Southern African Large Telescope. RSS Throughput Test Plan

Similar documents
Prime Focus Imaging Spectrograph

Southern African Large Telescope. Prime Focus Imaging Spectrograph. Instrument Acceptance Testing Plan

Eric B. Burgh University of Wisconsin. 1. Scope

Southern African Large Telescope. Prime Focus Imaging Spectrograph. Grating and Filter Specification Document

Spectrograph Lens Fabrication RFQ 22 Jan, 2003

Southern African Large Telescope. Prime Focus Imaging Spectrograph. Polarimetric Optics Design Study

Southern African Large Telescope. Prime Focus Imaging Spectrograph. Optics Design

The equipment used share any common features regardless of the! being measured. Electronic detection was not always available.

MULTI-ELEMENT LENSES. Don t see exactly what you are looking for? CVI Laser Optics specializes in prototype to volume production manufacturing!

Descriptions for Each Test

Southern African Large Telescope. RSS Observer s Guide

Southern African Large Telescope. Prime Focus Imaging Spectrograph. Optics Design

Southern African Large Telescope. Prime Focus Imaging Spectrograph. Optics Design

StarBright XLT Optical Coatings

MicroSpot FOCUSING OBJECTIVES

Using Stock Optics. ECE 5616 Curtis

FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION

Southern African Large Telescope. RSS CCD Geometry

Corundum C Axis Device for Sample Preparation Timothy Thomas, M.E., M.S.E.E. GIA Laboratory June 4, 2009

Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region

The LINOS Singlets. Our quality criteria:

Post PDR Optical Design Study. Robert Barkhouser JHU/IDG January 6, 2014

Chapter 8. The Telescope. 8.1 Purpose. 8.2 Introduction A Brief History of the Early Telescope

Improving the Collection Efficiency of Raman Scattering

capabilities Infrared Contact us for a Stock or Custom Quote Today!

Spectroscopy in the UV and Visible: Instrumentation. Spectroscopy in the UV and Visible: Instrumentation

LEOK-3 Optics Experiment kit

Spectroscopy of Ruby Fluorescence Physics Advanced Physics Lab - Summer 2018 Don Heiman, Northeastern University, 1/12/2018

PHYS General Physics II Lab Diffraction Grating

OPTICS DIVISION B. School/#: Names:

Image Slicer for the Subaru Telescope High Dispersion Spectrograph

Imaging Systems Laboratory II. Laboratory 8: The Michelson Interferometer / Diffraction April 30 & May 02, 2002

AST Lab exercise: aberrations

Optical System Design

Will contain image distance after raytrace Will contain image height after raytrace

Goodman Laboratory Technical Report

- Optics Design - Lens - Mirror - Window - Filter - Prism

A Narrow-Band Tunable Diode Laser System with Grating Feedback

High-brightness and high-efficiency fiber-coupled module for fiber laser pump with advanced laser diode

AgilOptics mirrors increase coupling efficiency into a 4 µm diameter fiber by 750%.

Chapter 14. Tunable Dye Lasers. Presented by. Mokter Mahmud Chowdhury ID no.:

Application Note (A11)

PHY 431 Homework Set #5 Due Nov. 20 at the start of class

1/8 m GRATING MONOCHROMATOR

Southern African Large Telescope. RSS UW Commissioning Activities,

Basic Optics System OS-8515C

Aberrations of a lens

A novel tunable diode laser using volume holographic gratings

PHYS 1112L - Introductory Physics Laboratory II

Optical design of MOIRCS

Measuring the throughput in spectrographs

Spectroscopy Lab 2. Reading Your text books. Look under spectra, spectrometer, diffraction.

arxiv: v1 [astro-ph.im] 26 Mar 2012

Maya2000 Pro Spectrometer

SIFS... SOAR Integral Field Spectrograph

TriVista. Universal Raman Solution

Photolithography II ( Part 2 )

Supplementary Materials

Optical Design. Instrument concept Foreoptics and slit viewer Spectrograph Alignment plan 3/29/13

