TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions

Similar documents
Automated dose control in multi-slice CT. Nicholas Keat Formerly ImPACT, St George's Hospital, London

diagnostic examination

1. Patient size AEC. Large Patient High ma. Small Patient Low ma

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

Image Quality and Dose. Image Quality and Dose. Image Quality and Dose Issues in MSCT. Scanner parameters affecting IQ and Dose

QC Testing for Computed Tomography (CT) Scanner

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D.

Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) MEDIX VOL. 61 P.8 P.11

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC?

Iterative Reconstruction

TORNIER BLUEPRINT. 3D Planning + PSI SCAN PROTOCOL

Iterative Reconstruction in Image Space. Answers for life.

Translating Protocols Between Scanner Manufacturer and Model

QC by the MPE in Belgium

The disclaimer on page 1 is an integral part of this document. Copyright November 30, 2017 by AAPM. All rights reserved.

NeuViz 16 Computed Tomography. Elevating routine imaging for exceptional results

Wide beam CT dosimetry. Elly Castellano

Diagnostic X-Ray Shielding

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

Detector technology in simultaneous spectral imaging

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Wide-Detector CT for TAVR Planning:

SOMATOM Esprit A Bundle of Energy

160-slice CT SCANNER / New Standard for the Future

Software and Hardware in CCTA. Elly Castellano PhD

CT Basics: Data Acquisition Module 3

Overview of Safety Code 35

COMPUTED TOMOGRAPHY 1

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

abc MHRA Philips Mx8000 IDT CT scanner technical evaluation September 2004 Best choice best practice nww.medical-devices.nhs.

SAFIRE. Sinogram Affirmed Iterative Reconstruction. Answers for life.

Digital radiography (DR) post processing techniques for pediatric radiology

Advanced Noise Reduction Processing for X-ray CT System with Iterative Processing. Koichi Hirokawa MEDIX VOL. 56 P.43 P.46

Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck

CT Basics: Equipment and Instrumentation Module 2

Computed Tomography. The Fundamentals of... THE FUNDAMENTALS OF... Jason H. Launders, MSc. Current Technology

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Maximum Performance, Minimum Space

Radiation Dose Modulation. the Multidetector CT Era: From Basics to Practice 1

CT Basics: Image Quality Module 6

Reducing Radiation Exposure from Survey CT Scans

T h e P h a n t o m L a b o r a t o r y

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Authors: Cabral, Ricardo 1 ; Carvoeiras, Pedro 2 ; Fatana, João 2, ; Alves, Rita 1. 1 Centro Hospitalar Lisboa Norte - Hospital de Santa Maria; 2

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Test Equipment for Radiology and CT Quality Control Contents

PD233: Design of Biomedical Devices and Systems

Model Based Iterative Reconstructions represent a paradigm shift - Imaging with almost no noise

An Activity in Computed Tomography

Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations

A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup

TISSUE EQUIVALENT PHANTOMS FOR EVALUATING IN-PLANE TUBE CURRENT MODULATED CT DOSE AND IMAGE QUALITY

Nuclear Associates , , CT Head and Body Dose Phantom

2D, 3D CT Intervention, and CT Fluoroscopy

Electronic Noise in CT Detectors: Impact on Image Noise and Artifacts

Y11-DR Digital Radiography (DR) Image Quality

Optimized CT metal artifact reduction using the Metal Deletion Technique (MDT)

CT Scanner Dose Survey

Improvement of CT image quality with iterative reconstruction idose4

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS

Quantitation of clinical feedback on image quality differences between two CT scanner models

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity)

An Activity in Computed Tomography

Teaching Digital Radiography and Fluoroscopic Radiation Protection

2012 :15th SESSION of ESMP

Maximizing clinical outcomes

Q3D. Speak to a 3D Specialist. CBCT 3D / Panoramic Imaging GENERAL DIMENSIONS. Suni Imaging Product Lines GET.

