Amorphous Selenium Direct Radiography for Industrial Imaging

Similar documents
10/26/2015. Study Harder

10/3/2012. Study Harder

PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY

Digital Detector Array Image Quality for Various GOS Scintillators

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Film Replacement in Radiographic Weld Inspection The New ISO Standard

Flat Panel Detectors in Industrial Radiography

radiography detector

Digital radiography: Practical advantages of Digital Radiography. Practical Advantages in image quality

IQI-Sensitivity and Applications of Flat Panel Detectors and X-Ray Image Intensifiers A Comparison

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

Setting up digital imaging department!

AUTOMATED AND QUANTITATIVE METHOD FOR QUALITY ASSURANCE OF DIGITAL RADIOGRAPHY IMAGING SYSTEMS

Unit thickness. Unit area. σ = NΔX = ΔI / I 0

Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC

TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist

Current technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016

CR Basics and FAQ. Overview. Historical Perspective

DIGITAL IMAGE PROCESSING IN X-RAY IMAGING

SYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology

Equivalent Penetrameter Sensitivity (EPS) for Performance Evaluation of Computed Radiography Systems Muzibur Khan * and Mike Brothers

Digital Radiography for the Inspection of Small Defects

Moving from film to digital: A study of digital x-ray benefits, challenges and best practices

Radiographic sensitivity improved by optimized high resolution X -ray detector design.

REAL-TIME X-RAY IMAGE PROCESSING; TECHNIQUES FOR SENSITIVITY

Preliminary Modulation Transfer Function Study on Amorphous Silicon Flat Panel System for Industrial Digital Radiography

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development

Examination of Pipe Welds by Image Plate Based Computed Radiography System

X-ray light valve (XLV): a novel detectors technology for digital mammography*

Photon Counting and Energy Discriminating X-Ray Detectors - Benefits and Applications

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

HIGH SPEED, HIGH RESOLUTION AND LOW COST DIGITAL RADIOGRAPHY

Essential Parameters for the Visibility of IQIs and Small Indications in Digital Radiography

Features and Weaknesses of Phantoms for CR/DR System Testing

Advancements In Digital Radiography: CR, DR, and DICONDE

High Performance Amorphous Silicon Image Sensor for X-ray Diagnostic Medical Imaging Applications

Computed Radiography

17th World Conference on Nondestructive Testing, Oct 2008, Shanghai, China

Digital Imaging Considerations Computed Radiography

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance

Minimum Requirements for Digital Radiography Equipment and Measurement Procedures by Different Industries and Standard Organizations

Conversion to Digital Radiography from Film Radiography

Acquisition, Processing and Display

Shad-o-Box HS Product Family

Digital Radiographic Inspection replacing traditional RT and 3D RT Development

Ayaz Jhanorwala, Rishikesh Kumar, Satish Tilva. L & T MHPS Boilers Pvt Ltd., Hazira, Surat , India

TDI Imaging: An Efficient AOI and AXI Tool

LARGE FORMATTED AND HIGH RESOLUTION CMOS FLAT PANEL SENSORS FOR X-RAY

DIGITAL RADIOGRAPHY. Digital radiography is a film-less technology used to record radiographic images.

A ghost story: Spatio-temporal response characteristics of an indirect-detection flat-panel imager

Hardware for High Energy Applications 30 October 2009

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

Strategies to improve the signal and noise performance of active matrix, flat-panel imagers for diagnostic x-ray applications

NON-DESTRUCTIVE EVALUATION UTILIZING IMAGING PLATES FOR FIELD RADIOGRAPHY APPLICATIONS

Introduction. Chapter 1

RADIOGRAPHIC EXPOSURE

Tomographic 3D-Radiometry for the Visualisation and Measurement of the Defects of Girth Seams

Digital Radiography : Flat Panel

Seminar 8. Radiology S8 1

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT. Exam Content Specifications 8/22/2012 RADT 3463 COMPUTERIZED IMAGING

Medical Device Manufacturing: Designing for X-ray Inspection. Gil Zweig, President Glenbrook Technologies.

