OPERATING MANUAL. 100 MHz CENTER FREQUENCY OFF AXIS ACOUSTO-OPTIC BEAM DEFLECTOR MODEL NUMBER: DEG-.51 DOCUMENT NUMBER: 51A12229A

Similar documents
OPERATING MANUAL. ACOUSTO OPTIC MODULATOR MODEL NUMBER: X-1.06-LTD X= 1, 2, or 3 mm DOCUMENT NUMBER: 51A14950A

OPERATING MANUAL. ACOUSTO OPTIC MODULATOR MODEL NUMBER: X-LTD X= 1, 2, or 3 mm DOCUMENT NUMBER: 51A00620D

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

OPERATING MANUAL VOLTAGE CONTROLLED OSCILLATOR MODEL NUMBER: 21XXX-YYY-ZASVCO DOCUMENT NUMBER: 51A18410

OPERATING MANUAL VOLTAGE CONTROLLED OSCILLATOR MODEL NUMBER: 21XXX-YYY-ZAMVCO DOCUMENT NUMBER: 51A18412A

Sintec Optronics Technology Pte Ltd 10 Bukit Batok Crescent #07-02 The Spire Singapore Tel: Fax:

OPERATING MANUAL DIGITALLY CONTROLLED FREQUENCY SYNTHESIZED OSCILLATOR MODEL NUMBER: ADSDFS-A DOCUMENT NUMBER: 51A19937C

OPERATING MANUAL CAVITY DUMPER / PULSE PICKER DRIVER MODEL NUMBER: 643ZZ.ZZZ-SYN-Y-X

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

Where λ is the optical wavelength in air, V a is the acoustic velocity, and f is the frequency bandwidth. Incident Beam

Precautions NEVER OPERATE THE DRIVER WITHOUT PROPER COOLING. THE MOUNTING FACE TEMPERATURE MUST NOT EXCEED 60*C.

MT80-A1-VIS/IR/1064, MT80-A1.5-VIS/IR/1064 MT110-A1-VIS/IR/1064, MT110-A1.5-VIS/IR/1064

ISOMET. Acousto-Optic Deflector Driver. Instruction Manual. D3x5-BS Series. Including: Basic Deflector Alignment. Models -

MT80-A1-VIS/IR/1064, MT80-A1.5-VIS/IR/1064 MT110-A1-VIS/IR/1064, MT110-A1.5-VIS/IR

MT80-A TECHNICAL DATA SHEET 2014

Base model features 1.0Vpp, 50ohm modulation input level and 24/28Vdc supply. L : +15V supply operation

Base model features 1.0Vpp, 50ohm modulation input level and 24/28Vdc supply.

ISOMET. Acousto-Optic Modulator Driver. Instruction Manual. 512c-m Series. Including: Basic Modulator Alignment

ISOMET. Acousto-Optic Modulator Driver. Instruction Manual. RFA9x0-41 Series. Including: Basic Modulator/Q-switch Alignment. May 07.

GOOCH & HOUSEGO NOVEL OPTICAL COMPONENTS FOR THE IR

ISOMET. Acousto-Optic Deflector Amplifier. Instruction Manual. RFA100 Series. Including: Basic Modulator Alignment. Apr 13.

ISOMET. Acousto-Optic Modulator Driver. Instruction Manual. RFA141 Series & RFA151 Series. Including: Basic Modulator Alignment for 1202 / 1208 series

Base model features TTL buffer compatible modulation input level and 24/28Vdc supply.

Including: Amplifier Connection & AO Deflector Alignment. 630C-fff-LP-m. Base model features 1.0Vpp, 50ohm modulation input level and 24/28Vdc supply.