Redefining Ergonomics

Dose Reduction in Helical CT: Dynamically Adjustable z-axis X-Ray Beam Collimation

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Multi-Access Biplane Lab

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development

X-RAYS - NO UNAUTHORISED ENTRY

Yinsheng Li 1, Peter Bannas 2, M.D., Perry Pickhardt M.D. 2, Meghan Lubner M.D. 2, Ke Li Ph.D. 1,2, and Guang-Hong Chen Ph.D. 1,2

January 2012 February 1, Release Date: Expiration Date:

PET: New Technologies & Applications, Including Oncology

Sarah Hughes, MS, DABR Radiation Safety Officer

Ludlum Medical Physics

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT)

Improved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques

Mammography: Physics of Imaging

of sufficient quality and quantity

FOREWORD. Acknowledgements

COCIR SELF-REGULATORY INITIATIVE FOR MEDICAL IMAGING EQUIPMENT COMPUTED TOMOGRAPHY MEASUREMENT OF ENERGY CONSUMPTION

Related topics Beam hardening, cupping effect, Beam hardening correction, metal artefacts, photon starvation

Features and Weaknesses of Phantoms for CR/DR System Testing

SOMATOM Sensation 4 Computed Tomography System for Multislice Spiral Scanning

PET/CT Instrumentation Basics

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

Nuclear Associates

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

Invisible sophistication. Visible simplicity. CS Welcome to the simplicity of compact panoramic imaging

1. Queries are issued to the image archive for information about computed tomographic (CT)

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

ELECTRONIC CONTROL CONCEPTS 160 Partition Street Saugerties, NY or local phone

Data. Take the Lead in CT SOMATOM Sensation 64

DIAGNOSTIC ACCREDITATION PROGRAM. Radiology and CT Quality Control Procedures Workbook

RADspeed Pro. EDGEpackage C501-E041C

Transcription:

CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions TOPICS: Computed Tomography Quick Overview CT Dosimetry Effects of CT Protocols on Image quality and Dose Importance of Understanding AEC in different CT scanners Frank N. Ranallo, Ph.D. Associate Professor of Medical Physics & Radiology University of Wisconsin School of Medicine & Public Health 1 Optimization of CT Scan Techniques for Dose & Image Quality Tailored to the Patient Size, Anatomy, and Clinical Task 2 Evolution to Helical/ Spiral CT Scanners Single Slice Helical/ Spiral CT Evolution to Multislice Scanners 2, 4, 8, 16, 64,? Data Acquisition 3 4 1

Evolution to Multislice Scanners 2, 4, 8, 16, 64,? Evolution to Multislice Scanners 2, 4, 8, 16, 64,? Definition of Pitch for Single-slice Helical / Spiral Scanning Pitch = Table travel per 360 tube rotation Nominal Slice Thickness Definition of Pitch for Multi-slice Helical / Spiral Scanning Pitch = Table travel per 360 tube rotation Total collimation width of all simultaneously collected slices 5 6 CT Protocols Scan & Reconstruction Parameters: From Recon 2: Sa & Co Reformat: Ave., 4.0 mm thick & Pediatric Routine Abdomen/ Pelvis 2.0 mm interval Patient Size: AP + Lateral 0-26 cm 27 31 cm 32 37 cm 38 43 cm 44 55 cm Approximate Age Newborn 6 mo 2.5 yrs 3 7 yrs 8 12 yrs 13-18 yrs Protocol Color Pink Red/Purple Yellow/White Blue/Orange Green/Black Series 1 - Scout Scout 1: kv / ma 80 / 10 80 / 10 80 / 10 80 / 10 100 / 10 Scout Plane 180 180 180 180 180 Scout 2: kv / ma 80 / 40 80 / 40 80 / 40 80 / 40 100 / 40 Scout Plane 90 90 90 90 90 WW/WL for Scout 600/50 600/50 600/50 600/50 600/50 Series 2 - Smart Prep Monitor Phase ma 20 40 40 40 40 Monitoring Delay (sec) 20.0 20.0 25.0 25.0 30.0 Monitoring ISD (sec) 2.0 2.0 2.0 2.0 2.0 Enhancement Threshold 50 50 50 50 50 Diagnostic Delay (sec) Min. Delay Min. Delay Min. Delay Min. Delay Min. Delay 7 CT Protocols Patient Size: 0-26 cm 27 31 cm 32 37 cm 38 43 cm 44 55 cm AP + Lateral Approximate Age Newborn 6 mo 2.5 yrs 3 7 yrs 8 12 yrs 13-18 yrs Protocol Color Pink Red/Purple Yellow/White Blue/Orange Green/Black Series 2 - Helical Scan Scan Type Helical Helical Helical Helical Helical Beam Collimation (mm) 20 20 20 40 40 Detector Rows 32 32 32 64 64 Detector Configuration 32 x 0.625 32 x 0.625 32 x 0.625 64 x 0.625 64 x 0.625 Scan FOV Small Body Small Body Small Body Small Body Medium Body Pitch 1.375 1.375 1.375 1.375 1.375 Speed (mm/rot) 27.5 27.5 27.5 55 55 Rotation Time (sec) 0.4 0.4 0.4 0.6 0.7 kv 80 80 80 80 100 Smart ma or Manual ma Smart ma Smart ma Smart ma Smart ma Smart ma Smart ma min/max ma Range 15-200 25-300 40-540 40-610 60-760 Noise Index 12 14 15.5 17.5 17 (Manual ma [Ref w/ ASiR]) (140) (210) (380) (430) (530) Dose Reduction Guidance (%) 0 0 0 0 0 Slice Thickness (mm) 3.75 3.75 3.75 3.75 5.0 Interval (mm) 2.00 2.00 2.00 2.00 3.0 8 2

