Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Similar documents
Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

LSM 710 Confocal Microscope Standard Operation Protocol

LSM 780 Confocal Microscope Standard Operation Protocol

LSM 800 Confocal Microscope Standard Operation Protocol

Zeiss LSM 880 Protocol

Zeiss 880 Training Notes Zen 2.3

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL

Zeiss 780 Training Notes

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s

Microscopy from Carl Zeiss

Operating Instructions for Zeiss LSM 510

Things to check before start-up.

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014

Zeiss LSM 780 Protocol

Training Guide for Carl Zeiss LSM 510 META Confocal Microscope

ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL

Zeiss Axiovert 135 Fluorescence Microscope Quick Guide / Operations Manual (v. 1.0 February 09)

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling

LSM 510 Meta Training Notes

Leica SP8 TCS Users Manual

ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL

Zeiss LSM880 Operating Instructions. UTMB Optical Microscopy Core Jan. 16, 2018

Contents. Introduction

Guide to Confocal 5. Starting session

LSM 510 Training Notes

Training Guide for Carl Zeiss LSM 880 with AiryScan FAST

OPERATING INSTRUCTIONS

LEICA TCS SP5 AOBS TANDEM USER MANUAL

Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope

Leica SP8 TCS Users Manual

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope

Quick Start Guide. Leica SP5 X

TRAINING MANUAL. Olympus FV1000

Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope

Nikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.

REMEMBER: You have 5GB of disk space on this microscope. Check before you start if you have room for your experiment. If not delete your old data.

Title: Leica SP5 Confocal User Manual

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope

Zeiss AxioImager.Z2 Brightfield Protocol

Cell Biology and Bioimaging Core

ZEISS LSM510META confocal manual

Nikon SIM-E & A1-R System

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope

Leica Sp5 II Confocal User Guide

CONFOCAL MICROSCOPE (Zeiss LSM 510 META v4.2)

Topics. - How to calibrate the LSM scanner. - How to clean the microscope. - How to adjust the pinhole alignment. - How to adjust the Collimator

Zeiss Axio Imager.A1 manual

Zeiss AxioObserver with ApoTome

Zeiss LSM 510 Multiphoton Confocal Microscope

Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software

Internal Medicine Imaging Core Emory University Department of Medicine

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging

Zeiss LSM 510 Multiphoton Confocal Microscope

Zeiss LSM 510 META Guide

QUICKSTART GUIDE: WIDEFIELD HWF1 Zeiss Cell Observer Live Cell Imaging System (HAMMERSMITH, L BLOCK, ROOM 314) Imperial College London

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017.

SPINNING DISK CSU-X1 USER MANUAL

Olympus Digital Microscope Camera (DP70) checklist

Practical work no. 3: Confocal Live Cell Microscopy

Widefield 1. Switching on

Title: Nikon A1R Confocal User Manual

Zeiss Deconvolution Microscope: A Quick Guide

Everest System / Slidebook Operating Procedures

DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands

Training Guide for Leica SP8 Confocal/Multiphoton Microscope

QUICKSTART GUIDE: WIDEFIELD WF3 Zeiss Cell Observer Live Cell Imaging System (SAF, ROOM 409) Imperial College London

DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions

Leica SPEII confocal microscope. Short Manual

Using the Nikon TE2000 Inverted Microscope

Zeiss Axioskop II. The AIF's "routine" light microscope. (Installed 8/24/04)AxioCam installed July 11th 2005

Leica TCS SP8 Quick Start Guide

Renishaw InVia Raman microscope

Nikon C1si Spectral Laser Scanning Confocal Microscope. User Guide

Microscope Confocal LSM510 META

The Zeiss AiryScan System, Confocal Four.

Nikon A1R. Multi-Photon & Laser Scanning Confocal Microscope. Kyle Marchuk Adam Fries Jordan Briscoe Kaitlin Corbin. April 2017.

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7

Leica TCS SL Confocal Training. Neuroscience Imaging Core Staff. Core Director. Facility Manager

Nikon E800 Operating Instructions.

Nasmyth Ultraview Vox User Protocol

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software:

Simplified Instructions: Olympus Widefield Microscope S1230

b. Turn the power switch and key to on position for blue laser.