Experiences of users in Digital Radiography

Real Time Linear Array Imaging. Brian Caccamise

Small Animal Radiographic Techniques and Positioning COPYRIGHTED MATERIAL

Digital Industrial Radiography

Do you have any other questions? Please call us at (Toll Free) or , or

X-Ray Medical Imaging and Pixel detectors

Investigation of Effective DQE (edqe) parameters for a flat panel detector

Determination of the detective quantum efficiency of a prototype, megavoltage indirect detection, active matrix flat-panel imager

Recommended Training Curriculum For Digital Radiography Personnel (Level II)

APPLICATION OF THE DIGITAL RADIOGRAPHY IN WELD INSPECTION OF GAS AND OIL PIPELINES

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

Charged Coupled Device (CCD) S.Vidhya

PRACTICAL CONSIDERATIONS AND EFFECTS OF METALLIC SCREEN FLUORESCENCE AND BACKSCATTER CONTROL IN GAMMA COMPUTED RADIOGRAPHY

Performance evaluation of a photon counting detector for high energy NDT applications

X-ray Tube and Generator Basic principles and construction

Basis of Computed Radiography & PACS

Essentials of Digital Imaging

BASICS OF FLUOROSCOPY

3084 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 60, NO. 4, AUGUST 2013

Advances in High Energy X-ray Digital Detector Arrays

ON THE WAY TO DIGITAL RADIOGRAPHY

DALLA LUCE VISIBILE AI RAGGI X: NUOVI RIVELATORI DI IMMAGINI PER RAGGI X A DISCRIMINAZIONE IN ENERGIA ED APPLICAZIONI

A COMPARATIVE STUDY ON THE PERFORMANCE OF DIGITAL DETECTOR SYSTEMS FOR HIGH ENERGY APPLICATIONS

Pre Conference Workshop on Advanced Digital imaging

DIGITAL IMAGING Recognise the importance of quality assurance

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

Spokane Industries Computed Radiography Implementation Experience. David Jolin Technical Sales

X-RAY FLUOROSCOPY IMAGING SYSTEMS. Dr Slavik Tabakov. Luminescence: Dept. Medical Eng. & Physics King s College London

Performance of Image Intensifiers in Radiographic Systems

ISO INTERNATIONAL STANDARD

Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT

Marten Bosma 1, Alex Fauler 2, Michael Fiederle 2 en Jan Visser Nikhef, Amsterdam, The Netherlands 2. FMF, Freiburg, Germany

CXDI-70C WIRELESS SPECIFICATIONS

ACQUISITION HARDWARE FOR DIGITAL IMAGING

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE

Radiology. Radiograph: Is the image of an object made with use of X- ray instead of light.

QUANTITATIVE COMPUTERIZED LAMINOGRAPHY. Suzanne Fox Buchele and Hunter Ellinger

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Transcription:

DGZfP Proceedings BB 67-CD Paper 22 Computerized Tomography for Industrial Applications and Image Processing in Radiology March 15-17, 1999, Berlin, Germany Amorphous Selenium Direct Radiography for Industrial Imaging Peter K. Soltani, Daniel Wysnewski, Liberty Technologies, Inc., Conshohocken, PA, USA; Ken Swartz, Direct Radiography Corporation, Newark, DE, USA I. Introduction Filmless radiation imaging tools have been available for many years, primarily through the use of real time fluoroscopic (radioscopy) systems with simple analog video outputs. These systems are usually employed in high volume inspection applications where inspection speed is critical or when inspection automation is required. Fluoroscopic systems overcame their inherently poor contrast and spatial resolution through long integration times and the use of microfocus x-ray tubes to achieve projection magnification. The evolution of such systems led to the development of intensified CCD cameras, where the analog video outputs could be digitized, allowing for further analysis and enhancement. However, the intrinsic image quality of radioscopy systems has never matched that of conventional film, making them a poor choice for film replacement in most radiography applications. Over the past decade, several new digital radiography methods have emerged which can achieve sufficient levels of image quality to allow for film replacement in some radiographic imaging applications. These methods include what can be referred to as indirect digital radiography technologies, such as photostimulable storage phosphors employed in computed radiography (CR), phosphor and scintillator coupled amorphous silicon (a-si) arrays. These systems are characterized by the use of turbid media that convert radiation into visible light, which is, in turn, electronically detected and digitized to form an image. Because of optical scattering within the media, spatial blurring and increased noise can be encountered which degrade image quality as compared to film. However, these systems offer superior performance relative to conventional radioscopy systems, while exhibiting faster overall image display times as compared to film. More recently, our group has investigated and introduced a direct method for acquiring digital radiographs which eliminates optical blurring and reduces noise in the x-ray conversion layer. This approach employs an amorphous selenium (a-se) photoconductor deposited on top of an a-si thin film transistor (TFT) array. It has been demonstrated that this method produces image quality which exceeds indirect digital methods and is comparable to fine grain radiographic film. This paper will describe the particular structure and operating principle of this device, along with performance characteristics relevant to industrial imaging applications. II. Comparison of different radiography methods Before discussing the particulars of the a-se detector array, it is worth reviewing other digital filmless radiography methods, namely those that employ an indirect x-ray to digital conversion process. The different methods are summarized in Figure 1. DGZfP Proceedings BB 67-CD 123 Paper 22

