4D dynamic imaging of the eye using ultrahigh speed SS- OCT

Similar documents
Ultrahigh speed volumetric ophthalmic OCT imaging at 850nm and 1050nm

MEMS tunable VCSEL light source for ultrahigh speed 60kHz - 1MHz axial scan rate and long range centimeter class OCT imaging

Axsun OCT Swept Laser and System

Visualization of human retinal micro-capillaries with phase contrast high-speed optical coherence tomography

Volumetric microvascular imaging of human retina using optical coherence tomography with a novel motion contrast technique

Handheld ultrahigh speed swept source optical coherence tomography instrument using a MEMS scanning mirror

Megahertz FDML Laser with up to 143nm Sweep Range for Ultrahigh Resolution OCT at 1050nm

Piezoelectric transducer based miniature catheter for ultrahigh speed endoscopic optical coherence tomography

Axsun Technologies Inc. Swept Laser based OCT Subsystems. May 2012

Extended coherence length megahertz FDML and its application for anterior segment imaging

Ultrahigh speed endoscopic optical coherence tomography using micro-motor imaging catheter and VCSEL technology

Ultrahigh Speed Spectral / Fourier Domain Ophthalmic OCT Imaging

High-speed imaging of human retina in vivo with swept-source optical coherence tomography

The TRC-NW8F Plus: As a multi-function retinal camera, the TRC- NW8F Plus captures color, red free, fluorescein

Optical coherence tomography

Spectral domain optical coherence tomography with balanced detection using single line-scan camera and optical delay line

Ultra High Speed Space Division Multiplexing OCT

some aspects of Optical Coherence Tomography

Simultaneously measuring ocular aberration and anterior segment biometry during accommodation

Swept source / Fourier domain polarization sensitive optical coherence tomography with a passive polarization delay unit

Fourier Domain (Spectral) OCT OCT: HISTORY. Could OCT be a Game Maker OCT in Optometric Practice: A THE TECHNOLOGY BEHIND OCT

Medical imaging has long played a critical role in diagnosing

Biology 70 Slides for Lecture 1 Fall 2007

A miniature all-optical photoacoustic imaging probe

Going beyond the surface of your retina OCT-HS100 OPTICAL COHERENCE TOMOGRAPHY

Going beyond the surface of your retina OCT-HS100 OPTICAL COHERENCE TOMOGRAPHY

Ultrahigh-resolution high-speed retinal imaging using spectral-domain optical coherence tomography

Static and dynamic crystalline lens accommodation evaluated using quantitative 3-D OCT

Going beyond the surface of your retina

Lecture 2 Slit lamp Biomicroscope

Phase-stabilized optical frequency domain imaging at 1-µm for the measurement of blood flow in the human choroid

Frequency comb swept lasers for optical coherence tomography

PHGY Physiology. SENSORY PHYSIOLOGY Vision. Martin Paré

OCT. Optical Coherence Tomography

Nature Neuroscience: doi: /nn Supplementary Figure 1. Optimized Bessel foci for in vivo volume imaging.

Cheng, KHY; Standish, BA; Yang, VXD; Cheung, KKY; Gu, X; Lam, EY; Wong, KKY

Kent Academic Repository

Image Modeling of the Human Eye

Real-Time Experimental Measurement of Swept Source VCSEL Properties Relevant to OCT Imaging

High-Coherence Wavelength Swept Light Source

3550 Aberdeen Ave SE, Kirtland AFB, NM 87117, USA ABSTRACT 1. INTRODUCTION

Transferring wavefront measurements to ablation profiles. Michael Mrochen PhD Swiss Federal Institut of Technology, Zurich IROC Zurich

Short Ring Cavity Swept Source Based on a Highly Reflective Chirped FBG

Frequency comb swept lasers

Frequency comb swept lasers

Design of a Swept-Source, Anatomical OCT System for Pediatric Bronchoscopy

60 MHz A-line rate ultra-high speed Fourier-domain optical coherence tomography

PHGY Physiology. The Process of Vision. SENSORY PHYSIOLOGY Vision. Martin Paré. Visible Light. Ocular Anatomy. Ocular Anatomy.

