LSM 710 Confocal Microscope Standard Operation Protocol

Similar documents
LSM 780 Confocal Microscope Standard Operation Protocol

LSM 800 Confocal Microscope Standard Operation Protocol

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Zeiss LSM 780 Protocol

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL

Zeiss LSM 880 Protocol

Zeiss 880 Training Notes Zen 2.3

Operating Instructions for Zeiss LSM 510

Leica SP8 TCS Users Manual

Microscopy from Carl Zeiss

Things to check before start-up.

Zeiss 780 Training Notes

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s

Guide to Confocal 5. Starting session

Leica SP8 TCS Users Manual

ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL

Contents. Introduction

Leica SPEII confocal microscope. Short Manual

Training Guide for Carl Zeiss LSM 510 META Confocal Microscope

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL

ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide

ZEISS LSM510META confocal manual

Quick Start Guide. Leica SP5 X

LSM 510 Training Notes

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope

Zeiss LSM880 Operating Instructions. UTMB Optical Microscopy Core Jan. 16, 2018

OPERATING INSTRUCTIONS

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling

LSM 510 Meta Training Notes

Cell Biology and Bioimaging Core

Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope

Training Guide for Leica SP8 Confocal/Multiphoton Microscope

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.

Title: Leica SP5 Confocal User Manual

Leica Sp5 II Confocal User Guide

LEICA TCS SP5 AOBS TANDEM USER MANUAL

TRAINING MANUAL. Olympus FV1000

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Nikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software:

REMEMBER: You have 5GB of disk space on this microscope. Check before you start if you have room for your experiment. If not delete your old data.

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014

Topics. - How to calibrate the LSM scanner. - How to clean the microscope. - How to adjust the pinhole alignment. - How to adjust the Collimator

Practical work no. 3: Confocal Live Cell Microscopy

Internal Medicine Imaging Core Emory University Department of Medicine

Zeiss LSM 510 Multiphoton Confocal Microscope

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope

Zeiss LSM 510 Multiphoton Confocal Microscope

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging

Training Guide for Carl Zeiss LSM 880 with AiryScan FAST

Leica TCS SP8 Quick Start Guide

Microscope Confocal LSM510 META

Leica TCS SP8 Quick Start Guide

The Zeiss AiryScan System, Confocal Four.

Nikon SIM-E & A1-R System

Leica TCS SL Confocal Training. Neuroscience Imaging Core Staff. Core Director. Facility Manager

Zeiss Deconvolution Microscope: A Quick Guide

Nikon A1R. Multi-Photon & Laser Scanning Confocal Microscope. Kyle Marchuk Adam Fries Jordan Briscoe Kaitlin Corbin. April 2017.

CONFOCAL MICROSCOPE (Zeiss LSM 510 META v4.2)

b. Turn the power switch and key to on position for blue laser.

Zeiss AxioImager.Z2 Brightfield Protocol

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each

Supplemental Method Information Zeiss LSM710

Leica SP8 Resonant Confocal. Quick-Start Guide

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO

Title: Nikon A1R Confocal User Manual

Instructions for the Leica SP5 II laser scanning confocal microscope

Axioscan - Startup. 1. Turn on the Axioscan (button to the left) and turn on the computer. 2. Log on and start the ZEN Blue software from the desktop

Zeiss LSM 510 META Guide

3. are adherent cells (ie. cells in suspension are too far away from the coverslip)

Nikon C1si Spectral Laser Scanning Confocal Microscope. User Guide

Overview. About other software. Administrator password. 58. UltraVIEW VoX Getting Started Guide

Zeiss Axio Imager.A1 manual

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017.

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions

User Guide to the IBIF Leica TCS SP8 MP Confocal Microscope

ZEN 2012 SP5 black edition Hotfix 12

Simplified Instructions: Olympus Widefield Microscope S1230

1 Co Localization and Working flow with the lsm700

Microscope Confocal Sp2 Upright.

