Low Value Impedance Measurement using the Voltage / Current Method

Similar documents
Measure Low Value Impedance Current Shunt Impedance

Power Supply Rejection Ratio Measurement

DC/DC Converter Stability Measurement

Equivalent Circuit Determination of Quartz Crystals

DC Biased Impedance Measurements MOSFET

DC/DC Converter Stability Measurement

Audio Amplifier Frequency Response

Measuring Power Line Impedance

Contactless RFID Tag Measurements

Transformer modelling

Loop Gain Measurement

DC/DC Converter Stability Measurement

Equivalent Circuit Determination of Quartz Crystals

Solar Cell Impedance Measurement using the Bode 100

Input Impedance Measurements for Stable Input-Filter Design

Invasive and Non-Invasive Stability Measurements

Making Invasive and Non-Invasive Stability Measurements

DC Biased Impedance Measurement

Measuring Impedance with the Bode 100. OMICRON Lab Webinar Nov. 2014

Power Supply Rejection Ratio Measurement

Power Supply Rejection Ratio Measurement

Battery Impedance Measurement

Power Supply Rejection Ratio Measurement

Smart Measurement Solutions. Bode 100. User Manual

Bode 100. User Manual

Measurement of the equivalent circuit of quartz crystals

Passive Component Analysis. OMICRON Lab Webinar Nov. 2015

Keysight E5061B-3L3/3L4/3L5 LF-RF Network Analyzer with Option 005 Impedance Analysis Function

Bode 100. User Manual. Smart Measurement Solutions

Exclusive Technology Feature. An Accurate Method For Measuring Capacitor ESL. ISSUE: April by Steve Sandler, Picotest, Phoenix, Ariz.

Bode 100. User Manual. Smart Measurement Solutions

How to Measure LDO PSRR

Application Note. Calibration of 7000 Series Precision LCR Meters

Lab 3: AC Low pass filters (version 1.3)

Experiment 8 Frequency Response

Agilent Accessories Selection Guide For Impedance Measurements. December 2008

Lab 9 Frequency Domain

Current Probe. Inspector Data Sheet. Low-noise, high quality measurement signal for side channel acquisition on embedded devices.

Opamp stability using non-invasive methods

The 2-Port Shunt-Through Measurement and the Inherent Ground Loop

Keysight Measuring High Impedance Sources Using the U8903B Audio Analyzer. Application Note

EE2210 Laboratory Project 1 Fall 2013 Function Generator and Oscilloscope

2. BAND-PASS NOISE MEASUREMENTS

ITT Technical Institute. ET275 Electronic Communications Systems I Onsite Course SYLLABUS

ET275P Electronic Communications Systems I [Onsite]

Accessories Selection Guide For Impedance Measurements. April 2005

S-Parameter Measurements with the Bode 100

Keysight E5061B-3L3/3L4/3L5 LF-RF Network Analyzer with Option 005 Impedance Analysis Function. Data Sheet

VVM measurement with E5061B for replacing 8508A vector voltmeter. May 2013 Agilent Technologies

ECE3204 D2015 Lab 1. See suggested breadboard configuration on following page!

Core Technology Group Application Note 1 AN-1

Evaluating DC-DC Converters and PDN with the E5061B LF-RF Network Analyzer. Application Note

Power Added Efficiency Measurement with R&S ZNB/ R&S ZVA

Core Technology Group Application Note 6 AN-6

A Walk Through the MSA Software Vector Network Analyzer Transmission Mode 12/18/09

LoadSlammer User Guide LS50 and LS1000

ECE 4670 Spring 2014 Lab 1 Linear System Characteristics

EXPERIMENT NUMBER 10 TRANSIENT ANALYSIS USING PSPICE

Automotive Scope/GMM/OBD II Code Reader

ECE4902 Lab 5 Simulation. Simulation. Export data for use in other software tools (e.g. MATLAB or excel) to compare measured data with simulation

Lab 4: Analysis of the Stereo Amplifier

Application Note 5525

ECE 2274 Lab 2. Your calculator will have a setting that will automatically generate the correct format.

