Digital Breast Tomosynthesis

Similar documents
Tomosynthesis and Motion

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Mammography: Physics of Imaging

Mammography Solution. AMULET Innovality. The new leader in the AMULET series. Tomosynthesis, 3D mammography and biopsy are all available.

Imaging Technique Optimization of Tungsten Anode FFDM System

2 nd generation TOMOSYNTHESIS

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Patient-Assisted Compression Impact on Image Quality and Workflow

ADVANCED MEDICAL SYSTEMS PTE LTD Singapore Malaysia India Australia

THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

4/19/2016. Quality Control Activities for the RadiologicTechnologist. Objectives. 3D Tomosynthesis QC differences

Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities

Introduction. Digital Mammography QA: Comparing the Manufacturers Recommendations. What is QC and why is it important? Review & compare QC tests

GE Healthcare. Senographe 2000D Full-field digital mammography system

Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities

Collimation Assessment Using GAFCHROMIC XR-M2

New spectral benefi ts, proven low dose

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

GE Healthcare. Essential for life. Senographe Essential Full-Field Digital Mammography system

Published text: Institute of Cancer Research Repository Please direct all s to:

Maximizing clinical outcomes

Protocol for the Quality Control of the Physical and Technical Aspects of Digital Breast Tomosynthesis Systems

Quality Control for Stereotactic Breast Biopsy. Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc

Protocol for the Quality Control of the Physical and Technical Aspects of Digital Breast Tomosynthesis Systems

MAMMOGRAPHY - HIGH LEVEL TROUBLESHOOTING

Quality Control of Full Field Digital Mammography Units

TECHNICAL DATA. GIOTTO IMAGE SDL/W is pre-arranged for Full Field Digital Biopsy examination with the patient in prone position.

TITLE: Development and Clinical Evaluation of Tomosynthesis for Digital Mammography

Performance and care. all in one

Image Quality. HTC Grid High Transmission Cellular Grid provides higher contrast images

Improved Tomosynthesis Reconstruction using Super-resolution and Iterative Techniques

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions

CONTRACTING ORGANIZATION: Massachusetts General Hospital Boston, Massachusetts 02114

Facility, Unit and Test Equipment Data

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

Aspire HD. Program Manual. 2nd Edition - October 2012

Image Quality. HTC Grid High Transmission Cellular Grid provides higher contrast images

diagnostic examination

Digital Mammography Quality Control for the Mammographic Physicist

Thermionic x-ray. Alternative technologies. Electron Field Emission. CNT Based Field Emission X-Ray Source

Tailoring automatic exposure control toward constant detectability in digital mammography

FFDM -FCRm QC Requirements- What You REALLY Need to Know

GE Healthcare. Performa. High-performance breast imaging

Protocol for the Quality Control of the Physical and Technical Aspects of Digital Breast Tomosynthesis Systems

Distributed source x-ray tube technology for tomosynthesis imaging

Pitfalls and Remedies of MDCT Scanners as Quantitative Instruments

TITLE: Automatic Exposure Control Device for Digital Mammography

Mammograph FFDM PRODUCT DATA

ACPSEM Position Paper RECOMMENDATIONS FOR A DIGITAL MAMMOGRAPHY QUALITY ASSURANCE PROGRAM V4.0

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

Do you have any other questions? Please call us at (Toll Free) or , or

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

(12) United States Patent

4.0 How to Turn On the Selenia Dimensions

KODAK DIRECTVIEW CR Mammography Feature User s Guide

Understanding Quantra 2.0 User Manual

Current technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016

radiography detector

X-ray Tube and Generator Basic principles and construction

7/20/2014. Outline. Outline. Disclosures. Learning Objectives. SBB: Practical Aspects of ACR Accreditation, QC and ACR On Site Surveys

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Breast Imaging Basics: Module 10 Digital Mammography

8/2/2017. Radiologist Responsibilities. Radiologist Responsibilities. Medical Physicist Mammography Equipment Evaluation and Annual Survey

Digital mammography: DQE versus optimized image quality in clinical environment an on site study

DRAFT Technical evaluation of Philips Microdose SI digital mammography system

COMPUTED TOMOGRAPHY 1

Threshold contrast visibility of micro calcifications in digital mammography

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT)

Overview of Safety Code 35

K-edge subtraction X-ray imaging with a pixellated spectroscopic detector

Optimization of Digital Mammography Resolution Using Magnification Technique in Computed Radiography 1