Temporal coherence characteristics of a superluminescent diode system with an optical feedback mechanism

Vision. The eye. Image formation. Eye defects & corrective lenses. Visual acuity. Colour vision. Lecture 3.5

Photoacoustic ophthalmoscopy for in vivo retinal imaging

PHY 431 Homework Set #5 Due Nov. 20 at the start of class

Split-spectrum amplitude-decorrelation angiography with optical coherence tomography

Optical Coherence Tomography Retina Scan Duo

Multi-Megahertz OCT: High quality 3D imaging at 20 million A-scans and 4.5 GVoxels per second

High Resolution Ocular Surface OCT to Directly Measure Tear Film Thickness in Human Eyes

AP PSYCH Unit 4.2 Vision 1. How does the eye transform light energy into neural messages? 2. How does the brain process visual information? 3.

High-speed ultra-broad tuning MEMS-VCSELs for imaging and spectroscopy

Customized Lasers for Specific Swept Source OCT Applications. Bill Ahern Axsun Technologies, Inc. June 20, 2013

Integre Pro Scan combines pattern scanning and multi-color photocoagulation in our unique all-in-one laser/slit lamp design.

Photoacoustic imaging using an 8-beam Fabry-Perot scanner

EnFocus Your Upgrade Path to High Performance Intrasurgical OCT

Introduction. Chapter Aim of the Thesis

Customized Correction of Wavefront Aberrations in Abnormal Human Eyes by Using a Phase Plate and a Customized Contact Lens

OPTICAL COHERENCE TOMOGRAPHY: OCT supports industrial nondestructive depth analysis

Wide-field high-speed space-division multiplexing optical coherence tomography using an integrated photonic device

OCULAR MEDIA* PHOTOGRAPHIC RECORDING OF OPACITIES OF THE. development by the control of diabetes, the supply of a deficient hormone

Ron Liu OPTI521-Introductory Optomechanical Engineering December 7, 2009

Pulsed illumination spectral-domain optical coherence tomography for human retinal imaging

Eye. Eye Major structural layer of the wall of the eye is a thick layer of dense C.T.; that layer has two parts:

Primary angle-closure glaucoma is a major cause of visual

Imaging the Subcellular Structure of Human Coronary Atherosclerosis Using 1-µm Resolution

III: Vision. Objectives:

Light has some interesting properties, many of which are used in medicine:

Diabetic Retinopathy Clinical Research Network (DRCR.net) UWF Optos 200Tx Imaging Protocol. Version 3.0 9/19/16

Optical Coherence Tomography. RS-3000 Advance

Diabetic Retinopathy Clinical Research Network (DRCR.net) UWF Optos Imaging Protocol. Version /14/14

2 The First Steps in Vision

Swept Wavelength Testing:

1272. Phase-controlled vibrational laser percussion drilling

High-speed spectral-domain optical coherence tomography at 1.3 µm wavelength

Understanding Optical Specifications

UC Davis UC Davis Previously Published Works

Single camera spectral domain polarizationsensitive optical coherence tomography using offset B-scan modulation

What s Fundus photography s purpose? Why do we take them? Why do we do it? Why do we do it? Why do we do it? 11/3/2014. To document the retina

The First True Color Confocal Scanner on the Market

Choices and Vision. Jeffrey Koziol M.D. Thursday, December 6, 12

Goldmann Visual Field. Humphrey Visual Field 4/25/2017. What s So Special About Special Testing?! Houston, we have a problem.

Akinori Mitani and Geoff Weiner BGGN 266 Spring 2013 Non-linear optics final report. Introduction and Background

Optical Coherence Tomography Systems and signal processing in SD-OCT

Off-axis full-field swept-source optical coherence tomography using holographic refocusing

Visual System I Eye and Retina

Swept source optical coherence tomography Gabor fusion splicing technique for microscopy of thick samples using a deformable mirror

Choices and Vision. Jeffrey Koziol M.D. Friday, December 7, 12

LMT F14. Cut in Three Dimensions. The Rowiak Laser Microtome: 3-D Cutting and Imaging