MT110-B50A1.5-IR-Hk + MDPS1C-xx

High-Frequency Electro-Optic Phase Modulators

COLLIMATORS AND FOCUSERS RECEPTACLE STYLE

Supplementary Materials

ISOMET. Acousto-Optic Modulator Driver. Instruction Manual. RFA2x1 Series. Including: Basic Modulator Alignment. Feb 17. Models - RFA2x1-z :

Broadband Power Amplifier

- BR : Brass heatsink, rear mounted water fittings only. (to suit brass or copper cased AO devices)

Kit for building your own THz Time-Domain Spectrometer

High-Frequency Electro-Optic Phase Modulators

Department of Electrical Engineering and Computer Science

INSTRUCTION MANUAL. Acousto-OptIc Modulator/Drfver Series 230. (Analog Modulation) ***PRECAUTIONS***

Design Description Document

DC-250 MHz Electro-Optic Phase Modulators Models 4001, 4002, 4003, 4004, 4061, 4062, 4063, 4064

How-to guide. Working with a pre-assembled THz system

Model Number Guide. M= Material. S= Apperture Size. P= Options

Model Series 400X User s Manual. DC-100 MHz Electro-Optic Phase Modulators

Coherent Laser Measurement and Control Beam Diagnostics

Fiber Pigtailed Variable Frequency Shifters Acousto-optic products

MTF characteristics of a Scophony scene projector. Eric Schildwachter

Polarization Experiments Using Jones Calculus

Laser Speckle Reducer LSR-3000 Series

User s Guide Modulator Alignment Procedure

Assembly and Experimental Characterization of Fiber Collimators for Low Loss Coupling

411LA Broadband Power Amplifier

Non-amplified Photodetectors

Fiber Optic Communications

The Lightwave Model 142 CW Visible Ring Laser, Beam Splitter, Model ATM- 80A1 Acousto-Optic Modulator, and Fiber Optic Cable Coupler Optics Project

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the

CLEANING FOR BETTER OBSERVATION AND PHOTOMICROGRAPHY

1.5µm PbSe Power Detector

Amplified Photodetectors

OBIS CellX. The Universal Light Engine FEATURES

Be aware that there is no universal notation for the various quantities.

Devices & Services Company

User s Guide Modulator Alignment Procedure

TIA-525 Optical/Electrical Converter Operating Instructions

EXPRIMENT 3 COUPLING FIBERS TO SEMICONDUCTOR SOURCES

Laser Diode Mounting Kits

R. J. Jones College of Optical Sciences OPTI 511L Fall 2017

Will contain image distance after raytrace Will contain image height after raytrace

Laser Telemetric System (Metrology)

CLEANING THE GALVO MIRRORS

Premier & Acculase Modulatable Laser Diode Modules

FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION

Vixar High Power Array Technology

Properties of Structured Light

Fast, Two-Dimensional Optical Beamscanning by Wavelength Switching T. K. Chan, E. Myslivets, J. E. Ford

Optical Components for Laser Applications. Günter Toesko - Laserseminar BLZ im Dezember

Optical Isolator Tutorial (Page 1 of 2) νlh, where ν, L, and H are as defined below. ν: the Verdet Constant, a property of the

Model BiConiLog Antenna. User Manual

Non-amplified High Speed Photodetectors

Exercise 1-3. Radar Antennas EXERCISE OBJECTIVE DISCUSSION OUTLINE DISCUSSION OF FUNDAMENTALS. Antenna types

ADVANCED OPTICS LAB -ECEN 5606

Optical Components - Scanning Lenses

End Capped High Power Assemblies

Displacement Follower

LASER ENHANCED REVOLVER GRIP OWNER S MANUAL RED LASER GREEN LASER

NON-AMPLIFIED PHOTODETECTOR USER S GUIDE

OxFAC-08 Angled Fiber Cleaver. User Manual. Issue 1.5

Machine Vision Lyte-MV 2

ADVANCED OPTICS LAB -ECEN Basic Skills Lab

MQH0XX-YYDS3-ZZZ-2S. High Power 2 Channel RF Driver: For 2 Acousto-optic Q-Switches. Description: Key Features: Applications:

TURNKEY, ULTRA STABLE, OEM LASER DIODE SOURCE OZ-1000 & OZ-2000 SERIES

25-Gbit/s, 850-nm VCSEL

Ox-RAC-08 Ribbon Angled Fiber Cleaver User Manual

Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy

The Reverse Polarity TNC(m) RF connector can be easily secured or removed from equipment in the field by a single gloved hand, no tools required.

