Advanced digital image processing for clinical excellence in fluoroscopy

Similar documents
Management of pediatric radiation dose using Philips fluoroscopy systems DoseWise: perfect image, perfect sense

Digital radiography (DR) post processing techniques for pediatric radiology

SYLLABUS. 1. Identification of Subject:

MUSICA Nerve Center. Artificial Intelligence. Intelligent tools for your Digital Radiography workflow. Fluoroscopy. Workflow Optimization

Detector technology in simultaneous spectral imaging

GE Healthcare. Senographe 2000D Full-field digital mammography system

Multi-Access Biplane Lab

Philips EasyDiagnost Eleva

Maximum Performance, Minimum Space

siemens.com/luminos-fusion Luminos Fusion The 2-in-1 system that fits your needs and fits your budget

Invisible sophistication. Visible simplicity. CS Welcome to the simplicity of compact panoramic imaging

Some operation methods show in the catalog reguire optional eguipment

Do you have any other questions? Please call us at (Toll Free) or , or

Maximizing clinical outcomes

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of

160-slice CT SCANNER / New Standard for the Future

R&F X-ray systsem. Savings With Every Exposure

GE Healthcare. Precision 500D. Digital radiography and fluoroscopy system

Visibility of Detail

X-RAYS - NO UNAUTHORISED ENTRY

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

GE Healthcare. Essential for life. Senographe Essential Full-Field Digital Mammography system

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II

INTRODUCTION TO FLEXIBLE BRONCHOSCOPY. Fluoroscopy Synopsis HENRI G COLT MD SECOND EDITION THE BRONCHOSCOPY EDUCATION PROJECT SERIES

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

10/15/2012 SECTION III - CHAPTER 6 DIGITAL FLUOROSCOPY RADT 3463 COMPUTERIZED IMAGING

Grid-like contrast enhancement for bedside chest radiographs acquired without anti-scatter grid

Truly flexible to meet your clinical needs

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity)

Digital Radiography : Flat Panel

YSF - 300/DAR i

Redefining Ergonomics

10/26/2015. Study Harder

10/3/2012. Study Harder

Performance and care. all in one

X-RAY MEDICAL EQUIPMENT

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing

ORTHOPANTOMOGRAPH OP 2D Quality and design

Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) MEDIX VOL. 61 P.8 P.11

LECTURE 1 The Radiographic Image

Control and confidence all around. Philips EP cockpit people focused solutions for heart rhythm care

Ysio Max. The most direct way to the image. Answers for life.

Fluoroscopy - Chapter 9

Philips XPER FD10C R7.0.4

Nuclear Associates

Beam-Restricting Devices

IBEX TECHNOLOGY APPLIED TO DIGITAL RADIOGRAPHY

Photomultiplier Tube

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

Lesson 06: Pulse-echo Imaging and Display Modes. These lessons contain 26 slides plus 15 multiple-choice questions.

SONIALVISION G4 Multi-purpose Digital R/F System C506-E075

SONIALVISION G4 Multi-purpose Digital R/F System C506-E075

FMT18 FLOOR MOUNTED SYSTEM

1-1. GENERAL 1-2. DISCOVERY OF X-RAYS

Luminos RF Classic. Where value meets performance.

SONIALVISION G4 Multi-purpose Digital R/F System

SAFIRE. Sinogram Affirmed Iterative Reconstruction. Answers for life.

Open. the Digitized world. Fuji Computed Radiography

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah

Medical Imaging. Digital R/F remote controlled table. radiology ahead

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Instant DR in Jordan

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

BASICS OF FLUOROSCOPY

OTC18 OVERHEAD TUBE CRANE SYSTEM

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

Iterative Reconstruction in Image Space. Answers for life.

SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT. Exam Content Specifications 8/22/2012 RADT 3463 COMPUTERIZED IMAGING

Image Quality. HTC Grid High Transmission Cellular Grid provides higher contrast images

CR Basics and FAQ. Overview. Historical Perspective

C501-E029C. RADspeed. GENERAL RADIOGRAPHIC SYSTEM Automatic

Brilliance in everything Philips CT products and services

China Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW, 150kV Radiography System - HF50-R

2D, 3D CT Intervention, and CT Fluoroscopy

The digital 2-in-1 solution for fluoroscopy and radiography. AXIOM Luminos drf.

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist

C506-E064. Full digital system. Printed in Japan A-NS

I AM CONFIDENT SOPIX SERIES TECHNICAL SPECIFICATIONS. A successful X-ray every time with minimal exposure to radiation

Luminos drf Max. Taking 2-in-1 to the MAX in radiography and fluoroscopy. siemens.com/luminos-drf-max

Digital radiography: Practical advantages of Digital Radiography. Practical Advantages in image quality

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development

PD233: Design of Biomedical Devices and Systems

A Short History of Using Cameras for Weld Monitoring

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT)

RADspeed Pro. EDGEpackage C501-E041C

Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck

Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare

THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY

Mammography Solution. AMULET Innovality. The new leader in the AMULET series. Tomosynthesis, 3D mammography and biopsy are all available.

Features and Weaknesses of Phantoms for CR/DR System Testing

X-ray Tube and Generator Basic principles and construction

DR _ solutions. We understand that customers don t need just products, they want. solutions

CXDI-70C WIRELESS SPECIFICATIONS

Image Quality. HTC Grid High Transmission Cellular Grid provides higher contrast images

Architecture of Quality Imaging Mary K. Henne, MS, CNMT, RDMS, RVT Ultrasound Education Specialist GE Healthcare

Overview. Professor Roentgen was a Physicist!!! The Physics of Radiation Oncology X-ray Imaging

Transcription:

Dynamic UNIQUE Digital fluoroscopy solutions Dynamic UNIQUE Advanced digital image processing for clinical excellence in fluoroscopy André Gooßen, PhD, Image Processing Specialist Dörte Hilcken, Clinical Specialist Detlef Mentrup, PhD, Clinical Scientist Mathias Schlüter, PhD, Image Processing Specialist Digital image processing challenges in fluoroscopy Modern fluoroscopy poses several challenges for digital image processing. First, low-dose settings are commonly used today in clinical practice (see side box and Ref. 1), particularly for children, who are most sensitive to the damaging effects of ionizing radiation. While a reduction in dose is always desirable 2,3, these settings inevitably result in images with high quantum noise, which may be difficult for radiologists to read. Second, because fluoroscopy means live viewing, relevant diagnostic information in the images must be visible straightaway at the correct level of brightness and contrast. Furthermore, because the human eye is very sensitive to temporal variations, the contrast and brightness impression must be stable over time. This is particularly challenging in view of the rapid temporal variations occurring in fluoroscopy, particularly with respect to: The body region under examination, e.g. when examining the esophagus The field-of-view, e.g. when moving the patient or modifying the collimated area Areas of direct radiation The distribution of the contrast agent The dose level Third, diagnostically relevant details such as catheters or small vessels must be visible with great clarity, even in challenging viewing conditions with large variations in tissue radio-density. Clarity is essential to reduce exam length and dose, and is particularly important for pediatric patients, as well as for those who are immobile or very sick. Low-dose fluoroscopy with unrivaled Philips GCF technology During the past two decades, the dose level in fluoroscopy has been reduced to only a fraction of the level needed with continuous fluoroscopy 4,5,6. This was achieved by hardware innovations enabling pulsed fluoroscopy. In the beginning, two variants existed: generator-pulsed and grid-pulsed fluoro. Both techniques include after-pulse adjustments for exposure optimization. A second major breakthrough happened in 1995, when Philips introduced in-pulse control for grid-controlled fluoroscopy (GCF) for the first time. GCF enables independent optimization of each pulse in real time by fast readout of the dose rate signal 7. Even today, Philips is the only vendor offering this technology.

