OPERATION MANUAL OF MODEL TH2681/A HIGH RESISTANCE TESTER

Similar documents
CD770 DIGITAL MULTIMETER INSTRUCTION MANUAL

OPERATION & SERVICE MANUAL FOR FC 110 AC POWER SOURCE

PRECISION INSTRUMENTS FOR TEST AND MEASUREMENT. HRRS 5kV AND 10kV SERIES. High Resistance, 5kV and 10kV Decade Substituter User and Service Manual

GAMMA 12 DIGITAL MULTIMETER

Dawson DDM181. Pocket-Size Autorange Digital Meter User s Manual

Pen Multimeter. Model

GAMMA 10 DIGITAL MULTIMETER

AMP-13 OPERATOR S MANUAL

AMP-12 OPERATOR S MANUAL

User s Guide. MultiView Series Digital MultiMeters Models: MV110 MV120 MV130

Wilcom MODEL T136BGMZW CIRCUIT TEST SET. Operating Instructions

Pen Multimeter. Model

T-IVA001S. Highly Sensitive Galvanometer Amplifier that can Measure Picoamperes. Operating Instruction Manual I V

User's Guide Mini Multimeter with Non-Contact Voltage Detector (NCV) Model EX330

SPM-50 RF Spectrum Power Meter PC Software User Manual

34134A AC/DC DMM Current Probe. User s Guide. Publication number April 2009

CONTENTS. SAFETY PRECAUTIONS: Before use, read the following safety precautions

Dawson DDM190. Digital Multimeter User s Manual

ELECTRONIC SHEAR PIN MODEL ESP 1 & 2 INSTRUCTION MANUAL

Wilcom MODEL T136BSBZW CIRCUIT TEST SET. Operating Instructions

Installation & Operation Manual SAGA1-K Series Industrial Radio Remote Control

User s Guide. Agilent 16441A R-Box. Agilent Part No Printed in Japan January Edition 4

R-X SERIES. Decade Resistor

2001A. 200KHz Function Generator Instruction Manual. 99 Washington Street Melrose, MA Phone Toll Free

User s Manual. MiniTec TM Series. Model MN26 (Model MN26T includes temperature probe) Mini Autoranging MultiMeter

Thank you again for choosing AstroAI, if you have any questions or concerns regarding your product, please contact us at

DIGIT & POINTER MULTIMETER

22 Series Relays. Technical Data...1. Specifications...2. Model Number Structure - Relays...3. Model Number Selection

Electrical safety testers

Passive Component LCR Meter

SERIES. Owners Manual & Installation Guide. Limited Warranty: 2-Channel High Speed Digital Stereo Amplifier

USER S MANUAL POWER SUPPLY - ISOLATOR - SIGNAL CONVERTER ZSP 41 POWER SUPPLY - ISOLATOR - SIGNAL CONVERTER SIGNAL MULTIPLIER ZSP 41/2

BROADBAND LINEAR AMPLIFIER Model P150

Dawson DDM230C. True RMS Multimeter with Bar Graph Display User s Manual

8730 Electric Parameter Meter User manual

OVA-50 Optical Variable Attenuator User s Manual

Digital Panel Meter K3TE. Model Number Structure. Model Number Legend. Easy-to-use, Low-cost Digital Panel Meter that Accepts DC Input K3TE

RHS SERIES RH Set Point Controller/Probe

Passive Component LCR Meter

SAFETY PRECAUTIONS: Before use, read the following safety precautions

DM-46 Instruction Manual

RIGOL Data Sheet DM3058/DM3058E Digital Multimeter Product Overview Applications Easy to Use Design Main Features owerful Measurement Functions

INSTALLATION AND OPERATING INSTRUCTIONS MODEL SIVF. KB Part No Signal Isolator for KBVF Controls. See Page 1 Pending

DIGITAL MULTIMETER AX-588B

JTY2 PULSE DURATION RECEIVER BEFORE USE... COMPONENT IDENTIFICATION

DSB810A Digital DC Servo Driver Manual V1.0

DDM350 Pen-Type Digital Multimeter User s Manual

Four-quadrant fast response bi-polar power supply

SUBJECT TO CHANGE WITHOUT NOTICE

LCR METER Model:

SI-125 Power Amplifier Manual 6205 Kestrel Road; Mississauga, Ontario; Canada; L5T 2A1 November 2016, Rev 0.5

Battery Analyzers BA6010 Series

ProfiScale MULTI Multimeter

LCT100. Load Cell Tester User s Guide (v1711) Anyload Transducer Co. Ltd Website:

RAIL MOUNTED NETWORK METER TYPE N27D USER S MANUAL

Autoranging Multimeter Extech EX503

DUAL CHANNEL BROADBAND LINEAR AMPLIFIER Model A800D

Instruction manual SUN-PM100L. SUN-PM100L Polishing Machine English

AM-510 Commercial / Residential Multimeter. AM-510-EUR Digital Multimeter. Users Manual

MS2109A AC/DC Clamp Meter. User Manual. Contents

SHI Series Pure Sine Wave Inverter (SHI2000 AND SHI3000) User s Manual

S-11 S-11. Digital Multimeter Operating Instruction. Digital Multimeter Operating Instruction

MODEL: W2DYS SEN TRONIC AG. Mini-MW Series

MS8250A/B OPERATION MANUAL MS8250A. Hz% FUNC REL RANGE REL HOLD OFF 10A. Hz% A NCV. Hz% COM. A ma 10A FUSED 600V CAT IV.

MODEL W Power Amplifier

NOTE: Fully read and understand this manual before using this Digital Multimeter.

EM420A/420B DIGITAL MULTIMETER OWNERS MANUAL Read this owners manual thoroughly before use

User s Manual. Model MS5 Potentiometer Converter (Free Range Type) MODEL AND SUFFIX CODES CAUTIONARY NOTES FOR SAFE USE OF THE PRODUCT

OPERATOR S MANUAL Model 77E Pneumatic Cable Stripper

TOS8830 DANGER OPERATION MANUAL

Model: &

Model UT20B: OPERATING MANUAL Table of Contents (1)

RIGOL. User s Guide. RP1000D Series High Voltage Differential Probe. Feb RIGOL Technologies, Inc

Turbine Temperature Test Set

This section is specifically about safety matters

Worldwide Manufacturer of Gas Detection Solutions. Transmitter EC 23. Operation Manual

INSTRUCTION MANUAL LKG 601 Electrical Safety Analyzer

AUTO RANGING DIGITAL MULTIMETER

Starbox. Installation guide

OPERATOR S INSTRUCTION MANUAL M-2625 AUTO RANGING DIGITAL MULTIMETER

VC835 DIGTAL MULTIMETER Operation Manual

HP 86290B RF PLUG-IN GHz HEWLETT PACKARD

SDM3045X Digital Multimeter. DataSheet

DM-45 Digital Multimeter

DVM98. True RMS Digital Multimeter. 1 Safety information. 1.1 Preliminary. 1.2 During use

Isolated Module TRPxxxxxP 系列产品

SDM3055 Digital Multimeter. DataSheet

731A seismic accelerometer and P31 power unit/amplifier Operating guide

User Guide. AC/DC Mini Clamp Meter. Model Introduction. Large 3 3/4 digit LCD display with 40 segment bargraph

Testers: / Ohmmeters Insulation Measuring Instruments. n Compact and rugged for service calls and laboratory use. Scope of delivery:

User's Guide. Extech 800 Series 1000 Amp Clamp Meters with IR Thermometer

series Oweners Manual Limited Warranty:

Model MV106J/MV116J. ±10nVdc to ±11Vdc Precision DC Voltage Standard Source. Operating Manual

DA560D COMPACT SERIES. INSTALLATION / OWNER'S MANUAL Mobile Power Amplifiers

Four-quadrant fast response bi-polar power supply

User s Guide. 400A AC/DC Clamp Meter. Model MA220

Model 5100F. Advanced Test Equipment Rentals ATEC (2832) OWNER S MANUAL RF POWER AMPLIFIER

