Image Quality, Artifacts and Hazards in Imaging. Laura Gruber, MBA, RT(R), RDMS, RVT Sr. Director Medical Imaging

Similar documents
Teaching Digital Radiography and Fluoroscopic Radiation Protection

Multi-Access Biplane Lab

INNOVATION BY DESIGN. Toshiba A History of Leadership REMOTE CONTROL R/F SYSTEM

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

of sufficient quality and quantity

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Truly flexible to meet your clinical needs

INTRODUCTION TO FLEXIBLE BRONCHOSCOPY. Fluoroscopy Synopsis HENRI G COLT MD SECOND EDITION THE BRONCHOSCOPY EDUCATION PROJECT SERIES

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Sarah Hughes, MS, DABR Radiation Safety Officer

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development

Ludlum Medical Physics

X-RAYS - NO UNAUTHORISED ENTRY

Digital radiography (DR) post processing techniques for pediatric radiology

SYLLABUS. 1. Identification of Subject:

Test Equipment for Radiology and CT Quality Control Contents

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

PD233: Design of Biomedical Devices and Systems

The Architecture of Medical Imaging

QC Testing for Computed Tomography (CT) Scanner

R/F. Comparison of Long View Radiography Systems. 1. Introduction. 2. Methods of Long View Radiography

Image Quality Artifacts in Digital Imaging

Advanced digital image processing for clinical excellence in fluoroscopy

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions

COMPUTED TOMOGRAPHY 1

Computed Tomography. The Fundamentals of... THE FUNDAMENTALS OF... Jason H. Launders, MSc. Current Technology

diagnostic examination

FOUR CATEGORIES OF SAFETY

CR Basics and FAQ. Overview. Historical Perspective

Overview of Safety Code 35

Maximum Performance, Minimum Space

Nuclear Associates

Maximizing clinical outcomes

1-1. GENERAL 1-2. DISCOVERY OF X-RAYS

GE Healthcare. Senographe 2000D Full-field digital mammography system

CHAPTER 6 QC Test For Fluoroscopic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question.

Digital Radiographic Inspection replacing traditional RT and 3D RT Development

Nuclear Associates

Fluoroscopy - Chapter 9

JEFFERSON COLLEGE COURSE SYLLABUS BET220 DIAGNOSTIC IMAGING. 3 Credit Hours. Prepared by: Scott Sebaugh Date: 2/20/2012

Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) MEDIX VOL. 61 P.8 P.11

NJDEP Medical Physicist s Radiographic QC Survey Registration Number:

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

2 nd generation TOMOSYNTHESIS

Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London

Nuclear Associates , , , , , ,

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

QC by the MPE in Belgium

RAD. Experiences Using the RADspeed Pro EDGE Package. 1. Hospital Description. 2. Background of Adoption. Hirohito Tanaka

China Resources Wandong Medical Equipment Co., Ltd. High Frequency 50kW, 150kV Radiography System - HF50-R

Essentials of Digital Imaging

Features and Weaknesses of Phantoms for CR/DR System Testing

Y11-DR Digital Radiography (DR) Image Quality

THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

Artefacts found in computed radiography

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

RADspeed Pro. EDGEpackage C501-E041C

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT)

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

While digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them.

Minnesota Rules, Chapter 4732 X-ray Revision

Key words: fluoroscopy, dose-area-product, kerma-area-product, calibration of KAP meters, patient exposure

Tomosynthesis and Motion

An Activity in Computed Tomography

Diagnostic X-Ray Shielding

DIGITAL RADIOGRAPHY ARTIFACTS

CS 9300 Family User Guide Including

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II

Ch. 223 VETERINARY MEDICINE CHAPTER 223. VETERINARY MEDICINE GENERAL PROVISIONS X-RAYS RADIOACTIVE MATERIAL. Authority

2217 US Highway 70 East Garner, NC Main: Fax:

Some operation methods show in the catalog reguire optional eguipment

C501-E029C. RADspeed. GENERAL RADIOGRAPHIC SYSTEM Automatic

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity)

Ultimate DR flexibility

10/3/2012. Study Harder

R&F X-ray systsem. Savings With Every Exposure

160-slice CT SCANNER / New Standard for the Future

T h e P h a n t o m L a b o r a t o r y

Wide-Detector CT for TAVR Planning:

Data. microcat +SPECT

Essentials of Digital Imaging

MaxRay Handheld X-ray Systems Operator Training Exam

2017 West Coast Educators Conference Orlando. Projection Geometry. 1. Review hierarchy of image qualities (amplified version):

