DIGITAL STORAGE OSCILLOSCOPES

Similar documents
3.2 Measuring Frequency Response Of Low-Pass Filter :

ENG 100 Lab #2 Passive First-Order Filter Circuits

UNIVERSITY OF CALIFORNIA, SANTA BARBARA Department of Electrical and Computer Engineering. ECE 2A & 2B Laboratory Equipment Information

Test No. 1. Introduction to Scope Measurements. Report History. University of Applied Sciences Hamburg. Last chance!! EEL2 No 1

LABORATORY 4. Palomar College ENGR210 Spring 2017 ASSIGNED: 3/21/17

B. Equipment. Advanced Lab

Faculty of Engineering, Thammasat University

DEPARTMENT OF INFORMATION ENGINEERING. Test No. 1. Introduction to Scope Measurements. 1. Correction. Term Correction. Term...

The University of Jordan Mechatronics Engineering Department Electronics Lab.( ) Experiment 1: Lab Equipment Familiarization

Combinational logic: Breadboard adders

Lab #5 Steady State Power Analysis

PHYSICS 171 UNIVERSITY PHYSICS LAB II. Experiment 4. Alternating Current Measurement

Notes on Experiment #1

Pre-Lab. Introduction

Laboratory Equipment Instruction Manual 2011

Sampling and Reconstruction

Oscilloscope and Function Generators

EXPERIMENT 2 DIGITAL STORAGE OSCILLOSCOPE

The Oscilloscope. Vision is the art of seeing things invisible. J. Swift ( ) OBJECTIVE To learn to operate a digital oscilloscope.

Frequency and Time Domain Representation of Sinusoidal Signals

Appendix A: Specifications

AC Magnitude and Phase

LAB I. INTRODUCTION TO LAB EQUIPMENT

LAB #7: Digital Signal Processing

Experiment 9 The Oscilloscope and Function Generator

Introduction to Lab Instruments

Introduction to basic laboratory instruments

ECE 2274 Lab 2 (Network Theorems)

Lab 3: AC Low pass filters (version 1.3)

Specifications for DS1000CA Series

EENG-201 Experiment # 4: Function Generator, Oscilloscope

N acquisitions, all channels simultaneously, N is selectable from 4, 16, 64, and 128 Inputs

Introduction to Oscilloscopes Instructor s Guide

2 : AC signals, the signal generator and the Oscilloscope

Agilent 33522A Function Arbitrary Waveform Generator. Tektronix TDS 3012B Oscilloscope

DSO 3000 Series Oscilloscope

Experiment 1.A. Working with Lab Equipment. ECEN 2270 Electronics Design Laboratory 1

LAB I. INTRODUCTION TO LAB EQUIPMENT

Tektronix Courseware. Academic Labs. Sample Labs from Popular Electrical and Electronics Engineering Curriculum

Sonoma State University Department of Engineering Science Spring 2017

Laboratory 1 Generating and viewing signals rev. 20e. Oscilloscope Settings

MSO-5000B Mixed Storage Oscilloscope User Manual

PHY152 Experiment 4: Oscillations in the RC-Circuits (Measurements with an oscilloscope)

How to Setup and Use an Oscilloscope

Operational Amplifiers 2 Active Filters ReadMeFirst

Chapter 5 : Specifications

Oscilloscope Fundamentals. For Electrical Engineering and Physics Undergraduate Students

AP034-OM-E Rev D ISSUED: January 2000 ²

DSO5000P Series Digital Storage Oscilloscope User Manual. (Version 1.1)

ECE 2274 Lab 2. Your calculator will have a setting that will automatically generate the correct format.

EC310 Security Exercise 20

USE OF BASIC ELECTRONIC MEASURING INSTRUMENTS Part II, & ANALYSIS OF MEASUREMENT ERROR 1

Measurement Bench. Accessories. Power supply. Wave form generator. Multimetre. Oscilloscope. Dr. L.Scucchia

Getting Started. MSO/DPO Series Oscilloscopes. Basic Concepts

ME 365 EXPERIMENT 1 FAMILIARIZATION WITH COMMONLY USED INSTRUMENTATION

UCE-DSO210 DIGITAL OSCILLOSCOPE USER MANUAL. FATIH GENÇ UCORE ELECTRONICS REV1

Lab #2: Electrical Measurements II AC Circuits and Capacitors, Inductors, Oscillators and Filters

University of Jordan School of Engineering Electrical Engineering Department. EE 204 Electrical Engineering Lab

EE 210: CIRCUITS AND DEVICES

Exp. #2-6 : Measurement of the Characteristics of,, and Circuits by Using an Oscilloscope

ENGR 1110: Introduction to Engineering Lab 7 Pulse Width Modulation (PWM)

ECE 2274 Lab 1 (Intro)

Tektronix digital oscilloscope, BK Precision Function Generator, coaxial cables, breadboard, the crystal earpiece from your AM radio kit.

