VDD PC-Based Digital Vibrometer

Similar documents
VDD PC-Based Digital Vibrometer

HSV-2000 High Speed Vibrometer

VibSoft Data Acquisition Software

HSV-2000 High Speed Vibrometer

VibSoft Data Acquisition Software

VibSoft Data Acquisition Software

OFV-5000 Xtra Laser Vibrometer. Xtra sensitivity and versatility Datasheet. OFV-5000 Xtra Laser Vibrometer

3D Optical Motion Analysis of Micro Systems. Heinrich Steger, Polytec GmbH, Waldbronn

Surface Vibration Measurement on Rotating Components

Laser-Based Evaluation of Cement Hydration. Laser-Based Evaluation of Cement Hydration Non-Contact Inspection of Curing Processes Application Note

Aircraft Health Monitoring. Aircraft Health Monitoring Laser Vibrometry for Damage Detection Using Lamb Waves Application Note

Fiber Optic Cables for the TopSens Series

Module 5: Experimental Modal Analysis for SHM Lecture 36: Laser doppler vibrometry. The Lecture Contains: Laser Doppler Vibrometry

VibroFlex Connect Configurable core of the modular vibration sensing system Preliminary datasheet. VibroFlex Connect

ArbStudio Arbitrary Waveform Generators

Model 865 RF / Ultra Low Noise Microwave Signal Generator

ArbStudio Arbitrary Waveform Generators. Powerful, Versatile Waveform Creation

Very High Frequency Calibration of Laser Vibrometer up to 350 khz

FFT Spectrum Analyzer

SIGNAL RECOVERY. Model 7265 DSP Lock-in Amplifier

Tektronix: Products > AWG 2040 Arbitrary Waveform Generator

R&S ZVT Vector Network Analyzer Specifications

648. Measurement of trajectories of piezoelectric actuators with laser Doppler vibrometer

Tutorial: 3D Scanning Vibrometry for. IMAC XXVII D. E. Oliver, Polytec, Inc.

Spectrum Analyzer. Distribution in the UK & Ireland

R&S ZNC Vector Network Analyzer Specifications

Microscopic Laser Doppler Vibrometer

CALIBRATION OF LASER VIBROMETER STANDARDS ACCORDING TO ISO

Fibre Laser Doppler Vibrometry System for Target Recognition

Arbitrary/Function Waveform Generators 4075B Series

R&S ZNB Vector Network Analyzer Specifications

Fiber Pigtailed Variable Frequency Shifters Acousto-optic products

LSV Laser Surface Velocimeter. LSV Laser Surface Velocimeter Non-contact speed and length measurement Product brochure

Keysight Technologies N9320B RF Spectrum Analyzer

Agilent N9320B RF Spectrum Analyzer

Vibration Transducer Calibration System

Laser Surface Velocimeter. LSV Series 6000

R&S ZNA Vector Network Analyzer Specifications

R&S ZVT Vector Network Analyzer Specifications

Instrument. Specifications OR34/35. SmartRouter. 2 to 8 Channels Multi-Analyzers. Controller Unit TECHNICAL SPECIFICATIONS

Vector Network Analyzers ZVB

m+p VibControl Sine Vibration Control

Dual Channel Function/Arbitrary Waveform Generators 4050B Series

DATA SHEET. LitePoint IQflex LitePoint Corporation. All rights reserved.

TEST & MEASURING INSTRUMENTS. Analyzer. (4 Ports) 4 Ports

DSA800. No.2 RIGOL TECHNOLOGIES, INC. All-Digital IF Technology 9 khz GHz Frequency Range

Model 745 Series. Berkeley Nucleonics Test, Measurement and Nuclear Instrumentation since Model 845-HP Datasheet BNC

R&S ZNBT8 Vector Network Analyzer Specifications

RF Signal Generators. SG380 Series DC to 2 GHz, 4 GHz and 6 GHz analog signal generators. SG380 Series RF Signal Generators

Spectrum Analyzer FSL

Datasheet RS Pro Arbitrary Waveform Generator 40MHz RS Stock Number : ENGLISH

TG1010A AIM & THURLBY THANDAR INSTRUMENTS. 10MHz programmable DDS function generator. Direct Digital Synthesis

Models 296 and 295 combine sophisticated

Experimental study of slider dynamics induced by contacts with disk asperities

LASER VIBROMETER CALIBRATION AT HIGH FREQUENCIES USING CONVENTIONAL CALIBRATION EQUIPMENT

Magnetic-Field Test System / Low-Frequency Test System for Emission and Immunity Tests / MTS-800

