Testing of the NSC Electronics Module with the GSI Clover Detector

Similar documents
K 223 Angular Correlation

Week 11: Chap. 16b Pulse Shaping

Fast first practical help -- detailed instructions will follow- preliminary Experiment F80

Activities in Electronics Lab Associates are: Mrs. Arti Gupta, K.S.Golda, S.Muralithar & Dr.R.K.Bhowmik

The Influence of Crystal Configuration and PMT on PET Time-of-Flight Resolution

CAEN. Electronic Instrumentation. CAEN Silicon Photomultiplier Kit

Purpose This experiment will use the coincidence method for time correlation to measure the lifetime in the decay scheme of 57

ORTEC. AN34 Experiment 14 Nuclear Lifetimes and the Coincidence Method. Equipment Needed from ORTEC. Equipment Required from Other Manufacturers

COMPTON SCATTERING. Purpose. Introduction. Fundamentals of Experiment

Traditional analog QDC chain and Digital Pulse Processing [1]

A digital method for separation and reconstruction of pile-up events in germanium detectors. Abstract

Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors

The (Speed and) Decay of Cosmic-Ray Muons

The 2017 IEEE NSS-MIC. Industrial Presentation

TECHNICAL REPORT QUAD TIMING FILTER AMPLIFEIR MODULE

ORTEC Experiment 13. Gamma-Gamma Coincidence with Angular Correlation. Equipment Required

Real-Time Digital Signal Processors with radiation detectors produced by TechnoAP

Digital Signal Processing for HPGe Detectors

Physics Laboratory Scattering of Photons from Electrons: Compton Scattering

--- preliminary Experiment F80

CAEN Tools for Discovery

Cosmic Rays in MoNA. Eric Johnson 8/08/03

Silicon Photomultiplier Evaluation Kit. Quick Start Guide. Eval Kit SiPM. KETEK GmbH. Hofer Str Munich Germany.

A user-friendly fully digital TDPAC-spectrometer

NIM INDEX. Attenuators. ADCs (Peak Sensing) Discriminators. Translators Analog Pulse Processors Amplifiers (Fast) Amplifiers (Spectroscopy)

Instructions for gg Coincidence with 22 Na. Overview of the Experiment

Highly Segmented Detector Arrays for. Studying Resonant Decay of Unstable Nuclei. Outline

Bipolar Pulsed Reset for AC Coupled Charge-Sensitive Preamplifiers

Nyquist filter FIFO. Amplifier. Impedance matching. 40 MHz sampling ADC. DACs for gain and offset FPGA. clock distribution (not yet implemented)

A high-performance, low-cost, leading edge discriminator

Simulation of Algorithms for Pulse Timing in FPGAs

SPECTROMETRIC CHARACTERISTIC IMPROVEMENT OF CdTe DETECTORS*

nanomca 80 MHz HIGH PERFORMANCE, LOW POWER DIGITAL MCA Model Numbers: NM0530 and NM0530Z

Method for digital particle spectrometry Khryachkov Vitaly

LIFETIME OF THE MUON

GAMMA-GAMMA CORRELATION Latest Revision: August 21, 2007

NIM. ADCs (Peak Sensing) Analog Pulse Processors Amplifiers (Fast) Amplifiers (Spectroscopy) Attenuators Coincidence/Logic/Trigger Units

PARTICLE DETECTORS (V): ELECTRONICS

A Prototype Amplifier-Discriminator Chip for the GLAST Silicon-Strip Tracker

Performance test of the tig10 to be used for ECBR measurements with TITAN

PACS codes: Qx, Nc, Kv, v Keywords: Digital data acquisition, segmented HPGe detectors, clock and trigger distribution

Charge Sensitive Preamplifiers (CSP) for the MINIBALL Array of Detectors

experiment no. 3.5 Anti-Compton Spectroscopy

Contents. The AMADEUS experiment at the DAFNE collider. The AMADEUS trigger. SiPM characterization and lab tests

Digital coincidence acquisition applied to portable β liquid scintillation counting device

nanodpp datasheet I. FEATURES

Electronic Instrumentation for Radiation Detection Systems

ORTEC. Research Applications. Pulse-Height, Charge, or Energy Spectroscopy. Detectors. Processing Electronics

LaBr 3 :Ce, the latest crystal for nuclear medicine

nanomca datasheet I. FEATURES

Detecting and Suppressing Background Signal

Coincidence Rates. QuarkNet. summer workshop June 24-28, 2013

Comparisons Between Digital Gamma-Ray Spectrometer (DSPec) and Standard Nuclear Instrumentation Methods (NIM) Systems