Laboratory Experiment #1 Introduction to Spectral Analysis

Time-Varying Signals

Fig. 1. NI Elvis System

LAB II. INTRODUCTION TO LAB EQUIPMENT

P a g e 1 ST985. TDR Cable Analyzer Instruction Manual. Analog Arts Inc.

2 Oscilloscope Familiarization

On-Line Students Analog Discovery 2: Arbitrary Waveform Generator (AWG). Two channel oscilloscope

Data Sheet. Digital Storage Oscilloscope. Features & Benefits. Applications. Ease-of-Use Feature DSO5202BMT DSO5102BMT DSO5062BMT

EE 3302 LAB 1 EQIUPMENT ORIENTATION

PHYS 235: Homework Problems

DST Series B Type Digital Storage Oscilloscope User Manual

Department of Electrical & Computer Engineering Technology. EET 3086C Circuit Analysis Laboratory Experiments. Masood Ejaz

UCE-DSO212 DIGITAL OSCILLOSCOPE USER MANUAL. UCORE ELECTRONICS

total j = BA, [1] = j [2] total

Laboratory equipments. Parameters of digital signals.

332:223 Principles of Electrical Engineering I Laboratory Experiment #2 Title: Function Generators and Oscilloscopes Suggested Equipment:

Introduction to basic laboratory instruments

Introduction to Basic Laboratory Instruments

ECE 3155 Experiment I AC Circuits and Bode Plots Rev. lpt jan 2013

EE 201 Function / Arbitrary Waveform Generator and Oscilloscope Tutorial

The object of this experiment is to become familiar with the instruments used in the low noise laboratory.

Department of Electrical and Computer Engineering. Laboratory Experiment 1. Function Generator and Oscilloscope

ENGR 210 Lab 6 Use of the Function Generator & Oscilloscope

Physics 310 Lab 2 Circuit Transients and Oscilloscopes

Group: Names: (1) In this step you will examine the effects of AC coupling of an oscilloscope.

Appendix A: Laboratory Equipment Manual

Lab #1 Lab Introduction

I = I 0 cos 2 θ (1.1)

ECE3204 D2015 Lab 1. See suggested breadboard configuration on following page!

DS1000B Series Digital Oscilloscopes

EECS 318 Electronics Lab Laboratory #2 Electronic Test Equipment

Experiment #2: Introduction to Lab Equipment: Function Generator, Oscilloscope, and Multisim

Product Channels Bandwidth Sampling Rate Memory Resolution

Laboratory 3 (drawn from lab text by Alciatore)

Frequency Response and Filters

3.003 Lab 3 Part A. Measurement of Speed of Light

1.5k. (a) Resistive Circuit (b) Capacitive Circuit

Transcription:

DIGITAL STORAGE OSCILLOSCOPES Electronic Measurements Lab Massimo Ortolano 2016 POLITECNICO DI TORINO

c 2011 2016 Massimo Ortolano Dipartimento di Elettronica e Telecomunicazioni (DET) Politecnico di Torino Corso Duca degli Abruzzi, 24 10129 Torino Italy Email: massimo.ortolano@polito.it This work is licensed under the Creative Commons Attribution-NonCommercial-ShareAlike 3.0 Unported License. To view a copy of this license, visit http://creativecommons.org/licenses/by-nc-sa/ 3.0/ or send a letter to Creative Commons, 444 Castro Street, Suite 900, Mountain View, California, 94041, USA.

Preliminaries In this lab you will practice with digital storage oscilloscopes (DSOs). Have a look at your workbench, you will need the following laboratory equipment (check here for specifications), components and cables: Bench DC power supply Signal generator board (LED product) Bench function generator Digital storage oscilloscope Breadboard 1 banana plug to banana plug cable (fig. 1) 1 three-wire power cable (fig. 2) 2 BNC male to BNC male cables (figg. 3 and 4) 1 crocodile clips to BNC male cable (fig. 5) 1 oscilloscope probe (fig. 6)

Figure 1: A banana plug. Figure 2: Three-wire power cable for the signal generator board. (Preliminaries)

Figure 3: A BNC male Figure 4: A BNC male to Figure 5: A crocodile (al- connector. BNC male cable. ligator in the US) clip. Figure 6: An oscilloscope probe. Figure 7: probe body. (Preliminaries) Oscilloscope

