SUSS Mask Aligner. Purpose: To expose photoresist on a wafer using a photomask

Similar documents
Electron Microscopy Sciences

DISCO DICING SAW SOP. April 2014 INTRODUCTION

Heidelberg µpg 101 Laser Writer

Bruker Dimension Icon AFM Quick User s Guide

(ksaligner & quintel resolution)

MicroPG 101 Pattern Generator Standard Operating Procedure Draft v.0.2

Figure 1 The Raith 150 TWO

UFNF YES Image Reversal & HMDS Oven Revision 6.0 1/22/2014 Page 1 of 5. YES Image Reversal and HMDS Oven SOP

Micro Automation- Model 1006 Dicing Saw Instructions. Serial # Rev 2 ( R.DeVito) Location Chase 1

ASSEMBLY AND ADJUSTMENT

Rotary Fixture M/V/X CLASS LASER SYSTEMS. Installation and Operation Instructions

EDGE2 DUAL MONITOR ARM

Installing Your Electronic Deadbolt

RH-412 STEEL DOORS INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS RH 412 STEEL DOORS

Installation Operation Care

Installation and Assembly: In-wall Mount for 32" to 71" Flat Panel Screens

Replacing the Reciprocator on the SWF Compact Series Machine (601C and 1201C)

Standard Operating Procedure

Part 5-1: Lithography

Strata. urniture. Adriana Instructions. Parts in the Arm Box: Parts in the Body Box: Watch our assembly videos at

Trebuchet Parts List and Overview

Lawson SSR Registration System Instructions

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE

Cap frame (Shaft-less type)

AZ 1512 RESIST PHOTOLITHOGRAPHY

The Useless Machine. Parts Only - Build Guide v0001

Pilaster with J-Channel Back Plate

JK Spartacus Stamped Front Bumper

Parts List (Continued) 18 each: #8 x 0.625" Truss Head Phillips Screw 20 each: #8 x 1.75" Flat Head Phillips Screw 4 each: #8 x 1.25" Flat Head Philli

Bruker Dimension Icon AFM Quick User s Guide

Instructions for Catalina Round/Square Extension Tables

Outline. 1 Introduction. 2 Basic IC fabrication processes. 3 Fabrication techniques for MEMS. 4 Applications. 5 Mechanics issues on MEMS MDL NTHU

1.2. Make sure the viewing screen is covered (exposure to liquid N 2 may cause it to crack).

Setting up and Using Digital Micrometer Controlled Lapping Fixtures

Basic Users Manual for Tecnai-F20 TEM

Nancy s Knit Knacks LLC 4 Yard Option Upgrade Kit Assembly Instructions and User Manual

Installation Instructions for Lambo Door Mechanism. 1. Lift the hood and remove left and right shield of front wheel.

Mechanical Frappe Press

MM540 Installation Instructions IMPORTANT SAFETY INSTRUCTIONS - SAVE THESE INSTRUCTIONS

POWER PET. Low-E Automatic Patio Pet Door Installation and Operating Instructions

Table Saw Small Parts Sled Instructions

Wedge Bonder --- West Bond E

The Useless Machine. DIY Soldering Edition. Instruction Guide v0004

ISONIC PA AUT Spiral Scan Inspection of Tubular Parts Operating Manual and Inspection Procedure Rev 1.00 Sonotron NDT

Strata. urniture. Mission Rim Instructions. Parts in the Arm Box: Parts in the Body Box:

STANDARD OPERATING PROCEDURE STAT-SITE M HGB

RAITH e-line OPERATING INSTRUCTIONS

LCD/Plasma Mount Manual

Installers guide Deadbolt 02.

1. Features Calibration Compensator Lock Button Circular bubble Line-of-sight...10

PFW 6875 Installation Guide Installationsanleitung, Guía de Instalacíon, Guida de Installazione, Guide d Installation, Installatie gids

Photolithography Technology and Application

S A F T E Y W A R N I N G

User Operation of JEOL 1200 EX II

18600 Angular Momentum

Vastex Manual Press Registration System Assembly and Operations Manual

How to Build a Game Console. David Hunt, PE

ZIERICK MANUFACTURING CORPORATION

98/9927. Devices covered by these instructions: Surface Vertical Rod Exit Device

KMPR 1010 Process for Glass Wafers

2. OPERATING PROCEDURES 3. WARNING AND SAFETY HAZARDS 4. QUALITY CONTROL 5. MAINTENANCE 6. INSTALLATION 7. WARRANTY AND LIMITATIONS

PACLEASE Hybrid Sacagawea VK-1222 Display Instructions / PAGE 1 - General Layout

Basic P-XRD instructions for Operating the Instrument

С 800 CASSIDA C 800 HIGH SPEED COIN COUNTER

i- Line Photoresist Development: Replacement Evaluation of OiR

SOP for Micro Mechanical Analyzer

MICRO AND NANOPROCESSING TECHNOLOGIES

TITAN2-EDGE Public Access Computer Station Dual Track

CABINETRY Assembly Instructions

Usage and Assembly Instructions

Assembly Instructions

Olympus Digital Microscope Camera (DP70) checklist

Technology for the MEMS processing and testing environment. SUSS MicroTec AG Dr. Hans-Georg Kapitza

Renishaw InVia Raman microscope

434 Cobi Chair - Seat Shell Replacement

SETUP INSTRUCTIONS If you would like to tell us about your experience with your setup instructions please us at

SWIFT SERIES M2252DGL MICROSCOPE

Lab 5: Brewster s Angle and Polarization. I. Brewster s angle

Sentinel Series Cigar Humidor End Tables

Installation and Assembly: In-wall Mount for 32" to 71" Flat Panel Displays

Rocketry Challenge - Technical Document I

FiLMiCPro v6 Quick Start Guide

APES HD-7700 Version Operator s Training Manual

Photolithography II ( Part 2 )