ORTEC Experiment 19. Gamma-Ray Decay Scheme and Angular Correlation for 60 Co. Equipment Required. Purpose. Introduction

PH2510 Nuclear Physics Laboratory Use of Scintillation Counters (NP5)

On the initiation of lightning in thunderclouds (Instrumentation, Supplementary information)

Experiment 10. The Speed of Light c Introduction Apparatus

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A

Contents. Why waveform? Waveform digitizer : Domino Ring Sampler CEX Beam test autumn 04. Summary

e t Development of Low Cost γ - Ray Energy Spectrometer

An ASIC dedicated to the RPCs front-end. of the dimuon arm trigger in the ALICE experiment.

AGATA Local Level Processing. - Pulse. Shape. Analysis - AGATA Week, LNL, September 15-19, 2003 Thorsten Kröll (TU München) for the AGATA PSA Team

AN-DPP-003 Rev A2: Using the DP5 with HPGe USING THE DP5 WITH GERMANIUM DETECTORS

PET Detectors. William W. Moses Lawrence Berkeley National Laboratory March 26, 2002

AGATA preamplifiers: issues and status

mesytec GmbH & Co. KG Wernher-von-Braun-Str Putzbrunn Germany Tel.: Fax: REPRESENTED BY:

MSCF-16-LN (Data sheet V5.0_01)

PoS(PhotoDet 2012)022

Development of Large Area and of Position Sensitive Timing RPCs

Spontaneous Fission Spectrum of Neutrons from 252 Cf with Kinetic Energies Less than 1 MeV. Suraj Bastola. A senior thesis submitted to the faculty of

Solid-State Photomultiplier in CMOS Technology for Gamma-Ray Detection and Imaging Applications

MASE: Multiplexed Analog Shaped Electronics

and N(t) ~ exp(-t/ ),

Development of a 256-channel Time-of-flight Electronics System For Neutron Beam Profiling

GRETINA. Electronics. Auxiliary Detector Workshop. Sergio Zimmermann LBNL. Auxiliary Detectors Workshop. January 28, 2006

A Novel Design of a High-Resolution Hodoscope for the Hall D Tagger Based on Scintillating Fibers

Front-End Electronics and Feature-Extraction Algorithm for the PANDA Electromagnetic Calorimeter

nanomca-sp datasheet I. FEATURES

High collection efficiency MCPs for photon counting detectors

Analog-to-Digital-Converter User Manual

Performance Assessment of Pixelated LaBr 3 Detector Modules for TOF PET

Evaluation of Multi-Channel ADCs for Gamma-ray Spectroscopy

Digital Signal Processing Electronics for Nuclear Physics Applications

Diamond sensors as beam conditions monitors in CMS and LHC

The software and hardware for the ground testing of ALFA- ELECTRON space spectrometer

Pixel hybrid photon detectors

Analog Peak Detector and Derandomizer

Physics 342 Laboratory. Scattering of Photons from Free Electrons: Compton Scattering

Model 673 Spectroscopy Amplifier and Gated Integrator Operating and Service Manual

SPECTROMETRIC DETECTION PROBE Model 310. Operator's manual

1 Purpose of This Lab Exercise:

A Readout ASIC for CZT Detectors

PHYSICS ADVANCED LABORATORY I COMPTON SCATTERING Spring 2002

Channel-Plate Photomultipliers

Development of a Multi-Channel Integrated Circuit for Use in Nuclear Physics Experiments Where Particle Identification is Important

Measurement of X-ray Photon Energy and Arrival Time Using a Silicon Drift Detector

Amptek sets the New State-of-the-Art... Again! with Cooled FET

Study of Silicon Photomultipliers for Positron Emission Tomography (PET) Application

Quad Analog-to-Digital Converter Technical Documentation

ORTEC Experiment 9. Time Coincidence Techniques Applied to Absolute Activity Measurements. Equipment Required. Purpose.

