TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO

Similar documents
ZEISS LSM510META confocal manual

Training Guide for Carl Zeiss LSM 510 META Confocal Microscope

Microscope Confocal LSM510 META

LSM 510 Meta Training Notes

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling

LSM 510 Training Notes

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope

Zeiss 880 Training Notes Zen 2.3

Things to check before start-up.

Training Guide for Leica SP8 Confocal/Multiphoton Microscope

Maria Smedh, Centre for Cellular Imaging. Maria Smedh, Centre for Cellular Imaging

Practical work no. 3: Confocal Live Cell Microscopy

3D light microscopy techniques

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Operating Instructions for Zeiss LSM 510

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014

Zeiss 780 Training Notes

LSM 510 META in Chang Gung University

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s

Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev. Microscopy course, Michmoret Dec 2005

Internal Medicine Imaging Core Emory University Department of Medicine

Zeiss LSM 510 Multiphoton Confocal Microscope

Zeiss LSM 510 Multiphoton Confocal Microscope

Leica SP8 TCS Users Manual

LSM 710 Confocal Microscope Standard Operation Protocol

Quick Start Guide. Leica SP5 X

3. are adherent cells (ie. cells in suspension are too far away from the coverslip)

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging

OPERATING INSTRUCTIONS

Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report. Introduction and Background

Confocal Microscopy. Kristin Jensen

Point Spread Function. Confocal Laser Scanning Microscopy. Confocal Aperture. Optical aberrations. Alternative Scanning Microscopy

Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope

Cell Biology and Bioimaging Core

MULTIPHOTON MICROSCOPY. Matyas Molnar Dirk Pacholsky

Leica TCS SP8 Quick Start Guide

Training Guide for Carl Zeiss LSM 880 with AiryScan FAST

Basics of confocal imaging (part I)

Boulevard du Temple Daguerrotype (Paris,1838) a busy street? Nyquist sampling for movement

Supplemental Method Information Zeiss LSM710

Leica TCS SL Confocal Training. Neuroscience Imaging Core Staff. Core Director. Facility Manager

Leica Sp5 II Confocal User Guide

Microscopy from Carl Zeiss

LSM 780 Confocal Microscope Standard Operation Protocol

Non-Descanned FLIM Detection in Multiphoton Microscopes

Leica TCS SP8 Quick Start Guide

Resolution. Diffraction from apertures limits resolution. Rayleigh criterion θ Rayleigh = 1.22 λ/d 1 peak at 2 nd minimum. θ f D

Zeiss LSM 780 Protocol

Zeiss LSM 880 Protocol

Shreyash Tandon M.S. III Year

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.

CONFOCAL MICROSCOPE (Zeiss LSM 510 META v4.2)

Examination, TEN1, in courses SK2500/SK2501, Physics of Biomedical Microscopy,

BASICS OF CONFOCAL IMAGING (PART I)

The Zeiss AiryScan System, Confocal Four.

RENISHAW INVIA RAMAN SPECTROMETER

TRAINING MANUAL. Olympus FV1000

Invitation for a walk through microscopy. Sebastian Schuchmann Jörg Rösner

Contents. Introduction

3D light microscopy techniques

b. Turn the power switch and key to on position for blue laser.

Leica SP8 TCS Users Manual

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL

Introduction to light microscopy

You won t be able to measure the incident power precisely. The readout of the power would be lower than the real incident power.

CMI STANDARD OPERATING PROCEDURE. Fluoview 300 laser scanning confocal microscope

LSM 800 Confocal Microscope Standard Operation Protocol

Bio 407. Applied microscopy. Introduction into light microscopy. José María Mateos. Center for Microscopy and Image Analysis

1 Co Localization and Working flow with the lsm700

Microscope Confocal Sp2 Upright.