Experiment 8: An AC Circuit

ECE 2274 Lab 2 (Network Theorems)

Standalone High Accuracy Transformer Analysis. Basic 0.02dB with class leading high frequency performance. RS232, USB and LAN as standard

PSM Soft. Features and Functions January PC Software Guide. Getting connected and Communication

12. Output Ripple Attenuator Module (MicroRAM )

Calsytech, # 38 North Mada Street Nandambakkam, Chennai, Tamil Nadu. Discipline Electro-Technical Calibration Issue Date

LAB 8: Activity P52: LRC Circuit

Grundlagen der Impedanzmessung

When you have completed this exercise, you will be able to determine the frequency response of an

Lab #5 Steady State Power Analysis

Experiment 1: Instrument Familiarization (8/28/06)

Multi-function Gain-Phase Analyzer (Frequency Response Analyzer) Model 2505

PHYSICS 330 LAB Operational Amplifier Frequency Response

Resonant Frequency of the LRC Circuit (Power Output, Voltage Sensor)

Integrators, differentiators, and simple filters

Circuit Characterization with the Agilent 8714 VNA


EENG-201 Experiment # 4: Function Generator, Oscilloscope

Homework Assignment 01

500 khz / 1 MHz Precision LCR Meter Models 894 & 895

Figure AC circuit to be analyzed.

AN-742 APPLICATION NOTE


Contents. CALIBRATION PROCEDURE NI PXIe-5668R 14 GHz and 26.5 GHz Signal Analyzer

Activity P52: LRC Circuit (Voltage Sensor)

LMV225/LMV226/LMV228 RF Power Detector for CDMA and WCDMA

ECE 231 Laboratory Exercise 6 Frequency / Time Response of RL and RC Circuits

Agilent 4285A Precision LCR Meter

Agilent 4294A Precision Impedance Analyzer, 40 Hz to 110 MHz. Configuration Guide

Week 8 AM Modulation and the AM Receiver

MAX2720/MAX2721 Evaluation Kits

Laboratory 4: Amplification, Impedance, and Frequency Response

Signal Injectors. Documentation. Version 1.00, October, Picotest Corp. All Rights Reserved.

Keysight Technologies Accessories Selection Guide For Impedance Measurements. Selection Guide

EXTEND YOUR REACH. Copper Mountain Technologies USB VNAs. S-parameter measurement solutions from 9 khz to 110 GHz Measured parameters from S 11

P a g e 1 ST985. TDR Cable Analyzer Instruction Manual. Analog Arts Inc.

Experiment P45: LRC Circuit (Power Amplifier, Voltage Sensor)

Transcription:

Low Value Impedance Measurement using the Voltage / Current Method By Florian Hämmerle & Tobias Schuster 2017 Omicron Lab V2.2 Visit www.omicron-lab.com for more information. Contact support@omicron-lab.com for technical support.

Page 2 of 12 Table of Contents 1 Introduction... 3 2 Measurement... 4 2.1 Measurement Setup... 4 2.2 Device Configuration... 6 2.3 Calibration... 7 2.4 Measurement Result... 8 2.5 Result Verification... 9 3 Conclusion... 11 Note: Basic procedures such as setting-up, adjusting and calibrating the Bode 100 are described in the Bode 100 user manual. You can download the Bode 100 user manual at www.omicron-lab.com/bode-100/downloads#3 Note: All measurements in this application note have been performed with the Bode Analyzer Suite V3.00. Use this version or a higher version to perform the measurements shown in this document. You can download the latest version at www.omicron-lab.com/bode-100/downloads

Page 3 of 12 1 Introduction The impedance of current sensing devices is generally very low, in the range of several milliohms. Measuring the impedance of such a part can be a challenging task. This document shows how to measure a 20 mω SMD resistor, using the voltage / current measurement method of Bode 100. The DUT 1 resistor is displayed in the picture below (resistor in red frame). Figure 1: 20 mω ± 1 % SMD resistor The 10 Ω resistor is used as current measurement resistor or reference resistor as explained in the following section. 1 Device Under Test

Page 4 of 12 2 Measurement 2.1 Measurement Setup The following measurement setup uses a reference resistor to measure the impedance of the DUT. Figure 2: Measurement setup Channel 1 measures the voltage drop V 1 along the reference resistor which is proportional to the current flowing through the reference resistor. The same current also flows through the DUT. The second input channel measures the voltage drop at the DUT (V 2 ). The impedance Z shunt of the device under test then equals the measured gain. Gain = CH2 CH1 = V 2 = V 2 V 1 I 10Ω = Z shunt 10 ( 1 ) Note: The ground path resistance of the channel 1 and 2 connections should be as low as possible to achieve accurate results! To do so, the test pins and the connections itself are soldered together in this measurement example. Note: The 10 Ω reference resistor was chosen because it shows resistive behavior up to a frequency of 1 MHz. When using a smaller value resistors the parasitic inductance lowers the usable frequency range. The following pictures show the used measurement setup. The reference resistor is directly soldered to the pin of the shunt. This point is the common ground point for the two input channels. Soldered test pins enable the use of clips for the connection of the Bode 100.