Usermanual for Leica SP8 confocal

Zeiss Axiovert 135 Fluorescence Microscope Quick Guide / Operations Manual (v. 1.0 February 09)

Using the Nikon TE2000 Inverted Microscope

SPINNING DISK CSU-X1 USER MANUAL

3 Choose the Channels button and set the Channel Settings. Set the Pinhole to 1 Airy unit.

Nikon. King s College London. Imaging Centre. N-SIM guide NIKON IMAGING KING S COLLEGE LONDON

1 Set up the confocal light path for imaging a green dye (Alexa488-EGFP). For example, the

Boulevard du Temple Daguerrotype (Paris,1838) a busy street? Nyquist sampling for movement

Use of the HSW5 Spinning Disk Confocal Microscope Updated last May 25, 2010 OK

Nikon A1Rsi Confocal Start-Up Sequence

Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope

Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7

Multifluorescence The Crosstalk Problem and Its Solution

RENISHAW INVIA RAMAN SPECTROMETER

OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING

Confocal Laser Scanning Microscopy

Imaging Beyond the Basics: Optimizing Settings on the Leica SP8 Confocal

Transcription:

LSM 710 Confocal Microscope Standard Operation Protocol Basic Operation Turning on the system 1. Switch on Main power switch 2. Switch on System / PC power button 3. Switch on Components power button 4. Switch on X-Cite fluorescent lamp 5. Sign on log sheet (and record the Fluoro Start time from X-Cite lamp) 6. (Skip this step if 458/488/514nm (yellow to green) excitation is not needed) Turn on Standby/On key switch for argon laser and Flip the Run-idle-switch to Run for argon laser 7. (Skip this step if two photons excitation is not needed) Turn On Standby/On key switch for MaiTai laser 8. (Please keep MaiTai laser off if you do not use it) 9. Turn On Computer 10. Green button on MP-285 stage controller If no yellow or green fluorescent needed For MaiTai laser (two photons) excitation only Starting the ZEN software 1. Log on computer with LSM User 2. Double click the ZEN 2011 icon on desktop 3. Choosing Start System for acquiring new images

Setting up the microscope and Locating the specimen 1. Use 10X objective first by turning the objective turret manually (wait several seconds for the automatic objective position adjustment) and slightly lower the stage by using coarse focus adjustment knob. 2. Put the sample on the mount on the stage (coverslip face up) 3. Move the area to be observed right under the objective lens (in X direction by moving the sample mount or stage controller and in Y direction by the stage controller) 4. Turn the shift knob to move the objective into WORK position. 5. In the Zen software, click on the Locate tab to access the microscope controls and press the Online button click on DAPI tab for blue emission channel observation click on GFP tab for green emission channel observation click on DsRed tab for red emission channel observation click on BF tab for transmitted light observation 6. Focus with coarse and then fine adjustment knob 7. Move the desired area to the center of view for further magnification using higher power objectives or for later image acquisition 8. Select a suitable high magnification objective by turning the objective turret manually (wait several seconds for the automatic objective position adjustment). Add corresponding refracting medium needed and do not move the stage in Z-direction if you have already used 10 X objective to focus. Note: Please make sure there is no oil and water on your slide when you are using any air objective and no oil on your slide when you are using water objective. Any inappropriate medium will damage these objectives) 9. Turn the shift knob slowly and carefully to lower the objective into the WORK position. Focus with the fine adjustment knob only. Note: Please be careful when you are focusing the sample. You may damage the objective and break your sample you keep moving objective down! 10. Check all the emission signals from different fluorophores in your sample. Center the view of interest and then click Offline in the Locate Panel to turn off all the light and go to Acquisition tab.