Optical Coherence Tomography. RS-3000 Advance 2

Dynamic Phase-Shifting Microscopy Tracks Living Cells

SUPPLEMENTARY INFORMATION

Version 1.0. th March 2011

Integrated photonic circuit in silicon on insulator for Fourier domain optical coherence tomography

Transcription:

4D dynamic imaging of the eye using ultrahigh speed SS- OCT The MT Faculty has made this article openly available. Please share how this access benefits you. Your story matters. Citation As Published Publisher Liu, Jonathan J., reneusz Grulkowski, Benjamin Potsaid, Vijaysekhar Jayaraman, Alex E. Cable, Martin F. Kraus, Joachim Hornegger, Jay S. Duker, and James G. Fujimoto. 4D Dynamic maging of the Eye Using Ultrahigh Speed SS-OCT. Edited by Fabrice Manns, Per G. Söderberg, and Arthur Ho. Ophthalmic Technologies XX (March 26, 2013). (2013) COPYRGHT Society of Photo-Optical nstrumentation Engineers (SPE) http://dx.doi.org/10.1117/12.2004369 SPE Version Final published version Accessed Thu Jun 07 13:06:43 EDT 2018 Citable Link Terms of Use Detailed Terms http://hdl.handle.net/1721.1/86411 Article is made available in accordance with the publisher's policy and may be subject to US copyright law. Please refer to the publisher's site for terms of use.

4-D dynamic imaging of the eye using ultrahigh speed SS-OCT Jonathan J. Liu 1, reneusz Grulkowski 1, Benjamin Potsaid 1,2, Vijaysekhar Jayaraman 3, Alex E. Cable 2, Martin F. Kraus 4, Joachim Hornegger 4, Jay S. Duker 5, and James G. Fujimoto 1 1 Department of Electrical Engineering and Computer Science & Research Laboratory of Electronics, Massachusetts nstitute of Technology, Cambridge, MA, USA 02139; 2 Advanced maging Group, Thorlabs nc., Newton, NJ, USA 07860; 3 Praevium Research nc., Santa Barbara, CA, USA 93111; 4 Pattern Recognition Lab & School of Advanced Optical Technologies, University Erlangen-Nuremberg, Erlangen, Germany 91054; 5 New England Eye Center & Tufts Medical Center, Tufts University, Boston, MA, USA 02116; ABSTRACT Recent advances in swept-source / Fourier domain optical coherence tomography (SS-OCT) technology enable in vivo ultrahigh speed imaging, offering a promising technique for four-dimensional (4-D) imaging of the eye. Using an ultrahigh speed tunable vertical cavity surface emitting laser (VCSEL) light source based SS-OCT prototype system, we performed imaging of human eye dynamics in four different imaging modes: 1) Pupillary reaction to light at 200,000 axial scans per second and 9 μm resolution in tissue. 2) Anterior eye focusing dynamics at 100,000 axial scans per second and 9 μm resolution in tissue. 3) Tear film break up at 50,000 axial scans per second and 19 μm resolution in tissue. 4) Retinal blood flow at 800,000 axial scans per second and 12 μm resolution in tissue. The combination of tunable ultrahigh speeds and long coherence length of the VCSEL along with the outstanding roll-off performance of SS-OCT makes this technology an ideal tool for time-resolved volumetric imaging of the eye. Visualization and quantitative analysis of 4-D OCT data can potentially provide insight to functional and structural changes in the eye during disease progression. Ultrahigh speed imaging using SS-OCT promises to enable novel 4-D visualization of realtime dynamic processes of the human eye. Furthermore, this non-invasive imaging technology is a promising tool for research to characterize and understand a variety of visual functions. Keywords: 4D, SS-OCT, VCSEL, ophthalmology, pupil, accommodation, tear film, retinal blood flow 1. NTRODUCTON Recent developments in swept-source and spectral domain optical coherence tomography (OCT) technologies have enabled ultrahigh speed OCT imaging at over 100,000 axial scans (A-scans) per second [1-4]. Novel ultralong range and ultrahigh speed imaging of the eye has been demonstrated using SS-OCT with vertical cavity surface emitting laser (VCSEL) technology [5]. n this paper, we demonstrate time resolved volumetric data using four-dimensional (4-D) OCT with speeds up to forty volumes per second. We imaged pupillary reactions, anterior eye focusing dynamics, tear film break up, and retinal blood flow. The real-time dynamics of the human eye is captured and analyzed for characterizing visual functions. 2. MATERALS AND METHODS We developed a VCSEL light source based prototype SS-OCT system (Figure 1) [5]. The laser was operated in four different modes with center wavelengths at around 1060 nm (Table 1). To image the pupillary reaction, the VCSEL light source was operated at an axial scan rate of 200 khz with 85 nm tuning range yielding 9 μm resolution and 5.6 mm Nyquist-limited imaging range in tissue. To image anterior eye accommodation dynamics, the axial scan rate was reduced to 100 khz to achieve a 10.3 mm Nyquist-limited imaging range in tissue. To image tear film break up, the VCSEL light source was operated at a reduced axial scan rate of 50 khz and reduced 30 nm tuning range for 19 μm Ophthalmic Technologies XX, edited by Fabrice Manns, Per G. Söderberg, Arthur Ho, Proc. of SPE Vol. 8567, 85670X 2013 SPE CCC code: 1605-7422/13/$18 doi: 10.1117/12.2004369 Proc. of SPE Vol. 8567 85670X-1