Digital image processing requirements in fluoroscopy Addressing these challenges requires new image processing technology that must comply with the real-time requirements of fluoroscopy. To enable consistent use of low-dose settings, this technology must allow efficient noise suppression and correct representation of diagnostic details even at a low dose level. To address the imperative for correct brightness and contrast, it must provide a robust, fully automatic stabilization of image brightness over time, as well as fast enhancement of the diagnostically important image content. To decrease the physical burden and manage the radiation exposure that long examinations and re-examinations place on patients, it must promote correct image representation that reduces the need for manual adjustments, enabling operators to focus on the patient and the procedure, rather than the device and its settings. More clarity and confidence with Dynamic UNIQUE Radiologist s challenges Difficulty distinguishing between very small anatomical structures/ details and noise. Low dose settings add to the difficulty and can result in longer fluoroscopy times and could make diagnosis more challenging. Difficulty observing the whole image immediately. The eyes must adjust, which adds time to diagnosis and causes eye fatigue. Difficulty distinguishing small, low-contrast structures from high-contrast structures, possibly resulting in more exposures. Philips answers Intelligent spatio-temporal noise suppression Temporal (inter-frame): Intelligent motion detection for signal averaging in regions without motion Spatial (intra-frame): Based on a physical noise model for consistent noise suppression Temporal brightness stabilization Real-time adjustment of contrast and brightness Multiscale image enhancement in real-time Harmonization of high-contrast and lowcontrast structures Very low computation time for processing every frame Benefits Reduced noise impression -improved detail visibility Anatomical details are preserved in every image No artifacts of moving structures, no lag effect No shadows of catheters, tubes and lines Less risk to miss important details More confidence Optimal image stability for fluoroscopy and spot images at any time Comfortable viewing at high quality Less risk to re-acquire Fast and stable Less lost time through less time needed to adapt to the luminosity Quicker workflow More confidence Less tiring for the eyes Optimal representation and visibility of diagnostically important information Enhancement of diagnostic details (e.g., small vessels, catheters) No enhancement artifacts No image clipping Less risk to miss important details More confidence With Dynamic UNIQUE, Philips offers a modern, elaborate, multi-scale digital image processing that addresses the diagnostic challenges of fluoroscopic X-ray examinations. Exploiting the speed and performance of modern data processing hardware to permit high-quality image processing in real-time. It features intelligent spatio-temporal noise suppression, temporal brightness stabilization, and real-time multiscale image enhancement. Conclusion Dynamic UNIQUE is a modern image processing technology, designed to support the operator in performing fluoroscopic examinations fast and safely, even at low dose levels. It meets the challenges of modern fluoroscopy by combining a robust, artifact-free, easy-to-use multiscale image enhancement with intelligent noise suppression. The diagnostic information of each frame is displayed in real time, with great clarity, the correct brightness, and very low latency. The image representation is consistent and stable even when the image content varies rapidly. This new level of image quality is also available when documenting clinical cases. All Philips fluoroscopy systems equipped with Dynamic UNIQUE offer as a standard feature the ability to record fluoroscopy sequences without restrictions on duration.

Intelligent spatio-temporal noise suppression Dynamic UNIQUE combines noise suppression by temporal averaging (inter-frame) and spatial noise suppression (intra-frame). Temporal noise suppression Temporal averaging, which accumulates the dose of successive frames, results in a depiction of static anatomical structures that mimics high-dose acquisition. However, it may fail in areas of motion. Dynamic UNIQUE suppresses temporal averaging when movement is detected, preventing lag and shadowing of moving structures. Spatial noise suppression * In moving regions of the image and if the sequences are very short, Dynamic UNIQUE applies spatial noise suppression. It uses a physical noise model to assess the contrast-to-noise ratio, which allows for structure-adaptive noise suppression within a single frame. The strength of the spatial noise suppression is adapted to the strength of the temporal noise reduction in adjacent areas. As a result, the image noise level is kept consistently low over the entire image. Figure 1: Example of spatio-temporal noise suppression. Noise Conventional noise suppression Dynamic UNIQUE noise suppression Lag effect No lag effect Original image without noise suppression. Noise suppression by simple temporal averaging produces a strong lag artifact of the moving catheter such that the actual catheter position cannot be determined. Dynamic UNIQUE reduces noise efficiently through the intelligent combination of spatial and temporal noise suppression, without image lag.

