Use of the HSW5 Spinning Disk Confocal Microscope Updated last May 25, 2010 OK

Similar documents
Widefield 1. Switching on

Simplified Instructions: Olympus Widefield Microscope S1230

Overview. About other software. Administrator password. 58. UltraVIEW VoX Getting Started Guide

LSM 510 Training Notes

Standard Operating Procedure (SOP) for Shared Equipment: Spinning Disk Confocal Microscope

LSM 710 Confocal Microscope Standard Operation Protocol

Zeiss LSM 780 Protocol

Guide to Confocal 5. Starting session

LSM 510 Meta Training Notes

Nikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol

Title: Leica SP5 Confocal User Manual

Leica SP8 Resonant Confocal. Quick-Start Guide

Training Guide for Leica SP8 Confocal/Multiphoton Microscope

Zeiss Axio Imager.A1 manual

Zeiss Axiovert 135 Fluorescence Microscope Quick Guide / Operations Manual (v. 1.0 February 09)

LSM 780 Confocal Microscope Standard Operation Protocol

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.

Practical work no. 3: Confocal Live Cell Microscopy

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging

Nasmyth Ultraview Vox User Protocol

Zeiss LSM 880 Protocol

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope

Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software

Nikon A1R. Multi-Photon & Laser Scanning Confocal Microscope. Kyle Marchuk Adam Fries Jordan Briscoe Kaitlin Corbin. April 2017.

Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope

Training Guide for Carl Zeiss LSM 510 META Confocal Microscope

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s

Zeiss 780 Training Notes

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017.

Operating Instructions for Zeiss LSM 510

Renishaw InVia Raman microscope

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling

Leica SP8 TCS Users Manual

Zeiss 880 Training Notes Zen 2.3

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

User manual for Olympus SD-OSR spinning disk confocal microscope

ZEISS LSM510META confocal manual

Nikon C1si Spectral Laser Scanning Confocal Microscope. User Guide

Microscopy from Carl Zeiss

LEICA TCS SP5 AOBS TANDEM USER MANUAL

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL

QUICKSTART GUIDE: WIDEFIELD HWF1 Zeiss Cell Observer Live Cell Imaging System (HAMMERSMITH, L BLOCK, ROOM 314) Imperial College London

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014

Swept-Field User Guide

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL

Contents. Introduction

Things to check before start-up.

TRAINING MANUAL. Olympus FV1000

SPINNING DISK CSU-X1 USER MANUAL

LSM 800 Confocal Microscope Standard Operation Protocol

Zeiss AxioImager.Z2 Brightfield Protocol

Leica Sp5 II Confocal User Guide

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Olympus Digital Microscope Camera (DP70) checklist

3. are adherent cells (ie. cells in suspension are too far away from the coverslip)

Leica SPEII confocal microscope. Short Manual

Leica TCS SP8 Quick Start Guide

Confocal Raman Microscopy (WITec Alpha 300R)

EPIFLUORESCENCE &/OR BRIGHTFIELD MICROSCOPY

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Olympus LEXT OLS 4000 Confocal Laser Microscope

Quick Start Guide. Leica SP5 X

OPERATING INSTRUCTIONS

ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide

Nikon. King s College London. Imaging Centre. N-SIM guide NIKON IMAGING KING S COLLEGE LONDON

WITec Alpha 300R Quick Operation Summary October 2018

b. Turn the power switch and key to on position for blue laser.

Olympus xcellence Software - basic user guide

REMEMBER: You have 5GB of disk space on this microscope. Check before you start if you have room for your experiment. If not delete your old data.

Nikon SIM-E & A1-R System

Brightfield Microscopy and Image Acquisition on Spotcam1. by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer

Cell Biology and Bioimaging Core

Before you start, make sure that you have a properly calibrated system to obtain high-quality images.