MUSICA Nerve Center. Artificial Intelligence. Intelligent tools for your Digital Radiography workflow. Fluoroscopy. Workflow Optimization

PET/CT Instrumentation Basics

Digital Imaging CT & MR

2D, 3D CT Intervention, and CT Fluoroscopy

Fabrício Sampaio Péres Kury Federal University of Rio de Janeiro Medical School

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

Learning Objectives: What s my motivation? (unknown screen actor) Workshop Overview

Visualization of sources of scattered radiation from x-ray equipment used for interventional radiology

Transcription:

Image Quality, Artifacts and Hazards in Imaging Laura Gruber, MBA, RT(R), RDMS, RVT Sr. Director Medical Imaging

Case 1 2

Case 1 What is the image quality issue in this picture? A.) Improper exposure/technique B.) Detector burn-in or ghosting artifact C.) Electromagnetic interference artifact (EMI) D.) Artifact from a thermoregulating blanket Case Courtesy of Judy Le, R.T. (CT) (R) 3

Case 1 What is the image quality issue in this picture? A.) Improper exposure/technique B.) Detector burn-in or ghosting artifact C.) Electromagnetic interference artifact (EMI) D.) Artifact from a thermoregulating blanket Case Courtesy of Judy Le, R.T. (CT) (R) 4

Case 1 Key Concepts: 1.) Thermo-regulating blankets can cause honey-comb like artifacts if they are left in place during xray. These artifacts do not obscure visibility of lines, however, they can be distracting when evaluating bowel gas and potential free air. 3.) When evaluating lines and lungs, thermoregulating blankets are acceptable to leave in if patient condition prohibits removal. If evaluating pathology in the abdomen, artifacts caused by these blankets may affect diagnostic interpretation. Case Courtesy of Judy Le, R.T. (CT) (R) 5

Case 2 6

Case 2 What is the image quality issue in this picture? A.) Baby should be prone B.) Immobilization board will create significant artifact C.) Lead shield in the fluoroscopic field of view D.) Using an image intensifier instead of flat panel fluoroscope 7

Case 2 What is the image quality issue in this picture? A.) Baby should be prone B.) Immobilization board will create significant artifact C.) Lead shield in the fluoroscopic field of view D.) Using an image intensifier instead of flat panel fluoroscope 8

Case 2 Key Concepts: 1.) All modern fluoroscopes uses automatic brightness control (ABC) which means the machine adjusts the xray tube output based on how much attenuation it detects. 2.) A lead gonad shield in the fluoroscopy field of view will cause the fluoroscopy system to substantially increase radiation to the patient AND compromise image quality. 3.) You should also not use lead shields in the field of view for other modalities that use automatic exposure control (ex. xray=phototiming, CT=dose modulation) 9

Case 3 10

Case 3 What is the image quality issue in this picture? A.) Improper exposure/technique B.) Detector burn-in or ghosting artifact C.) Electromagnetic interference artifact (EMI) D.) Artifact from a positioning device 11

Case 3 What is the image quality issue in this picture? A.) Improper exposure/technique B.) Detector burn-in or ghosting artifact C.) Electromagnetic interference artifact (EMI) D.) Artifact from a positioning device 12

Case 3 Key Concepts: 1.) Scoliosis exams are often performed with a series of 2 or 3 images exposing the same detector in rapid succession. 2.) Though rare, the burn-in artifact can be seen in high mas scoli exams, it is most often noted as a repeated image of the xray marker. 13

Case 4 14

Case 4 What is the image quality issue in this picture? A.) Use of Lap Shield B.) Room is not suitable for portable xray imaging C.) Source-to-image distance is too short D.) Portable xray tube should be under the table. 15

Case 4 What is the image quality issue in this picture? A.) Use of Lap Shield B.) Room is not suitable for portable xray imaging C.) Source-to-image distance is too short D.) Portable xray tube should be under the table. 16

Case 4 There is a ruler provided to help you determine the SID 17

Case 4 Key Concepts: 1.) Portable xray protocols have defined SID s, make sure you try to approximate the correct distance so the dose is appropriate and the image quality is consistent. 2.) Use the ruler provided by the system and pick the appropriate protocol (27, 40, or 72 ) 18

Case 5 19

Case 5 What is the image quality issue in this picture? A.) Improper exposure/technique B.) Detector burn-in or ghosting artifact C.) Electromagnetic interference artifact (EMI) D.) Artifact from a positioning device Case Courtesy of Judy Le, R.T. (CT) (R) 20