resolution and 60 mm Nyquist-limited imaging range in tissue. Finally, to image retinal blood flow, the VCSEL light source was operated at an axial scan rate of 800 khz and 50 nm tuning range for 12 μm resolution and 2.5mm Nyquista fixed 1 limited imaging range in tissue. A/D data acquisition cards (ATS9870 & ATS9350; AlazarTech) at either GSPS or a fixed 500 MSPS sampling rate with dual channel acquisition were used. The incident power on the cornea for all imaging modes was 1.9 mw, within the American National Standard nstitute (ANS) safety limits. Studies were performed under protocols approved by the MT Committee on n the Use of Humans as Experimental Subjects (COUHES) and written consent from participants was obtained. (A) Referernce MZ (B) VCSEL j Module r 80:20 'k \._ 8c Sample Arm Patient nterface SC 1 Anterior - Segment Full -Eye- Length nterface FT J Sc ox T i L _2 1 \ U S,H VTM RR 1 1 Reference Figure 1. (A) VCSEL based SS-OCT prototype setup. (B) Different interface configurations. Full-eye-length imaging was performed by using a longer focal length lens in the anterior segment configuration. Retinal imaging was performed by adding ocular lens to the anterior segment configuration and adjusting thee fixation target path. SC galvanometric scanners, FT fixation target, DM dichroic mirror, DC dispersion compensation glass, RR retroreflector, PDB balanced photodetectors, MZ Mach-Zehnder interferometer. Table 1. Configuration of VCSEL SS-OCT system. Parameter Effective imaging speed Wavelength tuning range 85 nm Axial resolution (tissue) Sample arm interface Transversee resolution Acquisition card Pupillary response 200,000 A-scans/s 100,000 A-scans/s 50,0000 A-scans/s 9 µm Anteriorr segment 30 µm ATS9870 (1 GSPS) Accommodat tion 85 nm 9 µm Anterior segment 30 µm ATS9870 (1 GSPS) Tear film break up 30 nm 19 µm Full eye length 73 µm ATS9870 (1 GSPS) Retinal blood flow 800,000 A-scans/s 50 nm 12 µm Retinal 25 µm ATS9350 (500 MSPS) Proc. of SPE Vol. 8567 85670X-2