CT Protocols Patient Size: 0-26 cm 27 31 cm 32 37 cm 38 43 cm 44 55 cm AP + Lateral Approximate Age Newborn 6 mo 2.5 yrs 3 7 yrs 8 12 yrs 13-18 yrs Protocol Color Pink Red/Purple Yellow/White Blue/Orange Green/Black Recon 1: DFOV 20 20 20 25 25 Recon Type Detail Detail Detail Detail Standard WW/ WL 550/100 550/100 550/100 550/100 400/60 Recon Option Plus Plus Plus Plus Plus ASiR Setup Slice 40% Slice 40% Slice 40% Slice 40% Slice 40% Recon 2: DFOV 20 20 20 25 25 Recon Type Detail Detail Detail Detail Detail WW/ WL 500/80 500/80 500/80 500/80 400/60 Recon Option Plus Plus Plus Plus Plus Recon Option IQ Enhance IQ Enhance IQ Enhance IQ Enhance IQ Enhance ASiR Setup Slice 40% Slice 40% Slice 40% Slice 40% Slice 40% Slice Thickness (mm) 1.25 1.25 1.25 1.25 1.25 Interval (mm) 0.625 0.625 0.625 0.625 0.625 9 Dose in Computed Tomography 10 Radiation scares people Skin Damage from CT All these skin burns/ hair loss were due to perfusion CT scans for stroke diagnosis these scans were performed incorrectly at about 8 or more times the proper dose. 11 12 3

RADIATION UNITS Absorbed Dose Units: Equivalent Dose Units: New SI Units Old Conventional Units gray (Gy) or rad (r) sievert (Sv) or rem Effective Dose Units: sievert (Sv) or rem 13 RADIATION UNITS Absorbed Dose gray (Gy) or rad [1 Gy = 100 rad] Energy absorbed per unit mass of tissue Important for skin effects/ effects on specific organs Equivalent Dose sievert (Sv) or rem [1 Sv = 100 rem] Equivalent dose = absorbed dose x radiation weighting factor (w R ) 14 RADIATION UNITS Effective Dose CTDI Phantoms The concept of effective dose takes into account the risk to the person exposed to radiation that is not uniform over the entire body. Different organs have different sensitivities to radiation This is expressed by the tissue weighting factor: w T 15 16 4

CTDI CTDI 100 measured at the center of the phantom is called CTDI 100 (center) CTDI 100 measured near the surface of the phantom is called CTDI 100 (surface) or CTDI 100 (peripheral) CTDI CTDI w is a weighted value of CTDI 100 that attempts to give the average volume dose in the central slices of an extended scan length and phantom (again, for contiguous axial slices or for helical scanning at a pitch of 1). It is defined as: CTDI w = 1/3 CTDI 100 (center) + 2/3 CTDI 100 (surface) 17 18 CTDI If you then take into account the effect of pitch on dose for a helical/ spiral scan then you have another version of CTDI: CTDI vol = CTDI w / Pitch This is an approximation that attempts to give the average volume dose in the central slices of an extended scan length and phantom. DLP A final dosimetry measure is the Dose Length Product (DLP) which is defined as: DLP = CTDI vol Scan Length and has units of mgy cm CTDI vol and DLP are the radiation units provided by the CT Scanner. 19 20 5

