OPERATION OF A SINGLE PASS, BUNCH-BY-BUNCH X-RAY BEAM SIZE MONITOR FOR THE CESR TEST ACCELERATOR RESEARCH PROGRAM*

Similar documents
Operation of a Single Pass, Bunch-by-bunch x-ray Beam Size Monitor for the CESR Test Accelerator Research Program. October 3, 2012

CESRTA Low Emittance Tuning Instrumentation: x-ray Beam Size Monitor

x-ray Beam Size Monitor

SIMULATION CODES. Proceedings of IBIC2014, Monterey, CA, USA

X-Ray Beam Size Monitor for CESRTA

ELECTRON CLOUD DENSITY MEASUREMENTS USING RESONANT MICROWAVES AT CESRTA

ELECTRON CLOUD MITIGATION INVESTIGATIONS AT CESR-TA

Measurements of Small Vertical Beamsize using a Coded Aperture at Diamond Light Source. C. Bloomer G. Rehm J.W. Flanagan

RESULTS ON FIELD MEASUREMENTS IN A FLAT POLE MAGNET WITH THE CURRENT CARING SHEETS

MEASUREMENT OF BEAM LOSSES USING OPTICAL FIBRES AT THE AUSTRALIAN SYNCHROTRON

Nonintercepting Diagnostics for Transverse Beam Properties: from Rings to ERLs

arxiv: v3 [physics.acc-ph] 4 Aug 2017

FIRST INDIRECT X-RAY IMAGING TESTS WITH AN 88-mm DIAMETER SINGLE CRYSTAL

Studies of Electron Cloud Growth and Mitigation in Wigglers Using Retarding Field Analyzers

Electron cloud effects, codes & simulations. K. Ohmi (KEK) ICAP12, Aug, 2012 Rostock

Light Source Diagnostics. Hywel Owen ASTEC Daresbury Laboratory

BEAM HALO OBSERVATION BY CORONAGRAPH

Introduction... 3 Slits for AIR Operation... 4 Slits in Vacuum Vessels... 5 Slits for High Vacuum Operation... 6 Custom Slits... 7 Steel Slits...

ITk silicon strips detector test beam at DESY

Supplementary Information

RF System Models and Longitudinal Beam Dynamics

X-Ray Transport, Diagnostic, & Commissioning Plans. LCLS Diagnostics and Commissioning Workshop

A Study of undulator magnets characterization using the Vibrating Wire technique

Examination of Microphonic Effects in SRF Cavities

RANDY W. ALKIRE, GEROLD ROSENBAUM AND GWYNDAF EVANS

SUPERCONDUCTING RF IN STORAGE-RING-BASED LIGHT SOURCES

Diagnostics for Electron Cloud Measurements in CESR

Implementation of A Nanosecond Time-resolved APD Detector System for NRS Experiment in HEPS-TF

Lawrence Berkeley Laboratory UNIVERSITY OF CALIFORNIA

LHCb Preshower(PS) and Scintillating Pad Detector (SPD): commissioning, calibration, and monitoring

High Rep-Rate KrF Laser Development and Intense Pulse Interaction Experiments for IFE*

HIGH POSITION RESOLUTION AND HIGH DYNAMIC RANGE STRIPLINE BEAM POSITION MONITOR (BPM) READOUT SYSTEM FOR THE KEKB INJECTOR LINAC TOWARDS THE SuperKEKB

Beam Infrared Detection with Resolution in Time

PArticles in an accelerator generally oscillate in directions

Check the LCLS Project website to verify 2 of 7 that this is the correct version prior to use.

ANALYSIS OF 3RD OCTAVE BAND GROUND MOTIONS TRANSMISSION IN SYNCHROTRON RADIATION FACILITY SOLARIS Daniel Ziemianski, Marek Kozien

X-Ray Detection Using SOI Monolithic Sensors at a Compact High-Brightness X-Ray Source Based on Inverse Compton Scattering

