Leica SP8 Resonant Confocal. Quick-Start Guide

Similar documents
Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software:

Leica SP8 TCS Users Manual

Quick Start Guide. Leica SP5 X

Leica Sp5 II Confocal User Guide

Things to check before start-up.

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.

LSM 780 Confocal Microscope Standard Operation Protocol

Leica TCS SP8 Quick Start Guide

Leica TCS SP8 Quick Start Guide

Leica SPEII confocal microscope. Short Manual

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Leica SP8 TCS Users Manual

Guide to Confocal 5. Starting session

Zeiss LSM 880 Protocol

LEICA TCS SP5 AOBS TANDEM USER MANUAL

Training Guide for Leica SP8 Confocal/Multiphoton Microscope

Usermanual for Leica SP8 confocal

Zeiss LSM 780 Protocol

TRAINING MANUAL. Olympus FV1000

3. are adherent cells (ie. cells in suspension are too far away from the coverslip)

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL

LSM 710 Confocal Microscope Standard Operation Protocol

Operating Instructions for Zeiss LSM 510

Nikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol

Cell Biology and Bioimaging Core

Zeiss 880 Training Notes Zen 2.3

Nikon A1Rsi Confocal Start-Up Sequence

Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope

Overview. About other software. Administrator password. 58. UltraVIEW VoX Getting Started Guide

Zeiss LSM880 Operating Instructions. UTMB Optical Microscopy Core Jan. 16, 2018

ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL

Title: Leica SP5 Confocal User Manual

Nikon C1si Spectral Laser Scanning Confocal Microscope. User Guide

Imaging Beyond the Basics: Optimizing Settings on the Leica SP8 Confocal

ZEISS LSM510META confocal manual

Zeiss 780 Training Notes

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017.

Simplified Instructions: Olympus Widefield Microscope S1230

1 Co Localization and Working flow with the lsm700

Microscope Confocal Sp2 Upright.

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL

LSM 510 Training Notes

Zeiss Axio Imager.A1 manual

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging

LSM 510 Meta Training Notes

Nikon A1R. Multi-Photon & Laser Scanning Confocal Microscope. Kyle Marchuk Adam Fries Jordan Briscoe Kaitlin Corbin. April 2017.

LSM 800 Confocal Microscope Standard Operation Protocol

Use of the HSW5 Spinning Disk Confocal Microscope Updated last May 25, 2010 OK

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling

b. Turn the power switch and key to on position for blue laser.

QUICKSTART GUIDE: WIDEFIELD HWF1 Zeiss Cell Observer Live Cell Imaging System (HAMMERSMITH, L BLOCK, ROOM 314) Imperial College London

Brief manual how to start and close the Leica sp2 Confocal. (TCS SP2 AOBS system mounted on a DM IRE2)

ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide

Title: Nikon A1R Confocal User Manual

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7

Zeiss Axiovert 135 Fluorescence Microscope Quick Guide / Operations Manual (v. 1.0 February 09)

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each

Nikon SIM-E & A1-R System

Microscope Confocal LSM510 META

Training Guide for Carl Zeiss LSM 510 META Confocal Microscope

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Swept-Field User Guide

Internal Medicine Imaging Core Emory University Department of Medicine

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope

Leica TCS SL Confocal Training. Neuroscience Imaging Core Staff. Core Director. Facility Manager

DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands

Zeiss Deconvolution Microscope: A Quick Guide

User Guide to the IBIF Leica TCS SP8 MP Confocal Microscope

Training Guide for Carl Zeiss LSM 880 with AiryScan FAST

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO

Last updated: May 2014 Y.DeGraaf

DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands

Microscopy from Carl Zeiss

Renishaw InVia Raman microscope

Contents. Introduction

QUICKSTART GUIDE: WIDEFIELD WF3 Zeiss Cell Observer Live Cell Imaging System (SAF, ROOM 409) Imperial College London

Leica TCS SP2 User Manual 1.1. Markus Friedrich and James Lee. Department of Biological Sciences Wayne State University

Leica Confocal - 2. Instructions for Leica SP2 Confocal Equipped with Visible Laser Lines

ScanArray Overview. Principle of Operation. Instrument Components

Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope

Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev. Microscopy course, Michmoret Dec 2005

SPINNING DISK CSU-X1 USER MANUAL

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope

Practical work no. 3: Confocal Live Cell Microscopy

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions

Diskovery Spinning Disk Guide

Operation Of The Leica SP8 Multiphoton Confocal System Using Single Or Multiple Fluorochromes

Widefield 1. Switching on

Topics. - How to calibrate the LSM scanner. - How to clean the microscope. - How to adjust the pinhole alignment. - How to adjust the Collimator

Thor. Nikon Inverted Structured Illumination (N-SIM) & A1 Laser Scanning Confocal Microscope User Guide. v. 1.0 (12/2015)

Basics of confocal imaging (part I)

Using the Nikon TE2000 Inverted Microscope

Nikon Ti-E Microscope Manual. Rightmire Hall Ohio State University. Director: Tony Brown Rightmire