CR systems have been in use for medical radiology since the early 1980 s and more recently for industrial imaging. This approach employs an imaging screen containing photostimulable storage phosphor crystals. The phosphor grains are coated on a flexible substrate and store the incident radiation energy in the form of trapped electrons - thus the name storage phosphor. The trapped electrons can be excited by laser stimulation to produce visible light from small pixel elements of the phosphor media. The light is detected electronically, digitized, and the digital signal level stored in computer memory. Once all available elements on the phosphor media have been stimulated and signal levels stored, the information is computed into a 2- dimensional image. As recent work has shown 1, CR systems are primarily limited by the blurring within the storage phosphor media and produce image quality comparable to coarse grain NDT radiographic films. Another method introduced over the past two years employs a fluorescent media, such as a medical phosphor intensifying screen or CsI crystal, to convert x-rays to light, which are then captured by an a-si photodiode array. As with CR, this process is primarily limited by the blurring within the phosphor or CsI, and though it will produce image quality somewhat better than CR systems, it still falls short of NDT film. The a-si arrays can exhibit a wide latitude as well as wide energy response, and are suitable for applications where lower image quality than medium speed NDT film is required. Figure 1. Comparison of indirect and direct digital radiography methods. DGZfP Proceedings BB 67-CD 124 Paper 22

III. a-se array description As shown in Figure 1, the primary difference between indirect methods and the a-se array is that x-rays are directly converted to charge in the latter. Due to the application of an electric field across the Se layer, it can also be seen that charge transport is constrained along vertical field lines, significantly reducing lateral information spread. This means that the detector array output closely matches the input radiation signal shape, with array resolution limited by the device geometry. The way in which the a-se array achieves the sharp input-output relationship results from its unique device structure 2,3, shown in Figures 2 and 3. The device employs a uniform and continuous a-se layer vapor deposited over a charge collection and field-effect transistor (FET) array forming a full 35 cm x 43 cm active area. The Se layer has a dielectric and top electrode layers applied to form a capacitor structure where electrons generated during exposure are collected under an applied bias. The charge collection and integration is accomplished with a 129 micron square charge collection (or so called mushroom ) electrode deposited over the FET array. The resulting pixel pitch is 139 microns, yielding a geometric fill factor of 86%; this is an important device parameter, which helps it achieve high charge collection efficiency, and thereby high image signal-to-noise ratios. The charge collection electrode is attached to a signal storage capacitor, which is in turn attached to the drain side of the FET. Figure 2. Cross section of the multi-layer a-se structure. DGZfP Proceedings BB 67-CD 125 Paper 22

Figure 3. Detail of pixel structure. Another key feature of the present device is that its structure, while producing a very high collection efficiency of the generated charge, prevents damage to the device at high exposure levels. As the device is exposed to radiation, the collection of charge reduces the applied bias. Charge collection efficiency is bias dependent, meaning that the electron-hole pair generation and separation efficiency is reduced. This prevents over charging of the signal storage capacitors to prevent damage to the structure at high exposures. IV. Image formation and readout As the array is exposed to ionizing radiation, electron-hole pairs are created within the a-se layer. The applied bias produces a 10V/micron field across the Se layer to separate the charge such that the holes are propagated to the charge collection electrode and the electrons collect at the top electrode interface. Additionally, the applied field prevents significant lateral spreading of the generated charge during exposure. Theoretical and experimental analysis of a-se layers under an applied field show that extremely high intrinsic spatial resolutions can be achieved 4. Unlike phosphor coupled a-si and CR systems whose resolution is phosphor limited, in the present case, the resolution is limited by the pixel geometry, not the a-se. Thus, very high modulation transfer function (MTF) would be expected in a-se arrays. During exposure, the FET s are in the off mode while charge is collected at the signal storage capacitor. At the end of the exposure, a positive pulse is applied to the FET gates starting with the first gate, G1. The data line, D1, connected to the FET source collects the charge from the drain, as shown in Figure 4. The signal charge is propagated to charge amplifiers connected to the data lines. The signals from the row are multiplexed and propagated to a 14 bit analog-to-digital converter and stored in computer memory. The readout continues row by row until all pixels have been read. DGZfP Proceedings BB 67-CD 126 Paper 22

