Nikon Ti-E Microscope Manual. Rightmire Hall Ohio State University. Director: Tony Brown Rightmire

Similar documents
DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014

DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands

Nikon E800 Operating Instructions.

Nikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions

A BRIEF INTRODUCTION TO MICROSCOPY The two key properties of a microscope that allow you to see microbes are resolution and magnification.

Motorized Axio Observer Start-up instructions

Laboratory Introduction

Using the Nikon TE2000 Inverted Microscope

Nikon E800 Microscope. Operating Instructions

Zeiss LSM 880 Protocol

Nikon E800 Operating Instructions.

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com

Basic Microscopy for Plant Biology

Match the microscope structures given in the left column with the statements in the right column that identify or describe them.

Zeiss Axio Imager.A1 manual

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Things to check before start-up.

TABLE OF CONTENTS. Safety notes i. Care and Maintenance. ii. 1. Components Illustration Installation of Components.. 4

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope

Characterization Microscope Nikon LV150

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.

Components of the Microscope

User Manual. Trinocular Metallurgical Microscope. MicroscopeNet.com

Guide to Confocal 5. Starting session

Leica TCS SL Confocal Training. Neuroscience Imaging Core Staff. Core Director. Facility Manager

Basic Microscopy. OBJECTIVES After completing this exercise, you should be able to do the following:

CALIBRATION OF MICROSCOPE EYEPIECE GRATICULE

LAB 1 Introduction to Microscopy

OMM300. Inverted Metallurgical Microscope

Marine Invertebrate Zoology Microscope Introduction

MICROSCOPE LAB. Resolving Power How well specimen detail is preserved during the magnifying process.

Zoom Stereo Microscope NYMCS-360 Instruction Manual

Practical Light Microscopy

STEINDORFF METALLURGICAL MICROSCOPE. NYMCS-620 Instruction Manual

Zeiss AxioImager.Z2 Brightfield Protocol

Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope

Nikon C1si Spectral Laser Scanning Confocal Microscope. User Guide

Lab: The Compound Microscope

Imaging Introduction. September 24, 2010

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7

Leica TCS SP2 User Manual 1.1. Markus Friedrich and James Lee. Department of Biological Sciences Wayne State University

Biology 29 Cell Structure and Function Spring, 2009 Springer LABORATORY 1: THE LIGHT MICROSCOPE

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017.

User instructions Compound laboratory microscope

Manual for BMS E1 eplan series, compound microscope

INSTRUCTIONS FOR COURSE WORK 4 (AxioVert) Instructor: Anne Vaahtokari (MIU) 1. Purpose of the work

Therefore, all descriptions and illustrations in this instruction manual, including all specifications are subject to change without notice.

Zeiss LSM 780 Protocol

Basic Users Manual for Tecnai-F20 TEM

Indian Institute of technology Madras Presents NPTEL NATIONAL PROGRAMME ON TECHNOLOGY ENHANCED LEARNING

EPIFLUORESCENCE &/OR BRIGHTFIELD MICROSCOPY

SWIFT SERIES M2252DGL MICROSCOPE

Quick Start Guide. Leica SP5 X

TEKSCOPE MICROSCOPE. Models N2 Series USER S MANUAL

MICROSCOPY and CELL STRUCTURE

User instructions Compound laboratory microscope

3. are adherent cells (ie. cells in suspension are too far away from the coverslip)

Bio 252: Microscopy Study THE COMPOUND MICROSCOPE


Microscope (and The Cell) Lab Exercise #1

Brief manual how to start and close the Leica sp2 Confocal. (TCS SP2 AOBS system mounted on a DM IRE2)

Microbiology Laboratory 2

Anatomy: Introduction to the Light Microscope

User Manual. Trinocular Infinity Compound LED Microscope. MicroscopeNet.com

Nikon SIM-E & A1-R System

Figure 1. Oil-immersion objectives available for use with the Lionheart FX.