User s Guide Modulator Alignment Procedure

PSW-002. Fiber Optic Polarization Switch. User Guide

POLARIZATION EXTINCTION RATIO METER

HIGH POWER COLLIMATORS AND FOCUSERS PIGTAIL STYLE

SUPERSEDED REVISION. Reasons for reissue of this instruction sheet are provided in Section 7, REVISION SUMMARY.

Amplified High Speed Photodetectors

Using Stock Optics. ECE 5616 Curtis

INSTRUCTION MANUAL OTOT-1000C-FQ WIDE-BAND 1550NM DIRECT-MOD QAM TRANSMITTER. 1,000 MHz Bandwidth / Optical Output Power is +9dBm (OTOT-1000C-09-FQ)

Application Note (A11)

Transcription:

OPERATING MANUAL 100 MHz CENTER FREQUENCY OFF AXIS ACOUSTO-OPTIC BEAM DEFLECTOR MODEL NUMBER: DOCUMENT NUMBER: 51A12229A Document approved for release: W Seale Date: 8/18/06 US OFFICE: NEOS Technologies, Inc. 4005 Opportunity Drive Melbourne, FL 32934 USA Tel: (321) 242-7818 Fax: (321) 242-1019 Email: neos@neostech.com UK OFFICE: Gooch & Housego The Old Magistrates Court Ilminster, Somerset TA19 0AB UK Tel: +44 1460 52271 Fax: +44 1460 54972 Email: sales@goochandhousego.com

TABLE OF CONTENTS SECTION DESCRIPTION PAGE I. INSPECTION PROCEDURE 3 II. DESCRIPTION 4 III. SPECIFICATIONS 5 IV. OUTLINE DRAWING 6 V. CALCULATIONS 7 VI. OPERATING PROCEDURE 8 VII. CLEANING PROCEDURE 10 2 of 10

SECTION I INSPECTION PROCEDURE Examine the shipping carton for damage. If the shipping carton or packing material is damaged it should be kept for the carrier s inspection. Notify the carrier and NEOS Technologies. Check the contents of the shipment for completeness, mechanical damage, and then test the equipment electronically. Operating procedures are contained in Section VI. If the contents are incomplete, or the equipment does not pass the electrical testing please notify NEOS Technologies. If there is any problem with the use of this equipment, or if the equipment fails to function as expected contact NEOS Technologies, do not try to trouble shoot or repair this equipment. Consult with a NEOS service engineer. If the equipment needs repair or replacement, contact NEOS Technologies, Inc for a Return Authorization Number. 3 of 10

SECTION II DESCRIPTION The off axis acousto-optic beam deflector (AOBD) is fabricated from Te0 2 crystals with LiNb0 3 slow shear wave transducers. The off axis deflector differs from the normal on axis deflector in that the acoustic beam is launched into the Te0 2 6.5 degrees off of the 110 propagating axis. This design eliminates a mid-band degeneracy that results in a loss of diffraction efficiency over a narrow frequency range in the operating band of 75 MHz and 125 MHz.. The AOBD can be driven by any good driver with a 50 ohm nominal output; however, it is recommended that a NEOS Technologies driver be used to drive this deflector to achieve optimum performance. The RF input power should not exceed 2.5 watts CW as NEOS Technologies will not warranty damage from RF power applied exceeding 2.5 watts 53B2024 4 of 10