Part I Basic usage 1 Getting started 1.1 A quick tour of the oscilloscope front panel Power up the oscilloscope, and take a look at the oscilloscope s front panel. Knobs and keys are divided in three main sections: vertical, horizontal and trigger. There are also other miscellaneous controls. Vertical In this area you will find the vertical controls, one set of controls for each vertical channel. The biggest knob can be used to change the vertical scale. Look at the front-panel or at the read-out: what are the minimum and maximum values (expressed in V/div) of the vertical scale? Try the vertical position knob ( position): what happens to the corresponding trace? The vertical menu key calls up a menu with the following main submenus: Coupling (DC, AC or GND), BW limit (the bandwidth of a vertical channel can be limited to 20 MHz to improve the display of noisy signals), Vertical scale adjustment (fine/coarse) and Probe (1, 10 etc.). Check the different options. Warning: In the following, be sure that the probe type is set to 1, otherwise voltage readings will be wrong. Be sure, too, that the vertical scale adjustment is set to coarse. Horizontal In this area you will find the horizontal (time-base) controls. The biggest

knob can be used to change the horizontal scale. Look at the front-panel or at the read-out: what are the minimum and maximum values (expressed in s/div) of the horizontal scale? Try the horizontal position knob ( position). Since there are no input signals, you will not see any change in the traces; look at the readout, instead: what happens? The horizontal menu key calls up a menu which allows the selection of a waveform segment to display (more in 4.5). Warning: In the following, be sure that the time-base is set to Main, in order to display the entire waveform. Be sure, too, that the trigger position indicator ( or T) is centred horizontally: if not, centre it by rotating the horizontal position knob. Trigger In this area you will find the trigger controls, which allow you to set the trigger operating mode and level (more in 2.3). Miscellaneous controls These include run controls (e.g. the Run/Stop key and the Run single key) which allow the user to start or stop an acquisition or to run single acquisitions. Other keys allow to change the acquisition mode or the sampling mode (Acquire), to activate cursors (Cursors) or to run automated measurements (Measure).

1.2 Powering the signal generator board The signal generator board requires a 12 V dual symmetrical power supply: 1 Turn on the DC power supply. Set the switches on the front panel to select the tracking/series mode of operation: in this mode of operation the output voltage of the slave section is equal to that of the master section and the two main outputs are connected in series internally. 2 By turning the voltage adjustment knob of the master section, adjust the output voltage to 12 V. Warning: Adjust the output voltage of the laboratory power supply before making any connection to the signal generator board. 3 Turn off the DC power supply. Using the three-wire power cable, connect the signal generator board to the power supply: red, 12 V; black, 12 V; and green, 0 V. Warning: Do not connect the green cable to the power supply s earth ground! The green cable must be connected to the common point between the two main power outputs (fig. 8). 4 Turn on the power supply.

0 30 0 30 + + Figure 8: Powering the signal generator board: at the end of 1.2, the power-supply set-up should look like the above. ( 1.2)

2 Oscilloscope s settings 2.1 Traces 1 Turn on the CH1 trace and turn off CH2. 2 Set the trigger mode to Auto, the trigger source to CH1 and the trigger coupling to AC. Set the trigger level around 0 V. 3 Set CH1 input coupling to AC. Centre the trace vertically on the screen. 4 Connect CH1 to the signal generator board with a BNC male to BNC male cable, to get signal A0 (output A, signal no. 0 on the display). 5 Set vertical scale for CH1 in order to expand the trace as much as possible within the display limits (the trace should occupy most of the vertical scale). 6 Set the horizontal scale in order to expand a waveform period as much as possible within the display limits (a single period should occupy most of the horizontal scale).

2.2 Acquisition modes A DSO can have several different acquisition modes, e.g. sample, peak detect, average. The sample mode is typically the default acquisition mode at turn-on. The acquisition mode can be changed by pressing the Acquire key. The peak-detect mode can be useful to detect random spikes, while the average mode can be useful to reduce the effect of noise and obtain a thinner trace. 1 With the oscilloscope set as in 2.1, change the acquisition mode to average. See what happens when changing the number of averages.

2.3 Trigger 1 Turn on the CH1 trace and turn off CH2. 2 Set the trigger mode to Auto, the trigger source to CH1 and the trigger coupling to DC. Set the trigger level around 0 V. 3 Set CH1 input coupling to DC. Centre the trace vertically on the screen. 4 Connect CH1 to signal A4. 5 Set CH1 vertical scale in order to expand the trace as much as possible within the display limits (the trace should occupy most of the vertical scale). 6 Set the horizontal scale in order to expand a waveform period as much as possible within the display limits (a single period should occupy most of the horizontal scale). 7 What happens if you change the trigger level? 8 Change the trigger slope from the trigger menu: what happens? 9 Change the trigger source to CH2: what happens? 10 Change the trigger mode to Normal: what happens? 11 Change the trigger source back to CH1: what happens?

