A Real-time Photoacoustic Imaging System with High Density Integrated Circuit

Similar documents
Non-contact Photoacoustic Tomography using holographic full field detection

Improving the Quality of Photoacoustic Images using the Short-Lag Spatial Coherence Imaging Technique

Derek Allman a, Austin Reiter b, and Muyinatu Bell a,c

Capacitive Micromachined Ultrasonic Transducers (CMUTs) for Photoacoustic Imaging

Synthetic-aperture based photoacoustic re-beamforming (SPARE) approach using beamformed ultrasound data

Photoacoustic imaging using an 8-beam Fabry-Perot scanner

Technical Datasheet UltraScope USB

15, Route de Douchy CHUELLES - Tél. : Fax : US-Scan.

Parametric Beamformer for Synthetic Aperture Ultrasound Imaging

EMBEDDED DOPPLER ULTRASOUND SIGNAL PROCESSING USING FIELD PROGRAMMABLE GATE ARRAYS

A miniature all-optical photoacoustic imaging probe

2. The design and realization of the developed system

Multiplierless sigma-delta modulation beam forming for ultrasound nondestructive testing

Simulation of Algorithms for Pulse Timing in FPGAs

FPGA Co-Processing Solutions for High-Performance Signal Processing Applications. 101 Innovation Dr., MS: N. First Street, Suite 310

A Delta-Sigma beamformer with integrated apodization

Implementing Logic with the Embedded Array

COMPUTER PHANTOMS FOR SIMULATING ULTRASOUND B-MODE AND CFM IMAGES

Ultrasound Beamforming and Image Formation. Jeremy J. Dahl

Design & Development of 4-channel Phased Array Control & Amplifier for EMAT based Phased Array UT System for Weld Joints

Ultrasound Bioinstrumentation. Topic 2 (lecture 3) Beamforming

Multimodal simultaneous photoacoustic tomography, optical resolution microscopy and OCT system

Linear arrays used in ultrasonic evaluation

FPGA-BASED CONTROL SYSTEM OF AN ULTRASONIC PHASED ARRAY

DESIGN OF HIGH-PERFORMANCE ULTRASONIC PHASED ARRAY EMISSION AND RECEPTION CON- TROLLING SYSTEM

Software Design of Digital Receiver using FPGA

The Simulation for Ultrasonic Testing Based on Frequency-Phase Coded Excitation

Multi-Element Synthetic Transmit Aperture Method in Medical Ultrasound Imaging Ihor Trots, Yuriy Tasinkevych, Andrzej Nowicki and Marcin Lewandowski

AN IMPLEMENTATION OF MULTI-DSP SYSTEM ARCHITECTURE FOR PROCESSING VARIANT LENGTH FRAME FOR WEATHER RADAR

A CMOS Analog Front-End for Driving a High-Speed SAR ADC in Low-Power Ultrasound Imaging Systems

All-optical endoscopic probe for high resolution 3D photoacoustic tomography

Ihor TROTS, Andrzej NOWICKI, Marcin LEWANDOWSKI

Journal of Engineering Science and Technology Review 9 (5) (2016) Research Article. L. Pyrgas, A. Kalantzopoulos* and E. Zigouris.

FTMS Booster X1 High-performance data acquisition system for FT-ICR MS

Imaging serial interface ROM

Novel fibre lasers as excitation sources for photoacoustic tomography and microscopy

Development and Application of 500MSPS Digitizer for High Resolution Ultrasonic Measurements

Nuove tecnologie per ecografia ad ultrasuoni: da 2D a 4D

Reconfigurable Arrays for Portable Ultrasound

Acoustic resolution. photoacoustic Doppler velocimetry. in blood-mimicking fluids. Supplementary Information

The Hardware Design of Partial Discharge Online Monitoring for Large Power Transformers System

STUDIES in oncology have shown that angiogenesis, the

Application of a Novel Software Algorithm for Information Reduction in High Frame Rate Ultrasonography

Light emitting diodes as an excitation source for biomedical photoacoustics

Design Implementation Description for the Digital Frequency Oscillator

arxiv: v2 [physics.med-ph] 29 Jan 2018

Low Complex, Programmable FPGA based 8-Channel Ultrasound Transmitter for Medical Imaging Researches

AREA EFFICIENT DISTRIBUTED ARITHMETIC DISCRETE COSINE TRANSFORM USING MODIFIED WALLACE TREE MULTIPLIER

Ultrasound-modulated optical tomography of absorbing objects buried in dense tissue-simulating turbid media

