Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities

Similar documents
Quality Control of Full Field Digital Mammography Units

Introduction. Digital Mammography QA: Comparing the Manufacturers Recommendations. What is QC and why is it important? Review & compare QC tests

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Acceptance Testing of a Digital Breast Tomosynthesis Unit

FFDM in the Field: Physicist's Role in the QC of Mammography Laser Printers May Carl R. Keener, Ph.D., DABMP, DABR

FFDM -FCRm QC Requirements- What You REALLY Need to Know

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

4/19/2016. Quality Control Activities for the RadiologicTechnologist. Objectives. 3D Tomosynthesis QC differences

Aspire HD. Program Manual. 2nd Edition - October 2012

Quality Control for Stereotactic Breast Biopsy. Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

8/2/2017. Radiologist Responsibilities. Radiologist Responsibilities. Medical Physicist Mammography Equipment Evaluation and Annual Survey

Mammography: Physics of Imaging

NJDEP Medical Physicist s Radiographic QC Survey Registration Number:

MAMMOGRAPHY - HIGH LEVEL TROUBLESHOOTING

Ansur TNT Users Manual. Plug-In

7/20/2014. Outline. Outline. Disclosures. Learning Objectives. SBB: Practical Aspects of ACR Accreditation, QC and ACR On Site Surveys

Overview of Safety Code 35

REQUIREMENTS FOR LICENCE HOLDERS WITH RESPECT TO QUALITY CONTROL TESTS FOR DIAGNOSTIC X-RAY IMAGING SYSTEMS

Collimation Assessment Using GAFCHROMIC XR-M2

Facility, Unit and Test Equipment Data

Nuclear Associates

ACPSEM Position Paper RECOMMENDATIONS FOR A DIGITAL MAMMOGRAPHY QUALITY ASSURANCE PROGRAM V4.0

Digital Mammography Quality Control for the Mammographic Technologist

Image Display and Perception

Breast Imaging Basics: Module 10 Digital Mammography

Quality control for digital mammography: Part II recommendations from the ACRIN DMIST trial

THE ART OF THE IMAGE: IDENTIFICATION AND REMEDIATION OF IMAGE ARTIFACTS IN MAMMOGRAPHY

MILADY. Product Data. Page 1 of 8

Essentials of Digital Imaging

Imaging Technique Optimization of Tungsten Anode FFDM System

A Practical Overview of the Clinical and Operational Impact of Computed Radiography(CR) Implementations. Shirley Weddle, RT(R)(M), CIIP, BBA

Test Equipment for Radiology and CT Quality Control Contents

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance

DIAGNOSTIC ACCREDITATION PROGRAM. Radiology and CT Quality Control Procedures Workbook

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter

The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System

GE Healthcare. Senographe 2000D Full-field digital mammography system

Digital Breast Tomosynthesis

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

Exposure System Selection

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

Safelight Fog does what to contrast and density on film?

History of digital imaging

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

Acquisition, Processing and Display

Compliance Guidance for RADIOGRAPHIC QUALITY CONTROL (5 th Edition)

Image Quality. HTC Grid High Transmission Cellular Grid provides higher contrast images

Patient-Assisted Compression Impact on Image Quality and Workflow

Learning Objectives. Outline. Getting Started with CR. Converting the Radiology Department from Film-Screen to Digital: Making the Transition

Introduction of Computed Radiography in Two Mammography Services: Image Quality and Dose Analysis

Features and Weaknesses of Phantoms for CR/DR System Testing

QC Testing for Computed Tomography (CT) Scanner

Digital Mammography Quality Control for the Mammographic Physicist

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Comparison of computed radiography and filmõscreen combination using a contrast-detail phantom

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II

PRACTICE GUIDELINE FOR DETERMINANTS OF IMAGE QUALITY IN DIGITAL MAMMOGRAPHY

Learning Objectives: What s my motivation? (unknown screen actor) Workshop Overview

New spectral benefi ts, proven low dose

Estimation of signal transfer property for wireless digital detector in different measurement schemes

Outline. Digital Radiography. Understanding Digital Modalities: Image Quality and Dose. Image Quality. Dose Control