3 Basic measurements 3.1 Amplitude measurements 1 Set the oscilloscope as in 2.1. Set the acquisition mode to average and the number of averages to 64. 2 Measure the peak-to-valley signal amplitude with the aid of the screen graticule. 3 Measure the peak-to-valley signal amplitude by using the cursors. 4 Measure the peak-to-valley signal amplitude by using the oscilloscope s automatic measurement mode, if available. 5 Determine the uncertainties of the above measurements (check the oscilloscope specifications).

3.2 Period and frequency measurements 1 Measure the period T of the signal with the aid of the screen graticule and calculate the corresponding frequency f = 1/T. 2 Measure the signal period and frequency by using the cursors. 3 Measure the signal period and frequency by using the oscilloscope s automatic measurement mode, if available. 4 Determine the uncertainties of the above measurements (check the oscilloscope s specifications).

3.3 Phase measurements 1 Turn on both CH1 and CH2. 2 Set the trigger mode to Auto, the trigger source to CH1 and the trigger coupling to AC. Set the trigger level around 0 V. 3 Set the input coupling of both channels to AC. Centre both traces vertically on the screen (be precise, otherwise you will get a measurement error). 4 Connect the two oscilloscope channels to signals A0 and B0. 5 Set the vertical scales of both channels in order to expand the traces as much as possible within the display limits. 6 Set the horizontal scale in order to expand a waveform period as much as possible within the display limits. 7 Determine the phase shift ϕ between the two signals according to the definition given in figure 9. 8 Determine the uncertainty of the phase shift.

v 1 (t) v 2 (t) O t T t ϕ = 360 t T Figure 9: Phase measurements: t is the time interval (with sign) between two nearby zero crossings; here, if signal 1 is taken as reference, t should be considered negative because the output signal lags (i.e. it is delayed) with respect to the input one.

4 Pulse wave measurements 4.1 Period 1 Turn on the CH1 trace and turn off CH2. 2 Set the trigger mode to Auto, the trigger source to CH1 and the trigger coupling to DC. Set the trigger level around 0 V. 3 Set CH1 input coupling to DC. Centre the trace vertically on the screen. 4 Connect CH1 to signal B4. 5 Set CH1 vertical scale in order to expand the trace as much as possible within the display limits (the trace should occupy most of the vertical scale). 6 Set the horizontal scale in order to expand a waveform period as much as possible within the display limits (a single period should occupy most of the horizontal scale; change the horizontal position for best results). 7 Measure the signal period T using the graticule, the cursors and the automated measurement mode. 8 Determine the uncertainty of T for each measurement.

4.2 Pulse width 1 Set the horizontal scale to expand the pulse as much as possible within the display limits. If necessary, change the horizontal position too. 2 Measure the pulse width t w using the graticule, the cursors and the automated measurement mode. 3 Determine the uncertainty of t w for each measurement.

4.3 Duty factor 1 From the results of 4.1 and 4.2, calculate the duty factor d = t w /T. 2 Determine the uncertainty of d. 3 The duty factor can also be measured with the following method. Set the oscilloscope as in 4.1 and determine d as the ratio of the pulse divisions to the period divisions. Determine the uncertainty of d in this second measurement. Are the two measurements compatible?

4.4 Levels and DC component 1 Measure the voltage V L corresponding to the pulse low level (recall that CH1 input coupling should be set to DC). 2 Measure the voltage V H corresponding to the pulse high level. 3 From V L, V H and the duty factor d measured in 4.3, calculate the DC component of the pulse wave, that is, its mean value over a period: t0 +T V dc = 1 T t 0 v(t) dt, where t 0 is an arbitrary initial time. 4 If you change the CH1 input coupling from DC to AC, the trace on the screen will shift vertically by the amount V dc. Measure the DC component in this way and check if this result is compatible with the one above.

4.5 Rise and fall times 1 Set the horizontal scale to expand the pulse rising edge as much as possible within the display limits. If necessary, change the horizontal position too. To expand the rising edge, you can also use the Window Zone/Window submenus of the horizontal menu. 2 Measure the pulse rise time. 3 Set the horizontal scale to expand the pulse falling edge as much as possible within the display limits. If necessary, change the horizontal position too. 4 Measure the pulse falling time.

