Testing a wavelet based noise reduction method using computersimulated

Similar documents
Use of Wavelet Multiresolution Analysis to Reduce Radiation Dose in Digital Mammography

Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities

Published text: Institute of Cancer Research Repository Please direct all s to:

Imaging Technique Optimization of Tungsten Anode FFDM System

A Study On Preprocessing A Mammogram Image Using Adaptive Median Filter

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

Features and Weaknesses of Phantoms for CR/DR System Testing

Dosepix Detector as kvp-meter in Radiology and Mammography: First steps

Differentiation of Malignant and Benign Masses on Mammograms Using Radial Local Ternary Pattern

Estimation of signal transfer property for wireless digital detector in different measurement schemes

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Modeling, validation and application of a mathematical tissue-equivalent breast phantom for linear slot-scanning digital mammography

Modeling detector performance in digital mammography using the linear cascaded systems approach

2012 :15th SESSION of ESMP

Tactile Sensation Imaging for Artificial Palpation

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing

The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography

Half value layer and AEC receptor dose compliance survey in Estonia

Dose conversion coefficients for electron exposure of the human eye lens

New spectral benefi ts, proven low dose

GE Healthcare. Senographe 2000D Full-field digital mammography system

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London

Improving registration metrology by correlation methods based on alias-free image simulation

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

A pre-processing technique to improve the performance of a computer-aided detection scheme in mammography

radiography detector

Acceptance Testing of a Digital Breast Tomosynthesis Unit

K-edge subtraction X-ray imaging with a pixellated spectroscopic detector

Electronic Noise in CT Detectors: Impact on Image Noise and Artifacts

PATIENT EFFECTIVE DOSES IN DIAGNOSTIC RADIOLOGY, NA

Pseudocolour Image Processing in Digital Mammography

Aquilion Precision Ultra-High Resolution CT: Quantifying diagnostic image quality

7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline

PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY

Mammography: Physics of Imaging

Report on the study and optimisation of chest and lumbar spine X-ray imaging

Your worldwide source of grids. What a difference a grid makes

Applied Sciences at Winterthur)

Iterative Reconstruction in Image Space. Answers for life.

SUSPENSION CRITERIA FOR IMAGE MONITORS AND VIEWING BOXES.

Improved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques

APJIMTC, Jalandhar, India. Keywords---Median filter, mean filter, adaptive filter, salt & pepper noise, Gaussian noise.

Calibration of KAP meters

Visibility of Detail

Comparison between film-screen and computed radiography systems in Brazilian mammography

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Abstract. ZEIGLER, GARY BOYCE, II. Direct Detection of Microcalcification Pairs in Simulated

Increasing the Number of Gray Shades in Medical Display Systems How Much is Enough?

Aim. Images for this section: Page 2 of 13

Synthesis Algorithms and Validation

Image Enhancement in Spatial Domain

High collection efficiency MCPs for photon counting detectors

Use of Back Scattered Ionizing Radiation for Measurement of Thickness of the Catalytic Agent Active Material

X-ray light valve (XLV): a novel detectors technology for digital mammography*

Digital Radiology with Photon Counting Detectors

Indicator of Small Calcification Detection in Ultrasonography using Decorrelation of Forward Scattered Waves

Amorphous Selenium Direct Radiography for Industrial Imaging

Digital Mammography: From Physical Performance Evaluation To Image Quality Analysis

While digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them.

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Electronic Noise Effects on Fundamental Lamb-Mode Acoustic Emission Signal Arrival Times Determined Using Wavelet Transform Results

Attikon, Rimini 1, , Athens, Greece , Athens, Greece , Athens, Greece

Evaluation of image quality of the compression schemes JPEG & JPEG 2000 using a Modular Colour Image Difference Model.