Page 5 of 12 The B-WIT and the Bode 100 are connected using short BNC to clip leads as shown in the picture below. Figure 3: Measurement Setup CH 1 B-WIT 100 CH 2 Figure 4: DUT and reference resistor with connections

Page 6 of 12 2.2 Device Configuration The measurement is performed as a Voltage/Current measurement type. Figure 5: Start menu The following settings are applied: Start Frequency: Stop Frequency: Sweep Mode: Number of Points: Level: Attenuator 1: Attenuator 2: Receiver Bandwidth: 10 Hz 1 MHz Logarithmic 201 or more 13 dbm 10 db 0 db 10 Hz Since a 10 Ω current sense resistor is used, which delivers 10 V / 1 A, the Probe 1 ratio has to be set to 1:-10 to get a correct impedance and inductance reading. 1:-10 instead of 1:+10 has to be set since we measure the current in the opposite direction. If this setting is positive, the inductance measurement would be negative. The probe ratio can be set in the Hardware Setup window. Figure 6: hardware setup button

Page 7 of 12 Figure 7: Hardware setup window 2.3 Calibration In this measurement, a calibration is not mandatory. If we set up the measurement like shown in Figure 8, we can check if a calibration is required. Both probes are applied to the current sense resistor and pick up the same signal. The result must be a flat line at 10 Ω since we have set the probe of CH1 to 1:-10 and a 10 Ω resistor is measured. Figure 8: Connection during THRU calibration & for calibration check

Page 8 of 12 2.4 Measurement Result After applying the settings and connecting the probes like in Figure 2, the measurement can be started by clicking on the single sweep button. Figure 9: measurement buttons The following graph shows the measurement results. Figure 10: Shunt impedance The impedance of the measured current sense resistor equals 20.1 mω below a frequency of approximately 30 khz. Above this frequency, the self-inductance of the shunt starts to influence the impedance magnitude. At 1 MHz, the impedance equals Z 40.5 mω. The inductance can be directly measured with the Bode 100. For this, we change the Trace 1 setting Format from Magnitude to Ls. Figure 11: Settings Trace 1

Page 9 of 12 The inductance can also be calculated with the imaginary value (Trace 1 set to Imaginary and Cursor on 1 MHz = 37,09 mω) which leads with the equation below, to the following result: L = 5,9 nh L = Imag(Z) 2 π f ( 2 ) This result equals the measurement as can be seen below where the cursor is set to 1 MHz. Figure 12: Measurement result 2.5 Result Verification To verify the measurement results, we perform a measurement without the Bode 100 by using standard multi-meters. Therefore, the DUT must be connected to two multi-meters and one power supply as shown below. The power supply is configured to drive 1 A thru the device under test. By measuring the voltage and current, we can calculate the DC value of the resistor as shown below: R = U I = 21.67mV 1A = 21.67 mω ( 3 )

Page 10 of 12 Figure 13: Measurement Setup Multi-meter Figure 14: Multi-meter measurement setup (from left: Ampere-meter, Power Supply and Voltmeter)

Page 11 of 12 3 Conclusion We have demonstrated how very low impedance values can be measured using the Bode 100 and additional accessories like the B-WIT 100 injection transformer. Impedance values down to several milliohms can be measured over a wide frequency range. This enables an easy way to estimate e.g. the self-inductance of a current sense resistor or the selfinductance of chip capacitors at high frequencies. If you need to measure low value impedance at higher frequencies, we recommend using the Shunt- Thru method.

Page 12 of 12 OMICRON Lab is a division of OMICRON electronics specialized in providing Smart Measurement Solutions to professionals such as scientists, engineers and teachers engaged in the field of electronics. It simplifies measurement tasks and provides its customers with more time to focus on their real business. OMICRON Lab was established in 2006 and is meanwhile serving customers in more than 40 countries. Offices in America, Europe, East Asia and an international network of distributors enable a fast and extraordinary customer support. OMICRON Lab products stand for high quality offered at the best price/value ratio on the market. The products' reliability and ease of use guarantee trouble-free operation. Close customer relationship and more than 25 years in-house experience enable the development of innovative products close to the field. Europe, Middle East, Africa OMICRON electronics GmbH Phone: +43 59495 Fax: +43 59495 9999 Asia Pacific OMICRON electronics Asia Limited Phone: +852 3767 5500 Fax: +852 3767 5400 Americas OMICRON electronics Corp. USA Phone: +1 713 830-4660 Fax: +1 713 830-4661 info@omicron-lab.com www.omicron-lab.com