Transcription:

Testing of the NSC Electronics Module with the GSI Clover Detector Rakesh Kumar 1, P. Queiroz 2, H.-J. Wollersheim 2 (Tutor) 1 Inter University Accelerator Centre Aruna Asaf Ali Marg Post Box No 10502 New Delhi 110067 India Summer Student Program 2005 2 Gesellschaft für Schwerionenforschung Planckstraße 1 D-64291 Darmstadt Germany This report refers to the testing of the NSC (Nuclear Science Centre, New Delhi) Electronics Module for Clover Ge-detectors performed at the RISING (Rare Isotope Spectroscopic INvestigation at GSI)) group in GSI. The aim of this investigation is to compare its performance with that of a conventional electronics apparatus. 1 Introduction This paper examines the results of the tests performed on the NSC Electronics Module for Clover Ge-detectors. These investigations were performed at GSI and their goal is to compare the behaviour of the NSC Module with that of a conventional electronic module. In the following sections, information regarding the setting, the electronics and the perfomance characteristics for both NSC Module and conventional electronics (energy and time resolutions) is to be found. 2 Materials and Methods The detector used was the EURISYS MESURES 4 fold segmented Super Clover detector N 101 of the VEGA-array (Versatile and Efficient GAmma Detector). The first Clover Detector was developed in France by EURISYS MESURES in the frame of the EUROGAM collaboration. It consists of four coaxial N-type Germanium detectors, arranged in a four leaf clover shape. Each Germanium crystal has a length of 140 mm and a square front face with two flat parts at 90 along the whole length and two tapered parts at an angle of 15. Its output signals characteristic values are presented in Table??. The detector was delivered to the GSI (Gesellschaft für Schwerionenforschung, Darmstadt) laboratory in April 1997 [?]. The NSC Module aims to replace the various electronics modules necessary to operate a Clover Detector in a double width NIM mode. This is achieved by placing all the necessary components into one single module capable of handling four channels, i.e., 4 TFA (Time Filter Amplifier), 4 CFD (Constant Fraction Discriminator), 4 main amplifiers (with 3 µs shaping constant), 4 gate generators and a coincidence unit [?]. The NSC Module arrived at GSI in 2005. The results for the Clover detector read out were obtained with the four channels of the NSC Module and with the conventional electronics channel. Each channel has to be connected to one of the four detector contacts. Table?? shows the correspondence between the detector s centre contacts (black, blue, green and red) and the electronics channels (both from the NSC Module and the conventional electronics). 1

2 Rakesh Kumar, P. Queirozi, H.-J. Wollersheim Tab. 1: Properties of the signals from the Clover Detector. Rise Fall Under- Detector Time Time shoot Contact ns µs % Black 120 ± 60 80 ± 10 2 ± 1 Blue 120 ± 60 80 ± 10 3 ± 1.5 Green 130 ± 70 80 ± 10 1.5 ± 0.5 Red 120 ± 60 80 ± 15 2 ± 1 The 60 Co γ-source (which has two well known transitions at 1173 and 1333 kev [?]) was attached to a plastic detector and placed before the front face of the crystals (at a 15-25 cm distance). The plastic detector served as a trigger for the electronics, as it has a very good time resolution ( 1 ns), although a very poor energy resolution. 60 Co β -decays into 60 Ni through a cascade of two E2 transitions: 4 + 1173 kev 2 + 1333 kev 0 + This process results in two almost simultaneous γ-rays being emitted, since the 4 + and the 2 + states are very short-lived (τ 1 ps). When one of the γ-rays hits the plastic detector and another one hits one of the Ge-detectors, the electronics are triggered (using the AND unit and TRIGGER 2, as in Figure??). A delayed Ge-detector signal is then used as a stop. Several different sets of measurements were taken, both by varying the threshold value and the CFD delay for the Ge-detectors. The threshold values used were the normal one (cutting just the noise) and the maximum one (cutting the Compton edge and leaving only the two photo-peaks). LEA software was used to acquire the spectra and to analyse them. 3 Results and Discussion 3.1 Resolution 3.1.1 Energy Resolution The energy resolution was measured with an ADC (Analog to Digital Converter). Its calibration was done by adjusting Gaussian curves Tab. 2: Correspondence between electronic channels, detector contacts and TDC and ADC inputs. Electr. Detector ADC TDC Channel Contact Input Input NSC A Black ENERGY1 TIME1 NSC B Blue ENERGY2 TIME2 NSC C Green ENERGY3 TIME3 NSC D Red ENERGY4 TIME4 conv. Black a ENERGY5 TIME5 a In some parts of the experiment, the conventional electronics channel was connected to different detector contacts. In those cases, this is clearly stated. Tab. 3: Calibration of the ADC and of the TDC for the various channels. Electronics ADC TDC Channel kev/channel ns/channel NSC A 1.44 0.24 NSC B 1.53 0.25 NSC C 1.16 0.26 NSC D 1.56 0.24 conv. 0.54 0.28 to the two photo-peaks of 60 Co and taking the distance between their mean values (given in ADC channels) as 160 kev (Table??). The results obtained for the energy resolution can be found in Table 4. It can be observed that there are no major variations for the two different threshold values. Initially there was an undershoot in the output signal and this was not adjustable from the front panel potentiometer. The polezero adjustment problem was corrected by replacing the fixed value of the feedback resistor 15kΩ in the shaping amplifier polezero circuit with a variable 50kΩ potentiometer. It was observed that the polezero could be corrected by varing the potentiometer to a value of 13.5kΩ.All the daughter cards were replaced by two 27kΩ resistors used in parallel to get 13.5kΩ. The resolution testing was done and it was found that the FWHM is same in both the cases. The Base Line Restorer (BLR) should be ajusted in order to get better resolution. Results for the Conven-