1. General Instructions 2 2. Safety 2 3. Lamp Starting Test Instrument LSTI 5 3

Model 9305 Fast Preamplifier Operating and Service Manual

INSULATION TESTER SERIES Selection guide

Transcription:

OPERATION MANUAL OF MODEL TH2681/A HIGH RESISTANCE TESTER Changzhou Tonghui Electronic Co., Ltd Addr: No.3 Tianshan Rd., New District, Changzhou Zip: 213022 Tel: 86-519-5132222 Fax: 86-519-59972 Http:// www.tonghuico.com E-mail: tonghui@pub.cz.jsinfo.net

CONTENTS I. Generals 3 II. Main Technical Data 4 III. Description of Panel Function 8 IV. Instruction of Operation 9 V. Calibration, Inspection and Maintenance of the Tester 11 VI. Warranty of the Tester 13 2

I. Generals 1.1 Brief Introduction of the Product Model TH2681/2681A High Resistance Tester is a measuring instrument used to measure insulation features of various kinds of electronic component, media materials and the complete set etc. It has features of fast measurement, good stability and easy operation, as well as the sorting function. This tester is in accord with the requirement in Group II of the Standard GB6587.1 General Outline for Electronic Measurement Instrument in Environmental Experiment of the Ministry of Electronic Industry. The rated conditions for application are as follows: A: Environment Temperature: (0~40) B: Relative Humidity: <70% C: Atmosphere Pressure: 86-6 kpa 1.2 Confirmation of Attachments before Application When receiving the tester, you are pleased to confirm following items so as to protect your right. A: If the appearance of the product is imperfection such as broken, scraped etc. B: If standard attachments listed in the table below are missing or not. Should above damage or insufficient attachments be found, please let our company or affiliates, office or agent know so as to provide you service immediately. Name of Item Materials Quantity Remarks Power Supply Wire 1 Measuring Wire 1 Normal measuring clip wire Operation Manual 1 Measuring Seat Option item 1.3 Ambient Environment of Application 1.3.1 Please do not use the tester in place where subjects to dust and vibration and those place under direct daylight or corrosive gas. You are pleased to use the tester in the place where ambient temperature is between 0-40 and relative humidity is less than 70%. 1.3.2 In order to prevent temperature rising inside the tester, take care of ventilation around it and keep it more than cm away from the wall or object. The left and right ventilation holes of Tester cannot be covered so as to maintain high accuracy of the instrument. 1.3.3 As the tester has some design against disturbance of AC power supply, it is still suggested to use the tester under the environment with less disturbance. Should disturbance not be avoided, please add the filter device of power supply. 1.3.4 The storage range of temperature for the instrument is between -25 ~70. Should the instrument not be used for a long period, it should be packed with the original packing materials or something similar and stored in the place where is dry and free of direct sunlight so as to maintain its high accuracy when it is reused. 3

1.4 Voltage of Power Supply Before switching on the power supply, be sure that the switch of power is in the OFF position and the applied voltage is within the range of specified voltage. The voltage of power supply is 220(1±%)V and the frequency of power supply is 50 Hz or 60 Hz. 1.5 Fuse There is a fuse holder on the rear panel of the instrument. Be attention when changing the fuse: 1.5.1 Make sure to switch OFF the power supply first and draw out the wire of power supply. 1.5.2 Visual inspection of the fuse cannot be confirmed. It is normal when measured resistance value is under 15Ω. 1.5.3 When changing the fuse, slightly press the outer cap of the fuse holder, which is projected from the rear panel, and turn it counterclockwise until the fuse pops out. 1.5.4 Take out the outer cap and replace the fuse with the new one (same specifications with that of the origin). 1.5.5 Slightly press the outer cap into the fuse holder and turn it clockwise. 1.6 Preheating Time All functions of the tester are available as long as the tester is switched on, however in order to obtain the specified accuracy, you are pleased to preheating it for about 30 minutes. II. Main Technical Data 2.1 Measuring Voltage: special demand for detailed voltage range is available upon request. TH2681: Voltage Voltage multiple V 0.1 25V 0.25 50V 0.5 0V 1 250V 2.5 500V 5 TH2681A: Voltage Voltage multiple V 0.1 50V 0.5 0V 1 250V 2.5 500V 5 00V The accuracy of measuring voltage: ±2% (open circuit state of measuring terminals); 4