System Architecture of an Experimental Synthetic Aperture Real-time Ultrasound System

Development of a 256-channel Time-of-flight Electronics System For Neutron Beam Profiling

FMC ADC 125M 14b 1ch DAC 600M 14b 1ch Technical Specification

4. Embedded Multipliers in the Cyclone III Device Family

DIGITAL FILTERING OF MULTIPLE ANALOG CHANNELS

Stratix II DSP Performance

FPGA-Based Control System of an Ultrasonic Phased Array Keywords: ultrasonic imaging, phased array, B-scan, FPGA

The Design and Simulation of Embedded FIR Filter based on FPGA and DSP Builder

Curved array photoacoustic tomographic system for small animal imaging

FYS3240 PC-based instrumentation and microcontrollers. Signal sampling. Spring 2015 Lecture #5

Analog front-end electronics in beam instrumentation

A DSP IMPLEMENTED DIGITAL FM MULTIPLEXING SYSTEM

A low dead time vernier delay line TDC implemented in an actel flash-based FPGA

Multi-depth photoacoustic microscopy with a focus tunable lens

APPLICATION OF PROGRAMMABLE LOGIC DEVICES FOR ACQUISITION OF ECG SIGNAL WITH PACEMAKER PULSES 1. HISTORY OF PROGRAMMABLE CIRCUITS

Acousto-optic imaging of tissue. Steve Morgan

Wideband Focused Transducer Array for Optoacoustic Tomography

Decision Based Median Filter Algorithm Using Resource Optimized FPGA to Extract Impulse Noise

A new Photon Counting Detector: Intensified CMOS- APS

4. Embedded Multipliers in Cyclone IV Devices

Resolution Enhancement and Frequency Compounding Techniques in Ultrasound.

Design of Multiplier Less 32 Tap FIR Filter using VHDL

FYS3240 PC-based instrumentation and microcontrollers. Signal sampling. Spring 2017 Lecture #5

A new Photon Counting Detector: Intensified CMOS- APS

Design and Implementation of Signal Processor for High Altitude Pulse Compression Radar Altimeter

Photoacoustic tomography of biological tissues with high cross-section resolution: Reconstruction and experiment

A10-Gb/slow-power adaptive continuous-time linear equalizer using asynchronous under-sampling histogram

Signal Processing and Display of LFMCW Radar on a Chip

Photoacoustic imaging with coherent light

Integrated 16-channel Transmit and Receive Beamforming ASIC for Ultrasound Imaging

Digital Logic, Algorithms, and Functions for the CEBAF Upgrade LLRF System Hai Dong, Curt Hovater, John Musson, and Tomasz Plawski

Session: 2A NEW ULTRASOUND SYSTEMS Chair: H. Ermert University of Bochum 2A-1 10:30 a.m.

Transmission- and side-detection configurations in ultrasound-modulated optical tomography of thick biological tissues

THIS work focus on a sector of the hardware to be used

Design of Electromagnetic Ultrasonic Data Acquisition and Analysis System Based on USB

An ultra-high ramp rate arbitrary waveform generator for communication and radar applications

Techniques for Implementing Multipliers in Stratix, Stratix GX & Cyclone Devices

MAKING TRANSIENT ANTENNA MEASUREMENTS

Implementation of Pixel Array Bezel-Less Cmos Fingerprint Sensor

Photoacoustic tomography imaging based on a 4f acoustic lens imaging system

A FPGA Based Platform for Multi-Frequency Eddy Current Testing

Time Matters How Power Meters Measure Fast Signals

ISSN: ISO 9001:2008 Certified International Journal of Engineering Science and Innovative Technology (IJESIT) Volume 2, Issue 4, July 2013

Comprehensive Ultrasound Research Platform

Appendix B. Design Implementation Description For The Digital Frequency Demodulator

Image acquisition. In both cases, the digital sensing element is one of the following: Line array Area array. Single sensor

JESD204A for wireless base station and radar systems

INDIAN INSTITUTE OF TECHNOLOGY-KANPUR (DEPARTMENT OF MECHANICAL ENGINEERING)

VC7300-Series Product Brief

Corona Current Data Acquisition Card Based on USB Bus in Extra High Voltage Environment

High Performance DSP Solutions for Ultrasound

Transcription:

2011 3 rd International Conference on Signal Processing Systems (ICSPS 2011) IPCSIT vol. 48 (2012) (2012) IACSIT Press, Singapore DOI: 10.7763/IPCSIT.2012.V48.12 A Real-time Photoacoustic Imaging System with High Density Integrated Circuit Naizhang Feng*, Mingjian Sun and Liyong Ma Department of Control Science and Engineering, Harbin Institute of Technology, Harbin 150001, China *Email: fengnz@yeah.net Abstract. Based on an array transducer, a real-time photoacoustic imaging system with high density integrated circuit is built. The system is composed of a short pulsed laser, an ultrasonic transducer array, a data acquisition circuit and a computer. The data acquisition circuit consists of 128-channel variable gain amplifiers, 128-channel anti-aliasing filters, 128-channel analog-to-digital converters, and field programmable gate arrays. Biological tissues are irradiated by the laser to generate the thermoelastic expansion. The transducer array is used to detect the photoacoustic signals generated from the tissue thermoelastic expansion. Then the photoacoustic signals are amplified and sampled by multiple parallel data acquisition channels. Subsequently, the sampled data set is transferred to the computer through an USB interface. The final photoacoustic image is reconstructed by the computer. The traditional delay-and-sum beamforming method is used as the reconstruction algorithm. An experiment with a phantom of two hairs is made. The result shows that two hairs can be displayed clearly in the reconstructed image and the imaging rate is the same as the pulse repetition frequency of the laser. Keywords: Photoacoustic Imaging; Real time; Data acquisition Channel; Beamforming 1. Introduction As a new technique, photoacoustic imaging (PAI) for biological tissues has been proposed and developed in recent years[1,2]. It has demonstrated great potential for visualization of the internal structures and function of soft tissue. When biological tissues are irradiated by a short pulsed laser, some energy is absorbed to generate the transient thermoelastic expansion. The expansion leads to a wideband ultrasonic emission. The generated ultrasonic waves are then detected by ultrasonic transducers to form images[3-5]. PA images combine high ultrasonic penetration and strong optical contrast in a single modality. For the scanning PAI system, hundreds or even thousands of scanning steps are required to form an image[6]. The scanning process usually takes several minutes and the long acquisition time will lead to movement artifacts. To realize the real-time imaging, an ultrasonic transducer array can be used to detect the PA signals[7-10]. For the PAI system based on an ultrasonic transducer array, a large number of parallel data acquisition (DAQ) channels are needed. In order to decrease the system cost and complexity, a real-time PA imaging system with high density integrated circuit is constructed in this paper. 2. Photoacoustic Imaging System The system diagram is illustrated in Figure 1, which is composed of a short pulsed laser, an ultrasonic transducer array, a data acquisition (DAQ) circuit and a computer. The laser transmits 7 ns pulses to irradiate the biological tissues at 15 Hz with wavelength 532nm. The transducer array (L10-5, 38 mm, 128 elements, 7.5-MHz central frequency, 75% -6dB bandwidth) is used to detect the PA signals generated from the tissue thermoelastic expansion. Then the PA signals are amplified and sampled by multiple parallel DAQ channels. 67

Subsequently, the sampled data set is transferred to the computer through an USB interface. The final PA image is reconstructed by the computer. Fig.1. Diagram of photoacoustic imaging system The data acquisition circuit includes 128 DAQ channels, which are composed of variable gain amplifiers (VGA), analog-to-digital converters (ADC) and field programmable gate arrays (FPGA). The specific chips include AD9272 and EP3C55. The chip AD9272 contains eight channels of a low noise preamplifier (LNA) with a VGA, an antialiasing filter (AAF), and a 12-bit, 10 MSPS to 80 MSPS ADC[11]. Its internal structure is illustrated in Figure 2. It is designed for low cost, low power, small size, and ease of use. Good noise performance relies on a proprietary ultra-low noise LNA at the beginning of the signal chain, which minimizes the noise contribution in the following VGA. Active impedance control optimizes noise performance for applications that benefit from input impedance matching. The anti-aliasing filter that the signal reaches prior to the ADC is used to reject dc signals and to band limit the signal for anti-aliasing. AD9272 uses a pipelined ADC architecture. The quantized output from each stage is combined into a 12-bit result in the digital correction logic. The pipelined architecture permits the first stage to operate on a new input sample and the remaining stages to operate on preceding samples. Sampling occurs on the rising edge of the clock. The ADC LVDS (Low Voltage Differential Signaling) outputs facilitate interfacing with LVDS receivers in custom ASICs (Application Specific Integrated Circuit) and FPGAs that have LVDS capability for superior switching performance in noisy environments. Here, two FPGAs are adopted to interface with ADCs and each FPGA records the data of 64 DAQ channels. Fig.2. Functional block diagram of AD9272 68