Instructions for Making On-Line Reservations for Microscopes in NB11-204

Standard Operating Procedure (SOP) for Shared Equipment: Spinning Disk Confocal Microscope

Focus Strategies with ZEN2. Sven Terclavers ZEISS Embedded 3D Imaging Specialist Harvard University,

Nikon A1Rsi Confocal Start-Up Sequence

Leica SP8 Resonant Confocal. Quick-Start Guide

Olympus xcellence Software - basic user guide

Brief manual how to start and close the Leica sp2 Confocal. (TCS SP2 AOBS system mounted on a DM IRE2)

Diskovery Spinning Disk Guide

Boulevard du Temple Daguerrotype (Paris,1838) a busy street? Nyquist sampling for movement

AxioVision 4.5 Brightfield Image Capture Procedure

Zeiss Axioplan 2 imaging microscope and Axiovision software

Leica DMi8A Quick Guide

Leica TCS SP8 Quick Start Guide

Transcription:

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center For any questions or concerns, please contact: Linda Nieman lnieman@mgh.harvard.edu Office: (617) 643-9684 Cell: (512) 565-8076 Chenyue Lu clu0@mgh.harvard.edu Cell: (413) 561-3316 Power Up Room code: 541 Turn on equipment in the following order: 1. Main System Power Make sure Key on right hand side in in the horizontal position. 1A. Flip MAIN SWTICH to ON position 1B. Flip SYSTEM/PC switch to ON position i. Wait for the Definite Focus module to display OFF 1C. Flip COMPONENTS switch to ON position i. Wait for components to initialize 2. Argon Laser Black Box Power Supply Turn the key to the right (clockwise) 3. Argon Laser Remote Controller Flip switch to RUN position 4. X-Cite widefield fluorescence light source Flip switch to ON position Write down the bulb hours on the sign in sheet 5. Turn on PC Login as LSM USER Wait for ~3 min for RTC to establish connection with computer 6. Zen software Double click ZEN black icon in center of desktop (Not Zen blue) When startup screen appears, click on the BOOT STATUS drop down menu Click START SYSTEM i. If the screen stalls at 2% or an error about the Real Time Controllers pops up. Press the RTC reset button on the box under the confocal for 5 seconds. Wait 30 seconds. If it fails to continue, close the software, press the button again, and then restart the ZEN software. If that fails, close Zen software, then power off the RTC. Wait 30 seconds. Power the RTC back on. Wait 1-2 minutes. Re-launch Zen. When startup screen appears, click on BOOT STATUS then START SYSTEM 1

Loading your sample 1. Push microscope head back. Do NOT push on gray HAL100 light box. 2. On Touchpad, choose objective. View using 10X first, then progress to higher magnification 3. Tap LOAD POSITION. (Drops down stage) 4. Load slide onto stage coverslip down (inverted microscope) 5. Bring microscope head back into position 6. On Touchpad, tap SET WORK POSITION Finding your sample 1. In Zen, go to the LOCATE tab 2. Click ONLINE to view through eyepieces 3. Choose configuration: a. BRIGHTFIELD b. DIC II For use with dry objectives i. 10x, 20x, 63x dry ii. Requires polarizers c. DIC III for use with wet objectives i. 40x oil, 40x water, 63x oil ii. Requires polarizers d. DAPI blue filter e. GFP green filter f. DsRED red filter 4. Focus on sample Too bright? Lower bright field transmission in ocular tab or using large black dial on XCite light box. 5. Adjust Kohler illumination, if needed: a. Close top aperture b. Focus with condenser head until a clear hexagon is formed c. Re-open top aperture 6. Click OFFLINE when finished a. This will turn off widefield lightsource 2

Setting confocal configuration 1. Click ACQUISITION tab to enter confocal mode NOTE: There are three different ways to setup a scan: a. Load saved configuration (These can be changed, so check before use.) b. Load template image (File / Open), then click REUSE c. Smart Setup (see instructions below) a. Load Configuration b. Re-use template image settings c. Smart Setup Smart setup 1. In the EXPERIMENT MANGER panel, choose SMART SETUP a. Configure your experiment. Choose dye / fluorophore and image display color b. Add additional dyes / fluorophores using + button. You can also delete dyes using the buttons 2. Choose imaging option a. FASTEST: i. Advantage: faster image acquisition ii. Disadvantage: potential cross-talk (spectral overlap) between channels. b. BEST SIGNAL: i. Advantage: Only one detector and one laser are switched on at any one time. This reduces crosstalk ii. Disadvantage: slower image acquisition (each color track imaged separately) c. BEST COMPROMISE - between both speed and best signal 3. Click APPLY 3

