AC/DC Resistance Standard DRR-112

Similar documents
R-X SERIES. Decade Resistor

1433 SERIES. High-Accuracy Decade Resistor User and Service Manual

MODEL 9050 EXTENDED SPECIFICATIONS. 50ppm TRANSPORTABLE CALIBRATOR

audionet amp II G2 Owner's Manual Mono - Amplifier

Available at Surplus Sales Contents INTRODUCTION... 1 SPECIFICATIONS... 3 OPERATION... 5 MAINTENANCE... 8 Figures

Model Quality in Test and Measurement Since 1949 OUTSTANDING PERFORMANCE DESCRIPTION APPLICATIONS ADDITIONAL FEATURES

Advanced Test Equipment Rentals ATEC (2832)

Model 7000 Low Noise Differential Preamplifier

audionet PAM G2 User s Manual Phono Preamplifier for MC / MM

High Performance Current Transducer IT 200-S ULTRASTAB = A. ε L

Model SR554 Transformer Preamplifier

TOS5300 SERIES Hipot Tester/Hipot Tester with Insulation Resistance Test

5790A Automated AC Measurement Standard

DVM98. True RMS Digital Multimeter. 1 Safety information. 1.1 Preliminary. 1.2 During use

Instruction Manual. TT1000 Tower Earth Tester. Rev 01

Quadrupole Power Supply Unit QPS400-3

Current clamps for AC current

Model 3102D 0-2 kv H.V. Power Supply

Group: Names: (1) In this step you will examine the effects of AC coupling of an oscilloscope.

AMP-13 OPERATOR S MANUAL

MODEL 9041 EXTENDED SPECIFICATIONS. 25ppm TRANSPORTABLE CALIBRATOR

ULTRA-PRECISE AIR RESISTANCE STANDARDS

Models 900CT & 900BT. Tunable Active Single Channel Certified Filter Instrument

USER MANUAL. DC 13kHz High Voltage Amplifier WMA V to +175V output. Stable with all capacitive loads, generates no overshoot

TPS TRANSDUCER POWER SUPPLY & LEM IT CURRNET TRANSDUCER

AC / DC STROOMTANG MULTIMETER NI 30 AC / DC CLAMP MULTIMETER NI 30. Handleiding / Manual

For the electronic measurement of current: DC, AC, pulsed..., with galvanic separation between the primary and the secondary circuit.

CL900. True RMS 1000V 2000A 60MΩ ENGLISH. INSTRUCTION MANUAL 2000A Digital Clamp Meter. Measurement Technology

MM700. INSTRUCTION MANUAL Auto-Ranging Digital Multimeter True RMS

TRANSMILLE 3010A PRECISION MULTIPRODUCT CALIBRATOR EXTENDED SPECIFICATIONS

PRECISION INSTRUMENTS FOR TEST AND MEASUREMENT. HRRS 5kV AND 10kV SERIES. High Resistance, 5kV and 10kV Decade Substituter User and Service Manual

MFJ-249B HF/VHF SWR ANALYZER

HIGH ACCURACY RESISTANCE DECADE BOXES TYPE RBB 4, 5 & 6 DECADES

DECLARATION OF CONFORMITY

Kaskode One Phono Preamplifier Owner s Manual

E Series 9.00 (1/2) E series

DM-46 Instruction Manual

External RF Driver. Electrical Driver for AOTF-crystal. Instruction Manual

MM V 10A ENGLISH. INSTRUCTION MANUAL Auto-Ranging DATA HOLD AUDIBLE CONTINUITY MIN / MAX TEMPERATURE DIODE TEST CAPACITANCE

High Speed Milliohm Resistance Meter

APPLICATION NOTE. Wide Range of Resistance Measurement Solutions from μω to PΩ

For the electronic measurement of current: DC, AC, pulsed..., with galvanic separation between the primary and the secondary circuit.