DLP Estimates of a patient s effective dose (E) can be derived from the values of DLP for an examination. Use the following equation containing a coefficient k appropriate to the examination: E = k DLP 21 Effective Dose from DLP Values of k for Adult scans: Region of Body k (msv / mgy cm) Phantom Head 0.0023 0.0021 16 cm Head & Neck 0.0031 16 cm Neck 0.0054 0.0059 32 cm Chest 0.017 0.014 32 cm Abdomen 0.015 0.015 32 cm Pelvis 0.019 0.015 32 cm From 2 different sources 22 Effective Dose from DLP Values of k for Pediatric scans: Region of 0 Year 1 Year 5 Year 10 Year Phantom Body k (msv / mgy cm) Head 0.0087 0.0054 0.0035 0.0027 16 cm Head & Neck 0.0100 0.0068 0.0048 0.0040 16 cm Neck 0.0210 0.0168 0.0121 0.0094 16 cm Effective Dose from DLP There s an app for that: Free: Chest 0.0739 0.0482 0.0739 0.0237 32 cm Abdomen 0.0841 0.0530 0.0357 0.0249 32 cm Pelvis 0.0701 0.0446 0.0300 0.0219 32 cm 23 24 6

Possibility of a large mistake (factor of 2 or more) in calculating pediatric effective doses: With some scanners the pediatric body doses were based on the DLP from a 16 cm phantom: this was the ACR recommended phantom. However with other scanners the pediatric body doses were based on the DLP from a 32 cm phantom 25 Thus you must pay attention to the phantom size used for the CTDI vol and the DLP value given on the scanner for the values to be meaningful. Most manufactures that did use the 16 cm phantom for the pediatric body are now switching to the use of the 32 cm phantom instead. So beware. 26 You have to be careful even when using these tables: The dose is calculated for a standard size patient and many (or most) patients are larger. This changes the dose estimate. This effect on the CTDI vol has been investigated by two AAPM task groups. 27 They used a method called Size-Specific Dose Estimates (SSDE) to provide dose corrections for the CTDI vol. See the following AAPM Reports: AAPM Report No. 204 & No 220. However these reports warn you against using these corrections to correct the DLP or the estimated dose for patient size The patient model for calculating the E DLP still uses a standard patient size. 28 7

Effect of CT Protocols on Image Quality and Dose 29 Artifacts Image Quality and Dose: Image Sharpness Modulation Transfer Function MTF (visibility of small high contrast objects) Dose visibility of smaller, lower contrast clinical objects Image Noise Low Contrast Detectability (visibility of large low contrast objects) 30 Axial Basic Scan & Reconstruction Techniques Affecting Image Quality & Dose kv mas ma & rotation time Slice thickness Helical Basic Scan & Reconstruction Techniques Affecting Image Quality & Dose kv mas ma & rotation time Slice thickness Pitch 31 32 8

Helical Scan Techniques Affecting Image Quality & Dose Definition of Pitch for Multislice Helical / Spiral Scanning: Pitch coll = Table travel per 360 tube rotation Total collimation width of all simultaneously collected slices Helical Scan Techniques Affecting Image Quality & Dose The image noise and patient dose for helical scanning is generally a function of ma x rotation time / pitch which is often referred to as Effective mas : Effective mas = mas / pitch 33 34 Helical Scan Techniques Affecting Image Quality & Dose Effective mas = mas / pitch Siemens and Toshiba scanners use the term Effective mas in their scan techniques. Phillips uses the term mas / slice, which means the same as effective mas. Helical Scan Techniques Affecting Image Quality & Dose Effective mas = mas / pitch You may change the ma, rotation time, or pitch values, but if the effective mas remains constant, so does the CTDI vol and the patient dose. If the effective mas remains constant the image noise will also remain constant or nearly so. 35 36 9

Manual vs. Automatic Exposure One deficiency of CT Scanners before 2001 They did not contain any type of phototimer or automatic exposure control (AEC) to assure a proper patient dose. Therefore, manual technique charts were needed for different patient sizes. Usually this was not done so that techniques more suited for larger patients were used on all patients resulting in unneeded radiation exposure. 37 Automatic Exposure Control in CT Scanners Modern CT scanners have some type of automatic exposure control (AEC) that changes the ma during the scan. There are two basic types of AEC that can be used separately or together: The scanner varies the ma at different axial positions of the patient. The scanner varies the ma as the tube rotates around the patient. It is optimal to use both types together if the scanner allows it (Most do allow it). 38 Automatic Exposure Control in CT Scanners Automatic Exposure Control in CT Scanners Caution: The methods used by different manufacturers to perform AEC in CT are very different and may achieve very different clinical results. The scanner varies the ma at different axial positions of the patient. The scanner varies the ma at different axial positions of the patient and also varies the ma as the tube rotates around the patient. 39 40 10

