions Installation instructions Orientation No limit Environmental conditions Ambient temperature -40 to 185 F (-40 to +85 C) for Ex-area, see Ex-certification or control drawing Storage temperature -40 to 212 F (-40 to +100 C) Climate class as per IEC 60 654-1, Class C Ingress protection NEMA 1 (IP 20) Shock and vibration resistance 4g / 2 to 150 Hz as per IEC 60 068-2-6 Electromagnetic compatibility (EMC) CE Electromagnetic Compatibility Compliance The device meets all requirements listed under IEC 61326 Amendment 1, 1998 and NAMUR NE 21. This recommendation is an uniform and practical way of determining whether the devices used in laboratory and process control are immune to interference with an objective to increase its functional safety. Discharge of static electricity IEC 61000-4-2 6 kv cont., 8 kv air Electromagnetic fields IEC 61000-4-3 80 to 1000 Hz 10 V/m Burst (signal) IEC 61000-4-4 2 kv Transient voltage IEC 61000-4-5 1 kv unsym. / 0.5 kv sym. HF coupling IEC 61000-4-6 0.15 to 80 MHz 10 V Condensation allowed 4

Mechanical construction Design, dimensions Values in inch (mm) Weight approx. 3.2 oz (90 g) Materials Terminals Housing: PC/ABS, UL 94V0 Keyed plug-in screw terminals, core size max. 16 AWG solid, or strands with ferrules Display and operating system Display elements Operating elements Illuminated yellow LED (0.08 in, 2 mm) signals device operation. There are no operating elements available on the transmitter to prevent from manipulation. Certificates and approvals CE-Mark Hazardous area approvals GL The measurement system fulfils the requirements demanded by the EU regulations. Endress+Hauser acknowledges successful unit testing by adding the CE mark. FM IS, Class I, Div. 1+2, Group A, B, C, D CSA IS, Class I, Div. 1+2, Group A, B, C, D ATEX II 2(1)G EEx ia IIC T4/T5/T6 ATEX II 3G EEx na IIC T4/T5/T6 Ship building approval (Germanischer Lloyd) UL Recognized component to UL 3111-1 Other standards and guidelines IEC 60529: Degrees of protection by housing (IP-Code) IEC 61010: Safety requirements for electrical measurement, control and laboratory instrumentation. IEC 61326: Electromagnetic compatibility (EMC requirements) NAMUR Standardization association for measurement and control in chemical and pharmaceutical industries. (www.namur.de) NEMA Standardization association for the electrical industry 5

Ordering information Product structure Temperature transmitter itemp RTD DIN rail TMT 127 for temperature measurement with Pt100; Analog output 4 to 20 ma, 2-wire techn.; Galv. isol., fail. mode to NAMUR NE 43; 22.5 mm wide, for 35 mm top hat DIN rail according to EN 50022; UL recognized, ship building approval GL Certification A Version for non hazardous areas, UL recognized, ship building approval GL B ATEX II2(1)G EEx ia IIC T4/T5/T6 C FM IS, Class I, Div.1+2, Group A,B,C,D D CSA IS, Class I, Div.1+2, Group A,B,C,D E ATEX II3G EEx na IIC T4/T5/T6 Transmitter connection 2 Connection RTD 2-wire 3 Connection RTD 3-wire 4 Connection RTD 4-wire Temperature sensor 1 Configuration Pt100-200 to 850 C, min. sp. 10 C Configuration range 1 LA Range -40 to 140 F LB Range -40 to 200 F MA Range -20 to 400 F NA Range 0 to 100 F NB Range 0 to 200 F NC Range 0 to 300 F ND Range 0 to 500 F NE Range 0 to 750 F NF Range 0 to 900 F NH Range 0 to 1200 F AA Range -100 to 50 C BB Range -50 to 200 C FG Range 0 to 150 C Model A B K Standard model Works calibration certificate (6 test points) Standard model, North America region TMT127 1 Order code (complete) 1 more ranges available on request Accessories No accessories are required for this device. Documentation System information Temperature measurement (SI 008R/24/ae) Operating manual itemp RTD/TC DIN rail TMT 127/128 (BA 157R/24/ae) Short operating manual itemp RTD/TC DIN rail TMT 127/128 (KA 140R/09/a3) Additional documentation for use in explosion-hazardous area: ATEX II2(1)G: XA 013R/09/a3 ATEX II3G: XA 018R/09/a3 Control Drawings: FM IS, NI 510 05062 CSA IS, NI 510 05575 6

For application and selection assistance, in the U.S. call 888-ENDRESS For total support of your installed base, 24 hours a day, in the U.S. call 800-642-8737 Visit us on our web site, www.us.endress.com United States Canada Mexico Endress+Hauser, Inc. 2350 Endress Place Greenwood, IN 46143 Phone: (317) 535-7138 888-ENDRESS FAX: (317) 535-8498 Endress+Hauser Canada Ltd. 1440 Graham s Lane Unit 1, Burlington ON, L7S 1W3 Phone: (905) 681-9292 800-668-3199 FAX: (905) 681-9444 Endress+Hauser Paseo del Pedregal No. 610 Col, Jardines del Pedregal 01900, Mexico D. F. Mexico Phone: (525) 568-2405 FAX: (525) 568-7459 TI 095R/24/ae/04.03 51006165 FM+SGML 6.0 ProMoDo