Zeiss LSM880 Operating Instructions. UTMB Optical Microscopy Core Jan. 16, 2018

Introduction to light microscopy

Leica SPEII confocal microscope. Short Manual

ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL

Light Microscopy. Upon completion of this lecture, the student should be able to:

LEICA TCS SP5 AOBS TANDEM USER MANUAL

Microscopy. Matti Hotokka Department of Physical Chemistry Åbo Akademi University

Leica SP8 Resonant Confocal. Quick-Start Guide

5/4/2015 INTRODUCTION TO LIGHT MICROSCOPY. Urs Ziegler MICROSCOPY WITH LIGHT. Image formation in a nutshell. Overview of techniques

ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide

VISUAL PHYSICS ONLINE DEPTH STUDY: ELECTRON MICROSCOPES

Guide to Confocal 5. Starting session

Nikon. King s College London. Imaging Centre. N-SIM guide NIKON IMAGING KING S COLLEGE LONDON

INTRODUCTION TO MICROSCOPY. Urs Ziegler THE PROBLEM

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software:

Confocal Laser Scanning Microscopy

Confocal 510 Tutorial

Introduction to light microscopy

Microscopic Structures

Title: Leica SP5 Confocal User Manual

Fundamentals of Light Microscopy II: Fluorescence, Deconvolution, Confocal, Multiphoton, Spectral microscopy. Integrated Microscopy Course

Imaging Retreat - UMASS Customized real-time confocal and 2-photon imaging

Multifluorescence The Crosstalk Problem and Its Solution

a) How big will that physical image of the cells be your camera sensor?

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL

Introduction to light microscopy

How to align your laser for two-photon imaging

長庚大學共軛焦顯微鏡課程 長庚大學共軛焦顯微鏡課程. Spot light 長庚大學

Olympus Confocal Microscope User Guide Last updated

Transcription:

TRAINING MANUAL Multiphoton Microscopy LSM 510 META-NLO September 2010

Multiphoton Microscopy Training Manual Multiphoton microscopy is only available on the LSM 510 META-NLO system. This system is equipped with the Chameleon laser, which is a Ti:Sa femtosecond laser. The pre-requisite for the multiphoton training is the Introduction to Confocal Microscopy training. Many of the steps for setting up and acquiring a multiphoton image is the same as for single photon imaging. Therefore this training only covers the steps that differ from the introductory training. Multiphoton Training includes: A. Introduction to Multiphoton Microscopy B. Start-up Procedure C. Turning on the Chameleon Laser D. Setting the Configuration E. Scan Parameters F. Laser Safety G. Shut-down Procedure Please refer to the LSM 510 META Training Manual for a more complete description of all of the steps.

A. INTRODUCTION - Multiphoton Microscopy Multiphoton microscopy is another type of laser scanning microscopy that also provides a single plane image, however multiphoton microscopy achieves this without using a pinhole. Instead, only the fluorescent molecules within the plane of focus are excited eliminating the need for the pinhole since there is no excitation and subsequent emission outside the focal plane. Excitation of only the fluorescent molecules in the focal plane is achieved by a femtosecond laser that sends photons of light with lower energy (longer wavelength) in rapid succession. Multiple photons of the lower energy light must be absorbed in order for the fluorochrome to reach the excited state and the focal plane is the only place where there is sufficient number of photons (in both space and time) for this to occur. Single Photon Multiphoton 2. Transition 2. Transition 1. Excitation 3. Emission 1. Excitation 3. Emission 1 photon 2 photons 1 photon with higher energy (Shorter wavelength) is required for excitation 2 photons with lower energy (Longer wavelength) is required for excitation Single Photon Multi Multiphoton Photon Although the wavelengths used for single and multiphoton excitation are different, the emission spectrum is the same. Emission Intensity (%) 100 Single photon excitation 75 50 25 Emission Intensity (%) 100 75 50 25 2 photon excitation 450 475 500 525 550 575 600 650 700 450 475 500 525 550 575 600 650 920 wavelength (nm) wavelength (nm) wavelength (nm)

A. INTRODUCTION Continued... Advantages of Multiphoton Microscopy: Longer wavelengths (Red or Near-Infrared light) Longer wavelength light is less biologically damaging Excitation is limited to the focal plane The only place where there is sufficient number of photons (in both space and time) is the focal plane therefore very little, if any, out-of-focus fluorescence occurs ( therefore a pinhole is not required) Photon Specimen Excitation is restricted to the focal plane due to photon crowding (In both space and time) Since excitation is limited to the focal plane, photobleaching is also limited to the focal plane Single Photon Excitation Multiphoton Excitation Coverslip } FOCAL PLANE Slide Photobleached Fluorochrome Photobleaching is limited to the focal plane with multiphoton excitation