Automatic Exposure Control in CT Scanners Some scanners (GE, Toshiba) try to keep the image noise constant as patient size increases: the automatic exposure control is adjusted by selecting the amount of noise that you wish in the image. This is done by selecting a Noise Index or SD (standard deviation). Typical values of Noise Index are 2.5 to 3.5 for a standard adult head scan and 12 to 20 for the body (for a 5 mm slice thickness). The scanner attempts to keep the image noise constant by adjusting the ma within set limits. 41 Automatic Exposure Control in CT Scanners For scanners that use a Noise Index or SD for AEC: The dose for a scan depends both on the Noise Index or SD AND the slice thickness selected for the first image reconstruction. Let s say you want to view reconstructed slice thicknesses of both 5 mm and 1.25 mm: Suppose the first image reconstruction has a slice thickness of 5 mm with a Noise Index of 12. If the first image reconstruction is switched to a slice thickness of 1.25 mm, the Noise Index needs to be changed to 24 to keep the dose constant. 42 Automatic Exposure Control in CT Scanners GE Example: Smart ma adds rotational variation of the ma to the axial variation performed in Auto ma without Smart ma. Therefore always press the Smart ma button when using Auto ma With GE scanners you must select whether you will be using manual techniques Manual ma or AEC techniques Auto ma. Manual ma uses an actual ma setting, Auto ma uses a Noise Index setting. Having one set correctly in a protocol does nothing to insure the other is properly set. 43 Automatic Exposure Control in CT Scanners Other scanners (Siemens, Philips) allow you to select the Effective mas, or the mas/ slice that you would use for an reference size patient. For Siemens scanners this selection is called the Quality reference mas. In AEC mode the scanner then automatically increases or decreases the effective mas for larger or smaller patients. This is done by varying the ma. Effective mas mas/ slice = (ma x rotation time) / pitch 44 11

Automatic Exposure Control in CT Scanners With Siemens Siemens: scanners you select the Eff. mas whether you will be using manual techniques OR AEC techniques. In manual mode this is the actual eff. mas used and in AEC mode it is the eff. mas that you would desire for an reference size patient. There is not the use of 2 different parameters for manual & AEC mode. 45 Automatic Exposure Control in CT Scanners Scanners that try to keep the image noise constant have the problem that they can quickly reach the maximum ma ceiling before getting to very large patients. Scanners that use a reference mas setting will generally allow the ma to increase only modestly with increased patient size, allowing the image noise to increase substantially for large patients. What is needed is a new AEC approach and the use of higher kv for larger patients. 46 Automatic Exposure Control in CT Scanners Automatic Exposure Control in CT Scanners A Concern with All CT Scanner: GE Siemens Proper centering of the patient is very important for the proper operation of the AEC system. A common problem is mis-centering the patient too low in the scan field. This can fool the AEC and also produce variable image quality over the patient. 47 48 12

Automatic Exposure Control in CT Scanners A Concern with All CT Scanners: Automatic Exposure Control in CT Scanners A Concern with All CT Scanner: Patient positioned 6 cm too low Midpoint of scout and scan field Patient positioned properly Midpoint of scout and scan field Most scanners use the last scout / topogram to adjust the ma modulation (though fine tuning can be done real time with some scanners.) Thus if the last scout/ topogram performed is a AP or PA a patient positioned too high or too low will fool the scanner into thinking that the lateral dimension is larger or small than reality. Thus a lateral scout / topogram should be the last performed 49 50 Image De-noising with Iterative Reconstruction Iterative Reconstruction (IR) is an additional step after performing Filtered Back Projection (FBP) which can reduce image noise Scanner manufacturers often make unrealistic claims on possible dose reductions based on the amount of noise reduction obtained with full strength IR However - Image noise reduction DOES NOT correlate well with actual improvements in Low Contrast Detectability (LCD) when using IR Image De-noising with Iterative Reconstruction One manufacturer claims a 40% dose reduction compared to FBP with the use of a moderate IR strength You will get the same image noise with this 40% dose reduction with this use of IR However the LCD will be substantially degraded Our tests indicate that you can only reduce the dose by 10% if you want the same LCD or the same diagnostic quality using this IR 51 52 13