Electronic Readout System for Belle II Imaging Time of Propagation Detector

HIGHER ORDER MODES FOR BEAM DIAGNOSTICS IN THIRD HARMONIC 3.9 GHZ ACCELERATING MODULES *

Superconducting RF System. Heung-Sik Kang

Cavity BPMs for the NLC

Experience with Insertion Device Photon Beam Position Monitors at the APS

Development of a Vibration Measurement Method for Cryocoolers

Pixel hybrid photon detectors

Supplementary Figure 1

The Architecture of the BTeV Pixel Readout Chip

UV/X-RAY DIFFRACTION RADIATION FOR NON-INTERCEPTING MICRON-SCALE BEAM SIZE MEASUREMENT

Upgrade of the ultra-small-angle scattering (USAXS) beamline BW4

LHC TRANSVERSE FEEDBACK SYSTEM: FIRST RESULTS OF COMMISSIONING. V.M. Zhabitsky XXI Russian Particle Accelerator Conference

Directly Chirped Laser Source for Chirped Pulse Amplification

The Ecloud Measurement Setup in the Main Injector

Supplementary Figure 1. Effect of the spacer thickness on the resonance properties of the gold and silver metasurface layers.

AIDA-2020 Advanced European Infrastructures for Detectors at Accelerators. Deliverable Report. CERN pixel beam telescope for the PS

DEVELOPMENT OF CAPACITIVE LINEAR-CUT BEAM POSITION MONITOR FOR HEAVY-ION SYNCHROTRON OF KHIMA PROJECT

Total Absorption Dual Readout Calorimetry R&D

PICO MASTER 200. UV direct laser writer for maskless lithography

Beam Diagnostics, Low Level RF and Feedback for Room Temperature FELs. Josef Frisch Pohang, March 14, 2011

Norbert Meyners, DESY. LCTW 09 Orsay, Nov. 2009

Electromagnetic characterization of materials for the CLIC Damping Rings and high frequency issues

Attosecond Diagnostics of Muti GeV Electron Beams Using W Band Deflectors

Silicon Photomultiplier Evaluation Kit. Quick Start Guide. Eval Kit SiPM. KETEK GmbH. Hofer Str Munich Germany.

Detection of Beam Induced Dipole-Mode Signals in the SLC S-Band Structures* Abstract

Short-Pulse X-ray at the Advanced Photon Source Overview

1 Detector simulation

Signal Reconstruction of the ATLAS Hadronic Tile Calorimeter: implementation and performance

BEAM SIZE MEASUREMENTS USING SYNCHROTRON RADIATION INTERFEROMETRY AT ALBA

PACS codes: Qx, Nc, Kv, v Keywords: Digital data acquisition, segmented HPGe detectors, clock and trigger distribution

New apparatus for precise synchronous phase shift measurements in storage rings 1

CALICE AHCAL overview

CSPADs: how to operate them, which performance to expect and what kind of features are available

Vertical External Cavity Surface Emitting Laser

TURN-BY-TURN BPM SYSTEM USING COAXIAL SWITCHES AND ARM MICROCONTROLLER AT UVSOR

Single bunch x-ray pulses on demand from a multi-bunch synchrotron radiation source. Resonant pulse picking and MHz Chopper

National Accelerator Laboratory

Multianode Photo Multiplier Tubes as Photo Detectors for Ring Imaging Cherenkov Detectors

A Modular Readout System For A Small Liquid Argon TPC Carl Bromberg, Dan Edmunds Michigan State University

A Synchrotron Phase Detector for the Fermilab Booster

Ultrafast instrumentation (No Alignment!)

Mass Spectrometry and the Modern Digitizer

FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION

Totem Experiment Status Report

Beam Size Monitors for KEKB, ILCDR. J.W. Flanagan ILC DR Workshop 19 Dec. 2007

Improving boundary conditions for microwave reflection to measure electron cloud density

Magnetic measurement system for superconducting final focus quadrupoles for SuperKEKB

Simulation of Algorithms for Pulse Timing in FPGAs

An ASIC dedicated to the RPCs front-end. of the dimuon arm trigger in the ALICE experiment.

Diamond sensors as beam conditions monitors in CMS and LHC

Position of the LHC luminous region

Influence of Distributed Ion Pump Voltage on the Anomalous Instability in CESR D.L. Hartill, T. Holmquist, J.T. Rogers, and D.C.