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software:

Leica TCS SP8 Quick Start Guide

Leica SP8 TCS Users Manual

Image-Pro Plus 7.0 Product Note

QUICKSTART GUIDE: WIDEFIELD WF3 Zeiss Cell Observer Live Cell Imaging System (SAF, ROOM 409) Imperial College London

Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope

Characterization Microscope Nikon LV150

RENISHAW INVIA RAMAN SPECTROMETER

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7

Olympus Time-lapse Microscope Basic operation

Horiba Jobin-Yvon LabRam Raman Confocal Microscope (GERB 120)

Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope

Training Guide for Carl Zeiss LSM 880 with AiryScan FAST

OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop

Zeiss LSM 510 Multiphoton Confocal Microscope

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO

Using Autofocus in NIS-Elements

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING

User Guide to the IBIF Leica TCS SP8 MP Confocal Microscope

Standard Operating Procedure of Atomic Force Microscope (Anasys afm+)

Internal Medicine Imaging Core Emory University Department of Medicine

Title: Nikon A1R Confocal User Manual

KEYENCE VKX LASER-SCANNING CONFOCAL MICROSCOPE Standard Operating Procedures (updated Oct 2017)

NIS-Elements: Grid to ND Set Up Interface

SOP: Polyvar Met Light Microscope

Transcription:

Use of the HSW5 Spinning Disk Confocal Microscope Updated last May 25, 2010 OK Getting Started: 2 Starting Micromanager and Loading a Configuration 3 The Main Micromanager GUI 3 Configuration Settings 3 LIVE BUTTON 4 Multi D Acq. 4 Modifying settings for a given Wavelength. 4 Setting up a Timelapse in Micromanager Using Multi D Acq. 5 Timeseries 6 Multiposition XY (Stage) 6 Z Series 6 Current at Start 6 Channels (Wavelengths) 7 Setting up for Save 7 Display Last Frame 7 Acquire 7 Stop 7 Setting up a Survey in Micromanager Using Multi D Acq. Using the Stanford Photonics Camera: 7 When Finished: 8 Between Users 8 End of Day 8 Troubleshooting Appendix 1: Preset Configuration Appendix 2: Filter Configurations Appendix 3: Additional Resources 1

Getting Started: 1. Turn Lasers On. #1 405nm: Turn key to On to power up laser. #2 Argon (488nm/514nm) : Turn key on laser to On to power up laser. When you see blue emission in the laser tube, flip the switch on the circuit board out of standby mode. Note: Start with laser on standby. If power insufficient for imaging application switch out of standby. #3 561nm: Turn key to On to power up laser. #4 640nm: Press toggle to I to power up laser. 2. Turn the key on the front of the spinning disk all the way to Laser On. 3. Turn on the following in any order A. The power strip to the right of the scope. B. The AOTF controller. On shelf above scope (to the right of the 488/514 laser controller) C. The stage heat controller (silver box, green switch) D. The camera chiller E. The computer 4. The scope will now be setup with the output of the sample going to the eyepiece. remember that for actual collection, you will rotate the knob on the left to redirect light to the camera. 5. Drop a very small drop of oil onto the objective and mount your sample in the appropriate sample carrier. (The 40x objective is a default and is oil immersion other lenses are available and see Omar/Pete to use these). Open micromanager and choose Contrast from the Wavelength Group and click Live to open this shutter and find your sample (see The Micromanager GUI, Wavelength below). Note about objectives: DO NOT add oil to any objective with a black top (generally 10x/20x). This means air objective for these particular objectives. 6. Turn on the camera temperature controller/power supply (black box above camera, on shelf) and Stanford Photonics camera controller (black box to the left of the scope) only AFTER camera chiller is down to temperature (about 5 10mins after chiller has been turned on). If the monitor starts to alarm, do not turn on the camera. Turn off the temperature monitor/power supply, wait 5 more mins to let the chiller get down to temperature then turn on the monitor again. MAKE SURE THE STANFORD PHOTONICS CAMERA CONTROLLER IS SET TO STANDBY PRIOR TO TURING TOGGLE SWITCH TO I (ON). 2

