AxioVision 4.5 Brightfield Image Capture Procedure

Similar documents
Olympus IX71 Microscope and DP71 Camera Instructions

Zeiss Axioskop II. The AIF's "routine" light microscope. (Installed 8/24/04)AxioCam installed July 11th 2005

Simplified Instructions: Zeiss Brightfield Microscope S1000

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions

Zeiss Deconvolution Microscope: A Quick Guide

Characterization Microscope Nikon LV150

Brightfield Microscopy and Image Acquisition on Spotcam1. by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE

Zeiss Axio Imager.A1 manual

Olympus Time-lapse Microscope Basic operation

Nikon E800 Microscope. Operating Instructions

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING

Using the Nikon TE2000 Inverted Microscope

Motorized Axio Observer Start-up instructions

Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope

Zeiss AxioImager.Z2 Brightfield Protocol

Nikon E800 Operating Instructions.

Olympus Digital Microscope Camera (DP70) checklist

Leica DB LB Research microscope and Studo Lite Imaging software

Zeiss LSM 880 Protocol

Nikon E800 Operating Instructions.

The ideal K-12 science microscope solution. User Guide. for use with the Nova5000

QUICKSTART GUIDE: WIDEFIELD HWF1 Zeiss Cell Observer Live Cell Imaging System (HAMMERSMITH, L BLOCK, ROOM 314) Imperial College London

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

EPIFLUORESCENCE &/OR BRIGHTFIELD MICROSCOPY

INSTRUCTIONS FOR COURSE WORK 4 (AxioVert) Instructor: Anne Vaahtokari (MIU) 1. Purpose of the work

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Takeoff Guide Tutorial to get started with the AxioVision Imaging System (Based on Release 4.8 June 2009)

Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software

Nikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol

Zeiss Axioplan 2 imaging microscope and Axiovision software

Zeiss AxioObserver with ApoTome

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7

OPERATING INSTRUCTIONS

Things to check before start-up.

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.

Nikon TE300 Eclipse Wide-Field Microscope

LSM 710 Confocal Microscope Standard Operation Protocol

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com

QUICKSTART GUIDE: WIDEFIELD WF3 Zeiss Cell Observer Live Cell Imaging System (SAF, ROOM 409) Imperial College London

Digital Design and Communication Teaching (DiDACT) University of Sheffield Department of Landscape. Adobe Photoshop CS5 INTRODUCTION WORKSHOPS

LSM 780 Confocal Microscope Standard Operation Protocol

Using the AmScope Microscope Cameras

Cell Biology and Bioimaging Core

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL

Leica TCS SP8 Quick Start Guide

Photography Protocol. Software: ZEN Blue 2012

User Guide for TWAIN / DirectX interface for GRYPHAX USB 3.0 cameras

User manual for Nikon Elements software

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL

Olympus xcellence Software - basic user guide

Zeiss Axiovert 135 Fluorescence Microscope Quick Guide / Operations Manual (v. 1.0 February 09)

Instructions for Making On-Line Reservations for Microscopes in NB11-204

LSM 800 Confocal Microscope Standard Operation Protocol

COMPACT MANUAL FOR GI USERS OF THE JEM 1400 FLASH BEGINNERS (For internal use only) Gray means additional information at the end of this mini-manual

DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands

Zeiss 880 Training Notes Zen 2.3

Nikon Ti-E Microscope Manual. Rightmire Hall Ohio State University. Director: Tony Brown Rightmire

Zeiss LSM 780 Protocol

Leica TCS SP8 Quick Start Guide

Widefield 1. Switching on

Everest System / Slidebook Operating Procedures

Optika ISview. Image acquisition and processing software. Instruction Manual

Leica SP8 TCS Users Manual

USING LEICA AS LASER MICRODISSECTION (LMD6000) MICROSCOPE Written By Jungim Hur

Zeiss 780 Training Notes

Microscopy from Carl Zeiss

Simplified Instructions: Olympus Widefield Microscope S1230

Operating Instructions for Zeiss LSM 510

ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide

LEICA TCS SP5 AOBS TANDEM USER MANUAL

Training Guide for Leica SP8 Confocal/Multiphoton Microscope

Guide to Confocal 5. Starting session

Introduction to 3D Printing. Activity 1: Design a keychain using computer-aided design software

Image Pro Ultra. Tel:

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION

Guide to AmScope SZM Series Dissecting Scope for Sample Imaging with MU Series Camera

ISCapture User Guide. advanced CCD imaging. Opticstar

Apple Photos Quick Start Guide

Brief manual how to start and close the Leica sp2 Confocal. (TCS SP2 AOBS system mounted on a DM IRE2)

Guide to Operating the Bruker FT-IR Microscope I. Basic Spectrum Acquisition with Vertex 70

Leica SP8 TCS Users Manual

Standard Operating Procedure

Leica DMi8A Quick Guide

AxioVision User's Guide. Release 4.1

DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands

Leica DMI 4000 tutorial

TRAINING MANUAL. Olympus FV1000

GlobiScope Analysis Software for the Globisens QX7 Digital Microscope. Quick Start Guide