3. RESULTSS AND DSCUSSON 3.1 Pupillary response The iris is a dynamic structure whose configuration regularly changes in response to light and during accommodation. Dynamic changes in intraocular structures caused by illumination or dilation are suggested to be risk factors for glaucoma development. Studying the dynamic response of the pupil to dark-light stimulus may provide a more comprehensive assessment of risks to primary-angle closure development and may help understand the pathophysiology of angle closure glaucoma [6-8]. We demonstrated 4-D imaging of the pupil response to light stimulus from an LED positioned adjacent to the eye, where sequential 150 x 150 axial-scan volumes over 17 mm x 17 mm were acquired at ~8 volumes per second for 5 seconds with an axial scan rate of 200 khzz enabling the visualization of changes in the iris in time and in three dimensions (Video 1). The pupillary response is shown in Figuree 2 where two time points in the 4-D data are displayed. The iris response to light stimulus can be quantitatively analyzed by measuring pupil size/area changes in time. As shown in Figure 2(A), the pupil area decreased drastically when light stimulus was applied. n contrast, the time constant of the pupil diameter recovery was longer than when stimulus was applied. Video 1. 4-D OCT imaging of the pupillary reflex. http://dx.doi.org/10.1117/12.2004369.1 Light stimulus.5 1 1.5 ;.5 4.5 sec M! JO' J 1 Figure 2. 4-D OCT imaging of the pupillary reflex. (A) Plot of pupil areaa vs time. (B) Volume#1 at 0.13 seconds. (C) Volume# #19 at 2.38 seconds. (D) Volume#38 at 4.5 seconds. The constriction of the iriss and the decrease in pupil area is apparent when comparing volume#1 and volume#19. The dilation of thee iris and increase in pupil area is visible when comparing volume#19 and volume #38. The cross-sectional OCT images show the iris structure while the enface OCT image depicts the pupil area. 1 Proc. of SPE Vol. 8567 85670X-3

3.2 Accommodation dynamics n the anterior eye, accommodation is the process by which the eye focuses on near objects. This occurs mainly through the deformation of the crystalline lens geometry. Studies of the dynamics of far to near focusing processes may enable understanding several aspects like the age-dependent loss of accommodative amplitude in presbyopia, the development of myopia, and the ageing of the crystalline lens as well as the mechanism of accommodation [9]. We performed 4-D imaging of the focusing process in the anterior eye acquiring 100 x 100 axial-scan volumes over 15mm x 15mmm at ~10 volumes per second for 5 seconds at an axial scan rate of 100 khz (Video 2). The target was moved in position from far to near and the subject was asked to focus on the target and accommodate the target s position change. Two time points are shown from the 4-D data in Figure 3. Pupil diameter decreasess when the eyee focuses at near positions. The iris becomes smaller while the crystalline lens becomes thicker with surfaces curvature changes. r L Video 2. 4-D OCT imaging of the anterior eye focusing on a near object. http://dx.doi.org/10.1117/12.2004369.2 Figure 3. 4-D OCT imaging of the anterior eye focusing on a near object. A) Volume#55 at 0.5 seconds. B) Volume #30 at 3 seconds. The decrease in pupil size and increase in lens thickness is observed and measured. 3.3 Tear film break up The tear film is the first refractive surface for light incident on the eye and plays an important role in the optical quality of the human eye. Tear film dynamics has been used to evaluate the tear system and diagnose dry eye. Tear film break up time (TBUT) is defined as the time interval between a complete blink and the appearancee of the first randomly distributed dry spot [10]. A TBUT of less than ten seconds is considered abnormal [11]. Fluorescein dye is typically Proc. of SPE Vol. 8567 85670X-4

used for clinical TBUT measurements. OCT allows for non-invasive tear film dynamics imaging without the need of fluorescein staining. We implemented a 4-D scan protocol with 300 x 300 axial-scan volumes over 8.5 mm x 8..5 mm at ~0.5 volumes per second for 20 seconds at an axial scan rate of 50 khz. n order to best visualize tear film break up, we examine the shadowing effect by summing the signal between thee anterior surface to the posterior surface of the crystalline lens (Video 3). Tear film break up can be clearly observedd in the frame-by-frame breakdown (Figure 4). a ii he region of enfa Video 3. llustration of th ace summation in a full-eye-length OCT dataset at a single time point for tear film visualization. http://dx.doi.org/10.1117/12.2004369.3 Figure 4. Tear film break up observed in enface images extracted from 4-D OCT data. beginningg at ~12 secondss where vertical streaking patterns appear. Tear film break up is observed 3.4 Retinal blood flow Retinal blood flow may play a role in the pathogenesis of diseases such as diabetic retinopathy, age-related macular degeneration, and glaucoma. The ability to monitor retinal blood flow may be important not only for treatment, but also for understanding the pathophysiology of these diseases [12-14]. We acquired 4-D data in the retina with 200 x 100 axial-scan volumes over 1 mm x 0.5 mm at ~40 volumes per second for 1 second at an axial scan rate of 800 khz (Video 4). Since the galvanometer scanners are driven at very high speeds, a conservative flyback pattern is chosen resulting in only a 0.5 mm x 0.5 mm evenly sampled region. mages from a single time point is extracted and shown in Figure 5(B)(C)(D). By taking the variance of maximum intensity projectionn images at alll time points from multiple volumes, we are able to visualize capillaries from the inner to the outer plexiform layers (Figure 5(E)). Proc. of SPE Vol. 8567 85670X-5