Setting up the Scanning Track/Channel Method 1 (only applied to single photon laser) 1. In Acquisition tab, click on the Smart Setup button 2. Click on the Arrow in the Dye list and choose the fluorophore in the specimen, add one by one down the rows 3. Choose the Pseudo Color according to you need 4. In Proposal, choose Fastest (simultaneous scanning) for fast live image acquisition but with (if any) excitation/ emission crosstalk among your fluorophore, or; Best signal (sequential scanning) to minimize crosstalk among each fluorophore but slow image acquisition or; Best Compromise (combined mode scanning) to simultaneously scan noncross talking fluorophore to speed up while other cross talking fluorophore are scan sequentially when necessarily Fastest Best signal Best Compromise 5. Press Apply Method 2 Load a predefined configuration of combination of dyes from the Load configuration list box. Method 3 Open any previous image with desire setting and click Reuse button in Dimension panel. Method 4 Setting up the configuration manually

Optimizing scanning parameters 1. Select one Tracks in the Channels tool 2. Select Live for continuous fast scanning 3. Activate Range Indicator. The scanned image appears in a false-color presentation. Red pixel = saturation (maximum intensity). Blue pixel = zero (minimum intensity). 4. Set the Pinhole to 1 AU (Airy Unit). For colocalization studies, adjust the pinhole of each channel to the same Optical Slice Thickness. 5. To get optimal intensity and background signal, adjust a. Increase Detector Gain (Maximum of 800) until a few red pixel (indicating saturation) appear in the image; b. Increase the Laser Power (too high cause bleaching), if increasing detector gain cannot achieve the saturation or desired intensity; c. Fine-tune the focus with the fine adjustment knob to the brightest or preferred z- position. And then adjust the detector gain and laser power to optimize the signal intensity d. Decrease the Digital Offset to reduce background signal until the desired background region is filled with blue pixel e. Increase Digital Gain to increase signal amplification if necessary. 6. Stop the Live scan process and uncheck the Range Indicator. 7. Repeat step 1-6 for other tracks. 8. Check all the channel boxes when finish adjusting. 1 5b 4 5a 5d 5e 2 3 6

Setting the scanning parameters 1. Select the Acquisition Mode tab in the middle column. 2. Select a suitable Area and Zoom Power to capture image of interest (or use Crop function in Dimension under the image container after a Live image). 3. Choose a Frame Size (change with Pixel size, thus resolution). 1024x1024 usually produces good results for general propose. Click on the Optimal button for best resolution which depending on objective N.A. and λ. 4. Adjust the scan speed with the Speed slider. 6 or 7 usually produce good results. For samples with high background noise, choose Speed 4 or 5. 5. Select the number of Average. Averaging (usually in 2 or 4) improves the image quality by increasing the signal-to-noise ratio but it also increases the scan time. 6. Select the dynamic range of 8 or 12 or 16 Bit (per pixel) in the Bit Depth. 12 or 16 Bit is recommended when doing quantitative measurements or other post-analysis. 7. Other setting can be kept as defaults (as in the right) which will be good enough for general propose. 8. Click Snap to acquire a single frame (multi-channel) image. Lower scan speed; or More averaging Larger frame size (Higher resolution)

Storing and exporting image data 1. To save the an acquired or processed images, click on that image to highlight it and then click on the Save Data button in the Images and Documents panel in the right or Click Save or Save As button in File menu. Warning Icon indicated unsaved or modified image 2. Create or choose your own folder in D:/user, enter a file name, select format as LSM 5 and click on the Save button 3. After saving the raw data as LSM5 format, Image can be export as various Image format and series images can be export as video. Go to File > Export, and choose the Format and Data as followed. Adjust Frames per seconds if necessary. Click Select file name and save and save in your own folder. Single Frame Series (Gallery of images; Time series movie or 3D animation) Format Date Format Date Tagged Image File JPEG File Interchange Format Portable Network Graphics Raw data single plane (without any overlay graphics) Contents of image window single plane (with overlay; as displayed as in image container; compressed) Full resolution image window single plane (with overlay; as displayed as in image container; uncompressed) Video of Window Apple Quick Time Raw data series (without any overlay graphics) Contents of image window series (with overlay; as displayed as in image container; compressed) Full resolution image window series (with overlay; as displayed as in image container; uncompressed)