PERFORMANCE MEASUREMENT OF MEDICAL IMAGING SYSTEMS BASED ON MUTUAL INFORMATION METRIC

Comparison of computed radiography and filmõscreen combination using a contrast-detail phantom

DRX Plus Detectors: Going from Good to Great

Development of a large-area CMOS-based detector for real-time x-ray imaging

Breast Ultrasound Images Enhancement Using Gray Level Co-Occurrence Matrices Quantizing Technique

ADVANCED MEDICAL SYSTEMS PTE LTD Singapore Malaysia India Australia

A Phantom Study for Assessing the Effect of Different Digital Detectors on Mammographic Texture Features

Bayesian Estimation of Tumours in Breasts Using Microwave Imaging

Quality assurance: a comparison study of radiographic exposure for neonatal chest radiographs at 4 academic hospitals

8/2/2017. Radiologist Responsibilities. Radiologist Responsibilities. Medical Physicist Mammography Equipment Evaluation and Annual Survey

2 nd generation TOMOSYNTHESIS

Digital Radiography. Selected Topics

Phase Contrast Imaging with X-ray tube

Evaluation of no-grid radiography using the digital scattered x-ray removal processing

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

Objective Evaluation of Edge Blur and Ringing Artefacts: Application to JPEG and JPEG 2000 Image Codecs

Investigation of Effective DQE (edqe) parameters for a flat panel detector

Evaluation of a Novel Wafer-Scale CMOS APS X- Ray Detector for Use in Mammography

Observer Performance of Reduced X-Ray Images on Liquid Crystal Displays

SAFIRE. Sinogram Affirmed Iterative Reconstruction. Answers for life.

X-ray Tube and Generator Basic principles and construction

Adaptive Optimum Notch Filter for Periodic Noise Reduction in Digital Images

Introduction of a Single Chip TLD System for Patient Dosimetry

Protocol for the Quality Control of the Physical and Technical Aspects of Digital Breast Tomosynthesis Systems

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

Characterization of photon counting CZT detectors for medical x-ray imaging and spectroscopy

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

Radiographic sensitivity improved by optimized high resolution X -ray detector design.

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance

Copyright 2002 by the Society of Photo-Optical Instrumentation Engineers.

Comparison of Reconstruction Algorithms for Images from Sparse-Aperture Systems

Threshold contrast visibility of micro calcifications in digital mammography

AN APPROXIMATION-WEIGHTED DETAIL CONTRAST ENHANCEMENT FILTER FOR LESION DETECTION ON MAMMOGRAMS

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

Applications of a Pixellated Detection System to Digital Mammography

Transcription:

Testing a wavelet based noise reduction method using computersimulated mammograms Christoph Hoeschen 1, Oleg Tischenko 1, David R Dance 2, Roger A Hunt 2, Andrew DA Maidment 3, Predrag R Bakic 3 1 GSF- National Research Center for Environment and Health Institute of Radiation Protection Working Group Medical Physics Ingolstaedter Landstrasse 1 85764 Neuherberg Germany 2 Department of Physics, The Royal Marsden NHS Foundation Trust London, London SW3 6JJ, United Kingdom 3 Department of Radiology, The Thomas Jefferson University of Pennsylvania, Philadelphia, PA 19104-4283 USA Email (principal author) Christoph.Hoeschen@gsf.de +49 89 3187 4560, FAX: + 49 89 3187 2517 ABSTRACT A wavelet based method of noise reduction has been tested for mammography using computer-simulated images for which the truth is known exactly. This method is based on comparing two images at different scales, using a crosscorrelation-function as a measure of similarity to define the image modifications in the wavelet domain. The computersimulated images were calculated for noise-free primary radiation using a quasi-realistic voxel phantom. Two images corresponding to slightly different geometry were produced. Gaussian noise was added with a mean value of zero and a standard deviation equal to 0.25% to 10% of the actual pixel value to simulate quantum noise with a certain level. The added noise could be reduced by more than 70% using the proposed method without any noticeable corruption of the structures for 4% added noise. The results indicate that it is possible to save 50% dose in mammography by producing two images (each 25% of the dose for a standard mammogram). Additionally, a reduction or even a removal of the anatomical noise might be possible and therefore better detection rates of breast cancer in mammography might be achievable. Medical Imaging 2005: Physics of Medical Imaging, edited by Michael J. Flynn, Proceedings of SPIE Vol. 5745 (SPIE, Bellingham, WA, 2005) 1605-7422/05/$15 doi: 10.1117/12.595806 969