NL-OPL Video 4. 4-D OCT imaging of the retina [15]. The video is slowed downn 5x with 0.2x playback speed. GCL: ganglion cell layer. PL-NL: inner plexiform layer to inner nuclear layer. NL-OPL: inner nuclearlayer to outer plexiform layer. http://dx.doi.org/10.1117/12.2004369.4 Figure 5. 4-D OCT imaging of retinal blood flow. A) Cross-sectional OCT image. B) Enface summation image in the axial direction of a single volume. C) Enface summation image of a region of interest in the axial direction of a single volume. D) Maximum projection image of a region of interest in the axial direction of a single volume. E) Variance image generated from maximum projection images (like E) of multiple volumes in the 4-DD data. The maximum projection image from a single volume shows moving highly reflective scatters. The variance image generated from multiple volumes highlight retinal capillaries. 4. CONCLUSONS n conclusion, 4-D OCT imaging was demonstrated in the human eyee using ultrahigh speed SS-OCT. Visualization and quantitative analysis of 4-D OCT imaging can potentially provide insight to functional and structural changes in the eye with disease. The tunable VCSEL light source enables different imaging modes for variable imaging ranges, speeds, and resolutions. Ultrahigh speed imaging using SS-OCT promises to enable novel 4-D visualization of real-time dynamics of the human eye and may benefit studiess by visualizing and characterizing a variety of visual functions. ACKNOWLEDGMENTS The authors want to thank Dr. Al-Hafeez Dhalla, WooJhon Choi, Chen D. Lu, Kathrin J. Molher, and Tsung-Hann Tsai at MT for technical discussions. Financial support from the National nstitutes of Health (NH R01-EY011289-27, R01- EY013178-12, R01-EY013516-09, R01-EY018184-05, R01-EY019029-04, R01-CA075289-14, R01-HL095717-04, R01-NS057476-05), Air Force Office of Scientific Research (AFOSRR FA9550-10-1-0063), Medical Free Electron Laser Program (FA9550-10-1-0551) and Deutsche Forschungsgesellschaft (DFG-GSC80-SAOT, DFG-HO-1791/ /11-1) is gratefully acknowledged. Dr. reneusz Grulkowski acknowledges partial support from KOLUMB Fellowship, Foundation for Polish Science (KOL/3/2010-). Dr. Jay S. Duker is supported by an unrestricted grant from Research to Prevent Blindness as well as the Massachusetts Lions Clubs. Dr. reneusz Grulkowski is a visiting scientist from the nstitute of Physics, Nicolaus Copernicus University, Torun, Poland. Proc. of SPE Vol. 8567 85670X-6