Full-field digital mammography with grid-less acquisition and software-based scatter correction: Investigation of dose saving and image quality

The Script of ZST + Presentation. MIS Upstream Marketing Team [ 日期 ]

PRACTICE GUIDELINE FOR DETERMINANTS OF IMAGE QUALITY IN DIGITAL MAMMOGRAPHY

RADspeed Pro. EDGEpackage C501-E041C

2017 West Coast Educators Conference Orlando. Projection Geometry. 1. Review hierarchy of image qualities (amplified version):

AeroDR X60. Motorized X-ray System

1.1.Clinical Artefacts

Image Quality in Digital Mammography: Image Acquisition

Image Display and Perception

LECTURE 1 The Radiographic Image

X-RAY MEDICAL EQUIPMENT

10/15/2012 SECTION III - CHAPTER 6 DIGITAL FLUOROSCOPY RADT 3463 COMPUTERIZED IMAGING

Wide-Detector CT for TAVR Planning:

Features and Weaknesses of Phantoms for CR/DR System Testing

Redefining Ergonomics

X-ray detectors in healthcare and their applications

12/21/2016. Siemens Medical Systems Research Agreement Philips Healthcare Research Agreement AAN and ASN Committees

GE Healthcare. Senographe Crystal The choice is crystal clear

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

ACR AAPM SIIM Practice Guideline for Determinants of Image Quality in Digital Mammography

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

1. Patient size AEC. Large Patient High ma. Small Patient Low ma

PET/CT Instrumentation Basics

RaySafe Xi. Specifications

Research Support. Dual-Source CT: What is it and How Do I Test it? Cynthia H. McCollough, Ph.D.

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

Transcription:

Digital Breast Tomosynthesis OLIVE PEART MS, RT(R) (M) HTTP://WWW.OPEART.COM 2D Mammography Not 100% effective Limited by tissue superimposition Overlapping tissue can mask tumors False negative Overlapping structures can mimic tumors False positive False Negative False Positive 1

Development Tomosynthesis first described in the 1960s The basic principles of tomosynthesis were reviewed 1971 However DBT was not possible until the development of digital detectors in the early 1990s 1 st Tomosynthesis Application Massachusetts General Hospital First to prove DBT could be applied to evaluation the whole breast MGH partnered with GE to develop digital detectors In 1996, Mass General physicist Loren Niklason, PhD, and Dr. Kopans filed a patent entitled "Tomosynthesis System for Breast Imaging." Described a digital breast tomosynthesis (DBT) system that had the tube moving across the breast Development Technique Initial experiments Imaged phantoms and mastectomy material by manually moving the x-ray gantry to collect projection images from different angles 2

DBT Options The x-ray tube may be moved while the breast and detector are held stationary The tube and detector may move The motion of the two can vary 1 st DBT System GE Healthcare built the first whole-breast DBT system for US under grant BC970208 from the U.S. Army Flat panel digital detector 40º angle, 9 views The first volunteer patient was imaged in 1999 GE Healthcare Performed the first studies Involved several hundred volunteers, beginning in 2000 Showed that DBT was able to greatly improve the detection rates and reduce the recall rate 3

Limited-Angle Tomography DBT a form of limited-angle tomography Low-dose full field projection images of the breast obtained from different angles X-rays pass through the breast from different directions Parallax Fundamental principle behind DBT takes advantage of parallax from the projection images obtained at varying angles The greater the angle through which the x-ray tube is moved, the better is the z resolution Breast Tomosynthesis Structures located at different depth will project at different locations Structures closer to the detector will appear to move over a shorter distance when compared to structures that are farther away from the detector 4

Synthesize Planes After image acquisition The computer aligns all of the projection images so that structures in the plane of interest all align precisely Structures in the plane of interest are registered and reinforced by the number of projection images obtained Structures not in the plane of interest are misregistered with one another and fade into the background Shift And Add The computer shifts the images and add them again to obtain a different plane All planes are synthesized through the entire breast from a small number of projection images The structures in each plane are more clearly visible without the interference of tissue in front or in back of the plane of interest Synthesized Slice All of the projection images are included in every synthesized slice The information is misregistered Out-of-plane information fades into the background with the plane of interest reinforced as many times as the number of projection images obtained 5

Projection versus Reconstruction Sequence of projection images are acquired Projections are reconstructed in 3D volume Reconstructed Volume The breast volume is reconstructed and displayed through planes parallel to detector surface Projection slices Computer reconstruction 3D Volume Dose Each projection image requires only a fraction of the total dose of a full 2D mammogram Reason: All of the projection images are added together to synthesize the planes DBT can be performed at approximately the same total radiation dose used for 2D mammography 6

