Surgical Technique 1

4.0 CANNULATED SCREW SYSTEM

MEDIALMAX System SURGICAL TECHNIQUE

OPERATIVE TECHNIQUE RIVAL REDUCE FRACTURE PLATING SYSTEM. foot & ankle trauma procedures

MTP Fusion Surgical Technique

Peanut Growth Control Plating System. Surgical Technique

ISO Plate SURGICAL TECHNIQUE

MaxTorque. surgical technique. Cannulated Screw System. Foot & Ankle. OrthoHelix Technology

5th Metatarsal Fracture System Surgical Technique

BioDrive Micro Screw System

OPERATIVE TECHNIQUE RIVAL VIEW PLATING SYSTEM. foot & ankle reconstruction procedures

Surgical Technique. Customer Service:

Distal Fibula Plate SURGICAL TECHNIQUE

Variable Angle LCP Tarsal Plates 2.4/2.7. Navicular Plate and Cuboid Plates.

DART-FIRE. Small Screw System SURGICAL TECHNIQUE

Distal Radius System 2.5

DART-FIRE. Small Screw System SURGICAL TECHNIQUE

Integra. Capture Screw System SURGICAL TECHNIQUE

Technique Guide. Occipito-Cervical Fusion System. Implants and instruments designed to optimize fixation to the occiput.

VariAx Fibula Locking Plate System

DART-FIRE. Small Screw System SURGIC A L T ECHNIQUE

Instructions for Use. LCP Locking Compression Plate. Combine without Compromise.

VARIABLE ANGLE LOCKING HAND SYSTEM

ACCS Anterior Cervical Compression System TECHNIQUE GUIDE

VariAx DistalFibula. Foot & Ankle. Locking Plate System. Operative Technique

Introduction TRIGEN META-TAN Nail specifications Surgical technique Patient positioning Opening the proximal femur Intramedullary reaming

VARIABLE ANGLE LOCKING HAND SYSTEM

MTP Set SURGICAL TECHNIQUE

1.5 MM LCP SYSTEM. For treatment of fractures and arthrodeses of canines and felines SURGICAL TECHNIQUE

WRIST SYSTEM. ARIX Volar Distal Radius Locking Plate System

Universal Humeral Nail

Technique Guide MTP Arthrodesis System

Variable Angle LCP Mesh Plate 2.4/2.7. Part of the Variable Angle LCP Forefoot/Midfoot System 2.4/2.7.

Integra. Stainless Headed Compression Screw System SURGICAL TECHNIQUE

Operasjonsteknikk. Retrograd Femur

Operative Technique Distal Fibula Fracture Repair Polyaxial Locking Technology Low Profile Design

VariAxFibula. Fibula Fractures. Locking Plate System. Operative Technique

Technique Guide. 2.4/2.7 mm Locking Tarsal Plates. Talus Plate, Navicular Plate and Cuboid Plate.

TwinFix Surgical Protocol. 3.2mm Cannulated Compression Screw System

Distal Medial Tibia Plate Surgical Technique

Ankle Fracture System. Surgical Technique STRENGTH FROM WITHIN

Small Fragment Locking Compression Plate (LCP ) System Stainless Steel and Titanium TECHNIQUE GUIDE

OPERATIVE TECHNIQUE RIVAL BITE HEADED CANNULATED AND HEADLESS COMPRESSION SCREWS. foot & ankle applications

Variable Angle LCP Forefoot/Midfoot System 2.4/2.7. Procedure specific plates for osteotomies, arthrodeses and fractures of the foot.

Technique Guide. Quadrilateral Surface Plates 3.5. Part of the Low Profile Pelvic System 3.5.

VECTRA SURGICAL TECHNIQUE. Anterior cervical plate system. This publication is not intended for distribution in the USA.

Small Plate and Screw System SURGICAL TECHNIQUE

VariAx Foot Locking Plate System

DISTAL RADIUS PLATES 3.5 mm / ANGULARLY STABLE. Distal radius plates 3,5 mm / angularly stable. Locking bone screws. Cortical bone screw

VECTRA. SURGICAL TECHNIQUE. Anterior cervical plate system. This publication is not intended for distribution in the USA.