Eyepieces KHW10X. Diopter Adjustment Ring. Binocular Body Inclined 30. Binocular Clamp Screw. Analyzer control Lever. Reflected Light Illuminator

STEINDORFF NYMC Polarizing Microscope

LEICA TCS SP5 AOBS TANDEM USER MANUAL

ML7520 ML7530 DIOPTER ADJUSTMENT RING BINOCULAR BODY, INCLINED 30. (a) Field Iris Control Lever. (c) Filter Slots EYEPIECES, KHW10X

Swift M10 Series Microscope Use and Care Manual

EXERCISE 3 The Microscope

Biological Microscope Manual

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah

Practical work no. 3: Confocal Live Cell Microscopy

Microscopy. The dichroic mirror is an important component of the fluorescent scope: it reflects blue light while transmitting green light.

LSM 710 Confocal Microscope Standard Operation Protocol

Leica SP8 TCS Users Manual

Easy Kohler Illumination Method

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

The microscope is useful in making observations and collecting data in scientific experiments. Microscopy involves three basic concepts:

MICROSCOPY MICROSCOPE TERMINOLOGY

Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400.

Swift M3600, M3700 Series Digital & Compound Microscope

OM FL400. Reflected Light Fluorescence Microscope. Instruction Manual. Please read instructions carefully before using microscope.

MANUAL EXC-350 MICROSCOPE SERIES

Microscopy. Matti Hotokka Department of Physical Chemistry Åbo Akademi University

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com

Olympus IX71 Microscope and DP71 Camera Instructions

User instructions Transmitted light phase contrast microscope

2/4/15. Brightfield Microscopy! It s all about Magnification..! or is it?!

Basics of Light Microscopy and Metallography

Microscope ECLIPSE 80i Instructions <Reference>

Microscopes. A guide to use, general Maintenance, and repair tailored to the Olympus CX-21 microscope

MANUAL EXC-400 MICROSCOPE SERIES. 73 Mall Drive, Commack, NY (P) (F)

Biological Microscope

Industrial Microscope ECLIPSE LV100DA Instructions

Transcription:

Nikon Ti-E Microscope Manual Rightmire Hall Ohio State University Director: Tony Brown Rightmire 060 292-1205 brown.2302@osu.edu Facility Manager: Paula Monsma Rightmire 062 293-0939 292-1367 monsma.1@osu.edu This manual prepared by Paula Monsma. 1

Table of Contents Getting to know the microscope... 4 View from left... 4 View from right... 5 Microscope controls... 6 Microscope control panel... 6 LCD display panel... 7 Left side controls... 8 Right side controls... 9 Changing filters... 10 Changing objectives... 11 Focus options... 12 To focus using the focusing knobs... 12 Changing the coarseness of the focusing... 13 Escape and Refocus buttons... 14 Eyepieces... 15 Adjusting parfocality... 15 Stage holders... 16 Changing stage holders... 17 Placing a slide on the universal stage... 18 Transmitted light imaging... 19 Adjust transmitted light brightness... 19 Koehler illumination... 20 Bright field observation... 22 Differential interference contrast (DIC)... 23 Using Perfect Focus System ( PFS )... 24 Using the ASI Stage... 27 Objectives... 29 OIL immersion... 29 2

WATER immersion... 29 Cleaning objectives... 30 Specimen preparation... 31 DOs and DON Ts... 32 3

Getting to know the microscope View from left Transmitted light illumination column (tilts back for better access to stage) Eyepieces s Condenser turret Condenser lens Objective turret Filter cube turrets Left side controls Focus knob Microscope control panel 4

View from right Eyepieces Side port Right side controls Stage focus Transmitted light controls X-Y stage movement Perfect Focus controls 5

Microscope controls Microscope control panel LCD display panel PORTS directs the light Eye 100% to the eyepieces L100 100% to the left port R100 100% to the right port L80 80% to the left port, 20% to the eyepieces Mag changer Perfect Focus controls 6

LCD display panel The LCD display panel can display different information depending on which setting is chosen. Pressing the display buttons will change the display setting. Indicates current objective Indicates current filter cube Indicates current Z-position of objective Indicates current status of Perfect Focus Display buttons choosing these buttons will change what information is displayed 7

Left side controls Objective turret control buttons Focus step buttons- Change coarseness of focusing Current setting is indicated with green light. Halogen light controls (disabled) 8

Right side controls Focus step buttons Filter cube turret buttons Focus knob Refocus button Escape button 9