Nuclear Associates EZ CR-DIN Phantoms

Nuclear Associates

NEXT Protocol for the Survey of Pediatric Chest Radiography

Ludlum Medical Physics

SUSPENSION CRITERIA FOR IMAGE MONITORS AND VIEWING BOXES.

diagnostic examination

X-ray Tube and Generator Basic principles and construction

of sufficient quality and quantity

Get more from your images with Symphony Image Processing

KODAK DIRECTVIEW CR Mammography Feature User s Guide

GE Healthcare. Essential for life. Senographe Essential Full-Field Digital Mammography system

DRAFT Technical evaluation of Philips Microdose SI digital mammography system

7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline

Protocol for the Quality Control of the Physical and Technical Aspects of Digital Breast Tomosynthesis Systems

Digital Imaging Considerations Computed Radiography

Tailoring automatic exposure control toward constant detectability in digital mammography

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC?

Practical Medical Physics Session: TG-151 Dose Monitoring. August 5, 2013 Katie Hulme, M.S.

Get more from your images with Symphony Image Processing

4.0 How to Turn On the Selenia Dimensions

Half value layer and AEC receptor dose compliance survey in Estonia

Protocol for the Quality Control of the Physical and Technical Aspects of Digital Breast Tomosynthesis Systems

Digital radiography (DR) post processing techniques for pediatric radiology

Evaluation of a quality control phantom for digital chest radiography

Initial Survey Results on the Accuracy of Dose Data in the DICOM Objects

Third Edition DECEMBER Collaborating institutes: QARAD, BE LUCK, BE ARCADES, FR LRCB, NL EMIFMA, BE NCCPM, UK. Contributors:

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development

SYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology

TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist

Image Quality. HTC Grid High Transmission Cellular Grid provides higher contrast images

Digital Quality Control Phantom for X-ray Mammography Image

X-RAY MEDICAL EQUIPMENT

Technical Paper CONSISTENT PRESENTATION OF MEDICAL IMAGES

GE Healthcare. Performa. High-performance breast imaging

Optimization of Digital Mammography Resolution Using Magnification Technique in Computed Radiography 1

Performance and care. all in one

Open. the Digitized world. Fuji Computed Radiography

FMT18 FLOOR MOUNTED SYSTEM

Transcription:

Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities Melissa Martin, M.S., FAAPM, FACR, FACMP AAPM Annual Meeting - Philadelphia July 19, 2010 MO-B-204C-1

Educational Objectives To cover the annual tests to be performed by the medical physicist for both film/screen and digital mammography units To discuss efficient methods of performing these required tests To discuss the required tests for laser printers and monitors or processors and view boxes

Film/Screen MQSA Required Tests Follow the tests outlined in the ACR 1999 QC Manual for Film/Screen Mammography Physicists must use the Summary Forms from the ACR Website: www.acr.org Download forms for both Equipment Evaluations and Annual Surveys

FILM/SCREEN Summary Sheet

FFDM Equipment Evaluations and Surveys At the present time, the FDA requires the physicist to use the FFDM Manufacturer s QC Manual and Procedures. Physicist must download the Summary Forms for both Equipment Evaluations and Annual Surveys from the ACR website for the particular manufacturer s unit tested.

FFDM Equipment Evaluations and Surveys Physicist must follow the latest version of the Manufacturers QC Manual procedures for the unit tested (may not be the latest version available from that particular manufacturer if the unit has not had the latest upgrade/update installed) Test results must meet manufacturer s performance standards

FFDM Equipment Evaluations and Surveys Failures must be fixed before use on patients or within 30 days as specified for each particular test GE, Fuji, Hologic and Siemens have received FDA Approval for alternative standards for some QC test results to allow 30 days to achieve compliance with manufacturer s specifications

same as FDA s ACR s Current FFDM QC Requirements ACR suggests using Manufacturer s data forms Medical Physicists MUST complete the ACR s Summary Forms If the medical physicist passes a test, ACR accepts it. If the medical physicist fails a test, ACR requires corrective action documentation. If the medical physicist says NA or see comments, the ACR requests further clarification.