Experiences of the QSBPM System on MAX II

Results from the Stanford 10 m Sagnac interferometer

High collection efficiency MCPs for photon counting detectors

Performance of 8-stage Multianode Photomultipliers

Precision RF Beam Position Monitors for Measuring Beam Position and Tilt Progress Report

Status of the PRad Experiment (E )

Instructions for the Experiment

Activities on Beam Orbit Stabilization at BESSY II

Study of the ALICE Time of Flight Readout System - AFRO

Testing the Electronics for the MicroBooNE Light Collection System

Encoding and Code Wheel Proposal for TCUT1800X01

Transcription:

OPERATION OF A SINGLE PASS, BUNCH-BY-BUNCH X-RAY BEAM SIZE MONITOR FOR THE CESR TEST ACCELERATOR RESEARCH PROGRAM* N.T. Rider, M. G. Billing, M.P. Ehrlichman, D.P. Peterson, D. Rubin, J.P. Shanks, K. G. Sonnad, CLASSE, Cornell University, Ithaca, NY 14853, U.S.A. M. A.Palmer, Fermilab, Batavia, IL 60510, U.S.A. J.W. Flanagan, KEK, Ibaraki 305-0801, Japan Abstract The CESR Test Accelerator (CESRTA) program targets the study of beam physics issues relevant to linear collider damping rings and other low emittance storage rings. This endeavour requires new instrumentation to study the beam dynamics along trains of ultra low emittance bunches. A key element of the program has been the design, commissioning and operation of an x-ray beam size monitor capable, on a turn by turn basis, of collecting single pass measurements of each individual bunch in a train over many thousands of turns. This new instrument utilizes custom, high bandwidth amplifiers and digitization hardware and firmware to collect signals from a linear InGaAs diode array. The instrument has been optimized to allow measurements with 3x10 9 to 1x10 11 particles per bunch. This paper reports on the operational capabilities of this instrument, improvements for its performance, and the methods utilized in data analysis. Examples of key measurements which illustrate the instrument s performance are presented. This device demonstrates measurement capabilities applicable to future high energy physics accelerators and light sources. INTRODUCTION The X-ray Beam Size Monitor (xbsm) provides experimenters in the CESRTA program with the ability to measure the vertical beam size of individual particle bunches on a turn-by-turn, single pass basis. At present two xbsm instruments have been installed in experimental areas of the Cornell High Energy Synchrotron Source (CHESS). One is used for positrons and the other is used for electrons. Each setup has its own x-ray source, which is a dipole magnet within the Cornell Electron Storage Ring (CESR). The critical energy is 0.6 kev during 2 GeV CESRTA operations. A set of invacuum optics focuses the photon flux onto the detector. The geometry of the beam line provides an optical magnification of 2.34 for the positron line and 2.52 for the electron line. A continuous vacuum vessel, containing optics elements and filters which are inserted into the x- ray beam, extends from the x-ray source to the detector. Images are collected via a custom data acquisition system. Figure 1 shows the functional layout of the positron line. The electron line is a near mirror image. *Work supported by NSF grant PHY-0734867, PHY-1002467 and DOE grant DE-FC02-08ER41538, DE-SC0006505 Figure 1: Layout of the xbsm positron line VACUUM A multi-stage differential pumping scheme has been implemented to allow windowless transmission path from the x-ray source to the detector. The scheme features a series of apertures, turbo pumps and gate valves along the x-ray beam line. This setup allows us to have electronics (printed circuit board, cables, etc) which are not vacuum compatible within the detector box without contaminating CESR. This is critical for the proper operation of the instrument. The vacuum system pressure is automatically controlled and monitored via programmable logic controllers. DETECTOR The detector is a vertical linear array of 32 InGaAs diodes with a 50 µm pitch and horizontal width of 400 µm. The InGaAs layer is 3.5 µm thick, and absorbs 73% of photons at 2.5 kev. The time response of the detector is sub-nanosecond. OPTICS The xbsm utilizes four different optical elements. For all beam energies, a vertically limiting slit (referred to as a pinhole) is available. For beam energies less than 2.5 GeV, a low energy Fresnel zone plate and a coded aperture are available. At 4 GeV and greater, a high power coded aperture can be inserted into the x-ray beam. The low energy optics are contained on one chip which is made from a 2.5 µm silicon substrate with a 0.7 µm layer of gold forming the optical features. The high energy coded aperture is made from a 625 µm silicon substrate with a 10 µm layer of gold. The coded aperture features of the high energy optics are the same but one half the scale of the low energy optics. Figure 2 shows the coded aperture features. 585