Polarization Experiments Using Jones Calculus

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the

Cascaded holographic spectrographs for astronomical applications

Better Imaging with a Schmidt-Czerny-Turner Spectrograph

Part 5-1: Lithography

LlIGHT REVIEW PART 2 DOWNLOAD, PRINT and submit for 100 points

Exp. No. 13 Measuring the runtime of light in the fiber

why TECHSPEC? From Design to Prototype to Volume Production

Oriel Flood Exposure Sources

Exercise 8: Interference and diffraction

Assembly and Experimental Characterization of Fiber Collimators for Low Loss Coupling

CHAPTER 7. Components of Optical Instruments

EUV Plasma Source with IR Power Recycling

White Paper: Modifying Laser Beams No Way Around It, So Here s How

1. Do any of the design changes adversely affect the ability of KOSMOS to meet the scientific capabilities called for in the ReSTAR report?

Astro 500 A500/L-8! 1!

Fresnel Lens Characterization for Potential Use in an Unpiloted Atmospheric Vehicle DIAL Receiver System

General Physics Experiment 5 Optical Instruments: Simple Magnifier, Microscope, and Newtonian Telescope

Phys 2310 Mon. Oct. 16, 2017 Today s Topics. Finish Chapter 34: Geometric Optics Homework this Week

ABSTRACT 1. INTRODUCTION

High Volume Stock optics

ECEN 4606, UNDERGRADUATE OPTICS LAB

SOAR Integral Field Spectrograph (SIFS): Call for Science Verification Proposals

Observational Astronomy

Physics 308 Laboratory Experiment F: Grating Spectrometer

Pre-Lab 10. Which plan or plans would work? Explain. Which plan is most efficient in regard to light power with the correct polarization? Explain.

ADVANCED OPTICS LAB -ECEN Basic Skills Lab

Physics 248 Spring 2009 Lab 1: Interference and Diffraction

Design Description Document

Snell s Law, Lenses, and Optical Instruments

Introduction to the operating principles of the HyperFine spectrometer

Application Note (A16)

Make Your Own Digital Spectrometer With Diffraction Grating

Option G 4:Diffraction

Performance Comparison of Spectrometers Featuring On-Axis and Off-Axis Grating Rotation

Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy

Chapter 7. Optical Measurement and Interferometry

High-Power, Passively Q-switched Microlaser - Power Amplifier System

P202/219 Laboratory IUPUI Physics Department THIN LENSES

Optical Design of the SuMIRe PFS Spectrograph

KOSMOS. Optical Design

Transcription:

Southern African Large Telescope RSS Throughput Test Plan Kenneth Nordsieck University of Wisconsin Document Number: SALT-3160AP0005 Revision 1.0 27 June, 2006 Change History Rev Date Description 1.0 27 June, 2006 Original Table of Contents 1 Scope...1 2 Throughput Shortfall Evidence...1 3 Theories...2 4 In-Situ Testing...3 5 Possible Outcomes and Fixes...4

RSS Throughput Test Plan Rev 1.0 June 27, 2006 1 1 Scope This document describes a plan for testing and fixing of the RSS throughput shortfall. 2 Throughput Shortfall Evidence Data collected in Nov 2005 June 2006 during commissioning of the Robert Stobie Prime Focus Imaging Spectrograph show evidence of a throughput shortfall which increases into the ultraviolet. Figure 1 shows some of this data. The bottom panel shows the apparent RSS Optics Detector efficiency, together with the lab QE of the three detector CCD s and the FPRD typical and minimum QE for the CCD s. The top chart shows the throughput shortfall : it divides the Optics Detector efficiency by the CCD #4 QE (the middle chip) and by the expected optics throughput, based on vendor coating efficiency measurements (the blanks are assumed to have 100% throughput). The data were obtained as follows 20031030 47 Tuc full field, full mirror imaging, 629 nm filter 20051122 EG 21 burst (4 segments) imaging, three filters 20051204 EG 21 full mirror slitless spectroscopy 3000 and 900 l/mm VPH gratings 200603xx 8 stars, 7 filters, imaging burst Figure 1. RSS throughput shortfall 20060612 HR5501, 340 and 380 nm interference filters, imaging burst The reduction of this data has assumed a SALT mirror efficiency of 60% (dirty mirrors) except for the June data, which used a clean segment (77%). For spectroscopy, we have removed lab VPH efficiency curves. The result is that the throughput is below specification, ranging from 95% expected at 900 nm to 50% at 400 nm, then decreasing to < 10% in the UV. The fact that the imaging agrees with spectroscopy verifies that the VPH grating throughput is as expected.