Scanning a Z-Stack 1. Check the Box for Z-Stack in the main tools area. 2. Open the Z-Stack tool panel by clicking on the Z-stack tab. 3. Choose a channel that have signal throughout the interested volume and then Click on Live. 1 3 & 6 4. Use the focus knob to locate one end of the specimen and click on the Set First button. 5. Then use the focus to locate another end of the specimen and click on the Set Last button 2 4 9 6. Stop the Live. Select all the channel needed for the Z-stack and then click the arrow for Optimize sectioning and Step. 7. Set the Z-stack Interval between each frame, users are suggested to set the interval less than the thickness of 1 Optical Section of the channel of the shortest wavelength (so that you can sample all the space in your specimen). Optimal value is calculated according to the emission wavelength, objective lens, and the pinhole diameter. Using the Optimal thickness mean double sampling (interval = ½ thickness of 1 optical section) and thus provide the best 3D image reconstruction. 5 7 7 1 Optical Section 8. For colocalization studies, adjust the Pinhole of each channel in the Channels panel to the same Optical Slice thickness (so that each voxel contain data from the same volume of the specimen). 9. Click on the Start Experiment button to start the recording of the Z-Stack. 10. After saving the LSM 5 format, a Maximum Projection images can be made. Image in all frames will projected onto a single plane to visualize all the signal for the specimen in 2D Click Create image to generate the projection. 11. A video of rotating 3D image can also be made. Activate the Series tap, select the rotating axis, number of frame and the rotation range (= Total Frame Difference angle). Click Apply and the series images will be generated in a new tab. Go to File > Export, choose video for Windows and then choose a suitable series Adjust the playing speed by Frame per Second and save in your own folder. 8 10 11

Setting up Time Series Experiment Check the Box for Time Series in the main tools area. 1. Open the Time Series tool panel by clicking on the Time Series tab. 2. Set the number of Cycle and time Interval between each frame. (The scanning of each frame is included in the countdown of the Interval, therefore Interval time should scanning time of one multi color frame.) 3. Set the channel and acquisition parameter if necessary and then click Start Experiment. 4. To quantifying changes in signal intensity after acquisition, click Mean ROI tab and create a ROI region with the drawing tool for you region of interest. 5. The intensity profile along the experiment duration will be shown in the graph while corresponding data will be shown in the table below. 6. After saving the LSM 5 format, video for the time series images can be exported. Go to File > Export, choose video for Windows and then a suitable series. Adjust the playing speed by Frame per Second and Select file name and save in your own folder.

Turning off the system Please check if the equipment will be used / booked by other users. Please switch off the system if no one is using within two sessions (1 hr) afterward. 1. Turn off MaiTai laser in software (if it is ON) 2. Exiting ZEN Software, transfer data through the Faculty Core Facility network storage server and then Shut down PC 3. Turn key switch 90 anti-clockwise for MaiTai (2 photon) laser to Standby 4. Flip the toggle of Run-idle-switch of argon laser to Idle 5. Turn key switch 90 anti-clockwise for argon laser to Standby Wait for 10 min to let Laser Fan OFF!!! While waiting, please Remove your sample and clean the stage Clean oil objective (with lens cleaning tissue only but NOT Kimwipe) 1. Remove residue oil from the objective lens by a dry lens cleaning tissue 2. Clean the lens with a new lens cleaning tissue with 100% absolute ethanol 3. Dry the lens with another new lens cleaning tissue Clean water objective (with lens cleaning tissue soaked with autoclaved Milli-Q water) Change objective to 10x; Record the fluorescent lamp hour and Sign on log sheet according to Actual usage time 6. Turn off X-Cite fluorescent lamp 7. Components button 8. System / PC button 9. Main power switch (Only after Argon Laser Fan OFF) 10. Compress Air (if necessary)