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

Supplemental Method Information Zeiss LSM710

Instructions for the Leica SP5 II laser scanning confocal microscope

OPERATING INSTRUCTIONS

Transcription:

Leica SP8 Resonant Confocal Quick-Start Guide

Contents Start-up Preparing for Imaging Part 1 On the scope Part 2 Software interface Part 3 Heat & CO2 incubation Part 4 Other hardware options Shut-down

Start-up: Turn on the Microscope / Computer Turn on 3 green switches from bottom to top Turn key to on position Turn on fluorescence excitation source for the eyepieces. Log-in using your ADS account name and password. For access to the network drive, select Run and then type: \\microscopy-nas1.nri.ucsb.edu Open LASX software Choose resonant scanner to be on or off

Part 1: on the scope Lean condenser/bright-field arm back Choose objective from LASX software Pay attention to immersion medium Place slide or dish in sample holder Use scope touchscreen to choose brightfield or fluorescence Be sure to turn off fluorescence shutter when not observing through eyepiece Small knob on left side will adjust fluorescence excitation intensity

Part 2: Software interface Acquisition Excitation and Emission Image Viewing

Part 2: Software interface Excitation Emission Click + button to access laser power menu Toggle from off to on for each laser Visible argon laser is mostly for bleaching Click Switch to Whitelight to see WLL interface Activate laser lines by clicking on 1,2,,8 Move line to desired wavelength Top number is laser % (5 is a good place to start) Click checkbox to turn on You can set which properties the control panel controls, and how coarse or fine the control is Dye assistant will set excitation and emission for chosen fluorophores Use the crosstalk column to decide if sequential scan is necessary For each detector you can set the starting and ending position of the emission window The only limitation is that the channel windows have to be arranged in order from left to right, 1 is leftmost, then 2 to the right of 1, etc.

Part 2: Software interface Scan Settings, Z-stack, timelapse Scan Settings Format: number of pixels, drop down menu gives choices, + button gives exact control. Speed: For resonant it is fixed at 8000. For galvo there are many options, speeds greater than 600 will limit field-of-view. Bidirectional: doubles scan speed, phase may need adjustment if image appears blurry. Pixel size: spatial size of pixels. To achieve best resolution zoom or increase pixel count to Nyquist limit. Button to the left of Format will set pixel count to Nyquist. Line Average: If low signal or noisy image increase line average to improve signal/noise. You will almost always need to use this with the resonant scanner. Z-stack Set Begin and End to desired positions System optimized: Nyquist z-spacing Z-Compensation: adjust excitation or detection gain to compensate for brightness changes with focal depth changes. Time Minimize: sets minimum time interval limited by image acquisition time Set length of timelapse in number of stacks, duration, or acquire until stopped

Part 2: Software interface Timelapse, Sequential, Autofocus Stage Tilescan: add a few positions with that button, then it will set a rectangular grid to include all of those positions. Mark and Find: track multiple fields-of-view for timelapses. For tilescan choose autostitching and blending method to have immediate stitching of grid of fields-of-view. Autofocus AFC tracks position of coverslip to eliminate z-drift over time. Best focus searches for most contrast within a certain z-range. Use this for samples that can move over time. Sequential Add sequences to separate excitation and emission for different fluorophores. Use this to eliminate crosstalk.

Part 3: Heat and CO 2 incubation stage Turn on switch on Okolab touch screen. Turn on switch on Lauda water bath. Open CO 2 valve (if needed). Set desired temperature and CO 2 percentage. Turn off airflow if CO 2 is not needed. Remove z-galvo stage, set aside. Two thumbscrews and one hex screw (red screwdriver). Move objective to lowest possible position. Slide in incubation stage so that spring contacts are used. Remember z-galvo stage is no longer attached so for your z- stack be sure to switch from z-galvo to z-wide.

Part 4: Other options HyD 1 is a special cooled detector. It can be used in any circumstance, but should be your first choice for low signal conditions. Auto immersion dispenser is available for long timelapses with the 40x water objective. Live Data Mode is available for automating more complicated tasks involving several different imaging jobs. Hyvolution is the deconvolution software. It can be run in automatic mode or manual mode. Manual mode will require some training.

Shut-Down Procedure Check the online schedule Shut-down if nobody is scheduled for the rest of the day. Leave the system on if somebody is using the system todaybut do the following. Log-off the computer Remove sample Wipe immersion oil off of objectives with lens paper Return to the 10x objective Adjust your online reservation end-time if you finished early or late Shut off the computer Turn off key, then each switch from top to bottom Turn off eyepiece excitation light. Put dust cover over microscope

Specifications 5 objectives installed 10x/0.3 air 20x/0.75 Multi-Immersion 40x/1.1 water motorized correction collar 63x/1.3 glycerol 63x/1.4 oil Available on request: 5x/0.15 air and 40x/1.3 oil 3 laser sources UV: 405 nm Argon: White Light Laser: 470-670 nm 4 HyD and 1 PMT detectors HyD 1 is cooled for better low signal detection