Pixel Manipulation To improve efficiency pixels are added together and averaged in some systems Pixel Binning Results in: Increased the pixel size Reduced spatial resolution Reduced visualization of calcifications Imaging Calcifications A cluster of calcifications may not be easily perceived with DBT Planes through the breast are presented as individual synthesized planes One plane may reveal a single calcification The next may contain two calcifications The next plane one calcification Consequently calcifications not visualized as a cluster Slabbing Slices are slabbed together Offer better visualization of calcifications The clusters become apparent in the volume 7

Imaging Typically, the number of images acquired ranges from approximately 10 to 25 Angle ranges from about 10 to 50 Hologic Digital Tomosynthesis 15 exposures in 4 sec Tube sweeps from -7 to 0 to +7 The x-ray beam is continuously on No movement of the patient DBT tomo.mp4 Hologic DBT Stats Tube Tungsten (W) Anode LFS: 0.3 mm; SFS: 0.1 m Detector a-se detector, 24 29 cm area 140 μm pixel size Reconstruction ~100 μm pixel size 1 mm thick slices NO GRID 8

Hologic DBT Stats Filtration Conv: 50 μm Rh; 50 μm Ag Tomo: 0.7 mm Al I View: 0.3 mm Cu Generator Max ma varies with kvp 20 39 kvp for 2D 20-49 kvp for tomosynthesis 200 ma max for LFS; 50 ma max for SFS 3D- Options The # of reconstruction images based on the breast thickness in mm Slices Slice #1 is closest to the detector. Highest slice # is closest to compression paddle Reconstruction is always in 1mm thick slices A breast 4 cm thick = 40 mm = 40 + 6 = 46 slices A breast 5cm thick =50 mm = 50 + 6 slices = 56 slices 9

Tomosynthesis and Motion Motion can occur at one point, multiple points or through-out the duration of the entire projection series Motion can occur at different areas of the breast, which may or may not impact breast tissue Repeats for motion increase radiation dose Potential to miss breast cancer Factors Contributing to Motion Unsharpness Inadequate Compression Poor Positioning Exposure Time Patient Movement Heart Motion 2D vs 3D Motion Unsharpness 2D Mammography Acquisition time is brief One image Technologists/ radiologists adept at detecting motion Repeats are left up to the Technologist 3D Tomosynthesis Longer acquisition time Multiple images acquired over a period of time Motion may go undetected 10

Appreciating 3D Motion: QC Review Projection Series Most efficient way to detect motion Review series at Selenia Dimensions System Tomosynthesis Reconstruction More difficult to detect/confirm motion on reconstruction Unsharpness in the tomosynthesis dataset Non-linear movement of calcifications Objects or lesions look sharp in one view, but not the opposing view Tomosynthesis and Motion 3D Motion in the Hologic System may be unrealized and unchecked Radiologists do not routinely review the projection dataset where motion can be confirmed or ruled-out Projection dataset may not be available to the radiologist (BTO) It is up to the technologist to detect motion and repeat when advised 2012 Hologic, Inc. All rights reserved. October 2012 rev001 PRE-00295 11

Appreciating 3D Motion Projection Series The x-ray tube moves in a path parallel to the chest wall The resulting breast image(s) and objects should move smoothly along this same pathway Medial to Lateral /Lateral to Medial Anterior/posterior movement of the breast images or objects indicates motion Appreciating 3D Motion Chest wall Movement of the Pectoral Muscle Structures that shift in and out of view Inframammary fold Abdomen motion Determine if it impacts the inferior and posterior breast Calcifications Should move in a straight line parallel to the chest wall More evident with large chunky calcifications Axilla Lymph Nodes shift back and forth or out of view 1 2 October 2012 2012 Hologic, Inc. All Rights Reserved rev001 PRE-00295 12

Reviewing Projection Images for Motion Review the 15 Projections Cine Mode Moderate to fast speed Reducing Motion Inform the patient Describe tube movement Explain how motion will affect the image Instruct the patient in the breathing technique Explain that STOP BREATHING means just that Patient SHOULD NOT take in a breath & hold it Breathing Technique Compress exposure controls While the x-ray tube is moving into position to start the tomosynthesis: Instruct patient to STOP breathing for the 3D acquisition At the conclusion of the tomosynthesis sweep Instruct patient to breathe As the tube moves to center, listen for the completion of the grid movement Then instruct the patient to stop breathing for the 2D acquisition 13