Figure 4. Top view of device. At the end of the readout cycle, a charge erase cycle is applied to prepare the array for the next exposure. This removes any residual charge from the interfaces and any charge in deep trapping states of the a-se, thus preventing ghost images between successive exposures. In order to maximize the signal to noise, and thus image quality, the device also performs background corrections. The first involves creating a correction look up table (LUT) which stores a reference image of the detector array response to a flat x- ray exposure. This LUT stores the resulting image containing variations in x-ray source uniformity, as well as slight variations in array response at different energies and exposure levels. These variations are due to slight differences in the gain of each pixel and charge amplifier output. Generally, it has been found that a calibration LUT can be created once for a given set of exposure conditions, and updated if the exposure conditions change significantly. In addition to the LUT, the system also acquires dark field reference images before and after each exposure. Amorphous Se, like many semiconductors, generates a small dark current at room temperature. Since the amount of dark current will vary with time during and between exposures and array temperature, the acquisition of dark field images before and after exposure is employed to simulate the background noise during exposure. The background correction task is programmed into the device firmware and is performed automatically. The two reference images are averaged to estimate the magnitude of the background during the exposure and the exposure image is corrected. The result of this correction is to improve system contrast sensitivity and consistency between exposures. DGZfP Proceedings BB 67-CD 127 Paper 22

V. a-se array performance In order to evaluate the usefulness of the a-se device for non-destructive testing applications, the following performance characteristics were evaluated: - Exposure and energy response - MTF - Penetrameter sensitivity - Speed - Throughput Figures 5 and 6 show the response of the array to different metal thicknesses and energies at 150 cm source to detector distance and approximately 12 ma-second exposure. The results are presented in digital counts as a function of metal thickness at 50, 100, and 200 kvp. They indicate that the system produces a large change in response for the different metal thickness, with each digital count equal to approximately 2000 charges. Thus, the system can simultaneously achieve both wide latitude and density contrast. Data for energies up to 450 kvp have also been generated and reported in this conference s proceedings by Schneberk, et. al. 100000 Digital Counts 10000 1000 100 kv 50 kv 100 0 0.2 0.4 0.6 0.8 1 1.2 Al Step Thickness (inches) Figure 5. Aluminum step wedge. Characteristic curve generated at 150 cm source to detector distance, 24 ma-second exposure. DGZfP Proceedings BB 67-CD 128 Paper 22

100000 10000 Digital Counts 1000 200 kv 100 100 kv 10 0 0.2 0.4 0.6 0.8 1 1.2 SS Step Thickness (inches) Figure 6. Stainless steel step wedge. Characteristic curve generated at 150 cm source to detector distance, 24 ma-second exposure. The system MTF was determined by exposing the array to a 50 micron thick Pb foil at 100 kvp, 36 ma-seconds, and 150 cm source to detector distance; 0.5 mm of Pb filtration was also applied to the x-ray tube to attenuate its intensity. The edge response of the foil was determined by averaging 33 adjacent lines across the foil. The line spread functions (LSF) was calculated by taking the derivative of the edge response, as shown in Figure 7. The MTF was obtained by taking the Fourier transform of the LSF. Digital Counts 13000 12500 12000 11500 11000 10500 10000 0 0.5 1 1.5 2 2.5 Position (mm) dc/dx 1.6 1.4 1.2 1 0.8 0.6 0.4 0.2 0 0-0.2 0.5 1 1.5 2 2.5 Position (mm) Figure 7. Edge spread response and LSF for a-se array. In order to compare the MTF of the a-se array with other digital radiography methods, similar data were generated for our CR system, and published data obtained for amorphous silicon photodiode arrays. The results are shown in Figure 8. The data for a-si are from an array employing fine grain fluorescent phosphor, which is expected to exhibit near the highest MTF possible for indirect arrays 5. The data of Figure 8 clearly show that the a-se exhibits a very high contrast, as compared to CR DGZfP Proceedings BB 67-CD 129 Paper 22