R&S ZV-Z5x Calibration Units Specifications

SHF Communication Technologies AG

Agilent 81980/ 81940A, Agilent 81989/ 81949A, Agilent 81944A Compact Tunable Laser Sources

R&S FSWP Phase Noise Analyzer Specifications

Handheld Spectrum Analyzer R&S FSH khz to 3 GHz

Advanced Test Equipment Rentals ATEC (2832)

Test System for Discrete Track Recording (DTR)

Agilent 86146B Optical Spectrum Analyzer Technical Specifications

R&S ZV-Z5x Calibration Units Specifications

Coherent Laser Measurement and Control Beam Diagnostics

SDG2122X SDG2082X SDG2042X

PicoSource PG900 Series USB differential pulse generators

Analog Devices Welcomes Hittite Microwave Corporation NO CONTENT ON THE ATTACHED DOCUMENT HAS CHANGED

R&S ESCI/ESCI7 EMI Test Receiver Specifications

EXHIBIT 10 TEST REPORT. FCC Parts 2 & 24

New Paradigm in Testing Heads & Media for HDD. Dr. Lutz Henckels September 2010

DC MHz 1 GS/s. Modular CompactPCI/PXI Digitizer with Oscilloscope Characteristics

HMC-T2240. Synthesized Signal Generator, 10 MHz to 40 GHz

R&S ZN-Z5x Calibration Units Specifications. Data Sheet V03.00

Reflectometer Series:

Datasheet SHF D Synthesized Clock Generator

Handheld Spectrum Analyzer R&S FSH3

Utilizzo del Time Domain per misure EMI

DSA800 RIGOL TECHNOLOGIES, INC.

Agilent 8614xB Optical Spectrum Analyzer Family Technical Specifications

HMC-T2220B. Portable & Battery Operated! Portable Synthesized Signal Generator, 10 MHz to 20 GHz. Instrumentation Product Support at

MSO Supplied with a full SDK including example programs Software compatible with Windows XP, Windows Vista and Windows 7 Free Technical Support

R&S ZN-Zxxx Calibration Units Specifications. Data Sheet V04.00

Arbitrary/Function Generator AFG1000 Series Datasheet

MRI & NMR spectrometer

DSA700 Series Spectrum Analyzer

WaveStation Function/Arbitrary Waveform Generators

Agilent 8560 EC Series Spectrum Analyzers Data Sheet

Handheld Spectrum Analyzer R&S FSH3

Unprecedented wealth of signals for virtually any requirement

The V-Line Airborne Laser Scanner RIEGL

Appendix A: Specifications

APPH6040B / APPH20G-B Specification V2.0

PXIe Contents CALIBRATION PROCEDURE. Reconfigurable 6 GHz RF Vector Signal Transceiver with 200 MHz Bandwidth

OPTICAL MEASURING INSTRUMENTS. MS9710B 0.6 to 1.75 µm GPIB OPTICAL SPECTRUM ANALYZER

Agilent U2701A and U2702A USB Modular Oscilloscope. Data Sheet

Chapter 6 Specifications

Berkeley Nucleonics Test, Measurement and Nuclear Instrumentation since 1963

430. The Research System for Vibration Analysis in Domestic Installation Pipes

Transcription:

VDD PC-Based Digital Vibrometer MODULAR VIBROMETER SYSTEM OFV-5000 Vibrometer Controller OFV-505/503 Standard Sensor Heads OFV-534 Compact Sensor Head OVF-551/552 Fiber Interferometers VDD PC-Based Digital Vibrometer VibSoft Software PICOMETER DISPLACEMENT SENSITIVITY New measurements for data storage, micro-electro-mechanical structures (MEMS) and hearing dynamics push a traditional analog LDV beyond its resolution limits. Polytec offers a choice of either the stand-alone VDD-E-600 Digital Front-End or a DD-600 equipped OFV-5000 Vibrometer Controller with analog and digital capability. The VDD PC-Based Digital Vibrometer utilizes 5 MSa/s, a multi-channel data acquisition and the VibSoft-VDD comprehensive software package for data processing. Benefits of Digital Vibrometers The sophisticated design of Polytec s Digital Vibrometers opens the door to many new possibilities. One of the most significant advantages of the digital LDV is the extremely low noise floor and the exceptional sub-picometer displacement resolution on mirror-like surfaces. Digital LDVs are best-suited when extreme accuracy is desired. With a maximum velocity of 810 mm/s and a 2 MHz bandwidth, a wide range of applications can be addressed. In addition, a digital LDV is an excellent choice when signals are expected to be saved, processed or exported to sophisticated signal processing software. The VibSoft-VDD Software package is cost effective, as it already includes a spectrum analyzer, an arbitrary waveform generator and a data acquisition system. VDD Digital Vibrometer measurement on a read/write head of a hard disk drive, with video image in VibSoft-VDD Software. Advancing Measurements by Light www.polytec.com