Image De-noising with Iterative Reconstruction Other problems with IR: You will get a substantially modified image noise, texture which can interfere with your ability to read through the noise Image blurring can occur, particularly with sharp algorithms. Real image textures give difficulties to the IR reconstruction, adversely affecting the image: The image noise may not be reduced as much Some of the image texture may be erased 53 Other newer innovations in CT scanners to improve image quality and/or dose The introduction of 70 kv for smaller pediatric patients: head or body scans. Auto kv selection that uses patient attenuation and clinical task to select optimal scan kv. Patient size affects the kv selection. The importance of iodine or bone in the imaging will also affect the kv selection, since this would lower the optimal kv. 54 Other newer innovations in CT scanners to improve image quality and/or dose ma modulation that decreases the ma over the anterior part of the patient to reduce dose to the anteriorly positioned organs (organ dose modulation) Attempts to reduce dose to the breast, lens of the eye. Reduced ma Limitations: Degree of ma reduction; effectiveness compared to simply reducing the effective mas by a small amount. 55 Optimizing CT Protocols: Misconceptions and Recommendations for Scan and Imaging Parameters 56 14

kv Misconceptions: Scanning at 140 kv will reduce patient dose for any type of CT scan: head, body, adult or pediatric. For head scans, 140 kv should be used through the posterior fossa region to reduce image artifacts from bone. kv If we ignore beam hardening artifact limitations and CT scanner power limitations: The theoretical optimal kv, for any CT imaging, is the kv that will give the highest ratio of contrast to noise at a given patient dose. 57 58 kv For all Head CT scans and all Head or Body Pediatric scans this theoretical optimal would be close to 80 kv. For Adult Body CT scans this theoretical optimal will range from 80 kv up to 140 kv. 59 kv Modern CT scanners now have higher x-ray power & much more efficient use of this power through multi-slice design. They also have improved beam hardening/ bone correction algorithms. These improvements allow you to use lower kv settings closer to the theoretical optimal. 60 15

Optimal kv Technique Setting for Axial or Helical Scanning kv - Head CT Peds and Adult Use 70-80 kv for Peds Head 0-2y w/wo IV contrast. Use 80 kv for Peds Head 2-6y w IV contrast. Use 100 kv for Peds Head 2-6y wo IV contrast. Use 100 kv for Adult Head w IV contrast. Optimal kv Technique Setting for Axial or Helical Scanning kv Body CT - Peds Use 80 kv for all Peds Body for whom the sum of lateral and AP dimensions is less than 44 cm w/wo IV contrast. If available, 70 kv can be used. Use 100 kv for Peds Body for whom the sum of lateral and AP dimensions is between 44-55 cm wo IV contrast. Use 120 kv for Adult Head wo IV contrast. Use Adult protocols for larger patients. 61 62 Optimal Technique Setting for Axial or Helical Scanning kv Body CT Adults wo IV contrast Use 100 kv for Small Adults for whom the sum of lateral and AP dimensions is less than 60 cm. Optimal Technique Setting for Axial or Helical Scanning kv Body CT Adults w IV contrast Use 80 kv for Small Adults for whom the sum of lateral and AP dimensions is less than 60 cm. Use 120 kv for Medium Size Adults. Use 100 kv for Medium Size Adults. Use 140 kv for Large Adults for whom the sum of lateral and AP dimensions is greater than 80 cm. Use 120 kv for Large Adults for whom the sum of lateral and AP dimensions is greater than 80 cm. 140 kv for Large Adults reduces image noise and provides better image quality without large exposure increases. 63 Note: the use of lower kv produces a significant increase in the contrast of iodine, with better optimization of contrast to noise. 64 16

kv For scanning the neck or upper thorax, the amount of lateral attenuation through the shoulders is a serious problem for average to large size patients. Artifact due to the patient extending outside the Scan Field of View; ALSO Stringy noise artifact It will cause some degree of horizontal streaking artifact through the shoulder, which is actually a noise effect. 65 66 kv Here the solution is to increase the kv from 120 kv to 140 kv in adults for shoulder imaging to reduce the amount of lateral attenuation through the shoulders as much as possible and thus reduce this noise streaking artifact. 67 kv and Pitch - Pediatric Misconceptions: Using 140 kv for children to reduce dose. On the contrary this will generally raise the dose for equal image quality and is not recommended. Using a pitch greater than 1.0 for children is often strongly recommended to reduce radiation dose. This is totally misguided, as we will see shortly. 68 17

