Instructions for easyscan Atomic Force Microscope

Similar documents
UNIVERSITY OF WATERLOO Physics 360/460 Experiment #2 ATOMIC FORCE MICROSCOPY

Standard Operating Procedure

University of MN, Minnesota Nano Center Standard Operating Procedure

Operating Instructions. easyscan 2 STM. Version 2.1

Nanosurf Easyscan 2 STM

ATOMIC FORCE MICROSCOPY

Atomic Force Microscopy (Bruker MultiMode Nanoscope IIIA)

Standard Operating Procedure

Standard Operating Procedure of Atomic Force Microscope (Anasys afm+)

Akiyama-Probe (A-Probe) guide

Nanoscience Instruments

How to Cut Out Photos

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL

Akiyama-Probe (A-Probe) guide

GlassSpection User Guide

Prepare Sample 3.1. Place Sample in Stage. Replace Probe (optional) Align Laser 3.2. Probe Approach 3.3. Optimize Feedback 3.4. Scan Sample 3.

Tribometrics. Version 2.11

Bruker Dimension Icon AFM Quick User s Guide

Digital Imaging - Photoshop

The ideal K-12 science microscope solution. User Guide. for use with the Nova5000

Atomic resolution of the graphite surface by STM

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING

Study of shear force as a distance regulation mechanism for scanning near-field optical microscopy

Basic methods in imaging of micro and nano structures with atomic force microscopy (AFM) Related Topics. Principle. Equipment

Basic methods in imaging of micro and nano structures with atomic force microscopy (AFM)

Introduction. Let s get started...

GXCapture 8.1 Instruction Manual

Imaging Carbon Nanotubes Magdalena Preciado López, David Zahora, Monica Plisch

AgilEye Manual Version 2.0 February 28, 2007

Ansoft Designer Tutorial ECE 584 October, 2004

Measurement calibration in Video ToolBox Software. An example using a ruler

Using LTSPICE to Analyze Circuits

ArbStudio Triggers. Using Both Input & Output Trigger With ArbStudio APPLICATION BRIEF LAB912

Bruker Dimension Icon AFM Quick User s Guide

User Guide. Version 1.2. Copyright Favor Software. Revised:

Plot cylinder pressure against crank angle

Personalize Your Napkins

iworx Sample Lab Experiment AN-2: Compound Action Potentials

User Guide. Version 1.4. Copyright Favor Software. Revised:

Keysight Technologies Using Non-Contact AFM to Image Liquid Topographies. Application Note

Photoshop CS6. Table of Contents. Image Formats! 3. GIF (Graphics Interchange Format)! 3. JPEG or JPG (Joint Photographic Experts Group)!

Introduction to Photoshop CS6

Optika ISview. Image acquisition and processing software. Instruction Manual

AFM AC Tips and Tricks Oregon State University. AC mode - This Manual Does Not Replace the Manufactures Manual and/or Proper Training!

Here is a video that is very helpful

For all question related to Photoshop that we cannot address in class, start by looking at the excellent Photoshop help: Help > Photoshop Help.

Studuino Icon Programming Environment Guide

RICOH Stereo Camera Software R-Stereo-GigE-Calibration

By Washan Najat Nawi

R I T. Title: Wyko RST Plus. Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 05/23/06 1 SCOPE 2 REFERENCE DOCUMENTS

Using the Advanced Sharpen Transformation

RENISHAW INVIA RAMAN SPECTROMETER

Investigate in magnetic micro and nano structures by Magnetic Force Microscopy (MFM)

User Manual Veterinary

LSM 710 Confocal Microscope Standard Operation Protocol

JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES

BIO 365L Neurobiology Laboratory. Training Exercise 1: Introduction to the Computer Software: DataPro

Color correction software for digital images

Using Adobe Photoshop

Create A Mug. Skills Learned. Settings Sketching 3-D Features. Revolve Offset Plane Sweep Fillet Decal* Offset Arc

