MAKING SENSE OF SLIT LAMP SERVICING

Similar documents
OMM300. Inverted Metallurgical Microscope

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com

User Manual. Trinocular Metallurgical Microscope. MicroscopeNet.com

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

Zoom Stereo Microscope NYMCS-360 Instruction Manual

Ophthalmic Slit-lamp Microscope (SLM 1ER) Operation. Instruction

TABLE OF CONTENTS. Safety notes i. Care and Maintenance. ii. 1. Components Illustration Installation of Components.. 4

STEINDORFF NYMC Polarizing Microscope

User Manual. Trinocular Infinity Compound LED Microscope. MicroscopeNet.com

Therefore, all descriptions and illustrations in this instruction manual, including all specifications are subject to change without notice.

OM FL400. Reflected Light Fluorescence Microscope. Instruction Manual. Please read instructions carefully before using microscope.

Compound Microscopes Instruction Manual

ILFORD SPORTSVIEW PROJECTOR INSTRUCTION BOOK

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com

SWIFT SERIES M2252DGL MICROSCOPE

MANUAL EXM-150 MICROSCOPE SERIES

Wide Angle Ophthalmoscope Instructions

User instructions Compound laboratory microscope

MANUAL MICROSCOPE SERIES. 73 Mall Drive, Commack, NY (P) (F)

STEINDORFF Digital LCD Biological Microscope

ML7520 ML7530 DIOPTER ADJUSTMENT RING BINOCULAR BODY, INCLINED 30. (a) Field Iris Control Lever. (c) Filter Slots EYEPIECES, KHW10X

Manual for BMS E1 eplan series, compound microscope

CLEANING FOR BETTER OBSERVATION AND PHOTOMICROGRAPHY

User instructions Compound laboratory microscope

Using a Compound Light Microscope

MANUAL EXC-350 MICROSCOPE SERIES

Microscopes. A guide to use, general Maintenance, and repair tailored to the Olympus CX-21 microscope

STEINDORFF METALLURGICAL MICROSCOPE. NYMCS-620 Instruction Manual

STEINDORFF NYMC C. Comparison Microscope. Operating Instructions. A. Features and Functions

Preface. General Requirements for Safety. Precautions

Swift M2250 Series Microscope Care and Use Manual

1. Features Calibration Compensator Lock Button Circular bubble Line-of-sight...10

TEKSCOPE MICROSCOPE. Models N2 Series USER S MANUAL

Instruction Manual T Binocular Acromat Research Scope T Trinocular Acromat Research Scope

Swift M2252DGL Series Microscope Use and Care Manual

Using a Microscope. Year Group: BVSc1 + Document number: CSL_L07

Biological Microscope

User instructions Metallurgical inverted microscope

Basic Microscopy for Plant Biology

Cleaning the cutter blade

User manual.

Nikon Ti-E Microscope Manual. Rightmire Hall Ohio State University. Director: Tony Brown Rightmire

Keeler Catalogue. All Keeler products are designed to offer the ultimate in professional performance.

Lecture 2 Slit lamp Biomicroscope

eflex 75x/300x Digital Microscope with Flexible Neck User Guide

User instructions Metallurgical microscope

Eyepieces KHW10X. Diopter Adjustment Ring. Binocular Body Inclined 30. Binocular Clamp Screw. Analyzer control Lever. Reflected Light Illuminator

Beautifully Crafted Slit Lamps

Components of the Microscope

Microbiology Laboratory 2

Basic Microscopy. OBJECTIVES After completing this exercise, you should be able to do the following:

Swift M10 Series Microscope Use and Care Manual

AGES 10+ INSTRUCTION MANUAL. 800x Power Advanced Microscope Biological Experiments Gear. x 2 NOT INCLUDED

Installation Instructions SKU# , Rated Voltage V~ 50Hz

Keeler Direct Ophthalmoscopes

Match the microscope structures given in the left column with the statements in the right column that identify or describe them.

11/10/2015. Haag Streit Topcon Zeiss Kowa Add On Systems- OIS/Escalon and Others. The Original Design. Photo Slit lamp Systems. Who Makes Them?