Testing of the NSC Clover Electronics Module 3 Fig. 1: Electronics block diagram. tionla electronics (conv) and for NSC s modules Module1 (NSC 1) and Module2 (NSC2) are tabulated in in Table 4. Tab. 4: Energy resolution for the conventional electronics and NSC s module with 60 Co Source. Energy resolution at 1332 KeV conv NSC 1 NSC 2 kev kev kev BLACK 3.2 3.2 3.2 BLUE 3.2 3.2 3.2 GREEN 2.9 2.9 2.9 RED 3.2 3.2 3.2 3.1.2 Time Resolution The time resolution was measured using first a TDC (Time to Digital Converter) and later a TAC (Time to Amplifier Converter). Both the TDC (Table??) and the TAC were calibrated using a pulse generator. The distance between the mean values of two peaks that are 24 ns appart was taken as a reference. In Table?? are the results for the TDC. The data for the maximum threshold value was not collected simultaneously, but for one NSC Module channel at a time. The conventional electronics channel was connected to the same detector contact as the NSC Module channel for every measurement. The conventional electronics channel presents a rather worse resolution for both cases. Concerning the normal threshold value, the possibility that the CFD of the Ge-detectors has a smaller ideal delay was investigated and the results can be found in Table??. For a 32 ns delay the resolution of the conventional electronics is similar to that of the NSC module. For 20 ns it worsens again. For the maximum threshold value another problem arises: the CFD starts working in a LE (Leading Edge) mode, which deteriorates the quality of the spectra. When the TDC delay is increased, the spectra deteriorates even further. The results for the TAC are very similar, and therefore not interesting. 3.2 Spectra 3.2.1 Energy Spectra Figures?? and?? are examples of some of the energy spectra collected with the conventional electronics. The spectra from the NSC module look essentialy the same.

4 Rakesh Kumar, P. Queirozi, H.-J. Wollersheim Tab. 5: Time resolution. For the maximum threshold value the convencional electronics channel (TIME5) was connected for every measurement to the same detector contact as the NSC Module channel. FWMH (ns) Channel Normal Maximum Thr. Thr. NSC conv. TIME1 11.4 6.7 12.1 TIME2 8.3 5.6 14.2 TIME3 10.2 5.2 13.1 TIME4 10.9 6.2 8.9 TIME5 15.1 - - Tab. 6: Time resolution from the conventional eletronics channel using different CFD delays and detector contacts. Detector CFD delay (ns) Contact Conv NSC 1 NSC 2 Black - 7.0 15.1 Blue - 7.0 - Green - 9.8 - Red 13.2 10.1 13.3 As desired, with the maximum threshold value (Figure??) the Compton edge is no longer present, but the two photo-peaks can still very clearly be seen. Fig. 3: Energy spectrum from the conventional electronics channel with maximum threshold. Figure?? refers to the conventional electronics. In Figure?? one can see unexpected smaller peaks in the NSC module normal threshold spectrum, which desappear in Figure?? with the setting of the maximum threshold. The tests are not conclusive enough to establish a reason for this behaviour. Fig. 4: Time spectrum from the conventional electronics channel with normal threshold and a TDC 64 ns delay. Fig. 2: Energy spectrum from the conventional electronics channel with normal threshold. 3.2.2 Time Spectra Figures??,?? and?? are examples of the time spectra collected. Fig. 5: Time spectrum from the NDC module channel D with normal threshold. References [1] Y. Kopach and H.-J. Wollersheim, Testing of the GSI Clover Detector under Experimental Conditions, Gesellschaft für Schwerionenforschung, Darmstadt (March 2000).

Testing of the NSC Clover Electronics Module 5 Fig. 6: Time spectrum from the NDC module channel D with maximum threshold. [2] S. Venkataramanan et al., Technical Report on Clover Electronics Module, Nuclear Science Centre, New Delhi (2003). [3] S. Eidelman et al., Particle Physics Booklet, Particle Data Group (July 2004).