2.2 Range of Measurement TH2681: Test voltage Range of measurement( 5 Ω ~ 5 12 Ω) V 0.1MΩ ~ 0GΩ 25V 0.25MΩ ~ 250GΩ 50V 0.5MΩ ~ 500GΩ 0V 1MΩ ~ 1TΩ 250V 2.5MΩ ~ 2.5TΩ 500V 5MΩ ~ 5TΩ TH2681A: Test voltage Range of measurement( 5 Ω~ 13 Ω) V 0.1MΩ ~ 0GΩ 50V 0.5MΩ ~ 500GΩ 0V 1MΩ ~ 1TΩ 250V 2.5MΩ ~ 2.5TΩ 500V 5MΩ ~ 5TΩ 00V MΩ ~ TΩ 2.3 Accuracy of Measurement Resistance Value < GΩ ±3 %±0.5 division Resistance Value >= GΩ ±6 %±0.5 division Resistance Value > 1TΩ ± %±0.5 division 2.4 Charging Time For electronic complete set, materials and resistance component: charging time < 0.1 second Capacitor: charging time from 0.5 second to seconds 2.5 Limit range: Limit = (1.0 ~ 9.9) selected multiples of voltage multiple switch (MΩ) Accuracy: ±3 %±0.5 division 2.6 Sorting Function This instrument has sorting function. The required value can be pre-selected by the dial switch. When measuring value is less than the preset value, No-Good lamp will light plus sound prompt (can be turned off) which means fail. When measuring value is larger than the preset value, No-Good lamp will go out and that means pass. 2.7 Select which to be indicated by Head Indication When Set/Measure button is released, the value indicated by the head is that preset by the dial. When Set/Measure button is pressed, the value indicated by the head is that to be measured. 5

2.8 Selection between Measuring and Discharging State When the band switch of voltage selection is in the discharge position, the measured component shall be discharged by a resistor of 5W/1kΩ inside the tester, and no voltage is applied to the terminals. 2.9 Conditions of Application: 2.9.1 Temperature: 0 ~40 ; Relative Humidity: less than 70% 2.9.2 Storage Temperature: -25 ~70 2.9.3 Voltage of Power Supply: 220(1±%)V Frequency: 50Hz or 60Hz Power: 50VA. 2. Dimensions and Weight 2..1 Dimensions: 300mm (length) 270mm (width) 180mm (height) 2..2 Weight: about 3 kg. 6

00V1VOPERATION MANUAL OF MODEL TH2681/A R TONGHUI ELECTRONICS 1 13 2 3 TH2681A Insulation Resistance Meter MULTIPLE VOLTAGE 0 1 2 3 4 5 DISCH 250V 1 2.5 DISCH DISCH 50V 0.5 500V 5 FAIL 0.1 DISCH DISCH 00V ADJ ON + - SET MEAS POWER RX! 11 12 8 9 7 6 5 4 7