VIPER SPOTTING SCOPE

XSP-100XSP-100SM/ OF BIOLOGICAL MICROSCOPE OPERATION MANUAL READ THIS MANUAL BEFORE USING THE MICROSCOPE

The Razor HD 65mm & 85mm Spotting Scopes

User instructions Transmitted light laboratory microscope

Marine Invertebrate Zoology Microscope Introduction

X-Mag. High-Magnification Workstation User s Guide Revision A

Laboratory Introduction

INSTRUCTION MANUAL. 150x Power Student Microscope Biological Experiments Gear 1.5V WARNING: WARNING: NOT INCLUDED

MANUAL EXC-400 MICROSCOPE SERIES. 73 Mall Drive, Commack, NY (P) (F)

User instructions Transmitted light phase contrast microscope

MVL-30 Video Light User Guide

Solid Sample Holder Accessory

LED Slit Lamp XL-1. Operations Manual. The contents of this Instruction Manual may be changed without prior notice. D06-S e Rev1.

OPERATION MANUAL MBM 207M MANUAL FOLDER

USER MANUAL FOR VISIBLE SPECTROPHOTOMETER

OPERATING AND MAINTENANCE MANUAL FLEXI MULTI-MOULDER (HAND FEED)

Clarinet/Bass Clarinet Care

AMCON LENSOMETER USER S MANUAL EQ Fax

AutoSeal FD 2006IL / FE 2006IL

O P E R A T I N G I N S T R U C T I O N S

Riverside. Windward Bay EntertaInment Wall System Assembly Instructions. Made In Viet Nam. Right Pier. Bridge & Back Panel.

Rabbit Series Laser Engraving Machine HX3040. User s Manual. 1 Jinan King Rabbit Technology Development Co. Ltd.

Ⅰ. Application. Ⅱ. Main Technical Specification. AS1 Biological microscope,which is widely used in medical and

CLEMENT CLARKE OPHTHALMIC

The light microscope

CALIBRATION OF MICROSCOPE EYEPIECE GRATICULE

Model S-520 Coin Counter / Sorter Operating Manual

MANUAL EXI-300 INVERTED MICROSCOPE SERIES. 73 Mall Drive, Commack, NY (P) (F)

O5: Lenses and the refractor telescope

SAFETY AND OPERATING MANUAL

Shutter Installation Instructions

Photometer Maintenance Guide

Swift M3600, M3700 Series Digital & Compound Microscope

GLASS GLOBE MOBILE LINEAR CHANDELIER

Therefore, all descriptions and illustrations in this instruction manual, including all specifications are subject to change without notice.

Epi-LED FL Fluorescence Microscope Instruction Manual

Beautifully Crafted Slit Lamps.

Bullnose Plane 05P42.01

Replacing the Reciprocator on an SWF Multi-head.

# R8 Mill Instruction Manual. Please read and understand all instructions before using this tool.

HEINE Direct Ophthalmoscopes

pc420 classic manual plus

Screwfeeder Troubleshooting Guide

Transcription:

MAKING SENSE OF SLIT LAMP SERVICING

HOW TO LOOK AFTER AND CARE FOR A SLIT LAMP The slit lamp is an essential and oftenused diagnostic instrument in ophthalmology. It provides illumination and magnification for the examination of many structures of the anterior segment. With complementary lenses, it is also used to examine the chamber angle and a signifi cant part of the retina. Its name derives from the fact that a narrow slit of light is used to illuminate the various structures being examined. By following these simple suggestions, you can ensure that a slit lamp performs optimally and remains functional for longer. LOCATION Place the slit lamp where it is easily accessible to both staff and patients, some of whom may have physical disabilities. An electrical outlet should be available nearby and the power cord should not be in the path of staff or patients. The slit lamp should not be exposed to excessive temperature extremes, such as those produced by direct sunlight or air conditioning. The slit lamp should be kept in a dry environment since there could be fungal growth (mould or mildew) on the optical components if they are exposed to humidity (combined heat and moisture). REPLACING THE BULB When handling or replacing a bulb, take care not to leave fingerprints on the bulb. Oil from your fi ngers can create hot spots on the bulb which will reduce its life. As a rule, handle bulbs with paper tissue or with cotton gloves. Check that you replace the bulb housing in the right position; otherwise the quality of the slit beam is compromised. Adjusting the position of the housing may correct a distorted beam. CLEANING The slit lamp should be cleaned weekly, at a minimum, or more often if in a dusty environment. The slit lamp housing should be cleaned with a cloth that has been slightly dampened with water. No other liquids or corrosive agents should be used. The exposed surfaces of the eyepiece optics (1) and the objective lens (2) should be cleaned using a soft optical dust brush. If, after being dusted, they still need additional cleaning, the lenses should then be wiped carefully with a lens cleaning cloth or with cotton swabs and lens cleaning solution. SPARE PARTS Spare bulbs and fuses should be kept within easy reach in order to avoid delays in patient care. The minimum recommended stock of bulbs and fuses is two of each per slit lamp. When a part is used, it should be restocked immediately. 1 2 2 SYMPHONY SLIT LAMP by KEELER www.keeler.co.uk

OPERATIONAL CHECKS The following functions should be checked weekly. The hospital s maintenance team or the service agent should be called if any problems are noticed during these checks. 12 Brightness control: should noticeably vary the bulb s brightness. Table top movement: should move up and down freely. 14 6 Chin rest adjustment (3): should move up and down freely. Joystick (4): should provide smooth motion up and down, forward and backward, and left and right. Slit controls: should smoothly vary the slit width (5), length (6), and inclination (6). Illumination rotation arm (7): should move smoothly and lock into position with the locking screw (8). Microscope rotation arm (9): should move smoothly and lock into position with the locking screw (10). Illumination tilting latch (11): should vary the illumination angle in stages. 4 1 13 9 11 5 3 7 8 10 Filter changing knob (12): should change the filters. Magnifi cation lever (13): should switch the magnification. The mechanism just behind the objectives that adjusts the pupillary distance (14) should move smoothly. 18 www.keeler.co.uk SYMPHONY SLIT LAMP by KEELER 3

OTHER TIPS FOR CARE AND MAINTENANCE If the clinic is subject to voltage fluctuations, the slit lamp should be plugged into a voltage stabiliser. When examining several patients in a row, the illumination should be maintained at a low level rather than switching it off between patients and then on again for each patient. This prolongs the bulb s life. 4 Moving the slit lamp should be avoided when the bulb is hot because the hot filament is more likely to break. 18 When not in use, the slit lamp should be covered with its plastic dust cover. If not provided, a simple cover can be made out of cloth the thicker/denser the better. If the slit lamp is stored in an environment prone to humidity, keep a sachet of silica gel drying agent or fungicidal (anti-mould) pellets within the dust cover, or use a dehumidifi er in the room. The forward and backward, and left and right movements of the slit lamp rely on the joystick (4), a rod () connecting the two geared wheels (16), and the two rails (17) which support the wheels. These mechanical devices may seize up and affect the smooth operation of the slit lamp. If this is the case, apply a light oil spray, such as WD40, to a piece of paper tissue and use the tissue to wipe the rod, the pad under the joystick (18), the wheels, and the rails. This should solve the problem. Oil should never be sprayed directly onto these parts. The information contained within this manual must not be reproduced in whole or part without the manufacturer s prior written approval. As part of our policy for continued product development we the manufaucturer reserves the right to make changes to specifications and other information contained in this document without prior notice. 17 16 Copyright Keeler Limited 2013. Published in the UK 2013. 4 SYMPHONY SLIT LAMP by KEELER www.keeler.co.uk

ACKNOWLEDGEMENTS/SOURCE MATERIAL Copyright 2011 Ismael Cordero, ORBIS International. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium for non-profit purposes, provided the original work is properly cited. 6 SYMPHONY SLIT LAMP by KEELER www.keeler.co.uk