VDD-E-600 Digital Front-End Alternative Solution: OFV-5000 Vibrometer Controller with DD-600 Digital Decoder Key Features Ultra-precise vibration measurements for data storage, MEMS, hearing dynamics or calibration applications True DC capabilities for switches and stages 2 MHz bandwidth and cutting-edge, sub-picometer resolution Outstanding linearity, accuracy and dynamic range Excellent stability, insensitive to drift and ageing effects (all digital) Upgrade any OFV-5000 based vibrometer system by adding a DD-600 Decoder Module Stand-alone VDD-E-600 Digital Front-End, can be used with any OFV Series Sensor Head Comprehensive VibSoft-VDD Software for data acquisition, processing and export, with open software interface (PolyFileAccess) Optional integrated arbitrary waveform generator Optional upgrade for precision MEMS displacement measurements using the MSV-400 Microscope Scanning Vibrometer and the MSA-400 Micro System Analyzer VDD System Configuration Front-Ends The vibrometer signal from the sensor head can go to one of two basic configurations the stand-alone, digital-only VDD-E-600 Front- End or, for the greatest flexibility, the DD-600 Digital Decoder board installed in the OFV-5000 Vibrometer Controller omplemented by a VDD- Z-011 Junction Box. This configuration enables both digital and analog demodulation, which is preferred by data storage and MEMS applications. The output from the front-end passes to the data management and acquisition system where the displacement, velocity and acceleration can be calculated and displayed. A reference channel input is available for acquiring additional signals for phase and frequency response function (FRF) measurements. The output from an internal arbitrary waveform generator is also available and can be used to drive the device under test. Sensor Heads Polytec offers five vibrometer heads the Dual Fiber OFV-552 Interferometer, the Single Fiber OFV-551 Interferometer, the OFV-534 Compact Sensor Head with optional video camera or the standard OFV-505 and OFV-503 Sensor Heads. Both the OFV-551 and OFV-552 are coupled to microscope based fixtures (MSV-050, MSV-100) for head Media, MEMS or inner ear dynamic testing. 2

Principle of Operation The output signal from the vibrometer sensor head contains information about the velocity (signal frequency) and displacement (signal phase) of the test object. The VDD decodes the phase of the signal to measure the dynamic displacement of the test object. In the VDD system a quadrature demodulator generates the sine and cosine components of the sensor signal. They are digitized by a high sampling rate ADC board inside the data management system. These two signals can be represented as a rotating vector whose phase is directly related to the displacement of the test object and the laser wavelength. By numerically decoding the phase with an arctan function, high displacement accuracy and resolution is achieved. Velocity and acceleration data are available through VibSoft-VDD Software by differentiation of the displacement signal. To learn more about digital vibrometry, please read our tutorial Basics of Digital Vibrometry at www.polytec.com/usa/vib-university. The I&Q Signal Format VibSoft-VDD Software Features The VibSoft-VDD Software package manages the data acquisition, signal decoding, arbitrary waveform generator and data display. Measurements can be presented in both time and frequency domains. Included with the software are an extensive number of functions including FRF, auto power spectra, cross power spectra, coherence and phase. Controls acquisition, signal processing, display and export of all data Calculates of FRFs, coherence and other signal properties Applies digital high, low and band pass filters; User can load custom ASCII format filters Differentiates and integrates in time and frequency domain Averages the input signal in the time domain Calculates FFT with 12,800 lines of resolution Controls integrated arbitrary waveform generator with common and user defined excitation waveforms Program automatic test routines and interface with other software using Visual Basic scripting (VBS) Export data in ASCII format, Universal file UFF, direct data access to binary data using Poly File Access software and optional binary ME Scope interface Access all settings of the OFV-5000 Controller via RS-232 interface Access live video images of the test object and laser beam position when using a sensor head equipped with an internal video camera, or an MSV-050/100 Microscope Adapter OFV-534 Compact Sensor OFV-505 Sensor Head OFV-551 Fiber Interferometer 3