EP3C55 is a FPGA chip of Cyclone III series made by Altera Incorporation[12]. Cyclone III device family includes a customer-defined feature set that is optimized for portable applications and offers a wide range of density, memory, embedded multiplier, and I/O options. Cyclone III device family provides the ideal solution for your high-volume, low-power, and cost-sensitive applications. EP3C55 owns 55,856 logic elements and 260 distributed memory blocks. The logic array block (LAB) consists of 16 logic elements and a LAB-wide control block. An LE is the smallest unit of logic in the Cyclone III device family architecture. Each LE has four inputs, a four-input look-up table (LUT), a register, and output logic. The four-input LUT is a function generator that can implement any function with four variables. Each memory block provides nine Kbits of on-chip memory capable of operating at up to 315 MHz. The embedded memory structure consists of M9K memory blocks columns that you can configure as RAM, first-in first-out (FIFO) buffers, or ROM. The ADC sampling rate is set to 40 MHz. Four memory blocks (buffer length is 1k 12bit) are used to record a DAQ channel data with maximum depth 4cm. Thus, 64 DAQ channels occupy 256 memory blocks. CY7C68013 is an USB interface chip with USB 2.0 protocol[13]. CY7C68013 fabricates 4KB FIFO to provide a high speed data transfer with 96 MBytes/s burst rate. The PA signals are sampled and recorded in distributed memory blocks. Then the distributed data sets are collected and transferred to the FIFO of CY7C68013, which is shown in Figure 3. At the same time, CY7C68013 reads data from the FIFO and transfer them to the internal memory of the computer. The total data volume of a frame image is 2(FPGA) 256(block) 512(Byte/block)=256KB. The pulse repetition frequency of the laser is 15Hz. To achieve the maximum imaging frame rate, it is necessary to transfer 256KB 15=1.25MB data per second. It is easy to attain since CY7C68013 can transfer more than 20MB data per second usually. Fig.3. Diagram of data transfer Based on the above description, the system hardware is built and shown in Figure 4. Fig.4. Picture of the photoacoustic imaging system In order to realize the real time imaging, a fast speed reconstruction algorithm is required. The beamforming method is an alternative one. 3. Photoacoustic Image Reconstruction In PAI, when the pulsed laser irradiates the testing tissue, the light quickly spreads throughout the tissue due to scattering. Therefore, the PA response is simultaneously produced everywhere in the tissue. All transducer elements are used to receive the PA signals at once. Based on the received data, every scan line can be reconstructed quickly by the traditional delay-and-sum beamforming method [14,15], shown in Figure 5. At the center of the transducer, the focusing aperture is symmetrical, shown in Figure 6. At the 69

edge of the transducer, the focusing aperture is asymmetrical, shown in Figure 7. The asymmetry of the focusing aperture can lead to the sidelobe rise. Transducer Transducer Focus Image Fig.5. Image reconstruction based on beamforming Transducer Image Fig.6. Focusing schematic at the center of the transducer Focus Image Fig.7. Focusing schematic at the edge of the transducer In order to improve the contrast resolution of the reconstructed image, the apodization weighting technology should be applied. Different weightings are selected for different scan lines to balance the lateral resolution of the reconstructed image. The apodization weighting curves at the edge and center of the transducer are plotted in Figure 8(a), and the corresponding beam patterns are plotted in Figure 8(b). It is seen that the mainlobes are comparable, while the sidelobe level at the edge of the transducer is obviously higher than that at the center of the transducer. As a result, the contrast resolution at the edge of the reconstructed image is worse than that at the center. Weighting 1 0.5 at the center at the edge 0 20 40 60 80 100 120 (a) Apodization curves (number) db 0-50 at the center at the edge -100 0 5 10 15 20 25 30 35 (b) Beam patterns (mm) Fig.8. Apodization and beam pattern 70