Starting Micromanager and Loading a Configuration 1. Double click Micromanager on the Desktop to open the application. This will also open Image J toolbar which is open throughout but is not necessary for any part of an acquisition. 2. A splashscreen will appear. Choose Default.cfg if you are a new user OR choose your configuration for established users (Note: Default.cfg will be located in the current version micromanager folder on the C: drive) The Main Micromanager GUI Opening the Micromanager GUI yields the Configuration Settings Screen as well as the Piper Control panel. Please do not adjust any of the settings in the Piper control panel unless you would like to use the bin controls. The Main GUI should look as follows: Configuration Settings (red arrow) Under Configuration Settings you should find several groups. They represent settings etc. of various hardware that you can access to define acquisition. Wavelength This is the definition of a wavelength combination excitation laser and its power, and emission filter choice. PIFOC Z (um) This always represents the current position of the PIFOC z motor in microns. It is a slider and will both change the current position and can be used during Live to set your z movement in acquisitions (aka Set in Multi D Acq.) 3

Camera Averaging This defines the number of frames averaged (or none) for the ICCD. This is important, particularly at high gain since the camera gets noisy. Ram Stacking This option allows the user to use frame addition to increase overall signal. Noise is also amplified with this setting. Users can turn this option On/Off and select how many frames to add. System The option houses the standard startup settings. Please do not adjust unless you are an advanced user. Live Button (green arrow) This opens the current shutter for the wavelength chosen in the Configuration Setting/Wavelength group (see above, when autoshutter is checked). It also opens a live window to view what that wavelength would look like in its current acquisition settings. Multi D Acq (yellow arrow) This is what you will click to setup multidimensional acquisitions. Modifying settings for a given Wavelength 1. Select the wavelength you plan to modify by clicking on the dropdown list that appears within the Wavelength row in Configuration Settings. 2. Click Edit / preset (purple arrow) (NOTE: Basic users should NOT edit groups ) 3. A window will appear showing you the details of that wavelength (see below). 4. For basic users, you will only edit the Laser Power setting. Typically you will do this in Live mode in order to get the intensity of illumination correct. Move the slider to the appropriate position. 5. To have variable taken, you MUST click outside of the box once to have the button set. To finish, click OK. 4

Setting up a Timelapse in Micromanager Using Multi D Acq. Click on the Multi D Acq button on the main micromanager page to open the Multi D Acq interface. It will look like this: There are subregions of the window that allow you to define a timelapse sequence. Each parameter defines a dimension of the acquisition. Use the settings defined below to set up the details of a timelapse. For Z Series and Channels, you probably want to click LIVE in the Micromanager main window to see the effects of your changes. Note, for changing illumination power specifically, you also will open the appropriate wavelength using Wavelength in the Configuration Settings area. 5

Timeseries Check this button to enable the use of the timeseries function. This will allow you to define the settings for the acquisition of images over time. Inputs in the Frames box allow the user to set the number of individual acquisitions taken (all channels are taken per frame) while inputs in the Interval box allow the user to determine the length of time between frame captures. Note: make sure that your interval time accounts for the time required to image your samples according to the exposure/averaging/ram stacking parameters (i.e. you should see/hear a period of shutter inactivity between intervals). Failure to do so will result in non linear timings of frame captures. Multiposition XY (Stage) Check this button to enable the function of the automated xy stage. Use the Prior stage controller (joystick) to manually move the stage from location to location in your sample and press the Mark button to save the coordinates into the acquisition manager. Z Series Checking this button enables the use of the PIFOC z axis controller. There are two modes for the operation of this device: Relative and Absolute. These are selectable from the dropdown menu at the botton of the Z Series box. Relative zeroes the PIFOC controller to the current position of the objective (set using coarse and fine focus knobs on the sides of the scope). From the current position the PIFOC will move up the amount (in um) set by the user in the Bottom and Top input boxes. This is performed by setting the Bottom to zero and the Top to the desired PIFOC height. Incremental movements can be adjusted using the appropriate input box. Absolute mode allows the user to adjust heights using the PIFOC controlling slider in the Configuration Settings window. The simplest method to use this function is to adjust the fine focus knob until the region of interest is just above the current focus, clicking the Set button adjacent to the Bottom input box, then adjusting the PIFOC height using the Configuration Settings PIFOC slider until the region of interest is just below the current focus and clicking the Set button adjacent to the Top input box. Above the Acquisition Summary box there is also a dropdown to select whether Slices or Wavelengths are acquired first. If Slices is selected, the acquisition will illuminate the sample with one wavelength, leaving the shutter open, and move the PIFOC the full range selected by the user after which, the next wavelength will be enabled and the process will repeat. Selecting Wavelengths moves the PIFOC only after all wavelengths selected by the user are taken. Current at Start For any channel that z is off, the acquisition will use the current z position when Acquire is pressed (at all subsequent timepoints). This means you should find the coarse focus for this single position and leave the knob and then use the PIFOC (in live mode, for each z channel) to define the bottom and top). 6