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah

GXCapture 8.1 Instruction Manual

Nikon C1si Spectral Laser Scanning Confocal Microscope. User Guide

Internal Medicine Imaging Core Emory University Department of Medicine

Title: Leica SP5 Confocal User Manual

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s

Progeny Imaging Veterinary

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope

CHAPTER1: QUICK START...3 CAMERA INSTALLATION... 3 SOFTWARE AND DRIVER INSTALLATION... 3 START TCAPTURE...4 TCAPTURE PARAMETER SETTINGS... 5 CHAPTER2:

Transcription:

AxioVision 4.5 Brightfield Image Capture Procedure 1. STARTING-UP PROCEDURE: Remove blue dust cover and place on shelf under microscope. Turn on the halogen lamp by pushing the switch at the back right of the microscope. Turn on the computer by pushing the button on the top front right side. Connect the camera computer cable to the back of the camera. Log on to the Cardiology network using your login name and password. 2. USING A CLEAN, DRY SLIDE: Clean your slide with 70% ethanol using Q-tips and Kimwipes if it has fingerprints or dust on it. Make sure AxioCam cable has been hooked up to camera for at least 15 minutes. Open the camera path. Select Position #5 on the filter cube turret (empty). 3. SETTING UP KOEHLER ILLUMINATION: Set the halogen lamp to 12 volts by turning up to the top value, "10, or push the black button beneath the voltage setting to achieve 3200 deg K (best). On the lower right side of microscope, set the neutral density filters: 0-1.5 Set to 1.5. (% transmission) 0-6 Set to 6. (% transmission) 0-25 Set to 0. You may use 25, but it is not necessary. Diffuser: Push to the back to swing the diffuser in. Be sure that the blue swing-in filter under the stage is in place. Make sure the moveable lens under the specimen stage is swung in when using all objectives higher than 5X. Place a high contrast specimen on the stage (H&E is good), and focus with the chosen objective. Raise the condenser (under the stage) to the upper stop position. Close the luminous field diaphragm (the large round lens at the base of the microscope) by turning the outer ring until the diaphragm is visible in the field of view. Lower the condenser until the luminous field diaphragm becomes a 12-sided figure with clear edges in focus. 1

Open the luminous field diaphragm until it almost fills the field of view but the edges are still visible. Center the luminous field diaphragm by using the two metal centering screws on the condenser carrier. Open the luminous field diaphragm until the edges exactly disappear beyond the field of view. Set the aperture diaphragm: Remove one eyepiece from the binocular tube. While looking down the eyepiece opening, slide the knob on the front of the condenser until the aperture diaphragm fills 2/3 to 4/5 of the diameter of the objective exit pupils. Because each objective has a different field size and objective aperture, you will have to set the Koehler Illumination for each change of objective. 4. STARTING THE SOFTWARE: Start AxioVision Rel. 4.5 on the computer and activate the Live image. Select the Speed of the camera. The Slow Speed gives the best detail and the Fast is most pixilated. At the bottom of the Live-AxioCam HR screen, these display icons are shown after the Snap, Camera Speed, Exposure, White balance analysis icons: The first icon on the left is the Best Fit (usually the most vivid). The next one is the Min/Max. Next is Linear (full range of possible values). 4th icon is Gamma preset display for color closest to what you see when you look through the microscope. Use the Gamma pre-set display to perform the Shading Correction and the White Balance. You may change to other display options later. 5. SETTING THE SHADING CORRECTION: Activate the Properties window from the Live-AxioCam HR window menu, from Acquisition in the menu bar at the top of the screen or from the Workflow bar on the left of the screen Select Adjust tab and set the exposure time to 20 to 50 ms. Increase as needed. Select Frame, choose RGB (or B/W). Select Resolution, usually 1300x1030 scanned, from drop-down menu. 2

Select General, set the Gain factor to "1," Index to "0. Perform the Shading Correction first, each time you change objectives. Focus on a clean, clear area of the slide. If there is no clean area, you may remove the slide completely. On Live-AxioCam HR/General, click on the Shading Correction to smooth out the background. Some color may be seen at this point, but it is temporary. A message will appear if it is too dark or too bright. Increase or decrease exposure time under Live Properties/Adjust. Click on the Shading Correction box again. Repeat until the background is even with no spots or artifacts. 6. SETTING THE WHITE BALANCE: Focus on the specimen and select Adjust again. Now do a White Balance to attain the best color representation of the specimen by one of two methods: (a) Select Interactive and hold the crosshairs on a white area. This will give a readout of the values for red, green, and blue. Choose the best spot and release the mouse. (b) If there is no white area on the specimen at the region of interest, click on 3200 K first, wait about 20 sec. for color change, then click Automatic. You may also click on the White Balance icon at the bottom of the Live window. 7. ADJUSTING THE EXPOSURE: After the White Balance is done, increase the exposure as needed by increasing the exposure setting under the Adjust tab until the background is an appropriate shade of white for your specimen. You may also do an automatic exposure measurement by checking Automatic and then click Measure. Change the exposure time if necessary until the background is white enough for your application. Disable (uncheck) the automatic to use your own settings or the software will use automatic. You must repeat the SHADING CORRECTION and WHITE BALANCE every time you change objectives if you are capturing images, even if you are going back to one done previously. 3