ECE 2274 Lab 2. Your calculator will have a setting that will automatically generate the correct format.

Advanced Test Equipment Rentals ATEC (2832)

DC Current Transducers CT-200 CT-300 CT-400 User s Manual All Rights Reserved CAEN ELS d.o.o. Rev. 1.0 November 2014

HF Amateur SSB Receiver

ECE 2274 Lab 2 (Network Theorems)

Voltage Probe Manual and Data North Star High Voltage, Inc. Rev January 2016

Green LED (right of 'D' type connector)

Experiment 5: Grounding and Shielding

MM700. True RMS ENGLISH. INSTRUCTION MANUAL Auto-Ranging. Measurement Technology

For ultra-high precision measurement of current: DC, AC, pulsed..., with galvanic separation between primary and secondary. Applications.

DS2000ICLA. Specification highlights Symbol Unit Min Typ Max. Features. Applications: 1 ppm linearity. MPS for particles accelerators.

Variable-Gain High Speed Current Amplifier

POWER AMPLIFIERS 4 QUADRANTS 3x500 VA to 3x1500 VA - THREE-PHASES

Current Probes. User Manual

Primary-standard resistance thermometry bridge Model CTR9000

Model Operating Manual

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE EXT-02 B EXTRACELLULAR AMPLIFIER

Experiment 4: Grounding and Shielding

Accessories Selection Guide For Impedance Measurements. April 2005

12604 N New Reflection Dr Marana, AZ, FAX

INSTALLATION AND MAINTENANCE MANUAL FOR GROUND MONITOR GM-250 COPYRIGHT 1983 AMERICAN MINE RESEARCH, INC.

Model 4800 O P E R AT I O N M A N U A L L O A D C E L L S U M M I N G T R A N S M I T T E R

LM675 Power Operational Amplifier

MFJ-219/219N 440 MHz UHF SWR Analyzer TABLE OF CONTENTS

Agilent U1253B True RMS OLED Multimeter. Quick Start Guide

NOTE: Fully read and understand this manual before using this Digital Multimeter.

For ultra-high precision measurement of current: DC, AC, pulsed..., with galvanic separation between primary and secondary. Applications.

Technician License Course Chapter 4. Lesson Plan Module 9 Antenna Fundamentals, Feed Lines & SWR

audionet AMP 1 V2 User s Manual Stereo - Amplifier

79/26 Series III Multimeter

Page 1 of IR Port 5 External power port 6 Calibration button port 7 DUT + jack 8 DUT - jack 9 Shielding grounding jack

Agilent Accessories Selection Guide For Impedance Measurements. December 2008

ATK Clamp Meter

USER. manual. Falco Systems WMA-100. High Voltage Amplifier DC - 500kHz

USER'S MANUAL DMR-6700

2 Interfacing. 2.1 Introduction. The C 411 III pickup is a condenser transducer and therefore needs a power supply. Important!

Non-Contact Capacitance Gauging Instrument & Series 2800 Capacitive Probes

GFL-1000 User Manual Ground Fault Locator

I PM. Current Transducer ITZ 5000-SB FLEX ULTRASTAB = 5000 A

5500A. Multi-Product Calibrator. Extended Specifications 2005

User s Manual for Integrator Short Pulse ISP16 10JUN2016

DVM1190 DIGITAL MULTIMETER

DC Resistance Meters 2840 Series

For ultra-high precision measurement of current: DC, AC, pulsed..., with galvanic separation between primary and secondary. Applications.

Model: &

WHALETEQ. ESU Neutral Electrodes Impedance Tester. Model: HFPA150. User Manual

SI-125 Power Amplifier Manual 6205 Kestrel Road; Mississauga, Ontario; Canada; L5T 2A1 November 2016, Rev 0.5

Multifunction Digital

BARTINGTON INSTRUMENTS. How to use this Manual... 3 Symbols Glossary Introduction to the Mag Vector Measurements and Conventions...