RSS Throughput Test Plan Rev 1.0 June 27, 2006 2 3 Theories Explanations for the throughput shortfall break down into the following categories Serious absorption ( color centers ) in the optical blanks: Fused Silica, NaCl, or CaF 2 AR Coating deficiency Fold Flat coating deficiency Detector QE degradation We have focused on the first alternative so Figure 2. NaCl spare blank transmission far, since vendor test data exist for the other three. So far, serious absorption in the NaCl crystals has been eliminated by having Janos Inc (the RSS lens manufacturers) polish and measure the transmission of one of the spare NaCl blanks. In Figure 2, the thick line shows the total transmission of the blank, and the thin line removes two Fresnel surface reflections to give the internal transmission. There may be a slight absorption between 300 and 400 nm, but this would be much too small to explain the observations of Figure 1. The total thickness of NaCl in RSS is about 50 mm. Another possibility which has come to our attention is a process by which the coating of CaF 2 lenses may cause crystal damage (through electrostatic acceleration of ions in the coating plasma into the material). Hans Dekker of ESO reported that this problem occurred with UVES on VLT, and led to very poor UV throughput until it was corrected, by Winlight System of France, through UV flooding of the coated lenses. We have verified this account with Philippe Godefroy of Winlight, and have since found a patent application by two Japanese researchers which tells a similar story. However, we have not found anything similar in the literature. The expected absorption should be similar to color centers caused by irradiation. Figure 3 shows a CaF 2 color center absorption spectrum (A Smakula Figure 3. CaF 2 color center absorption spectrum 1950: Phys Rev 77, 408). One can see a sharp absorption band starting at about 420 nm and peaking at 400 and 340 nm. A secondary broad peak at 650 nm could explain the overall low RSS throughput in the yellow and red, and the absorption does disappear above about 800 nm, as is seen in RSS. The curve does not exactly match the RSS throughput curve, where the sharp absorption begins below 400 nm. However it is possible that the precise shape of the absorption curve is a function of the cause of the damage. Thus this remains a promising lead in the

RSS Throughput Test Plan Rev 1.0 June 27, 2006 3 investigation. There are 6 RSS coated CaF 2 elements, using three different coatings (see Figure 4): 1) Element 2 in Field Lens: MgF2 coating 2) Element 1 in the Collimator triplet: MgF2/ solgel 3) Element 3 in the Collimator triplet: MgF2 4) Element 2 in the Collimator doublet: multilayer 5) Element 4 in the Camera quartet: MgF2/ solgel 6) Camera singlet: MgF2/ solgel both sides Figure 4. RSS Optics It seems unlikely that this process would be a problem with the fused silica elements, since fused silica color centers are found exclusively below 250 nm. It is not a problem with the NaCl, since both NaCl elements are the central elements of triplets, and hence uncoated. We must still keep an open mind on the remaining three theories: coatings, mirror, and detector. The following test process does not assume the CaF 2 color center theory. It will test the detector and fold flat theory and if necessary isolate optical elements. If the CaF 2 theory holds, it would be confirmed by isolating the problem to multiplets containing CaF 2 with common coatings. If it is coatings, we would expect elements with common coatings, but including both fused silica and CaF 2. 4 In-Situ Testing There are many fault scenarios possible with the above theories. Some of these scenarios would require a major, costly dismantling of the optics to gain access to individual elements, and some would not. We suggest the following in-situ testing apparatus which would narrow down the possibilities sufficiently to determine this. It is relatively easy to place a mirror in three places in the instrument, the filter, the focal plane (a slitmask), and at the 1/4 waveplate slide. Measuring the intensity of a return beam from a sufficiently collimated UV light source placed in the accessible collimated beam (Fig 4) would obtain a ~10% accurate measurement of the throughput (in double pass) of the camera, the collimator, and the collimator excluding the field lens. Because of the length of the beam, it seems likely that the best light source is a laser, a relatively expensive item. The shortest wavelength CW laser currently available is a 375 nm diode laser,