8. FOCUSING ON THE SPECIMEN: While looking at the computer screen, adjust the fine focus knob slowly, then pause a couple seconds before continuing to adjust the focus. Right click on the live image to obtain the Spot meter/focus bar. When the green color comes to the red line, the software thinks best focus is achieved. Trust your eyes, not the focus bar. When the Spot meter is activated, a green frame appears around the live window. You may use this frame to enclose the region of interest (ROI) on which you want to focus by dragging the handles of the frame. 9. SELECTING THE OBJECTIVE AND RESOLUTION: On the Live-AxioCam HR window menu, select the Objective and Resolution from the dropdown menu. This will be saved with your image, and you will be able to add a scale bar by just selecting the scale bar icon. 10. CAPTURING THE IMAGE: Capture the image by using the "Snap" at the bottom of the Live- AxioCam HR image window, or by clicking any Camera Icon. You may do post-processing of your image before you save it, but the best practice is to save your image as captured as a.zvi. Then, after processing, save it with a new name or add "edited" to the name. Do not save images on the C Drive. Save on the D Drive or local network. 11. SAVING THE IMAGE: Save the image using File/Save As from the File menu. You may select from these four formats to store the images:.zvi This format allows you to store all data microscope, user, date, annotations and measurement together in a single file. You should save your original best image as a.zvi and export as a.tif. Keep an archival copy of both..tif With these external formats, AxioVision now stores all additional.jpg information separately from the actual image data. To do this, an.bmp XML file is created for every image file that contains other information such as additional data, annotations, and interactive measurement data. 4

12. USING THE SPLITTER SCREEN: Compare saved images side by side in the Splitter display: Select from the menu at the top of the screen View/Windows/ Splitter. All the opened images will be in the Gallery. In the Splitter display choose from two to eight screens. Click on Gallery. Select from the small gallery window at the top of each screen by clicking the image. Use the small red frame in the lower right corner of the Splitter display to change the size and region of the images in the Splitter Display. Right click to zoom up or back to change the size of the images. All the images in the Splitter display may be saved as a combined single image by clicking on "Create Image. 13. ADDING AND EDITING SCALE BARS: You may add a Scale Bar before or after the merged image is saved. If you selected the appropriate objective in the Live window dropdown menu, just select the scale bar icon and place it on the image. Otherwise, you must select from the top menu Measure/Scalings. To edit the Scalebar, go to the Attributes tab of Show Properties, available from the bottom menu of the image window, or from the workflow icons on the left of the screen. Highlight Scalebar. The line, color and font sizes can be edited here and saved as a default setting for all your scale bars. Undo is available from the Edit menu at the top of the screen. 14. EXPORTING IMAGES: Export by going to File/Export. An export window will appear. Export as a.tif image to a local folder, network folder, CD, or memory stick. Check "Burn in annotations" to save scalebars, etc. You will not be able to remove these annotations once they have been exported. Export as a batch by clicking on Batch, then click Add Files. Select files to be exported. The file names will appear in the window on the right. Click on Run Batch. As the files are exported, an "OK" message or an error message appears. Error messages are usually from incorrect syntax in the file name. Exported files should be saved to your folder in drive D or in a local network folder. 5

Set up the path before you export! 15. POST-PROCESSING YOUR IMAGES: Post-processing can be done on the saved images by selecting Show Properties and Display or by activating Processing from the menu at the top of the screen, then using its drop-down menu. The editing done with this dropdown menu takes some time to appear. There are several measuring tools available in this software that you may want to use. You may also measure your images with Image Pro 5.2 on the computer in WMB 306. You must get the USB dongle from Debby Martinson in WMB 303 or Lu Hilenski in WMB 329 to use the Image Pro software. 6

16. SHUTTING-DOWN PROCEDURE: Exit the AxioVision Rel. 4.5 from the File menu. Check to see if the next person is signed up to do fluorescence imaging. If so, then leave the lamp turned on and the camera plugged in. Log off the computer but do not turn the computer off. If you are the last person to use the system today, shut down the computer. Turn off the computer now a one step process. From the start menu, choose SHUTDOWN. The new monitor can be left on. Disconnect the camera computer cable. Swing a low power objective (5X or 10X) into the light path. Turn off the power supply for the mercury lamp by pushing the on/off switch. Put the blue dust cover back on the microscope, but not over the hot lamp. CLEAN UP THE AREA. You may log on the local Cardiology network and save directly to your own local shared folder. Please delete your files from the computer! 7