For ultra-high precision measurement of current: DC, AC, pulsed..., with galvanic separation between primary and secondary. Applications.

Agilent U3400 Series 4½ and 5½ Digit Digital Multimeters

TENTATIVE PP800D120-V01

Digital Clamp Meter Model: &

PRECISION CURRENT TRANSDUCERS. DC Current Transducers CT-100 CT-150. User s Manual. All Rights Reserved CAEN ELS d.o.o. Rev. 1.

Signal Conditioning Amplifier

Experiment 1: Instrument Familiarization (8/28/06)

5075 Precision Digital Multimeter Time Electronics Calibration, Test & Measurement

MODEL 5002 PHASE VERIFICATION BRIDGE SET

Transcription:

AC/DC Resistance Standard DRR-112 Operator's Manual DTM Instruments, LLC www.dtminstruments.com Version 2.0 November 2013

Introduction The AC/DC resistance standard DRR-112 is used to calibrate resistance meters with an accuracy of 1000ppm or better. It contains eleven inductance-free, highly stable and accurate resistors, which can be selected with a dial switch at the front panel. It is especially designed to calibrate AC earth resistance meters, but it can be used to test and calibrate all types of DC and AC resistance meters with an AC test frequency of up to 100kHz and a test power of up to 10 Watts. For maximum accuracy, the DRR-112 offers four shielded terminals. Shielded and unshielded measurements with four-, three- and two-wire connections can be performed. A built-in guard amplifier can be used for measurements of up to 100kHz. Operator's manual DRR-112 page 2 of 16 Manual DRR-112 V20

Content of delivery One AC/DC resistance standard DRR-112 One switched mode power supply 12V/0.5A One RG-58/U BNC cable, 1ft. Operator's manual Calibration certificate Operator's manual DRR-112 page 3 of 16 Manual DRR-112 V20

Table of Contents 1. Safety instructions 6 1.1. Grounding 6 1.2. Maximum voltage and current values 6 2. Operation of the DRR-112 7 2.1. Preparation for operation 8 2.2. Taking out of operation 8 3. Connecting to the device under test 9 3.1. Four-wire (Kelvin) connection 9 3.2. Four-wire connection with guard 9 3.3. Three-wire connection 10 3.4. Two-wire connection 11 4. Reducing measurement errors 12 4.1. Connections 12 4.2. Thermal offset voltage 12 4.3. Measurement errors caused by inductive coupling 13 4.4. Measurement errors caused by stray capacitance 13 4.4.1. Using guard terminals 14 4.4.2. Grounding of the guard shield 14 4.4.3. Using the guard amplifier 15 5. Technical data 16 Operator's manual DRR-112 page 4 of 16 Manual DRR-112 V20

1.Safety instructions 1.1.Grounding Grounding is not required for the operation of the AC/DC resistance standard DRR-112. Nevertheless, it is strongly recommended to ground the housing via the grounding post to protect the operator from dangerous voltages, which might be created by faulty equipment. 1.2.Maximum voltage and current values maximum voltage: 100V maximum current: 0.5 A maximum power dissipation per resistor: 10 W The values in the following table are derived from the above limits: Selected Resistor Max. Current Max. Voltage Dissipated Power 1 Ω 500 ma 0.5 V 0.25 W 5 Ω 500 ma 2.5 V 1.25 W 10 Ω 500 ma 5 V 2.5 W 50 Ω 447 ma 22.3 V 10 W 100 Ω 316 ma 31.6 V 10 W 500 Ω 141 ma 70.7 V 10 W 1,000 Ω 100 ma 100 V 10 W 1,500 Ω 66.7 ma 100 V 6.7 W 1,900 Ω 52.6 ma 100 V 5.3 W 10,000 Ω 10.0 ma 100 V 1.0 W 18,000 Ω 5.6 ma 100 V 0.56 W Operator's manual DRR-112 page 5 of 16 Manual DRR-112 V20