C-View Option DBT imaging No added radiation or imaging Reconstructed 2D image Generating 2D Images Perform a standard tomosynthesis scan Hologic Proprietary - For Educational Use Only - rev002 PRE-00380 14

Generating 2D Images Perform a standard tomosynthesis scan Reconstruct tomosynthesis slices 55 Tomosynthesis Slices* Reconstruction Algorithm *Average slices based on 5 cm compressed breast 15 Projection Images Hologic Proprietary - For Educational Use Only - rev002 PRE-00380 Generating 2D Images Perform a standard tomosynthesis scan Reconstruct tomosynthesis slices Synthesize 2D image Available in any tomosynthesis projection *Average slices based on 5 cm compressed breast 55 Tomosynthesis Slices* Software Algorithm Hologic Proprietary - For Educational Use Only - rev002 PRE-00380 June 2013 Generated 2D Images Advantages of C-View Scan Time Reduction 4 sec vs.12 sec Lowers Risk of Patient Motion Patient Dose Reduction 1.45 mgy vs. 2.65mGy 15

Image Comparison: Case 1 2D Tomosynthesis Slice Generated 2D (C-view) Hologic Proprietary - For Educational Use Only - rev002 PRE-00380 Radiation Dose The glandular dose deposited in any one region can vary considerably Glandular tissue and fatty tissue distribution Glandular tissue concentrated towards center of breast Mean glandular dose (MGD) or Average glandular dose (AGD) assumes a homogeneous mixture of adipose and glandular tissue surrounded by a layer of skin Average Glandular Dose ESE for a typical single exposure 800-1200mrad (8-12mGy) Glandular dose 100mrad (1.0mGy) ACR recommends 0.3 rad (300mrad or 3mGy) with a grid 0.1 rad (100mrad or 1mGy) without a grid 16

Radiation Dose 1.2 mgy 2D 1.45 mgy 3D 2.65 mgy COMBO Imaging implants in 3D = more radiation to patient because lower kvp used Drawbacks of Hologic DBT Motion artifacts hard to detect at radiologist workstation Projection vs reconstruction images Radiologist views reconstruction images Large number of images Degraded imaging of calcifications Slabbing will help Tomo not possible for FB, Mag and if the breast is more than 24.5 cm Total exposure time for COMBO = 12 sec GE SenoCare 9 exposures to acquire 3D MLO 25 degree scan angle Mo/Rh x-ray tube Reproduced with permission from http://www3.gehealthcare.com/static/senoclaire/index.html 17

FDA Approved Sequence Two 2D CC Two 3D MLO Radiation dose similar to 2D mode In the step-and-shoot method after each image the movement resumes and x-ray tube moves to the next position Comparison GE & Hologic While the gantry moves the x-ray beam is pulsed Results in longer exposure time (10 seconds) More chance of patient motion Reproduced with permission from http://www3.gehealthcare.com/static/senoclaire/index.html GE SenoCare Stats Image size 24 x 30 cm Angular range +/- 12.5 Target/filter Mo/Mo, Mo/Rh & Rh/Rh Filter material/thickness Mo: 0.03mm Rh: 0.025mm kvp Mo/Mo: 24-30 Mo/Rh: 26-32 Rh/Rh: 26-40 18

DBT and Grid Use Uses a step-andshoot system Eliminates focal spot motion blur Uses a moving grid Offers improved image quality by reducing scatter Imaging Calcifications Offers reconstruction of slabs (10 mm) and planes (0.5 mm or 1 mm) are reconstructed Uses ASIRDBT, an iterative reconstruction algorithm with a calcification artifact correction Motorized Tomo Device 100 micron pixel size no binning of pixels 5:1 Anti Scatter grid Not available for Magnification, CESM, or negative angle exposures 19

Technologist AWS Only 9 acquired images available to review positioning Reconstruction takes place at Review Workstation Drawbacks of GE SenoCare Motorized Tomosynthesis Device (MTD) attachment weighs over 12 kg (26.5 lb) and is not counterbalanced Must be attached for TOMO Large number of images Face plate moves with tube Attaching the DBT Device 20

Performing DBT V-Preview 2D image generated form the raw DBT projection data set 2D 3D V-Preview Reproduced with permission from http://www3.gehealthcare.com/static/senoclaire/index.html GE Senographe Pristina Digital detector technology: GEMS amorphous silicon matrix with CsI scintillator. Linear focused grid specially designed for mammography with Fiber interspaced Source to Image Distance (SID) of 660mm 21