REFERENCES [1] B. Potsaid,. Gorczynska, V. J. Srinivasan, Y. Chen, J. Jiang, A. Cable and J. G. Fujimoto, "Ultrahigh speed spectral / Fourier domain OCT ophthalmic imaging at 70,000 to 312,500 axial scans per second," Opt Express 16(19), 15149-15169 (2008) [2] M. Gora, K. Karnowski, M. Szkulmowski, B. J. Kaluzny, R. Huber, A. Kowalczyk and M. Wojtkowski, "Ultra high-speed swept source OCT imaging of the anterior segment of human eye at 200 khz with adjustable imaging range," Opt Express 17(17), 14880-14894 (2009) [3] B. Potsaid, B. Baumann, D. Huang, S. Barry, A. E. Cable, J. S. Schuman, J. S. Duker and J. G. Fujimoto, "Ultrahigh speed 1050nm swept source/fourier domain OCT retinal and anterior segment imaging at 100,000 to 400,000 axial scans per second," Opt Express 18(19), 20029-20048 (2010) [4] T. Klein, W. Wieser, C. M. Eigenwillig, B. R. Biedermann and R. Huber, "Megahertz OCT for ultrawide-field retinal imaging with a 1050 nm Fourier domain mode-locked laser," Opt Express 19(4), 3044-3062 (2011) [5]. Grulkowski, J. J. Liu, B. Potsaid, V. Jayaraman, C. D. Lu, J. Jiang, A. E. Cable, J. S. Duker and J. G. Fujimoto, "Retinal, anterior segment and full eye imaging using ultrahigh speed swept source OCT with vertical-cavity surface emitting lasers," Biomed Opt Express 3(11), 2733-2751 (2012) [6] C. K. Leung, C. Y. Cheung, H. Li, S. Dorairaj, C. K. Yiu, A. L. Wong, J. Liebmann, R. Ritch, R. Weinreb and D. S. Lam, "Dynamic analysis of dark-light changes of the anterior chamber angle with anterior segment OCT," nvest Ophthalmol Vis Sci 48(9), 4116-4122 (2007) [7] C. Y. Cheung, S. Liu, R. N. Weinreb, J. Liu, H. Li, D. Y. Leung, S. Dorairaj, J. Liebmann, R. Ritch, D. S. Lam and C. K. Leung, "Dynamic analysis of iris configuration with anterior segment optical coherence tomography," nvest Ophthalmol Vis Sci 51(8), 4040-4046 (2010) [8] Y. Lee, K. R. Sung, J. H. Na and J. H. Sun, "Dynamic changes in anterior segment (AS) parameters in eyes with primary angle closure (PAC) and PAC glaucoma and open-angle eyes assessed using AS optical coherence tomography," nvest Ophthalmol Vis Sci 53(2), 693-697 (2012) [9] M. Ruggeri, S. R. Uhlhorn, C. De Freitas, A. Ho, F. Manns and J. M. Parel, "maging and full-length biometry of the eye during accommodation using spectral domain OCT with an optical switch," Biomed Opt Express 3(7), 1506-1520 (2012) [10] M. A. Lemp and J. R. Hamill, Jr., "Factors affecting tear film breakup in normal eyes," Arch Ophthalmol 89(2), 103-105 (1973) [11] M. A. Lemp, C. H. Dohlman, T. Kuwabara, F. J. Holly and J. M. Carroll, "Dry eye secondary to mucus deficiency," Trans Am Acad Ophthalmol Otolaryngol 75(6), 1223-1227 (1971) [12] B. Baumann, B. Potsaid, M. F. Kraus, J. J. Liu, D. Huang, J. Hornegger, A. E. Cable, J. S. Duker and J. G. Fujimoto, "Total retinal blood flow measurement with ultrahigh speed swept source/fourier domain OCT," Biomed Opt Express 2(6), 1539-1552 (2011) [13] Y. Jia, O. Tan, J. Tokayer, B. Potsaid, Y. Wang, J. J. Liu, M. F. Kraus, H. Subhash, J. G. Fujimoto, J. Hornegger and D. Huang, "Split-spectrum amplitude-decorrelation angiography with optical coherence tomography," Opt Express 20(4), 4710-4725 (2012) [14] W. Choi, B. Baumann, J. J. Liu, A. C. Clermont, E. P. Feener, J. S. Duker and J. G. Fujimoto, "Measurement of pulsatile total blood flow in the human and rat retina with ultrahigh speed spectral/fourier domain OCT," Biomed Opt Express 3(5), 1047-1061 (2012) [15] T. Schmoll, C. Kolbitsch and R. A. Leitgeb, "Ultra-high-speed volumetric tomography of human retinal blood flow," Opt Express 17(5), 4166-4176 (2009) Proc. of SPE Vol. 8567 85670X-7