Lorad Selenia

LORAD Selenia QC Tests QC Manual tests: Medical Physicist Annually Radiologic Technologist Daily Weekly Biweekly Monthly Quarterly Semiannually

Medical Physics Annual Tests Mammographic Unit Assembly Evaluation Collimation Assessment Artifact Evaluation kvp Accuracy and Reproducibility HVL Measurement Evaluation of System Resolution Breast Entrance Exposure and Glandular Average Dose Radiation Output Rate Phantom Image Quality Evaluation Signal-To-Noise and Contrast-To-Noise Measurements Viewbox Luminance and Room Illuminance Softcopy Workstation QC

ACR Required Tests The ACR does require specific tests, common to all vendors: Weekly: Viewboxes and Viewing Conditions Monthly: Visual Checklist Semi-Annual: Compression force Taken from the ACR Mammography Quality Control Manual 1999 13

Daily and Weekly QC Tests (per Vendor) Test GE Fuji CR Siemens Hologic Fischer Monitor Cleaning Daily Laser Image Quality Test Weekly Daily Flat Field Calibration Weekly Weekly Weekly Weekly Artifact Evaluation Weekly Weekly Phantom Image Quality Weekly Weekly Daily Weekly Weekly Phantom Acquisition Test Signal-to-noise-ratio Weekly Weekly Weekly Contrast-to-noise ratio Weekly Weekly Weekly Weekly Viewbox and Viewing Conditions Weekly Weekly Compression Thickness Indicator Test Bi-weekly 14

Monthly and Quarterly QC Tests Test GE Fuji CR Siemens Hologic Fischer MTF Measurement DS/Essential weekly 2000D - monthly AOP Mode and SNR Monthly SNR Weekly SNR Weekly System Resolution System Operation Monthly Monthly Visual Checklist Monthly Monthly Monthly Repeat Analysis Quarterly Quarterly Quarterly Quarterly Quarterly Compression Force Soft Copy Workstation QC Laser Film Printer QC Semi-annually FDA guidance Per printer mfg Semiannually Semi-annually Semi-annually FDA guidance FDA guidance FDA guidance FDA guidance Per printer mfg Daily Weekly Mobile Unit QC At every move Unknown At every move At every move Image Plate Fog Semiannually Semiannually At every move 15

Laser Printer - must be able to print images on mammography laser film

Laser Printer Quality Control SMPTE Pattern from Acquisition Workstation Measure 10%, 40% and 90% Density Patches Daily Performance 40% density tracks within ±15% (40% - 10%) density tracks within ±15% 90% density tracks within ±15% NOTE: Printer uses linear LUT

OD Requirements for Hi-Resolution Laser Imagers D max > 3.5 OD Mid-density > 1.5 OD

Laser Processor QC Kodak daily sensitometry Base + Fog Density Difference OD closest to 2.20 minus OD closest but not less than 0.45 Mid-density step closest to but not less than 1.20 Action Limits: D max MD & DD + 0.15 OD B+Fog = 0.03 D max + 0.25

FDA Required Laser Printer QC GE 2000D & Senographe DS: Follow the laser printer manufacturer s QC manual Fischer Senoscan: Follow the laser printer manufacturer s QC manual Lorad Selenia: Follow the Lorad Selenia QC Manual Siemens Mammomat Novation DR: Follow the laser printer manufacturer s QC manual but conduct QC every day that images are printed. Fuji MammoCR System: Follow the laser printer manufacturer s QC manual

Viewing Conditions for Monitors From the ACR QC Manual, 1999, the room illuminance must be below 50 lux. From the February 2006 version of the Medical Physicist Equipment Evaluation forms from ACR, room illuminance for soft copy displays must be below 20 lux.

Hologic QC tests on SecurView Required testing is determined by monitor type CRT LCD (Flat panel) CRT Manually perform QC tests using photometer on each monitor to document the testing LCD QC tests are set up to run automatically at night Technologist must check results and record 23

Hologic QC tests on SecurView CRT: Cathode Ray Tube Weekly Calibration Settings Check Measure Display White Measure Display Black Monthly Measure Display Black Measure Display White Quarterly Uniformity Measurement LCD: Liquid Crystal Display Weekly Automatic I-Guard Check Calibration Settings Check Measure Display White Monthly Measure Quality Level 24

Softcopy Workstation QC 25

13. Diagnostic Review Workstation QC Date: Review Station Model: 18-May-10 SVDX-00400 Review Station SN: 29808071104 Left Monitor Right Monitor Comments Monitor Serial Number 1890097709 1890097714 Barco 5K Monitors White Level Performance 599.57 599.58 Target = 600 cd/sq m Black Level Performance NA - LCD Monitors NA - LCD Monitors Quality Level Performance OK OK Within 3% of 100% target Uniformity Performance NA - LCD Monitors NA - LCD Monitors Action Limit: The white level shall not be different from the monitor calibration level by more than 6%. The quality level performance shall not exceed 10%.