A. INTRODUCTION Continued... Advantages of Multiphoton Microscopy Continued... Image thicker specimens Light Scatters Both single and multi-photon excitation and resulting fluorescence can be scattered in the specimen The extent of scattering depends on the specimen Generally the laser power is increased to compensate for scatter } FOCAL PLANE In both single and multi-photon microscopy the excitation laser is scattered in thick specimens. The laser intensity is increased to compensate for this scatter. Single Photon Multiphoton Emission in multiple planes with single photon excitation { FOCAL Out-of-focus emission scattered and passes through the pinhole to the detector FOCAL PLANE quality (contrast) of the image by collecting out-of focus fluorescence PLANE In-focus emission scattered and blocked by the pinhole in emission collected from the focal plane In-focus emission passed through the pinhole and is collected Excitation only occurs in the focal plane because the absorption of 2 photons at the same time only happens in the focal plane. Therefore, emission only originates from the focal plane Since there is no pinhole in multiphoton microscopy scattered emission (along with nonscattered) is able to reach the detector Detector Detector

A. INTRODUCTION Continued... Some Disadvantages or Limitations of Multiphoton Microscopy Less lateral and axial resolution When light from the various points of a specimen passes through the objective and is reconstituted as an image, the various points of the specimen appear in the image as small patterns (not points) known as Airy patterns. This is caused by diffraction or scattering of the light as it passes through the specimen and the circular back aperture of the objective Airy Pattern Resolution the minimum detectable distance between two closely space specimen points the limit at which 2 airy disks can be resolved into separate entities (Rayleigh Criterion) These points can be resolved 3 critical characteristics set the resolution limit of the microscope These 2 points cannot be resolved as separate points NA at a fixed, size of airy disk, resolution Numerical Aperture: the light cone size decreases with NA (at a fixed ) and produces a corresponding decrease in the size of the Airy disks ( resolution) Refractive Index: numerical aperture can be dramatically increased by immersion medium such as oil, glycerin or water Wavelength: an increase in the illumination wavelength at a fixed NA will result in an increase in Airy pattern ( resolution) multiphoton microscopy uses longer excitation compared to single photon (therefore multiphoton microscopy has resolution) Increasing illumination Increasing NA illumination size of airy disk resolution

A. INTRODUCTION Continued... Some Disadvantages or Limitations of Multiphoton Microscopy Continued... Excitation spectra difficult to predict and measure few sources of multiphoton excitation information available twice the wavelength of the single photon excitation may not necessarily be the most efficient for multiphoton excitation Single Photon Excitation Multiphoton Excitation FITC Rhodamine Absorption 400 500 600 700 800 900 1000 1100 Wavelength Thermal damage to specimen Water molecules absorb more energy from longer wavelengths of light (NIR and IR) compared to shorter wavelengths (VIS and UV) Heat is generated as water molecules absorb the energy from NIR and IR light. Since the excitation wavelength in multiphoton microscopy is in the NIR and IR range, thermal damage can be a problem as water molecules (both intracellular and Extracellular) in the sample absorb the excitation light generating heat

B. START-UP PROCEDURE - Multiphoton Microscopy 1. If the chiller is in Standby Mode, press the Standby/Run button to switch the chiller from o Standby Mode to Run Mode. (The chiller is set to ~25 C and the current temperature will be displayed.) NOTE: DO NOT OPERATE THE CHAMELEON LASER WITH THE CHILLER IN STANDBY MODE!!!! THE CHILLER MUST ALWAYS BE TURNED ON WHEN USING THE CHAMELEON LASER!!!! 2. Turn on the mercury burner 3. Turn both the PC and Components buttons on the REMOTE CONTROL switch to the ON position 4. Turn on the computer 5. Log into user profile - press Ctrl, Alt and Delete simultaneously - enter Userid and password 6. Double click on the LSM 510 icon start the LSM software 7. In the LSM 510 Switchboard window select Scan New Images and then Start Expert Mode The system will go through an instrument initialization and then the main LSM Toolbar will appear