Nanosurf easyscan 2 Your Modular Scanning Probe Microscopy System

Scanning Procedure Using SilverFast Ai Software (6.6.1r2b) 1

ISONIC PA AUT Spiral Scan Inspection of Tubular Parts Operating Manual and Inspection Procedure Rev 1.00 Sonotron NDT

Lab 3 Introduction to SolidWorks I Silas Bernardoni 10/9/2008

Experiment 2: Electronic Enhancement of S/N and Boxcar Filtering

Outline: Introduction: What is SPM, history STM AFM Image treatment Advanced SPM techniques Applications in semiconductor research and industry

NSF Project: Cataclysms And Catastrophes LEARNING EXPERIENCE 2: STUDENT HANDOUT HURRICANES

Introduction to Lab Instruments

Operating Instructions. easyscan E-STM Version 2.0

BIM - ARCHITECTUAL IMPORTING A SCANNED PLAN

Using Curves and Histograms

Chanalyzer Lab. Chanalyzer Lab by MetaGeek USER GUIDE page 1

A graph is an effective way to show a trend in data or relating two variables in an experiment.

GIMP (GNU Image Manipulation Program) MANUAL

Before you start, make sure that you have a properly calibrated system to obtain high-quality images.

Experiment P55: Light Intensity vs. Position (Light Sensor, Motion Sensor)

digitization station DIGITAL SCRAPBOOKING 120 West 14th Street

Photoshop Elements 3 First Steps

Magnetic Levitation System

PATTERN MAKING FOR THE INFINITY WAND

Mid_Term_Review_PhotoShop_Design Test B Name

Fuzzy Image Editor. User Manual

Congratulations on your decision to purchase the Triquetra Auto Zero Touch Plate for All Three Axis.

EKA Laboratory Muon Lifetime Experiment Instructions. October 2006

Operating Rausch ScanCam within POSM.

Recording EPR Spectra using ER 4102ST Resonator

Adobe Lightroom CC Tutorial

Using the Microscope for a NANSLO Remote Web-based Science Lab Activity

Image Editing and Measurement Software. User Manual

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM

Agility Course Designer Users Guide. Agility Course designer program for Windows PC

1. Hand Calculations (in a manner suitable for submission) For the circuit in Fig. 1 with f = 7.2 khz and a source vin () t 1.

Remote Sensing 4113 Lab 08: Filtering and Principal Components Mar. 28, 2018

Brightfield Microscopy and Image Acquisition on Spotcam1. by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer

ThermaViz. Operating Manual. The Innovative Two-Wavelength Imaging Pyrometer

ezafm OPERATING MANUAL

Photoshop: Manipulating Photos

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL

Section 4 Digital Imagery/ Photo

GIMP WEB 2.0 ICONS. Web 2.0 Icons: Paperclip Completed Project

Transcription:

UVA's Hands-on Introduction to Nanoscience Instructions for easyscan Atomic Force Microscope (revision 8 November 2012) NOTE: Instructions assume software is pre-configured per "UVA Instructor Guide for easyscan 2 STM and AFM" Start up Electronics and Software (manual pages 27 & 37): 1) Open the log book and record your Name(s), Date, Time, and a short sample description. Initial and date. All Users MUST use the log books or face removal from lab 2) Turn on the control electronics (long hexagonal black box). Indicator lights should blink yellow 3) After the controller lights have settled, on the laptop PC click the Nanosurf easyscan2" icon. Indicator lights on controller should then become steady yellow, except: - STM scan head = dark (because it instead sees an AFM scan head) - Video Module (unless are using AFM #4 which DOES have video) - Probe Status may remain red at this point White light should go on at AFM cantilever position 4) Click Menu: File / Workspace / Load Go to the folder: C:/Program Files/Nanosurf easyscan2/config Select and load the file: UVA_AFM.gui By this point the bottom margin of the program must have these entries: The numbers in the 4th entry (##-##-###.hed) MUST match the label on top of the ATM If your entries are different, the software is confused as to if it is controlling an AFM or STM, or it lacks vital calibration and control data - ASK THE INSTRUCTOR TO FIX!! The preloaded files SHOULD set most of the required AFM parameters. But, because these files are easily corrupted, it is essential that you verify settings: For dynamic (tapping) mode, find the cantilever resonant frequency: 5) In the "Operating Mode Panel" In the Operating Mode sub-bar: 1