INTRODUCTION The purpose of this study was to test a newly developed method of reducing the contribution of non-correlated noise components to radiographic images [1]. The method is based on comparing two images obtained with slightly different geometry, but can also be used for two images sharing the same geometry. The images are compared at different scales, using a cross correlation function as a measure of similarity to define the necessary image modifications in the wavelet domain. Formally, a wavelet coefficient at some scale is given weighting proportional to the value of the corresponding correlation matrix in the given location [1]. This method is based on the idea of measuring anatomical noise [2, 3]. If the method is used on two images resulting from two detection processes of the same exposure, these should only deviate by means of quantum and detector noise. By using the proposed method on such images the quantum noise and the detector noise should be reduced greatly without reducing anatomical structures as long as these structures are not corrupted by noise. If two detection processes of two exposures with slightly different geometric conditions are the basis of this noise reduction, quantum noise, detector noise and anatomical noise are reduced by this method. Theoretically, the method could in this case disturb structures. It is the case which implies more difficulties to the noise reduction method based on correlations. Therefore, it has to be proven that there is no disturbance of the structures. This study tests the proposed method using computer simulated mammograms, for which the truth is known exactly. Also the noise reduction possibilities should be quantified. If the noise reduction method is even able to deal with this case of different imaging geometries it can be used for reducing technical noise but also anatomical noise. MATERIAL AND METHODS To quantify the effect of noise reduction and to prove the de-noising possibility a set of images with known noise components is needed. Such sets of images have been produced for this study by using computer-simulated images [4] of an artificial high-resolution voxel phantom of the breast (see figure 1) [5]. Due to the method of creation, this phantom is a very good representation of the structures and tissue compositions within the breast, and is used in this paper for simulation of the breast anatomical background. Simulated images without quantum noise have been generated using Monte-Carlo simulation with a pixel size of 100 um assuming noise-free primary radiation. Two such generated synthetic mammograms, corresponding to focus-detector distances of 650 mm and 600 mm, are shown in figures 2 and 3, respectively. The synthetic mammograms do not contain highly realistic anatomical noise, due to the lack of fine, small-scale tissue structures in the phantom. We added to each pixel Gaussian noise with a mean of zero and a variance per pixel proportional to the pixel value, which simulated the appearance of quantum noise. 970 Proc. of SPIE Vol. 5745

Figure 1: A cross-section of the phantom, which have been used for the Monte-Carlo simulation. Proc. of SPIE Vol. 5745 971

Figure 2: Simulated mammogram using the phantom shown in figure 1 for a Monte-Carlo code using noise-free primary radiation with a Source focus detector distance (SDD) of 650 mm. Figure 3: Simulated mammogram using the phantom shown in figure 1 for a Monte-Carlo code using noise-free primary radiation with a Source focus detector distance (SDD) of 600 mm. Gaussian noise was added to each pixel with a mean of zero and a variance per pixel proportional to the pixel value, which simulates the appearance of quantum noise. It should be mentioned here that the simulation is not referring to any detector noise or any frequency filtration caused by a possible detector as it could be described according to the model introduced by Cunningham [6]. 972 Proc. of SPIE Vol. 5745

A series of image pairs were obtained with noise levels in the glandular region of each image ranging from 0,25 % to 10 %. We evaluated for this study the four pairs with 1%, 2%, 4% and 10% additional noise to simulate something like a real mammogram (2% additional noise) and mammograms collected with much lower doses. The evaluation for the images with 1% added noise was selected for evaluation to illustrate the limitations in noise reduction and to give an impression of the maximal difference between the original image and the de-noised image. In this case, the noise in the difference image is expected not to be very prominent and thus better shows effects on the simulated anatomical structures than those with higher noise amounts. This assumption is correct as it can be seen in figure 9c versus figures 4d, 6d and 8c RESULTS The effect of the de-noising as described in [1] is shown in the ROIs of images with 10% added noise (figure 4). In figure 5 the power spectra of the difference images (figure 4 D) are demonstrated compared to those of the added noise itself, which can be obtained by calculating the difference between images B and A. The effect of de-noising on the appearance of the image is very evident. The noise can be reduced by 70% or more. This value results from taking the integrals under the curves of the power spectra of the added noise and the reduced noise. For this reduction a visual comparison of the re-constructed and the original images shows no noticeable corruption of the structures within the images. There might be only small effects of the noise reduction on the structures, which can also be seen on the appearance of the structures. However, it is very difficult to recognize the difference. It should be stated that 10% added noise is about a factor of 5 higher noise properties than what would be expected in real mammographic images. This corresponds to a dose reduction for a single image (two are needed for the image creation used as the de-noising process) of a factor of 25. We are not claiming to reduce dose in mammography by a factor like this, but it could according to these images be assumed that the relation between image quality and dose can be improved by a factor of 5 to 10 at least. A B C Proc. of SPIE Vol. 5745 973