C. TURNING ON CHAMELEON LASER - Multiphoton Microscopy Before turning on the Chameleon laser: make sure the chiller is in Run Mode (not in Standby Mode) THE CHILLER MUST BE ON AT ALL TIMES WHEN THE CHAMELEON LASER IS TURNED ON!!!!! make sure the key on the Chameleon power supply (under the table) is in the On position If the chiller and Chameleon is in Standby Mode, go to step #1 in the Multiphoton Microscopy Start-up Procedure (Section A) 1. click on Chameleon to select the Chameleon laser 2. Click the On button to turn the laser on Mode Locked 3. Click on Modify to open the Laser Modify Control window 4. Enter the excitation wavelength (720-930 nm only) for the Chameleon laser and then click on Store. Wait a few seconds until Mode Lock is re-established Close the laser control window if all of the lasers needed are turned on DO NOT adjust the fine tuning AO-Frequency!!!

D. CONFIGURATIONS - Multiphoton Microscopy All of the steps for setting the configurations are the same for both single photon and multiphoton microscopy. There are only a few things to keep in mind. multiphoton can be used in combination with single photon (ie. Argon & HeNe lasers) use as a single or multi-track use Ch S, Ch2 or Ch3 always use a KP (short pass) main dichroic beam splitter always use either a BP or KP emission filter (never use a LP filter) Refer to the General Confocal Microscopy Training Manual for a more detailed description and instruction on the configurations 1. Select Channel Mode 2. Select Single Track or Multi Track 3. Click on Excitation to open the Laser Control Window 4. Activate the excitation wavelengths (check box) and setting of excitation intensities (slider) NEVER SET THE CHAMELEON LASER >10%!!!! * * 5. Click on main dichroic beam splitter (HFT) to open the list of beam splitters and select the appropriate dichroic for the excitation laser line(s) you have activated. The main dichroic beam splitter separates excitation from emission NT 80/20 = no filter * HFT = main dichroic beam splitter KP = short pass (LSM 510 META - NLO only) - use with the Chameleon laser Always use a KP beam splitter with the Chameleon laser

D. CONFIGURATIONS Continued... 6. Click on the first secondary beam splitter (NFT) to open the list of options for this position. This beam splitter directs the appropriate wavelengths of emission to Channel S and/or Channel 2 and 3. Select the appropriate mirror or filter. None = all emission will pass through to Channel S Mirror = all emission will be directed to Channel 2 and 3 Plate = all emission will pass through to Channel S NFT = secondary dichroic beam splitter KP = short pass filter 7. Click on the second secondary beam splitter to open the list or options for this position. This beam splitter directs the appropriate wavelengths of emission to Channel 2 and/or Channel 3. Select the appropriate mirror or filter. Mirror = all emission will be directed to Channel 2 NFT 490 = all emission with wavelengths less than 490 nm will be directed to Channel 2 and emission with wavelengths greater than 490 nm will pass through to Channel 3 NFT 545 = all emission with wavelengths less than 545 nm will be directed to Channel 2 and emission with wavelengths greater than 545 nm will pass through to Channel 3 BG 39 = all emission will pass through to Channel 3

D. CONFIGURATIONS Continued... 8. Click on Ch S and select the adjust the sliders or click on the Emission Filter for Channel 2 and Channel 3 to open the list of emission filters available. 9. Select the appropriate emission filter and then check the box to turn on the corresponding channel. Use either a KP or BP filter for multiphoton microscopy. Do not use a LP filter. Band Pass (BP) - permit only wavelengths of emission within the range indicated to pass through to the PMT - Band pass filters can be used with all of the lasers Short Pass (KP) - permit emission with wavelengths below the threshold indicated to pass through to the PMT Channel 2 Emission Filters Channel 3 Emission Filters X X - Short pass filters are used only with the Chameleon laser and not the Argon or HeNe lasers. Long Pass (LP) - Long pass filters permit emission with wavelengths above the threshold indicated to pass through to the PMT - Long pass filters can be used with the Argon and HeNe lasers, but should not be used with the Chameleon laser 10. Click on the colour box to o select the appropriate pseudo-colour for that particular channel. 11. Add another track and repeat steps #1-11 if necessary.

