STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

Similar documents
Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Quality Control for Stereotactic Breast Biopsy. Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc

7/20/2014. Outline. Outline. Disclosures. Learning Objectives. SBB: Practical Aspects of ACR Accreditation, QC and ACR On Site Surveys

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities

Quality Control of Full Field Digital Mammography Units

Introduction. Digital Mammography QA: Comparing the Manufacturers Recommendations. What is QC and why is it important? Review & compare QC tests

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance

Nuclear Associates

Ansur TNT Users Manual. Plug-In

REQUIREMENTS FOR LICENCE HOLDERS WITH RESPECT TO QUALITY CONTROL TESTS FOR DIAGNOSTIC X-RAY IMAGING SYSTEMS

NJDEP Medical Physicist s Radiographic QC Survey Registration Number:

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Y11-DR Digital Radiography (DR) Image Quality

4/19/2016. Quality Control Activities for the RadiologicTechnologist. Objectives. 3D Tomosynthesis QC differences

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

Overview of Safety Code 35

8/2/2017. Radiologist Responsibilities. Radiologist Responsibilities. Medical Physicist Mammography Equipment Evaluation and Annual Survey

Aspire HD. Program Manual. 2nd Edition - October 2012

Nuclear Associates

FFDM -FCRm QC Requirements- What You REALLY Need to Know

FFDM in the Field: Physicist's Role in the QC of Mammography Laser Printers May Carl R. Keener, Ph.D., DABMP, DABR

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

Test Equipment for Radiology and CT Quality Control Contents

Facility, Unit and Test Equipment Data

TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Mammography: Physics of Imaging

MILADY. Product Data. Page 1 of 8

Collimation Assessment Using GAFCHROMIC XR-M2

QC Testing for Computed Tomography (CT) Scanner

Ludlum Medical Physics

Features and Weaknesses of Phantoms for CR/DR System Testing

MAMMOGRAPHY - HIGH LEVEL TROUBLESHOOTING

Quality control for digital mammography: Part II recommendations from the ACRIN DMIST trial

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter

Published text: Institute of Cancer Research Repository Please direct all s to:

Beam-Restricting Devices

Introduction of Computed Radiography in Two Mammography Services: Image Quality and Dose Analysis

Half value layer and AEC receptor dose compliance survey in Estonia

Exposure System Selection

diagnostic examination

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of

DIAGNOSTIC ACCREDITATION PROGRAM. Radiology and CT Quality Control Procedures Workbook

QUALITY CONTROL TESTS IN SOME DIAGNOSTIC X-RAY UNITS IN BANGLADESH

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions

QC in Diagnostic Radiology. Main steps for a QC survey in Diagnostic Radiology

GE Healthcare. Performa. High-performance breast imaging

Learning Objectives: What s my motivation? (unknown screen actor) Workshop Overview

Overview. Professor Roentgen was a Physicist!!! The Physics of Radiation Oncology X-ray Imaging

THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY

TECHNICAL DATA. GIOTTO IMAGE SDL/W is pre-arranged for Full Field Digital Biopsy examination with the patient in prone position.

Image Display and Perception

Digital radiography (DR) post processing techniques for pediatric radiology

Breast Imaging Basics: Module 10 Digital Mammography

ACPSEM Position Paper RECOMMENDATIONS FOR A DIGITAL MAMMOGRAPHY QUALITY ASSURANCE PROGRAM V4.0

Safelight Fog does what to contrast and density on film?

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Nuclear Associates

7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline

Unit thickness. Unit area. σ = NΔX = ΔI / I 0

The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System

New spectral benefi ts, proven low dose

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by

Imaging Technique Optimization of Tungsten Anode FFDM System

Image Quality. HTC Grid High Transmission Cellular Grid provides higher contrast images

Protocol for the Quality Control of the Physical and Technical Aspects of Digital Breast Tomosynthesis Systems

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II

of sufficient quality and quantity

LECTURE 1 The Radiographic Image

KODAK DIRECTVIEW CR Mammography Feature User s Guide

Diagnostic x-ray equipment compliance and facility survey

Assessment of Beam Alignment, Collimation and Half Value Layer of Some Selected X-Ray Machines in Plateau State, Nigeria