Phantom Image - Hologic Selenia FFDM Units Review Weekly Phantom Image Results Measured film densities Phantom objects seen (printer, monitors) 5 fibers 4 speck groups 4 masses

SNR and CNR Measurements ACR Accreditation Phantom with Disc (Weekly) Measure SNR and CNR from Image Selenia FFDM Units

SNR and CNR Measurements Passing Criteria - Selenia FFDM Units SNR at least equal or greater than 40 CNR should stay within ±15% of measurement obtained during acceptance testing of system

GE Senographe FFDM Units

QC Manuals for GE Senographe FFDM Units

GE Display Systems

CNR Test for Senographe 2000D

MTF Measurement - 2000D

Artifacts Physicist must test all Target/Filter combinations and both sizes of focal spots for artifacts

GE Senographe Essential Large and Small Fields Require Collimation Assessment of all three positions of small field plus large field for both target materials

GE Senographe Essential Control Panel Laterality must be selected prior to each exposure

GE Senographe Essential Measure Entrance Skin Exposure with Ion Chamber at 5 cm from outside edge of detector area.

Senographe DS & Essential

Seno DS & Essential - IQST Phantom

GE Senographe DS & QA Phantom Overview Essential Uniform area for noise power spectrum measurement Step wedge for contrast measurement Mesh for resolution uniformity measurement Edge object for MTF measurement Rulers for measuring distance of detector from chest wall edge of Bucky Tabs to position IQST with respect to chest wall edge of image receptor. Rails to position the IQST left-to-right

AOP Mode and SNR Check

AOP Test Materials

AOP Phantom Techniques

AOP Expected Parameters

CR Plates Fuji CRm System CR Plate Reader

Fuji CRm System S-Value Confirmation Measure Exposure Rate at Surface of Breast Support

Dynamic Range Test Fuji CRm System

Fuji CRm System - CNR Test CNR PER OBJECT THICKNESS Image Mode: Auto Time Focal Spot: 3 Density Control: 0 AEC Mode: Auto Time Phantom Thickness Target Filter kvp mas (Auto) mas (Manual) CNR CNR (relative to 4cm PMMA) Acceptable level 2 cm Mo Mo 25.0 27.2 28.0 17.197 114.4% >110% 4 cm Mo Mo 25.0 135.0 130.0 15.030 100.0% 100% 6 cm Mo Rh 30.0 282.0 280.0 13.765 91.6% >90% PASS

Inter Plate Consistency Exposure Unite ID: Lorad M-III Exposure conditions AECmode Tar get Filter kvp Auto Time Mo Mo 25 Group: Small Cassettes mas limit is plus or minus 10 % from mean. SNR limit is plus or minus 15 % from mean. Cassette ID ma s Accept able A456, #3 132 Yes A234, #2 132 Yes A123, #1 132 Yes A222, #4 132 Yes A111, #5 132 Yes SN R Accept able 70.9 46 Yes 70.9 52 Yes 70.1 52 Yes 70.1 66 Yes 69.5 21 Yes A777, #6 133 Yes 69.6 62 Yes

Fuji CRm - Monitor Tests Site: Med. Img Ctr of So Cal Date of Survey: 4/19/2007 Room #: Reading Room Date of Installation: Apr-07 DOME E5 Left Monitor Right Monitor Serial Numbers 703PNKN00011 703PNKN00028 Luminance Value Setting (cd/sq-m) 499.2 Target Value = 500 cd/sq-m 500.88 Is the CXtra icon present in the taskbar and shown as a green check mark? Manual Conformance report for both displays attached? Is DICOM calibration Graph for both displays attached? Yes Yes Yes No No No

Thank You!!! Melissa C. Martin, M.S., FACR Therapy Physics Inc. 879 West 190 St., Ste 419 Gardena, CA 90248 Office Phone: 310-217-4114 e-mail: Melissa@TherapyPhysics.com