III. Description of Front Panel 3.1 The Head of the Tester: Model TH2681/TH2681A Insulation Resistance Head (also 0μA Current Head); 3.2 Multiple Selection Switch: there are 6 grades for multiple selection: 0, 1, 2, 3, 4, 5 ; 3.3 Voltage Selection Switch: there are 11 grades in which 6 for voltage selection and 5 for discharging: TH2681: V/25V/50V/0V/250V/300V TH2681A: V/50V/0V/250V/500V/00V 3.4 Set/Measurement Switch: selection for indication state of the head; Set/Measurement Switch in Set position: the value indicated by the head is that preset by the dial; Set/Measurement Switch in Measurement position: the value indicated by the head is that to be measured. 3.5 Setting of the Dial It consists of two bits, and is used to preset the limit value of component for sorting. 3.6 Output Terminals of Voltage It outputs negative high voltage and is connected to one end of measuring wire. Take care of safety due to output of high voltage. 3.7 Measuring terminals: to be connected to the other end of measuring wire. 3.8 Adjusting Potentiometer It can adjust range of each multiples to make the pointer of the insulation tester indicate. 3.9 Grounding Terminal 3. Failure Indicating Lamp When it lights, it means the insulating resistance of measured component is less than the preset value, and the component is not good. When it doesn t light, it means the insulating resistance of measured component is larger than the preset value, and the component is good one. 3.11 Indicating Lamp of Power Supply 3.12 Switch of Power Supply 3.13 Model, name and Trade Mark Model of the Tester: TH2681 or TH2681A Name: High Resistance Tester Trade Mark: trade mark of product of Tonghui Electronic Co., Ltd 8

IV. Instruction of Operation It is suggested that customer should read this operation manual carefully before application so as to be familiar with its operating method and working principle. 4.1 Measuring Method of the Object to be Measured!!! NOT ICE 1. When measuring, take care of safety due to high voltage on both ends of the object to be measured. 2. Take car of correct connection if the object to be measured has polarity. 3. In order to obtain correctness and stability of measurement, isolation of environment and control of temperature and humidity should be done well. 4. The shielded wire on the measuring end (+) should not be connected with the casing of the instrument. 4.2 Notice before Measuring 4.2.1 Multiple Input Resistance 0 1kΩ 1 kω 2 0kΩ 3 1MΩ 4 MΩ 5 0MΩ 4.2.2 Above drawing is equivalent circuit in which the input resistance has relation with the multiple and is shown as the table. The measurement range of the object to be measured has relation with voltage and multiple and shown as follows: TH2681: Unit: Ω multiple range voltage 0 1 2 3 4 5 V 0K-1M 1M-M M-0M 0M-1G 1G-G G-0G 25V 250K-2.5M 2.5M-25M 25M-250M 250M-2.5G 2.5G-25G 25G-250G 50V 500K-5M 5M-50M 50M-500M 500M-5G 5G-50G 50G-500G 0V 1M-M M-0M 0M-1G 1G-G G-0G 0G-1T 250V 2.5M-25M 25M-250M 250M-2.5G 2.5G-15G 25G-250G 250G-2.5T 500V 5M-50M 50M-500M 500M-5G 5G-50G 50G-500G 50G-5T 9

TH2681A: OPERATION MANUAL OF MODEL TH2681/A Unit: Ω multiple range voltage 0 1 2 3 4 5 V 0K-1M 1M-M M-0M 0M-1G 1G-G G-0G 50V 500K-5M 5M-50M 50M-500M 500M-5G 5G-50G 50G-500G 0V 1M-M M-0M 0M-1G 1G-G G-0G 0G-1T 250V 2.5M-25M 25M-250M 250M-2.5G 2.5G-15G 25G-250G 250G-2.5T 500V 5M-50M 50M-500M 500M-5G 5G-50G 50G-500G 50G-5T 00V M-0M 0M-1G 1G-G G-0G 0G-1T 1T-T M: 6 G: 9 T: 12 Data Table of Voltage, Multiple and Measuring Range 4.2.3 There is an adjuster in the front panel, which is used to make sure that head pointer should point to for each multiple range when test terminals are open circuit. For the multiple range of 5 : a. When measuring terminals have no measuring fixture or no measuring wires connected, the head pointer should point to. b. When measuring terminals have measuring fixture or measuring wires connected but no object Rx is connected, the head pointer may have some drift, because the feedback resistance of this range is as larger as 0MΩ, which is caused from disturbance signal received by the antenna, i.e. measuring wires. When Rx is connected, external stray signals will disappear, and the correctness of measurement in this range will not be affected. 4.3 Set up Measurement Data 4.3.1 Voltage select According to technical requirement of product, set the voltage band switch to corresponding voltage grade. Take care that the measuring fixture is live now and is dangerous. 4.3.2 Multiple select Set relative multiple according to the relation drawing in Section 4.2.2. 4.3.3 Upper Limit set up According to technical requirement of product, set the value of the dial switch, which should correspond to the value of the head pointer. For example, if the value is 2.0, the head pointer should point to 2.0 as well, but the lower limit of insulating resistance is not 2.0. For example, if selected voltage is 250V, selected multiple is 2, selected value of the dial is 2.0, then the lower limit shall be: (2.0 2 2.5)MΩ = 500 MΩ.