Pitch Misconceptions Scanning at higher pitch should be used as a strategy to reduce adult or pediatric patient dose and is always the best way to reduce scan time and motion artifact and blur. WRONG!!! Pitch Misconceptions: A pitch of less than one over-irradiates the patient due to scanning overlap, and thus wastes radiation dose. Thus one should avoid using a pitch less than one, particularly in pediatric scans. WRONG!!! 69 70 Pitch Journal of the American College of Radiology Volume 12, Number 4, April 2015 71 Pitch Changing the pitch from 1.0 to 0.5 increases the patient dose by a factor of 2 but also decreases image noise. These effects on dose and noise are the same as increasing the ma or the rotation time by a factor of 2. But decreasing the pitch has the added advantage of decreasing helical artifacts. 72 18

Pitch The effect of increased dose at lower pitch is easily countered by reducing the rotation time or ma in manual mode. There is NO increase in dose when decreasing pitch in AEC mode since the AEC mode in all scanners will keep the dose constant as the pitch is changed. Pitch Lowering the pitch and decreasing the exposure time by the same factor will keep the patient dose and exam time constant, but provide better image quality you get something for nothing! 73 74 Example: Pitch Change a 1.0 sec rotation time and a pitch of 1.6 to a 0.5 sec rotation time and a pitch of 0.8 Pitch For head scanning ALWAYS use a pitch of less than 1.0 to minimize helical artifact for adult or pediatric scanning. Best results are usually obtained with a pitch just above 0.5: 1. 75 76 19

Pitch For body scanning use a pitch of less than 1.0 whenever possible to minimize helical artifacts and allow more radiation for the adequate imaging of larger patients. When decreasing pitch in body scans, you need to be aware of breath hold limitations and contrast considerations. 77 Proper Use of Higher Pitch When to increase pitch (greater than 1.0): For pediatric and adult body scanning a shorter total scan time may allow a reduction in contrast volume. You may find that using a pitch greater than 1.0 allows a shorter total scan time with the available scan rotation times. Important in pediatric body scanning. 78 Axial vs. Helical Scanning Misconceptions: Heads should always be scanned using the axial rather than the helical mode or you will get a lower quality image. 79 Axial vs. Helical Scanning Helical scanning will almost always allow an exam with equal or better image quality than an axial scan if you have a CT scanner with 16 or more slices and select proper scan techniques. Axial scanning is still useful if required for positioning of the patient to avoid artifacts, since tilting the gantry is not allowed with helical scanning. 80 20

Axial vs. Helical Scanning and slice reconstruction interval Advantages of Helical scanning: Shorter total scan time with less chance for patient motion during the scan. The ability to reconstruct slices at intervals less than the slice thickness. VERY IMPORTANT! 81 Axial vs. Helical Scanning and slice reconstruction interval With axial scanning, the slice reconstruction incrementation is normally equal to the slice thickness. 82 Axial vs. Helical Scanning and slice reconstruction interval With helical scanning, the slice reconstruction incrementation can be set at any value. The best z-resolution is obtained by reconstructing at intervals ½ of the actual slice thickness this particularly helps with multiplanar reformatting. This is a significant advantage of helical scanning that is often not utilized. 83 Axial vs. Helical Scanning and slice reconstruction interval When creating slices for reformating of axial images to a modified axial plane, or for sagittal or coronal images, ALWAYS use thin slices as the source images, if this is not done automatically by the scanner. DO NOT USE 5 mm slices! For soft tissue recons use 1.0 to 1.5 mm slice thickness. For bone or high res recons use 0.5 to 0.75 mm slice thickness. 84 21

Axial vs. Helical Scanning and slice reconstruction interval Optimizing CT Protocols This talk has discussed some the most important ideas in CT protocol optimization. Slice thickness = 3mm increment = 3mm Slice thickness = 3mm increment = 1mm A final thought: After optimizing all other parameters: Slice thickness = 1mm increment = 1mm Slice thickness = 5mm increment = 5mm Reduce the patient dose to a level that produces consistently diagnostic scans and no lower! 85 85 86 22