SECTION III DEVICE SPECIFICATIONS PARAMETER SPECIFICATION Interactive Material TeO 2 Acoustic Mode Operating Wavelength Window Configuration Shear Wave 440 to 530nm AR Coated Static Transmission >95 % Operating Frequency Intensity Variation Diffraction Efficiency Light Polarization Acoustic Aperture Size 75 to 125 MHz. 2dB >70% Midband Linear, parallel to acoustic propagation 5 mm Process Time 7.5 µs Resolution (T.BW product) Deflection Angle Deflection Angle RF Power Level Impedance VSWR Package: Acceptance Test Procedure: Acceptance Test Results form: Recommended Drivers: 375 spots with no less than 60 µs scan time and full illumination of the aperture 39 mrad @ 514 nm 78 mrad @ 100 MHz, 514 nm 2 watt maximum 50 ohms <2:1 across band 53B2024 42A12652 52A11841 Analog System Digital Frequency Synthesizer: 64010-200-2ADSDFS-A Analog Module Digital Frequency Synthesizer: 64010-200-2ADMDFS-A Analog System Voltage Controlled Oscillator: 21075-125-2ASVCO Analog Module Voltage Controlled Oscillator: 21075-125-2AMVCO 5 of 10

SECTION IV OUTLINE DRAWING 53B2024 Dimensions are in inches Tolerances: Decimal:.xx =.01.xxx =.005 Dimensions in [ ] are in mm. Milimeter:.xx =.25mm.xxx =.127mm Angle: = + 30' 6 of 10

SECTION V CALCULATIONS The maximum aperture of each device is 5 millimeters by 5 millimeters. The input laser beam diameter should fully illuminate the aperture in order to achieve the proper number of resolvable spots. The access time ( T) can be determined from the following: The process time ( T) is equal to the Beam diameter (d 0 ) in the acoustic direction divide by the Velocity of sound (V) in the material. Where: d 0 = 5 mm V = 0.66 mm/µs d T = 0 5 = = 7.5 µs V 0. 66 The number of resolvable spots (TBW) for Acousto-Optic Beam Deflector is the product of f and T Where: the RF bandwidth ( f) of the device is 50 MHz. TBW = f T = 50 MHz X 7.5 µs = 375 375 resolvable spots with uniform illumination of the aperture and no less than 60 µs chirp time to avoid lensing effects. The actual number of resolvable spots (N) are dependent on the uniformity of the illumination of the aperture (truncation of the laser beam) and scan chirp time. T T f N = (1- ) ( ) T + T a Where: T = Chirp time T = process time a = A parameter for uniformity of illumination. Where: a = 1 for uniform illumination. 1 a = 1.34 for gaussian illumination clipped at the 2 intensity points. e The RF chirp applied to the AOBD causes a lensing effect as well as a deflection of the laser beam. The focal length (FL a ) of the acoustic cylinder lens, lensing effect is: dfa FL a = V 2 a / ( λ) dt dfa Where: is the slope of the frequency change vs time. dt The angle between the diffracted and the zero order beam, the deflection angle, d, is defined by: λf d = 2θ Bragg = =. V λf 0.66mm/ µ sec Where: f = RF frequency in MHz λ = optical wavelength θ = Bragg angle of the Acousto-Optic Beam Deflector The frequencies "f "are from 75 to 125 MHz with the center frequency = 100 MHz. 7 of 10