and indirect arrays. The MTF data for a-se are much closer to that of NDT film than any of the other digital methods shown. These results confirm that both the spatial and contrast resolution of indirect methods are severely limited by the optical blurring which take place in the fluorescent screens, whereas a-se exhibits no optical blurring. 1 MTF 0.9 0.8 0.7 0.6 0.5 0.4 0.3 0.2 0.1 fluorescent phosphor storage phosphor a -Se array 0 0 1 2 3 4 5 frequency (Lp/mm) Figure 8. MTF for medium grain CR, fine grain medical fluorescent phosphor screen, and amorphous selenium array. In addition to MTF, the penetrameter sensitivity of the a-se array was also evaluated. This was done by exposing a 25 mm thick stainless steel bock to x-rays at 200 kvp with 1%, 2%, and 4% plaque type penetrameters. The exposure was varied and the resulting image optimized to evaluate the visibility of the penetrameter holes. The exposure at which various penetrameter holes could be seen was noted. These results are summarized in Figure 9 for a-se, along with data for CR, medical fluorescent phosphors employed with a-si indirect arrays, and NDT film 1. These show that the a-se array achieves penetrameter sensitivities comparable to fine grain NDT film, but at about 1/50 th the exposure. Furthermore, the data show that a-se exhibits superior performance relative to both CR and indirect arrays employing fluorescent phosphor screens. VI. Areas of application It is important to note that even though it is possible to achieve better than 2% penetrameter sensitivity with both CR and indirect arrays, these methods still fall short of the sensitivity achieved with conventional NDT film, as well the a-se array. It is also important to note that simply achieving a 2% penetrameter sensitivity does not necessarily mean that flaws can be detected with equal probability. In fact, penetrameters were not designed to provide a measure of flaw detection efficiency, but rather to ensure the quality of the radiographic procedure when using film. However, since film has intrinsically high resolution which exceeds what is needed in many imaging applications, parameters such as film MTF do not need to be considered. Unlike film, however, digital detectors do have a finite resolving DGZfP Proceedings BB 67-CD 130 Paper 22

capability, which may be on the order of the flaw size being investigated. Therefore, simply employing a 2% sensitivity to assess the performance of a digital system is not sufficient. The system MTF must be evaluated along with penetrameter sensitivity to assess the performance of different digital systems and their ability to detect flaws of a particular size and density. The MTF data of Figure 8 clearly show very significant differences in resolving capability between the various systems, which directly impacts the ability to detect flaws. Another interesting point is that the use of fluorescent screens which might otherwise improve exposure speed are typically forbidden for NDT use; the reason is the blurring which takes place, resulting in loss of both resolution and contrast needed to detect flaws. This means that care should be taken when employing indirect methods; only those applications where high spatial resolution and contrast sensitivity are not needed should be targeted for the use of indirect methods. Conversely, the a-se array offers a suitable choice for applications where high contrast sensitivity is needed, and resolution requirements do not significantly exceed the pixel pitch. 10 a -Se % EPS indirect digital NDT film 1 1 10 100 1000 10000 Relative Exposure Figure 9. Penetrameter sensitivity for different imaging methods; 200 kvp, 150 cm source to detector distance, 25 mm stainless steel. In summary, it has been found that though the pixel pitch of a-se direct conversion flat panel arrays is limited to 139 microns, the contrast sensitivity is very high and comparable to fine grain NDT film. Furthermore, it is found that because of the systems high signal-to-noise ratio, the system can detect the presence of flaws smaller than the pixel pitch as long as sufficient object contrast is produced. Overall, the a-se array offers significant performance improvements over indirect digital methods, making it useful for a wider range of film replacement applications. DGZfP Proceedings BB 67-CD 131 Paper 22

VII. References 1. R. Kochakian, B. Vassen, and P. Willems, Application Limitations for Digital Radiography, Proceedings of the 1 st Pan-American Conference, September 14-18, 1998, pp. 15. 2. D. L. Lee, L. K. Cheung, E. F. Palecki, and L. S. Jeromin, A Discussion of the Resolution and Dynamic Range of Se-TFT Direct Digital Radiographic Detector, SPIE Vol. 2708, Physics of Medical Imaging, pp. 511, 1996. 3. D. L. Lee, L. K. Cheung, L. S. Jeromin, A New Digital Detector for Projection Radiography, SPIE Vol. 2432, Physics of Medical Imaging, pp. 237, 1995. 4. W. Que and J. A. Rowlands, X-ray Imaging Using Amorphous Selenium: Inherent Spatial Resolution, Med. Phys. 22, 364-374 (1995). 5. R. L. Weisfield, R. A. Street, R. Apte, A. Moore, An Improved Page-Size 127 mm Pixel Amorphous Silicon Image Sensor for X-ray Diagnostic Medical Imaging Applications, SPIE Proceedings, Medical Imaging 1997, 22-28 February 1997. DGZfP Proceedings BB 67-CD 132 Paper 22