Digital Compression Screw

Integra. HALLU -Lock M.T.P. Arthrodesis System SURGICAL TECHNIQUE

Treace Medical Concepts Plating System

Technique Guide. Variable Angle LCP Opening Wedge Plates 2.4/2.7. Part of the Variable Angle LCP Forefoot / Midfoot System 2.4 / 2.7.

ACLP Anterior Cervical Locking Plate System TECHNIQUE GUIDE

MatrixMANDIBLE Preformed Reconstruction Plates. Preshaped to the mandibular anatomy.

Surgical Technique Guide

Vectra, Vectra-T and Vectra-One. Anterior cervical plating for spinal fusion.

Technique Guide. Variable Angle LCP 1 st MTP Fusion Plates 2.4/2.7. Part of the Variable Angle LCP Forefoot / Midfoot System 2.4 / 2.7.

Orthopedic Bone Nail System Universal Humeral Nail

SURGICAL TECHNIQUE GUIDE

ORTHOLINK. 2 - Hole Wedge Osteotomy Plate SURGICAL TECHNIQUE

Cervical Solutions. Lineum OCT. Spine System. Surgical Technique Guide

Knee Nail for Retrograde Femoral Mode

Reflex Hybrid System Overview

MaxAn Anterior Cervical Plate System

Occipito-Cervical Fusion System. Implants and instruments designed to optimize fixation to the occiput.

LCP Pilon Plate 2.7/3.5. Surgical Technique

Technique Guide. Modular Sternal Cable System. Flexibility and strength in sternal closure and repair.

VariAx. Foot & Ankle. Foot Locking Plate System

DLS Dynamic Locking Screw. Combined with LCP Locking Compression Plate.

TORNIER MAXLOCK EXTREME. Clavicle Plating System SURGICAL TECHNIQUE

Anterior Cervical Plate SURGICAL TECHNIQUE GUIDE. Surgeon Driven Innovation

CSLP Variable Angle. For Use with the Cervical Spine Locking Plate System TECHNIQUE GUIDE. Self-drilling Screw. Variable Screw Angulation

Lag Screw Device TECHNIQUE GUIDE. Indicated for symphyseal fracture fixation of the mandible. Instruments and implants approved by the AO Foundation

Fibula Plating System

Humeral Nail System Procedural Steps.

surgical technique addendum

Fixation screw ENGLISH

OCCIPITO-CERVICAL FUSION SYSTEM Implants and instruments designed to optimize fixation to the occiput

Interlagos Retractor System Surgical Technique

Table of Contents 2-6. Introduction. Indications Surgical Technique. Ordering Information 15-24

LCP Pilon Plate 2.7/3.5

Technique Guide. LCP Dynamic Helical Hip System (DHHS). Part of the Synthes Large Fragment LCP System.

Occipito-Cervical Fusion System

3.5 mm Cannulated Screw Technique Guide

CHARLOTTE. Multi-Use Compression Screw

Distal Volar Radius Plate Procedure Steps.

Integra. surgical Technique. Total Wrist Fusion System. Products for sale in Europe, Middle-East and Africa only

LCP Pilon Plate 2.7/3.5

Technique Guide. 4.5 mm Cannulated Screws. Part of the Synthes Cannulated Screw System.

3.5 mm and 4.5 mm Curved Locking Compression Plates (LCP )

OPERATIVE TECHNIQUE CENTURION POSTERIOR OCCIPITAL CERVICO-THORACIC (POCT) SYSTEM

CETRA ANTERIOR CERVICAL PLATE

Zimmer Natural Nail System

Virage OCT Spinal Fixation System

BLACKBIRD Spinal System

CHARLOTTE CLAW Plate Compression Locking Arthrodesis by Wright SURGICAL TECHNIQUE

Ulna Shortening System 2.5

Transcription:

Surgical Technique 1

D-RAD SMART PACK Single-Use Volar Distal Radius Plating System Surgical Technique Table of Contents Indications... 3 Contraindications... 3 D-RAD SMART PACK product overview... 4 Instrumentation... 5 Surgical technique... 6 OR set up... 6 Patient positioning... 7 Incision... 7 Fracture reduction... 7 Single-Use Kit selection... 7 Plate positioning... 7 Screw hole preparation... 8 Screw selection... 9 Screw insertion... 10 Guide Block removal... 10 Final imaging... 10 Closure... 10 Catalog information... 11 Nota Bene The technique description herein is made available to the healthcare professional to illustrate the author s suggested treatment for the uncomplicated procedure. In the final analysis, the preferred treatment is that which addresses the needs of the specific patient. 2