Figure 4: A: A part (ROI) of the original noise free simulated image; B: The same ROI with 10% added noise; C: The same ROI from the de-noised image; D: Difference image between C and A. D Figure 5: The power spectra of the difference images in figure 4 D compared to the noise spectra of the corresponding added noise. Figure 6 shows the same ROI as analyzed in Figure 4 and its evaluation for the case with 4% added noise (Figure 7). Again the appearance of the image shows the effect of the de-noising very obvious. Hardly any corruption of the structures can be seen. The noise can again be removed by 70% or more if one is referring to a noise quantity for example like the integral under the determined power spectra as shown in figure 7. This figure shows also that the noise reduction is even more effective for higher frequencies. For the lowest frequency band the noise reduction has not been performed, instead the information from this band has been used to characterize the translation process needed for correlating the images in an optimal way. The assumption for this was, that the low frequency noise is not disturbing the detection of structures as much as the high frequency noise as can be concluded from results in a paper by Burgess and Judy [7] and as described in a paper by Hoeschen et al. [8]. 974 Proc. of SPIE Vol. 5745

A B C D Figure 6: A: A part (ROI) of the original noise free simulated image; B: The same ROI with 4% added noise; C: The same ROI from the de-noised image; D: Difference image between C and A. Figure 7: The power spectra of the difference images in figure 6 D compared to the noise spectra of the corresponding added noise. Figure 8 and 9 show another ROI of the images for 2% and 1% added noise with the same impression. Noise is reduced, while structures are preserved. Figure part A for these two figures shows the image with the added noise while part B Proc. of SPIE Vol. 5745 975

shows the de-noised image. As it has been expected for the images with 1% added noise the residuum (figure 9c) does show some edges of the structures as well. It can be clearly seen in the reconstructed image (figure 9b) that the structures themselves are preserved but the appearance of the edges is slightly stronger, which corresponds to a certain edge enhancement which is possible due to the noise reduction. This difference in the representation of the edges can be seen in the residuum image because the power of the noise is much lower than for higher noise levels, so that the differences in the edges have a visible contrast which is not the case for the residua calculated for higher noise levels. A B C Figure 8: A: A part (ROI) of the simulated image with 2% added noise; B: The same ROI from the de-noised image; C: Difference image (residuum) between B and A. A B C Figure 9: A: A part (ROI) of the simulated image with 1% added noise; B: The same ROI from the de-noised image; C: Difference image (residuum) between B and A. CONCLUSIONS The results of this study with images for which the truth is known exactly have shown that a recently proposed method of noise reduction [9] allows the reduction of the technical noise (quantum and detector noise) so effectively that it is possible to save at least 50% dose in mammography. This can be achieved by producing two images (each at 25% of dose). It might be possible to reduce dose even further but there seem to be theoretical limitations. These can be explained as follows: 976 Proc. of SPIE Vol. 5745