2.Operation of the DRR-112 Fig 1 Operators Panel DRR-112 1: Power Supply connector, 12V/0.5A 2: Main power switch 3: Resistance selector switch 4.1: High-side current terminal 4.2: High-side voltage sensing terminal 4.3: Low-side voltage sensing terminal 4.4: Low-side current terminal 5: Shielded BNC terminals for 4.1 4.4 6: Grounding terminal 7: Power switch for guard amplifier 8: BNC jack for guard amplifier 9.1: Indicator Overload for guard ampl. 9.2: Indicator Power for guard ampl. Operator's manual DRR-112 page 6 of 16 Manual DRR-112 V20

2.1.Preparation for operation Connect the external 12V power supply to the power supply terminal at the DRR-112 (1) and plug it into the power outlet. Connect the device under test to the appropriate terminals (4.1 4.4), see chapter 3. Switch on the mains power (2), a warm-up time is not required if the DRR-112 is already at room temperature. Select the Short setting at the selector dial (3) and check that the device under test reads Zero. Select the desired resistance settings at the selector dial (3) and record the readout of the device under test. Check that the current and voltage output of the device under test does not exceed the limits given in chapter 1.2 2.2.Taking out of operation Switch off the device power (2, 7) Remove the external power supply from the mains power outlet and the power supply terminal at the DRR-112 (1) Remove the test leads Close the lid and store the resistance standard DRR-112 in a dry place. Operator's manual DRR-112 page 7 of 16 Manual DRR-112 V20

3.Connecting to the device under test 3.1.Four-wire (Kelvin) connection Reference DRR-112 Fig 2 4-Wire Connection The four wire connection is a straight one-to-one connection of the meter terminals with the terminals of the resistance standard DRR-112. Sometimes it is difficult to figure out which terminal corresponds to which. Naming of the terminals seems to follow no standard. Several common terminal names are printed on the front plate to guide with the connections. In doubt, check the manual of the resistance meter or contact the manufacturer. 3.2.Four-wire connection with guard For very accurate measurements it is recommended to use shielded connection cables to connect to the BNC jacks instead of the binding posts at the resistance standard DRR- 112. See chapter 4.44.4 for more information. Operator's manual DRR-112 page 8 of 16 Manual DRR-112 V20

ReferenceDRR-12 470pF470pF 3.3.Three-wire connection Reference DRR-112 Fig 3 3-Wire Connection The three-wire method of measuring resistances is less accurate than the four-wire method. Still, many earth resistance meters use this method. The standard connection setup is shown in Fig 5. The terminals of most three-wire meters are labeled as shown in the picture. At low values, the reference resistors used in the DRR-112 have additional resistance when used in the three-wire and two-wire connection. Refer to the calibration certificate for the actual resistance values between 1Ω and 50Ω. To avoid the error caused by the additional resistance of the relay contacts, all resistors are hard wired and no relay contacts are used at the terminal C1. The terminal C1 serves as the common ground connector. Operator's manual DRR-112 page 9 of 16 Manual DRR-112 V20

3.4.Two-wire connection Reference DRR-112 Fig 4 2-Wire Connection Digital Multimeter and many other test systems have only two connectors to measure the resistance. In this case, all wiring and contact resistances add to the measurement result. To keep these errors as small as possible, it is recommended to short connect the terminals C1 and P1 as well as C2 and P2 as shown in Fig 6. If used without this short, special care must be taken since some terminals of the fourterminal resistors exhibit an additional resistance. At low values, the reference resistors used in the DRR-112 have additional resistance when used in the three-wire and two-wire connection. Refer to the calibration certificate for the actual resistance values between 1Ω and 50Ω. Operator's manual DRR-112 page 10 of 16 Manual DRR-112 V20