Senographe Pristina Stats Dual track anode Molybdenum 0.03 mm Mo Rhodium 0.03 mm A Silver filter & Cesium Iodine detector Self-Compression Self-Compression Handheld wireless remote control that patients can use to adjust the compression force after breast positioning Senographe Pristina Operating kv-ranges Mo/Mo: 22-32 kv Rh/Ag: 27-40 kv Mo/Cu and Rh/Cu: 40-49 kv Designed to support future functionalities Contrast Enhanced Spectral Mammography Biopsy Mobile imaging 22

GE screening protocol 3D CC/MLO V-Preview CC/MLO When positioning for MLO the tube head can be moved to a parked position away from the technologist s head Automatic Modes AEC operating mode of Senographe Pristina AOP (Automatic Optimization of Parameters) AOP STD+: Provides a higher image contrast to noise ratio (CNR) at the cost of a higher dose AOP STD: Provides balanced CNR and dose choices, AOP DOSE-: Delivers a lower dose at the cost of a reduced CNR Siemens Mammomat Inspiration (PRIME & DBT) 50 Tomosweep (Continuous Scan) 25 projections <25 sec W/Rh only Grid removed a-se detector 23

Siemens MAMMOMAT Inspiration Offers the largest angular range -50 Highest number of projections images 25 Results: 25 projections for 3D reconstruction 50% dose by using the Tungsten/Rhodium Siemens Stats 3 different dose levels: Mo/Mo, Mo/Rh, or W/Rh. Direct-to-digital amorphous selenium (ase) High Detective Quantum Other Options Individualized OpComp Stops compression automatically Compresses only as long as a woman s breast is soft and pliable PRIME: Progressive Reconstruction Intelligently Minimizing Exposure allows less dose without compromising image quality 24

Contrast Enhancement Drawbacks of MAMMOMAT Inspiration Large number of images Face plate moves with tube Wide sweep FujiFilm Aspire DBT 25

Target/Filter Combo Target/filter: W/Rh and W/Al kvp range: 23-40 mas range: 2 600 Focal Spot: 0.1mm & 0.3mm Filter: 50 µm Rh Grid: 6:1, 40 lines /cm FujiFilm Aspire Stats Tube current: max 200mA Detector ase + TFT (hexagon pixels) Detector size: 24 x 30 cm Output Pixel size: 50 µm Exposure time: ST = less than 4 seconds HR = less than 10 seconds Fujifilm ASPIRE Cristalle The standard (SD) FFDM detector offers 3 Dose Modes H-(High), N (Normal) or L (Low) Only N-Mode is Released in the United States DBT images of the breast are acquired in the ST (standard) DBT mode with N-mode dose setting only 26

Standard (ST) Mode Good for assessment of calcification clusters and general assessment of breast architecture AGD 1.2 mgy for ST mode N-dose mode FujiFilm Aspire DBT HR mode image can visualize spiculation and marginal structure of tumor clearly due to higher depth resolution and higher plane resolution. ST mode offers less depth resolution when compared with an HR mode image The depth directional structure of lesions such as microcalcification is superior in ST mode. Processing: Pattern 1 enhances spiculations and calcifications while keeping maximum contrast for the viewing of masses within the glandular tissue. Pattern 2 maximizes the visualization of fine spiculations and calcifications. DBT Imaging DBT images are displayed individually or dynamically in a cine mode. The angular range is ±7 degrees (15 degrees total) 4 sec exposure time Slices are 1-mm thick Continuous tube movement 15 pulsed low dose exposures Slice range compression thickness + 5 mm Good for assessment of calcifications 27

Compression System Automatic compression 50N to 200N Manual compression 0N to 200N The images are acquired and reconstructed into a series of high-resolution 1 millimeter slices Ompression plate compression Automatic Exposure Control Intelligent AEC (iaec)auto mode: kvp selected based on compression thickness, mas determined by pre-exposure Semi-Auto: kvp selected by used, mas determined by pre-exposure Manual: all factors set by user 28

Radiation Dose PMMA Thickness (mm) AGD (mgy) 20 mm 1.0 mgy 30 mm 0.9 mgy 40 mm 1.3 mgy 45 mm 1.6 mgy 50 mm 1.7 mgy 60 mm 2.6 mgy 70 mm 3.2 mgy Thank You! 29