ABOVETABLE X-RAY SOURCE FLUOROSCOPIC AND SPOT-FILM SYSTEMS

X-ray Tube and Generator Basic principles and construction

CHAPTER 6 QC Test For Fluoroscopic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC

History of digital imaging

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Essentials of Digital Imaging

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

Calibration of KAP meters

Image Quality. HTC Grid High Transmission Cellular Grid provides higher contrast images

Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities

Digital Breast Tomosynthesis

GE AMX 4+ Portable X-Ray

Seminar 8. Radiology S8 1

Unfors EDD-30 Radiation Protection in Fluoroscopy

ODYSSEY HF SERIES. ULTRA High Frequency X-Ray Technology. Precision... Performance... Power. Innovations in Digital Imaging.

T h e P h a n t o m L a b o r a t o r y

Nuclear Associates

Radiographic Techniques, Contrast, and Noise in X-Ray Imaging

Title: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS

Protocol for the Quality Control of the Physical and Technical Aspects of Digital Breast Tomosynthesis Systems

Outline. Digital Radiography. Understanding Digital Modalities: Image Quality and Dose. Image Quality. Dose Control

Comparison of computed radiography and filmõscreen combination using a contrast-detail phantom

Mammograph FFDM PRODUCT DATA

FMT18 FLOOR MOUNTED SYSTEM

Digital Radiography System

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London

Teaching Digital Radiography and Fluoroscopic Radiation Protection

Transcription:

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS JAMES A. TOMLINSON, M.S. Diagnostic Radiological Physicist American Board of Radiology Certified Medical Physics Consultants, Inc.

Bio 28 yrs experience 100% Dx Rad >100 mammo systems/yr 15 are dedicated stereotactic units

SBB-Testing Requirements Currently exempt from MQSA certification and ACR accreditation (voluntary). Some states require testing, all tests that are applicable. This focus is on dedicated units with digital IR.

Testing Procedures The ACR-SBB manual contains testing procedures for technologists and physicists. There are options for filmscreen units and those with digital receptors.

SBB Units Dedicated prone table units are made by Fischer and Lorad. Other re-salers have put their label on the Lorad unit. Add-on stereotactic devices are also used. We will only discuss the prone table units.

FISCHER MAMMOTEST

Multi-Angle Biopsy

Image Receptor 9-10 lp/mm 1024 2 Min-R screen 5x5cm Fiber-optic taper CCD 1 x1 1024x1024

MTF

LORAD STEREOGUIDE

Breast support Detector assembly

Image Receptor ~5 lp/mm 512 2 7-10 lp/mm 1024 2 Lens Mirror CCD 1 x1 1024x1024 Lanex screen

EQUIPMENT TESTING 1. Unit assembly evaluation 2. Collimation assessment 3. Focal spot performance/system resolution 4. kvp performance 5. Beam quality (HVL) 6. AEC or manual exposure performance

EQUIPMENT TESTING 7. Digital image uniformity 8. Breast radiation dosimetry 9. Image quality 10. Artifact evaluation 11. Localization accuracy

1. Unit Assembly Evaluation Free-standing dedicated unit is mechanically stable All moving parts move smoothly All locks and detents work properly Image receptor holder assembly is free from vibrations Image receptor is held securely by assembly Compressed breast thickness scale accurate to 5 mm Patient or operator not exposed to sharp or rough edges or other hazards

2. Collimation Assessment Radiation to digital image alignment Radiation beam confinement to image receptor

Coin Method

Monitor Film

Collimator Tool Method

Monitor Film

Collimation Tolerance The radiation field should extend beyond the image receptor But not more than 5 mm on any edge

Radiation Beam Confinement Often, the rad field will exceed the image receptor at the chest wall breast unnecessary radiation radiation field image field metal compression paddle

3. Focal Spot Performance The focal spot performance is tested with the res pattern and film (no screen)

Focal Spot Performance 11 lp/mm in the direction perpendicular to the anode cathode axis 13 lp/mm in the direction parallel to the anode cathode axis 13 parallel 11 perp