. SECTION VI. OPERATING PROCEDURE Connect to the AOBD, a RF source that will provide nominally 2 Watt of CW RF power between the frequencies between 75 and 125 MHz with a 50 ohm cable. Project a collimated light beam, linear polarized, parallel to acoustic propagation into the aperture of the AOBD on the input side (closest to the BNC connector ), as this device is unidirectional. The light beam should fully fill the aperture in order achieve the full output resolution. Rotate the AOBD Bragg angle so that the beam is normal to the input window of the Te02 crystal. Apply the RF power at 100 MHz to the AOBD. Adjust the Bragg angle as shown in figure 1 and view the output light at a distance of about 1 meter from the output side of the AOBD, as an array of spots will result when approaching the Bragg angle. Use a IR viewing card to see the output. When this array of spots becomes evident, maximize the intensity of the diffracted (-) first order beam (toward the BNC connector), by varying Bragg angle, the vertical and horizontal position of the AOBD. Refer to figure 1. The vertical positioning of the deflector is important to ensure that the light beam is centered in the acoustic column. In most AOBD's you may also see some diffracted light in the positive first and negative second orders, however the intensity of these orders are generally very low when the AOBD is set at the Bragg angle. Next, check the diffraction efficiency at 75 MHz and 125 MHz RF drive frequencies or sweep the RF over this 50 MHz band. Slightly adjust the three adjustments to obtain best flatness diffracted light intensity across the band of 75 to 125 MHz. Next, the RF power should be adjusted to achieve the maximum required diffraction efficiency. However, in no case should the RF power exceed 2.5 watts, NEOS will not warranty any failure resulting from the application of too much RF power. This device has been produced and tested to meet or exceed the specifications. Note: The diffracted output beam will be rotated in polarized to be perpendicular to the acoustic propagation. This rotation in polarization is an advantage when using the device in the X-Y, 2D, AOBD configuration. 8 of 10

Figure 1 AOBD Bragg Angle Adjustment BRAGG ADJUSTMENT BRAGG ANGLE EXAGGERATED INPUT BEAM HORIZONTAL POLARIZATION VERTICAL POLARIZATION + 1ST ORDER 0 ORDER - 1ST ORDER BRIGHTEST BEAM - 2ND ORDER INPUT OUTPUT 45A12233 Figure 2 BRAGG ADJUSTMENT BRAGG ANGLE EXAGGERATED INPUT BEAM HORIZONTAL POLARIZATION HORIZONTAL POLARIZATION LENS DC BEAM BLOCK OPTICAL DETECTOR INPUT AOBD -1ST ORDER BEAM OUTPUT VERTICAL POLARIZATION SCAN PLANE F 1 F 2 RF FREQUENCY SWEEP H V FREQUENCY SOURCE OSCILLOSCOPE OPTICAL METER 45A18969 9 of 10

SECTION VII. CLEANING PROCEDURE Periodic cleaning of the AOBD's optical windows is a normal part of maintaining an optical system. When the AOBDs are installed in an optical system, make sure that there is access to allow cleaning or removal for cleaning of the AOBDs and it's mount from the system. When removed from the system, follow the alignment procedure in this manual to reinstall, realign and, adjust the AOBDs. Make sure that the AOBDs are reinstalled in the correct position as the devices are unidirectional. To clean the AOBD device, remove the screws that hold the cover to the device. Caution must be used when placing a screw driver near the window opening in the cover, as it is best to protect the opening with tape or cover the opening with your finger (without touching the crystal) to protect it. NEOS will not warrant any damage or scratches caused by inserting the screwdriver into the window opening. Remove the protective cover. Blow off any visible dust with canned air. Do not use an air gun unless it is filtered and water and oil free! Fold (4 times) a new lens tissue into a triangle to make a cleaning tool. Dip the tip of the lens tissue into fresh acetone or spray fresh acetone from a squeeze bottle onto it. Then shake excess fluid out of the lens tissue. Do not handle the wet area of the tissue, as your finger oil will be absorbed and contaminate the optical surface of the crystal. Wipe (only once) across the crystal window in an even motion, starting near the transducer and drawing the tissue across the optical surface toward the other end. Do not damage the bond wires! Do not reuse the tissue as the mounting silver epoxy may be spread onto the window of the crystal. Repeat with a new tissue each time and for each surface that needs cleaning. Replace the AOBD cover and screws. Realign the AOBDs in your system and adjust the Bragg angle for maximum diffraction efficiency. Notes: The lens tissue must be lint free and the best grade available. Only use each tissue once, for only one surface. Do not reuse the tissue, as it will redistribute the removed dust or mounting silver epoxy. The acetone must be electronic grade. The acetone must be fresh from a new bottle, as the acetone will absorb water from the air and cause streaks. Discard any acetone, which has been exposed to the air for more than 4 hours. If the bottle is half- empty, do not use. 10 of 10