SmartView Mounting Frame 3 Wide x 3 Deep Video Wall Display Installation Guide

RB12 AERO - SINGLE INSTALLATION MANUAL. April 2017

TMA812 Tilt Mount Adapter for NEAR A8 and A12 Loudspeakers Installation Instruction Manual

Standard Operating Procedure of Atomic Force Microscope (Anasys afm+)

Deluxe Exterior Solar Shades

Installation And Care Instructions. Vertical Honeycomb Shades

Closet Wall Drawer Unit

STEINBERGER TRANSTREM (TYPE 2) TECHNICAL DOCUMENT

Disc Sander Angle Jig Instructions

STERILIZATION The following are GUIDELINES for product sterilization.

Installation Guide. SN-LM42 Brilliance Series Photoluminescent Area of Rescue Signs

Gardall Locks Operating Instructions

INSTALLATION OF THE TRACK FOR THE STRAIGHT SIDE STEEL LADDER

Installation tutorial for Console Customs Xbox Mode Dual Button (RFX-5B) Rapid fire Microchip for all Wired and Wireless controllers

Mini Cooper Lock Actuator

Transcription:

SUSS Mask Aligner Purpose: To expose photoresist on a wafer using a photomask Overview This SOP will go over how to use the machine for basic exposures. This will include commonly used controls and frequently modified settings of the machine. Preparation and Precautions: 1. Locate the SUSS mask aligner in the photolithography bay 2. Ensure that the mask aligner is on. The aligner is usually left in change mask mode. It is from this state that the SOP operates from.

3. Complete the Dehydration SOP, Photoresist Spin SOP, and Soft Bake SOP before using the mask aligner. Also make sure to bring the photomask being used in the process. Procedure 1. Place the photomask on the mask holder located to the left of the machine. The photomask should be placed feature side up (reddish side up). Press the photomask up and to the left so that it rests against the screws to ensure proper placement. 2. Press the Enter key to activate the photomask vacuum that will ensure it is held into place. Also, press down on the metal clip below the photomask causing the black bar to pring up and help lock the mask into place.

3. Take the photomask assembly, flip it over, and move it into the SUSS aligner as shown in the picture. Once it is in place press the Change Mask button to inform the machine that there is a photomask in the aligner. 4. Press the Load button to begin the loading procedure for your wafer. Pull out the slide located under the photomask assembly.

5. Locate the three small pins on the wafer holder. Place the flat of your wafer against the two pins at the bottom and move the wafer up against the last one on the left. Once the wafer is in position, push the assembly into the machine and press the Enter button. 6. The aligner will go through a process where it will force the wafer and mask into contact before separating the two by a small amount. Once this is done we can begin aligning the wafer. We will now go over some of the alignment controls, but if this is a first exposure or one that is not aligning to existing features then skip ahead to step 9. 7. Alignment is done in two steps, first we have to find our alignment mark in the photomask and then in the wafer. This step will show and describe all of the controls to help locate the alignment marks in the photomask:

a. This knob on the top of the machine allows us to change the focus on the front side alignment microscope. Your first task should be to make sure the photomask is in focus on the screen by adjusting this knob. b. This small knob located just above either microscope will allow us to move the microscope independently in the horizontal direction. Use this to help in finding the alignment mark.

c. This knob on top of the microscope assembly will rotate the two microscopes with respect to each other. d. These knobs will adjust the light level for the microscope allowing us the best possible view.

d. These controls will move the entire microscope assembly in one piece. 8. Alignment is done in two steps, first we have to find our alignment mark in the photomask and then in the wafer. This step will show and describe all of the controls to help locate the alignment marks in the wafer: a. This small knob on the right of the machine will allow us to rotate our wafer without moving the photomask.

b. This large knob located on the right of the machine will move the wafer in the horizontal direction c. This large knob located on the left of the machine will move the wafer in the vertical direction. 9. Now that our wafer is aligned (if necessary), we are ready to set up the exposure. You will want to press the Edit Parameters button shown below.

10. The key feature here that we will need to change is the exposure time. Press the left and right button until that parameter is being edited. Use the up and down button to enter in the correct exposure time. Due to frequent light bulb replacements common exposure times cannot be written here, but reference the log sheet accompanying the machine to know how long to expose your photoresist. Shipley 1813 needs an exposure of roughly 115 and an exact exposure time can be calculated using a power meter. Press Edit Parameters again when you have the correct value. 11. Now we should be ready to start the exposure. Hit the Exposure button to start the sequence. Be sure not to look at the light during exposure because it can be potentially damaging to the eyes.

12. Once the exposure is complete, the aligner will notify you that it is safe to pull out the wafer assembly and retrieve your wafer. Do so and push the assembly back into place. 13. Before we retrieve our mask we have to make sure that the machine is returned to its default settings so the next user can easily use the machine: a. Rotate the knobs on either side of the stage so that they read a value of 10-0 as shown in the picture. This is the midpoint of the translational motion of the wafer and will ensure that the next user will have a full range of motion in any direction b. Adjust the wafer rotation using the small knob next to the stage. Look to the tab below the wafer loading handle and try to leave it parallel to the machine (as in the first picture). c. Lower the light levels using these knobs to ensure long life for the light bulbs.

14. Press the Change Mask button and remove the mask assembly and place it over to the left side of the machine. Press Enter to release the vacuum and pull the arm back below the photomask until it locks into place. Retrieve your photomask and proceed to the development SOP.