System Resolution (Digital) The resolution test pattern is used with the digital receptor to find limiting resolution of the 50x50 mm image size. For 512 image, pixel size is = 0.098 mm = 5.1 lp/mm For 1024 image, pixel size is 0.049 mm = 10.2 lp/mm Typically see 7-10 lp/mm

Q1- The maximum limiting system resolution for a dedicated SBB device in 1024 mode would be: 20% 20% 20% 20% 20% 1. 1 lp/mm 2. 3 lp/mm 3. 5 lp/mm 4. 7 lp/mm 5. 10 lp/mm 10

5. 10 lp/mm Bushberg, Jerrold T, Ph.D., et al; The Essential Physics of Medical Imaging, Second Edition; Lippincott, Williams & Wilkins, 2002, p. 284

4. KVp Performance

kvp Accuracy and Reproducibility Mammography units should have clinically used kvp accuracy to within 5% of the indicated kvp From four measurements at most commonly used kvp, compute mean, s.d., and coefficient of variation (sd/mean) C.V. should be within 0.02

Q2- Tube potential accuracy should be within: 20% 20% 20% 20% 20% 1. 1 % 2. 1 kvp 3. 2 % 4. 2 kvp 5. 5 % 10

5. 5 % American College of Radiology; Stereotactic Breast Biopsy Quality Control Manual; 1999, p. 65

5. Beam Quality (HVL) The HVL requirements are slightly different from filmscreen mammo, because the image port is open (no paddle attenuation) The HVL should be above HVL kvp/100 (mm Al) The ACR suggests an upper limit of HVL < kvp/100 + 0.12 (mm Al) for Mo/Mo At 28 kvp, 0.40 HVL 0.28

Half-Value Layer High-purity aluminum

Half-Value Layer Logarithmic interpolation can be are used to calculate HVL Using the logarithmic method, where E 0 =0 mm, E a = first added filtration, E b =second total added filtration HVL = t b ln (2E a / E o ) - t a ln (2E b / E o ) ln ( E a / E b )

Q3- The half-value layer (HVL) at 28 kvp must be greater than or equal to: 20% 20% 20% 20% 20% 1. 0.26 mm Al 2. 0.28 mm Al 3. 0.30 mm Al 4. 0.32 mm Al 5. 0.35 mm Al 10

2. 0.28 mm Al American College of Radiology; Stereotactic Breast Biopsy Quality Control Manual; 1999, p. 68

6. AEC Performance (if used) For the performance test, 4, 6, and 8 cm of lucite is evaluated with clinical technique factors

AEC System Performance For digital systems: read the signal value in the center of the image signal from 6 and 8 cm should be within ± 20% of the 4 cm reading if not, a technique chart should be developed and posted

Technique Chart Posted techniques should be evaluated and adjusted to keep image quality high, yet exposure times < 2 seconds

7. Digital Image Uniformity 4 cm lucite image Signal analysis in center and quadrants Signal Standard deviation (sd) SNR Calculate quadrant deviance from the center values Systems without ROI capability, use wire mesh pattern

Lorad Tolerances SNR in each corner ± 15% of center (early manuals show 128x128 box). I ve never found a Lorad unit that will pass with this protocol. Later test protocols use a 32x32 box and specifies the corner areas to sample. The specified tolerance was that the SNR in each corner not exceed ± 20% of the central SNR. Most units will pass with these parameters.

6% of total area 0.4% of total area 128x128 512 image 32x32 512 image

Newer Lorad Units On the latest units, the operator manual indicates to take signal in the center of 1024 image with 32x32 box (0.1% of total area). This also requires significantly more radiation to keep signal at 4k, when quadrupling image matrix. Signal in each quadrant should be within 15% of the center. All of them appear to be able to pass this protocol.