Applications The VDD PC-Based Digital Vibrometer is an excellent tool to characterize very small dynamic displacements. Both researchers and engineers use the large bandwidth, flat response and fine resolution to measure displacement in applications where traditional analog vibrometers lack sufficient resolution. A few of these situations are described briefly in what follows. Runout Testing of Precision Spindles The hard disk drive (HDD) industry requires higher density storage in each successive product generation. With track densities approaching 250,000 TPI, a HDD manufacturer must control spindle runout so that tracking isn t lost. The VDD Digital Vibrometer can measure data that allows the calculation of both the repetitive runout (RRO) and non repetitive runout (NRRO) with the fine precision necessary to verify the tight tolerances needed (NRRO < 5 nm). The low noise (high resolution) of the digital LDV between DC and 20 khz enables this application. The acquired data is easily transferred to a computer for proprietary RRO and NRRO analysis. Frequency Response Measurement Frequency response testing on actuators and head gimbal assemblies (HGA s) has traditionally relied on data from a LDV processed by a 2- channel FFT analyzer. With the VDD, Polytec provides a complete solution that eliminates the cost of an external spectrum analyzer and improves the resolution of the measurement. Measuring the resonance response of a part has never been easier. Slider Displacement Development HDD researchers are interested in the precise, time-domain displacement history of a slider. The off-track motion of the slider is very small. For real time measurements, a high resolution analog decoder is often the best choice, but when even higher resolution is required, the post processed digital output is the only alternative. Displacement Stable flying heights for a disk drive head are determined with a digital LDV. A digital vibrometer reduces signal-to-noise by one or two orders of magnitude compared to analog vibrometers and can easily detect the motion. Using the VDD system, it is possible to make a direct comparison between the time dependent slider position and the position error signal (PES). Using this comparison, contributions to the PES can be identified and minimized. Head/Media Interface Studies The high resolution of the digital LDV permits accurate measurements of head/media interface dynamics. Engineers can characterize a given HGA and slider configuration with a particular media product. If not satisfactory, other drive component combinations can be tried until an appropriate solution is found that offers stable flight and minimal registration errors. Calibration of Transducers and Sensors The digital LDV does a direct and highly precise mathematical phase demodulation comparable to the international standard ISO 16063-11 (Method 3). The measurement results are referenced to the HeNe laser wavelength due to its accuracy and stability. The digital LDV s performance is so outstanding that it can be used as a reference standard in primary calibration systems for shock and vibration transducers. VibSoft-VDD provides an FFT function for switching from time to frequency domain. The repetitive spindle runout (RRO) is obvious in the frequency domain. 5

Technical Data Front-End Specifications VDD-E-600 Stand-Alone Digital Front-End Power 100 VAC... 240 VAC, max. 75 W Dimensions [W x L x H] 235 mm x 320 mm x 150 mm (9.3 in x 12.6 in x 5.9 in) Weight 5.7 kg (12.6 lbs) Signal inputs REF1/REF2 ± 200 mv... ± 10 V, into 1 MΩ Input coupling AC or DC, adjustable by software Inputs TRIG, GATE, AUX TTL/CMOS Output signal Internal generator output, max. ±10 V Outputs SYNC, AUX TTL/CMOS DD-600 Signal Converter Board for OFV-5000 Vibrometer Controller Decoder type Output signal Auxiliary decoder (I&Q converter) Baseband Doppler signal in I&Q format for external VDD decoding VDD-Z-011 Junction Box Signal input REF ±200 mv... ±10 V, into 1 MΩ Input coupling AC or DC, adjustable by software Inputs/outputs as specified for VDD-E-600 VDD Performance Specifications Frequency range Max. number of FFT lines 12,800 Maximum number of samples Max. velocity for frequency, f, and sample time, t 0 Hz 2 MHz Time mode: 64 MSa FFT mode: 32,768 Frequency f 51.2 khz Frequency f > 51.2 khz t < 1 s t = 1 s... 2 s t > 2 s Velocity resolution 1), depending on the vibration frequency f Max. displacement Displacement resolution 1) Measurement error for displacement Total harmonic distortions Amplitude frequency response Spurious signals 810 mm/s 810 mm/s 810 mm/s 405 mm/s 405 mm/s 405 mm/s f = 2 khz f = 20 khz f = 200 khz f = 2 MHz 0.005 µm s -1 / Hz 0.05 µm s -1 / Hz 0.5 µm s -1 / Hz 5 µm s -1 / Hz unlimited < 0.4 pm µm s -1 / Hz ± (1 % of RMS reading) for 1 nm 100 nm ± (0.2 % of RMS reading + 1 nm) for 100 nm < 0.1 % up to 50 khz (sinusoidal vibration) Displacement: typ. ± 0.1 % Velocity: typ. ± 0.2 % Acceleration: typ. ± 0.2 % < 20 pm 1) Resolution is defined as the signal amplitude (rms) at which the signal-to-noise ratio is 0 db in a 1 Hz resolution bandwidth (RBW), measured on 3M Scotchlite Tape (reflective film). The velocity resolution values refer to the settings for a maximum velocity of 810 mm/s. 4