According to data in the literature e.g. [10] one can assume that any detail in an image which should be detected and recognized by a human observer needs a signal to noise ratio of at least 3:1 for each single frequency component that should be analysed by the observer. There are data in the literature also, which would imply higher signal-to-noiseratios needed for detection, however we used the lowest value, because this will give us a conservative expectation of the possible dose reduction. The method which is used for the de-noising process needs locally and frequency dependent a signal to noise ratio of 1:1 or higher in each of the images of the image pair to detect a structure without noticeable errors. When the signal to noise ratio gets smaller for a certain frequency, the algorithm used might detect a statistical correlation of the noise properties instead of the structures, which would be suppressed in this case. If the signal to noise ratio is larger than one this should not happen according to the process being used for the de-noising. If one assumes that the noise is only caused by quantum noise, than a three times smaller signal to noise ratio as needed with the method giving the same image quality would correspond to a dose reduction by a factor of 9 for each single image of the two images needed, assuming a Poisson statistic for the quantum noise. However the detector noise is getting more important [11] for lower doses. This would mean that the amount of dose which can be saved is smaller. On the other hand it has been shown, that for some examinations noise effects due to overlaying structures in the patient which cannot be distinguished are of the same magnitude as the technical noise [3]. These effects are scaling with the dose which means they would be much smaller in the case of smaller doses and would allow an even further reduction. The reduction of the technical noise by 70% might be suggesting a 3-fold increase of in the effective SNR, resulting again in a possible dose reduction of a factor of 9 for each single image in a two image procedure; however this value was gained using the whole frequency range. Assuming those frequencies with the most interesting structures in mammography and the biggest difficulties to detect, a much better noise reduction was achieved which implies that an even higher dose reduction would be possible. The proposed noise reduction method [9] also allows the removal of this anatomical noise and thus has the potential to improve breast cancer detection rates for mammography as it has been stated in [12]. The reduction of anatomical noise could be proven by comparing the results presented here with simulated mammograms gained with a high resolution voxel phantom produced with using a real breast specimen. This voxel phantom is currently under development. [13]. It could be shown in this study that the structures are not significantly disturbed even when using different geometric imaging conditions for the two images. On the other hand the method can be used to increase image quality as it has been shown on the 2% and 1% added noise images for example by using the implemented automated edge enhancement without suffering from increased highfrequency noise. This enhancement is again free of parameters and is automatically applied by the optimisation of the proposed method using the optimisation for the k-values as described in [9]. It should be stated here again very clearly, that by using this method the DQE of the detecting systems is not changed. The method just takes advantage of an additional information layer which has not been previously used. This information layer is the correlation between the relevant information content in both images. ACKNOWLEDGEMENT This work was partly funded and carried out in the EU 5 th Framework Programme (1998-2002), Nuclear Fission and Radiation Protection under Contract N FIGM-CT-2000-00036. Proc. of SPIE Vol. 5745 977

REFERENCES [1] O Tischenko, C Hoeschen, E Buhr: Reduction of anatomical noise in medical X-ray images. To be published in Radiation Protection Dosimetry. [2] O Tischenko, C Hoeschen, O Effenberger, S Reissberg, E Buhr, W Döhring: Measurement of the noise components in the medical X-ray intensity pattern due to overlaying, nonrecognizable structures, Proceedings of SPIE 5030 (I), 422-432 (2003). [3] C Hoeschen, O Tischenko, E Buhr, H Illers: Comparison of technical and anatomical noise in digital thorax X-ray images. To be published in Radiation Protection Dosimetry. [4] RA Hunt, DR Dance, PR Bakic, ADA Maidment, M Sandborg, G Ullman, G Alm Carlsson: Calculation of the properties of digital mammograms using a computer simulation. To be published in Radiation Protection Dosimetry. [5] PR Bakic, M Albert, D Brzakovic, ADA Maidment: Mammogram synthesis using a 3D simulation. II. Evaluation of synthetic mammogram texture. Med. Phys. 29 (9), 2140-2151 (2002). [6] I A Cunningham: Linear-systems modeling of parallel cascaded stochastic processes: the NPS of radiographic screens with reabsorption of characteristic x-radiation. SPIE 3326, 220-230 (1998). [7] A E Burgess, P F Judy: Detection in power-law noise: spectrum exponents and CD diagram slopes. SPIE 5034, 57 62 (2003). [8] C Hoeschen, S Reissberg, W Döhring: The importance of Optimizing the Image Processing for Different Digital X- ray Detectors to Get as Much Information as Possible from the Radiographs. SPIE 4682, 828-838 (2002) [9] O Tischenko, C Hoeschen, E Buhr: An artefact-free, structure-saving noise reduction method using the correlation between two images for threshold determination in the wavelet domain, to be published in the proceedings SPIE Medical Imaging 2005, conference on image processing. [10] P G J Barten: Contrast Sensitivity of the Human Eye and its Effects on Image Quality., here p. 8-11, SPIE Press (1999). [11] H Illers, E Buhr, C Hoeschen, D Bergmann: Measurement of the detective quantum efficiency (DQE) of digital x- ray imaging devices according to the forthcoming standard IEC 62220-1, SPIE 5368, 177-87 (2004). [12] C Hoeschen, O Tischenko, D R Dance, R A Hunt, A D A Maidment, P R Bakic: Evaluation of a novel method of noise reduction using computer-simulated mammograms. To be published in Radiation Protection Dosimetry. [13] C Hoeschen, U Fill, M Zankl, W Panzer, D Regulla, W Döhring: A high-resolution voxel phantom of the breast for dose calculations in mammography To be published in Radiation Protection Dosimetry. 978 Proc. of SPIE Vol. 5745