The focusing delay parameters can be computed in advance and the focus data can be indexed quickly by the focusing delay parameters. Then, the focus points of multi-channels are weighted and accumulated. Since the delay-and-sum beamforming algorithm has no complex computation, the PA image can be reconstructed in real time by the algorithm. Hair 2 Hair 1 artifact Fig.9. The experiment result of a two hair phantom An experiment is made based on the constructed hardware platform shown in Figure 4. The phantom of two hairs is irradiated by the short-pulsed laser. The generated PA signals are detected by the array transducer. Then they are amplified, sampled and transferred to the computer. The PA image is reconstructed on the computer and shown in Figure 9. From the experiment result, it is seen that the two hairs can be detected and displayed clearly. Also, some artifacts are displayed in the image, which are caused by the multiple reflections of the PA signals from the transducer surface. The imaging rate is the same as the pulse repetition frequency of the laser. 4. Conclusion In order to decrease the system cost and complexity, a real-time photoacoustic imaging system with high density integrated circuit is built in this paper. The system consists of a short pulsed laser, an ultrasonic transducer array, a data acquisition circuit and a computer. The data acquisition circuit includes 128 high integrated channels which is composed of 128 LNAs, 128 VGAs, 128 ADCs and 2 FPGAs. As a result, the PA signals can be detected by 128 transducer elements simultaneously. Subsequently, the sampled data set is transferred to the computer through an USB interface. The final PA image is reconstructed by the computer. The traditional delay-and-sum beamforming method is used to reconstruct the PA image quickly. The imaging rate can attain to the pulse repetition frequency of the laser. Based on the built system platform, an experiment with a phantom of two hairs is made. The result shows that two hairs can be displayed clearly in the reconstructed image. However, some artifacts are still displayed in the image, which are caused by the multiple reflections of the PA signals from the transducer surface. 5. Acknowledgment This work is supported by the National Natural Science Foundation of China (No. 30800240), Shandong Provincial Promotive Research Fund for Excellent Young and Middle-aged Scientists (BS2010DX001) and Weihai City Science & Technology Development Project (2010-3-96). 6. References [1] L. V. Wang, Prospects of photoacoustic tomography, Medical Physics, vol.35, no.12, 2008, pp. 5759-5767. [2] Z. Xu, C. Li, L. V. Wang, Photoacoustic tomography of water in phantoms and tissue, Journal of Biomedical Optics, vol.13, no.3, 2010, pp.036019:1-6. [3] B. E. Treeby, B. T. Cox, k-wave: MATLAB toolbox for the simulation and reconstruction of photoacoustic wave fields, Journal of Biomedical Optics, vol.15, no.2, 2010, pp.021314:1-12. 71

[4] H. F. Zhang, K. Maslov, G. Stoica, et al, Functional photoacoustic microscopy for high-resolution and noninvasive in vivo imaging. Nature Biotechnology, vol.24, no.7, 2006, pp.848-851. [5] M. Xu, L. V. Wang, Photoacoustic imaging in biomedicine, Review of Scientific Instruments, vol.77, 2006, pp.041101:1-21. [6] Z. Guo, C. Li, L. Song, et al, Compressed sensing in photoacoustic tomography in vivo, Journal of Biomedical Optics, vol.15, no.2, 2010, pp.021311:1-6. [7] J. J. Niederhauser, M. Jaeger, R. Lemor, et al, Combined ultrasound and ptoacoustic system for real-time highcontrast vascular imaging in vivo, IEEE Trans. Med. Imag., vol.24, no.4, 2005, pp.436-440 [8] Y. Zeng, D. Xing, Y. Wang, et al, Photoacoustic and ultrasonic coimage with a linear transducer array, Opt. Lett., vol.29, no.15, 2004, pp.1760-1762 [9] S. Park, S. Mallidi, A. B. Karpiouk, et al, Photoacoustic imaging using array transducer, Proc. of SPIE, vol.6437, 2007, pp.643714:1-7 [10] R. G. M. Kolkman, P. J. Brands, W. Steenbergen, et al, Real-time in vivo photoacoustic and ultrasound imaging, Journal of Biomedical Optics, vol.13, no.5, 2008, pp. 050510:1-3. [11] AD9272-Octal LNA/VGA/AAF/ADC and Crosspoint Switch Datasheet, 2009, Analog Devices Incorporation. [12] Cyclone III Device Handbook, 2009, Altera Corporation. [13] CY7C68013 EZ-USB FX2 USB Microcontroller High-speed USB Peripheral Controller Datasheet, 2002, Cypress Semiconductor Corporation. [14] S. Park, S. R. Aglyamov, S. Y. Emelianov, Beamforming for photoacoustic imaging using linear array transducer, IEEE Ultrasonics Symposium, 2007, pp.856-859 [15] S. Park, A. B. Karpiouk, S. R. Aglyamov, et al, Adaptive beamforming for photoacoustic imaging using linear array transducer, IEEE Ultrasonics Symposium, 2008, pp.1088-1091 72