Indications The D-RAD SMART PACK Single-Use Volar Distal Radius Plating System is indicated for the fixation of fractures involving the distal radius. Contraindications 1. Physical conditions that would preclude adequate implant support or retard healing, such as, blood supply impairment, insufficient bone quality or quantity, previous infection, obesity, severe bow or gross distortion of the radius. 2. Mental conditions that preclude cooperation with the rehabilitation regimen. For additional information, including warnings and precautions, please review the Instructions for Use included with the product. 3

D-RAD SMART PACK product overview 4-Hole Volar Plate 1.4mm K-wire holes for distal plate positioning Anatomically contoured to respect the watershed line while providing reconstructive articular support Slotted 1.4mm K-wire hole or slotted screw hole to allow for proximal/distal plate adjustments Volar tilt 24º VLP locking technology enables the use of Cortex Screws, Locking Screws or Locking Pegs within a 30º cone 56mm 22mm 27mm 1.4mm K-wire hole to enable centering of the plate on the diaphysis Standard Wide Standard and wide options accommodate varying patient anatomy 1.6mm Screws/Pegs 2.4mm Cortex Screws 2.4mm Variable-Angle Locking Screws (10 24mm) (10 24mm) 1.8mm Variable Angle Locking Pegs for bicortical support of the metaphysis while minimizing dorsal soft tissue irritation (12 24mm) 4

Instrumentation Self-retaining T7 screw driver 1.4mm K-wires assist with plate positioning and reference screw trajectories Fixed-angle drill guide predetermines screw trajectories Variable-angle drill guide enables intraoperative screw placement flexibility Preassembled guide block and plate holder Hooked, laser-etched depth gauge grabs the dorsal cortex 1.8mm drill for cortex screws, locking screws and locking pegs 5

Surgical technique OR setup Single-Use Kits are packaged sterile with a left or right 4-hole volar distal radius plate and include single-use instruments needed for implantation. Left and right plates are available in two widths (standard and wide) to accommodate varying patient anatomy. In addition to Single-Use Kits, the Screw and Template Tray containing a variety of screws, pegs, and templates will be needed to conduct the procedure. The Screw and Template Tray should be positioned outside of the sterile field as the unit is non-sterile and should not be sterilized. The lid can be removed and nested underneath the tray. Single-Use Kits include either a Left Standard, Left Wide, Right Standard, or Right Wide Plate. Left Standard Left Wide Note: Soft tissue retractors, fracture reduction instruments and longer plates are not available in the D-RAD SMART PACK Single-Use Volar Distal Radius Plating System. Note: OR setup should include the appropriate wire driver(s) needed to utilize 1.4mm K-wires for plate positioning and a 1.8mm Drill Wire for screw preparation. Screw and Template Tray Plate Templates are available in left and right, standard and wide options and exactly match the width, length and contour of their respective Plates. Open the appropriate Plate Template box (right or left) and place the single-use standard and wide plate templates on the back table in preparation for the Single-Use Kit selection step. Plate Templates: Left Standard and Left Wide Note: A Single-Use Kit should only be opened after the appropriate plate size has been determined by the Single-Use Plate Templates during the Single-Use Kit Selection step of the procedure. 6

Patient positioning The patient should be placed in the supine position with the affected limb positioned to expose the surgical site. Incision A longitudinal incision is made in Henry s interval at the wrist (flexor carpi radialis and radial artery). Carry the dissection to the pronator quadratus. Elevate the pronator quadratus to reveal the fracture site. Fracture reduction After exposure and debridement of the fracture site, the fracture is reduced and provisionally fixed under fluoroscopy, if necessary. Single-Use Kit selection Using the previously opened Plate Templates, confirm the plate width that will best accommodate the patient s anatomy and fracture pattern by applying the distal portion of each Template to the approximate watershed line. Once the proper plate width has been determined via the Plate Templates, open the corresponding Single-Use Kit and introduce the sterile tray into the sterile field. Plate Template on the distal radius Plate positioning Remove the Plate and preassembled Guide Block from the Single-Use Kit. Place them into the wound and onto the affected distal radius. The Plate has been designed with a variety of 1.4mm K-wire holes to assist with plate positioning. A wire driver should be used to advance the provided 1.4mm K-wires through the plate to provisionally position it on the bone. Plate and Guide Block positioned on the distal radius 7