Channels (Wavelengths) Press the New button to add more wavelengths to the planned acquisition. Each wavelength has a Z Series enable/disable function, Exposure Length adjustability (in ms), and Color setting (this will assign a gradient table based on one color for the wavelength). As a note, when using the 1x1 and 2x2 bin modes the camera acquires in 33.33ms and 16.66ms intervals respectively. That is, if you input a 40ms exposure in a 1x1 bin the program will round to the nearest interval, in this case 33.33ms. Setting up for Save Check this box if you would like to save the files from the planned acquisition. User has the ability to select a save destination as well as name the acquisition files and add comments. Display IMPORTANT: This should ALWAYS be set to Last Frame from the dropdown menu if the user is setting up a timeseries acquisition. Failure to do so may fill the ImageJ memory buffer and terminate the acquisition prior to its completion. Acquire Press this button to begin acquisition. Stop Early Stop an acquisition by pressing the red X button in the image acquisition window or by pressing Stop below the Acquire button. Micromanager saves images as they are taken, thus if the save option was checked all files will be retained at the selected destination even if the acquisition is prematurely terminated. Using the Stanford Photonics Camera: There are essentially only three physical controls the standard user will be concerned with on the camera controller: the Power I/O toggle switch, the Camera On/Standby switch, and the Gain potentiometer. Powering On: Make sure the cooling unit (below laser table, to the left of the scope) has been on for at least 5 min prior to turning on temperature monitor/power supply (black box above camera on shelf). Turn on Camera in STANDBY mode after monitor is running. If the monitor starts to alarm, do not turn on the camera. Turn off the temperature monitor, wait 5 more mins to let the chiller get down to temperature then turn on the monitor again. Switch the camera control to enable to stream images to the camera. 7

Adjusting Gain: Adjust the camera gain using the potentiometer on the left side of the box. Do not adjust the potentiometer on the right or toggle any other switches. Signal to noise ratio is highest between 0.800V 1.350V. DO NOT EXCEED 1.400V. Optimizing image quality: This camera/software system s ability to detect images is limited by the following parameters: 1) Gain set by potentiometer Gain is physically controlled and thus applied to every wavelength at the selected voltage. 2) RAM Stacking/Averaging of frames Frame modifiers are controlled by the configuration settings window and are thus applied to every wavelength as well. The best method to acquire images is to set the Gain potentiometer and RAM Stack/Averaging values to fit the dimmest channel. Then adjust the exposure rates of the other channels and their laser voltages to fit the other channels. Tinkering with these values will be a must for this setup as sometimes a sample will require a combination of additive (RAM) and averaged frames to reduce noise and maximize signal as well as laser power corrections and exposure experimentation to minimize laser illumination at the sample. With that said, please allocate an appropriate amount of time (30mins) to adjust the settings to your liking. When Finished: Between Users: 1. Clean oil off the objective, and surrounding surfaces. Failure to do so can damage objective springs. 2. Shut down Micromanager Application and save data to CD/server/hard drive. 3. Do not shut off lasers if there are users using the scope after you (please check calendar). End of Day: 1. Clean oil off of the objective, and surrounding surfaces. Failure to do so can damage objective springs. 2. Turn off the electronics in reverse order of turning on lasers off LAST. Laser fan will continue to cool for about five minutes and lights will stay on on the front panel of the laser power supply these are on all the time. 8