Proceedings of IBIC2012, Tsukuba, Japan Figure 2: Coded aperture. The low energy coded aperture provides a series of transmitting and opaque features which range from 10 µm to 40 µm. Analysis of the coded aperture interference pattern yields information about both beam size and photon energy spectrum. Due to the variation in feature size, the coded aperture is sensitive to a range of x-ray energies and very small beam sizes. This, coupled with the relatively high level of transparency, allows the codedd aperture to be useful for low energy and low photon count experiments. Analysis of the Fresnel zone plate has not been fully developed. It has proven to have a very broad background component when used with a wide spectrum x-ray beam, and the introduction of a monochromato or limits the number of photons to an unusable level. The pinhole consists of a vertically limiting set of adjustable tungsten blades which allow for a configurable slit height. This allows the height to be optimized for different particle beam energies. FILTERS/SLITS In order to empirically investigate and control the detector response, we have created a movable stage whichh contains a variety of horizontally limiting slits and filters. These slits and filters are inserted into the x-ray beam between the optics and the detector. The horizontally limiting slits are useful in reducing the amount of flux during high energy and highh current operation. Slit sizes from 35 µm to 171 µm have been tested. The filters are used to change the spectral content of the x-ray beam which reaches the detector, and thus derive the energy spectrum based on the response. We have tested 6 µm diamond and 2 µm molybdenum filters and have plans to test a 6 µm aluminium filter. DATA ACQUISITION The xbsm instrument has an independent data acquisition channel for each of the 32 detector diodes. Each channel consists of a transimpedance amplifier, a variable gain amplifier (-4 db to 20 db) ), a 12 bit 3000 MSPS analog to digital converter, a field programmable gate array (FPGA), a local 1 Msample SRAM buffer and a programmable sample clock delay. Figure 3 shows a single channel. 586 Figure 3: Data acquisition channel Each off these data acquisition channels is synchronized to the storage ring via a 24 MHz instrumentation clock. This clockk is derived from the CESR master oscillator and contains encoded triggers and turns markers. A local timing board receives this 24 MHz clock and utilizes a voltage controlled oscillator, phase detection circuit and programmable delays to generate the sample clocks for all 32 channels. These individual sample clocks have adjustablee delays in 10 ps increments to allow for fine phase adjustments. This timing control capability is used to peak sample the signal generated when x-rays are absorbed by the diode detector. Figure 4 shows the signals generated by two 4 ns spaced bunches and their corresponding sample points. Figure 4: Peak sampling of 4 ns spaced bunches. A single channel samples each of these bunches when clocked at 250 MHz with the sampling clock edges aligned with the peak of the signal. Note that the image shown is a composite made by shifting the sample point in 10 ps increments. All 32 channels are synchronized with the CESR turn marker and gated with a bunch pattern which matches the CESR filll pattern. The peak value detected on all 32 channels is collected and used to form a detector image.

SOFTWARE A Matlab based graphical user interface is used to control the data acquisition and motorized features of the instrument. All data is written to disk and is available to both real time and offline analysis programs. The control program has limited analysis functionality that provides fitting and display of a sample of the turn by turn beam size measurements in real time and allows for the possibility of real time particle beam tuning. Offline analysis is performed using the xbsmrootfitter. This locally developed, C++ based analysis software uses the CERN ROOT package to book, store, fit and display histograms. It leverages MINUIT for fitting and FFTW for frequency analysis. It provides a variety of plots and analysis. xbsmrootfitter provides systematic compensations such as pedestal subtraction, channel by channel calibrations, gain range scaling, bunch to bunch crosstalk compensation and bad channel suppression. PINHOLE IMAGE ANALYSIS The pinhole is a simple optical device which has a single slit and an opaque masking material. In order to generate a model for the image we apply the derived x- ray spectrum to a numerical calculation of Fraunhofer diffraction. The broad energy spectrum smooths out the outer diffraction features and allows us to approximate the image as a sum of two gaussians plus a flat background. Figure 5 shows the model (line with points) and the two gaussian function (solid red line). Figure 5: Pinhole model and function. This function is used to fit the detector image data and calculate a beam size and image offset on the detector. Figure 6 shows a single bunch, single turn detector image with a fit applied. Figure 6: Pinhole detector image with fit. CODED APERTURE IMAGE ANALYSIS The coded aperture iss a multi-slit optical device which generates a diffraction pattern at the detector. This diffractionn pattern is calculable using the derived x-ray spectrum and the detailed geometry and materials of the coded aperture. We parameterize the diffraction pattern as the sum of 12 gaussians. Figure 7 shows the model smoothedd to a source size of 7 µm (line with dots), the 12 gaussians (various colors, solid lines), and the resulting function (single solid red line). Figure 7: Code aperture model and function. This function is used to fit the detector image data and calculate a beam size and image offset on the detector. Figure 8 shows a single bunch, single turn, detector image with a fitt applied whichh corresponds to a source size of 9.85 µm. 587