Screw hole preparation The Plate may be implanted using either a shaft-first or distal-first fixation method. The Guide Block combined with the VLP locking mechanism offers users the option of fixed, predesigned screw trajectories through the use of the Fixed Angle Drill Guide or variable-angle screw trajectories by using the Variable Angle Drill Guide. Fixed Angle Drill Guide The star shaped distal tip of the Fixed Angle Drill Guide must be seated into the star shaped hole of the plate to facilitate Locking Screw engagement. The conical shape of the Fixed Angle Drill Guide restricts angulation when placed through the Guide Block and thus centers the Drill for fixed predesigned screw trajectories. Note: It is still possible to misplace screws into the radiocarpal joint using the variable-angle capability of the plate even when the K-wires appear to be well placed. Always confirm safe screw placement with fluoroscopy. Note: The Guide Block limits screw trajectories to a 30º cone or 15º in any direction when using the Variable Angle Drill Guide. Screw angles should not exceed this in order for the VLP locking mechanism to work properly. Additionally, exceeding 15º off axis may cause unacceptable screw head prominence. Variable Angle Drill Guide The star shaped distal tip of the Variable Angle Drill Guide must be seated into the star shaped hole of the plate to facilitate Locking Screw engagement. When placed through the Guide Block, the Variable Angle Drill Guide allows for drilling within a 30º cone or 15º in any direction. Note: Regardless of which drill guide is used (Fixed Drill Guide or Variable Angle Drill Guide), fluoroscopy should always be used to confirm safe drill placement due to the variability of patient anatomy and fracture patterns. Insert the star tip of the Fixed or Variable Angle Drill Guide into the desired hole and drill accordingly with the 1.8mm Drill Wire. Screw length can be determined by reading the exposed laser-etched line on the 1.8mm Drill Bit through the window of the Drill Guide. Alternatively, a Depth Gauge with a laseretched line has been provided and can be used through the Drill Guide to determine proper screw length. 8 Drilling with the 1.8mm Drill Wire Determine screw hole depth with the Depth Gauge through the Drill Guide The laser-etched line on either the 1.8mm Drill Wire or the Depth Gauge can be read through the window of the Drill Guide to determine the appropriate Screw length.

Note: Drilling multiple screw holes without the plate firmly secured to the bone can cause misalignment between the Locking Screw/Peg heads and the Plate. This can lead to engagement difficulties. It s advised that a Screw/Peg be placed in a prepared hole prior to additional holes being drilled. Screw selection 2.4mm Cortex Screws, 2.4mm Locking Screws and 1.8mm Locking Pegs are contained in sterile tamper-evident vials and are organized by type and size within the Screw and Template Tray. Once the appropriate screw or peg has been identified, twist the non-sterile cap off of the vial and allow the sterile inner container to drop into the sterile field. Flip the cap of the sterile container up and remove the screw or peg with the provided T7 Self-Retaining Driver by engaging the head. Alternatively, the screw or peg can be removed by turning the inner container upside-down and allowing the screw or peg to fall out. D-RAD SMART PACK screws and pegs are packaged sterile thus eliminating the need to confirm length before insertion. However, a screw/peg scale has been provided in the Single-Use Kit should one be needed. Screw/Peg Vial Sterile Inner Container Note: When using the screw/peg scale provided on the Single-Use Kit tray, ensure that the top surface of the screw/peg head is in contact with the top surface of the scale. The T7 Self-Retaining Driver can be used to remove the Screw/Peg from the Sterile Inner Container Screw/Peg scale 9