Changing filters Use the motorized fluorescent filter cube changer on the right side of the microscope: Press left button to rotate filter cube turret left Press right button to rotate filter cube turret right The current filter will display on the LCD display panel. The filter cube turret contains four filters, the DIC analyzer, plus an empty slot GFP : cube for green fluorochromes RFP : cube for red fluorochromes CFP : cube for cyan fluorochromes YFP : cube for yellow fluorochromes ANALYZER : DIC analyzer -- : empty position Notes: The empty position is used for confocal imaging and for transmitted light observation through the eyepieces 10

Changing objectives Use the objective turret control buttons on the left side of the microscope Press left key to increase magnification Press right key to decrease magnification Current objective indicated on LCD display panel 11

Focus options To focus using the focusing knobs One way to focus is using the focusing knobs located on the left and right side of the microscope. Turning the knob so that your thumb moves away from you focuses down Turning the knob so that your thumb moves toward from you focuses up Focusing knobs (turn in direction of arrows to focus objective down) 12

Changing the coarseness of the focusing The focusing is electronic and has three settings. Use the buttons on either the right or left side of the microscope. Please use the buttons on the right side of the microscope to avoid accidentally changing the objective. A light will display by the current setting. Press the upper button to change to a coarser focus. Press the lower button to change to a finer focus. 13

Escape and Refocus buttons The Escape and Refocus buttons are located on the Right Side Controls. Refocus button pressing this button will raise the objective to the last position before the Escape button was pressed. Improper use could result in objective damage. Be very careful when using this button! Escape button pressing this button will lower the objective to the escape position. To raise the objective after this, press the Refocus button. 14

Eyepieces Adjusting interpupillary distance: Adjust eyepieces to match interpupillary distance by moving the eyepieces closer together or further apart Adjusting parfocality Focus on specimen using electronic focusing controls Close right eye and adjust the left eyepiece so that the image appears in focus to your left eye Close the left eye and adjust the right eyepiece so that the image appears in focus to your right eye The eyepieces are now parfocal 15

Stage holders The microscope has two stage holders: Oko Lab heated chamber (maintains a constant temperature, as well as providing humidified air with or without CO 2 ) Universal stage holder (holds dishes and slides at room temperature) 16

Changing stage holders To remove either stage holder: Lower the objective all the way down Lift stage holder out of the stage. There is a springloaded tab that holds the chamber in place using pressure. When removing the Oko Lab chamber, carefully place it in the plastic container to the right of the microscope and place the plastic lid loosely on the plastic container. Pay special attention so that the wires do not crimped. Be careful when placing the wires and tubing here. Take care not to touch condenser lens surface when removing and replacing the stages! 17

Placing a slide on the universal stage To avoid touching or damaging the objective, lower the objective turret all the way using the focus knob 1 1 2 2 Position the slide holder clips as shown above Insert slide into the holder, pushing in the side supports so that the slide is supported on either side (1) Slide clips inward onto slide to secure the slide (2) 18

Transmitted light imaging Adjust transmitted light brightness Use the dial on the Prior LED to the right of the microscope. Prior LED This light will be on if the Prior LED is on. Press this button to turn the Prior LED on or off. 19

Koehler illumination Koehler illumination is essential to obtain good transmitted light images Select objective Open the condenser aperture diaphragm (move lever to the right) Focus on specimen Close the field aperture diaphragm (twist ring to the left) Focus condenser using condenser focusing knobs until the image of the aperture is sharp If necessary center the field diaphragm in the field of view using the two centering screws located on the front of the condenser Open field aperture diaphragm until it just disappears from the field of view Close down the condenser aperture diaphragm until the desired contrast is achieved 20

Field aperture diaphragm Condenser focusing knobs DIC polarizer Centering screws Condenser turret Condenser aperture diaphragm 21

Bright field observation Bright field observation means observation with transmitted light using no contrast enhancement method (e.g. no phase contrast or DIC) For bright field observation: rotate condenser turret to empty position switch filter cube turret to empty position Condenser turret in empty position 22