E. INITIAL SCAN PARAMETERS Set the initial Scan Parameters the same as for single photon with the exception of setting the pinhole: When using the Chameleon laser click on Max to open the pinhole multiphoton microscopy does not require a pinhole emission only occurs when 2 or more photons hit the fluorophore at the same time this only occurs within the focal spot F. LASER SAFETY Before you put your sample on the stage there are some very important laser safety precautions that need to be taken. But first, a little about the hazards and classification of lasers: Lasers are classified into 4 groups according to the maximum power or energy of the beam, wavelength output, and pulse repetition Generally, the classification indicates the capability of the laser or laser system to produce injury with higher class numbers indicating a greater potential hazard. Class 1 - lasers or laser systems that do not pose a hazard under normal operating conditions Class 2 - low power visible lasers that are incapable of causing eye injury unless viewed directly for extended periods of time Class 3A - lasers or laser systems that have a low risk of injury but normally would not injure the eye unless directly viewed through binoculars or similar optical devices Class 3B - lasers or laser systems that can produce accidental injuries to the eye from viewing the direct beam or a specular reflection, but normally is not a hazard from diffuse reflection (unless viewed through an optical instrument) Class 4 - lasers that have eye and skin hazards from the direct beam, specular reflections and diffuse reflections. This is a very brief description of the lasers classes. Please refer to the UWO Laser Safety Manual for a more detailed description.

F. LASER SAFETY Continued... Specifically for our systems, the Argon, HeNe and 405nm Laser Diode are all Class 3 lasers whereas the CHAMELEON LASER IS A CLASS 4 LASER!!!! When enclosed within the microscope system, all of the lasers pose very little risk and are therefore considered Class 1. Regardless, there still very important safety steps that must be followed when using the Chameleon laser. Since the Argon, HeNe and 405nm Laser Diode lasers emit in the visible range, the laser beam can be seen when running. The Chameleon Laser however emits in the near infrared range, THEREFORE YOU CANNOT SEE THE LASER BEAM!! Class 4 lasers can damage your eyes and skin from the direct beam and REFLECTED BEAM!!! NEVER MANIPULATE ADJUST YOUR SLIDE WITH THE LASER ON to avoid potentially reflecting the laser beam into your eyes and/or exposing your skin to the laser beam. There are 3 main steps to take before you do any manipulation of your sample on the stage to ensure the laser is turned off: 1. Click Stop in the Scan Control Window 2. Click Vis in the main toolbar 3. Tilt the light carrier back to disconnect the laser circuit. There are also goggles and shields available for your protection

G. SHUT-DOWN PROCEDURE 1. Close all open windows, except the main LSM Toolbar 2. Open the Laser Control box Cool the Argon Laser on Standby highlight the argon laser click on Standby and then Off wait until cooling in the status bar switches to connected IMPORTANT: the argon laser must be cooled before turning the system off Turn the NeNe lasers off highlight the appropriate laser click Off Turn off the Chameleon laser highlight the Chameleon laser click Off turn key on the Chameleon power supply (under the table) to Standby 3. Click on File in the main LSM Toolbar and then Exit. Click Exit in the main switchboard. ( A window will open reminding you the lasers are on and do not turn the system off. Click OK) 4. Log off the computer (ctl + alt + del) and record use in the log sheet. 5. CLEAN THE OBJECTIVE!!!! if not cleaned properly. These objectives are very expensive and are damaged very easily To clean: wipe off any oil on the objective with lens paper use lens paper damp with lens cleaner to clean the objective dry the objective with a clean and dry piece of lens paper IF YOU DO NOT CLEAN THE OBJECTIVES PROPERLY, YOUR MICROSCOPE PRIVILEGES WILL BE TAKEN AWAY.