Proceedings of IBIC2012, Tsukuba, Japan Figure 11: Image Offset Over 1024 turns. Fig 8: Coded aperture detector image with fit. SIZE AND OFFSET MEASUREMENTS The xbsm is capable of bunch by bunch, turn by turn, single shot measurements of beam size and offset of the image on the detector. Figures 9 through 12 are representative measurements of a single bunch of positrons in CESR. The size of a single bunch is plotted over 1024 turns along with the frequency spectrum of the size variations. Figure 12: Spectrum of image offset over 1024 turns Figure 9: Single bunch size over 1024 turns. Figure 10: Frequency spectrum of bunch size of a single bunch. The offset of the image on the detector can also be plotted, giving the experimenter beam motion information. The peak in the spectrum of offsets in Figure 12 appears at the vertical betatron tune. 588 PHYSICS MEASUREMENTS The xbsm is in regular use during CESRTA experimental runs. The following are two examples of experiments which have collected and analysed xbsm data. Thee physics phenomena which are evident in these plots are outside the scope of this paper. Work is on- going to decouple systematic instrumentation effects from beam physics. Intra beam scattering (IBS) is a phenomenon of current dependentnt beam size in small beams. The xbsm is used to measure the beam size and detector offset of individual bunches versus beam current over many turns. Figure 13 representss an IBS experiment using a single bunch of positrons undergoing decay from 10 11 particles to 10 9 particles. Data sets were collected under identical conditionss using the pinhole and codedd aperture. Figure 13: Positron IBS experiment beam sizes versus current.

During the IBS experiments, the image offset is also analysed. Figure 14 is an image offset spectrum plot for a similar current decay as above. A current dependent tune shift can be observed moving from approximately 210 khz to 235 khz. Figure 16: Bunch slicing. Figure 14: IBS image offset spectrum. Experiments in the beam dynamics program study electron induced multi-bunch phenomenonn in CESR. The xbsm is used to measure the beam size and detector offset for individual bunches along trains of bunches over many turns. The spacing between bunches can be varied in 4 ns increments while the bunch current is held constant at approximately 1 ma. Figure 15 shows a comparison of different bunch spacing and the effect of the electron cloud on the bunch size along the train. Figure 15: Beam size along a train of bunches. FUTURE MEASUREMENTS One future extension of the xbsm concept involves measuring the instabilities and head-tail oscillations of a single bunch. Figure 16 demonstrates a possible bunch oscillation and four slices of the bunch size. This measurement will utilize high speed diode detectors with rise and fall times on the order of 35 ps. This response time coupled with a four channel phased digitizer will allow forr four longitudinal slices of the bunch to be measured. SUMMARY The xbsm is routinely used in the CESRTA experimental program for precision measurement of beams with vertical sizee as small as 9 µm. There is an on- to going effort to understand systematic effects and optimize the analysis of images. The device has proven essential to the CESRTAA program to understand intrabeam scatteringg and the emittance diluting effect of the electron cloud. REFERENCES [1] J.P. Alexander et al, TH5RFP026, PAC09 [2] J.P. Alexander et al, TH5RFP027, PAC09 [3] J.W. Flanagan et al, TH5RFP048, PAC09 [4] D. P. Peterson et al, MOPE090, PAC10 [5] J.W. Flanagan et al, MOPE007, PAC10 [6] N.T. Rider et al, MOP304, PAC11 589