Fischer Tolerances SNR in each corner ± 15% of center (ACR manual shows 128x128 box) Most units will pass this tolerance

Q4- The ACR digital receptor uniformity tolerance for signal value is: 20% 20% 20% 20% 20% 1. 5 % 2. 10 % 3. 15 % 4. 20 % 5. 25 % 10

3. 15% American College of Radiology; Stereotactic Breast Biopsy Quality Control Manual; 1999, p. 75

8. Breast Radiation Dosimetry Find technique factors that give manufacturer recommended signal values for the ACR phantom or the mas given from an AEC technique. Position ion chamber in port at 4.5 cm from breast support. Measure manual exposure 4 times. For AEC techniques, use closest mas. Calculate average, s.d., and c.v. < 0.05

Breast Dose

Breast Dosimetry For AEC technique, calculate ESE from: ESE aec = ESE man (mas aec /mas man ) Calculate the average glandular dose from look-up table in ACR manual (kvp vs HVL) for Mo/Mo. Typical conversion factor is 0.18 at 28 kvp. Dose should be less than 3.0 mgy (300 mrad). (Michigan requires 2.0 mgy for all mammography)

9. Image Quality Make an exposure of the ACR phantom using the kvp and density normally used clinically. For the standard ACR phantom, image in each quadrant to determine IQ score. For the ACR mini-phantom, only one image is needed.

ACR Phantom Screen-film: Digital: 4.0 fibers, 3.0 specks, 3.0 masses 5.0 fibers, 4.0 specks, 3.5 masses

Mini-Phantom

Mini-Phantom Screen-film: Digital: 2.0 fibers, 2.0 specks, 2.0 masses 3.0 fibers, 3.0 specks, 2.5 masses

ACR Phantom Scoring Fibers are scored down to the smallest full or half fiber visible. A full point is given for full fiber, 0.5 for half. Specks are scored down to the smallest visible group. A full point is given if specks 4, 5, & 6 are visible, 0.5 if only 2 & 3 are visible, and 0 if only 1 speck is visible. Masses are scored down to the smallest visible. A full point is given if round and in the right location, a 0.5 point if in the right location but not round.

Score Subtraction for Artifacts If a fiber-like artifact is as apparent as an actual fiber, it is subtracted from the last half or full fiber. If speck-like artifacts are present, they are deducted on a one for one basis from the last group counted. If a mass-like artifact is as apparent as an actual mass, it is subtracted from the last real mass.

Mag Tool

Q5- The digital image quality required for the dedicated mini-phantom: 20% 20% 20% 20% 20% 1. 4.0 fibers, 4.0 specks, 4.0 masses 2. 4.0 fibers, 3.5 specks, 3.5 masses 3. 3.5 fibers, 3.5 specks, 3.5 masses 4. 3.0 fibers, 3.0 specks, 3.0 masses 5. 3.0 fibers, 3.0 specks, 2.5 masses 10

5. 3.0 fibers, 3.0 specks, 2.5 masses American College of Radiology; Stereotactic Breast Biopsy Quality Control Manual; 1999, p. 91

10. Artifact Evaluation For digital IR image lucite sheet large enough to cover IR Adjust the window width and level to look for defects in the image

Very narrow window

11. Localization Accuracy

Mammotome

Scout Image (0 0 )

Stereo Scouts (+15 0, -15 0 )

Oh Baloney!

Thin metal pieces

Stereo Scouts

Pre-Fire

Post-Fire

Technologist s QC Zero Alignment - before each patient Localization Accuracy (XYZ) - each day of use Phantom Imaging - each week of use Visual Checklist - monthly Hardcopy QC - monthly (if applicable) Compression - semiannual Repeat Analysis - semiannual

Coordinates Calibration (XYZ)

Hardcopy QC (monthly) Print SMPTE pattern from console Print SMPTE from printer or Radiograph a step-wedge and print

Q6- The localization accuracy (in air) and phantom image quality should be conducted by the technologist: 20% 20% 20% 20% 20% 1. Daily, daily 2. Daily, weekly 3. Each day of use, weekly 4. Each day of use, each week of use 5. Weekly, weekly 10

4. Each day of use, each week of use American College of Radiology; Stereotactic Breast Biopsy Quality Control Manual; 1999, pp. 17&21

SUMMARY SBB-ACR accreditation is voluntary, for now Even though exempt from MQSA testing regimen, dedicated units should be tested routinely by the Medical Physicist, and may be required by state agencies.