Temporal brightness stabilization When the dose level or the field-of-view is modified (e.g., by modifying its size, by shifting it from the body center to the periphery, or by moving the patient), Dynamic UNIQUE automatically determines the signal level in the anatomical region and maps it to an appropriate brightness level. Areas of direct radiation are reliably detected and excluded from the brightness adjustment *. In contrast, simpler methods include areas of direct radiation in adjustments, resulting in dark images and obscured anatomy, as well as annoying flickering. Figure 2: Example of brightness stabilization in a moving patient. Fluoroscopy sequence without temporal brightness stabilization. The anatomical details are visible only after correct positioning of the anatomy. Dynamic UNIQUE determines the signal level in the anatomical region reliably detecting areas of direct radiation. It maps this signal level to an appropriate brightness level regardless of positioning.

Real-time multiscale image enhancement Dynamic UNIQUE is a second-generation multiscale image enhancement that provides a high level of detail, according to the diagnostic need, in real time. Each and every single frame is fully processed, without significant additional computation time (below 14 ms for fluoroscopy, below 55 ms for exposures), even at very high frame rates (up to 30 fps). Therefore, the diagnostically important image content is optimally enhanced according to the clinical task. Dynamic UNIQUE features a perfect harmonization of areas filled with contrast agent, soft tissue, and bones, even in challenging viewing conditions with large variations of the radio-density of the tissue. Contrary to traditional multiscale enhancement algorithms, the novel Dynamic UNIQUE processing strictly separates global and local contrast enhancement, preventing undesired enhancement artifacts *. If desired, it is easily customizable according to the diagnostic need.. Figure 3: Example of representation of the diagnostic details for a lateral myelography. Too bright Corrected low density area Acceptable Too dark Dense area made transparent Dose-proportional detector values are mapped to gray values without real-time multiscale image processing. Image information is lost in the dark and bright areas. Dynamic UNIQUE provides a much higher level of detail in all image areas in real-time.

References 1 Federal Office for Radiation Protection. Publication of updated diagnostic reference levels for diagnostic and interventional X-ray examinations (2016). Available at http://www.bfs.de/en/topics/ion/medicine/ diagnostics/reference-values/reference-values.html 2 Goske MJ et al. The Image Gently Campaign: Working Together to Change Practice. American Journal of Roentgenology 190:2, 273-274 (2008) 3 Brink JA, Amis ES. Image Wisely: A Campaign to Increase Awareness about Adult Radiation Protection. Radiology 257(3): 601-602 (2010) 4 Hernandez RJ, Goodsitt MM.Reduction of radiation dose in pediatric patients using pulsed fluoroscopy. American Journal of Roentgenology. 167(5):1247-53 (1996). 5 Brown PH, Silberberg PJ, Thomas RD, Strife JL, Towbin RB. A multihospital survey of radiation exposure and image quality in pediatric fluoroscopy. Pediatric Radiology 30:236-242 (2000). 6 Ward VL et al. Radiation exposure reduction during voiding cystourethrography in a pediatric porcine model of vesicoureteral reflux. Radiology 238:96-106 (2006). 7 Stueve D. Management of pediatric radiation dose using Philips fluoroscopy systems - DoseWise: perfect image, perfect sense. Pediatric Radiology 36:216-220 (2006). * Patent pending. 2017 Koninklijke Philips N.V. All rights reserved. Specifications are subject to change without notice. Trademarks are the property of Koninklijke Philips N.V. or their respective owners. How to reach us Please visit www.philips.com healthcare@philips.com 4522 991 30771 * Nov 2017