Image capture settings 1. In the ACQUISITION MODE panel, set the following parameters: a. SCAN MODE = FRAME b. FRAME SIZE (i.e. pixel density) = 1024x1024 c. SCAN SPEED should be set between 7-9 i. NOTE: faster scanning gives smaller time interval between images, but lower signal-tonoise ratio. Slower scanning gives longer time intervals, but higher signal-to-noise ratio d. AVERAGING = 1 i. NOTE: if you have a dim sample with white noise, you may want to consider averaging at least 4 frames. Note that adding frames adds to the total scan time. e. BIT DEPTH = 12 for publication images (or 16 bit for high dynamic range images) i. NOTE: increasing bit depth increases number of gray scale levels (2 n ) 2. Set PINHOLE to desired spatial resolution NOTE: For best resolution, click 1 AU NOTE: There is an inherent tradeoff between spatial resolution and image brightness. If your sample is dim and high resolution is not critical, you may want to set pinhole larger 4

Setting the detector gain 1. Click SET EXPOSURE a. (System will find optimal gain setting for each color channel) 2. Click LIVE b. NOTE: live continuously scans sample at low resolution to allow quick adjustments. The SNAP image will be captured with Image capture settings above 3. Adjust focus, if needed 4. Click STOP 5. Click TRACK1 a. Desired track will be highlighted in light gray. Other color channels can be unchecked 6. Click LIVE 7. Enable SINGLE CHANNEL a. (Active channel will be highlighted blue) 8. Enable RANGE INDICATOR a. Red shows pixels that are saturated b. Black are pixels that are underexposed c. Blue are pixels that have optimal exposure d. Aim for an image that is has few red pixels 5

9. Adjust GAIN (MASTER) a. A good initial setting is ~650 b. The maximum gain is 1200 *** Aim for an image that is has few red pixels *** 10. Click STOP in EXPERIMENT MANAGER 11. To view another track (i.e. color channel), click on TRACK2 in the CHANNELS panel. 12. Adjust GAIN (MASTER) for TRACK2 13. Repeat until all color channels give good images 14. Click STOP in EXPERIMENT MANAGER 6

Acquiring a z-stack 1. Check the Z-STACK box 2. Click LIVE 3. Focus (see the blue plane on the screen) manually on the plane closest to the objective, then click SET FIRST 4. Focus (see the blue plane on the screen) manually on the plane furthest to the objective, then click SET LAST 5. Click STOP to stop scanning 6. Select either SMALLEST slice interval or select the number of SLICES desired 7. Click START EXPERIMENT to begin acquiring z- stack 7

Graphics / Overlays 1. Add in: a. Text b. Scale bar c. Lines d. Arrows e. Boxes f. Circles g. Free draw Saving your images 1. Click disk icon to save 2. Save to Data / LAB PI NAME / YOUR NAME if saving locally 3. Please do not store data permanently on computer. It is not backed up and data folders are periodically deleted 4. Images can be opened using FIJI (ImageJ) 5. Images can be exported as TIFF from the FILE / EXPORT menu Finishing up Turn off equipment in following order: 6. Close ZEN software 5. Shut down PC 4. Turn off X-Cite light source 3. Argon Laser Remote controller: Flip switch to OFF Position 2. Argon Laser Black Box Power Supply: Turn key to left (counter-clockwise) 1. Main system power: 1C. Flip COMPONENTS switch to OFF position 1B. Flip SYSTEM / PC switch to OFF position *** WAIT 5 MINUTES TO LET LASER COOL *** (you will hear fan shut off when laser is cool enough) 1A. Flip MAIN SWITCH to OFF position 1. Clean work area 2. Sign out 3. Clean any immersion objectives that you used with lens paper DO NOT USE KIMWIPES! 8

ZEN lite for Windows Download ZEN lite, your free copy of the powerful ZEN imaging software. Use this free microscope software to try out key features of the ZEN core package or simply install it as a viewer for your CZI files. Use ZEN lite to: Control Axiocam microscope cameras Create microscope images and simple video sequences Measure distances interactively Manage images/videos and export them into various formats View meta information relating to CZI image files Use simple report functions Expand ZEN lite with selected modules: create impressive multichannel and time lapse images of your specimens or use extended measurement functions. To download go to: http://www.zeiss.com/microscopy/en_us/products/microscope-software/zen-lite.html 9