5a) Following your AFM's label/tag verify that Mounted Cantilever is set to EITHER NCLR or ACLA 5b) Verify that Operating Mode = Dynamic Force 5c) Expand the Mode Properties sub-bar which should have appeared 5d) Verify that Free vibration amplitude = 100-300 mv (preset = 199 mv) Use 50-100 mv for relatively flat samples Use 200-300 mv for rough, steep or deep samples (e.g., CD, IC...) IMPORTANT: Later if you change this amplitude setting, you must stop data collection, withdraw the cantilever (for ~ 5 seconds), then re-approach the sample. This allows the unit to detect the cantilever s new free vibration resonant frequency (which changes with amplitude). WARNING: Later indication of improper setting: Uniform orange image + absolutely flat/smooth line scan + red or blinking red Probe Status light on black electronics module. 5e) Verify that box is checked for Display Sweep Chart 5f) Click the Set button The AFM will now search for resonant frequency of the cantilever of the type specified in step 5a. If it succeeds you will see two charts pop up, first with a sharp peak, and then a second chart with a greatly expanded peak IF PLOT IS NOT AS ABOVE and Probe Status LED is not now green: STOP! The cantilever is not oscillating as expected and required due to: Wrong operating mode Improperly mounted cantilever Wrong cantilever type Damaged cantilever 2

Optimize feedback loop for type of sample you are examining: 6) In the "Z-controller Panel" 6a) Verify that Set point = 50% 6b) Verify that P gain = 10000 6c) Verify I gain = 1000 Approach the sample to be measured (manual page 39): 7) In the "Positioning Window" 7a) If necessary expand the Approach" sub bar to read full wording of buttons: 7b) With no sample in place OR with the AFM well above the sample: Use and buttons to raise or lower the head to a starting position flush to 1mm below body of AFM (as shown at left below) CORRECT: Head flush to 1mm below body WRONG: Head several mm below body SAMPLE LOADING: 8a) Turn micrometers to pull X-Y sample stage out from under AFM (as far as it will go) 8b) Place sample on stage but DO NOT PUSH TWEEZERS OR SAMPLE UNDER AFM 8c) INSTEAD use X-Y stage micrometers to move sample back under AFM (Later also remove sample using stage instead of placing tweezers under AFM) 3

COARSE MANUAL APPROACH: CRITICAL THAT FOLLOWING IS DONE RIGHT! 9a) WHILE PRESSING DOWN ON AFM body so that is does not move: Take turns adjusting the three screws on top until the cantilever is 1-2 mm from surface and the head is level (judge this by eye view from the front of unit) sample As AFM is lowered, keep its body ~ level (a bubble level is provided to check) 9b) Get a side view of cantilever in right lens (or with AFM #4 Click: for TV image) 9c) Using this side view, continue adjusting the three screws until the shadow of cantilever enters the view as shown below: Upper arrow at right: Side view of cantilever Lower arrow at right: Reflected image of cantilever Vague dark shadow of cantilever assembly DON T GO ANY CLOSER THAN THIS MANUALLY!! MOTORIZED FINE APPROACH (using buttons to control motor) YOU controlling AFM (fairly coarse): Advance = Turn on motor (while button is depressed) to move closer to sample Retract = Turn on motor (while button is depressed) to move away from sample AFM doing things automatically (very fine and slow): Approach = An automatic slow final approach to sample with AFM sensing contact Withdraw = Automatic slow movement away from sample 10a) Use Advance button to move toward sample until cantilever tip ALMOST meets its reflection 4