Part II Other measurements, oscilloscope probe 5 Step response of an RC low-pass filter 5.1 Assembling the circuit 1 Calculate the nominal cut-off frequency f H and the nominal step-response risetime t r for the circuit of figure 10. 2 Assemble the circuit of figure 10 on the breadboard. Figure 11 shows a possible way of assembling the circuit.

R 15 kω V i 270 pf C V o Figure 10: Circuit diagram of the RC filter. Figure 11: A possible way of assembling the RC filter of figure 10.

5.2 Rise-time and cut-off frequency (Coaxial cable) 1 Using a BNC-to-BNC cable, a BNC T adapter (fig. 12) and a coaxial cable with crocodile clips, connect the function generator output to the oscilloscope CH1 input and then to the circuit input (V i ). Warning: Recall that the black clip is the ground. 2 Using a coaxial cable with crocodile clips, connect the circuit output (V o ) to the oscilloscope CH2 input. 3 Set the input coupling of both inputs to DC. 4 Using a BNC-to-BNC cable, connect the trigger output of the bench function generator to the external trigger input of the oscilloscope. Set the trigger source to External. 5 Set the function generator output for a square wave with a frequency of 1 khz. 6 Measure the step-response rise time and calculate the cut-off frequency from the formula f H = 0.35/t r. Compare these values with those obtained in 5.1. Do they agree? How do you explain the result obtained? Figure 12: A BNC T adapter

5.3 Rise-time and cut-off frequency (Oscilloscope probe) 1 Disconnect the coaxial cable from the circuit output and the CH2 input and replace it with an oscilloscope probe (1 : 10 attenuation). 2 Compensate the probe (ask for a suitable screwdriver). 3 Measure the step-response rise time and calculate the cut-off frequency from the formula f H = 0.35/t r. Compare these values with those obtained in 5.1. Is there any difference with the values obtained in 5.2? 4 Correct the values obtained in the previous point for the errors caused by the oscilloscope intrinsic rise time and by the function generator output resistance. State if these errors are negligible or not.

6 Frequency response of the RC low-pass filter A generic linear filter (fig. 13) is characterised by its (complex) transfer function H(f ) = V o V i, where V i is the phasor associated with a sinusoidal input voltage at frequency f, and V o is the phasor associated with the corresponding sinusoidal output voltage. V i H(j2πf ) V o Figure 13 H(f ) is thus a complex number which changes with f. The Bode plots are a graphical representation of the quantities G(f ) = 20 log H(f ) and ϕ(f ) = arg H(f ) as a function of frequency (fig. 14). G(f ) is the magnitude of the transfer function, expressed in decibel (db), and ϕ(f ) is the phase. The frequency axis is logarithmic.

GH f LdB 0. -5. -10. -15. -20. f -25. 104 105 Hz j deg 0. -20. -40. -60. -80. f 104 105 Figure 14: Example Bode plots. Hz

1 Measure G(f ) and ϕ(f ) for the filter assembled in 5 over a frequency range from 500 Hz to 500 khz (see also figure 15 for a definition of the parameters). I suggest to make measurements at the following frequency points, but you can also make other choices: 500 Hz, 800 Hz, 1200 Hz, 2000 Hz, 3000 Hz, 5 khz, 8 khz, 12 khz, 20 khz, 30 khz, 50 khz, 80 khz, 120 khz, 200 khz, 300 khz and 500 khz. 2 Search the 3 db cut-off frequency by continuously varying the input frequency; compare this value with the one obtained in 5.3. 3 Draw the Bode plot of the measured frequency response.

v i (t) v o (t) G(f ) = V o V i O t T t 2 V o 2 V i ϕ(f ) = 360 t T Figure 15: This figure shows an example of input (black) and output (red) voltage waveforms. t is the time interval (with sign) between two nearby zero crossings. Here t should be considered negative because the output signal lags (i.e. it is delayed) with respect to the input one.

R 1 100 kω C 1 1 nf V i C 2 270 pf R 2 100 kω V o Figure 16 7 An uncompensated voltage divider 1 Assemble the circuit of figure 16. Figure 17 shows a possible way of assembling the circuit. 2 Measure the frequency response (magnitude and phase) for the filter assembled in 5 over a frequency range from 200 Hz to 20 khz (see also figure 15 for a definition of the parameters). I suggest to make measurements at the following frequency points, but you can also make other choices: 200 Hz, 350 Hz, 630 Hz, 1.1 khz, 2 khz, 3.5 khz, 6.3 khz, 11 khz and 20 khz. 3 Draw the Bode plot and compare it with the theoretical one (homework).

Figure 17: A possible way of assembling the voltage divider of figure 17.