Case 5 What is the image quality issue in this picture? A.) Improper exposure/technique B.) Detector burn-in or ghosting artifact C.) Electromagnetic interference artifact (EMI) D.) Artifact from a positioning device Case Courtesy of Judy Le, R.T. (CT) (R) 21

Case 5 Key Concepts: 1.) Gel pillows are filled with a non-toxic water based gel that can provide the support needs of micro-preemie, preemie, and full-term infants 2.) Removal of gel pillows is highly recommended for any indication. Artifacts caused by gel pillows can mimic pathology, such as free air, which can lead to misinterpretation by Radiologists. 3.) These gel positioning devices can lead to additional x-rays to confirm artifact vs. pathology, resulting in increased radiation dose to the patient. Case Courtesy of Judy Le, R.T. (CT) (R) 22

Case 6 23

Case 6 What is the image quality issue in this picture? A.) Chest chair is too large B.) Baby should have arms down C.) Holding is never acceptable in pediatric radiology D.) Poor collimation 24

Case 6 What is the image quality issue in this picture? A.) Chest chair is too large B.) Baby should have arms down C.) Holding is never acceptable in pediatric radiology D.) Poor collimation 25

Case 6 Key Concepts: 1.) Poor collimation is a source of poor image quality in digitally processed images. 2.) Differences in collimation introduces variability in the source images and inconsistent results on processed images. 2.) Collimation is also very effective at reducing our patient s exposure (and it also reduces the exposure of anyone who is holding the child). Always collimate as much as reasonable! 26

Case 7 27

Case 7 What is the image quality issue in this picture? A.) Patient should be feet first first B.) Patient is off iso-center C.) Incorrect pediatric table insert in place D.) Incorrect position on the table in the head-to-foot direction 28

Case 7 What is the image quality issue in this picture? A.) Patient should be feet first first B.) Patient is off iso-center C.) Incorrect pediatric table insert in place D.) Incorrect position on the table in the head-to-foot direction 29

Case 7 Automatic dose control requires good isocentering. 30

Case 7 Magnification = SID/SOD 31

Z-axis dose modulation Bushberg JT, et al. The Essential Physics of Medical Imaging, Wolters Kluwer, Philadelphia, 3 rd Edition, 2012 32

Z-axis dose modulation Bushberg JT, et al. The Essential Physics of Medical Imaging, Wolters Kluwer, Philadelphia, 3 rd Edition, 2012 33

Case 7 Key Concepts: 1.) Use your lasers to ensure very good centering of the patient in the AP and RL directions. 2.) If you do notice you are off-isocenter, do not move your planning box to accommodate. Go back in the room and physically reposition your table and acquire another topogram. 34

Case 8 35

Case 8 What is the image quality issue in this picture? A.) Cosmetics containing glitter may lead to tissue heating and discomfort during MRI B.) Cosmetics containing glitter may lead to imaging artifacts in MRI C.) Glitter tattoos can lead to bright spots in the MRI image D.) Both A and B 36

Case 8 What is the image quality issue in this picture? A.) Cosmetics containing glitter may lead to tissue heating and discomfort during MRI B.) Cosmetics containing glitter may lead to imaging artifacts in MRI C.) Glitter tattoos can lead to bright spots in the MRI image D.) Both A and B 37

Case 8 38

Case 8 39

Case 8 40

41

Case 9 42

Case 9 What is the image quality issue in this picture? A.) Artifact due to patient motion B.) Iterative reconstruction artifact C.) Artifact due to contrast extravasation D.) Ring artifact from detector failure 43

Case 9 What is the image quality issue in this picture? A.) Artifact due to patient motion B.) Iterative reconstruction artifact C.) Artifact due to contrast extravasation D.) Ring artifact from detector failure 44

Case 9 Key Concepts: 1.) Contrast extravasation means that the scan quality will be compromised and possibly non-diagnostic. 2.) Always check with an attending before continuing a scan where extravasation is suspected. 45

Case 10 46

Case 10 What is the image quality issue in this picture? A.) Post-processing error B.) Improper exposure/technique C.) Electromagnetic interference artifact (EMI) D.) Artifact from a positioning device TG Sandridge, M Karastanovic, Radiologic Technology, January/February 2015, Volume 86, Number 3 47

Case 10 What is the image quality issue in this picture? A.) Post-processing error B.) Improper exposure/technique C.) Electromagnetic interference artifact (EMI) D.) Artifact from a positioning device TG Sandridge, M Karastanovic, Radiologic Technology, January/February 2015, Volume 86, Number 3 48