Differential interference contrast (DIC) For DIC, you need: A DIC objective (i.e. any objective with a magnification greater than 10x) A condenser prism in the condenser turret A polarizer above the condenser An analyzer in the filter cube position An objective prism beneath the objective (always in position) To set up DIC: Select a DIC objective (any objective except the 10x) The objective DIC prism required for that objective is already installed underneath the objective Manually rotate condenser turret to the N2 position (all of the DIC objectives use the N2 condenser prism) Insert polarizer into light path Change to the analyzer position in the filter cube turret Polarizer Condenser turret 23

Using Perfect Focus System ( PFS ) The microscope is equipped with a Perfect Focus System. When this system is engaged, the microscope will maintain a specific focus by utilizing a very long wavelength beam of light to determine what the offset from the coverslip should be. The PFS will only work within a small distance of the coverslip, so it will not be usable for every sample. To use PFS, there are several things that need to happen. The PFS prism needs to be in the in position PFS prism lever 24

The sample needs to be in the right focal range. The Focus green light will be on when the sample is in the PFS range. The display indicates that PFS is off. The FOCUS green light indicates that the sample is in the PFS range. When the Focus green light is on, the PFS button can be pressed, turning PFS on. The display indicates that PFS is on. The ON button lights up when pressed. 25

Use the PFS focus controls to change focus. The focusing knobs on the microscope and the ASI stage will not do anything when the PFS is on. PFS focus 26

Using the ASI Stage The microscope is equipped with an ASI motorized stage. The stage can be used to move the sample in the X-Y direction, as well as changing the focus. ASI Stage Controls Stage X-Y Joystick - Move the joystick to move the sample. Push the button on top of the joystick to change from fast to slow movement. Stage Information Panel Stage Focus 27

The Stage Information Panel displays the current position of the stage. The coordinates of the stage position can be recorded, allowing a user to easily return to a specific position on a slide. 28

Objectives OIL immersion: Clean coverslips before using immersion objectives Apply oil to the objective or to the coverslip before placing your slide on the stage Use only NIKON IMMERSION OIL Use the minimum amount of oil necessary If you use too much oil it may run down the side of the objective and damage the optics! To apply oil to an objective: tilt the oil bottle so that oil gathers in the dropper squeeze gently so that oil reaches the end of the dropper (hold against a piece of lens paper to avoid a mess) apply oil to objective by gently squeezing over the lens without introducing bubbles DO NOT TOUCH THE LENS DIRECTLY! WATER immersion: The 40x and 60x water immersion objectives use water rather than oil as the immersion medium. Contact Paula for instructions on using these objectives. 29

Cleaning objectives This microscope is equipped with the highest quality objectives, with a total value exceeding $30,000! You must exercise great care to preserve these objectives To clean oil off immersion objectives: Remove specimen from stage Blot (not wipe!) off excess oil using lens tissue Use a fresh area of the lens tissue for each blot Repeat blotting until no more oil comes off onto the lens tissue Wipe the metal housing around the lens using lens tissue It typically takes 2 or 3 sheets of lens tissue to properly clean a single objective What to do if you accidentally get oil on a dry objective: Blot off excess with lens tissue Then contact Paula, who will clean the lens with special cleaning fluid (do not attempt this yourself!) Lens tissue alone will not get all the oil off the lens and this will interfere with the optics 30

Specimen preparation Some general tips: Select fluorochromes that are optimally excited by the confocal laser lines For multiple labeling, the less overlap between the excitation and emission spectra the better Always use #1.5 coverslips Mount coverslip to slide securely and seal with nail enamel or mounting medium that solidifies Do not observe on the microscope until the nail enamel or mounting medium has hardened! Secure the slide with the clips when acquiring Z series 31

DOs and DON Ts DOs Use only lens tissue to clean objectives Use Kimwipes to clean coverslips and slides DON Ts Never touch any optical surface with anything other than lens tissue Never clean the objectives with any kind of solvent Never use Kimwipes to clean objectives or any other optics Never use any liquid cleaners or solvents to clean the microscope optics or any part of the microscope! When to ask for Paula s assistance Adding or removing objectives Adding or removing filter cubes Cleaning oil off 10x or 20x dry objectives Cleaning oil off the 40x or 60x water objectives Using the objective warmer Using the Okolab heated chamber Any other time you are unsure what you are doing! 32