4.Reducing measurement errors Every measurement is subject to interference which introduces errors to the measurement result. These errors must be kept small enough so that they don't falsify the calibration results. The higher the required accuracy of the calibration and the higher the frequency of the AC measurement signal, the more important are the following guidelines. 4.1.Connections It should go without saying that all connectors have to be clean and free of oxides and other residues as well as that the connection leads don't have broken wires or loose plugs. It is also important to keep the length of the connection leads as short as practically possible. A long lead will pick up more interference, have a higher inductance and stray capacitance than a shorter one. For the calibration of earth testers and other resistance meters with up to 0.1% accuracy, unshielded connection cable is usually sufficient. For very sensitive and accurate measurements (e.g. to calibrate the resistance standard DRR- 112 itself) it is recommended to use short shielded cable (e.g. RG-58/U) connected to the BNC jacks. 4.2.Thermal offset voltage If a junction of different metals is on a different temperature than the rest of the metals, a thermal voltage will be produced. This junction acts as a tiny battery which creates a DC offset voltage within the test circuit. Only measurements with DC voltage are affected by thermal voltages. To test for thermal offset voltages, connect the voltage measurement terminals to the P1 and P2 terminal at the resistance standard DRR-112 and short the current terminals of the resistance meter (Fig 7). A reading different than Zero indicates a thermal offset voltage. To verify that the thermal offset voltages is created within the resistance standard box, swap the voltage leads between the P1 and P2 terminal and observe the change in the readout. Reference DRR-112 Fig 5 Test for Thermal Offset Operator's manual DRR-112 page 11 of 16 Manual DRR-112 V20

4.3.Measurement errors caused by inductive coupling Every closed electric circuit exhibits an inductance, it could be seen as a choke with one turn. It is of concern only for test with AC voltages and currents (e.g. earth tester), DC is not affected. This inductance creates an additional voltage drop that affects two- and three-wire measurements with AC signals. It also creates a magnetic coupling between the current and potential leads that falsifies the measured voltage at three- and four-wire measurements. Inside the resistance standard DRR-112 the connections are carefully laid out to minimize the inductive effect and inductance-free resistors are used. The external connection leads should be as short as practically possible and twisted to minimize the inductance and the magnetic field coupling (see Fig 8 and Fig 9). Reference DRR-112 Reference DRR-112 Fig 6 4-Terminal Connection to minimize interference Fig 7 3-Terminal Connection to minimize interference The interference created by the circuit inductance and magnetic coupling is affecting the measurement more at low resistance values and higher test signal frequency. To determine if the measurement setup is affected by the circuit inductance, select the Short position at the selector switch. The instrument should read Zero within its measurement accuracy. 4.4.Measurement errors caused by stray capacitance Electrical leads with a different electrical potential create a small capacitor, the so-called stray capacitance. This capacitor lies in parallel to the reference resistor and tends to reduce the overall AC impedance (the measured resistance ). The longer the leads get and the closer they are, the larger the stray capacitance becomes. Stray-capacitance affects only AC measurements. The higher the signal frequency and the larger the reference resistor is, the more the measurement will be affected by the parasitic stray capacitance. As an example, at 10kHz and 18kΩ, the internal stray Operator's manual DRR-112 page 12 of 16 Manual DRR-112 V20