4.4 Example One factory needs to sort a lot of metal-film capacitor, its specification is CBB-250V- 2.2μF and insulation resistance Rx>0G. 4.4.1 Turn on the power supply, the power lamp lights. Preheat for 15 minutes before test. 4.4.2 Turn the voltage select switch to 250V position, then 250V voltage will be outputted from measuring terminals. Based on Table 4.2.2, select multiple switch at 4 grade, Lower limit is 0GΩ, and the dial switch value will be: 0 9 / (2.5 4 6 ) = 4.0 4.4.3 Connect the measuring fixture provided by our company and begin measurement. 4.4.4 Discharge the measured capacitor. It is suggested to discharge outside the instrument as it needs certain discharging time due to that there is a 1 kω discharge resistor inside the instrument. V. Calibration, Inspection and Maintenance of the Tester 5.1 Creation and Check of DC voltage 5.1.1 Creation of V It is provided by resistors R18, R9 and adjuster W1. The accuracy can be adjusted by adjusting W1. 5.1.2 Creation of other voltages (For example of TH2681A) 5V voltage is created by the voltage stabilizer V20. As the pin 3 of N1 is a virtual ground, so its level is 0V. The constant current is determined by R17 and W2 and can be adjusted to 1mA by adjusting W2. When 1mA current flows7 through,different voltages can be acquired from R51, R52, R53, R54, R55, R56, R57 and R58 in which 50V voltage from R51, 0V from R52, 250V from R53 and R54, 500V from R55 and R56, 00V from R57and R58. 5.1.3 Should voltage created has larger error or voltage cannot be created, check all components in Section 5.1.2 to see if they are correct. Check components N1, V27, V28 and V29 and resistors and diodes around them to see if they are damaged, and replace them if necessary. 5.1.4 Check connection wires as the drawing below. Its measurement should meet requirement stipulated in Section 2.1. Should it not met, check feature and accuracy of components in Sections 5.1.1 to 5.1.3. 11

5.2 Inspection of Maximum Charging Current 5.2.1 When the tester make charge over measured component, it is a constant current and I = 20mA which is decided by circuit N2, R44, 425 and R2. The value I = R2 Vcc/(R12 R44). 5.2.2 The inspection of maximum charging current is as the drawing below. At this time the voltage grade is at 0V position. The switch K cannot be set at On position for longer time and seconds is suggested. When measurement is completed, set the switch K at OFF position. Should charging current has larger error, customer can make inspection as per Section 5.2.1, and make modification for charging current as per Section 5.2.1 if needed, and only the ratio value of resistors R2 and R12 should be modified. 5.3 Inspection of Insulation Resistance 5.3.1 Make inspection as per drawing and the table below. Its accuracy should meet requirement stipulated in Section 2.3. Measuring Voltage Standard Resistor Actually Measured Resistance 0V 1M M 0M 1G G 0G 12

Standard Resistor Measuring Voltage Actually Measured Resistance 0M V 25V 50V 0V 500V 00V VI. Warranty This instrument product is warranted against defects in material and workmanship for a period of two years from the date of shipment. The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by buyer, buyer-supplied unauthorized modification or misuse, operation outside of the environmental specifications for the product, or improper site preparation or maintenance. Address: No.3, Tianshan Road, Changzhou New District, Jiangsu, China Tel: 86-519-5132222 5113342 59592 Fax: 86-519-59972 Postcode: 213022 Website: www.tonghuico.com E-mail: tonghui@pub.cz.jsinfo.net Sales@tonghui.com.cn 13