DON T LEAVE THE MOUSE CURSOR OVER Or tip may start moving even when the mouse button not pushed RESULT: Crashes cantilever into sample, breaking it! 10b) Complete approach by clicking automatic final approach button False Approach: Sometimes the AFM thinks it sensed a change in the cantilever s resonant frequency - and thus sample contact - before it has actually occurred (due to electrical noise). If this occurs you get data like that to the right: To fix: 1) Stop the scan; 2) Click the Approach button again. Starting a measurement (manual page 43): 11) In the "Imaging Window" use START, STOP buttons as with STM Use the data fitting parameters to your advantage: For the best images we have set up the program to use DERIVED DATA But RAW DATA and LINE FIT might work better for certain samples. MOVE the scanned area by clicking the image in the Topography Scan Forward pane. - Then click the MOVE button immediately above the image. - Click on image and draw an arrow giving the direction and distance you wish to move - Finally, under the Tool Results Panel on the left, click the MOVE button 5

ZOOM into a portion of your current image using one or the other of these techniques: 1) To zoom into the center of your current image area, in the Imaging Panel click the down arrow next to the Image Size box. This will zoom in to ½ the scan size (i.e. magnify by two). 2) Should you instead need to zoom into a specific non-center area: a) Click on the developing image in the Topography Scan Forward window pane b) Click the ZOOM button immediately above the image c) Click and drag a box around the area of interest d) Under the Tool Results Panel on the left, click the ZOOM button Capturing and saving your images: 1) During any measurement, click on the Topography Scan Forward window 2) Click on the Photo button immediately above/right of this image (it will turn blue) When scan reaches the top or the bottom, a new window with that saved image will pop up Program will continue saving images as long as "Photo" button is blue LATER, to save your data to a file, click on its pop-up window: - To save ALL information: Go to FILE / Save (as an "nid" Nanoscience Inc. format file) - To save ONLY a bitmap image (which cannot later be reprocessed, e.g. in 3D): Go to FILE, then EXPORT, then CURRENT DOCUMENT AS bmp, png or tiff file For a crude 3D map of your image, select its window then: 1) On the top outermost program bar, click "Line Graph" icon and switch "3D View" 2) Rotate 3D view as desired 3) Under the Tools menu you can also apply "Glitch" and "Noise" filters to smooth the data 4) Go to FILE, then EXPORT, then CURRENT DOCUMENT AS in the format you want For a FAR BETTER 3D map with noise filtering, smoothing & other enhancements: 1) Load your nid format file into program Nanosurf Reports by clicking "File/Open a Studiable" 2) Apply "Operators / Level" to level an image higher at one edge (i.e. different color / lighter) 3) Click "Studies/ Continuous Axonometric" (and rotate 3D plot as desired) 6

Click OK in window that pops up 4) Adjust "Height Amplification" (rainbow colored volcano icon, below "Reserves" and "Options") 5) Try adjusting color "Palette" (icon at far left, under the menu bar) 6) Try applying other "Operators," for instance: - "Waviness / Roughness" to filter out higher frequency noise - "Threshold" to cut out regions of low signal 7) Try applying "Studies / Photo Simulation" 8) To save EVERYTHING (in proprietary format): File / Save the Document (mmt format file) 9) To export the currently selected image: File / Export Reserve Image (choose tiff, gif, jpeg...) FOR CARBON NANOTUBE SAMPLES: For all of our other measurements you should use Dynamic Force (tapping) mode measurement as detailed above. However, for the carbon nanotube sample you will instead need to use Static Force (continuous contact) mode. 1) Have the instructor install a different static force type of AFM probe 2) Enter that probe s description into the AFM software 3) Change the operating mode to Static Force. 4) For the approach, set the force to 10 nn. 5) When the probe contacts the sample (and begins to scan), change the force setting to 2 nn. 7

Ending a Measurement - Click Withdraw button to automatically pull back a little ways from sample. - Click to then pull farther back Final Notes & Cantilever Replacement - Before shutting down AFM software & collecting tools, record all required information in log book - Under no circumstances are students to replace cantilevers. - Nor are you to modify or enable hardware/software settings in order to access the Internet - Immediately notify an instructor of any damage to the equipment UVA's Hands-on Introduction to Nanoscience 8