capacitance of 52.5pF will create a measurement error of approximately 1700ppm (0.17%). To minimize the stray capacitance, the internal wiring, relay switches and circuit board are equipped with a guard shield. The guard shield is accessible at the outside of the BNC connectors. The stray capacitance inside the reference resistors can not be shielded out. Without shielding, a stray capacitance of 250pF can be roughly estimated. With this, a measurement error below 1000ppm (0.1%) will be reached if the AC test frequency is below 1.6kHz. For earth resistance meters, this is usually the case and guarding is not necessary. For more accurate measurements, e.g. to calibrate the resistance standard DRR-112, guarding might be required. To determine if guarding is necessary, connect two capacitors 470pF/100V to the terminals P1 and P2. Compare the reading of the 18kΩ resistor with and without the two capacitors connected (see Fig 10). If the reading is less when the capacitors are connected, guarding is strongly recommended. Reference DRR-112 Reference DRR-112 470pF 470pF 470pF 470pF Fig 8 Stray Capacitance Test Fig 9 Guard Grounding Test 4.4.1.Using guard terminals The preferred way of guarding out the stray capacitance uses the guard circuit of the resistance meter. Reference resistance meters and bridges usually use guard circuits. Simply connect the shielded terminals of the resistance meter with the BNC connectors at the resistance standard DRR-112 via shielded cables. The outside of the BNC connectors is connected to the internal guard shields of the high and low side. 4.4.2.Grounding of the guard shield If no guard output is available at the resistance meter, maybe a ground terminal (or a DC charger terminal) is accessible. Connect the center point of the two 470pF capacitors to Operator's manual DRR-112 page 13 of 16 Manual DRR-112 V20

the ground terminal and see if the 18kΩ reading goes back to the value without the capacitors (Fig 11). If so, remove the capacitors and connect the ground terminal to the outer BNC conductor on the left (P2) or right (P1) side. Check which connection gives the highest reading, this is the best connection for the guard shield. 4.4.3.Using the guard amplifier If the other two methods do not work, the built-in guard amplifier can be used. Switch on the guard amplifier and check that the LED Power (green) is on. Connect the guard amplifier BNC output to the BNC connectors P1 or P2 via a short BNC cable. Also connect the center point of the two 470pF capacitors to the outer conductor of the BNC connector and see if the 18kΩ reading goes back to the value without the capacitors. If so, remove the capacitors and check which connection gives the highest reading without the LED Overload (red) lighting up. This is the best connection for the guard shield. Troubleshooting: If the LED Power (green) does not light up, the guard amplifier might be damaged by input over voltages. The internal operational amplifier can be replaced by opening the front cover and exchanging it with a new OP249GP. If the LED Overload (red) lights up, this is an indication that the guard amplifier output is saturating. Try connecting the ground terminals at the DRR-112 and the device under test or use the other P1/P2 terminal. Operator's manual DRR-112 page 14 of 16 Manual DRR-112 V20

5.Technical data DRR-112 Box Dimensions: 29 x 25 x 16 cm (11.5 x 10 x 6.25 inch) Weight: 2.7kg (6 lbs) Operating temperature: 10 30ºC (50 86 F), 30 70 %RH Storage temperature: 0 40ºC (32 104 F), not condensing Supply current: 130mA (Relays: 50mA, Guard Amplifier 80mA) Reference Resistors maximum circuit current: 0.5A maximum circuit voltage: 100V maximum applied power: 10W annual stability: 100ppm (0.01%) p.a. temperature coefficient: ±5ppm/K (±2.8ppm/F) resistance material: NiCr-Foil frequency range: 0 10 khz accuracy: Range Accuracy 1.0 Ω 5.0 Ω 1000ppm (0.1%) 10 Ω 1.9 kω 500ppm (0.05%) 10 kω 18 kω 750ppm (0.075%) Operator's manual DRR-112 page 15 of 16 Manual DRR-112 V20

stray capacitance: Resistance C Stray 10kHz error 100kHz error 500Ω 300pF 44ppm 0.4% 1,000Ω 208pF 85ppm 0.8% 1,500Ω 165pF 121ppm 1.2% 1,900Ω 170pF 206ppm 2.0% 10,000Ω 68.5pF 925ppm 8.1% 18,000Ω 52.5pF 1760ppm 14.0% Guard Amplifier Output voltage swing: ±12V Operating frequency: 30Hz 300kHz External Power Supply Input Voltage: 100 240 V, 50 60 Hz, 0.3A Output Voltage: 12 V DC, 0.5A DC Power connector: 5.5/2.1mm Barrel Plug, center positive Operator's manual DRR-112 page 16 of 16 Manual DRR-112 V20