Screw insertion Use the T7 Self-Retaining Driver to insert the appropriate length 2.4mm Cortex Screws, 2.4mm Locking Screws and 1.8mm Locking Pegs. Screws and Pegs can be inserted through the Guide Block. Note: A moderate amount of downward pressure should be applied through the driver when the head of a Locking Screw or Locking Peg begins to engage the Plate. Note: The non-locking screw hole in the metaphyseal area of the plate has been designed to only accept a cortex screw. Note: Fluoroscopy should always be used to confirm safe screw placement due to the variability of patient anatomy and fracture patterns. Guide block removal Once distal screw insertion is complete, use the T7 Self-Retaining Driver to turn the Plate Holder counter-clockwise and subsequently remove the Guide Block from the Plate. Final imaging Obtain final radiographic images to confirm optimal screw placement and length. Be certain that the plate fits dorsal to the volar lip of the radius to help avoid tendon injury. Closure If possible, repair the pronator quadratus over the plate. Wound closure as per surgeon preference. 10

Catalog D-RAD SMART PACK Single-Use Volar Distal Radius Plating System Cat No Description Qty 7115-8016 D-RAD Screw and Template Tray 1 7115-8023 D-RAD Screw and Template Tray Lid 1 7115-8019 D-RAD SMART PACK 4H Left Standard Plate 2 7115-8020 D-RAD SMART PACK 4H Left Wide Plate 1 7115-8021 D-RAD SMART PACK 4H Right Standard Plate 2 7115-8022 D-RAD SMART PACK 4H Right Wide Plate 1 7115-8017 D-RAD Plate Templates Left 3 7115-8018 D-RAD Plate Templates Right 3 7467-1812 D-RAD TI 1.8mm x 12mm Locking Peg T7 3 7467-1814 D-RAD TI 1.8mm x 14mm Locking Peg T7 8 7467-1816 D-RAD TI 1.8mm x 16mm Locking Peg T7 7 7467-1818 D-RAD TI 1.8mm x 18mm Locking Peg T7 7 7467-1820 D-RAD TI 1.8mm x 20mm Locking Peg T7 7 7467-1822 D-RAD TI 1.8mm x 22mm Locking Peg T7 4 7467-1824 D-RAD TI 1.8mm x 24mm Locking Peg T7 3 7468-2410 D-RAD TI 2.4mm x 10mm Cortex Screw T7 S-T 7 7468-2412 D-RAD TI 2.4mm x 12mm Cortex Screw T7 S-T 8 7468-2414 D-RAD TI 2.4mm x 14mm Cortex Screw T7 S-T 7 7468-2416 D-RAD TI 2.4mm x 16mm Cortex Screw T7 S-T 7 7468-2418 D-RAD TI 2.4mm x 18mm Cortex Screw T7 S-T 7 7468-2420 D-RAD TI 2.4mm x 20mm Cortex Screw T7 S-T 7 7468-2422 D-RAD TI 2.4mm x 22mm Cortex Screw T7 S-T 4 7468-2424 D-RAD TI 2.4mm x 24mm Cortex Screw T7 S-T 3 7469-2410 D-RAD TI 2.4mm x 10mm Locking Screw T7 S-T 7 7469-2412 D-RAD TI 2.4mm x 12mm Locking Screw T7 S-T 8 7469-2414 D-RAD TI 2.4mm x 14mm Locking Screw T7 S-T 7 7469-2416 D-RAD TI 2.4mm x 16mm Locking Screw T7 S-T 7 7469-2418 D-RAD TI 2.4mm x 18mm Locking Screw T7 S-T 7 7469-2420 D-RAD TI 2.4mm x 20mm Locking Screw T7 S-T 7 7469-2422 D-RAD TI 2.4mm x 22mm Locking Screw T7 S-T 4 7469-2424 D-RAD TI 2.4mm x 24mm Locking Screw T7 S-T 3 7446-2402 VLP MINI-MOD 1.8mm Drill Short AO QC 1 11

Smith & Nephew, Inc. 1450 Brooks Road Memphis, TN 38116 USA www.smith-nephew.com Telephone: 1-901-396-2121 Information: 1-800-821-5700 Orders/Inquiries: 1-800-238-7538 Trademark of Smith & Nephew. Reg. US Pat. & TM Off. 2015 Smith & Nephew, Inc. All rights reserved. 00662 V3 71081167 REV0 01/15