Case 10 Key Concepts: 1.) Some devices such as this thermoregulation unit can interfere electromagnetically with digital detectors. 2.) Electromagnetic interference artifacts often manifest as parallel lines and can be eliminated if the source of the interference can be removed or turned off. TG Sandridge, M Karastanovic, Radiologic Technology, January/February 2015, Volume 86, Number 3 49

Case 10 Thermoregulation Unit ON Thermoregulation Unit OFF TG Sandridge, M Karastanovic, Radiologic Technology, January/February 2015, Volume 86, Number 3 50

Case 11 51

Case 11 What is the hazard or compliance issue in this picture? 52

Case 11 Answer: Objects depressing the fluoroscopy foot-pedal! 53

Case 11 Key Concepts: 1.) Always remember to practice safety with foot-pedal operated devices. Keep them clear of equipment which can fall on to the pedal. Inhibit xrays if possible when the device is not in use. 2.) Become familiar with the beam-on alerts for your device, both visual and audio. 3.) Always wear your radiation badge in case of unexpected exposures. 54

Case 12 55

Case 12 What is the hazard or compliance issue in this picture? 56

Case 12 Answer: Someone ripped off the spacer cone! 57

Case 12 Answer: The spacer cone is there to protect the patient s skin. 58

Case 12 Key Concepts: 1.) Never remove the spacer cone. This is in place to ensure the entrance skin dose to the patient is not too high and is regulated by the State and Joint Commission. 2.) Remember that the xray radiation is highest directly at the tube output and decreases by 1/distance^2. Entrance skin dose is highest if anatomy is closer to tube output, so the spacer cone is there to reduce the likelihood of skin burns. 59

Case 13 60

Case 13 What is the hazard or compliance issue in this picture? 61

Case 13 Answer: This physician is entering the MRI room with his pager and stethoscope! 62

Case 13 Key Concepts: 1.) Most metal object MRI accidents occur because of untrained or absent-minded people entering with common items. 2.)As the most trained staff in the hospital, it s our job to always be vigilant with screening (even if it is a doctor or supervisor who is entering the MRI room). 63

Case 14 64

Case 14 What is the hazard or compliance issue in this picture? 65

Case 14 Answer: A magnet quench may cause vaporized helium (a white smoky substance) to fill the MRI scanner room. 66

Case 14 MRI scanners are built with a exhaust for vaporized helium. 67

Case 14 Example of vaporized helium exhaust from a quench. Failure of the exhaust system can lead to very high gas pressures in the scanner room. 68

Case 14 Key Concepts: 1.) MRI system use superconducting technology that must be kept at cryogenic temperatures. Any disturbance to this cryo system can lead to a quench. 2.) Quenches happen more often than you might think. Generally there is a audible loud bang, an immediate loss of field strength, and the room will fill with helium gas. 2.)Immediately evacuate all personnel from the area to avoid breathing in helium vapor and contact HTM. 69

Case 15 70

Case 15 What is the hazard or compliance issue in this picture? 71

Case 15 Answer: Drinking cups in the Nuclear Medicine Hotlab! 72

Case 15 Key Concepts: 1.) Eating, drinking, applying cosmetics or creams can introduce radio-isotopes to your body and lead to internal exposure. 2.) Refrain from these activities in all radioactive material work areas and avoid touching your eyes, nose, and lips. 73

Case 16 74

Case 16 What is the hazard or compliance issue in this picture? 75

Case 16 Answer: This staff member is standing in the highest radiation scatter zone! 76

Case 16 Scatter profile for CT scanners is forward along the table. 77

Case 16 The side of the CT scanner has lower scatter and is a preferable place to stand than next to the table. 78

Case 16 Key Concepts: 1.) The CT scanner tend to throw radiation forward and backward along the length of the patient bed. 2.) If possible, it s better for staff and family member to sneak around to the side of the scanner during the acquisition to reduce their own exposure. 79

Case 17 80

Case 17 What is the hazard or compliance issue in this picture? 81

Case 17 Answer: Staff member is using frontal apron but has back turned to source of radiation! 82

Frontal Drop Lead Apron Type 83

Skirt Vest Type Apron 84

Case 17 Key Concepts: 1.) Be careful if you are wearing a front apron that you are facing the source of radiation and do not get your body rotated. 2.) Wear a skirt/vest style apron in circumstances when you cannot always face the procedure (ie. if you are circulating). 3.) If you have no other option, wear two front lead aprons, one front-ward, the other back-ward to protect your organs from all sides. 85

Acknowledgements Jackie Choragwicki Merima Karastanovic Judy Le THANK YOU! Christina L. Sammet, Ph.D., DABR Medical Physicist Radiation and Laser Safety Officer 86