Technical specifications are subject to change without notice. LM_DS_VDD_2006_05_3000_E VDD Data Acquisition System Acquisition board 4 channels (2 for I/Q signal, 2 for reference); 5 MSa/s, 12 bits Computer (optional) Signal generator (optional) External signal generator VibSoft-VDD Software Vib-S-VDD VIB-S-GateIn VIB-S-FFT128 VIB-S-VBEng VIB-S-ExpUFF VibSoft-VDD Options VIB-S-SIG-M VIB-S-ExtSig VIB-S-ExpME VIB-S-SigPro VIB-S-VIDEO PC-D Desktop PC or PC-I Industrial PC (from Polytec) System requirements: AMD Athlon 3000 XP+ or Intel Pentium 4; 3.0 GHz or higher; 512 MByte RAM; Windows XP Professional or Windows 2000 operating system 1 channel, 512 khz, arbitrary waveforms HP-33250A (option from Polytec) or HP-33120A (discontinued model), connected to GPIB via USB adapter Digital data acquisition and processing, time & frequency domain displays, time & frequency domain averaging, digital filter program, time domain integration & differentiation. Input channels for vibrometer & reference (VDD-E-600: 2 references) TTL Gate input software Supports 12,800 FFT lines Compliance with Standards Electrical safety IEC/EN 61010 Creates Visual Basic Scripts for automation of measurements and data presentation. Includes User Defined Data Sets (UDDS) for adding external or post processed data Converts data to Universal File Format (UFF) Internal arbitrary waveform generator Controls external arbitrary waveform generator via optional adapter Binary data interface for data exchange with ME scope software Polytec Signal Processor, the user interface to the math library included in the VibSoft Software Video option for VibSoft and VDD Software EMC IEC/EN 61326 Emission: FCC Class B, IEC/EN 61000-3-2 and 61000-3-3 Immunity: IEC/EN 61000-4-2 to 61000-4-6 and IEC/EN 61000-4-11 Laser safety IEC/EN 60825-1 (CFR 1040.10, CFR 1040.11) For more information and applications of Polytec s Digital Vibrometers please contact your local Polytec sales engineer or visit our website at www.polytec.com/usa/lm-digital. Windows and Visual Basic are registered trademarks of Microsoft Corp. Laser Radiation Do not stare into beam Class 2 Laser Product According to IEC/EN 60825-1 (2001) Complies with 21 CFR 1040.10 and 1040.11 except for deviations pursuant to Laser Notice no. 50, dated 26 July 2001 P 1 mw/cw; λ = 620-700 nm Polytec GmbH (Germany) Polytec-Platz 1-7 76337 Waldbronn Tel. + 49 (0) 7243 604-0 Fax + 49 (0) 7243 69944 info@polytec.de Polytec-PI, S.A. (France) 32 rue Délizy 93694 Pantin Tel.+33(0)148103934 Fax + 33 (0) 1 48 10 09 66 info@polytec-pi.fr Lambda Photometrics Ltd. (Great Britain) Lambda House, Batford Mill Harpenden, Herts AL5 5BZ Tel. + 44 (0)1582 764334 Fax + 44 (0)1582 712084 info@lambdaphoto.co.uk Polytec KK (Japan) Hakusan High Tech Park 1-18-2 Hakusan, Midori-ku Yokohama-shi, 226-0006 Kanagawa-ken Tel. +81(0) 45 938-4960 Fax +81(0) 45 938-4961 info@polytec.co.jp Polytec, Inc. (USA) North American Headquarters 1342 Bell Avenue, Suite 3-A Tustin, CA 92780 Tel. +1 714 850 1835 Fax +1 714 850 1831 info@polytec.com Midwest Office 3915 Research Park Dr. Suite A-12 Ann Arbor, MI 48108 Tel. +1 734 662 4900 Fax +1 734 662 4451 East Coast Office 25 South Street, Suite A Hopkinton, MA 01748 Tel. +1 508 544 1224 Fax +1 508 544 1225 Advancing Measurements by Light www.polytec.com