RSS Throughput Test Plan Rev 1.0 June 27, 2006 4 which is (barely) well enough into the UV drop-off of the throughput curve to be sensitive to the throughput problem. The power available is ~3mW. A possible alternative is a AlGaN UV LED, which can be purchased in 320 and 340 nm versions. This has the advantage that the wavelength is at the bottom of the UV throughout curve, and so would be very sensitive. However, LED s are poorly collimated, and a simple A-S argument suggests that one would get less than 1% of the available power into the return beam. Starting at an LED power of 0.5 mw, this is 3 orders of magnitude down from the laser, and (especially if the optics has poor transmission), it would be very difficult to find the return beam. So we suggest the following apparatus: 375 nm laser, CrystalLaser, BCL-005-375, $6450 (lowest of 3 quotes, awaiting a 4th) UV Photodiode, OSI Optoelectronics, 1 cm 2 active area; available on loan from Rutgers. 380/ 25 nm interference filter. Andover, $242 (for stray light rejection) Electrometer for photodiode; borrow lab instrument. UV converter plate, UVP, 21 26 cm, $245 (for visual alignment of return beam) Mirror blank for filter: glass < 8mm thick, < 130x90mm + spare filter holder Mirror blank for slitmask: blank longslit or coated microscope slide Mirror blank for 1/4 wave slide. Use existing spare blank Figure 5, In-situ measurement apparatus The laser and photodiode would be mounted on a fixture that fits into the RSS grating holder (Fig 5). The photodiode would be adjustable up and down, and the RSS grating rotator would be used to adjust the return beam placement in the horizontal plane. A UV to white light convertor plate would be used to find the return beam and help with the alignment. 5 Possible Outcomes and Fixes Figure 6 illustrates the test/ decision tree. The in-situ throughput measurement determines whether the problem is in the camera and/or the collimator optics, or in the detector (we assume there are not two independent problems, with the optics and the detector). If there is a camera problem, it will be removed from the instrument (it is relatively accessible), taken to a clean room, and disassembled and tested there. If there is a collimator problem, it may be in any or all of four assemblies, the field lens, main group, fold flat, and doublet. If it is only in the field lens,

RSS Throughput Test Plan Rev 1.0 June 27, 2006 5 a third in-situ measurement with a return mirror at the 1/4 waveplate slide can determine that. If there are problems in the rest of the collimator, the doublet, and, if necessary, the fold mirror can be fairly easily removed and tested. If there is a problem in the field lens or main group, the instrument will have to be partially or completely removed from the telescope to gain access. Reassembly and alignment of the instrument with either of these elements removed will require the removal of the instrument for realignment. Because of the difficulty of the latter, it may be wise to eliminate or fix problems in the doublet or fold mirror so that a Figure 6. Test/ decision tree repeat collimator in-situ test may be performed to determine whether any remaining field lens or main group problem is serious enough to merit the risk. Bench testing equipment will consist of the same photodiode/ electrometer detector used for the in-situ testing, with a more flexible UV light source, either UV LED s, or a fiber light source loaned from Rutgers. UV LED s cost about $200. If the camera or the collimator main group need to be disassembled, this will be done by Alan Schier of Pilot Group, the original assembler of these optics. We would hope to use the SAAO clean room. Reassembly and alignment would also be done by Alan Schier. If testing confirms the CaF 2 color center theory, repair will be done in the same clean room, using a mercury curing lamp as a UV flood, and the UV transmission device as a monitor. The flooding process takes less than an hour. We would hope to do the flooding process without disassembling multiplets, though this will be the subject of analysis. The time from disassembly to reassembly in this scenario should be on the order of a month. If the problem resides in coatings, repair will require disassembling multiplets, sending them to the manufacturer to have the coatings polished off, then to the coater, then reassembled. The timescale for this repair would be a minimum of 3-6 months.